3M™ GMI™ PS200 PORTABLE GAS MONITOR

OPERATIONS & MAINTENANCE MANUAL

64171, Rev. 10

WARNING: ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY FOR

USING, MAINTAINING, OR SERVICING THIS PRODUCT MUST READ THIS

ENTIRE MANUAL CAREFULLY. FAILURE TO USE THIS EQUIPMENT

PROPERLY COULD RESULT IN SERIOUS INJURY OR DEATH.

64171, Rev. 10/May 2019

loading

Related Manuals for 3M GMI PS200 Series

Summary of Contents for 3M GMI PS200 Series

  • Page 1
    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATIONS & MAINTENANCE MANUAL 64171, Rev. 10 WARNING: ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY FOR USING, MAINTAINING, OR SERVICING THIS PRODUCT MUST READ THIS ENTIRE MANUAL CAREFULLY. FAILURE TO USE THIS EQUIPMENT PROPERLY COULD RESULT IN SERIOUS INJURY OR DEATH.
  • Page 2
    Reformatted and revised entire manual. Copyright © 3M™ GMI™, May 2019 The information contained within is for use only with the 3M™ GMI™ PS200 Portable Gas Moni- tor. Reproduction, in whole or in part, including utilization in machines capable of reproduction or retrieval without written permission of 3M™…
  • Page 3
    GMI™ PS200 Portable Gas Monitor (or “the monitor”). LIABILITY Every care has been taken in the preparation of this O&M manual, but 3M™ and GMI™ do not accept any responsibility for errors or omissions and their consequences. Information in this O&M manual is subject to change without notice.
  • Page 4
    3M™ GMI™ PS200 PORTABLE GAS MONITOR This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 5: Table Of Contents

    3M™ GMI™ PS200 PORTABLE GAS MONITOR TABLE OF CONTENTS TABLE OF CONTENTS Sect. # Section Title Page # ABOUT THIS GUIDE ……………………..1-1 1.1. Guide Conventions ……………………… 1-1 1.2. Certifications and Approvals………………….1-2 1.3. General Safety Information………………….1-4 1.4. Additional Safety Details — CSA ONLY ………………

  • Page 6
    3M™ GMI™ PS200 PORTABLE GAS MONITOR TABLE OF CONTENTS Sect. # Section Title Page # 6.4. Quick/Full Bump Test ……………………6-3 6.5. Confirm Alarms ……………………..6-4 RECHARGING THE BATTERIES ………………… 7-1 7.1. General…………………………7-1 7.2. Recharge Monitor Using the Charging/Comms Clip …………7-1 OPERATOR MAINTENANCE ………………….8-1…
  • Page 7
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF FIGURES LIST OF FIGURES Fig. # Figure Title Page # PS200 Portable Gas Detector ………………….2-1 Display Example (4-Gas) …………………..2-2 Display Examples ……………………..2-2 OK Display Example (4-Gas) ………………….2-2 Function Buttons ……………………..3-1 Quick Configuration Display ………………….3-2…
  • Page 8
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF FIGURES Fig. # Figure Title Page # 5-12 Calibration Required……………………5-5 BUMP TEST Confirmation Screen ………………..6-1 Apply Gas Screen (4-Gas Model) ………………..6-2 Regulator Valve Choice …………………….6-2 Bump Test Kit……………………….6-3 Confirm Alarms ……………………..6-4 Bump Test Pass ……………………..6-4…
  • Page 9
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF TABLES LIST OF TABLES Table # Table Title Page # Certifications and Approvals………………….1-2 10-1 Supplied Accessories……………………10-1 10-2 Optional Accessories ……………………10-1 10-3 Charging Accessories……………………10-1 10-4 Manual Bump & Field Calibration Kit ………………10-1 10-5 Automatic Bump &…
  • Page 10
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF TABLES This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 11: About This Guide

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE 1. ABOUT THIS GUIDE This guide instructs gas detection personnel on the features and usage of the PS200 Portable Gas Monitor (or “the monitor”), including information on operation, configuration, maintenance, specifi- cations and troubleshooting.

  • Page 12: Certifications And Approvals

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE 1.2. Certifications and Approvals The monitor has the following approvals available: Table 1-1: Certifications and Approvals Mark II 2G Ex ia d IIC T4 Gb (Ta = -20ºC to +50ºC) ATEX IECEx Ex ia d IIC T4 Gb (Ta = -20ºC to +50ºC) or…

  • Page 13
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE Table 1-1: Certifications and Approvals Mark TP TC 012/2011 TP TC 020/2 1.2.1. LABEL Always refer to the rear label for the actual certification of your monitor (see Figure 1-1: Identification…
  • Page 14: General Safety Information

    INSTRUCTIONS MAY ALSO RESULT IN SERIOUS INJURY OR DEATH. 3M™ GMI™ can take no responsibility for use of its equipment if it is not used in accordance with these instructions. If further operational or maintenance details are required but not provided in this guide, contact 3M™…

  • Page 15
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE ATTENTION: TOUT LECTURE DÈPASSANT RAPIDEMENT LE MAXIMUM DE L’ÊCHELLE ET SUIVIE PAR UNE DIMINUTION OU UNE LECTURE ERRONÉE, INDIQUE UNE CONCETRATION DE GAZ SUPÉRIEURE À LA VALEUR MAXIMALE DE CETTE ÉCHELLE. CETTE LECTURE N’EST PASSIGNIFICATIVE.
  • Page 16
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 17: Introduction

    2. INTRODUCTION 2.1. General Description 3M GMI PS200 Portable Gas Monitor series combines quality, ruggedness and advanced technol- ogy in a user friendly, portable gas detector. It is compact, lightweight, water resistant, extremely robust, and is suitably certified to recognized International Standards.

  • Page 18: Display Example (4-Gas)

    3M™ GMI™ PS200 PORTABLE GAS MONITOR INTRODUCTION All gases being monitored are displayed. A 4-gas monitor display is illustrated in Figure 2-2: Display Example (4-Gas) Figure 2-2: Display Example (4-Gas) NOTE: Character size increases if fewer sensors are configured, as illustrated in…

  • Page 19: Features

    2.3.1. Accessing Logged Data Download data from the monitor to a PC/Laptop using the standard charging/comms cable and additional software. Contact the 3M GMI Sales Department for further details. 2.4. Filters Filters protect the monitor from water and dust ingress. Check filters regularly and replace as neces- sary (refer to Section 8.2.

  • Page 20
    3M™ GMI™ PS200 PORTABLE GAS MONITOR INTRODUCTION This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 21: Operation

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 3. OPERATION Figure 3-1: Function Buttons 3.1. Operating Procedure CAUTION: The monitor can be supplied with a flammable gas sensor. This sensor is designed for use in concentrations of gas not exceeding the Lower Explosive Limit (LEL).

  • Page 22: Quick Configuration Display

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Additional: • Perform regular leak checks on pumped instruments by placing your thumb over the sample inlet nozzle and ensuring the monitor displays FLOW FAULT. • Perform regular bump tests. 3.2. Quick Configuration Display…

  • Page 23: Flash Screen

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-3: Flash Screen 3.3.1. MONITOR IDENTIFICATION During warm-up, the LCD identifies the serial number, software version, battery status, and calibra- tion gas information, as illustrated in Figure 3-4: Monitor Identification Display Figure 3-4: Monitor Identification Display 3.3.2.

  • Page 24: Date And Time

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-6: Date and Time 3.3.4. BUMP DUE DATE (OPTION) This configurable option is disabled by default but can be configured to: • indicate briefly that bump test is due, as illustrated in Figure 3-7: Bump Due Date •…

  • Page 25: Calibration Due Date

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 3.3.5. CALIBRATION DUE DATE The calibration due date feature (shown in ) is configurable, including Figure 3-9: Calibration Due Date the factory-set option to display the number of days remaining until the next calibration is due.

  • Page 26: Cal Gas Selection

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-11: Cal Gas Selection When this option is displayed (see ), the original gas used to calibrate the Figure 3-11: Cal Gas Selection monitor is identified between two arrowheads. NOTE: The calibration certificate also identifies the calibration gas type used.

  • Page 27: Switch The Display Backlight On/Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-13: 4-Gas Normal Operating Display 3.4. Switch the Display Backlight On/Off The display backlight can be manually switched ON when working in poor lighting conditions. Press the R button once to switch ON. After 20 seconds it will automatically switch OFF.

  • Page 28: Manual Datalog

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 6. MAX/MIN readings can be reset by pressing and holding the R button for 2 seconds when either of the MAX/MIN screens are displayed. After resetting, the monitor will return to the normal operating screen.

  • Page 29: Self Test

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-16: Sample Line Connection 3.8.1. PUMP OPERATION Press and hold the R button to start or stop the pump. When the pump is running, a pump symbol (as highlighted in ) rotates in Figure 3-17: Pump Symbol the display.

  • Page 30: Switch The Monitor Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-18: Self Test The current username, if configured, will appear at the top of the display. 3.10. Switch the monitor OFF Press and hold both the L and R buttons to switch the monitor OFF.

  • Page 31: Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS 4. ALARMS 4.1. Gas Alarms When an alarm set-point has been reached, the audible, visual, and vibrating alarms will activate to alert the users. All gas alarms are configurable to meet the specific needs of users and only operate once the monitor warm-up is complete.

  • Page 32: Acknowledge Gas Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS Figure 4-2: LEL HIHI Alarm 4.2. Acknowledge Gas Alarms After the gas readings have returned to a safe level, press and hold the R button to clear all alarms. 4.3. Mute Alarms If configured, muting the alarm will silence the audible alarm for 60 seconds, as follows: •…

  • Page 33: Get Out/High Gas Timer

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS Figure 4-3: Over Range Alarm 4.4.1. 10 SECOND SWITCH OFF The monitor must be returned to clean air. NOTE: To avoid accidental switch-off in this dangerous state, the off cycle is increased to 10 seconds.

  • Page 34
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 35: Warnings & Faults

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS 5. WARNINGS & FAULTS 5.1. Low Battery LOW BATTERY flashes when approximately 30 minutes of run time remains, as shown in Figure 5-1: . The display turns red, the audible alarm sounds once every two seconds, and LOW BATTERY Warning the red LED’s flash.

  • Page 36: Lel Zero Fail

    Figure 5-4: LEL ZERO FAULT Return the monitor to clean air and switch it OFF then ON. If the ZERO FAULT persists, return the monitor to a 3M GMI approved service/repair facility. NOTE: The monitor can still detect and alarm on all other sensors.

  • Page 37: Sensor Faults

    S, as illustrated in , allow Figure 5-7: CO/H S SENSOR FAULT the monitor to run in clean air for 20 minutes. If the fault persists, return the monitor to a 3M GMI approved service/repair facility. Figure 5-7: CO/H S SENSOR FAULT 5.5.

  • Page 38: Memory Fault

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS Figure 5-8: FLOW FAULT 5.6. Memory Fault During warm-up, if the ‘MEMORY FAULT’ screen is displayed, as illustrated in Figure 5-9: Memory , the monitor has detected a memory fault. Fault Figure 5-9: Memory Fault The monitor must be returned for service.

  • Page 39: Cal Expired

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS Alternately, press the R button once to acknowledge the warning, cancel the audible/visual alarm and continue. Every 30 seconds, a CAL EXPIRED warning, will flash as shown in Figure 5-11: CAL…

  • Page 40
    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 41: Manual Bump Test Options

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6. MANUAL BUMP TEST OPTIONS 6.1. Bump Testing A bump test verifies sensor response and alarm operation by exposing the monitor to a known con- centration of gas. The monitor has 2 manual bump test options: •…

  • Page 42: Apply Gas Screen (4-Gas Model)

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS Figure 6-2: Apply Gas Screen (4-Gas Model) 6.2.1. PUMPED INSTRUMENTS — REGULATOR SELECTION If the monitor has a pump enabled, the user is asked if the pump will be used during the bump test, as…

  • Page 43: Applying Test Gas

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6.3. Applying Test Gas Apply the test gas to the monitor via the direct flow regulator set to 0.5 l/min, as shown in Figure 6-4: Bump Test Kit Figure 6-4: Bump Test Kit 6.4.

  • Page 44: Confirm Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6.5. Confirm Alarms The user is prompted to confirm if the audible and visual alarms were activated, as shown in Figure 6-5: Confirm Alarms Figure 6-5: Confirm Alarms 6.5.1. Bump Test – Pass…

  • Page 45: Bump Test Fail — Switch Monitor Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS After selecting YES or NO, BUMP TEST FAIL is displayed and the user is instructed to switch the monitor off, as shown in Figure 6-8: Bump Test Fail — Switch Monitor OFF Figure 6-8: Bump Test Fail — Switch Monitor OFF If a monitor fails a bump test, then a full re-calibration is necessary.

  • Page 46
    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 47: Recharging The Batteries

    3M™ GMI™ PS200 PORTABLE GAS MONITOR RECHARGING THE BATTERIES 7. RECHARGING THE BATTERIES 7.1. General Use only 3M GMI chargers to recharge monitor. WARNING: CHARGING IS ONLY PERMITTED IN A NON-HAZARDOUS AREA. CAUTION: Switch the monitor off when charging. The battery should be recharged when: •…

  • Page 48: Connect Cable To Charging/Comms Clip

    3M™ GMI™ PS200 PORTABLE GAS MONITOR RECHARGING THE BATTERIES Charging/Comms Clip Mini-USB to USB Cable Figure 7-2: Connect Cable to Charging/Comms Clip 3. Connect the other end of the charging cable to the USB Power adapter (or suitable USB socket).

  • Page 49: Operator Maintenance

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATOR MAINTENANCE 8. OPERATOR MAINTENANCE 8.1. Cleaning CAUTION: Do not use polishes that contain silicon or solvent to clean the monitor as these may damage any flammable gas sensor fitted. Do not use abrasive chemicals or strong, volatile chemical solutions as these could damage the casing.

  • Page 50: Sample Inlet (Dust) Filter Replacement

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATOR MAINTENANCE • NOTE: Ensure that the locating pegs in the monitor filter recess align with the mating pin-holes in the hydrophobic filter. • Place the filter cover over the filter recess and slide it towards the display until the locating lugs engage in the locating slots.

  • Page 51: Calibration

    9.1. General Description The monitor has been calibrated for particular gases. Where any doubt exists, return the monitor to 3M GMI or an authorized distributor for calibration. WARNING: ONLY AUTHORIZED PERSONNEL MAY CALIBRATE THE MONITOR. Various calibration options are available: 1.

  • Page 52: Quick Calibration

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION 9.2. Quick Calibration This feature, if enabled, allows calibration without using the monitor’s set-up menu and passcodes. Quick Calibration uses predefined calibration gas cylinder values saved in the monitor. The default values are: •…

  • Page 53: Apply Gas Screen

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION NOTE: The test gas must match the predefined cylinder values stored in the monitor. 5. Once gas is applied, press and hold the R button to start the calibration. The 60 second count- down will commence (refer to ).

  • Page 54: Calibration Validity

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION A. UPDATE: The monitor saves the calibration data in the memory and automatically updates the calibration due date based on the option stored in the monitor set-up menu. B. SAVE ONLY: The monitor saves the calibration data in the memory but does not update the calibration due date.

  • Page 55: Accessories

    64052 Automatic Bump & Calibration Station (6mm fittings; includes PSU/USB/Softeware) 64052Q Automatic Bump & Calibration Station (1/4” fittings; includes PSU/USB/Software) 99118 On Demand Flow Regulator *Refer to https://gasdetection.3m/en/ps200 for the complete range of PS200 accessories. 64171, Rev. 10/May 2019 10-1…

  • Page 56
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ACCESSORIES* This page is intentionally left blank. 64171, Rev. 10/May 2019 10-2…
  • Page 57: Operating Parameters & Sensor Types

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATING PARAMETERS & SENSOR TYPES APPENDIX A. OPERATING PARAMETERS & SENSOR TYPES A.1. Typical Operating Parameters Range Resolution Response Time 0 to 100% 10 secs. (IEC 60079-26-1) Oxygen (O 0 to 25% 0.1% 12 secs. (BSEN 50104)

  • Page 58: Lel Sensor Types

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATING PARAMETERS & SENSOR TYPES A.2. LEL Sensor Types There are several LEL sensor types available for the monitor. The following table shows the benefits of each sensor and the combustible gases it will detect:…

  • Page 1
    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATIONS & MAINTENANCE MANUAL 64171, Rev. 10 WARNING: ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY FOR USING, MAINTAINING, OR SERVICING THIS PRODUCT MUST READ THIS ENTIRE MANUAL CAREFULLY. FAILURE TO USE THIS EQUIPMENT PROPERLY COULD RESULT IN SERIOUS INJURY OR DEATH.
  • Page 2
    Reformatted and revised entire manual. Copyright © 3M™ GMI™, May 2019 The information contained within is for use only with the 3M™ GMI™ PS200 Portable Gas Moni- tor. Reproduction, in whole or in part, including utilization in machines capable of reproduction or retrieval without written permission of 3M™…
  • Page 3
    GMI™ PS200 Portable Gas Monitor (or “the monitor”). LIABILITY Every care has been taken in the preparation of this O&M manual, but 3M™ and GMI™ do not accept any responsibility for errors or omissions and their consequences. Information in this O&M manual is subject to change without notice.
  • Page 4
    3M™ GMI™ PS200 PORTABLE GAS MONITOR This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 5: Table Of Contents

    3M™ GMI™ PS200 PORTABLE GAS MONITOR TABLE OF CONTENTS TABLE OF CONTENTS Sect. # Section Title Page # ABOUT THIS GUIDE ……………………..1-1 1.1. Guide Conventions ……………………… 1-1 1.2. Certifications and Approvals………………….1-2 1.3. General Safety Information………………….1-4 1.4. Additional Safety Details — CSA ONLY ………………

  • Page 6
    3M™ GMI™ PS200 PORTABLE GAS MONITOR TABLE OF CONTENTS Sect. # Section Title Page # 6.4. Quick/Full Bump Test ……………………6-3 6.5. Confirm Alarms ……………………..6-4 RECHARGING THE BATTERIES ………………… 7-1 7.1. General…………………………7-1 7.2. Recharge Monitor Using the Charging/Comms Clip …………7-1 OPERATOR MAINTENANCE ………………….8-1…
  • Page 7
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF FIGURES LIST OF FIGURES Fig. # Figure Title Page # PS200 Portable Gas Detector ………………….2-1 Display Example (4-Gas) …………………..2-2 Display Examples ……………………..2-2 OK Display Example (4-Gas) ………………….2-2 Function Buttons ……………………..3-1 Quick Configuration Display ………………….3-2…
  • Page 8
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF FIGURES Fig. # Figure Title Page # 5-12 Calibration Required……………………5-5 BUMP TEST Confirmation Screen ………………..6-1 Apply Gas Screen (4-Gas Model) ………………..6-2 Regulator Valve Choice …………………….6-2 Bump Test Kit……………………….6-3 Confirm Alarms ……………………..6-4 Bump Test Pass ……………………..6-4…
  • Page 9
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF TABLES LIST OF TABLES Table # Table Title Page # Certifications and Approvals………………….1-2 10-1 Supplied Accessories……………………10-1 10-2 Optional Accessories ……………………10-1 10-3 Charging Accessories……………………10-1 10-4 Manual Bump & Field Calibration Kit ………………10-1 10-5 Automatic Bump &…
  • Page 10
    3M™ GMI™ PS200 PORTABLE GAS MONITOR LIST OF TABLES This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 11: About This Guide

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE 1. ABOUT THIS GUIDE This guide instructs gas detection personnel on the features and usage of the PS200 Portable Gas Monitor (or “the monitor”), including information on operation, configuration, maintenance, specifi- cations and troubleshooting.

  • Page 12: Certifications And Approvals

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE 1.2. Certifications and Approvals The monitor has the following approvals available: Table 1-1: Certifications and Approvals Mark II 2G Ex ia d IIC T4 Gb (Ta = -20ºC to +50ºC) ATEX IECEx Ex ia d IIC T4 Gb (Ta = -20ºC to +50ºC) or…

  • Page 13
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE Table 1-1: Certifications and Approvals Mark TP TC 012/2011 TP TC 020/2 1.2.1. LABEL Always refer to the rear label for the actual certification of your monitor (see Figure 1-1: Identification…
  • Page 14: General Safety Information

    INSTRUCTIONS MAY ALSO RESULT IN SERIOUS INJURY OR DEATH. 3M™ GMI™ can take no responsibility for use of its equipment if it is not used in accordance with these instructions. If further operational or maintenance details are required but not provided in this guide, contact 3M™…

  • Page 15
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE ATTENTION: TOUT LECTURE DÈPASSANT RAPIDEMENT LE MAXIMUM DE L’ÊCHELLE ET SUIVIE PAR UNE DIMINUTION OU UNE LECTURE ERRONÉE, INDIQUE UNE CONCETRATION DE GAZ SUPÉRIEURE À LA VALEUR MAXIMALE DE CETTE ÉCHELLE. CETTE LECTURE N’EST PASSIGNIFICATIVE.
  • Page 16
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ABOUT THIS GUIDE This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 17: Introduction

    2. INTRODUCTION 2.1. General Description 3M GMI PS200 Portable Gas Monitor series combines quality, ruggedness and advanced technol- ogy in a user friendly, portable gas detector. It is compact, lightweight, water resistant, extremely robust, and is suitably certified to recognized International Standards.

  • Page 18: Display Example (4-Gas)

    3M™ GMI™ PS200 PORTABLE GAS MONITOR INTRODUCTION All gases being monitored are displayed. A 4-gas monitor display is illustrated in Figure 2-2: Display Example (4-Gas) Figure 2-2: Display Example (4-Gas) NOTE: Character size increases if fewer sensors are configured, as illustrated in…

  • Page 19: Features

    2.3.1. Accessing Logged Data Download data from the monitor to a PC/Laptop using the standard charging/comms cable and additional software. Contact the 3M GMI Sales Department for further details. 2.4. Filters Filters protect the monitor from water and dust ingress. Check filters regularly and replace as neces- sary (refer to Section 8.2.

  • Page 20
    3M™ GMI™ PS200 PORTABLE GAS MONITOR INTRODUCTION This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 21: Operation

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 3. OPERATION Figure 3-1: Function Buttons 3.1. Operating Procedure CAUTION: The monitor can be supplied with a flammable gas sensor. This sensor is designed for use in concentrations of gas not exceeding the Lower Explosive Limit (LEL).

  • Page 22: Quick Configuration Display

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Additional: • Perform regular leak checks on pumped instruments by placing your thumb over the sample inlet nozzle and ensuring the monitor displays FLOW FAULT. • Perform regular bump tests. 3.2. Quick Configuration Display…

  • Page 23: Flash Screen

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-3: Flash Screen 3.3.1. MONITOR IDENTIFICATION During warm-up, the LCD identifies the serial number, software version, battery status, and calibra- tion gas information, as illustrated in Figure 3-4: Monitor Identification Display Figure 3-4: Monitor Identification Display 3.3.2.

  • Page 24: Date And Time

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-6: Date and Time 3.3.4. BUMP DUE DATE (OPTION) This configurable option is disabled by default but can be configured to: • indicate briefly that bump test is due, as illustrated in Figure 3-7: Bump Due Date •…

  • Page 25: Calibration Due Date

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 3.3.5. CALIBRATION DUE DATE The calibration due date feature (shown in ) is configurable, including Figure 3-9: Calibration Due Date the factory-set option to display the number of days remaining until the next calibration is due.

  • Page 26: Cal Gas Selection

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-11: Cal Gas Selection When this option is displayed (see ), the original gas used to calibrate the Figure 3-11: Cal Gas Selection monitor is identified between two arrowheads. NOTE: The calibration certificate also identifies the calibration gas type used.

  • Page 27: Switch The Display Backlight On/Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-13: 4-Gas Normal Operating Display 3.4. Switch the Display Backlight On/Off The display backlight can be manually switched ON when working in poor lighting conditions. Press the R button once to switch ON. After 20 seconds it will automatically switch OFF.

  • Page 28: Manual Datalog

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION 6. MAX/MIN readings can be reset by pressing and holding the R button for 2 seconds when either of the MAX/MIN screens are displayed. After resetting, the monitor will return to the normal operating screen.

  • Page 29: Self Test

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-16: Sample Line Connection 3.8.1. PUMP OPERATION Press and hold the R button to start or stop the pump. When the pump is running, a pump symbol (as highlighted in ) rotates in Figure 3-17: Pump Symbol the display.

  • Page 30: Switch The Monitor Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATION Figure 3-18: Self Test The current username, if configured, will appear at the top of the display. 3.10. Switch the monitor OFF Press and hold both the L and R buttons to switch the monitor OFF.

  • Page 31: Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS 4. ALARMS 4.1. Gas Alarms When an alarm set-point has been reached, the audible, visual, and vibrating alarms will activate to alert the users. All gas alarms are configurable to meet the specific needs of users and only operate once the monitor warm-up is complete.

  • Page 32: Acknowledge Gas Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS Figure 4-2: LEL HIHI Alarm 4.2. Acknowledge Gas Alarms After the gas readings have returned to a safe level, press and hold the R button to clear all alarms. 4.3. Mute Alarms If configured, muting the alarm will silence the audible alarm for 60 seconds, as follows: •…

  • Page 33: Get Out/High Gas Timer

    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS Figure 4-3: Over Range Alarm 4.4.1. 10 SECOND SWITCH OFF The monitor must be returned to clean air. NOTE: To avoid accidental switch-off in this dangerous state, the off cycle is increased to 10 seconds.

  • Page 34
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ALARMS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 35: Warnings & Faults

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS 5. WARNINGS & FAULTS 5.1. Low Battery LOW BATTERY flashes when approximately 30 minutes of run time remains, as shown in Figure 5-1: . The display turns red, the audible alarm sounds once every two seconds, and LOW BATTERY Warning the red LED’s flash.

  • Page 36: Lel Zero Fail

    Figure 5-4: LEL ZERO FAULT Return the monitor to clean air and switch it OFF then ON. If the ZERO FAULT persists, return the monitor to a 3M GMI approved service/repair facility. NOTE: The monitor can still detect and alarm on all other sensors.

  • Page 37: Sensor Faults

    S, as illustrated in , allow Figure 5-7: CO/H S SENSOR FAULT the monitor to run in clean air for 20 minutes. If the fault persists, return the monitor to a 3M GMI approved service/repair facility. Figure 5-7: CO/H S SENSOR FAULT 5.5.

  • Page 38: Memory Fault

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS Figure 5-8: FLOW FAULT 5.6. Memory Fault During warm-up, if the ‘MEMORY FAULT’ screen is displayed, as illustrated in Figure 5-9: Memory , the monitor has detected a memory fault. Fault Figure 5-9: Memory Fault The monitor must be returned for service.

  • Page 39: Cal Expired

    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS Alternately, press the R button once to acknowledge the warning, cancel the audible/visual alarm and continue. Every 30 seconds, a CAL EXPIRED warning, will flash as shown in Figure 5-11: CAL…

  • Page 40
    3M™ GMI™ PS200 PORTABLE GAS MONITOR WARNINGS & FAULTS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 41: Manual Bump Test Options

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6. MANUAL BUMP TEST OPTIONS 6.1. Bump Testing A bump test verifies sensor response and alarm operation by exposing the monitor to a known con- centration of gas. The monitor has 2 manual bump test options: •…

  • Page 42: Apply Gas Screen (4-Gas Model)

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS Figure 6-2: Apply Gas Screen (4-Gas Model) 6.2.1. PUMPED INSTRUMENTS — REGULATOR SELECTION If the monitor has a pump enabled, the user is asked if the pump will be used during the bump test, as…

  • Page 43: Applying Test Gas

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6.3. Applying Test Gas Apply the test gas to the monitor via the direct flow regulator set to 0.5 l/min, as shown in Figure 6-4: Bump Test Kit Figure 6-4: Bump Test Kit 6.4.

  • Page 44: Confirm Alarms

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS 6.5. Confirm Alarms The user is prompted to confirm if the audible and visual alarms were activated, as shown in Figure 6-5: Confirm Alarms Figure 6-5: Confirm Alarms 6.5.1. Bump Test – Pass…

  • Page 45: Bump Test Fail — Switch Monitor Off

    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS After selecting YES or NO, BUMP TEST FAIL is displayed and the user is instructed to switch the monitor off, as shown in Figure 6-8: Bump Test Fail — Switch Monitor OFF Figure 6-8: Bump Test Fail — Switch Monitor OFF If a monitor fails a bump test, then a full re-calibration is necessary.

  • Page 46
    3M™ GMI™ PS200 PORTABLE GAS MONITOR MANUAL BUMP TEST OPTIONS This page is intentionally left blank. 64171, Rev. 10/May 2019…
  • Page 47: Recharging The Batteries

    3M™ GMI™ PS200 PORTABLE GAS MONITOR RECHARGING THE BATTERIES 7. RECHARGING THE BATTERIES 7.1. General Use only 3M GMI chargers to recharge monitor. WARNING: CHARGING IS ONLY PERMITTED IN A NON-HAZARDOUS AREA. CAUTION: Switch the monitor off when charging. The battery should be recharged when: •…

  • Page 48: Connect Cable To Charging/Comms Clip

    3M™ GMI™ PS200 PORTABLE GAS MONITOR RECHARGING THE BATTERIES Charging/Comms Clip Mini-USB to USB Cable Figure 7-2: Connect Cable to Charging/Comms Clip 3. Connect the other end of the charging cable to the USB Power adapter (or suitable USB socket).

  • Page 49: Operator Maintenance

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATOR MAINTENANCE 8. OPERATOR MAINTENANCE 8.1. Cleaning CAUTION: Do not use polishes that contain silicon or solvent to clean the monitor as these may damage any flammable gas sensor fitted. Do not use abrasive chemicals or strong, volatile chemical solutions as these could damage the casing.

  • Page 50: Sample Inlet (Dust) Filter Replacement

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATOR MAINTENANCE • NOTE: Ensure that the locating pegs in the monitor filter recess align with the mating pin-holes in the hydrophobic filter. • Place the filter cover over the filter recess and slide it towards the display until the locating lugs engage in the locating slots.

  • Page 51: Calibration

    9.1. General Description The monitor has been calibrated for particular gases. Where any doubt exists, return the monitor to 3M GMI or an authorized distributor for calibration. WARNING: ONLY AUTHORIZED PERSONNEL MAY CALIBRATE THE MONITOR. Various calibration options are available: 1.

  • Page 52: Quick Calibration

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION 9.2. Quick Calibration This feature, if enabled, allows calibration without using the monitor’s set-up menu and passcodes. Quick Calibration uses predefined calibration gas cylinder values saved in the monitor. The default values are: •…

  • Page 53: Apply Gas Screen

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION NOTE: The test gas must match the predefined cylinder values stored in the monitor. 5. Once gas is applied, press and hold the R button to start the calibration. The 60 second count- down will commence (refer to ).

  • Page 54: Calibration Validity

    3M™ GMI™ PS200 PORTABLE GAS MONITOR CALIBRATION A. UPDATE: The monitor saves the calibration data in the memory and automatically updates the calibration due date based on the option stored in the monitor set-up menu. B. SAVE ONLY: The monitor saves the calibration data in the memory but does not update the calibration due date.

  • Page 55: Accessories

    64052 Automatic Bump & Calibration Station (6mm fittings; includes PSU/USB/Softeware) 64052Q Automatic Bump & Calibration Station (1/4” fittings; includes PSU/USB/Software) 99118 On Demand Flow Regulator *Refer to https://gasdetection.3m/en/ps200 for the complete range of PS200 accessories. 64171, Rev. 10/May 2019 10-1…

  • Page 56
    3M™ GMI™ PS200 PORTABLE GAS MONITOR ACCESSORIES* This page is intentionally left blank. 64171, Rev. 10/May 2019 10-2…
  • Page 57: Operating Parameters & Sensor Types

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATING PARAMETERS & SENSOR TYPES APPENDIX A. OPERATING PARAMETERS & SENSOR TYPES A.1. Typical Operating Parameters Range Resolution Response Time 0 to 100% 10 secs. (IEC 60079-26-1) Oxygen (O 0 to 25% 0.1% 12 secs. (BSEN 50104)

  • Page 58: Lel Sensor Types

    3M™ GMI™ PS200 PORTABLE GAS MONITOR OPERATING PARAMETERS & SENSOR TYPES A.2. LEL Sensor Types There are several LEL sensor types available for the monitor. The following table shows the benefits of each sensor and the combustible gases it will detect:…

File Specifications:

1478/1478156-gmi_ps200_series.pdf file (24 Apr 2023)

Accompanying Data:

3M GMI PS200 Series Gas Detectors, Measuring Instruments PDF Quick Operation Manual (Updated: Monday 24th of April 2023 05:32:00 AM)

Rating: 4.2 (rated by 21 users)

Compatible devices: Touch Screen Controller SX112, PL300, PROTEGE PLUS, 1200, Digital Walldisplay 9000PD Plus, Edge eg5, Touch Screen Controller RX141, Dynapro ET 3200 Series.

Recommended Documentation:

Quick Operation Manual (Text Version):

(Ocr-Read Summary of Contents of some pages of the 3M GMI PS200 Series Document (Main Content), UPD: 24 April 2023)

  • 1, Quick Operation Guide Quick Zero & Calibration (Optional) 10s Quick Config Display Switch On Alarm AcknowledgeSelf Test 1s1s PumpManual Datalog 1s 2s Switch O 3s Calibration Overdue https://gasdetection.3m.com/en/ps200 This Quick Operation Guide is NOT a substitute for the Operations & Maintenance Manual which MUST be read and understood, prior to operating, by anyone using or servicing the monitor. Failure to do…

  • 2, Part Number: 64191 Issue 6 Backlight MAX / MIN / STEL / LTEL Reset MAX / MIN 20s Zero Fault / Flow Fault 1s Charging 4h

  • 3M GMI PS200 Series User Manual

  • 3M GMI PS200 Series User Guide

  • 3M GMI PS200 Series PDF Manual

  • 3M GMI PS200 Series Owner’s Manuals

Recommended: MWB 850, CX-5, EEH 6300.8, Accu-Z Razor 357-044M

Links & Tools

Operating Impressions, Questions and Answers:

background image

PS200

Series

QUICK OPERATING INSTRUCTIONS

Switch ON

Press and hold the Right Hand button

for one second to switch the instrument ON and
initiate a warm-up routine.

The display will show the serial number,

software version, battery status and calibration
gas. Next, the date and time are shown, followed
by the Calibration Due date. If expired, press the

button to acknowledge.

Lastly, a sensor check is run. If sensors are

working OK and were zeroed correctly, a

symbol is displayed adjacent to the gas type. If a
sensor fails the zero check, the display will be
red, the audible / visual alarms activate and a
symbol is displayed adjacent to the faulty gas type.
To acknowledge this fault, press the

button.

Switch ON — Manual Bump Test

Press and hold the

button to switch the

instrument ON and initiate a Manual Bump Test
(after the warm-up routine is complete). When
concluded, the instrument will operate as normal.

Refer to Essential Health and Safety Requirement (EHSR) and PS200

Series User Handbook before first use.

Backlight / MAX & MIN

Press the

button once to switch the backlight

ON. (After 20 seconds it will automatically switch
OFF.) Press again to view MAX readings and a further
press to view the MIN readings. To reset MAX / MIN
readings, press and hold the

button for 2

seconds when either MAX / MIN screen is displayed.

Alarms Reset / Acknowledge

In alarm condition, the display will be red and the
audible and visual alarms are activated. When the
gas readings (of a Latching Alarm) are safe, press
and hold the

button to reset / acknowledge the

alarm. (If alarms are Non-Latching, the audible and
visual alarms reset automatically when gas readings
are safe.)

Normal Operation

The standard operating screen for the 4 Gas model
is shown here. Other models will report the relevant
gases they are capable of detecting.

Switch Pump ON / OFF

Press and hold the

button to toggle the pump ON / OFF

(Pumped instruments only).

Manual Datalog

Press the

button once to perform a manual datalog, briefly indicated on

the display as ‘LOGGING’.

Switch OFF

Press and hold both

and

buttons, for 3 seconds, to switch the

instrument OFF.

Портативный газоанализатор PS200

Предназначен для одновременного обнаружения и контроля концентрации горючих газов (Ех), кислорода O2, сероводорода H2S, угарного газа CO.

Регистрация данных – это стандартная функция всех приборов серии PS200. Внутренняя память может хранить до 180 записей испытаний с контрольным выбросом.

Для упрощения работы с прибором ps200 system sensor была разработана станция калибровки и автоматической проверки методом контрольного выброса.

Прибор легко устанавливается на место, при этом станция автоматически начинает регистрировать результаты проверки и калибровки, что занимает всего 60 и 90 секунд соответственно.

Пользователь может применять станцию gmi ps200 как автономное устройство, с которого легко скачать данные ps 200 для последующего считывания и просмотра в доступном виде с регистрацией по времени.

Данная информация может быть сохранена для обеспечения непрерывной регистрации событий. Результаты сохраняются в столбцах и могут быть легко преобразованы в электронные таблицы.

Комплект поставки прибора PS200 включает следующее:

  • зажим для зарядки/связи,
  • универсальный разъем питания,
  • компакт-диск с руководством пользователя,
  • краткое руководство,
  • 3-метровая трубка с быстроразъемным соединителем (только для исполнения со встроенным насосом).

Основные характеристики PS200:

Система сигнализации тройная: визуальная, звуковая и вибрационная

Время работы от аккумулятора: 14 ч – со стандартным датчиком без насоса,8ч – с насосом, до 80ч – с новым высокотехнологичным датчиком LEL.

Отображение данных токсичных газов в ppm или в мг/м3: настройка через меню

Память: 24 часа с настраиваемым интервалом от 1 сек до 10 мин

Меню: на русском языке.

Масса: 215 г(без насоса)

Класс водо — пыленепроницаемости: IP 67

Температурный режим работы: от -40С до +50С

Встроенный насос: опция

Гарантия: 2 года.

Диапазон измерений: LEL 0-100%/1%; CO 0-1000ppm/1 ppm; H2S 0-100ppm/1 ppm; O2 0-25%/1%;

Дополнительные аксессуары: зарядная станция на 5 приборов, автомобильное ЗУ на 12/24В, резиновые чехлы 7 цветов, автоматическая калибровочная станция.

Дополнительные аксессуары PS200:

  • зарядная станция на 5 приборов ps 200
  • автомобильное зарядное устройство 12/24В
  • крэдл для зарядки
  • резиновые чехлы 7 цветов,
  • сменные фильтры
  • ручной аспиратор
  • кейс для переноски
Определяемые газы

Сероводород (H2S),

Угарный Газ (CО),

Метан (CH4),

Кислород (O2)

Диапазоны измерений

Lel (горючие газы — СН4 и др.): от 0 до 100% НПВ или от 0 до 5% об.,

H2S — (Сероводород): от 0 до 100 ppm,

CO (Угарный газ): от 0 до 1000 ppm,

O2 (Кислород): от 0 до 25%

Рабочая температура

От -20 до 50 °C

Степень защиты

IP 67

Маркировка взрывозащиты

Ex ia d IIBT3

Вес

215 гр.

Размеры

12,1 x 5,9 x 3,2 мм

Питание

Литий-ионный перезаряжаемый аккумулятор

Время зарядки батареи

Не более 4 часов

Время работы батареи

> 14 часов (прокачка > 8 часов)

Предупреждающая сигнализация

Хорошо заметный мигающий светодиод, резкий звуковой сигнал > 90 дБ

Гарантия

Полная 2-летняя гарантия

Материал корпуса

Высокопрочный прорезиненный поликарбонат

Понравилась статья? Поделить с друзьями:
  • Profibus tester 5 bc 700 pb инструкция на русском
  • Руководство французской революции
  • Мануал хонда инсайт 2009 на русском
  • Optibelt tt инструкция на русском языке
  • Инструкция по охране труда при работе с кофемашиной