Puritan bennett руководство

Covidien Puritan Bennett 540 User Manual

  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

Puritan Bennett 540™

Ventilator

User’s Manual

loading

Related Manuals for Covidien Puritan Bennett 540

Summary of Contents for Covidien Puritan Bennett 540

  • Page 1
    Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 2
    Copyright information Copyright 2009 Nellcor Puritan Bennett LLC. All rights reserved. Puritan Bennett 540™ is a trademark of Nellcor Puritan Bennett LLC and/or its affiliates. COVIDIEN, COVIDIEN with Logo and ™ marked brands are trademarks of Covidien AG or an affiliate.
  • Page 3: Table Of Contents

    Connecting to an External DC Power Source ……..4–5 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 4
    Pneumatic …………..A–12 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 5
    …………….Index–7 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 6
    This page intentionally blank…
  • Page 7
    Puritan Bennett 540™ Portable Ventilator ……..
  • Page 8
    This page intentionally blank…
  • Page 9
    List of Circuits…………..E–2 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 10
    This page intentionally blank…
  • Page 11: Preface

    Bennett. Extended Service The Puritan Bennett 540™ Portable Ventilator offers extended service contracts/warranties for purchase when the ventilator is purchased. Please contact your local Covidien/Puritan Bennett Sales or Service Representative for additional information. Preface-1 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 12
    540™ Ventilator and other Puritan Bennett products 24 hours a day, 7 days a week. If you require further assistance, contact your local Puritan Bennett representative or call Puritan Bennett Technical Support at 1.800.255.6774. Preface-2 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 13: Safety Information

    Indicates points of particular emphasis, that make operation of the ventilator more efficient or convenient. It is essential to read, understand and follow these instructions before using the Puritan Bennett 540™ Ventilator. In order to use the ventilator correctly and efficiently and to help prevent incidents, please pay particular atten- tion to sections 1.2,…

  • Page 14: Warnings

    Refer to chapter “Cleaning”. • Handle the ventilator with care during and after use, particularly when room temperatures are high. Some ventilator surfaces may become hot, even if safety specifications are not exceeded. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 15
    This classification and regulatory requirements may vary depending upon the country and mode of transport. In addition, air transport of the Puritan Bennett 540™ Ventilator is only allowed as carry-on baggage, which may include up to two (2) individually packaged spare Lithium-ion batteries.
  • Page 16
    Batteries should be disposed of according to environmental legislation in your country and locality. • Never expose any batteries to direct flame. • Never touch the ventilator’s internal components, including the battery, and the patient simultaneously. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 17
    When using non-invasive ventilation (niv), use a non-vented nose or face mask. • Before using the Nurse Call system, ensure that its connections are secure and it operates properly. For more information, contact Puritan Bennett. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 18
    A continuous alarm condition will be activated if the ventilator power switch is turned off while ventilation is in progress. When the power switch is turned back on again, the ventilation will resume without having to press the VENTILATION ON/OFF button. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 19
    (see section 10, “Routine Maintenance in the Clinician’s Manual). This is particularly important when the ventilator is installed on a wheelchair, because environmental conditions may cause the filter to become dirty more rapidly. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 20
    • The Puritan Bennett 540™ Ventilator does not include an oxygen analyzer. Always measure the delivered oxygen with a calibrated oxygen analyzer that features a high and low concentration alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.
  • Page 21: Cautions

    General Precautions For Use Caution • To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only qualified personnel should attempt to service the Puritan Bennett 540™ Portable Ventilator. General Precautions For Installation Caution Environment: •…

  • Page 22: Symbols And Markings

    Symbols and Markings Table 1-1. Ventilator Symbols Symbols Descriptions It is essential to read, understand, and follow these instructions before using the Puritan Bennett 540™ Ventilator (ISO 7000-0434A). This symbol appears on the ventilator’s back panel, and internal battery; see Table 1-2, item 11.

  • Page 23
    Figure 4-9 on page 4-9. Oxygen inlet. This marking appears on the back panel of the ventilator, adjacent to the oxygen inlet port; Figure 1-4, item 3, and Figure 4-14 on page 4-14. 1-11 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 24
    Preferences menu. WEEE (Waste Electrical and Electronic Equipment). This symbol means that this product must not be disposed of with household waste. Observe local ordinances for proper disposal. Refer to Table 1-2, item 6. 1-12 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 25: Labels / Identification And Instruction Information

    Use the item numbers in the following tables to locate the labels in Figure 1-1 through Figure 1-5. 1-13 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 26: Ventilator Labels And Markings

    5. Exhalation Limb 6. Air Inlet Label 7. Exhaled Gas Outlet Label Connection of Patient Circuit – (Figure 1-4) (Figure 1-2, Figure 1-4) Single Use Exhalation Block Label (Figure 1-1, Figure 1-2, Figure 1-5) 1-14 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 27
    19. AC Mains Warning (Located Inside Connector Marking 18. ESD (Located inside the Device) the Device) (Figure 1-4) Note: The item number callouts in the following figures refer to those listed in Table 1-2. 1-15 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 28: Figure 1-1. Locations Of Labels — Top-Front View

    Safety Information Figure 1-1. Locations of Labels – Top-Front View Figure 1-2. Locations of Labels – Front-Left View 1-16 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 29: Figure 1-3. Location Of Labels — Front-Right View

    Labels / Identification and Instruction Information Figure 1-3. Location of Labels – Front-Right View Figure 1-4. Location of Labels and Markings – Back-Left View 1-17 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 30: Figure 1-5. Location Of Labels — Bottom-Front View

    Safety Information NOTE: To view labels (items 11 and 14), Battery Cover and Internal Battery must be removed. Refer to section 6.6, “Replacing the Battery. ” Figure 1-5. Location of Labels – Bottom-Front View 1-18 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 31: Ventilator Overview

    Nurses • Home care providers • Patient and patient’s families For more details on the knowledge and skill requirements for operating the Puritan Bennett 540™ Ventilator Ventilator, refer to Appendix A, “Patient/Caregiver Checklist” in the Clinician’s Manual. WARNING Federal law restricts this device to sale by or on the order of a licensed physician.

  • Page 32: Operational Use

    Ventilator Overview Operational Use The Puritan Bennett 540™ Ventilator uses a micro-turbine and the patient circuit with integral exhalation valve to provide ventilatory support to patients. Clinicians may use a variety of interfaces to connect patients to the ventilator: nasal or full face masks; endotracheal or tracheotomy tubes.

  • Page 33: Front Panel

    Exhaled Gas Outlet– Exhalation Valve connects monitoring proximal patient pressure. here. a. If exhaled volume monitoring is required, use the double-limb circuit for exhaled tidal volume or minutes volume monitoring. Figure 2-1. Front Panel Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 34: Back Panel

    USB functionality to allow for system upgrades. AC power cable. Not currently used. Access cover for the internal battery. Air Inlet Filter: Filters air as it enters the ventilator. DC power cable connector with key. Figure 2-2. Back Panel Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 35: Control Panel

    DC POWER indicator lit: DC power source connected. • Access to a sub-menu. • INTERNAL BAT indicator lit continuously: Ventilator powered by internal battery (no external power source connected.) • INTERNAL BAT indicator flashing: battery charging. Figure 2-3. Control Panel Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 36: Ventilation Menu

    I:E ratio when the patient actively triggers a Refer to chapter 3, “Alarms and breath. Troubleshooting” for details. Figure 2-4. Ventilation Menu Display Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 37: Alarm Menu

    Detected symbol appears date and end-of-event adjacent to the monitored I:E time. ratio when the patient actively triggers a breath. Refer to chapter 3, “Alarms and Troubleshooting” more information. Figure 2-5. Alarm Menu Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 38: Waveforms Menu

    Keep in mind that troubleshooting information is available in this manual to assist you in the event of a problem. Refer to chapter 3, “Alarms and Troubleshooting”. If you cannot determine the cause of a problem, contact your equipment supplier or Puritan Bennett. Puritan Bennett Technical Services (USA): 1-800-255-6774 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 39: Alarms And Troubleshooting

    3 Alarms and Troubleshooting WARNING Setting Alarm limits to extreme values can cause the ventilator alarms to malfunction. The alarms or faults generated by your Puritan Bennett 540™ Ventilator are classified into two categories: • Ventilation (or utilization) alarms •…

  • Page 40: Alarm Display

    There are currently no Low Priority (LP) Alarms. When an alarm is triggered, if the current menu displayed is not the Ventilation parameters or Alarm menu, the display automatically switches to one of these menus to display the alarm message. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 41: Alarm Logs Menu

    The display appears as follows: 3. Press the ENTER key. The Alarm Logs screen is displayed. Note: When no alarm has been activated, “NO DATA” is displayed on the screen (see graphic below). Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 42: Silencing The Audible Portion Of Alarms

    Note: Only qualified service personnel may access all alarms and events recorded by the ventilator. Qualified personnel should refer to the Puritan Bennett 540™ Service Manual for further information. Silencing the Audible Portion of Alarms You may silence the audible portion of alarms for 60 seconds at a time. This is referred to as the Audio Paused function.

  • Page 43: Resetting/Pausing Alarms

    The High Pressure alarm must be manually reset. Refer to section 3.7, “Overview of Alarms, ” on page 3- To manually reset the High Pressure Alarm, proceed as follows: Press the ALARM CONTROL key twice. • The visual alarms will be reset. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 44: Re-Activating Alarms

    • The messages of all active alarms are displayed in a loop in the Ventilation and Alarm menus. • The Audio Paused symbol disappears (if it was displayed). • The Alarm Paused symbol disappears. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 45: Overview Of Alarms

    Consequence: battery charging *IF PERSISTS stops. REPLACE VENT Failure of one calibration point of the internal exhaled flow sensor. CAL FAULT Consequence: failed calibration SEE USER MNL point is replaced by the default point. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 46
    CHECK Consequence: SETTINGS • Locking Key disabled • Out-of-range settings are replaced by their default values CHK EXH VALV RESTART VENT* • Faulty pressure valve sensor signal at ventilation start. *IF PERSISTS REPLACE VENT Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 47
    Detection of a fault in sensor DEVICE FAULT7 voltage measurement. REPLACE VENT Inspiratory flow is constant (+/- 1 lpm) with normal turbine DEVICE FAULT8 temperature and speed REPLACE VENT conditions. Contact your service representative for assistance. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 48
    HIGH INSP VOLUME Alarm activation occurs: • After three consecutive breaths. Minute volume higher than High MinVol set during three consecutive breaths. HIGH MINUTE Alarm activation occurs: VOLUME • After three consecutive breaths. 3-10 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 49
    LOW BATTERY < 30 min. or 5%. Expired tidal volume less than Low VTE set during three consecutive breaths (in double- limb setup). LOW EXH VOLUME Alarm activation occurs: • After three consecutive breaths. 3-11 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 50
    Alarm activation occurs: *IF PERSISTS • REPLACE VENT After three consecutive breaths. Cut-off of the external DC power ON INTERNAL supply. BATTERY Consequence: switch-over to (with Periodic the internal battery. Audible Signal) 3-12 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 51
    Contact your SPKR FAULT3 service representative for REPLACE VENT assistance. Buzzer Battery Failure. The Battery Buzzer Voltage is too SPKR FAULT4 low, disabling the buzzer. REPLACE VENT Contact your service representative for assistance. 3-13 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 52: Troubleshooting

    • To ensure proper servicing of the ventilator and reduce the possibility of physical injury, only qualified personnel should attempt to service the Puritan Bennett 540™ Ventilator. 3.8.1 Alarms WARNING When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator.

  • Page 53
    BATT FAULT4 Battery charging impossible. cannot be involuntarily disconnected. REPLACE VENT In the event the internal battery capacity is low, use an alternate device to ventilate the patient. Call your customer service representative. 3-15 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 54
    EXH VALVE1 * phase. Have a qualified technician replace the *IF PERSISTS Contaminated or defective expiratory flow defective component(s) and call your REPLACE VENT sensor. customer service representative. 3-16 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 55
    Reminder: the internal battery can be overextended. charged only when the ventilator connected to an AC power supply. Check and properly connect the patient EXHSENS FAULT OR Leak in the patient circuit. circuit connections. CIRCUIT FAULT 3-17 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 56
    Treat patient’s cough. Patient inspiratory resistance or Have physician determine new ventilator compliance changes. settings. Defective internal circuits of the machine or Replace the ventilator and call your pressure sensor. customer service representative. 3-18 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 57
    Replace the defective component(s) and Defective expiratory flow sensor. Call your customer service representative or clinician. Under the guidance of a Clinician modify Adjustment of the Low VTE level too high. the Low VTE level. 3-19 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 58
    Replace the ventilator and call your Ventilator’s current-limiting fuse blown. customer service representative. Internal problem in the electrical power Replace the ventilator and call your POWER FAULT supply. customer service representative. REPLACE VENT 3-20 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 59: Other Problems

    Replace the ventilator and call your Internal technical problem that prevents SPKR FAULT4 customer service representative. the battery warning buzzer from operating. REPLACE VENT 3.8.2 Other Problems Table 3-3 provides other possible ventilator problems, causes, and corrective actions. 3-21 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 60: Table 3-3. Additional Problems And Corrective Actions

    Remove obstructions from all blocked Excessive heat inlets of the casings. ventilator air inlets and outlets. emitted Replace the ventilator and call your Condensation Liquid entered the device. customer service representative. inside the device 3-22 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 61: Installation And Assembly

    The use of a bacterial filter at the ventilator’s outlet—or both ports if a double-limb circuit is used—is recommended. Installing the Ventilator To install your Puritan Bennett 540™ Ventilator correctly we recommend that you proceed in the following manner: •…

  • Page 62: Connecting To External Ac Power

    The use of any accessory other than those specified by Puritan Bennett, may lead to an increase in electromagnetic emissions or a decrease in the equipment’s insulation against electromagnetic emissions. • The Puritan Bennett 540™ Ventilator requires special precautions for electromagnetic compatibility and should be installed and started according to the recommendations stated in Appendix A, “Specifications”.

  • Page 63: Figure 4-2. Inserting The Power Cable Holder Into The Notch

    The only time the AC POWER and INTERNAL BAT indicators are illuminated at the same time is when the ventilator is connected to an AC supply and the battery is charging (INTERNAL BAT indicator is flashing). Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 64
    2. Disconnect the AC power cable from the ventilator’s AC connector at the rear of the device. 3. Grasp the AC power cable at the level of the power cable holder and turn the cable counterclockwise while lifting it upwards and out of the holder. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 65: Connecting To An External Dc Power Source

    Connect the external DC power source by first connecting the power cable to the ventilator and then to the external DC source. Follow the reverse procedure to disconnect the device from the external DC power source. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 66: Patient Circuit

    The ventilator and patient circuit should always be placed so as to provide a secure, comfortable fit for the patient. • The patient circuit should always be positioned to avoid hindering the patient’s movements, to prevent accidental disconnection, and to minimize the risk of patient strangulation. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 67: Choosing The Patient Circuit Type

    Always assure that the patient is receiving the appropriate inspired volume when altering the breathing circuit configuration. • Users must always possess an additional circuit and valve while using the Puritan Bennett 540™ Ventilator. 4.4.1 Choosing the Patient Circuit Type Single limb circuits are used with breathing modes where spirometry measurements are not required, and double limb circuits are used with breathing modes where spirometry is required.

  • Page 68
    8. Ensure that the water trap (item 3) on the expiratory limb of the patient circuit (item 7) is securely installed between the patient wye (item 5) and the exhalation valve (item 9). Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 69: Figure 4-9. Close-Up Of Exhalation Valve Tube And Proximal Pressure Tube

    LOW PRESSURE DISCONNECT triggering threshold by doing one of the following: set the high VTI alarm limit for pressure modes or set low VTE alarm limit for all ventilation modes if using a dual limb circuit. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 70: Filters

    Air Inlet Filter Consisting of foam and fine particle filter media and located at the rear of the ventilator, this filter cleans the air as it enters the ventilator. Refer to the figure below. 4-10 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 71: Humidifier

    4-12) adds moisture (water vapor) and warms the air in the patient circuit. It is inserted into the patient circuit between the main outlet and the patient (refer to Figure 4-7 Figure 4-8). 4-11 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 72: Exhalation Block

    • Each time an exhalation block is installed, the expiratory flow sensor must be recalibrated before using the ventilator. Caution Ensure that the exhalation block has been completely dried after it is cleaned. 4-12 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 73: Oxygen

    Since the factors that affect administered oxygen flow may change over time, you must ensure that these settings always correspond to the current oxygen therapy objectives specified by the physician. Note: To monitor patient Oxygen levels use an external sensor/alarm. 4-13 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 74: Connecting The Oxygen Supply

    Inspect the oxygen coupler (Figure 4-15, item 2) before use to ensure it has its black O-ring attached and in good condition. Do not use an oxygen coupler with a missing, damaged, or worn O-ring. 4-14 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 75: Figure 4-15. Connecting The Oxygen Supply System

    3. Disconnect the oxygen supply’s oxygen connector by pulling it towards you. The ventilator’s oxygen connector’s locking stud (Figure 4-15, item 2) will then extend outwards, which is required before the oxygen connector can be reconnected. 4-15 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 76: Mounting The Ventilator On A Wheelchair

    Thoroughly air the room to bring the oxygen concentration down to normal. • To prevent any interference with the internal sensors of the Puritan Bennett 540™ Ventilator, do not install a humidifier upstream of the ventilator. Caution •…

  • Page 77: Mounting The Ventilator On The Utility Cart

    Before using the ventilator on the internal battery, ensure that it is charged and that the charge holds (refer to chapter 6, “Internal Battery”). The Dual Bag accessory consists of a carrying bag that allows the Puritan Bennett 540™ Ventilator to be both mounted onto a wheelchair and carried as a backpack (see Figure 4-17).

  • Page 78: Connecting The Nurse Call Cable

    Puritan Bennett. Caution • To connect the Puritan Bennett 540™ Ventilator to a nurse call device, first contact Puritan Bennett to verify device compatibility and to obtain a suitable connection cable. • Do not use Nurse Call devices that operate based on the closure of an electrical circuit, because the devices often do not take into account possible cable disconnection or a total loss of power.

  • Page 79
    A remote alarm is activated when an alarm condition occurs, unless either of the following is true: • The audio paused function is active. • The ventilator power switch is OFF. • The remote alarm port is an 8-pin female connector; allowable current is 100mA at 24VDC(max). 4-19 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 80
    Installation and Assembly This page intentionally blank 4-20 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 81: Operating Procedures

    Turning on The Ventilator Caution After storage of the Puritan Bennett 540™ Portable Ventilator in an atmosphere that differs significantly in temperature from where it is installed—typically, ± 36 °F (20 °C)—stabilize the device at room temperature for at least two (2) hours prior to starting it.

  • Page 82: Starting Ventilation

    Starting Ventilation Before starting ventilation, refer to Appendix B, “Operational Verification Checklist”, and set the parameter values in the Preferences menu. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 83
    The blue light indicator, at the upper right of the VENTILATION ON/OFF (see Figure 5-4, item 2), turns off. • A “beep“ sounds. • The ventilation starts. • The values of the monitored parameters are displayed in the right-hand window. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 84: Stopping Ventilation

    1. Hold down the VENTILATION ON/OFF (Figure 5-4, item 1) for about three (3) seconds. • A message prompting the user to keep the button pressed appears on the monitoring window, as shown in the graphic below: Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 85: Turning Off The Ventilator

    AC POWER indicator is illuminated), the internal battery continues charging. WARNING Handle the ventilator with care after use, particularly when room temperatures are high. Some ventilator surfaces may be very hot, even if safety specifications are not exceeded. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 86
    Operating Procedures This page intentionally blank Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 87: Internal Battery

    6 Internal Battery WARNING • Because the internal battery of the Puritan Bennett 540™ Ventilator contains more than 8 grams (equivalent) lithium, it is categorized as Dangerous Goods (DG) Class 9—even though the ventilator meets current safety standards. The ventilator, because of this categorization, is subject to strict transport conditions under the…

  • Page 88: Battery Operation

    • Battery reserve capacity is displayed on the right of the symbol. • The “ “ indicator at the top left of the ventilator’s front panel is continuously lit INTERNAL BAT (Figure 6-1). Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 89
    If the ventilator is running, the internal battery reserve is momentarily displayed as a percentage. Then, after the ventilator calculates the battery time remaining (which takes about two minutes, Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 90: Testing The Battery

    Recharging the Battery In the event that the battery charge level is considered insufficient, as per the reserve capacity display, recharge of the internal battery is necessary. In general, it is recommended that the ventilator be Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 91: Storing The Battery

    If the device is to be stored for an extended period of time, it is not necessary to remove the battery. However, the battery should be stored in cool, dry, well-ventilated environment, as follows: Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 92: Replacing The Battery

    You remove the protective “pull tabs” (if they are present) from the battery’s contacts. • The battery’s contacts align with those of the ventilator, according to the drawing on the label on the back of the battery cover. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 93
    7. Install the three Torx screws (Figure 6-3, item 1) to secure the battery cover (Figure 6-3, item 2) to the ventilator and tighten the screws with the Torx T10 screwdriver. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 94
    This page intentionally blank…
  • Page 95: Cleaning

    Table 7-1. Approved Cleaning Solutions for Exterior Ventilator Surfaces Description Product Names Tap water – Saline – Mild dishwashing detergent Joy® (or equivalent) 70% isopropyl alcohol (rubbing alcohol) – 10% chlorine bleach (90% tap water) Clorox® (or equivalent) Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 96: Cleaning The Accessories

    Caution Never use a liquid cleaner inside the patient circuit, or on any component of a gas or air pathway. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 97: Routine Maintenance

    The fine particle side of the filter faces outwards, away from the ventilator. b. The filter is properly installed in its housing. Proper installation of the filter prevents particles from entering the device. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 98: Recommended Schedule Of Maintenance

    See manufacturer’s recommendation Exhalation Block 3 months (for each new patient) Note: For more information regarding parts and accessories for the Puritan Bennett 540 Ventilator contact your service representative or www.puritanbennett.com. WARNING • Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator. Replace it when necessary—even before the recommended replacement period has elapsed, and particularly…

  • Page 99: Service Assistance

    Puritan Bennett. Corporate Address: Technical Service Phone Numbers: Covidien Troubleshooting your portable ventilator: 1-800-255-6774 6135 Gunbarrel Avenue Servicing your portable ventilator: 1-800-255-6774 Boulder, CO 80301 U.S.A. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 100
    Routine Maintenance This page intentionally blank Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 101: A Specifications

    Maximum allowable flow resistance: 4cmH2O at Requirement 60 lpm Electrical Table A-2. AC Electrical Supply Voltage Frequency Consumption 100 VAC to 240 VAC 50 Hz / 60 Hz 180 VA max. 12 VDC 8.3 A 30 VDC 3.3 A Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 102: Indicators And Alarms

    The remote alarm port is an 8-pin female remote supply -(not connector. Allowable current is 100 mA at used) 24 VDC (maximum). RX Signal (not used) TX Signal (not used) remote supply +(not used) not used Indicators and Alarms Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 103: Performance

    Table A-9. Volume Performance Test Results R-Rate (bpm) Insp Time VTI min. (ml) C50 R5 VTI max. (ml) 2000 2000 1750 1200 VTI min. (ml) C20 R20 VTI max. (ml) 1000 VTI min. (ml) C20 R20 VTI max. (ml) Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 104: Monitored Parameters

    Expiratory Minute Volume (Min VE) 0 to 99.9 l +/- (20ml + 20%) in CPAP mode above 200ml a. The PB 540 does not have the capability to reduce pressure below the PEEP pressure during the expiratory phase. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 105: Range, Resolution, And Accuracy

    Accuracy: +/- (4 lpm +10%) of target exhalation flow based on Exh Sense Default value: 25% In CPAP, V SIMV, and P SIMV, Exh Sens is fixed at 25% and is not adjustable. Ventilator Settings (Cont’d) Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 106
    Default value: 300 Depends on: High VTI High Inspired Tidal Volume Range: 80 mL to 3000 mL (High VTI) Resolution: 10 mL Default value: 2000 mL Depends on: Low VTI Pressure A/C Mode (Cont’d) Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 107
    Resolution: 10 mL Default value: OFF Depends on: High VTE High exhaled tidal volume Range: 80 mL to 3000 mL (High VTE) Resolution: 10 mL Default value: OFF Depends on: Low VTE PSV/CPAP Mode (Cont’d) Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 108
    Default value: OFF Depends on: Vt, High VTI High Inspired Tidal Volume Range: 80 mL to 3000 mL (High VTI) Resolution: 10 mL Default value: 2000 mL Depends on: Vt, Low VTI V SIMV Mode (Cont’d) Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 109
    Resolution: 10 mL Default value: OFF Depends on: High VTE High Exhaled Tidal Volume Range: 80 mL to 3000 mL (High VTE) Resolution: 10 mL Default value: OFF Depends on: Low VTE P SIMV Mode (Cont’d) Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 110
    Default value: OFF Depends on: Vt, Low VTE Low Minute Volume Range: 0.5 L to 50 L (Low Min Vol) Resolution: 0.5 L Default value: OFF Depends on: Vt, R-Rate, High Min vol A-10 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 111
    Resolution: 0.1 s Accuracy: ± 50 ms or 5%, whichever is greater Expiratory time (E time) Range: 0 s to 59.9 s Resolution: 0.1 s Accuracy: ± 50 ms or 10%, whichever is greater A-11 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 112: Environmental

    0.057 cmH O (5.6 pascals) at 24 lpm flow a. All testing conducted on 3-inch diameter test piece Table A-17. Oxygen Inlet Specifications Maximum pressure Maximum flow 7.25 psi (50 kPa) 15 lpm A-12 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 113: Manufacturer’s Declaration

    Table A-19. Electromagnetic Emissions The Puritan Bennett 540™ Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment.

  • Page 114: Table A-20. Electromagnetic Immunity

    IEC 61000-4-8 or hospital environment. Note: U is the AC mains voltage prior to application of the test level. A-14 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 115: Table A-21. Electromagnetic Immunity — Conducted And Radiated Rf

    , should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol: A-15 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 116: Table A-22. Recommended Separation Distances

    Legendair XL2 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Puritan Bennett 540™. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

  • Page 117: Standards Compliance And Iec Classification

    Lung Ventilators for Medical Use- Particular Requirements for Basic Safety and Essential Performance Part 2: Home Care Ventilators for Ventilator-Dependent Patient ISO10651-2 2004. • Standard Specification for Ventilators Intended for Use in Critical Care ASTM F 1100 1997. A-17 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 118
    Specifications This page intentionally blank A-18 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 119: B Operational Verification Checklist

    Perform the Functioning Alarms Test. Pass Verify the alarm volume is adapted to the patient environment. Pass Ensure that you know how to charge the internal battery. Refer to Pass chapter 6, “Internal Battery”. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 120
    If exhaled volume monitoring is required, use the Pass double-limb circuit for exhaled tidal volume or minute volume monitoring. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 121: C Unpacking And Preparation

    C Unpacking and Preparation The Puritan Bennett 540™ Ventilator is shipped from the factory with the following items: (1) Printed User’s Manual (1) Clinician’s Manual on CD (a print copy is available upon request by the customer) (1) Patient circuit and valve in one of the following configurations (as requested by the customer): •…

  • Page 122: Figure C-1. Puritan Bennett 540™ Portable Ventilator

    Puritan Bennett. 5. Clean the ventilator with a mild soap solution, if necessary (refer to chapter “Cleaning”). 6. Ensure that the air inlet filter is installed. Figure C-1. Puritan Bennett 540™ Portable Ventilator Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 123
    Figure C-2. Dual Bag Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 124
    Unpacking and Preparation This page intentionally blank Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 125: D Modes And Breath Types

    D Modes and Breath Types Modes of Ventilation This chapter is a general description of the various modes of ventilation and breath types available with the Puritan Bennett 540™ Ventilator. Note: The default ventilation mode setting is PRES A/C; for more information, see below.

  • Page 126: Psv/Cpap Mode

    (Pressure Support) to each of the patient’s breaths. This has the same benefits as CPAP, with the additional benefit of assisting the patient in moving air into his or her lungs. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 127: E Parts And Accessories

    E Parts and Accessories Table E-1 provides a list of accessories that are available for the Puritan Bennett 540™ Ventilator. To order parts or accessories, contact your equipment supplier or Puritan Bennett representative. Note: The ventilator is shipped from the factory with the following items: a printed User’s Manual; a CD with Clinician’s Manual (printed copy available upon request);…

  • Page 128
    Allegiance™/Cardinal™/Airlife™ single limb patient circuit, 6463-H08 ADULT Hudson™, Allegiance™, Cardinal™, and Airlife™ are the property of their respective owners. For more information regarding parts and accessories for the Puritan Bennett 540™ Ventilator contact your service representative or www.puritanbennett.com. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 129: F Glossary

    “Alarm Silence. ” Back Up Rate Rate of control cycles in PSV or SIMV modes during apnea phase. Battery Level Display of the remaining battery capacity; located adjacent to the battery symbol. Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 130
    Exhalation sensitivity The exhalation sensitivity (Exh Sens) level is a percentage of peak flow at which a pressure-supported breath will be terminated. Expiratory Tidal Volume (VTI) Volume exhaled by the patient at each expiratory phase. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 131
    Counter for the total ventilation time since manufacture or the last CPU board change. Mains AC power supply. Mean Airway Pressure. Average pressure during a ventilation cycle. Minimum Exhalation Time Minimum exhalation time before allowing the patient inspiratory trigger. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 132
    Pressure is maintained until inspiratory flow is reduced to a percentage of peak flow that depends on the expiratory sensitivity setting for the inspiration, when the ventilator cycles into exhalation. Available in SIMV or Spontaneous modes. Pounds Per Square Inch. Rebreathing The patient breathes his/her exhaled gas. Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 133
    The Puritan Bennett 540™ Ventilator is flow-triggered, with sensitivity levels in the range from 1 to 5: the lower the number, the more sensitive the trigger.
  • Page 134
    Glossary This page intentionally blank Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 135: Index

    CAL FAULT 3-16 silencing CALIBRATION FAIL 3-14 3-8, 3-16 Troubleshooting CHECK PROXIMAL LINE1 Alarms and troubleshooting CHECK PROXIMAL LINE2 3-8, 3-16 Alarms, utilization CHECK SETTINGS 3-8, 3-9, 3-16 Alarms, ventilation CHECK/CHANGE EXH VALVE1 Index-1 Puritan Bennett 540™ Ventilator User’s Manual…

  • Page 136
    CHK EXH VALV RESTART VENT alarm message EMPTY BATTERY alarm message 3-16 Environment 3-9, CHK VOL/INSP TIME SETTINGS alarm message suitable for use of ventilator 3-17 A-12 Environmental specifications 4-12 Classification of device Exhalation block Cleaning replacement interval Index-2 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 137
    OCCLUSION CHECK CIRCUIT alarm message 3-20 testing 3-12, 3-20, Internal battery, replacing (figure) ON INTERNAL BATTERY alarm message 3-11, 3-19 INVERSED IE RATIO alarm message I/O switch (figure) Operational verification checklist Operator/Users Index-3 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 138
    Standards, compliance, and IEC clasiification specifica- A-17 installation tions environment Starting ventilation maintenance Stopping ventilation 1-10 oxygen Storing the internal battery 4-14 Precautions for use, warnings Stud, oxygen connector general Index-4 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 139
    Turning on unpacking and preparation 1-3, 7-1 Ventilator, and liquid ingress (Warning) Warnings definition of general list of Preface-1 Warranty Waveforms menu Welcome Menu screen display of skipping Index-5 Puritan Bennett 540™ Ventilator User’s Manual…
  • Page 141
    Nellcor Puritan Bennett LLC 6135 Gunbarrel Avenue, Boulder, CO 80301 U.S.A. Telephone Toll Free 1.800.255.6774 © 2009 Nellcor Puritan Bennett LLC. All rights reserved. 10028295Rev B-0109…

………………………………………………………..……..

Operator’s Manual

Part No. G-061874-00
Rev. D
September 2000

Copyright Information

Copyright 2000 Mallinckrodt Inc. EasyCart, EasyNeb, 740, 760, and 700 Series are
trademarks of Mallinckrodt Inc. All rights reserved. The 700 Series
(including the 740
proprietary information, covered by one or more of the following U.S. Patents and foreign
equivalents: 5,524,615; 5,540,222; 5,596,984; 5,632,270; 5,664,560; and 5,673,689.

The information contained in this manual is the sole property of Mallinckrodt Inc. and may not
be duplicated without permission. This manual may be revised or replaced by Mallinckrodt
Inc. at any time and without notice. You should ensure that you have the most current
applicable version of this manual; if in doubt, contact the Technical Publications Department
of Mallinckrodt Inc. While the information set forth herein is believed to be accurate, it is not
a substitute for the exercise of professional judgment.

The ventilator should be operated and serviced only by trained professionals. Mallinckrodt’s
sole responsibility with respect to the ventilator, and its use, is as stated in the limited
warranty provided.

Nothing in this manual shall limit or restrict in any way Mallinckrodt’s right to revise or
otherwise change or modify the equipment (including its software) described herein, without
notice. In the absence of an express, written agreement to the contrary, Mallinckrodt Inc. has
no obligation to furnish any such revisions, changes, or modifications to the owner or user of
the equipment (including its software) described herein.

Ô

and 760Ôventilators) are manufactured in accordance with Mallinckrodt

Ô

Ventilator System

Definitions

This manual uses these special indicators to convey information of a specific
nature:

Warning

Indicates a condition that can endanger the patient or the ventilator
operator.

Caution

Indicates a condition that can damage the equipment.

NOTE:

Indicates points of particular emphasis that make operation of the
ventilator more efficient or convenient.

ii

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

……………………………………………………………..…..

Warnings, cautions, and notes

Please take the time to familiarize yourself with the following safety
considerations, special handling requirements, and regulations that govern the use
of the 700 Series Ventilator System.

Warning

To avoid an electrical shock hazard while servicing the ventilator, be
sure to remove all power to the ventilator by disconnecting the power
source and turning off all ventilator power switches.

Warning

To avoid a fire hazard, keep matches, lighted cigarettes, and all other
sources of ignition (e.g., flammable anesthetics and/or heaters) away
from the ventilator and oxygen hoses.

Do not use oxygen hoses that are worn, frayed, or contaminated by
combustible materials such as grease or oils. (Textiles, oils, and other
combustibles are easily ignited and burn with great intensity in air
enriched with oxygen.)

In case of fire or a burning smell, immediately disconnect the ventilator
from the oxygen supply and electrical power source.

Warning

Patients on life-support equipment should be appropriately monitored
by competent medical personnel and suitable monitoring devices.

The 700 Series Ventilator is not intended to be a comprehensive
monitoring device and does not activate alarms for all types of
dangerous conditions for patients on life-support equipment.

Warning

Check the ventilator periodically as outlined in the service manual; do
not use if defective. Immediately replace parts that are broken,
missing, obviously worn, distorted, or contaminated.

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

iii

Warning

An alternative source of ventilation should always be available when
using the 700 Series Ventilator System.

Warning

To ensure proper servicing and avoid the possibility of physical injury,
only qualified personnel should attempt to service or make authorized
modifications to the ventilator.

The user of this product shall have sole responsibility for any ventilator
malfunction due to operation or maintenance performed by anyone not
trained by Mallinckrodt staff.

Warning

For a thorough understanding of ventilator operations, be sure to read
the 700 Series Ventilator System Operator’s Manual in its entirety
before attempting to use the system.

Warning

Before activating any part of the ventilator, be sure to check the
equipment for proper operation and, if appropriate, run the self­diagnostic short self test (SST) program described in this manual.

Caution

U.S. Federal law restricts this device to sale by or on the order of a
physician.

Warranty

The 700 Series Ventilator System is warranted against defects in material and
workmanship in accordance with Mallinckrodt Medical Equipment Warranty for
a period of one year from the time of sale. To ensure the validity of the warranty,
be sure to keep a maintenance record.

iv

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

……………………………………………………………..…..

Year of manufacture

The 700 Series Ventilator System’s year of manufacture is indicated by the fifth
and sixth digits of the serial number which is located at the lower edge of the
ventilator front panel.

Manufacturer

anufactured by

Nellcor Puritan Bennett Ireland
A subsidiary of Mallinckrodt Inc.
Mervue, Galway, Ireland

Phone: +353.91.753.771
Fax: +353.91.753.922

European Headquarters

Mallinckrodt Europe BV
Hambakenwetering 1
5231 DD ’s-Hertogenbosch
The Netherlands

Phone: +31.73.6485200
Fax: +31.73.6410915

Electromagnetic susceptibility

The 700 Series Ventilator System complies with the requirements of IEC 60601­1-2 (EMC Collateral Standard), which includes E-field susceptibility and ESD
requirements. However, even though the device is compliant at the levels of
immunity specified in the standard, certain transmitting devices (cellular phones,
walkie-talkies, cordless phones, paging transmitters, etc.) emit radio frequencies
that could interrupt ventilator operation if located in a range too close to the
ventilator. It is difficult to determine when the field strength of these devices
becomes excessive. Practitioners should be aware that radio frequency emissions
are additive, and that the ventilator must be located a sufficient distance from
transmitting devices to avoid interruption. Do not operate the ventilator in a
magnetic resonance imaging (MRI) environment. The Alarm handling section of
this manual describes possible ventilator alarms and what to do if they occur.
Consult with your institution’s biomedical engineering department in case of
interrupted ventilator operation, and before relocating any life support equipment.

Customer assistance

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

For further assistance contact your local Mallinckrodt representative.

v

vi

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Contents

…………………………………………………………..……….

1 Introduction

1.1 Functionaldescription ………………………………..1-2

1.2 Symbolsandlabels ………………………………….1-8

1.3 Keyboard ………………………………………..1-13

1.3.1 VENTILATORSETTINGS …………………….. 1-14

1.3.2 PATIENTDATA ……………………………. 1-24

1.3.3 VENTILATORSTATUS ……………………….1-28

2 Setting up the ventilator

2.1 Connectingandusinginternalandexternalbatteries …………..2-2

2.2 Connectingtheelectricalsupply…………………………. 2-6

2.3 Connecting the oxygen supply . . . . . . . . …………………… 2-8

2.4 Connectingtheventilatorbreathingcircuit …………………. 2-10

2.5 Installingthecollectorvial ……………………………. 2-13

2.6 Installingtheflexarm ………………………………..2-14

2.7 Installingthehumidifier………………………………. 2-16

2.8 Usingtheventilatorcart ………………………………2-17

3 Getting started

3.1 Poweringuptheventilator ……………………………..3-1

3.2 Selectingventilatorsettings ……………………………. 3-4

3.3 Viewingandchangingalarmsettings………………………3-6

3.4 Enteringandexitingstandbymode ………………………. 3-8

4 Self tests (SST and EST)

4.1 Shortselftest(SST) …………………………………. 4-3

4.2 Extended self test (EST) . . . . . . . . . . . . ………………….. 4-13

5 Once ventilation begins

5.1 Changingsettings:aquickreview ………………………..5-1

5.1.1 Changingsettings……………………………. 5-1

5.1.2 SwitchingbetweenVCV,PCV,andPSV ……………. 5-2

5.1.3 Changingthemode …………………………..5-2

5.2 Viewingandchangingalarmsettings:aquickreview ………….. 5-3

5.3 Adjustingapneaparameters …………………………… 5-4

5.3.1 Adjustingtheapneainterval ……………………..5-6

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

vii

5.4 Viewingpatientdata…………………………………. 5-6

5.5 The 100% O2 and MANUAL INSP keys . . . . . ……………… 5-10

5.6 The EXP PAUSE and INSP PAUSE keys (760 only) …………. 5-11

6 The MENU key

6.1 Moreactivealarms………………………………….. 6-5

6.2 Autoresetalarms …………………………………… 6-6

6.3 Selftests ………………………………………… 6-7

6.4 Usersettings ……………………………………… 6-7

6.4.1 Endotrachealtube …………………………… 6-7

6.4.2 Humidifiertype ……………………………… 6-8

6.4.3 Dateandtimeset ……………………………. 6-8

6.4.4 Apneainterval(Ta)…………………………… 6-9

6.4.5 VCVflowpattern ……………………………. 6-9

6.4.6 Speakingvalvesetup ………………………… 6-10

6.4.7 Alarmvolume ……………………………… 6-16

6.4.8 PCVtimingsetting ………………………….. 6-16

6.4.9 VolumeLEDbar……………………………. 6-16

6.5 Oxygensensor……………………………………. 6-17

6.6 Standby mode . . . . . . . . . . . . …………………………. 6-19

6.7 Batteryinfo ……………………………………… 6-20

6.8 Softwarerevision ………………………………….. 6-20

6.9 Servicesummary ………………………………….. 6-21

6.10 Nebulizer ………………………………………. 6-21

Contents

7 Alarm handling

7.1 Autoresetalarms …………………………………… 7-3

7.2 Alarmsilence ……………………………………… 7-3

7.3 Alarmreset……………………………………….. 7-4

7.4 Clinicalandtechnicalalarms …………………………… 7-5

7.5 Poweralarm……………………………………… 7-17

7.5.1 LossofACPower ………………………….. 7-17

7.5.2 LossofPower …………………………….. 7-18

Appendix A Maintenance

A.1 Cleaning,disinfection,andsterilization …………………….A-2

A.1.1 Cleaning: general guidelines . . . . . . ……………….A-4

A.1.2 Disinfectionandsterilization …………………….A-4

A.2 Preventivemaintenance ………………………………A-5

viii

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Contents

……………………………………………………………..…..

A.2.1 Daily or as required:

inspiratoryandexpiratorybacteriafilters………….. A-8

A.2.2 Dailyorasrequired:collectorvial ……………….. A-8

A.2.3 Dailyorasrequired:in-linewatertraps ……………. A-9

A.2.4 Asnecessary:oxygensensorcalibration…………… A-9
A.2.5 Every 250 hours (or 1 month of use): cooling fan filter . . . A-10
A.2.6 Every 1000 hours (or 3 months of use): air intake filter . . . A-11

A.2.7 Every2years:devicechecks …………………. A-12

A.2.8 Storage ………………………………… A-13

A.2.9 Repacking ………………………………. A-13

Appendix B Part numbers

Appendix C Specifications

C.1 Physical ………………………………………… C-2

C.2 Environmental ……………………………………. C-3

C.3 Power …………………………………………. C-3

C.4 Complianceandapprovals …………………………… C-5

C.5 Technical ……………………………………….. C-5

Appendix D Breath delivery

D.1 A/Cmode ……………………………………….. D-3

D.2 SPONTmode ……………………………………. D-3

D.3 SIMVmode ……………………………………… D-4

D.3.1 Breathtiming ……………………………… D-4

Appendix E Alarm testing

Appendix F Pneumatic schematic

Appendix G Glossary

Index

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

ix

Contents

x

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Figures

…………………………………………………………..……….

Figure 1-1 Block diagram: 700 Series Ventilatorfunction …………1-5

Figure 1-2 740 VentilatorSystemkeyboard………………….1-13

Figure 1-3 760 VentilatorSystemkeyboard………………….1-14

Figure2-1 Liftingtheventilator ………………………….. 2-2

Figure2-2 Internalbatterychargeindicator ………………….. 2-3

Figure2-3 Pluggingtheexternalbatteryintotheventilator ………. 2-5

Figure2-4 Disconnectingtheexternalbattery …………………2-6

Figure2-5 Connectingtheventilatorpowercord ………………. 2-7

Figure2-6 Storingthepowercordontheventilator …………….. 2-7

Figure2-7 Connectingtheoxygensupply …………………… 2-9

Figure2-8 Connectingtheventilatorbreathingcircuit …………..2-12

Figure2-9 Installingthecollectorvial …………………….. 2-13

Figure2-10 Installingtheflexarm …………………………2-14

Figure2-11 Shorteningtheflexarm ………………………. 2-15

Figure2-12 Installingthehumidifier ……………………….2-16

Figure2-13 Lockingandunlockingthecart’sfrontwheels ………..2-17

Figure3-1 Turningthepowerswitchon(the“I”position)…………. 3-2

Figure5-1 Viewingpatientdata …………………………..5-8

Figure 5-2 Volume bar graph (760 Ventilatoronly) …………….. 5-9

Figure 6-1 Using the More active alarms menu function

to view active alarms 6-5

Figure 6-2 Using the Autoreset alarms menu function

to view autoreset alarms 6-6

Figure7-1 Viewingactivealarms …………………………7-2

FigureA-1 Removing/replacingthecollectorvial ……………… A-9

FigureA-2 Coolingfancover ………………………….. A-10

FigureA-3 Airintakefilter …………………………….. A-12

FigureB-1 Ventilatoraccessories ……………………….. B-9

Figure C-1 Recommended ventilator breathing circuit configurations . . C-8

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

xi

Figures

FigureD-1 Flowwaveform …………………………….. D-2

FigureD-2 SIMVbreathperiodintervals …………………… D-4
Figure D-3 Synchronizing breath intervals with patient effort . . . . . . . . . D-5

FigureD-4 SpontaneousbreathsduringSIMV ………………. D-5

FigureD-5 MandatorybreathsduringSIMV…………………. D-5

FigureD-6 ManualinspirationduringSIMV …………………. D-6

xii

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Ta bl e s

…………………………………………………………..……….

Table 1-1 Mode/breath type availability on 740/760 Ventilators ……..1-2
Table 1-2 Changes to current settings in occlusion cycling mode . . . . . . . 1-7
Table 1-3 700 Series Ventilator keyboards: VENTILATOR SETTINGS . 1-15

Table 1-4 700 Series Ventilatorkeyboards:PATIENTDATA ………1-24
Table 1-5 700 Series Ventilator keyboards: VENTILATOR STATUS . . . 1-29

Table 4-1 700 Series Ventilatorselftests …………………….4-2

Table4-2 SSTsequenceoftests…………………………..4-9

Table4-3 OverallSSTresults ……………………………4-12

Table4-4 ESThardwarerequirements ……………………..4-13

Table4-5 ESTsetupmessages………………………….. 4-14

Table4-6 ESTtestsequence…………………………….4-16

Table4-7 KeyfunctionsduringEST ………………………. 4-21

Table4-8 PromptsduringEST……………………………4-22

Table4-9 ESTcompletionstatus …………………………4-23

Table5-1 Breathtypeavailability …………………………..5-2

Table6-1 Menufunctionsummary…………………………. 6-2

Table7-1 Clinicalalarms ……………………………….. 7-5

Table7-2 Technicalalarms ……………………………..7-12

TableA-1 Cleaning,disinfection,andsterilization ……………… A-3

TableA-2 Preventivemaintenanceschedule ………………… A-7

TableB-1 Ventilatoraccessories …………………………. B-2

TableC-1 Physicalspecifications …………………………. C-2

TableC-2 Environmentalspecifications …………………….. C-3

TableC-3 Powerspecifications ………………………….. C-3

TableC-4 Complianceandapprovals ……………………… C-5

TableC-5 Technicalspecifications ………………………… C-5

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

xiii

Tables

xiv

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

SECTION

Introduction 1

1

…………………………………………………………..……….

The 700 Series Ventilator System (including the 740 and 760 Ventilators)
provides respiratory support for a wide range of pediatric to adult patients for a
wide variety of clinical conditions. The ventilator’s mixing technique allows it to
ventilate critically ill patients at adjustable oxygen concentrations without the
need for a blender, compressor, or hospital-grade wall air.

The 700 Series Ventilator System can be mains- or battery-powered. Each
ventilator includes two microcontrollers: one for breath delivery (which controls
ventilation), and one for the user interface (which monitors ventilator and patient
data). Each microcontroller verifies that the other is functioning properly. Using
two independent microcontrollers in this fashion prevents a single fault from
causing a simultaneous failure of controlling and monitoring functions.

The 700 Series Ventilator System supplies mandatory or spontaneous breaths
with a piston-based pneumatic system. Table 1-1 summarizes the modes and
breath types offered by the 740 and 760 Ventilators. Mandatory breaths can be

volume control ventilation (VCV, available on 740 and 760 Ventilators) or
pressure control ventilation (PCV, available on the 760 Ventilator only). VCV

delivers breaths to the patient at a preset tidal volume, peak flow, waveform, and
oxygen concentration at a minimum respiratory rate. PCV delivers breaths to the
patient at a preset inspiratory pressure, I:E ratio or inspiratory time, rise time
factor (how quickly inspiratory pressure rises to achieve the set inspiratory
pressure), and oxygen concentration at a minimum respiratory rate. A
spontaneous breath allows the patient inspiratory flows of up to 300 L/min, with
or without pressure support ventilation (PSV). On the 760 Ventilator, you can set
the rise time factor and exhalation flow sensitivity (that is, the point at which the
ventilator cycles from inspiration to exhalation) in PSV.

The ventilator begins apnea ventilation if no breath (patient- , ventilator-, or
operator-initiated) is delivered within the selected apnea interval. Apnea
ventilation is active during all modes. On the 740 Ventilator, only VCV breaths
are available in apnea ventilation. On the 760 Ventilator, VCV or PCV breaths are
available in apnea ventilation.

The 760 Ventilator also offers the ability to perform respiratory mechanics
calculations and maneuvers as a standard feature using the EXP PAUSE (to

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

1-1

1 Introduction

calculate auto-PEEP) and INSP PAUSE (to calculate patient resistance and
compliance) keys.

Table 1-1: Mode/breath type availability on 740/760 Ventilators

Mode/breath type 740 Ventilator 760 Ventilator

VCV breath type

PCV breath type

PSV breath type (support pressure
setting)

PSV (rise time factor and exhalation
sensitivity settings)

SIMV mode

Apnea ventilation (VCV breath type)

Apnea ventilation (choice of VCV or
PCV breath type)

Respiratory mechanics (EXP PAUSE
and INSP PAUSE)

••

This manual tells you how to operate and perform simple maintenance for the 700
Series Ventilator. Mallinckrodt recommends that you become familiar with this

manual and accompanying labels before attempting to operate or maintain the
ventilator. If you need additional copies of this manual, contact your Mallinckrodt
representative.

To ensure optimum performance of the 700 Series Ventilator System,
Mallinckrodt recommends that a qualified service technician perform periodic
maintenance on the ventilator. For more information, contact your Mallinckrodt
representative.

1.1 Functional description

By pressing keys and turning the knob on the ventilator keyboard, the operator
gives initial instructions and data to the ventilator
(Figure 1-1). The user interface microcontroller processes this information and

1-2

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

stores it in the ventilator’s memory. The breath delivery microcontroller uses this
stored information to control and monitor the flow of gas to and from the patient.

The 700 Series Ventilator uses a flow trigger to recognize patient effort. The
trigger monitors flow from the piston during exhalation. When the patient
inhales, patient circuit pressure drops very slightly below end-expiratory
pressure. At the same time, the piston moves forward to deliver flow to the
ventilator breathing circuit and maintain the preset PEEP/CPAP level. The level
of flow depends on the patient’s effort. If this flow exceeds the user-set level, the
ventilator triggers. By design, the ventilator attempts to maintain PEEP in the
presence of a circuit leak. Since a leak drives the piston to deliver flow to make
up for pressure losses, a circuit leak can require an increase in the flow trigger
level to avoid autocycling.

During exhalation, the ventilator’s piston retracts and draws air and oxygen into
the cylinder. The ventilator uses room air, which means the ventilator can operate
without a compressor or wall air source. Room air enters the ventilator through a
protected user-replaceable air intake filter just inside the ventilator cabinet. This
filter captures airborne particles.

Oxygen from a cylinder or wall supply enters the ventilator through a hose and
oxygen fitting (the fitting is available in several versions). Once inside the
ventilator, the oxygen is regulated to a pressure the ventilator can use, then mixed
with air, according to the selected % O

The flow-triggered piston/cylinder system and motor controller circuit control the
flow of gas to the patient. On the 760 Ventilator in PCV or PSV, the rate of flow
is also determined by the preset rise time factor. This system is designed with a
minute gap (about the size of a thin sheet of paper) between the piston and the
cylinder wall. This design eliminates the friction between the piston and cylinder,
allowing it to respond more rapidly than a “sealed” system.

.

2

A small amount of gas leaks through the gap between the piston and cylinder.
Ventilator software and a continuous forward motion of the piston compensate for
this leak.

The piston delivers the mixed air and oxygen through the inspiratory manifold
system, and out to the patient. The oxygen concentration and temperature of the
delivered gas are monitored here, using a galvanic oxygen sensor and a
thermistor. The galvanic sensor generates a voltage proportional to the partial
pressure of oxygen, from which the oxygen concentration is calculated. The
ventilator alarms if the monitored oxygen concentration is more than ten
percentage points above or below the % O

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

setting. The inspiratory manifold

2

1-3

1 Introduction

system also includes a safety valve to relieve patient pressure if necessary (for
example, if the ventilator breathing circuit is kinked or occluded).

The patient system includes the components external to the ventilator that route
gas between the ventilator and the patient. These components include the
inspiratory filter (which protects against contamination between the ventilator
and patient), a humidification device, ventilator breathing circuit (the tubing
through which the gas travels), collector vial (which protects the exhalation
system from moisture in the exhaled gas, and can be emptied without losing
circuit PEEP), and an expiratory filter (which limits the bacteria in the patient’s
exhaled gas from escaping to room air or contaminating the ventilator).

1-4

700 Series Ventilator Operator’s Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

+

740 Ventilator

Keyboard

Regulator

Piston/cylinder

system

Exhalation/

PEEP/CPAP

system

Filter

Oxygen

Room air

Inspiration
manifold

Inspiratory
filter

Gas intake
system

Expiratory

filter

Humidification
device

Collector

vial

Ventilator breathing
circuit (inspiratory limb)

Patient

Ventilator breathing

circuit (expiratory limb)

7-00017

Figure 1-1. Block diagram: 700 Series Ventilator function

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operator’s Manual

1-5

1 Introduction

The heated exhalation system monitors the flow of the patient’s exhaled gas using
a differential pressure transducer. The patient exhales through the exhalation
valve. During exhalation, the PEEP/CPAP system maintains user-selected
pressure in the ventilator breathing circuit.

Throughout the respiratory cycle, pressure transducers monitor inspiratory,
expiratory, and atmospheric pressures. The temperatures of the pneumatic
compartment and inspiratory gas are also monitored. Information from these
transducers is continuously used to update the calculations that control
ventilation. (Appendix F provides a diagram of the ventilator’s pneumatic system
and ventilator breathing circuit.)

Power to operate the ventilator comes from ac mains (wall) or battery power. The
power supply is designed to protect against excessive voltages, temperatures, or
current draws. A power cord retainer prevents the cord from accidental
disconnection.

The ventilator includes an internal battery, and accommodates an optional
external battery. Depending on the ventilator settings, battery backup power can
be supplied for up to 2 ½
the external battery. Both batteries are recharged during operation from ac power.
If both are installed, the external battery is used first when ac power is not
present. If the external battery is depleted or not installed, the internal battery
supplies power to the ventilator when ac power is not available. The keyboard
indicates the source of power and battery charge level of the internal battery at all
times.

hours using the internal battery, and up to 7 hours using

1-6

Emergency modes: The ventilator declares a ventilator inoperative (VENT INOP)
condition if a hardware failure or critical software error that could compromise

safe ventilation occurs. In case of a ventilator inoperative condition, the VENT
INOP indicator lights and the ventilator enters the safety valve open (SVO) state.
To correct a ventilator inoperative condition, the ventilator must be turned off,
then powered on again; at power-on, the operator must run extended self-test
(EST). The ventilator must pass EST before normal ventilation can resume.

The safety valve allows the patient to breathe room air unassisted when the
ventilator is in the SVO state. The ventilator remains in the SVO state until
power-on self-test (POST) verifies that power levels to the ventilator are
acceptable and that the motor controller and microcontrollers are functioning
correctly, and until the user has confirmed ventilator settings.

If the ventilator enters the SVO state and POST is not running, the SAFETY
VALVE OPEN indicator lights and a high-priority alarm sounds. The ventilator
enters the SVO state if it detects a hardware or software failure that could

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

compromise safe ventilation. In case of a malfunction that prevents software from
opening the safety valve, there is also an analog circuit that opens the safety valve
when system pressure exceeds 115 cmH

If the ventilator detects an occlusion or a continuous high inspiratory pressure
condition, it opens the safety and exhalation valves to vent excess pressure, then
shuts them and begins occlusion cycling mode. In occlusion cycling mode the
ventilator uses current settings except for those summarized in Table 1-2. If the
ventilator again detects an occlusion or continuous high pressure condition, it
again opens the safety and exhalation valves then resumes occlusion cycling
mode. If the operator presses the alarm reset key or the ventilator does not detect
an occlusion or continuous high pressure condition, it reverts to normal
ventilation using the most recently accepted settings.

Table 1-2: Changes to current settings in occlusion cycling mode

Setting Change to setting

O (113 hPa).

2

HIGH PRESSURE
alarm (VCV breath
type)

PEEP (all modes) Set to 0 cmH2O

(all modes) Set to 100%

%O

2

SPONT mode (PSV
breath type)

SUPPORT
PRESSURE (PSV
breath type)

SUPPORT
PRESSURE (PCV
breath type)

RISE TIME FACTOR
(PCV breath type)

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

Setto30cmH

Breaths are delivered at a rate of 12/min with an inspiratory time
of 2 seconds.

If less than 15 cmH
If 15 cmH

If less than 15 cmH
If 15 cmH2O or above: the current setting is used (no change).

Set to 70%

O

2

O:setto15cmH2O.

2

O or above: the current setting is used (no change).

2

O:setto15cmH2O.

2

1-7

1 Introduction

1.2 Symbols and labels

These symbols and labels appear on the 700 Series Ventilator System:

Power switch positions per IEC 601-1.”Irepresents ON position; “O
represents OFF position.

7-00421

Refer to manual per IEC 601-1. When this symbol appears on
product, it means Refer to documentation for information.

7-00418

Potential equalization point, per IEC 601-1

7-00416

External battery connection

7-00426

Circuit breaker

1-8

7-00414

Serial number

SN

ac current

7-00427

Type B equipment, per IEC 601-1

7-00415

Indicates the degree of protection provided by enclosure (drip-proof).

IPX1

7-00403

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Signifies compliance with the Medical Device Directive, 93/42/EEC

7-00412

CSA and NRTL (Nationally Recognized Testing Laboratory)
certification, granted by CSA

8-00417

The TUV Rheinland logo signifies TUV Rheinland Type Test approval
to Annex III of the Medical Device Directive

7-00420

Exhaust port connector

EXHAUST

Inspiratory limb connector

TO

PATI EN T

Expiratory limb connector

FROM

PATI EN T

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

1-9

1 Introduction

Oxygen inlet port label

O

2

.

V

max

61006

Air intake label

61010

Cooling fan label

General life support equipment warning label

1-10

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

61009

61011

Introduction 1

………………………………………………………………….

740 Ventilator back panel label

61012

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

1-11

1 Introduction

760 Ventilator back panel label

1-12

61952

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

1.3 Keyboard

Caution

To avoid damaging the keyboard, do not press on it with sharp objects.

The keyboard (Figure 1-2 shows the 740 keyboard and Figure 1-3 shows the 760
keyboard) is grouped into three sections:

VENTILATOR SETTINGS: Where you set breath delivery variables.

PATIENT DATA: Where you set alarm limits and view the monitored
pressures, breath timing, and volumes.

VENTILATOR STATUS: Where you see the alarm status and operating
condition of the ventilator.

Patient data Ventilator status

Ventilator settings

Figure 1-2. 740 Ventilator System keyboard

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

7-00048

1-13

1 Introduction

Patient data Ventilator status

1-14

Ventilator settings

7-00123

Figure 1-3. 760 Ventilator System keyboard

1.3.1 VENTILATOR SETTINGS

The VENTILATOR SETTINGS section of the keyboard allows you to select the
ventilation mode, breath type, and settings. For more detail on ventilation modes
and breath delivery, see Appendix D.

To change the mode and settings, select the mode, then the breath type, and then
the ventilator settings. The keys flash during setup and mode changes to ensure
that you review all pertinent settings. The keyboard is designed to minimize
accidental or unintentional changes.

Table 1-3 summarizes the functions of the keys, knob, and indicators in the
VENTILATOR SETTINGS section of the keyboard. Ventilator settings are also
limited by these breath delivery boundaries:

I:E ratio £ 4:1 for PCV (760 Ventilator only), £ 3:1 for all other breath types

Inspiratory time = 0.2 to 8 seconds (excluding plateau)

Expiratory time ³ 0.2 seconds

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

PEEP/CPAP + SUPPORT PRESSURE or INSPIRATORY PRESSURE £ 80
cmH

O (80 hPa)

2

NOTE:

Maximum SUPPORT PRESSURE is 70 cmH
INSPIRATORY PRESSURE is 80 cmH

2

SUPPORT PRESSURE or INSPIRATORY PRESSURE + PEEP/CPAP
< HIGH PRESSURE — 2 cmH

O(2hPa)

2

HIGH PRESSURE (in A/C and SIMV modes) > PEEP/CPAP +
7cmH

O(7hPa)

2

HIGH PRESSURE (in SPONT mode) > PEEP/CPAP + SUPPORT
PRESSURE + 2 cmH

O(2hPa)

2

HIGH PRESSURE > LOW INSP PRESSURE

Minute volume £ 50 L/min at an I:E ratio of 2:1

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS

O.

O, maximum

2

Key/indicator Specifies… Range

Mode/breath type settings

A/C Assist/control mode VCV (volume control ventilation)

and PCV (pressure control
ventilation) breath types.

(PCV available on 760 Ventilator
only.)

SIMV Synchronous intermittent

mandatory ventilation mode

SPONT Spontaneous mode PSV breath type

VCV VCV breath type VCV available on 740 and 760

PCV PCV breath type PCV available on 760 Ventilator

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

VCV, PCV (760 only), and PSV
(pressure support ventilation)
breath types.

Ventilators in A/C or SIMV
modes.

only in A/C or SIMV modes.

1-15

1 Introduction

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

PSV PSV breath type PSV available in SIMV or

SPONT modes.

Mandatory (VCV) settings

RESPIRATORY
RATE

TIDAL
VOLUME

PEAK FLOW Maximum flow of gas delivered

PLATEAU (s) Length of inspiratory pause after a

Mandatory (PCV) settings (760 Ventilator only)

RESPIRATORY
RATE

The minimum number of
mandatory breaths the patient
receives per minute. During apnea
ventilation the minimum
RESPIRATORY RATE setting is 6 /
minute.

Volume delivered to the patient
during a mandatory breath,
compliance-compensated and
corrected to body temperature and
pressure, saturated (BTPS).

during a mandatory breath (BTPS).
(Combined with tidal volume, peak
flow defines the active portion of
inspiratory time.)

mandatory breath has been
delivered, during which no gas is
delivered.

The minimum number of
mandatory breaths the patient
receives per minute. During apnea
ventilation the minimum
RESPIRATORY RATE setting is 6 /
minute.

1to70/minute
Accuracy: ± (0.1 + 1%) /minute

40 to 2000 ml
Accuracy: ± (10 ml + 10% of

setting)

3to150L/min
Accuracy: ± (5 + 10% of setting)

L/min

0.0to2.0second
Accuracy: ± 0.05 second

1to70/minute
Accuracy: ± (0.1 + 1%) /minute

INSPIRATORY
PRESSURE
(760 only)

1-16

Pressure above PEEP during the
inspiratory phase of a PCV breath.

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

5to80cmH
Accuracy: ± (3 + 2.5% of setting)

O

cmH

2

O(5to80hPa)

2

Introduction 1

………………………………………………………………….

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

TI/I:E RATIO
(760 only)

RISE TIME
FAC TOR
(760 only)

NOTE:

To help determine the correct setting for inspiratory time, during PCV the message
window displays peak inspiratory flow, end inspiratory flow, and end exhalation flow
in L/min.

YoucanusetheMENUkeyto
select inspiratory time (T
ratio as the breath timing setting for
a PCV breath. You can change the
selected breath timing setting (T
I:E ratio), but the setting remains
constant when you change the
respiratory rate in PCV.

Selecting I:E ratio makes the set
ratio of inspiratory time to
expiratory time for a PCV breath.

Thetimeforinspiratorypressureto
rise from 0 to 95% of the target
pressure level during a PCV
breath.

A setting of 100 = a 100-ms rise
time, and a setting of 5 = 80% of
the inspiratory time or 2500 ms,
whichever is less.

When you adjust this setting, the
message window shows the actual
time (in seconds) to reach 95% of
target pressure.

)orI:E

I

or

I

Inspiratory time (T

0.2 to 8 seconds
Accuracy: ± 0.05 second

I:E ratio:
Accuracy: ± (0.1 + 2%)

5to100

£ 4:1

):

I

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

1-17

1 Introduction

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

Spontaneous (PSV) settings

SUPPORT
PRESSURE

RISE TIME
FAC TOR
(760 only)

Pressure above PEEP maintained
during spontaneous inspiration.
Support pressure is terminated
when inspiratory flow falls to 25%
of peak inspiratory flow, or to the
exhalation sensitivity setting (760
only), or 10 L/min or 25% of peak
flow, whichever is lower (740 only).
Maximum inspiratory time is 3.5
seconds for adults, and 2.5
seconds for pediatric patients.

Thetimeforinspiratorypressureto
rise from 0 to 95% of the target
pressure level during a PSV breath.

A setting of 100 = a 100-ms rise
time, and a setting of 5 = 80% of
the inspiratory time or 1500 ms
(when adult ventilator breathing
circuit is selected) or 600 ms (when
pediatric circuit is selected),
whichever is less.

When you adjust this setting, the
message window shows the actual
time (in seconds) to reach 95% of
target pressure.

0to70cmH
Accuracy: ± (3 + 2.5% of setting)

O

cmH

2

5to100

O(0to70hPa)

2

1-18

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

EXH
SENSITIVITY
(760 only)

Common settings

PEEP/CPAP Positive end expiratory pressure/

The percent of peak expiratory flow
at which the ventilator cycles from
inspiration to exhalation for
spontaneous breaths. The flow at
which the ventilator cycles from
inspiration to exhalation for PSV
breaths. Exhalation begins when
the inspiratory flow is less than the
set value.

To help set EXH SENSITIVITY
appropriately, the peak inspiratory
flow and end inspiratory flow are
displayed in the message window
in PSV. (To ensure accurate
estimates of these flows, it is
important to run SST so that tubing
compliance calculations are
correct.)

continuous positive airway
pressure. Minimum pressure
maintained during inspiratory and
expiratory phases.

1to80%

0to35cmH
Accuracy:

± (2 cmH

O + 4% of setting)

2

O(0to35hPa)

2

TRIGGER
SENSITIVITY
(L/min)

%O

2

Inspiratory flow required to trigger
the ventilator to deliver a breath.

Percentage of inspired oxygen of
the gas delivered to the patient.

1to20L/min

21 to 100%
Accuracy: ± 3% full scale

NOTE:

It may take several minutes for the oxygen percentage to
stabilize.

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

1-19

1 Introduction

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

Other keys, knobs, and indicators

APNEA
PA RA MS k ey

AllowsyoutoselectVCVorPCV(760 Ventilator only) apnea ventilation,
and the apnea ventilation settings. The apnea interval is adjustable from
10 to 60 seconds. Apnea ventilation is available in all modes. For apnea
parameter ranges, see mandatory VCV settings and mandatory PCV
(760 Ventilator only) settings.

MENU Allows you to view active and reset alarms, run SST and EST, adjust

certain settings (including alarm volume, PCV timing, volume bar graph
display, endotracheal tube size, and date and time), access oxygen
sensor functions (calibrate, enable or disable % O

alarm limits, and

2

enable or disable display of oxygen sensor reading), enter standby
mode, view battery information, display software revision, display
service information, and enter EasyNeb nebulizer functions. (Section 6
tells you how to use the menu function.)

100% O

2

Switches the % O2to 100% for 2 minutes, then returns to the current %

setting. The 2-minute interval restarts every time you press 100%

O

2

O

. Once the 100% O2has started, you can press CLEAR to stop the

2

maneuver (unless you have entered a MENU key function or selected a
setting).

MANUAL INSP Delivers one mandatory breath to the patient according to the current

mandatory settings (in A/C or SIMV) or the current apnea parameters (in
SPONT). You can deliver a MANUAL INSP at any time during the
exhalation phase of a breath as long as the exhaled flow is less than
30% of the peak exhaled flow, except during apnea ventilation.

1-20

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

EXP PAUSE
(760 only)

Allows you to measure the patient’s auto-PEEP. An EXP PAUSE
maneuver causes the ventilator to close the exhalation valve at the end
of the expiratory phase, and not deliver the next mandatory breath. At
the end of the maneuver, the message window shows the calculated
value for auto-PEEP (expiratory pressure at the beginning of the
maneuver minus expiratory pressure at the end of the maneuver) and
total PEEP for 30 seconds.

The message window shows the end expiratory flow in L/min at the
beginning of each breath. If there is expiratory flow when the ventilator
delivers the next breath, auto-PEEP is present.

The EXP PAUSE continues as long as you hold down the key, and
should last only until expirator y pressure stabilizes. An EXP PAUSE
maneuver ends when you release the key, the patient initiates a breath,
an alarm occurs, the expiratory phase (including the maneuver) lasts
more than 20 seconds, or the ventilator detects a leak.

Auto-PEEP:
Range: 1 to 35 cmH
Accuracy: ± (1 cmH
The EXP PAUSE maneuver is unavailable when the RESPIRATORY

RATE setting is less than 3 /minute.

O.

2

O + 3% of reading).

2

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

1-21

1 Introduction

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

Other keys, knobs, and indicators (continued)

INSP PAUSE
(760 only)

Allows you to measure the patient’s compliance and resistance. An
extended inspiratory pause also allows you to expand the patients lungs
forupto10seconds.

Pressing INSP PAUSE momentarily causes the ventilator to wait until
the end of the inspiratory phase of the current or next mandatory breath
(in SPONT mode, the ventilator delivers a mandatory breath using the
MANUAL INSP key according to the apnea settings), stop breath
delivery, and keep the exhalation valve closed. The INSP PAUSE
continues until the ventilator detects a stable plateau pressure or 2
seconds have elapsed. An INSP PAUSE maneuver ends when a stable
plateau is reached or an alarm occurs.

Pressing INSP PAUSE for 2 or more seconds after the pause begins
causes the ventilator to deliver an inspiratory pause for as long as you
hold down the key. An extended INSP PAUSE maneuver ends when you
release the key or 10 seconds have elapsed.

You can press CLEAR or release the INSP PAUSE key at any time to
cancel an INSP PAUSE maneuver, and alarms cancel the maneuver.

At the end of the breath, the message window shows the calculated
value for compliance and resistance (if the mandatory breath was a VCV
breath) or compliance (if the mandatory breath was a PCV breath) for 30
seconds.

Compliance:
Range: 1 to 150 mL/H
Accuracy: ± (1 mL/cmH

O.

2

O + 20% of reading).

2

Resistance:
Range: 0 to 150 cmH
Accuracy: ± (3 cmH

O/L/second.

2

O/L/second + 20% of reading).

2

CLEAR Pressing CLEAR before accepting a setting cancels the proposed

setting. Pressing CLEAR does not cancel accepted settings.
Pressing CLEAR twice returns the ventilator to its previous state (unless

you have entered a MENU key function that requires you to press
CLEAR more times).

Pressing CLEAR during a 100% O2maneuver cancels the maneuver.

1-22

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS (continued)

Key/indicator Specifies… Range

Other keys, knobs, and indicators (continued)

ACCEPT Makes changes to settings effective. If you dont press ACCEPT within

30 seconds of proposing a new setting, the user interface returns to its
previous state.

Knob Adjusts the value of a setting or selects a menu option. A setting value

that flashes means that the knob is linked to that setting. Turning the
knob clockwise increases the value, and turning the knob
counterclockwise decreases the value.

CURRENT Lights when the ventilator is operating according to the displayed

settings, or during apnea ventilation. (There is one indicator for
mandatory breaths, and one for spontaneous breaths.)

PROPOSED Lights when you propose a mode or breath type, or you are setting

apnea parameters. Once a proposed setting is accepted, it becomes
effective at the next breath.

APNEA
PA RA MS
indicator

Message
window

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

Lights when apnea ventilation is active.
Lights with PROPOSED indicator when you are setting apnea

parameters, and both indicators turn off once apnea parameters are
accepted.

Shows up to four lines of information (20 characters per line).
First line: Reserved for the highest-priority active or autoreset alarm. On

the 760 Ventilator only, if no alarm is active and the display of the oxygen
sensor reading is enabled, the % O
VCV is the current or proposed breath type and there are no active or
autoreset alarms, shows the selected VCV flow pattern (ramp or
square).

Second line: Information about the menu function or settings, alarm
silence time remaining, or current date and time. During normal
ventilation, shows Flow (L/min).”

Third and fourth lines: Reserved for other messages. For every breath
type, peak and end inspiratory flows are displayed on the third line, and
end expiratory flow is displayed on the fourth line (except that inspiratory
flow is not displayed during VCV breaths or VCV apnea ventilation).

is displayed here. Otherwise, if

2

1-23

1 Introduction

1.3.2 PATIENT DATA

The PATIENT DATA section of the keyboard allows you to view the pressure,
breath timing, and volume of the patient’s breath. You can also view the alarm
settings. A lighted key indicates that a measurement is selected, and its value
appears in the display window. Values are continuously displayed and updated
during ventilation.

Table 1-4 summarizes the functions of the keys and indicators in the PATIENT
DATA section of the keyboard.

Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA

Key/

indicator

Pressure

MEAN
PRESSURE

PEAK
PRESSURE

PLATEAU
PRESSURE

(760 only)

PEEP/CPAP
(760 only)

Function Range

Shows the calculated value of ventilator
breathing circuit pressure over an entire
respiratory cycle. Updated at the
beginning of each breath.

Shows the maximum pressure measured
during inspiration. Updated at the
beginning of each expiratory phase.
(Default pressure display.)

Shows the pressure measured at the end
of the plateau period of a mandatory
inspiration (whether the inspiration is in a
regular VCV breath or is part of an
inspiratory pause maneuver). Updated at
the beginning of each expiratory phase.
The PRESSURE display shows a blank if
the ventilator does not detect a stable
plateau pressure.

Shows the pressure measured at the
expiratory limb before any inspiratory
effort. Updated at the beginning of each
inspiratory phase.

0to99cmH
Accuracy:

± (3 + 4% of reading) cmH2O

0to140cmH
Accuracy:

± (3 + 4% of reading) cmH

0to140cmH
Accuracy:

± (3 + 4% of reading) cmH

0to140cmH
Accuracy:

± (3 + 4% of reading) cmH

O(0to99hPa)

2

O(0to140hPa)

2

O(0to140hPa)

2

O(0to140hPa)

2

2

2

2

O

O

O

1-24

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)

Key/

indicator

Breath timing

RATE (/min) Shows the calculated value of the total

respiratory rate, based on the previous 60
seconds or 8 breaths (whichever interval
is shorter). Updated at the beginning of
each breath. (Default breath timing
display.)

The calculation is reset (and display is
blank) when ventilation starts, when
apnea ventilation starts or autoresets,
when you change the mode, breath type,
or RESPIRATORY RATE setting, and
whenyoupressthealarmresetkey.

I:E RATIO Shows the ratio of measured inspiratory

time to measured expiratory time.
Updated at the beginning of each breath.

INSP
TIME(s)
(760 only)

The measured inspiratory time, including
breaths that are truncated due to a HIGH
PRESSURE alarm. Updated at the
beginning of each expiratory phase.

Function Range

1 to 199 /minute
Accuracy:

± (0.1 +1% of reading)/minute

1:99.9 to 9.9:1
Accuracy:

± (0.1 + 2%)

0.1 to 9.90 seconds
Accuracy:

± 0.05 seconds

Vol ume

EXHALED
VOLUME
(ml)

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

Shows the patients measured expiratory
tidal volume averaged over the last 5
breaths (for A/C VCV breaths, ventilator­initiated PCV breaths, and PCV apnea
breaths) or for the just-completed breath
(for all other breaths). Corrected to BTPS
and compliance-compensated. Updated
at the beginning of each inspiration.
(Default volume display.)

The calculation is reset when ventilation
starts, when apnea ventilation starts or
autoresets, when you change the mode or
breath type, and when you press the
alarm reset key.

0to9L
Accuracy:

± (10 ml + 10% of reading)

1-25

1 Introduction

Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)

Key/

indicator

Volume (continued)

TOTAL
MINUTE
VOLUME (L)

DELIVERED
VOLUME
(ml)
(760 only)

SPONT
MINUTE
VOLUME (L)
(760 only)

Shows the patients measured expiratory
minute volume, based on the previous 60
seconds or 8 breaths (whichever interval
is shorter). Updated at the beginning of
each breath.

The calculation is reset when ventilation
starts, when apnea ventilation starts or
autoresets, when you change the mode or
breath type, and when you press the
alarm reset key.

Shows the measured inspiratory tidal
volume for the just-completed PCV or PSV
breath. Corrected to BTPS and
compliance-compensated. Updated at the
beginning of each inspiration for PCV and
PSV breath types.

Shows the patients measured expiratory
minute volume for all spontaneous
breaths, based on the previous 60
seconds or 8 breaths (whichever interval
is shorter). Updated at the beginning of
each breath.

The calculation is reset when ventilation
starts, when apnea ventilation starts or
autoresets, when you change the mode or
breath type, and when you press the
alarm reset key.

Function Range

0to99L
Accuracy:

± (10 ml + 10% of reading)

0 to 3000 ml
Accuracy:

± (10 ml + 10% of reading)

0to99L
Accuracy:

± (10 ml + 10% of reading)

1-26

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Introduction 1

………………………………………………………………….

Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)

Key/

indicator

Alarm settings

HIGH RATE An active alarm indicates that measured

respiratory rate is higher than the alarm
setting.

HIGH TIDAL
VOLUME

LOW INSP
PRESSURE

LOW TIDAL
VOLUME

HIGH
PRESSURE

An active alarm indicates that exhaled
volume for three out of four consecutive
breaths was above the alarm setting.

An active alarm indicates that monitored
circuit pressure is below the alarm setting
at the end of inspiration. Inactive in for any
spontaneous breath.

An active alarm indicates that delivered
volume for three out of four consecutive
breaths were below the alarm setting. (If
this alarm is set to 0 ml and breath type is
PCV, an active alarm indicates that
delivered volume is less than 3 ml for three
out of four consecutive breaths.)

An active alarm indicates that two
consecutive breaths were truncated
because circuit pressure reached the
alarm setting.

Function Range

3 to 100 /minute
Accuracy:

± (0.1 +1% of setting)/minute

20 to 6000 ml
Accuracy:

± (10 ml + 10% of setting)

3to60cmH
Accuracy:

± (1 + 3% of setting)

0 to 2000 ml
Accuracy:

± (10 ml + 10% of setting)

10 to 90 cmH
(10to90hPa)

Accuracy:
± (1 + 3% of setting)

O(3to60hPa)

2

O

2

LOW
MINUTE
VOLUME

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

An active alarm indicates that monitored
minute volume is less than the alarm
setting, based on an eight-breath running
average or the previous minute, whichever
is less.

0to50L
Accuracy:

± (10 ml + 10% of setting)

1-27

1 Introduction

Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA (continued)

Key/

indicator

Other indicators

Pressure bar
graph

Volume bar
graph
(760 only)

MAND Lights at the start of each breath to

Shows real-time pressures in centimeters
of water (cmH2O) or hectopascals (hPa).

LEDs show the current HIGH PRESSURE
alarm setting and the peak pressure of the
last breath during exhalation.

Shows real-time exhaled volume in
milliliters (ml). Volumes are compliance­compensated and corrected to BTPS.

The active scale is determined by the
HIGH TIDAL VOLUME alarm setting.
LEDs show the current HIGH TIDAL
VOLUME and LOW TIDAL VOLUME
alarm settings. You can use the MENU key
to enable or disable the volume bar graph.
During exhalation, LEDs show the
maximum exhaled volume of the last
breath.

indicate a ventilator- or operator-initiated
(time or manually triggered) mandatory
breath is being delivered.

Function Range

-10to90cmH
(-10to90hPa)

Resolution: 1 cmH2O(1hPa)

If HIGH TIDAL VOLUME
setting < 500 ml: 0 to 500 ml

Resolution: 5 ml

If HIGH TIDAL VOLUME
setting >

Resolution: 20 ml

Not applicable

O

2

500ml:0to2000ml

ASSIST Lights at the start of each breath to

indicate a patient-initiated mandatory (flow
triggered) breath is being delivered.

SPONT Lights at the start of each breath to

indicate a patient-initiated spontaneous
(flow triggered) breath is being delivered.

1.3.3 VENTILATOR STATUS

The VENTILATOR STATUS section of the keyboard shows the operating
condition of the ventilator, and is continuously updated during operation. Table

1-28

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Not applicable

Not applicable

Introduction 1

………………………………………………………………….

1-5 summarizes the functions of the keys and indicators in the VENTILATOR
STATUS section of the keyboard.

Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS

Key/indicator Color Function

ALARM Red

(high priority)

CAUTION Yellow

(medium priority)

NORMAL Green Lights when no alarm condition is present.

VENT INOP Red

(high priority)

SAFETY VALVE
OPEN

Red
(high priority)

Flashes when a high-priority alarm is active. A
repeating sequence of three, then two beeps
sounds. Lights steadily when a high-priority
alarm has been autoreset.

Flashes when a medium-priority alarm is active.
A repeating sequence of three beeps sounds.
Lights steadily when a medium-priority alarm
has been autoreset.

Lights to indicate that the ventilator is
inoperative, and the ventilator safety valve is
open. A qualified service technician must run
and pass the extended self-test (EST) before
normal ventilation can resume.

If the condition that caused the safety valve to
open no longer exists, and the VENT INOP
indicator is off, press the alarm reset key to
resume ventilation.

Lights when the ventilators safety valve and
exhalation valve open and only room air is
available to the patient.

Can indicate that the ventilator is inoperative, or
there is an occlusion in the ventilator breathing
circuit. If possible, the message window shows
the alarm that triggered the safety valve open
condition and how much time has elapsed since
the last breath was triggered.

ON AC/
BATTERY
CHARGING

ON INTERNAL
BATTERY

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

Green Lights when the ventilator is running on ac

power and the battery is charging.

Yellow Flashes when the ventilator is running on the

internal battery.

1-29

1 Introduction

Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS (continued)

Key/indicator Color Function

ON EXTERNAL
BATTERY

INTERNAL
BATTERY
LEVEL

7-00423

7-00424

Yellow Flashes when the ventilator is running on the

external battery.

Green Shows the relative charge level of the internal

battery.

Ye l l o w Alarm silence: Silences the alarm sound for 2

minutes from the most recent key press.

Not applicable Alarm reset: Reestablishes all alarm indicators,

cancels the alarm silence period, and resets the
patient data displays. If the condition that
caused the alarm still exists, the alarm
reactivates. Cancels apnea ventilation, if active.
Reestablishes previous settings and ventilation
resumes, unless the ventilator is inoperative.

1-30

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

SECTION

Setting up the ventilator 2

2

…………………………………………………………..……….

This section tells you how to set up the ventilator, including:

Connecting and using the internal and external batteries

Connecting the electrical supply

Connecting the oxygen supply

Connecting the ventilator breathing circuit

Installing the collector vial

Installing the flex arm

Installing the humidifier

Using the ventilator cart

Warning

To avoid tipping or damaging the ventilator, do not stack other
equipment on the ventilator. 700 Series Ventilators are designed
to be mounted on either a cart or a shelf by a qualified service
technician. When lifting the ventilator, lift from the base, and use
assistance and appropriate safety precautions. Figure 2-1 shows
proper lifting technique.

To avoid the possibility of injury to the patient and to ensure
proper ventilator operation, do not attach any device to the port
labeled EXHAUSTunless the device is specifically authorized
by Mallinckrodt.

To minimize the increased risk of fire due to an oxygen-enriched
environment, do not use the ventilator in a hyperbaric chamber.

To avoid raising the oxygen concentration of room air, use the
ventilator in an adequately ventilated room.

Caution

Do not obstruct the cooling fan

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

.

2-1

2 Setting up the ventilator

Use two people to lift

Lift ventilator from base only

7-00053

Figure 2-1. Lifting the ventilator

2.1 Connecting and using internal and external batteries

Warning

A Mallinckrodt battery must always be installed in the ventilator.
Without a battery, the ventilator is not protected against low or lost
ac power. Do not use the ventilator unless a battery with at least
minimal charge is installed.

2-2

NOTE:

If the ventilator has been stored for an extended period, allow it to
acclimate to its environment before turning it on. This helps ensure
that the ventilator powers up correctly.

Every 700 Series Ventilator System includes an internal battery as a backup
power supply that comes standard with the ventilator. The internal battery can
provide up to 2

½ hours of backup power, depending on ventilator settings and

battery charge level. The ventilator operates on backup power when ac power is
lost or drops below minimum. The internal battery’s charge level is continuously

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Setting up the ventilator 2

…………………………………………………………………….

indicated on the keyboard (see Figure 2-2). The ventilator alarms when it
determines that only 5 minutes of power remain using the current settings.

Internal battery level

7-00054

Figure 2-2. Internal battery charge indicator

The optional external battery supplies up to 7 additional hours of backup power in
case ac power is lost or falls below minimum. (See Appendix B for external
battery ordering information.) If installed, the external battery is the first source
of backup power. The ventilator uses the internal battery if the external battery is
depleted or is not connected.

NOTE:

When the ventilator switches off because battery charge is
inadequate, turn off the power switch to prevent the battery from
being fully depleted. Connect the ventilator to ac power as soon as
possible.

The ventilator charges the internal and external batteries during ac power
operation, or in standby mode when attached to ac power. (See Appendix C for
battery specifications.) When the ventilator is operating on battery, you can use
the MENU key to check the estimated operational time remaining until recharge
is required for the internal and external batteries. (Section 6 tells you how to use
the MENU key.) The ventilator must operate for 2 minutes before it can estimate
battery time remaining.

You can disconnect and connect the external battery during normal operation.
When operating the ventilator on battery power, you can install the external

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

2-3

2 Setting up the ventilator

battery on the cart for easy transport. Follow these steps to connect the external
battery:

1. Mount the external battery on the base of the cart (or other suitable
location).

2. Plug the external battery’s cable into the connector (Figure 2-3).

3. Figure 2-4 shows you how to disconnect the external battery.

To ensure that the batteries retain their charge, store the ventilator in standby
mode, with its power switch on, and connected to ac power. If the ventilator is not
stored in this manner, check battery charge levels before using the ventilator.

Caution

If you plan to store the ventilator for more than 6 months, remove
batteries before storage. Replace the internal battery before using
the ventilator again. A qualified service technician must replace the
battery according to the instructions in the 700 Series Ventilator
System Service Manual.

If you turn on the ventilator after it has been unplugged for an extended period,
the LOW EXT BATTERY and LOW INT BATTERY alarms may become active.
If so, recharge the internal battery by leaving the ventilator plugged in and turned
on in standby mode (with no patient connected) for up to 8 hours.

2-4

If the LOW EXT BATTERY, LOW INT BATTERY, or BAT NOT CHARGING
alarm is still active after 8 hours, replace the batteries. If you turn off the
ventilator when the battery is absent or inadequately charged, a LOSS OF
POWER alarm sounds for at least 2 minutes.

NOTE:

The ON BATTERY indicator flashes when the ventilator is operating
on battery power.

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Setting up the ventilator 2

…………………………………………………………………….

External

battery

connector

External

battery

7-00055

Figure 2-3. Plugging the external battery into the ventilator

G-061874-00 Rev. D (09/00) 700 Series Ventilator Operators Manual

2-5

2 Setting up the ventilator

7-00051

Figure 2-4. Disconnecting the external battery

2.2 Connecting the electrical supply

Warning

To avoid electrical shock hazard, connect the ventilator power
cord into a grounded ac power outlet. If the integrity of the ac
ground is in question, operate the ventilator from the internal or
external battery.

If used in the U.S., connect the ventilator to an ac receptacle
marked Hospital Onlyor Hospital Gradeto ensure grounding
reliability.

2-6

Every 700 Series Ventilator System is supplied with a power cord. A power cord
retainer covers the connector and socket to protect against liquid spills or
accidental disconnection.

When the ventilator is ready for operation on ac power, connect the power cord to
ac power (see Figure 2-5).

700 Series Ventilator Operators Manual G-061874-00 Rev. D (09/00)

Loading…

  • About
  • Blog
  • Projects
  • Help
  • Donate

    Donate icon
    An illustration of a heart shape

  • Contact
  • Jobs
  • Volunteer
  • People

Bookreader Item Preview

texts

Puritan Bennett 840 Service Manual

Puritan Bennett 840 Service Manual

Addeddate
2020-05-19 23:46:40
Classification
Clinical;Ventilator;Puritan Bennett Ventilator;Puritan Bennett 840
Identifier
manual_Puritan_Bennett_840_Service_Manual
Identifier-ark
ark:/13960/t2s55fd77
Ocr
ABBYY FineReader 11.0 (Extended OCR)
Ppi
300
Scanner
Internet Archive Python library 1.9.0

comment

Reviews

There are no reviews yet. Be the first one to
write a review.

SIMILAR ITEMS (based on metadata)

В наличии документация на медицинское оборудование фирмы Nellcor Puritan Bennett (в т.ч. на русском языке):

Modification: 17 September 2023

Руководство оператора (Operators Guide) на 700 Series (Rev. A 09.1998) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Руководство оператора Operators Guide на 700 Series (Rev. A 09.1998) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 700 Series (Rev. A 11.1999) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 700 Series (Rev. A 11.1999) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 700, 740, 760 Rev. A 1999 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 700, 740, 760 Rev. A 1999 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 700, 740, 760 Rev. B 2009 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 700, 740, 760 Rev. B 2009 [Nellcor Puritan Bennett]

Инструкция по эксплуатации (Operation (Instruction) manual) на 7200 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Инструкция по эксплуатации Operation (Instruction) manual на 7200 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 7200 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 7200 [Nellcor Puritan Bennett]

Руководство оператора (Operators Guide) на 740, 760 Дополнение: коммуникационный блок [Nellcor Puritan Bennett] ИВЛ-Анестезия

Руководство оператора Operators Guide на 740, 760 Дополнение: коммуникационный блок [Nellcor Puritan Bennett]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на 840 (800) (Rev. С 2014-09) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Техническое описание, инструкция по эксплуат. Technical description, instructions на 840 (800) (Rev. С 2014-09) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 840 (Rev. A August, 2003) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 840 (Rev. A August, 2003) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 840 (Rev. A October 1998) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 840 (Rev. A October 1998) [Nellcor Puritan Bennett]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на 840 (Rev. C November 1998) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Техническое описание, инструкция по эксплуат. Technical description, instructions на 840 (Rev. C November 1998) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 840 (Rev. E June 2007) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 840 (Rev. E June 2007) [Nellcor Puritan Bennett]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на 840 (Rev. G Oktober 2006) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Техническое описание, инструкция по эксплуат. Technical description, instructions на 840 (Rev. G Oktober 2006) [Nellcor Puritan Bennett]

Техническое описание, инструкция по эксплуат. (Technical description, instructions) на 840 (Rev. H September 2009) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Техническое описание, инструкция по эксплуат. Technical description, instructions на 840 (Rev. H September 2009) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 840 (Rev. С June 2005) [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 840 (Rev. С June 2005) [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на 840 General information [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на 840 General information [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Achieva [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на Achieva [Nellcor Puritan Bennett]

Техническая документация (Technical Documentation/Manual) на Achieva [Nellcor Puritan Bennett] ИВЛ-Анестезия

Техническая документация Technical Documentation/Manual на Achieva [Nellcor Puritan Bennett]

Инструкция пользователя (User manual) на Achieva [Nellcor Puritan Bennett] ИВЛ-Анестезия

Инструкция пользователя User manual на Achieva [Nellcor Puritan Bennett]

Инструкция пользователя (User manual) на Achieva [Nellcor Puritan Bennett] ИВЛ-Анестезия

Инструкция пользователя User manual на Achieva [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Infant Star model 100 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на Infant Star model 100 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на LP6 Plus, LP10, LP20 [Nellcor Puritan Bennett] ИВЛ-Анестезия

Сервисная инструкция Service manual на LP6 Plus, LP10, LP20 [Nellcor Puritan Bennett]

Инструкция оператора (Operator manual) на N-4000C [Nellcor Puritan Bennett] Мониторы

Инструкция оператора (Operator manual) на NPB-4000, 4000C [Nellcor Puritan Bennett] Мониторы

Сервисная инструкция (Service manual) на NPB-4000, 4000C (Mallinckrodt) [Nellcor Puritan Bennett] Мониторы

Схема электрическая (Electric scheme (circuit)) на Датчик сатурации кислорода DS-100A фирмы Nellcor [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция (Service manual) на Капнограф-пульсоксиметр NPB-40 (2001, 55стр.) [Nellcor Puritan Bennett] Диагностика

Руководство оператора (Operators Guide) на Капнограф-пульсоксиметр NPB-75 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция (Service manual) на Капнограф-пульсоксиметр OxiMax NPB-40 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция (Service manual) на Капнограф-пульсоксиметр OxiMax NPB-75 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция (Service manual) на Пульсоксиметр N-395 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр N-395 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Пульсоксиметр NPB-190 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр NPB-190 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Пульсоксиметр NPB-195 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр NPB-195 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Пульсоксиметр NPB-290 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр NPB-290 [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Пульсоксиметр NPB-295 [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр NPB-295 [Nellcor Puritan Bennett]

Инструкция оператора (Operator manual) на Пульсоксиметр OxiMax N-550 [Nellcor Puritan Bennett] Диагностика

Инструкция оператора Operator manual на Пульсоксиметр OxiMax N-550 [Nellcor Puritan Bennett]

Инструкция по эксплуатации (Operation (Instruction) manual) на Пульсоксиметр OxiMax N-595 [Nellcor Puritan Bennett] Диагностика

Инструкция по эксплуатации Operation (Instruction) manual на Пульсоксиметр OxiMax N-595 [Nellcor Puritan Bennett]

Инструкция по эксплуатации (Operation (Instruction) manual) на Пульсоксиметр OxiMax N-600x [Nellcor Puritan Bennett] Диагностика

Инструкция по эксплуатации Operation (Instruction) manual на Пульсоксиметр OxiMax N-600x [Nellcor Puritan Bennett]

Сервисная инструкция (Service manual) на Пульсоксиметр OxiMax N-600x [Nellcor Puritan Bennett] Диагностика

Сервисная инструкция Service manual на Пульсоксиметр OxiMax N-600x [Nellcor Puritan Bennett]

Инструкция по эксплуатации (Operation (Instruction) manual) на Система обогрева растворов WarmFlo Модель FW-588 [Nellcor Puritan Bennett] Разное


Архив технической документации медицинского оборудования © 2006 … 2023


3.0

Rated 3 out of 5

3 out of 5 stars (based on 1 review)

Your overall rating

COVIDEN PURITAN BENNETT 980 (01) PDF MANUAL

Click here to download COVIDEN PURITAN BENNETT 980 (01) PDF MANUAL


COVIDEN PURITAN BENNETT 980 (01) PDF MANUAL

FREE ENGLISH PDF

OPERATING INSTRUCTIONS

USER GUIDE – USER MANUAL

OWNER GUIDE – OWNER MANUAL

REFERENCE GUIDE – REFERENCE MANUAL

INSTRUCTION GUIDE – INSTRUCTION MANUAL


Your overall rating

  • Twitter
  • YouTube
  • Reddit
  • Facebook
  • Instagram
  • Pinterest

COVIDEN PURITAN BENNETT 980 (01) PDF MANUAL

COVIDEN PURITAN BENNETT 980 (01) PDF MANUAL

Понравилась статья? Поделить с друзьями:
  • Ветбицилин 5 инструкция по применению для коров
  • Руководство по ремонту маз 5440а8
  • Лактоферон для кошек инструкция по применению цена
  • Kenwood hdv 810 инструкция на русском
  • Ман d2066 мануал по ремонту двс