Tuttnauer 5075 Technician Manual

Tuttnauer 5075 Technician Manual (134 pages)

Laboratory Vertical Steam Sterilizers ELV, ELVC, ELVPRC

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 2.78 MB

Table of Contents
  • Table of Contents

    3

  • 1 Introduction

    6

  • 2 Symbol Description

    6

  • 3 Insta Llation Instructions

    7

    • Installation Site

      8

    • Connection of the Pressure Regulator

      8

    • Rear View Model 2540

      9

    • Rear View Models 3150, 3170

      10

    • Rear View Models 3850, 3870

      11

    • Rear View Models 5050, 5075

      12

  • 4 Tests

    13

    • Installation Tests

      13

    • Periodical Tests

      13

  • 5 Water Quality

    14

    • Water for Generating Steam

      14

    • Water for the Vacuum Pump and Drain Cooling

      14

    • Reverse Osmosis

      14

  • 6 Description of the Control System

    15

    • Digital Board DIG — T2

      15

    • Control System Block-Diagram

      15

    • DIG- T2 Board

      16

      • Analog Board ANL-T2

        17

    • ANL — T2 Board

      17

      • AC — Board — AC-T1

        18

    • AC-T1 Board

      18

  • 7 Calibration of Temperature and Pressure

    19

    • Method of Calibration

      19

    • Temperature Calibration Procedure

      20

    • Pressure Calibration Procedure

      22

  • 8 Test Points

    23

  • 9 Software Programming Procedures

    24

    • General

      24

    • Changing Cycle Parameters

      24

    • Resetting the Autoclave

      32

    • Test Mode

      32

  • 10 Replacing Elements

    33

    • Safety Tests after Repair

      33

    • Replacing the Circuit Breaker

      34

    • Replacing the Safety Valve

      35

    • Replacing the Door Gasket

      36

    • Replacing the Electronics Board (Control Panel)

      37

    • Replacing the Electronics Box

      38

    • Cut-Out Thermostat

      41

    • Replacing Heating Elements

      42

    • Safety Door Locking System for Models 2540, 3150, 3170, 3850, 3870

      43

    • Safety Door Locking System for Models 5050, 5075

      45

    • Disassembling and Assembling the Door Cover, Models 2540, 3150, 3170

      46

    • Disassembling and Assembling the Door Cover, Models 5050, 5075

      48

    • Replacing the Locking Microswitch (Models 5050, 5075)

      51

    • Replacing the Safety-Opening Valve (Models 5050, 5075)

      52

    • Replacing the «DOOR CLOSED» Microswitch (Models 5050, 5075)

      53

    • Replacing the Solenoid Lock (Models 5050, 5075)

      54

    • 10.17 Cleaning and Replacing the Water Level Electrodes

      55

    • 10.18 Replacing the Printer

      56

    • Replacing the Door Switch (Models 2540, 3150, 3170, 3850, 3870)

      58

    • Replacing the Locking Device (Models 2540, 3150, 3170, 3850, 3870)

      59

    • 10.21 Replacing the Plunger or Coil of the BACCARA Solenoid Valve

      60

  • 11 In-Out Test

    61

  • 12 Troubleshooting

    62

    • Displayed Messages and Reference to Trouble Shooting

      62

    • Pre-Process Malfunction

      63

    • In Process Malfunction

      69

    • Mechanical Malfunction

      88

  • 13 List of Spare Parts

    89

    • Vessel Assembly Models 2540, 3150, 3170, 3850, 3870

      94

    • Vessel Assembly Models 5050, 5075

      95

    • Door Tightening Bolt — Assembly

      96

    • Door Locking System

      97

    • Water Outlet Strainer

      98

  • 14 Pressure Vs Temperature for Saturated Steam

    99

  • 15 XPCS Manual

    103

  • 16 Valves Numbering

    116

    • Piping Drawing: 2540 ELV &, ELVPRC

      118

    • Piping Drawing: 2540 ELVC

      119

    • Piping Drawing: 3150, 3170, 3850, 3870 ELV & ELVPRC

      120

    • Piping Drawing: 3150, 3170, 3850, 3870 ELVC

      121

    • Piping Drawing: 5050, 5075 ELV & ELVPRC

      122

    • Piping Drawing: 5050, 5075 ELVC

      123

    • Electrical Wiring Diagram for Autoclave Model 2540 ELV

      124

    • Electrical Wiring Diagram for Autoclave Models 3150/3170 ELV

      125

    • Electrical Wiring Diagram for Autoclave Models 3850/3870 ELV 3X400V

      126

    • Electrical Wiring Diagram for Autoclave Models 3850/3870 ELV 3X230V

      127

    • Electrical Wiring Diagram for Autoclave Models 3850/3870 ELV 3X208V

      128

    • Electrical Wiring Diagram for Autoclave Models 5050/5070 ELV 3X400V

      129

    • Electrical Wiring Diagram for Autoclave Models 5050/5070 ELV 3X230V

      130

    • Electrical Wiring Diagram for Autoclave Models 5050/5070 ELV 3X208V

      131

    • Switch Box Wiring Diagram

      132

    • Open Switch Box

      133

    • Closed Switch Box

      134

Advertisement

Tuttnauer 5075 Operation And Maintenance Manual

Tuttnauer 5075 Operation And Maintenance Manual (95 pages)

Electronic Laboratory Autoclaves

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 8.41 MB

Table of Contents
  • Table of Contents

    3

  • General

    6

  • Incoming Inspection

    6

  • Warranty

    6

  • Warranty Statement

    6

  • Accessories

    7

  • Introduction

    7

  • Safety Instructions

    10

  • Technical Data

    12

  • Operating Conditions

    12

  • Directives and Standards

    12

  • Environmental Emission Information

    13

  • Electrical Data

    13

  • Water Quality

    14

  • Specifications

    15

  • Construction

    16

  • Utilities

    16

  • Stickers Description

    18

  • Front View

    19

  • Front View (WR Configuration, Inside Reservoir, Only Specific Items Are Shown)

    20

  • Rear View

    21

  • Rear View (WR Configuration, Inside Reservoir, Only Specific Items Are Shown)

    22

  • Overall Dimensions — Model 5050ELV (MM)

    23

  • Overall Dimensions — Model 5075ELV

    24

  • Overall Dimensions — Model 5075ELV WR Configuration

    24

  • Safety Features Include Built-In Safety Features such as

    25

  • Description of Operation

    25

  • Control Panel

    27

  • Control Panel Drawing

    27

  • Description and Functions of the Front Panel Keyboard

    28

  • Displayed Error Messages / Symbols

    30

  • Displayed Operational Messages / Symbols

    30

  • Sterilization Programs

    32

  • Program 1: Glass

    35

  • Program 2: Plastic

    36

  • Program 3: Liquid a

    37

  • Program 4: Liquid B — Waste

    38

  • Program 5: Liquid a — Cooling (C — Cooling Models Only)

    39

  • Program 6: Liquid B — Waste Cooling (C — Cooling Models Only)

    40

  • Program 7: Bio Hazard 1 (BH, Bio-Hazard Models Only)

    41

  • Program 8: Bio Hazard 2 (BH, Bio-Hazard Models Only)

    42

  • Program 9: Bio Hazard Liquids (BH, Bio-Hazard Models Only)

    43

  • Program 10: Vacuum Test (PV, Pre-Vacuum Models Only)

    45

  • Program 11: Warm-Up

    46

  • Program 12: Isothermal

    47

  • Program 13: Air Mixture

    48

  • Program 14: Glass Test

    49

  • Program 15: Durham

    50

  • Checking and Changing Parameters and Other Data

    52

  • Browsing through the Menus

    52

  • Changing a Parameter

    53

  • Quick Options Screen

    55

  • Logging in and Entering the Main Menu

    59

  • Changing Parameters

    60

  • System Parameters

    61

  • Maintenance

    62

  • Cycle Parameters

    63

  • Screens

    65

  • Screens Following a Complete Successfully Cycle — «Cycle Ended

    65

  • Screens Following Aborted Cycles after Complete Sterilization Stage

    65

  • Screens Following a Fail Cycle

    66

  • Printer

    68

  • Printer Output

    68

  • Printer Handling

    70

  • Preparation before Sterilization

    73

  • Instruments

    73

  • Liquids

    74

  • Loading

    74

  • Operation

    75

  • Draining and Filling the Water Reservoir (WR Only)

    75

  • Turning on the Autoclave

    76

  • Opening the Door — MD Models (Mechanical Door)

    77

  • Opening the Door — D Models (Automatic Door)

    77

  • Loading

    78

  • Selecting Program

    79

  • Closing and Locking the Door — MD Models (Mechanical Door)

    79

  • Closing and Locking the Door — D Models (Automatic Door)

    79

  • Starting a Cycle

    81

  • Unloading

    81

  • Stopping the Process and Cancelling the ERROR Message

    81

  • Start Cycle by Clock Mode

    82

  • Moving the Autoclave

    82

  • Loading and Unloading the Device

    82

  • Service and Maintenance

    84

  • Preventive Maintenance

    84

  • Replacing the Safety Valve — PED Type

    85

  • Replacing the Safety Valve — ASME Approved Type

    87

  • Checking the Safety Valve

    87

  • Troubleshooting

    89

  • Baskets and Containers

    94

  • Spare Parts List

    95

  • Accessories

    95

Tuttnauer 5075 Operation & Maintenance Manual

Tuttnauer 5075 Operation & Maintenance Manual (87 pages)

Laboratory Vertical Steam Sterilizers

Brand: Tuttnauer
|
Category: Steriliser
|
Size: 5.71 MB

Table of Contents
  • Table of Contents

    3

  • General

    7

    • Introduction

      7

    • Incoming Inspection

      8

    • Warranty

      8

    • Warranty Statement

      9

  • Safety Instructions

    10

  • Technical Data

    12

    • Stand — by Heating Mode

      12

    • Storage Conditions

      12

    • Operating Conditions

      12

    • Utilities

      12

    • Construction

      13

    • Environment Emission Information

      13

    • Directives and Standards

      13

    • Electrical Data

      14

    • Overall Dimensions Models 5050, 5075 el

      15

    • Symbol Description

      16

    • Water Quality

      17

    • Safety Features

      18

  • Control Panel

    21

    • Control Panel Drawing

      21

    • Description and Functions of the Front Panel Keyboard

      22

    • Displayed Error Messages / Symbols

      24

    • Displayed Operational Messages / Symbols

      24

    • Description of Operation

      25

    • Leakage Test

      26

    • Exhaust

      26

    • Drying (PV — Pre-Vacuum Nodel Only)

      26

    • Vacuum Test Program (PV — Pre-Vacuum Model Only)

      26

  • Sterilization Programs

    27

    • Program 1: Glass

      29

    • Program 2: Plastic

      30

    • Program 3: Liquid a

      31

    • Program 4: Liquid B — Waste

      32

    • Program 5: Liquid a — Cooling (ELC — Cooling Models Only)

      33

    • Program 6: Liquid B — Waste Cooling (ELC — Cooling Models Only)

      34

    • Program 7: Waste

      35

    • Program 8: Hollow Load

      36

    • Program 9: Vacuum Test (PV, Pre-Vacuum Model Only)

      37

    • Program 10: Bio Hazard Liquid Cooling (BH, Bio-Hazard Models Only)

      38

    • Program 11: Bio Hazard Liquid (BH, Bio-Hazard Models Only)

      39

    • Program 12: Bio Hazard (BH, Bio-Hazard Models Only)

      40

    • Program 13: Warm-Up

      41

    • Program 14: Isothermal

      42

  • Checking and Changing Parameters and Other Data

    43

    • Directories and Subdirectories

      43

    • Entering the Main Menu

      43

    • System Parameters

      45

    • History

      47

    • Maintenance

      49

    • Advanced Options

      53

    • Version Information

      56

  • Screens

    59

    • Screens Following a Completely Successful Cycle — «Cycle Ended

      59

    • Screens Following Aborted Cycles after Complete Sterilization Stage

      60

    • Screens Following a Fail Cycle

      60

  • Printer

    62

    • Printer Output

      62

    • Printer Handling

      64

    • Treating the Thermal Papers

      66

  • Preparation before Sterilization

    67

    • Instruments

      67

    • Liquids

      68

    • Tubing

      69

    • Loading

      69

  • 10 Operating Instructions

    70

    • Turning on the Autoclave

      70

    • Opening the Door

      70

    • Safety

      71

    • Loading

      71

    • Selecting the Program

      71

    • Closing the Door

      71

    • Starting Cycle

      71

    • Unloading

      73

    • Stopping the Process and Cancelling the ERROR Message

      73

    • 10.10 Start Cycle by Clock Mode

      74

  • 11 Service and Maintenance Instructions

    75

    • Preventive and Scheduled Maintenance

      75

    • Checking the Safety Valve

      76

    • Moving the Autoclave

      81

  • 12 Troubleshooting

    82

  • 13 Baskets and Containers

    86

  • Spare Parts List — 5075

    87

  • 15 Accesories List

    87

Advertisement

Tuttnauer 5075 Operation & Maintenance Manual

Tuttnauer 5075 Operation & Maintenance Manual (85 pages)

Laboratory Vertical Steam Sterilizers

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 1.85 MB

Table of Contents
  • Table of Contents

    3

  • General

    6

  • Incoming Inspection

    6

  • Warranty

    6

  • Warranty Statement

    6

  • Accessories

    7

  • Safety Instructions

    8

  • Technical Data

    10

  • Introduction

    10

  • Operating Conditions

    11

  • Directives and Standards

    11

  • Environment Emission Information

    12

  • Electrical Data

    12

  • Specifications

    12

  • Construction

    12

  • Loading Capacities

    13

  • Utilities

    13

  • Symbol Description

    14

  • Front View

    15

  • Rear View

    16

  • OVERALL DIMENSIONS DRAWING (MM)

    17

  • Water Quality

    18

  • Safety Features

    19

  • Description of Operation

    20

  • Control Panel

    21

  • Control Panel Drawing

    21

  • Description and Functions of the Front Panel Keyboard

    22

  • Displayed Error Messages / Symbols

    24

  • Displayed Operational Messages / Symbols

    25

  • Sterilization Programs

    26

  • Program 1: Unwrapped Instruments

    27

  • Program 2: Wrapped Instruments

    28

  • Program 3: Unwrapped Pouches

    29

  • Program 4: Wrapped Pouches

    30

  • Program 5: Unwrapped Delicate Instruments

    31

  • Program 6: Wrapped Delicate Instruments

    32

  • Program 7: Custom 1

    33

  • Program 8: Custom 2

    34

  • Program 9: Liquid a

    35

  • Program 10: Liquid B

    36

  • Checking and Changing Parameters and Other Data

    37

  • Directories and Subdirectories

    37

  • Table of Parameters

    38

  • Entering the Main Menu

    39

  • Cycle Parameters (Custom 1)

    41

  • System Parameters

    47

  • History

    50

  • Maintenance

    52

  • Advanced Options

    57

  • Version Information

    60

  • Screens

    63

  • Screens Following a Complete Successfully Cycle — «Cycle Ended

    63

  • Screens Following Aborted Cycles after Complete Sterilization Stage

    64

  • Screens Following a Fail Cycle

    65

  • Printer

    66

  • Printer Output

    66

  • Printer Handling

    68

  • Preparation before Sterilization

    71

  • Instruments

    71

  • Liquids

    72

  • Loading

    72

  • Operation

    73

  • Turning on the Autoclave

    73

  • Opening the Door

    73

  • Loading

    74

  • Selecting Program

    74

  • Closing the Door

    74

  • Starting Cycle

    74

  • Unloading

    74

  • Stopping the Process and Cancelling the ERROR Message

    75

  • Start Cycle by Clock Mode

    75

  • 10.10 Moving the Autoclave

    75

  • 10.11 Loading and Unloading the Device

    75

  • Door Safety System

    77

  • Pressure Switch and Locking Solenoid

    77

  • Service and Maintenance

    78

  • Preventive Maintenance

    78

  • Checking the Safety Valve

    79

  • Troubleshooting

    81

  • Baskets and Containers

    84

  • Spare Parts List

    85

  • Accessories

    85

Tuttnauer 5075 Operation & Maintenance Manual

Tuttnauer 5075 Operation & Maintenance Manual (68 pages)

Electronic Table — Top Autoclaves (With cooling option)

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 1.37 MB

Table of Contents
  • Table of Contents

    3

  • 1 General

    6

    • Incoming Inspection

      6

    • Warranty Statement

      6

  • 2 Safety Instructions

    8

  • 3 General Information

    10

    • Stand — by Heating Mode

      10

    • Operating Conditions

      11

    • Utilities

      11

    • Load on Counter

      11

    • Construction

      11

    • Environment Emission Information

      12

    • Directives and Standards

      12

    • Electrical Data

      12

    • Overall Dimensions Model 2540 el

      13

    • Overall Dimensions Models 3150 el

      14

    • Overall Dimensions Models 3850, 3870 el

      15

    • Overall Dimensions Models 5050, 5075 el

      16

    • Loading Capacities

      17

    • Symbol Description

      17

    • Technical Specifications el

      18

    • Water Quality

      19

  • Front View

    21

  • Rear View Model 2540

    22

  • Rear View Models 3150, 3850, 3870

    23

  • Rear View Models 5050, 5075

    24

  • 4 Sterilization Programs

    25

    • Program 1 — Unwrapped Instruments (1 — Instruments)

      25

    • Program 2 (2- Instruments)

      26

    • Program 3 (3 -Waste)

      27

    • Program 4 (4 — Liquids)

      28

    • Program 5 (5- Liquids)

      29

    • Program 6 (6 — Liq+Cool)

      30

  • Front Panel Keyboard

    31

    • Description and Functions of the Front Panel Keyboard

      32

    • Description of the Display

      35

    • Description of Displayed Messages and Safety Measures

      37

  • 5 KEYBOARD (Keys and Display)

    31

  • 6 Printer

    39

    • Printer Operation

      39

    • DPU-20 Printer Handling

      41

    • DPU 30 Printer Handling

      42

  • 7 Preparation before Sterilization

    43

    • Filling the Water Reservoir

      46

  • 8 Operating Instructions

    47

    • Operation

      47

    • Loading and Unloading the Device

      48

  • 9 Maintenance Instructions

    50

    • Preventive and Scheduled Maintenance

      50

    • Draining the Reservoir

      51

    • Cleaning Water Strainer

      52

    • Checking the Safety Valve

      53

    • Replacing the Door Gasket

      55

    • Door Safety System for Models 2540, 3150, 3850, 3870

      56

    • Door Safety System for Models 5050, 5075

      57

    • Cleaning Table Top Autoclaves with Chamber Brite

      60

  • 10 Troubleshooting

    62

  • Pouch Rack

    64

  • Tray — Standard

    65

  • Tray, Special ELC Version

    66

  • Tray Holder, Special ELC Version

    66

  • Baskets and Containers

    67

  • 11 Spare Parts List

    68

  • 12 Acssesories List

    68

Tuttnauer 5075 Operation & Maintenance Manual

Tuttnauer 5075 Operation & Maintenance Manual (59 pages)

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 1.56 MB

Table of Contents
  • Table of Contents

    3

  • 1 General

    6

    • Incoming Inspection

      6

    • Warranty

      6

    • Warranty Statement

      6

  • 2 Safety Instructions

    8

  • 3 Technical Data

    10

    • Introduction

      10

    • Stand — by Heating Mode

      11

    • Operating Conditions

      11

    • Utilities

      11

    • Load on Counter

      12

    • Construction

      12

    • Environment Emission Information

      12

    • Directives and Standards

      12

    • Electrical Data

      13

    • Overall Dimensions Models 5050, 5075 el

      14

    • Symbol Description

      15

    • Technical Specifications EL-D

      16

    • Water Quality

      17

  • 4 Control Panel

    21

    • Description and Functions of the Front Panel Keyboard

      22

    • Displayed Error Messages / Symbols

      24

    • Displayed Operational Messages / Symbols

      24

  • 5 Description of Programs

    26

    • Operations Sequence

      26

    • Vacuum Test Program (ELPV Model Only)

      26

  • 6 Sterilization Programs

    27

    • Program 1: Glass

      29

    • Program 2: Plastic

      30

    • Program 3: Liquid a

      31

    • Program 4: Liquid B — Waste

      32

    • Program 5: Vacuum Test (ELPV Model Only)

      33

  • 7 Screens

    34

    • Screens Following a Completely Successful Cycle — «Cycle Ended

      34

    • Screens Following Aborted Cycles after Complete Sterilization Stage

      35

    • Screens Following a Fail Cycle

      35

  • 8 Printer

    37

    • Printer Output

      37

    • Printer Handling

      39

    • Treating the Thermal Papers

      41

  • 9 Preparation before Sterilization

    42

    • Instruments

      42

    • Tubing

      43

    • Sterilization of Liquids

      44

    • Filling the Water Reservoir

      44

  • 10 Operating Instructions

    45

    • Turning on the Autoclave

      45

    • Opening the Door

      45

    • Safety

      45

    • Loading

      46

    • Selecting the Program

      46

    • Closing the Door

      46

    • Starting Cycle

      46

    • Unloading

      47

    • Start Cycle by Clock Mode

      47

  • 11 Service and Maintenance Instructions

    48

    • Preventive and Scheduled Maintenance

      48

    • Draining the Reservoir

      49

    • Cleaning Water Strainer

      50

    • Checking the Safety Valve

      51

    • Moving the Autoclave

      54

  • 12 Troubleshooting

    55

  • 13 Spare Parts List

    59

  • 14 Acssesories List

    59

Tuttnauer 5075 Installation, Operation & Maintenance Manual

Tuttnauer 5075 Installation, Operation & Maintenance Manual (59 pages)

Laboratory Vertical Steam Sterilizers ELV, ELVC, ELVPRC

Brand: Tuttnauer
|
Category: Laboratory Equipment
|
Size: 0.99 MB

Table of Contents
  • Table of Contents

    2

  • 1 General

    5

    • Incoming Inspection

      5

    • Warranty

      5

    • Warranty Statement

      5

    • Ordering Information

      6

  • 2 Safety Instructions

    7

  • 3 Technical Data

    9

    • Introduction

      9

    • Operating Conditions

      10

    • Directives and Standards

      10

    • Environment Emission Information

      10

    • Electrical Data

      11

    • Specifications

      11

    • Loading Capacities

      12

    • Utility

      12

    • Symbol Description

      13

  • Overall Dimensions 2540 ELV

    14

  • Overall Dimensions Drawing for the 3150 / 3170 ELV

    15

  • Overall Dimensions Drawing for the 3850 / 3870 ELV

    16

  • Overall Dimensions Drawing for the 5050/ 5075 ELV

    17

    • Water Quality

      18

    • Safety Features

      19

    • Description of Operation

      20

  • 4 Sterilization Programs

    21

    • Program 1 — Unwrapped Instruments (1 — Instruments)

      21

    • Program 2 (2- Instruments)

      22

    • Program 3 (3 -Waste)

      23

    • Program 4 — (4 — Liquid)

      24

    • Program 5 (5- Liquid)

      25

    • Program 6 (6-Liquid + Cool) — Option on ELVC Only

      26

  • Front Panel Keyboard

    27

    • Description and Functions of the Front Panel Keyboard

      28

    • Description of the Operational Messages

      31

    • Displayed Error Messages and Safety Measures

      33

  • 5 KEYBOARD (Keys and Display)

    27

  • 6 Printer

    35

    • Printer Operation

      35

    • DPU-20 Printer Handling

      37

    • DPU 30 Printer Handling

      38

  • 7 Preparation before Sterilization

    39

    • Instruments

      39

    • Tubing

      40

    • Liquids

      40

    • Loading

      40

  • 8 Operation

    41

    • Operating Instructions

      41

    • Moving the Autoclave

      42

    • Loading and Unloading the Device

      42

  • 9 Door Safety System

    44

    • Solenoid Locking Device

      44

    • Door Safety System for Models 5050, 5075

      44

    • Piston Lifting Device for 3850/3870/5050/5075 ELV

      48

  • 10 Service and Maintenance

    49

    • Preventive Maintenance

      49

    • Replacing the Air Filter

      50

    • Replacing the Door Gasket

      51

    • Checking the Safety Valve

      51

    • Cleaning Water Outlet Strainer

      54

  • 11 Troubleshooting

    55

  • Baskets and Containers

    57

  • 12 Spare Parts List

    59

  • 13 Accessories

    59

Advertisement

Related Products

  • Tuttnauer 5050

  • Tuttnauer 5075 HSG

  • Tuttnauer 5075 EL

  • Tuttnauer 5075 GS-2A

  • Tuttnauer 50125 GS-2A

  • Tuttnauer 5050 ELV-D

  • Tuttnauer 5075 ELV-D

  • Tuttnauer 5075 EL-MD

  • Tuttnauer 5075 EL-D

  • Tuttnauer 5050 EL

Tuttnauer Categories

Laboratory Equipment

Medical Equipment

Steriliser

Accessories

Cleaning Equipment

More Tuttnauer Manuals

  • LEEC W100

    1 USER MANUAL LEEC WARMING CABINET W100, W157 and W330 CONTENTS 1) GENERAL DESCRIPTION 2) INSTALLATION AND SWITCHING ON WARMING CABINET 3) PROGRAMMING TEMPERATURE 4) OVERTEMPERATURE CUT-OUT 5) WARMING CABINET USEAGE 6) CLEANING & DISINFE …

    W100 Laboratory Equipment, 8

  • Interacoustics Sera

    D-0122934-A – 2019/02 Page 1 of 2 Quick Guide SeraTM Cart Assembly About this quick guide This quick guide provides instructions about how to place and attach the SeraTM and accessories on the cart. Follow the cart manufacturer’s instructions to assemble the cart prior to installing the SeraTM and accessories. …

    Sera Measuring Instruments, 2

  • Dynojet Power commander III

    i303-411 www.powercommander.com 1998-2003 Suzuki TLR1000 — PCIII USB — 11998-2003 Suzuki TLR1000Installation InstructionsDynojet Research 2191 Mendenhall Drive North Las Vegas, NV 89081 (800) 992-4993 www.powercommander.comParts List1 Power Commander1 USB Cable1 CD-ROM1 Installation Guide1 Power Adapter2 Power …

    Power commander III Motorcycle Accessories, 4

  • Barnstead Thermolyne Corporation Lab-Line 4628

    (217) 352-9330 | [email protected] | artisantg.com-~ ARTISAN® ~I TECHNOLOGY GROUP Your definitive source for quality pre-owned equipment. Artisan Technology Group Full-service, independent repair center with experienced engineers and technicians on staff. We buy your excess, underutilized, and idle equipment along w …

    Lab-Line 4628 Laboratory Equipment, 26

  • Thermo Scientific MaxQ 4000

    MaxQ 4000 Incubated andRefrigerated ShakersOperating Manual and Parts List 057-287-00 Rev. 1_________________________________________________________________________________Part of Thermo Fisher Scientific …

    MaxQ 4000 Laboratory Equipment, 44

  • Gilson ASPEC GX-274

    Gilson,Inc. | PO Box 620027 | Middleton, WI 53562-0027 | Tel: 608-836-1551 OR 800-445-7661 | Fax: 608-831-4451GilsonS.A.S. | 19, avenue des Entrepreneurs | BP 145, F-95400 Villiers-le-bel, FRANCEwww.gilson.com | [email protected] | [email protected] | [email protected] Qualication ProceduresGX-27 …

    ASPEC GX-274 Laboratory Equipment, 5

  • Julabo HT60-M2

    19533727-V1.docx Print date: 13.01.14 Operating Manual High Temperature Circulator HT60-M2 HT60-M3 C.U.-cooling unit 1.953.3727-V1 us 01/14 Forte HT with cooling unit Version: V 4.01-J JULABO USA, Inc. 884 Marcon Boulevard  Allentown, PA 18109 Phone: +1(610) 231-0250  Fax: +1(610) 231 …

    HT60-M2 Laboratory Equipment, 67

  • TESTO 875i

    Brief instructions entesto 875i · Thermal imagerContactGREAT BRITAIN / IRELANDTesto Ltd.Alton, Hampshire GU34 2QJ Tel. (14 20) 54 44 33, Fax (14 20) 54 44 E-mail: [email protected] Inc.Flanders, NJ. 07836Tel. (973) 2 52 17 20, Fax (973) 2 52 17 29E-mail: [email protected] Pty LtdCroydon South VIC 313 …

    875i Thermal cameras, 2

В наличии документация на медицинское оборудование фирмы Tuttnauer (в т.ч. на русском языке):

Modification: 06 September 2023

Инструкция по применению и обслуживанию (User and Service manual) на Автоклав Model 4483 1R — EP [Tuttnauer] Стерилизаторы

Инструкция по применению и обслуживанию User and Service manual на Автоклав Model 4483 1R - EP [Tuttnauer]

Эксплуатационная и сервисная документация (Operating and Service Documentation) на Автоклав Model 5596 EP-1 [Tuttnauer] Стерилизаторы

Эксплуатационная и сервисная документация Operating and Service Documentation на Автоклав Model 5596 EP-1 [Tuttnauer]

Инструкция по экспл. и обслуживанию (Operating and Service Documentation) на Автоклав Model 5596 SP-1 [Tuttnauer] Стерилизаторы

Инструкция по экспл. и обслуживанию Operating and Service Documentation на Автоклав Model 5596 SP-1 [Tuttnauer]

Эксплуатационная и сервисная документация (Operating and Service Documentation) на Автоклав Model 66120 [Tuttnauer] Стерилизаторы

Эксплуатационная и сервисная документация Operating and Service Documentation на Автоклав Model 66120 [Tuttnauer]

Эксплуатационная и сервисная документация (Operating and Service Documentation) на Автоклав Model 69150 [Tuttnauer] Диагностика

Эксплуатационная и сервисная документация Operating and Service Documentation на Автоклав Model 69150 [Tuttnauer]

Эксплуатационная и сервисная документация (Operating and Service Documentation) на Автоклавы модели 1730,2340,2540,3850,3870 серии E, EK, EA & EKA [Tuttnauer] Стерилизаторы

Эксплуатационная и сервисная документация Operating and Service Documentation на Автоклавы модели 1730,2340,2540,3850,3870 серии E, EK, EA & EKA [Tuttnauer]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на Модели 1730,2540,3850,3870,5050,5075 EL [Tuttnauer] Стерилизаторы

Техническое описание, инструкция по эксплуат. Technical description, instructions на Модели 1730,2540,3850,3870,5050,5075 EL [Tuttnauer]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на Модели 2540 3850 3870 5050 5075 ELV [Tuttnauer] Стерилизаторы

Техническое описание, инструкция по эксплуат. Technical description, instructions на Модели 2540 3850 3870 5050 5075 ELV [Tuttnauer]

Эксплуатационная и сервисная документация (Operating and Service Documentation) на Модели 2540 3850 3870 5050 5075 MLV [Tuttnauer] Стерилизаторы

Эксплуатационная и сервисная документация Operating and Service Documentation на Модели 2540 3850 3870 5050 5075 MLV [Tuttnauer]


Архив технической документации медицинского оборудования © 2006 … 2023

В связи с нестабильным курсом валют, цены могут отличаться. Обращайтесь за уточнениями к менеджерам.

  • КОД: ЦБ-00012814
  • Артикул:
  • Бренд: Tuttnauer

Товар снят с производства

Его сняли с производства и скорее всего у других продавцов
его
больше нет.

Доставка по всей России

  • Самовывоз — бесплатно
  • В офис или домой — от 250 ₽
  • До ПВЗ — от 250 ₽

Условия доставки

Характеристики

Артикул:

Бренд: Tuttnauer

Страна:
Флаг
Израиль

Описание

Серия автоклавов Tuttnauer ELV D-Line – это современные многофункциональные паровые стерилизаторы с большим количеством настраиваемых параметров и функций, разработаны для покрытия широкого спектра задач лабораторий, исследовательских институтов и биотехнологических учреждений.

ОСОБЕННОСТИ:

  • температурный диапазон стерилизации от 105°C до 138°C
  • удобная система управления с цветным дисплеем
  • до 30 программ стерилизации
  • два гибких датчика температуры PT100 внутри камеры обеспечивают контроль температуры стерилизуемых жидкостей, чувствительных к воздействию температур и времени стерилизации
  • защитное устройство, предотвращающее открывание крышки автоклава если камера находится под давлением
  • защитное устройство, предотвращающее начало цикла стерилизации при условии открытой крышки
  • рабочая камера оснащена предохранительным клапаном, позволяющем выпускать избыток давления в случае превышения давления в камере выше установленного значения
  • защита встроенного парогенератора: система слежения за уровнем воды поддерживает постоянный уровень воды для безопасной работы нагревательных элементов
  • камера автоклава из нержавеющей стали 316L
  • высококачественные лабораторные автоклавы Tuttnauer выполнены в соответствии со строгими международными директивами и стандартами

ДОПОЛНИТЕЛЬНЫЕ ОПЦИИ:

  • система быстрого охлаждения за счёт контура водяного охлаждения
  • вакуумный насос для предвакуумного удаления воздуха и поствакуумного удаления влаги. Для более эффективной сушки используется в комбинации с парогенератором (EL модели)
  • независимый встроенный парогенератор
  • система контроля процесса стерилизации F0 для чувствительных материалов
  • дезинфекционный/изотермический цикл от 60°C до 94°C для мягкого нагрева и охлаждения агаризованных сред
  • цикл прогрева при температуре от 95°C до 104°C
  • система биологической фильтрации с фильтром 0,2 мкм для очистки выводимого из автоклава воздуха
  • встроенный принтер
  • внешний независимый самописец
  • удалённое управление
  • камера автоклава из нержавеющей стали 316Ti вместо стандартной 316L

АКСЕССУАРЫ:

  • контейнеры с вентиляционными отверстиями
  • корзины для вертикальных автоклавов
  • передвижное подъемное устройство для вертикальных автоклавов
  • ролики для вертикальных автоклавов

Отзывов пока нет, но ваш может стать первым

о
товаре 5075 ELV D-Line — лабораторный автоклав с вертикальной загрузкой, объём камеры 160 л

Оформление заказа

Оформление заказа

Укажите ваши данные

Запросить аналог товара

Введите поисковый запрос

Описание Лабораторный автоклав вертикальный Tuttnauer 5075 ELV-WR D-line 

Лабораторный автоклав вертикальный Tuttnauer 5075 ELV-WR D-line. Новая линейка вертикальных автоклавов Tuttnauer WR со
встроенным резервуаром для обессоленной воды разработана для случаев, когда
отсутствует возможность подключения стерилизатора к воде и дренажной системе.
Вода из резервуара автоматически подается в камеру, а конденсат и отработанная
жидкость после завершения цикла сливаются обратно в резервуар.

Автоклавы D-line оборудованы простой русскоязычной системой
управления, которая полностью соответствует современным требованиям и позволяет
вам успешно решать

Области применения: любые задачи в современной лаборатории. Лабораторный автоклав вертикальный Tuttnauer 5075 ELV-WR D-line купить в Тольятти по приемлемой цене.

Лабораторный автоклав с вертикальной загрузкой
5075 ELV-WR
оснащён резервуаров для обессоленной воды, микропроцессорной
панелью управления и камерой объёмом 160 литров,
изготовленной из нержавеющей стали.

Области применения:

— Исследовательские лаборатории;

— Медико-биологические лаборатории;

— Учебные заведения;

— Производство.

Преимущества автоклавов серии EL D-Line:

— Стерилизация жидкости с различными опциями охлаждения;

— Стерилизация пипеток и стекол микроскопа;

— Стерилизация инструментов (обернутых и необернутых);

— Контроль биологической опасности и стерилизация отходов;

— Приготовление агара;

— Лабораторные автоклавы оснащены рядом независимых механических
и цифровых средств защиты. Защитное устройство не позволяет оператору открыть
дверцу, пока камера находится под давлением. При открытой дверце пар в камеру
не попадает.

Если дверца открыта или не заблокирована надлежащим образом,
цикл не запускается.

Дверца не может быть разблокирована до тех пор, пока
температура жидкости не достигнет предварительно установленной конечной
температуры. Дверца не может быть разблокирована до тех пор, пока давление в
камере не достигнет комнатного давления;

— Двойной независимый контроль – это сочетание электронного
и механического контроля, который обеспечивает оператора двумя независимыми
средствами контроля давления;

— Камера оснащена предохранительными клапанами: если
давление превышает допустимое предельное значение, предохранительные клапаны
будут разгружены.

Особенности 5075 ELV-WR D-line:

— Продвинутая система управления. Информация о 200 последних
циклах автоклавирования хранится во встроенной памяти; есть два температурных
сенсора; есть возможность удаленного мониторинга с ПК;

— Автоматический замок двери;

— Цветной дисплей;

— Русскоязычное меню;

— До 10 фиксированных программ стерилизации;

— До 20 настраиваемых программ стерилизации;

— Два погружных температурных датчика, которые обеспечивают
контроль температуры стерилизуемой жидкости, что особенно важно при
стерилизации растворов, чувствительных к температурному режиму (например,
микробиологических сред);

— USB-порт для передачи данных.

Максимальная загрузка колб Эрленмайера или флаконов
с винтовой крышкой в автоклаве 5075 ELV-WR:

Объём колб или
флаконов, мл

250

500

1000

2000

3000

5000

Количество колб

3×21

3×14

3×8

2×5

2×4

1×2

Количество флаконов

3×32

3×21

3×15

2×8

2×4

Дополнительные опции:

— Встроенный принтер EL-D/ELV-D/ELV-WR/EL;

— Ролики для вертикальных автоклавов;

— Независимое записывающее устройство для автоклавов
Tuttnauer;

— Дополнительный температурный датчик PT100 для автоклавов
Tuttnauer;

— Распределительная коробка для автоклавов Tuttnauer серии
D-Line;

— Передвижное подъемное устройство для вертикальных
автоклавов;

— Микробиологический фильтр для биол. опасных
материалов BH для настольных автоклавов Tuttnauer;

— Камера из нержавеющей стали 316 Ti для настольных
автоклавов Tuttnauer;

— Программа R.PC.R для создания отчетов на ПК;

— Защита информации 21 CFR part 11;

— Пакет валидационных протоколов IQ/OQ/PQ;

— Загрузочные корзины из нержавеющей стали для вертильных
автоклавов Tuttnauer серии D-Line;

— Загрузочные контейнеры из нержавеющей стали для
вертикальных автоклавов Tuttnauer серии D-Line.

Технические характеристики 5075 ELV-WR D-line:

— Полезный объем камеры – 160 л;

— Диаметр/глубина камеры – 500×750 мм;

— Габариты, Д×В×Ш – 770×1090×770 мм;

— Вес – 171
кг;

— Расход воды на цикл – 20 л;

— В/Мощность – 208/230/380-400, 9 кВт.