Configuration and Use Manual
MMI-20019028, Rev AB
March 2018
Micro Motion® Model 1700 Transmitters with Analog Outputs
Configuration and Use Manual
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
•Worldwide: flow.support@emerson.com
•Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America |
Europe and Middle East |
Asia Pacific |
||||
United States |
800-522-6277 |
U.K. |
0870 240 1978 |
Australia |
800 |
158 727 |
Canada |
+1 303-527-5200 |
The Netherlands |
+31 (0) 704 136 666 |
New Zealand |
099 |
128 804 |
Mexico |
+41 (0) 41 7686 111 |
France |
0800 917 901 |
India |
800 |
440 1468 |
Argentina |
+54 11 4837 7000 |
Germany |
0800 182 5347 |
Pakistan |
888 |
550 2682 |
Brazil |
+55 15 3413 8000 |
Italy |
8008 77334 |
China |
+86 |
21 2892 9000 |
Central & Eastern |
+41 (0) 41 7686 111 |
Japan |
+81 |
3 5769 6803 |
||
Russia/CIS |
+7 495 981 9811 |
South Korea |
+82 |
2 3438 4600 |
||
Egypt |
0800 000 0015 |
Singapore |
+65 |
6 777 8211 |
||
Oman |
800 70101 |
Thailand |
001 |
800 441 6426 |
||
Qatar |
431 0044 |
Malaysia |
800 |
814 008 |
||
Kuwait |
663 299 01 |
|||||
South Africa |
800 991 390 |
|||||
Saudi Arabia |
800 844 9564 |
|||||
UAE |
800 0444 0684 |
|||||
Contents
Contents
Part I |
Getting started |
|||
Chapter 1 |
Before you begin ………………………………………………………………………………………………. |
3 |
||
1.1 |
About this manual …………………………………………………………………………………………………………. |
3 |
||
1.2 |
Transmitter model code …………………………………………………………………………………………………. |
3 |
||
1.3 |
Communications tools and protocols ……………………………………………………………………………….. |
4 |
||
1.4 |
Additional documentation and resources ………………………………………………………………………….. |
4 |
||
Chapter 2 |
Quick start |
……………………………………………………………………………………………………….. |
5 |
|
2.1 |
Power up the transmitter ………………………………………………………………………………………………… |
5 |
||
2.2 |
Check meter status ………………………………………………………………………………………………………… |
6 |
||
2.2.1 ……………………………………………………………………… |
Transmitter status reported by LED |
6 |
||
2.3 |
Make ……………………………………………………………………..a startup connection to the transmitter |
7 |
||
2.4 |
(Optional) ………………………………………………………………Adjust digital communications settings |
7 |
||
2.5 |
Verify …………………………………………………………………………………………mass flow measurement |
7 |
||
2.6 |
Verify ……………………………………………………………………………………………………………….the zero |
8 |
||
2.6.1 …………………………………… |
Terminology used with zero verification and zero calibration |
9 |
Part II |
Configuration and commissioning |
|||
Chapter 3 |
Introduction to configuration and commissioning ………………………………………………… |
13 |
||
3.1 |
Configuration flowchart ……………………………………………………………………………………………….. |
13 |
||
3.2 |
Default values and ranges ……………………………………………………………………………………………… |
15 |
||
3.3 |
Enable access to the off-line menu of the display ………………………………………………………………. |
15 |
||
3.4 |
Disable write-protection on the transmitter configuration …………………………………………………. |
15 |
||
3.5 |
Restore the factory configuration …………………………………………………………………………………… |
16 |
||
Chapter 4 |
Configure process measurement ……………………………………………………………………….. |
17 |
||
4.1 |
Configure mass flow measurement ………………………………………………………………………………… |
17 |
||
4.1.1 |
Configure Mass Flow Measurement Unit ……………………………………………………………. |
17 |
||
4.1.2 |
Configure Flow Damping …………………………………………………………………………………. |
20 |
||
4.1.3 |
Configure Mass Flow Cutoff ……………………………………………………………………………… |
21 |
||
4.2 |
Configure volume flow measurement for liquid applications ………………………………………………. |
22 |
||
4.2.1 |
Configure Volume Flow Type for liquid applications ……………………………………………… |
23 |
||
4.2.2 |
Configure Volume Flow Measurement Unit for liquid applications ………………………….. |
23 |
||
4.2.3 |
Configure Volume Flow Cutoff …………………………………………………………………………. |
26 |
||
4.3 |
Configure GSV flow measurement ………………………………………………………………………………….. |
27 |
||
4.3.1 |
Configure Volume Flow Type for gas applications ………………………………………………… |
28 |
||
4.3.2 |
Configure Standard Density of Gas ……………………………………………………………………. |
28 |
||
4.3.3 |
Configure Gas Standard Volume Flow Unit …………………………………………………………. |
30 |
||
4.3.4 |
Configure Gas Standard Volume Flow Cutoff ………………………………………………………. |
33 |
||
4.4 |
Configure Flow Direction ……………………………………………………………………………………………… |
34 |
||
4.4.1 |
Options for Flow Direction ………………………………………………………………………………. |
35 |
||
4.5 |
Configure density measurement ……………………………………………………………………………………. |
39 |
||
4.5.1 |
Configure Density Measurement Unit ……………………………………………………………….. |
39 |
Configuration and Use Manual |
i |
Contents
4.5.2 |
Configure two-phase flow parameters ……………………………………………………………….. |
40 |
||
4.5.3 |
Configure Density Damping …………………………………………………………………………….. |
42 |
||
4.5.4 |
Configure Density Cutoff …………………………………………………………………………………. |
43 |
||
4.6 |
Configure temperature measurement …………………………………………………………………………….. |
44 |
||
4.6.1 |
Configure Temperature Measurement Unit ………………………………………………………… |
44 |
||
4.6.2 |
Configure Temperature Damping …………………………………………………………………….. |
44 |
||
4.6.3 |
Effect of Temperature Damping on process measurement ……………………………………. |
45 |
||
4.6.4 |
Configure Temperature Input ………………………………………………………………………….. |
45 |
||
4.7 |
Configure pressure compensation ………………………………………………………………………………….. |
46 |
||
4.7.1 |
Configure pressure compensation using ProLink III ……………………………………………… |
46 |
||
4.7.2 |
Configure pressure compensation using the Field Communicator ………………………….. |
47 |
||
4.7.3 |
Options for Pressure Measurement Unit …………………………………………………………….. |
49 |
||
Chapter 5 |
Configure device options and preferences …………………………………………………………… |
51 |
||
5.1 |
Configure the transmitter display …………………………………………………………………………………… |
51 |
||
5.1.1 |
Configure the language used for the display ……………………………………………………….. |
51 |
||
5.1.2 |
Configure the process variables and diagnostic variables shown on the display …………. |
51 |
||
5.1.3 |
Configure the number of decimal places (precision) shown on the display ……………….. |
53 |
||
5.1.4 |
Configure the refresh rate of data shown on the display ………………………………………… |
54 |
||
5.1.5 |
Enable or disable automatic scrolling through the display variables ………………………… |
54 |
||
5.1.6 |
Enable or disable the display backlight ……………………………………………………………….. |
55 |
||
5.1.7 |
Enable or disable Status LED Blinking …………………………………………………………………. |
55 |
||
5.2 |
Enable or disable operator actions from the display …………………………………………………………… |
56 |
||
5.2.1 |
Enable or disable Totalizer Start/Stop from the display ………………………………………….. |
56 |
||
5.2.2 |
Enable or disable Totalizer Reset from the display ………………………………………………… |
57 |
||
5.2.3 |
Enable or disable the Acknowledge All Alerts display command ……………………………… |
57 |
||
5.3 |
Configure security for the display menus …………………………………………………………………………. |
58 |
||
5.4 |
Configure response time parameters ……………………………………………………………………………… |
59 |
||
5.4.1 |
Configure Update Rate ……………………………………………………………………………………. |
60 |
||
5.4.2 |
Configure Response Time ………………………………………………………………………………… |
61 |
||
5.5 |
Configure alert handling ……………………………………………………………………………………………….. |
62 |
||
5.5.1 |
Configure Fault Timeout …………………………………………………………………………………. |
62 |
||
5.5.2 |
Configure Status Alert Severity …………………………………………………………………………. |
63 |
||
5.6 |
Configure informational parameters ………………………………………………………………………………. |
67 |
||
5.6.1 |
Configure Sensor Serial Number ……………………………………………………………………….. |
67 |
||
5.6.2 |
Configure Sensor Material ……………………………………………………………………………….. |
67 |
||
5.6.3 |
Configure Sensor Liner Material ………………………………………………………………………… |
68 |
||
5.6.4 |
Configure Sensor Flange Type ………………………………………………………………………….. |
68 |
||
5.6.5 |
Configure Descriptor ………………………………………………………………………………………. |
69 |
||
5.6.6 |
Configure Message ………………………………………………………………………………………… |
69 |
||
5.6.7 |
Configure Date ………………………………………………………………………………………………. |
69 |
||
Chapter 6 |
Integrate the meter with the control system ………………………………………………………… |
71 |
||
6.1 |
Configure the transmitter channels ………………………………………………………………………………… |
71 |
||
6.2 |
Configure the mA Output ……………………………………………………………………………………………… |
72 |
||
6.2.1 |
Configure mA Output Process Variable ……………………………………………………………… |
72 |
||
6.2.2 |
Configure Lower Range Value (LRV) and Upper Range Value (URV) …………………………. |
74 |
||
6.2.3 |
Configure AO Cutoff ……………………………………………………………………………………….. |
75 |
||
6.2.4 |
Configure Added Damping ………………………………………………………………………………. |
77 |
||
6.2.5 |
Configure mA Output Fault Action and mA Output Fault Level ………………………………. |
78 |
||
6.3 |
Configure the Frequency Output ……………………………………………………………………………………. |
79 |
||
6.3.1 |
Configure Frequency Output Polarity ………………………………………………………………… |
80 |
ii |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Contents |
||||
6.3.2 |
Configure Frequency Output Scaling Method ……………………………………………………… |
81 |
||
6.3.3 |
Configure Frequency Output Fault Action and Frequency Output Fault Level …………… |
82 |
||
6.4 |
Configure the Discrete Output ………………………………………………………………………………………. |
83 |
||
6.4.1 |
Configure Discrete Output Source ……………………………………………………………………. |
84 |
||
6.4.2 |
Configure Discrete Output Polarity …………………………………………………………………… |
86 |
||
6.4.3 |
Configure Discrete Output Fault Action ……………………………………………………………… |
86 |
||
6.5 |
Configure events …………………………………………………………………………………………………………. |
87 |
||
6.5.1 |
Configure a basic event ……………………………………………………………………………………. |
88 |
||
6.5.2 |
Configure an enhanced event …………………………………………………………………………… |
88 |
||
6.6 |
Configure digital communications …………………………………………………………………………………. |
90 |
||
6.6.1 |
Configure HART/Bell 202 communications ………………………………………………………… |
90 |
||
6.6.2 |
Configure HART/RS-485 communications ………………………………………………………….. |
95 |
||
6.6.3 |
Configure Modbus/RS-485 communications ………………………………………………………. |
96 |
||
6.6.4 |
Configure Digital Communications Fault Action ………………………………………………….. |
98 |
||
Chapter 7 |
Complete the configuration …………………………………………………………………………….. |
101 |
||
7.1 |
Test or tune the system using sensor simulation ……………………………………………………………… |
101 |
||
7.1.1 |
Sensor simulation …………………………………………………………………………………………. |
102 |
||
7.2 |
Back up transmitter configuration ………………………………………………………………………………… |
103 |
||
7.3 |
Enable write-protection on the transmitter configuration ………………………………………………… |
103 |
||
Part III Operations, maintenance, and troubleshooting |
||||
Chapter 8 |
Transmitter operation ……………………………………………………………………………………. |
107 |
||
8.1 |
Record the process variables ……………………………………………………………………………………….. |
107 |
||
8.2 |
View process variables ………………………………………………………………………………………………… |
108 |
||
8.2.1 |
View process variables using the display …………………………………………………………… |
108 |
||
8.2.2 |
View process variables and other data using ProLink III ……………………………………….. |
109 |
||
8.2.3 |
View process variables using the Field Communicator ………………………………………… |
109 |
||
8.3 |
View transmitter status using the status LED ………………………………………………………………….. |
110 |
||
8.4 |
View and acknowledge status alerts ……………………………………………………………………………… |
111 |
||
8.4.1 |
View and acknowledge alerts using the display …………………………………………………. |
111 |
||
8.4.2 |
View and acknowledge alerts using ProLink III …………………………………………………… |
113 |
||
8.4.3 |
View alerts using the Field Communicator ……………………………………………………….. |
114 |
||
8.5 |
Read totalizer and inventory values ………………………………………………………………………………. |
114 |
||
8.6 |
Start and stop totalizers and inventories ………………………………………………………………………… |
114 |
||
8.6.1 |
Start and stop totalizers and inventories using the display …………………………………… |
115 |
||
8.7 |
Reset totalizers ………………………………………………………………………………………………………….. |
116 |
||
8.7.1 |
Reset totalizers using the display …………………………………………………………………….. |
116 |
||
8.8 |
Reset inventories ……………………………………………………………………………………………………….. |
118 |
||
Chapter 9 |
Measurement support ……………………………………………………………………………………. |
119 |
||
9.1 |
Options for measurement support ……………………………………………………………………………….. |
119 |
||
9.2 |
Use Smart Meter Verification (SMV) ……………………………………………………………………………… |
120 |
||
9.2.1 |
SMV requirements ………………………………………………………………………………………… |
120 |
||
9.2.2 |
SMV test preparation …………………………………………………………………………………….. |
120 |
||
9.2.3 |
Run SMV ……………………………………………………………………………………………………… |
121 |
||
9.2.4 |
View test data ………………………………………………………………………………………………. |
125 |
||
9.2.5 |
Schedule automatic execution of the SMV test ………………………………………………….. |
129 |
||
9.3 |
Use PVR, TBR, and TMR ……………………………………………………………………………………………….. |
132 |
||
9.3.1 |
PVR, TBR, and TMR applications ………………………………………………………………………. |
133 |
||
9.4 |
Piecewise linearization (PWL) for calibrating gas meters …………………………………………………… |
134 |
Configuration and Use Manual |
iii |
Contents
9.5 |
Zero the meter ………………………………………………………………………………………………………….. |
134 |
|
9.6 |
Validate the meter ……………………………………………………………………………………………………… |
135 |
|
9.6.1 |
Alternate method for calculating the meter factor for volume flow ……………………….. |
136 |
|
9.7 |
Perform a (standard) D1 and D2 density calibration …………………………………………………………. |
137 |
|
9.7.1 |
Perform a D1 and D2 density calibration using ProLink III …………………………………….. |
138 |
|
9.7.2 |
Perform a D1 and D2 density calibration using the Field Communicator ………………… |
139 |
|
9.8 |
Perform a D3 and D4 density calibration (T-Series sensors only) ………………………………………… |
140 |
|
9.8.1 |
Perform a D3 or D3 and D4 density calibration using ProLink III ……………………………. |
140 |
|
9.8.2 |
Perform a D3 or D3 and D4 density calibration using the Field Communicator ……….. |
141 |
|
9.9 |
Perform temperature calibration ………………………………………………………………………………….. |
142 |
|
9.9.1 |
Perform temperature calibration using the display …………………………………………….. |
143 |
|
9.9.2 |
Perform temperature calibration using ProLink III ………………………………………………. |
143 |
|
9.9.3 |
Perform temperature calibration using the Field Communicator ………………………….. |
145 |
Chapter 10 Troubleshooting ……………………………………………………………………………………………. |
147 |
|
10.1 |
Status LED states ……………………………………………………………………………………………………….. |
148 |
10.2 |
Status alerts, causes, and recommendations ………………………………………………………………….. |
148 |
10.3 |
Flow measurement problems ……………………………………………………………………………………… |
159 |
10.4 |
Density measurement problems ………………………………………………………………………………….. |
161 |
10.5 |
Temperature measurement problems …………………………………………………………………………… |
162 |
10.6 |
Milliamp output problems …………………………………………………………………………………………… |
163 |
10.7 |
Frequency Output problems ………………………………………………………………………………………… |
164 |
10.8 |
Using sensor simulation for troubleshooting ………………………………………………………………….. |
165 |
10.9 |
Check power supply wiring ………………………………………………………………………………………….. |
165 |
10.10 |
Check sensor-to-transmitter wiring ………………………………………………………………………………. |
166 |
10.11 |
Check grounding ……………………………………………………………………………………………………….. |
167 |
10.12 |
Perform loop tests ……………………………………………………………………………………………………… |
167 |
10.12.1 Perform loop tests using the display ………………………………………………………………… |
167 |
|
10.12.2 Perform loop tests using ProLink III ………………………………………………………………….. |
169 |
|
10.12.3 Perform loop tests using the Field Communicator ……………………………………………… |
170 |
|
10.13 |
Check the HART communication loop …………………………………………………………………………… |
171 |
10.14 |
Check HART Address and mA Output Action ………………………………………………………………….. |
172 |
10.15 |
Check HART burst mode ……………………………………………………………………………………………… |
173 |
10.16 |
Check the trimming of the mA Output ………………………………………………………………………….. |
173 |
10.17 |
Check Lower Range Value and Upper Range Value ………………………………………………………….. |
173 |
10.18 |
Check mA Output Fault Action …………………………………………………………………………………….. |
173 |
10.19 |
Check for radio frequency interference (RFI) …………………………………………………………………… |
174 |
10.20 |
Check Frequency Output Scaling Method ……………………………………………………………………… |
174 |
10.21 |
Check Frequency Output Fault Action …………………………………………………………………………… |
174 |
10.22 |
Check Flow Direction …………………………………………………………………………………………………. |
175 |
10.23 |
Check the cutoffs ………………………………………………………………………………………………………. |
175 |
10.24 |
Check for two-phase flow (slug flow) …………………………………………………………………………….. |
175 |
10.25 |
Check the drive gain …………………………………………………………………………………………………… |
176 |
10.25.1 Collect drive gain data …………………………………………………………………………………… |
177 |
|
10.26 |
Check the pickoff voltage ……………………………………………………………………………………………. |
177 |
10.26.1 Collect pickoff voltage data ……………………………………………………………………………. |
178 |
|
10.27 |
Check for internal electrical problems …………………………………………………………………………… |
178 |
10.27.1 Check the sensor coils ……………………………………………………………………………………. |
179 |
|
10.28 |
Check the core processor LED ………………………………………………………………………………………. |
181 |
10.28.1 Core processor LED states ………………………………………………………………………………. |
184 |
|
10.29 |
Perform a 700 core processor resistance test ………………………………………………………………….. |
186 |
iv |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Contents
Appendices and reference |
|||
Appendix A |
Using the transmitter display …………………………………………………………………………… |
189 |
|
A.1 |
Components of the transmitter interface ………………………………………………………………………. |
189 |
|
A.2 |
Use the optical switches ……………………………………………………………………………………………… |
190 |
|
A.3 |
Access and use the display menu system ……………………………………………………………………….. |
191 |
|
A.3.1 Enter a floating-point value using the display …………………………………………………….. |
192 |
||
A.4 |
Display codes for process variables ……………………………………………………………………………….. |
195 |
|
A.5 |
Codes and abbreviations used in display menus ……………………………………………………………… |
196 |
|
Appendix B |
Using ProLink III with the transmitter ………………………………………………………………… |
201 |
|
B.1 |
Basic information about ProLink III ……………………………………………………………………………….. |
201 |
|
B.2 |
Connect with ProLink III ……………………………………………………………………………………………… |
202 |
|
B.2.1 Connection types supported by ProLink III ………………………………………………………… |
202 |
||
B.2.2 Connect with ProLink III to the service port ……………………………………………………….. |
203 |
||
B.2.3 Make a HART/Bell 202 connection …………………………………………………………………… |
204 |
||
B.2.4 Make a HART/RS-485 connection …………………………………………………………………….. |
209 |
||
B.2.5 Connect with ProLink III to the RS-485 port ……………………………………………………….. |
212 |
||
Appendix C |
Using a Field Communicator with the transmitter ……………………………………………….. |
217 |
|
C.1 |
Basic information about the Field Communicator …………………………………………………………… |
217 |
|
C.2 |
Connect with the Field Communicator …………………………………………………………………………. |
218 |
|
Appendix D |
Default values and ranges ……………………………………………………………………………….. |
221 |
|
D.1 |
Default values and ranges ……………………………………………………………………………………………. |
221 |
|
Appendix E |
Transmitter components and installation wiring ………………………………………………… |
227 |
|
E.1 |
Installation types ……………………………………………………………………………………………………….. |
227 |
|
E.2 |
Power supply terminals and ground ……………………………………………………………………………… |
230 |
|
E.3 |
Input/output (I/O) wiring terminals ………………………………………………………………………………. |
231 |
|
Appendix F |
NE 53 history ………………………………………………………………………………………………… |
233 |
|
F.1 |
NE 53 history …………………………………………………………………………………………………………….. |
233 |
Configuration and Use Manual |
v |
Contents
vi |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Getting started
Part I
Getting started
Chapters covered in this part:
•Before you begin
•Quick start
Configuration and Use Manual |
1 |
Getting started
2 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Before you begin
1 Before you begin
Topics covered in this chapter:
•About this manual
•Transmitter model code
•Communications tools and protocols
•Additional documentation and resources
1.1About this manual
This manual helps you configure, commission, use, maintain, and troubleshoot
Micro Motion Model 1700 transmitters with analog outputs.
Important
This manual assumes that the following conditions apply:
•The transmitter has been installed correctly and completely according to the instructions in the transmitter installation manual
•The installation complies with all applicable safety requirements
•The user is trained in local and corporate safety standards
1.2Transmitter model code
You can verify that this manual pertains to your transmitter by ensuring the model code on the transmitter tag matches the format.
Example:
The transmitter has a model number of the following form: 1700(R/I/E/B/C/M/P)**A******
R 4-wire remote-mount with aluminum housing
IIntegral mount
E 4-wire remote mount transmitter with 9-wire remote enhanced core processor
B4-wire remote mount transmitter with 9-wire remote core processor
C 9-wire remote-mount with integral core processor and aluminum housing
M 4-wire remote mount with stainless steel housing
P9-wire remote mount transmitter with integral core processor and stainless steel housing
A Analog outputs option board
Configuration and Use Manual |
3 |
Before you begin
1.3Communications tools and protocols
You can use several different communications tools and protocols to interface with the transmitter, use different tools in different locations, or use different tools for different tasks.
Tool |
Supported protocols |
|
ProLink III |
• |
HART/RS-485 |
• |
HART/Bell 202 |
|
• |
Modbus/RS-485 |
|
• |
Service port |
|
Field Communicator |
HART/Bell 202 |
|
For information about how to use the communication tools, see the appendices in this manual.
Tip
You may be able to use other communications tools, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.emerson.com.
1.4Additional documentation and resources
Topic |
Document |
Hazardous area installa- |
See the approval documentation shipped with the transmitter, or |
tion |
download the appropriate documentation at www.emerson.com. |
Product Data Sheet |
Micro Motion Series 1000 and Series 2000 Transmitters with MVD™ Tech |
nology Product Data Sheet |
|
Sensor |
Sensor documentation |
Transmitter installation |
Micro Motion® Model 1700 and 2700 Installation Manual |
All documentation resources are available at www.emerson.com or on the user documentation DVD.
4 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Quick start
2 Quick start
Topics covered in this chapter:
•Power up the transmitter
•Check meter status
•Make a startup connection to the transmitter
•(Optional) Adjust digital communications settings
•Verify mass flow measurement
•Verify the zero
2.1Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
1.Ensure that all transmitter and sensor covers and seals are closed.
DANGER!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
2.Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. The transmitter is self-switching and will automatically detect the supply voltage. When using DC power, a minimum of 1.5 amps of startup current is required. During this period, Alert 009 is active. The diagnostic routines should complete in approximately
30 seconds. For transmitters with a display, the status LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED exhibits different behavior, an alert is active.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
Configuration and Use Manual |
5 |
Quick start
2.2Check meter status
Check the meter for any error conditions that require user action or that affect measurement accuracy.
1.Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
2.Check the status LED on the transmitter.
Related information
View and acknowledge status alerts
2.2.1Transmitter status reported by LED
Table 2-1: Transmitter status reported by status LED
LED state |
Description |
Recommendation |
Solid green |
No alerts are active. |
Continue with configuration or process meas- |
urement. |
||
Flashing green (if ena- |
Unacknowledged corrected condition (no |
Continue with configuration or process meas- |
bled) |
alert) |
urement. Acknowledge the alert if you choose. |
Solid yellow |
One or more low-severity alerts are active. |
A low-severity alert condition does not affect |
measurement accuracy or output behavior. |
||
You can continue with configuration or proc- |
||
ess measurement, but Micro Motion still rec- |
||
ommends identifying and resolving the alert |
||
condition. |
||
Flashing yellow (if ena- |
Calibration in progress, or Known Density Ver- |
A low-severity alert condition does not affect |
bled) |
ification in progress. |
measurement accuracy or output behavior. |
One or more low-severity alerts are active and |
You can continue with configuration or proc- |
|
have not been acknowledged. |
ess measurement, but Micro Motion still rec- |
|
ommends identifying and resolving the alert |
||
condition. |
||
Solid red |
One or more high-severity alerts are active. |
A high-severity alert condition affects meas- |
urement accuracy and output behavior. Re- |
||
solve the alert condition before continuing. |
||
Flashing red (if ena- |
One or more high-severity alerts are active |
A high-severity alert condition affects meas- |
bled) |
and have not been acknowledged. |
urement accuracy and output behavior. Re- |
solve the alert condition before continuing. |
||
Acknowledge the alert if you choose. |
||
If Status LED Blinking is disabled, all LEDs will show a solid color rather than flashing.
6 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Quick start
2.3Make a startup connection to the transmitter
For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool |
Connection type to use |
Instructions |
ProLink III |
HART/RS-485 |
Appendix B |
Modbus/RS-485 |
||
Service port |
||
Field Communicator |
HART/Bell 202 |
Appendix C |
2.4(Optional) Adjust digital communications settings
Change the communications parameters to site-specific values.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Procedure
1.To change the communications parameters using ProLink III, choose Device Tools > Configuration > Communications.
2.To change the communications parameters using the Field Communicator, choose On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
2.5Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
•Read the value for Mass Flow Rate on the transmitter display.
•Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables panel.
•Connect to the transmitter with the Field Communicator and read the value for Mass Flow Rate.
Configuration and Use Manual |
7 |
Quick start
On-Line Menu > Overview > Primary Purpose Variables
Postrequisites
If the reported mass flow rate is not accurate:
•Check the characterization parameters.
•Review the troubleshooting suggestions for flow measurement issues.
2.6Verify the zero
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
•The zero is required by site procedures.
•The stored zero value fails the zero verification procedure.
Procedure
1.Allow the flowmeter to warm up for at least 20 minutes after applying power.
2.Run the process fluid through the sensor until the sensor temperature reaches the normal process operating temperature.
3.Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if available.
4.Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full of process fluid.
5.From ProLink III, choose Device Tools > Calibration > Zero Verification and Calibration > Verify Zero and wait until the procedure completes.
6.Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero or Field Verification Zero value. If the average value is close to 0, you should not need to zero the meter.
7.If the zero verification procedure fails:
a.Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is completely full of process fluid.
b.Verify that the process fluid is not flashing or condensing, and that it does not contain particles that can settle out.
c.Remove or reduce sources of electromechanical noise if appropriate.
8 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Quick start
d.Repeat the zero verification procedure.
e.If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
2.6.1Terminology used with zero verification and zero calibration
Term |
Definition |
Zero |
In general, the offset required to synchronize the left pickoff and the right |
pickoff under conditions of zero flow. Unit = microseconds. |
|
Factory Zero |
The zero value obtained at the factory, under laboratory conditions. |
Field Zero |
The zero value obtained by performing a zero calibration outside the fac- |
tory. |
|
Prior Zero |
The zero value stored in the transmitter at the time a field zero calibration |
is begun. May be the factory zero or a previous field zero. |
|
Manual Zero |
The zero value stored in the transmitter, typically obtained from a zero |
calibration procedure. It may also be configured manually. Also called |
|
“mechanical zero” or “stored zero”. |
|
Live Zero |
The real-time bidirectional mass flow rate with no flow damping or mass |
flow cutoff applied. An adaptive damping value is applied only when the |
|
mass flow rate changes dramatically over a very short interval. Unit = con- |
|
figured mass flow measurement unit. |
|
Zero Stability |
A laboratory-derived value used to calculate the expected accuracy for a |
sensor. Under laboratory conditions at zero flow, the average flow rate is |
|
expected to fall within the range defined by the Zero Stability value (0 ± |
|
Zero Stability). Each sensor size and model has a unique Zero Stability val- |
|
ue. Statistically, 95% of all data points should fall within the range defined |
|
by the Zero Stability value. |
|
Zero Calibration |
The procedure used to determine the zero value. |
Zero Time |
The time period over which the Zero Calibration procedure is performed. |
Unit = seconds. |
|
Field Verification Zero |
A 3-minute running average of the Live Zero value, calculated by the |
transmitter. Unit = configured mass flow measurement unit. |
|
Zero Verification |
A procedure used to evaluate the stored zero and determine whether or |
not a field zero can improve measurement accuracy. |
|
Configuration and Use Manual |
9 |
Quick start
10 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
•Introduction to configuration and commissioning
•Configure process measurement
•Configure device options and preferences
•Integrate the meter with the control system
•Complete the configuration
Configuration and Use Manual |
11 |
Configuration and commissioning
12 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Introduction to configuration and commissioning
3 Introduction to configuration and commissioning
Topics covered in this chapter:
•Configuration flowchart
•Default values and ranges
•Enable access to the off line menu of the display
•Disable write protection on the transmitter configuration
•Restore the factory configuration
3.1Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual. If you are using the Weights & Measures application, additional configuration and setup are required.
Configuration and Use Manual |
13 |
Introduction to configuration and commissioning
Figure 3-1: Configuration flowchart
Configure process measurement
Configure mass flow measurement
Configure volume flow meaurement
Volume flow type
Define gas properties
Configure flow direction
Configure density measurement
Configure temperature measurement
Configure pressure compensation (optional)
Configure PVR, TMR,
or TBR (if available)
Configure device options and preferences
Configure display parameters
Configure fault handling parameters
Configure sensor parameters
Configure device parameters
Integrate device with control system
Configure the channel(s)
Configure the mA
output(s)
Configure the frequency output(s)
Configure the discrete output(s)
Configure events
Configure digital communications
Test and move to production
Test or tune transmitter using sensor simulation
Back up transmitter configuration
Enable write-protection on transmitter configuration
Done
14 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Introduction to configuration and commissioning
3.2Default values and ranges
See Section D.1 to view the default values and ranges for the most commonly used parameters.
3.3Enable access to the off-line menu of the display
Display |
OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY |
ProLink III |
Device Tools > Configuration > Transmitter Display > Display Security |
Field Communicator |
Configure > Manual Setup > Display > Offline Variable Menu Features |
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.
3.4Disable write-protection on the transmitter configuration
Display |
OFF-LINE MAINT > CONFG > LOCK |
ProLink III |
Device Tools > Configuration > Write-Protection |
Field Communicator |
Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect |
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.
Configuration and Use Manual |
15 |
Introduction to configuration and commissioning
3.5Restore the factory configuration
Display |
Not available |
ProLink III |
Device Tools > Configuration Transfer > Restore Factory Configuration |
Field Communicator |
Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration |
Overview
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Important
You cannot restore factory configurations with a 700 core.
Tip
Restoring the factory configuration is not a common action. You may want to contact customer support to see if there is a preferred method to resolve any issues.
16 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
4 Configure process measurement
Topics covered in this chapter:
•Configure mass flow measurement
•Configure volume flow measurement for liquid applications
•Configure GSV flow measurement
•Configure Flow Direction
•Configure density measurement
•Configure temperature measurement
•Configure pressure compensation
4.1Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
4.1.1Configure Mass Flow Measurement Unit
Display |
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Mass Flow Unit |
Overview
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The unit used for mass total and mass inventory is derived from this unit.
Any selected measurement unit, (mass, volume or gas standard volume), is automatically applied to both the mA and Frequency Outputs.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual |
17 |
Configure process measurement
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label |
|||
Display |
ProLink III |
Field Communica- |
|
Unit description |
tor |
||
Grams per second |
G/S |
g/sec |
g/s |
Grams per minute |
G/MIN |
g/min |
g/min |
Grams per hour |
G/H |
g/hr |
g/h |
Kilograms per second |
KG/S |
kg/sec |
kg/s |
Kilograms per minute |
KG/MIN |
kg/min |
kg/min |
Kilograms per hour |
KG/H |
kg/hr |
kg/h |
Kilograms per day |
KG/D |
kg/day |
kg/d |
Metric tons per minute |
T/MIN |
mTon/min |
MetTon/min |
Metric tons per hour |
T/H |
mTon/hr |
MetTon/h |
Metric tons per day |
T/D |
mTon/day |
MetTon/d |
Pounds per second |
LB/S |
lbs/sec |
lb/s |
Pounds per minute |
LB/MIN |
lbs/min |
lb/min |
Pounds per hour |
LB/H |
lbs/hr |
lb/h |
Pounds per day |
LB/D |
lbs/day |
lb/d |
Short tons (2000 pounds) per |
ST/MIN |
sTon/min |
STon/min |
minute |
|||
Short tons (2000 pounds) per |
ST/H |
sTon/hr |
STon/h |
hour |
|||
Short tons (2000 pounds) per |
ST/D |
sTon/day |
STon/d |
day |
|||
Long tons (2240 pounds) per |
LT/H |
lTon/hr |
LTon/h |
hour |
|||
Long tons (2240 pounds) per |
LT/D |
lTon/day |
LTon/d |
day |
|||
Special unit |
SPECL |
special |
Spcl |
18 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Define a special measurement unit for mass flow
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow > Special Units |
Field Communicator |
Configure > Manual Setup > Measurements > Special Units > Mass Special Units |
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1.Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.Calculate Mass Flow Conversion Factor as follows:
a.x base units = y special units
b.Mass Flow Conversion Factor = x ÷ y
The original mass flow rate value is divided by this value.
4.Enter Mass Flow Conversion Factor.
5.Set Mass Flow Label to the name you want to use for the mass flow unit.
6.Set Mass Total Label to the name you want to use for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1.Set Base Mass Unit to Pounds (lb).
2.Set Base Time Unit to Seconds (sec).
3.Calculate Mass Flow Conversion Factor:
Configuration and Use Manual |
19 |
Configure process measurement
a.1 lb/sec = 16 oz/sec
b.Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4.Set Mass Flow Conversion Factor to 0.0625.
5.Set Mass Flow Label to oz/sec.
6.Set Mass Total Label to oz.
4.1.2Configure Flow Damping
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Flow Damping |
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table.
Update Rate setting |
Damping range |
Normal |
0 to 51.2 seconds |
Special |
0 to 40.96 seconds |
The value you enter is automatically rounded off to the nearest valid value. For example, if the damping is currently set to 0.8 seconds, any value entered up to 1.2 seconds will be rounded down to 0.8 seconds, and any value entered from 1.21 to 1.59 seconds will be rounded up to 1.6 seconds.
Update Rate setting |
Valid damping values |
Normal |
0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2 |
Special |
0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56, |
5.12, 10.24, 20.48, 40.96 |
|
20 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported via the mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.1.3Configure Mass Flow Cutoff
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff |
Overview
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended value is 0.5% of the nominal flow rate of the attached sensor. See the sensor specifications. Leaving Mass Flow Cutoff at 0.0 g/sec is not recommended.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Configuration and Use Manual |
21 |
Configure process measurement
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
•mA Output Process Variable: Mass Flow Rate
•Frequency Output Process Variable: Mass Flow Rate
•mA Output Cutoff: 10 g/sec
•Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
•mA Output Process Variable: Mass Flow Rate
•Frequency Output Process Variable: Mass Flow Rate
•mA Output Cutoff: 15 g/sec
•Mass Flow Cutoff: 10 g/sec
Result:
•If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
—The mA Output will report zero flow.
—The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all internal processing.
•If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all internal processing.
4.2Configure volume flow measurement for liquid applications
The volume flow measurement parameters control how liquid volume flow is measured and reported.
22 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Note
If you need to switch from gas standard volume to liquid volume, polling for base density will automatically be disabled.
4.2.1Configure Volume Flow Type for liquid applications
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Liquid |
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
•Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Liquid.
4.2.2Configure Volume Flow Measurement Unit for liquid applications
Display |
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Volume Flow Unit |
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Configuration and Use Manual |
23 |
Configure process measurement
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Label |
|||
Unit description |
Display |
ProLink III |
Field Communicator |
Cubic feet per second |
CUFT/S |
ft3/sec |
Cuft/s |
Cubic feet per minute |
CUF/MN |
ft3/min |
Cuft/min |
Cubic feet per hour |
CUFT/H |
ft3/hr |
Cuft/h |
Cubic feet per day |
CUFT/D |
ft3/day |
Cuft/d |
Cubic meters per second |
M3/S |
m3/sec |
Cum/s |
Cubic meters per minute |
M3/MIN |
m3/min |
Cum/min |
Cubic meters per hour |
M3/H |
m3/hr |
Cum/h |
Cubic meters per day |
M3/D |
m3/day |
Cum/d |
U.S. gallons per second |
USGPS |
US gal/sec |
gal/s |
U.S. gallons per minute |
USGPM |
US gal/min |
gal/min |
U.S. gallons per hour |
USGPH |
US gal/hr |
gal/h |
U.S. gallons per day |
USGPD |
US gal/day |
gal/d |
Million U.S. gallons per day |
MILG/D |
mil US gal/day |
MMgal/d |
Liters per second |
L/S |
l/sec |
L/s |
Liters per minute |
L/MIN |
l/min |
L/min |
Liters per hour |
L/H |
l/hr |
L/h |
Million liters per day |
MILL/D |
mil l/day |
ML/d |
Imperial gallons per second |
UKGPS |
Imp gal/sec |
Impgal/s |
Imperial gallons per minute |
UKGPM |
Imp gal/min |
Impgal/min |
24 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement |
||||
Label |
||||
Unit description |
Display |
ProLink III |
Field Communicator |
|
Imperial gallons per hour |
UKGPH |
Imp gal/hr |
Impgal/h |
|
Imperial gallons per day |
UKGPD |
Imp gal/day |
Impgal/d |
|
Barrels per second(1) |
BBL/S |
barrels/sec |
bbl/s |
|
Barrels per minute(1) |
BBL/MN |
barrels/min |
bbl/min |
|
Barrels per hour(1) |
BBL/H |
barrels/hr |
bbl/h |
|
Barrels per day(1) |
BBL/D |
barrels/day |
bbl/d |
|
Beer barrels per second(2) |
BBBL/S |
Beer barrels/sec |
bbbl/s |
|
Beer barrels per minute(2) |
BBBL/MN |
Beer barrels/min |
bbbl/min |
|
Beer barrels per hour(2) |
BBBL/H |
Beer barrels/hr |
bbbl/h |
|
Beer barrels per day(2) |
BBBL/D |
Beer barrels/day |
bbbl/d |
|
Special unit |
SPECL |
special |
Spcl |
|
(1)Unit based on oil barrels (42 U.S. gallons).
(2)Unit based on U.S. beer barrels (31 U.S. gallons).
Define a special measurement unit for volume flow
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow > Special Units |
Field Communicator |
Configure > Manual Setup > Measurements > Special Units > Volume Special Units |
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1.Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
Configuration and Use Manual |
25 |
Configure process measurement
3.Calculate Volume Flow Conversion Factor as follows:
a.x base units = y special units
b.Volume Flow Conversion Factor = x ÷ y
4.Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5.Set Volume Flow Label to the name you want to use for the volume flow unit.
6.Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1.Set Base Volume Unit to Gallons (gal).
2.Set Base Time Unit to Seconds (sec).
3.Calculate the conversion factor:
a.1 gal/sec = 8 pints/sec
b.Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4.Set Volume Flow Conversion Factor to 0.1250.
5.Set Volume Flow Label to pints/sec.
6.Set Volume Total Label to pints.
4.2.3Configure Volume Flow Cutoff
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Volume Flow Cutoff |
Overview
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is 0.
26 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. mAO Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
•mA Output Process Variable: Volume Flow Rate
•Frequency Output Process Variable: Volume Flow Rate
•AO Cutoff: 10 l/sec
•Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
•mA Output Process Variable: Volume Flow Rate
•Frequency Output Process Variable: Volume Flow Rate
•AO Cutoff: 15 l/sec
•Volume Flow Cutoff: 10 l/sec
Result:
•If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
—The mA Output will report zero flow.
—The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all internal processing.
•If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all internal processing.
4.3Configure GSV flow measurement
The gas standard volume (GSV) flow measurement parameters control how volume flow is measured and reported in a gas application.
Configuration and Use Manual |
27 |
Configure process measurement
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
4.3.1Configure Volume Flow Type for gas applications
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas |
Volume |
|
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
•Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2Configure Standard Density of Gas
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > GSV > Gas Ref Density |
Overview
The Standard Density of Gas value is the gas density at standard reference conditions. Use it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for Standard Density of Gas.
Procedure
From the Source field, choose the method to supply gas base density data and perform the required setup.
28 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement |
||
Option |
Description |
|
Fixed Value or Digital |
A host writes gas base density data to the meter at appropriate intervals. |
|
Communications |
Continue to Configure fixed value or digital communications. |
|
Poll for external value |
The meter polls an external HART device for gas base density data in order |
|
to then compute gas standard volume from the mass flow and gas base |
||
density. |
||
Continue to Poll for external value. |
||
Configure fixed value or digital communications
Prerequisites
Section 4.3.2
Procedure
1.Set Standard Density of Gas to the standard reference density of the gas you are measuring.
Note
ProLink III provides a guided method that you can use to calculate your gas base density, if you do not know it.
2.Continue to Section 4.3.3.
Poll for external value
Prerequisites
Section 4.3.2
Procedure
1. |
Set Polling Slot to an available slot. |
|
2. |
Set Polling Control n as one of the following options: |
|
The n is the value you selected in the Polling Slot field. |
||
If there is another master, and if that master is primary, then set this field to |
||
secondary. If the other master is secondary, then set this field to primary. |
||
Option |
Description |
|
Poll as Primary |
No other HART masters will be on the network. |
|
Poll as Secondary |
Other HART masters will be on the network. |
|
3. |
Set External Device Tag n to the HART tag of the device being polled. |
|
The n is the value you selected in the Polling Slot field. |
Configuration and Use Manual |
29 |
Configure process measurement
•The device being polled (slave) cannot have special units set for density. Otherwise, the master will reject the base density and report the following alarm:
A115: No External Input or Polled Data Alert
•On the slave side, setup the HART Primary Variable for Base Density. The master will reject anything other than Base Density for the HART Primary Variable and trigger an A115 alarm.
•The density units on the transmitter and the polled device can be different as long as they can be classified as density units; for example, kg/m3 and g/cm3. The transmitter converts the polled units into compatible specified units.
For wiring and setup instructions for a polled device, refer to the Micro Motion Gas Density Meters (GDM) Installation manual or the Micro Motion Specific Gravity Meters (SGM) Installation manual.
4. Continue to Section 4.3.3.
4.3.3Configure Gas Standard Volume Flow Unit
Display |
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > GSV |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > GSV > GSV Flow Unit |
Overview
Gas Standard Volume Flow Unit specifies the unit of measure that will be displayed for the gas standard volume flow. The measurement unit used for the gas volume total and the gas volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
For polling, the first transmitter (master) requests density from a second transmitter (slave) via HART communications. Special units for GSV are allowed on the master side, but the device being polled (slave) cannot have special units set for density, otherwise the master will reject the base density and report an A115: No External Input or Polled Data Alert.
Procedure
Set Gas Standard Volume Flow Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Unit is SCFM (Standard Cubic Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
30 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Options for Gas Standard Volume Flow Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label |
|||
Unit description |
Display |
ProLink III |
Field Communicator |
Normal cubic meters per second |
NM3/S |
Nm3/sec |
Nm3/sec |
Normal cubic meters per minute |
NM3/MN |
Nm3/sec |
Nm3/min |
Normal cubic meters per hour |
NM3/H |
Nm3/hr |
Nm3/hr |
Normal cubic meters per day |
NM3/D |
Nm3/day |
Nm3/day |
Normal liters per second |
NLPS |
NLPS |
NLPS |
Normal liters per minute |
NLPM |
NLPM |
NLPM |
Normal liters per hour |
NLPH |
NLPH |
NLPH |
Normal liters per day |
NLPD |
NLPD |
NLPD |
Standard cubic feet per second |
SCFS |
SCFS |
SCFS |
Standard cubic feet per minute |
SCFM |
SCFM |
SCFM |
Standard cubic feet per hour |
SCFH |
SCFH |
SCFH |
Standard cubic feet per day |
SCFD |
SCFD |
SCFD |
Standard cubic meters per second |
SM3/S |
Sm3/sec |
Sm3/sec |
Standard cubic meters per minute |
SM3/MN |
Sm3/min |
Sm3/min |
Standard cubic meters per hour |
SM3/H |
Sm3/hr |
Sm3/hr |
Standard cubic meters per day |
SM3/D |
Sm3/day |
Sm3/day |
Standard liters per second |
SLPS |
SLPS |
SLPS |
Standard liters per minute |
SLPM |
SLPM |
SLPM |
Standard liters per hour |
SLPH |
SLPH |
SLPH |
Standard liters per day |
SLPD |
SLPD |
SLPD |
Special measurement unit |
SPECL |
special |
Special |
Define a special measurement unit for gas standard volume flow
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow > Special Units |
Field Communicator |
Configure > Manual Setup > Measurements > Special Units > Special GSV Units |
Configuration and Use Manual |
31 |
Configure process measurement
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1.Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.Calculate Gas Standard Volume Flow Conversion Factor as follows:
a.x base units = y special units
b.Gas Standard Volume Flow Conversion Factor = x ÷ y
4.Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
5.Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow unit.
6.Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1.Set Base Gas Standard Volume Unit to SCF.
2.Set Base Time Unit to minutes (min).
3.Calculate the conversion factor:
a.1 thousands of standard cubic feet per minute = 1000 cubic feet per minute
b.Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001 standard
4.Set Gas Standard Volume Flow Conversion Factor to 0.001.
5.Set Gas Standard Volume Flow Label to MSCFM.
32 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
6. Set Gas Standard Volume Total Label to MSCF.
4.3.4Configure Gas Standard Volume Flow Cutoff
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > GSV > GSV Cutoff |
Overview
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no upper limit.
Interaction between Gas Standard Volume Flow Cutoff and mA Output Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported via outputs and the gas standard volume flow values used in other transmitter behavior (e.g., events defined on gas standard volume flow).
mA Output Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
•mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
•Frequency Output Process Variable: Gas Standard Volume Flow Rate
•mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
•Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0 will be used in all internal processing.
Configuration and Use Manual |
33 |
Configure process measurement
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
•mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
•Frequency Output Process Variable: Gas Standard Volume Flow Rate
•mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
•Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
•If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
—The primary mA Output will report zero flow.
—The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all internal processing.
•If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be used in all internal processing.
4.4Configure Flow Direction
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Flow |
Field Communicator |
Configure > Manual Setup > Measurements > Flow > Flow Direction |
Overview
Flow Direction controls how forward flow and reverse flow affect flow measurement and reporting.
Flow Direction is defined with respect to the flow arrow on the sensor:
•Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
•Reverse flow (negative flow) moves in the direction opposite to the flow arrow on the sensor.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of the Flow Direction parameter.
Procedure
Set Flow Direction to the value you want to use.
The default setting is Forward.
34 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
4.4.1Options for Flow Direction
Flow Direction setting |
Relationship to Flow Direction arrow |
|
ProLink III |
Field Communicator |
on sensor |
Forward |
Forward |
Appropriate when the Flow Direction ar- |
row is in the same direction as the major- |
||
ity of flow. |
||
Reverse |
Reverse |
Appropriate when the Flow Direction ar- |
row is in the opposite direction from the |
||
majority of flow. |
||
Absolute Value |
Absolute Value |
Flow Direction arrow is not relevant. |
Bidirectional |
Bi directional |
Appropriate when both forward and re- |
verse flow are expected, and forward |
||
flow will dominate, but the amount of re- |
||
verse flow will be significant. |
||
Negate Forward |
Negate/Forward Only |
Appropriate when the Flow Direction ar- |
row is in the opposite direction from the |
||
majority of flow. |
||
Negate Bidirectional |
Negate/Bi-directional |
Appropriate when both forward and re- |
verse flow are expected, and reverse flow |
||
will dominate, but the amount of for- |
||
ward flow will be significant. |
||
Effect of Flow Direction on mA Outputs
Flow Direction affects how the transmitter reports flow values via the mA Outputs. The mA Outputs are affected by Flow Direction only if mA Output Process Variable is set to a flow variable.
Flow Direction and mA Outputs
The effect of Flow Direction on the mA Outputs depends on Lower Range Value configured for the mA Output:
•If Lower Range Value is set to 0, see Figure 4 1.
•If Lower Range Value is set to a negative value, see Figure 4 2.
Configuration and Use Manual |
35 |
Configure process measurement
Figure 4-1: Effect of Flow Direction on the mA Output: Lower Range Value = 0
Flow Direction = Forward |
Flow Direction = Reverse, Negate Forward |
Flow Direction = Absolute Value, Bidirectional, |
|||||||||
Negate Bidirectional |
|||||||||||
20 |
20 |
20 |
|||||||||
mA output |
12 |
mA output |
12 |
mA output |
12 |
||||||
4 |
4 |
4 |
|||||||||
-x |
0 |
x |
-x |
0 |
x |
-x |
0 |
x |
|||
Reverse flow |
Forward flow |
Reverse flow |
Forward flow |
Reverse flow |
Forward flow |
•Lower Range Value = 0
•Upper Range Value = x
Figure 4-2: Effect of Flow Direction on the mA Output: Lower Range Value < 0
Flow Direction = Forward |
Flow Direction = Reverse, Negate Forward |
Flow Direction = Absolute Value, Bidirectional, |
|||||||||
Negate Bidirectional |
|||||||||||
20 |
20 |
20 |
|||||||||
mA output |
12 |
mA output |
12 |
mA output |
12 |
||||||
4 |
4 |
4 |
|||||||||
-x |
0 |
x |
-x |
0 |
x |
-x |
0 |
x |
|||
Reverse flow |
Forward flow |
Reverse flow |
Forward flow |
Reverse flow |
Forward flow |
•Lower Range Value = −x
•Upper Range Value = x
Example: Flow Direction = Forward and Lower Range Value = 0
Configuration:
•Flow Direction = Forward
•Lower Range Value = 0 g/sec
•Upper Range Value = 100 g/sec
36 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Result:
•Under conditions of zero flow, the mA Output is 4 mA.
•Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA Output varies between 4 mA and 20 mA in proportion to the flow rate.
•Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the mA Output will be proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
Example: Flow Direction = Forward and Lower Range Value < 0
Configuration:
•Flow Direction = Forward
•Lower Range Value = −100 g/sec
•Upper Range Value = +100 g/sec
Result:
•Under conditions of zero flow, the mA Output is 12 mA.
•Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA Output varies between 12 mA and 20 mA in proportion to (the absolute value of) the flow rate.
•Under conditions of forward flow, if (the absolute value of) the flow rate equals or exceeds 100 g/sec, the mA Output is proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
•Under conditions of reverse flow, for flow rates between 0 and −100 g/sec, the mA Output varies between 4 mA and 12 mA in inverse proportion to the absolute value of the flow rate.
•Under conditions of reverse flow, if the absolute value of the flow rate equals or exceeds 100 g/sec, the mA Output is inversely proportional to the flow rate down to 3.8 mA, and will be level at 3.8 mA at higher absolute values.
Example: Flow Direction = Reverse
Configuration:
•Flow Direction = Reverse
•Lower Range Value = 0 g/sec
•Upper Range Value = 100 g/sec
Result:
•Under conditions of zero flow, the mA Output is 4 mA.
•Under conditions of reverse flow, for flow rates between 0 and +100 g/sec, the mA Output level varies between 4 mA and 20 mA in proportion to the absolute value of the flow rate.
•Under conditions of reverse flow, if the absolute value of the flow rate equals or exceeds 100 g/sec, the mA Output will be proportional to the absolute value of the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher absolute values.
Configuration and Use Manual |
37 |
Configure process measurement
Effect of flow direction on Frequency Outputs
Flow direction affects how the transmitter reports flow values via the Frequency Outputs. The Frequency Outputs are affected by flow direction only if Frequency Output Process Variable is set to a flow variable.
Table 4-1: Effect of the flow direction parameter and actual flow direction on
Frequency Outputs
Actual flow direction |
|||
Flow Direction setting |
Forward |
Zero flow |
Reverse |
Forward |
Hz > 0 |
0 Hz |
0 Hz |
Reverse |
0 Hz |
0 Hz |
Hz > 0 |
Bidirectional |
Hz > 0 |
0 Hz |
Hz > 0 |
Absolute Value |
Hz > 0 |
0 Hz |
Hz > 0 |
Negate Forward |
0 Hz |
0 Hz |
Hz > 0 |
Negate Bidirectional |
Hz > 0 |
0 Hz |
Hz > 0 |
Effect of flow direction on Discrete Outputs
The flow direction parameter affects the Discrete Output behavior only if Discrete Output Source is set to Flow Direction.
Table 4-2: Effect of the flow direction parameter and actual flow direction on
Discrete Outputs
Actual flow direction |
|||
Flow Direction setting |
Forward |
Zero flow |
Reverse |
Forward |
OFF |
OFF |
ON |
Reverse |
OFF |
OFF |
ON |
Bidirectional |
OFF |
OFF |
ON |
Absolute Value |
OFF |
OFF |
ON |
Negate Forward |
ON |
OFF |
OFF |
Negate Bidirectional |
ON |
OFF |
OFF |
Effect of flow direction on digital communications
Flow direction affects how flow values are reported via digital communications. The following table describes the effect of the flow direction parameter and actual flow direction on flow values reported via digital communications.
38 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Table 4-3: Effect of the flow direction on flow values
Actual flow direction |
||||
Flow Direction setting |
Forward |
Zero flow |
Reverse |
|
Forward |
Positive |
0 |
Negative |
|
Reverse |
Positive |
0 |
Negative |
|
Bidirectional |
Positive |
0 |
Negative |
|
Absolute Value |
Positive(1) |
0 |
Positive(1) |
|
Negate Forward |
Negative |
0 |
Positive |
|
Negate Bidirectional |
Negative |
0 |
Positive |
|
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Effect of flow direction on flow totals
Flow direction affects how flow totals and inventories are calculated.
Actual flow direction |
|||
Flow Direction setting |
Forward |
Zero flow |
Reverse |
Forward |
Totals increase |
Totals do not change |
Totals do not change |
Reverse |
Totals do not change |
Totals do not change |
Totals increase |
Bidirectional |
Totals increase |
Totals do not change |
Totals decrease |
Absolute Value |
Totals increase |
Totals do not change |
Totals increase |
Negate Forward |
Totals do not change |
Totals do not change |
Totals increase |
Negate Bidirectional |
Totals decrease |
Totals do not change |
Totals increase |
4.5Configure density measurement
The density measurement parameters control how density is measured and reported.
4.5.1Configure Density Measurement Unit
Display |
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS |
ProLink III |
Device Tools > Configuration > Process Measurement > Density |
Field Communicator |
Configure > Manual Setup > Measurements > Density > Density Unit |
Overview
Density Measurement Unit controls the measurement units that will be used in density calculations and reporting.
Configuration and Use Manual |
39 |
Configure process measurement
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Label |
|||
Unit description |
Display |
ProLink III |
Field Communicator |
Specific gravity(1) |
SGU |
SGU |
SGU |
Grams per cubic centimeter |
G/CM3 |
g/cm3 |
g/Cucm |
Grams per liter |
G/L |
g/l |
g/L |
Grams per milliliter |
G/mL |
g/ml |
g/mL |
Kilograms per liter |
KG/L |
kg/l |
kg/L |
Kilograms per cubic meter |
KG/M3 |
kg/m3 |
kg/Cum |
Pounds per U.S. gallon |
LB/GAL |
lbs/Usgal |
lb/gal |
Pounds per cubic foot |
LB/CUF |
lbs/ft3 |
lb/Cuft |
Pounds per cubic inch |
LB/CUI |
lbs/in3 |
lb/CuIn |
Degrees API |
D API |
degAPI |
degAPI |
Short ton per cubic yard |
ST/CUY |
sT/yd3 |
STon/Cuyd |
(1) Non standard calculation. This value represents line density divided by the density of water at 60 °F.
4.5.2Configure two-phase flow parameters
Display |
Not available |
|
ProLink III |
Device Tools > Configuration > Process Measurement > Density |
|
Field Communicator |
• Configure > Manual Setup > Measurements > Density > Slug Low Limit |
|
• |
Configure > Manual Setup > Measurements > Density > Slug High Limit |
|
• |
Configure > Manual Setup > Measurements > Density > Slug Duration |
|
Overview
The two-phase flow parameters control how the transmitter detects and reports twophase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is also referred to as slug flow.
40 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Configure process measurement
Procedure
1.Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density measurement.
The default value for Two-Phase Flow Low Limit is 0.0 g/cm³. The range is 0.0 to
10.0g/cm³.
2.Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value. Values above this will cause the transmitter to post Alert A105 (Two-Phase Flow).
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density measurement.
The default value for Two-Phase Flow High Limit is 5.0 g/cm³. The range is 0.0 to
10.0g/cm³.
3.Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase flow condition to clear before posting the alert.
The default value for Two-Phase Flow Timeout is 0.0 seconds, meaning that the alert will be posted immediately. The range is 0.0 to 60.0 seconds.
Detecting and reporting two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues. By configuring the two-phase flow parameters appropriately for your application, you can detect process conditions that require correction.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value.
A two-phase flow condition occurs whenever the measured density goes below Two-Phase Flow Low Limit or above Two-Phase Flow High Limit. If this occurs:
•A two-phase flow alert is posted to the active alert log.
•All outputs that are configured to represent flow rate hold their last pre alert value for the number of seconds configured in Two-Phase Flow Timeout.
If the two-phase flow condition clears before Two-Phase Flow Timeout expires:
•Outputs that represent flow rate revert to reporting actual flow.
Configuration and Use Manual |
41 |
Configure process measurement
•The two-phase flow alert is deactivated, but remains in the active alert log until it is acknowledged.
If the two-phase flow condition does not clear before Two-Phase Flow Timeout expires, the outputs that represent flow rate report a flow rate of 0.
If Two-Phase Flow Timeout is set to 0.0 seconds, the outputs that represent flow rate will report a flow rate of 0 as soon as two-phase flow is detected.
4.5.3Configure Density Damping
Display |
Not available |
ProLink III |
Device Tools > Configuration > Process Measurement > Density |
Field Communicator |
Configure > Manual Setup > Measurements > Density > Density Damping |
Overview
Density Damping controls the amount of damping that will be applied to the line density value.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Density damping affects all process variables that are calculated from line density.
Procedure
Set Density Damping to the value you want to use.
The default value is 1.6 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table:
Update Rate setting |
Damping range |
Normal |
0 to 51.2 seconds |
Special |
0 to 40.96 seconds |
Tips
•A high damping value makes the process variable appear smoother because the reported value changes slowly.
•A low damping value makes the process variable appear more erratic because the reported value changes more quickly.
42 |
Micro Motion Model 1700 Transmitters with Analog Outputs |
Loading…
Page 1 — Model 1700 and 2700
Installation Manual 20001700, Rev CC April 2013Micro Motion® Model 1700 and 2700Installation Manual
Page 2
The transmitter’s model number is on a tag on the side of the transmitter. You can use themodel number to determine the transmitter’s output
Page 3 — Contents
• You cannot configure the combination of Channel B as discrete output and Channel C asfrequency output.9.2 mA/HART wiring9.2.1 Basic mA output wiring
Page 4
HART/analog single loop wiringFigure 9-2: ABA.820 Ω maximum loop resistanceB. HART-compatible host or controller9.2.3HART multidrop wiringTipFor opt
Page 5 — 1 Planning
HART multidrop wiringFigure 9-3: ABCDEFA. 250–600 Ω resistanceB. HART-compatible host or controllerC. HART-compatible transmittersD. Model 2700 conf
Page 6
Output voltage versus load resistanceFigure 9-5: 16141210864200 500 1000 1500 2000 2500Load resistance (Ohms)High level output voltage (Volts)Maximu
Page 7
CAUTION!Exceeding 30 VDC can damage the transmitter. Terminal current must be less than 500 mA.Recommended pull-up resistor versus supply voltageFigur
Page 8
Internally powered frequency output wiring on Channel CFigure 9-8: AA. CounterOutput voltage versus load resistanceFigure 9-9: 16141210864200 1000
Page 9 — 1.1.2 Maximum cable lengths
Externally powered frequency output wiring on Channel CFigure 9-10: ABCA. Pull-up resistorB. 3–30 VDCC. CounterCAUTION!Exceeding 30 VDC can damage t
Page 10 — Planning
9.4 Discrete output wiring9.4.1 Internally powered discrete output wiring on Channel BInternally powered discrete output wiring on Channel BFigure 9-1
Page 11 — Power requirements
9.4.2 Externally powered discrete output wiring on Channel BExternally powered discrete output wiring on Channel BFigure 9-14: ABA. 3–30 VDCB. Pull-
Page 12 — 1.6 Orientation
Recommended pull-up resistor versus supply voltageFigure 9-15: 360032002800240020001600120080005 10 15 20 25 30Supply voltage (Volts)External pull-u
Page 13 — Installation Manual 9
Output options for Model 1500 and Model 2500 transmittersTable 1-4: Letter DescriptionA Analog outputs – one mA, one frequency, one RS-485B Configur
Page 14 — 10 Micro Motion
Output voltage versus load resistanceFigure 9-17: 16141210864200 1000 2000 3000 4000 5000Load resistance (Ohms)High level output voltage (Volts)Maxi
Page 15 — D. Sensor
CAUTION!Exceeding 30 VDC can damage the transmitter. Terminal current must be less than 500 mA.Recommended pull-up resistor versus supply voltageFigur
Page 16 — (optional)
Internally powered discrete input wiringFigure 9-20: AA. Switch9.5.2 Externally powered discrete input wiringExternally powered discrete input wirin
Page 17
10 SpecificationsTopics covered in this chapter:•Electrical connections•Input/output signals•Local display•Environmental limits•Physical specification
Page 18
10.2 Input/output signalsInput/output signals – Model 1700 transmitter with analog outputsTable 10-2: Type DescriptionOutput variables • Mass flow•
Page 19 — 3.1 Mounting options
Input/output signals – Model 2700 transmitter with intrinsically safeoutputsTable 10-5: Type DescriptionOutput variables • Mass & volume flow• N
Page 20 — 16 Micro Motion
Local display (standard)Table 10-7: Type DescriptionLocal display Standard user interface with 2-line LCD panel• Two optical switches for local oper
Page 21 — D. Mounting bracket
10.4 Environmental limitsEnvironmental specificationsTable 10-9: Type ValueAmbient temperature limits –40 to +140 °F (–40 to +60 °C)Humidity limits
Page 22 — 18 Micro Motion
4-wire remote mount transmitter dimensions (painted aluminumhousing)Figure 10-1: 2 13/16(71)2 13/16(71)4 × Ø3/8(10)3 11/16(93)3 × 1/2″–14 NPTor
Page 23 — Prepare the 4-wire cable
4-wire and 9-wire remote mount transmitter dimensions (stainless steelhousing)Figure 10-3: 2 13/16(71)4 x Ø3/8(10)2 13/16(71)5 7/16(139)2 × 1/2″
Page 24 — 20 Micro Motion
• 18 to 100 VDC, 6 watts typical, 11 watts maximum• Complies with low voltage directive 2006/95/EC per EN 61010-1 (IEC 61010-1) withamendment 2, and I
Page 25
Remote core processor dimensionsFigure 10-4: 2 13/16(71)2 13/16(71)4 × Ø3/8(10)6 3/16(158)2 1/4(57)4 9/16(116)wall mount5 1/2 (140)To centerline of
Page 26
Index4-wire cablepreparation 19, 54types 21, 56user-supplied 21, 569-wire cableconnecting to sensor 37, 40, 65, 69preparation 31, 59types an
Page 27 — Installation Manual 23
frequency output 86frequency output with galvanic isolator 91hazardous area 87, 92mA output hazardous area 89MmA outputhazardous area wiring
Page 29
*20001700*20001700Rev CC2013Micro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301T +1 303-527-5200T +1 800-522-6277
Page 30
1.7 Accessibility for maintenanceMount the flowmeter in a location and orientation that satisfies the following conditions:• Allows sufficient clearan
Page 32 — 28 Micro Motion
2 Mounting and sensor wiring forintegral installationsTopics covered in this chapter:•Mounting and sensor wiring•Rotate the transmitter on the sensor
Page 33 — 4.1 Mounting options
1. Loosen each of the four cap screws (4 mm) that fasten the transmitter to the base.2. Rotate the transmitter counter-clockwise so that the cap screw
Page 34 — A. Mounting bracket
Display componentsFigure 2-2: A. Transmitter housingB. Sub-bezelC. Display moduleD. Display screwsE. End-cap clampF. Cap screwG. Display cover1. Shu
Page 35 — Prepare the 9-wire cable
10. Turn the display cover clockwise until it is snug.11. Replace the end-cap clamp by reinserting and tightening the cap screw.12. Restore power to t
Page 36 — 32 Micro Motion
3 Mounting and sensor wiring for 4-wire remote installationsTopics covered in this chapter:•Mounting options•Prepare the 4-wire cable•Wire the transmi
Page 37 — Installation Manual 33
Safety messagesSafety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefullybefore procee
Page 38
Components of 4-wire remote mount transmitter (aluminumhousing)Figure 3-1: A. End capB. Cap screwsC. TransmitterD. Mounting bracketMounting and sens
Page 39 — Installation Manual 35
Components of a 4-wire remote mount transmitter (stainless steelhousing)Figure 3-2: ABDCA. End capB. Cap screwsC. TransmitterD. Mounting bracket2. A
Page 40 — 36 Micro Motion
Components of 4-wire remote mount transmitter (aluminumhousing)Figure 3-3: A. End capB. Cap screwsC. TransmitterD. Mounting bracketMounting and sens
Page 41
Components of a 4-wire remote mount transmitter (stainless steelhousing)Figure 3-4: ABDCA. End capB. Cap screwsC. TransmitterD. Mounting bracket2. A
Page 42
4-wire cable preparationFigure 3-5: Cable layoutRun conduit to sensorMetal conduitWrap the drain wires twice around the shield and cut off the exces
Page 43 — Installation Manual 39
4-wire cable shieldingFigure 3-6: Assemble the Gland1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring.2.
Page 44
• Twisted pair construction.• Applicable hazardous area requirements, if the core processor is installed in ahazardous area.• Wire gauge appropriate f
Page 45
Wiring path for transmitters with aluminum housingFigure 3-7: A. 4-wire cableB. Mating connectorMounting and sensor wiring for 4-wire remote install
Page 46
Wiring path for transmitters with stainless steel housingFigure 3-8: AVDC+VDC–RS-485ARS-485BBA. 4-wire cableB. Mating connector3.4Rotate the user in
Page 47
Display componentsFigure 3-9: A. Transmitter housingB. Sub-bezelC. Display moduleD. Display screwsE. End-cap clampF. Cap screwG. Display cover1. Shu
Page 49
10. Turn the display cover clockwise until it is snug.11. Replace the end-cap clamp by reinserting and tightening the cap screw.12. Restore power to t
Page 50
Transmitter external grounding screwFigure 3-11: Mounting and sensor wiring for 4-wire remote installationsInstallation Manual 27
Page 51
Mounting and sensor wiring for 4-wire remote installations28 Micro Motion® Model 1700 and 2700
Page 52 — 48 Micro Motion
4 Mounting and sensor wiring for 9-wire remote installationsTopics covered in this chapter:•Mounting options•Prepare the 9-wire cable•Wire the transmi
Page 53 — 5.1 Mounting options
Components of 9-wire remote mount transmitterFigure 4-1: ABCA. Mounting bracketB. Cap screwsC. Transmitter2. Attach the mounting bracket to the wall
Page 54 — 50 Micro Motion
Components of 9-wire remote mount transmitterFigure 4-2: ABCA. Mounting bracketB. Cap screwsC. Transmitter2. Attach the mounting bracket to an instr
Page 55
Preparing jacketed cableFigure 4-3: 1. Trim 4 inches (100 mm) of cable jacket.2. Remove the clear wrap and filler material.3. Remove the foil that i
Page 56 — 52 Micro Motion
Preparing shielded or armored cableFigure 4-4: 1. Without cutting the shield, strip 9 inches (225 mm) of cable jacket.2. Strip 8 ½ inches (215 mm) o
Page 57
Cable typesMicro Motion supplies three types of 9-wire cable: jacketed, shielded, and armored. Notethe following differences between the cable types:•
Page 60
Cross-section view of shielded cableFigure 4-6: AC (1)BDE (4)F (4)G (5)A. Outer jacketB. Tin-plated copper braided shieldC. Foil shield (1 total)D.
Page 61
4.3 Wire the transmitter to the sensor usingjacketed cableFor ATEX installations, the jacketed cable must be installed inside a user-supplied sealedme
Page 62 — 58 Micro Motion
Sensor and transmitter terminal designationsTable 4-5: Wire color Sensor terminal Transmitter terminal FunctionBlack No connection 0 Drain wiresBrow
Page 63
F-Series, Model D, and Model DL sensor terminalsFigure 4-9: Mounting and sensor wiring for 9-wire remote installationsInstallation Manual 39
Page 64 — 60 Micro Motion
Model 2700 transmitter terminalsFigure 4-10: ABCDEFGHIJKA. BrownB. VioletC. YellowD. OrangeE. GrayF. BlueG. WhiteH. GreenI. RedJ. Mounting screwK. G
Page 65 — Installation Manual 61
CAUTION!Improperly sealed housings can expose electronics to moisture, which can cause measurementerror or flowmeter failure. Install drip legs in con
Page 66
Cross-section of assembled cable gland with cableFigure 4-12: ABCDE FGAA. CableB. Sealing nutC. SealD. Compression nutE. Braided shieldF. Brass comp
Page 67 — Installation Manual 63
Sensor and transmitter terminal designations (continued)Table 4-6: Wire color Sensor terminal Transmitter terminal FunctionGray 8 8 Right pickoff –W
Page 68 — 64 Micro Motion
F-Series, Model D, and Model DL sensor terminalsFigure 4-14: Mounting and sensor wiring for 9-wire remote installations44 Micro Motion® Model 1700 a
Page 69
Model 2700 transmitter terminalsFigure 4-15: ABCDEFGHIJKA. BrownB. VioletC. YellowD. OrangeE. GrayF. BlueG. WhiteH. GreenI. RedJ. Mounting screwK. G
Page 70
1 PlanningTopics covered in this chapter:•Flowmeter components•Outputs option identification•Environmental limits•Hazardous area classifications•Power
Page 71 — Installation Manual 67
Display componentsFigure 4-16: A. Transmitter housingB. Sub-bezelC. Display moduleD. Display screwsE. End-cap clampF. Cap screwG. Display cover1. Sh
Page 72
10. Turn the display cover clockwise until it is snug.11. Replace the end-cap clamp by reinserting and tightening the cap screw.12. Restore power to t
Page 73
Transmitter external ground screwFigure 4-18: Mounting and sensor wiring for 9-wire remote installations48 Micro Motion® Model 1700 and 2700
Page 74
5 Mounting and sensor wiring forremote core processor with remotesensor installationsTopics covered in this chapter:•Mounting options•Mount the remote
Page 75 — Installation Manual 71
Components of 4-wire remote mount transmitter (aluminumhousing)Figure 5-1: A. End capB. Cap screwsC. TransmitterD. Mounting bracketMounting and sens
Page 76 — 72 Micro Motion
Components of a 4-wire remote mount transmitter (stainless steelhousing)Figure 5-2: ABDCA. End capB. Cap screwsC. TransmitterD. Mounting bracket2. A
Page 77 — Installation Manual 73
Components of 4-wire remote mount transmitter (aluminumhousing)Figure 5-3: A. End capB. Cap screwsC. TransmitterD. Mounting bracketMounting and sens
Page 78
Components of a 4-wire remote mount transmitter (stainless steelhousing)Figure 5-4: ABDCA. End capB. Cap screwsC. TransmitterD. Mounting bracket2. A
Page 79
a. Loosen each of the four cap screws (4 mm).b. Rotate the bracket so that the core processor is oriented as desired.c. Tighten the cap screws, torqui
Page 80
4-wire cable preparationFigure 5-6: Cable layoutRun conduit to sensorMetal conduitWrap the drain wires twice around the shield and cut off the exces
Page 81 — 6 Wiring the power supply
Integral installationFigure 1-1: TransmitterSensor• High-temperature flexible conduit – Some high-temperature meters comepreinstalled with a flexibl
Page 82 — 78 Micro Motion
4-wire cable shieldingFigure 5-7: Assemble the Gland1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring.2.
Page 83 — 7.1 Basic analog wiring
• Twisted pair construction.• Applicable hazardous area requirements, if the core processor is installed in ahazardous area.• Wire gauge appropriate f
Page 84 — 80 Micro Motion
4. Connect wires to the appropriate terminals on the mating connector.ImportantNever ground the shield, braid, or drain wire(s) at the transmitter.Tip
Page 85
Wiring path for transmitters with stainless steel housingFigure 5-10: AVDC+VDC–RS-485ARS-485BBA. 4-wire cableB. Mating connector5.5Prepare the 9-wir
Page 86 — 82 Micro Motion
Preparing jacketed cableFigure 5-11: 1. Trim 4 inches (100 mm) of cable jacket.2. Remove the clear wrap and filler material.3. Remove the foil that
Page 88 — External resistor R
Cable typesMicro Motion supplies three types of 9-wire cable: jacketed, shielded, and armored. Notethe following differences between the cable types:•
Page 89
Bend radii of armored cableTable 5-5: Jacket material Outside diameter Minimum bend radiiStatic (no load) condition Under dynamic loadPVC 0.525 inch
Page 90 — 86 Micro Motion
Cross-section view of shielded cableFigure 5-14: AC (1)BDE (4)F (4)G (5)A. Outer jacketB. Tin-plated copper braided shieldC. Foil shield (1 total)D.
Page 91 — 8.5 Hazardous area wiring
5.6 Wire the remote core processor to the sensorusing jacketed cableFor ATEX installations, the jacketed cable must be installed inside a user-supplie
Page 92
4-wire remote installation – painted aluminum housingFigure 1-3: SensorCore processorTransmitter4-wire cable4-wire remote installation – stainless s
Page 93 — Installation Manual 89
Sensor and remote core processor terminal designations(continued)Table 5-6: Wire color Sensor terminal Remote core processor terminal FunctionOrange
Page 94
ELITE, H-Series, T-Series, and some F-Series sensor terminalsFigure 5-16: DIHFEABCGA. VioletB. YellowC. OrangeD. BrownE. WhiteF. GreenG. RedH. GrayI
Page 95 — Installation Manual 91
Model DT sensor terminals (user-supplied metal junction box withterminal block)Figure 5-18: 198765432AA. Earth groundRemote core processor terminals
Page 96 — 92 Micro Motion
5.7 Wire the remote core processor to the sensorusing shielded or armored cableFor ATEX installations, shielded or armored cable must be installed wit
Page 97 — Installation Manual 93
3. Screw the nipple into the conduit opening for the 9-wire cable. Tighten it to one turnpast hand-tight.4. Slide the compression ring, compression nu
Page 98 — 94 Micro Motion
Sensor and remote core processor terminal designationsTable 5-7: Wire color Sensor terminal Remote core processor terminal FunctionBlack No connecti
Page 99 — 9 I/O wiring for Model 2700
ELITE, H-Series, T-Series, and some F-Series sensor terminalsFigure 5-22: DIHFEABCGA. VioletB. YellowC. OrangeD. BrownE. WhiteF. GreenG. RedH. GrayI
Page 100 — 9.2 mA/HART wiring
Model DT sensor terminals (user-supplied metal junction box withterminal block)Figure 5-24: 198765432AA. Earth groundRemote core processor terminals
Page 101 — HART multidrop wiring
5.8 Rotate the user interface on the transmitter(optional)The user interface on the transmitter electronics module can be rotated 90º or 180° fromthe
Page 103 — Installation Manual 99
9-wire remote installation typeFigure 1-5: TransmitterJunction boxSensor9-wire cable• Remote core processor with remote sensor – A remote core proce
Page 104 — 100 Micro Motion
Transmitter external grounding screwFigure 5-28: 3. Ground the remote core processor according to applicable local standards, usingthe remote core p
Page 105 — Installation Manual 101
6 Wiring the power supply6.1 Wire the power supplyA user-supplied switch may be installed in the power supply line. For compliance with low-voltage di
Page 107 — 9.4 Discrete output wiring
7 I/O wiring for Model 1700 andModel 2700 transmitters with analogoutputsTopics covered in this chapter:•Basic analog wiring•HART/analog single loop w
Page 108 — 104 Micro Motion
7.2 HART/analog single loop wiringNoteFor HART communications:•600 Ω maximum loop resistance•250 Ω minimum loop resistanceHART/analog single loop wiri
Page 109 — Installation Manual 105
7.3 RS-485 point-to-point wiringRS-485 point-to-point wiringFigure 7-3: ABCA. Other devicesB. Primary controllerC. Multiplexer7.4HART multidrop wiri
Page 110 — 106 Micro Motion
HART multidrop wiringFigure 7-4: ABCDEFA. 250–600 Ω resistanceB. HART-compatible host or controllerC. HART-compatible transmittersD. Model 1700 or M
Page 111 — 9.5 Discrete input wiring
8 I/O wiring for Model 1700 andModel 2700 transmitters withintrinsically safe outputsTopics covered in this chapter:•Safe area mA output wiring•Safe a
Page 112
Safe area mA output load resistance valuesFigure 8-2: 1000900800700600500400300200100012 14 16 18 20 22 24 26 28 30OPERATING REGIONSupply voltage VD
Page 113 — 10 Specifications
Safe area mA output load resistance valuesFigure 8-4: 1000900800700600500400300200100012 14 16 18 20 22 24 26 28 30OPERATING REGIONSupply voltage VD
Page 114 — 10.2 Input/output signals
Remote core processor with remote sensor installation typeFigure 1-6: Core processorTransmitter4-wire cable9-wire cableSensorJunction box1.1.2 Maxim
Page 115 — 10.3 Local display
Safe area HART multidrop wiringFigure 8-5: ABCDEFA. 250–600 Ω resistanceB. HART-compatible host or controllerC. HART-compatible transmittersD. Model
Page 116 — 112 Micro Motion
Safe area frequency output/discrete output load resistance valuesFigure 8-7: 1000090008000700060005000400030002000100005 7 9 11 13 15 17 19 21 23OPE
Page 118 — Specifications
8.5.1 Hazardous area mA output wiringHazardous area mA output wiringFigure 8-8: ABCDEHazardous area Safe areaA. VinB. VoutC. GroundD. RloadE. Rbarri
Page 119 — Installation Manual 115
Safe area mA output load resistance valuesFigure 8-9: 1000900800700600500400300200100012 14 16 18 20 22 24 26 28 30OPERATING REGIONSupply voltage VD
Page 120 — 116 Micro Motion
8.5.2 Hazardous area frequency/discrete output wiring usinggalvanic isolatorHazardous area frequency/discrete output wiring using galvanicisolatorFigu
Page 121 — Installation Manual 117
8.5.3 Hazardous area frequency/discrete output wiring usingbarrier with external load resistanceHazardous area frequency/discrete output wiring using
Page 122 — 118 Micro Motion
Safe area frequency output/discrete output load resistance valuesFigure 8-12: 1000090008000700060005000400030002000100005 7 9 11 13 15 17 19 21 23OP
Page 123
I/O wiring for Model 1700 and Model 2700 transmitters with intrinsically safe outputs94 Micro Motion® Model 1700 and 2700
Page 124 — *20001700*
9 I/O wiring for Model 2700transmitters with configurable input/outputsTopics covered in this chapter:•Channel configuration•mA/HART wiring•Frequency
(Ocr-Read Summary of Contents of some pages of the Emerson Micro Motion 1700 Document (Main Content), UPD: 05 July 2023)
-
38, • The device being polled (slave) cannot have special units set for density. Otherwise, the master will reject the base density and report the following alarm: A115: No External Input or Polled Data Alert • On the slave side, setup the HART Primary Variable for Base Density. The master will reject anything other than Base Density for the HART Primary Variable and trigger an A115 alarm. • The density units on the transmitter and the polled …
-
157, Alert num- ber Alert title Possible cause Recommended actions A002 RAM Error (Core Pro- cessor) The core processor has experienced a memory error. • Cycle power to the meter. • Replace the core processor. • Contact customer support. A003 No Sensor Response The transmitter is not receiving one or more basic electrical signals from the sensor. This alert often occurs in conjunc- tion with Alert 102. • Check the drive gain…
-
183, Emerson Micro Motion 1700 10.22 Check Flow Direction If Flow Direction is set inappropriately for your process, the transmitter may report flow data that is not appropriate for your requirements. The Flow Direction parameter interacts with actual flow direction to affect flow values, flow totals and inventories, and output behavior. For the simplest operation, actual process flow should match the flow arrow that is on the side of the sensor case. Procedure 1. Verify the actual direction of process flow through the …
-
93, Emerson Micro Motion 1700 Important If you assign Flow Switch to the Discrete Output, you should also configure Flow Switch Variable, Flow Switch Setpoint, and Hysteresis. Related information Configure an enhanced event Fault indication with the Discrete Output Configure Flow Switch parameters Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > CONFIG FL SW ProLink III Device Tools > Configuration > I/O > Outputs > Disc…
-
81, • If you have configured Display Variable 1 to track mA Output Process Variable, be aware that changing the configuration of mA Output Process Variable will change the contents of Display Variable 1. Procedure Set mA Output Process Variable as desired. The default setting is Mass Flow Rate. Postrequisites If you changed the setting of mA Output Process Variable, verify the settings of Lower Range Value (LRV) and Upper Range Value (URV). Options …
-
91, Emerson Micro Motion 1700 Options for Frequency Output Fault Action Options for Frequency Output Fault Action Table 6-5: Label Frequency Output behavior Upscale Goes to configured Upscale value: • Range: 10 Hz to 15000 Hz • Default: 15000 Hz Downscale 0 Hz Internal Zero 0 Hz None (default) Tracks data for the assigned process variable; no fault action CAUTION! If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to …
-
63, Option Description Enabled The display automatically scrolls through each display variable as specified by Scroll Rate. The operator can move to the next display variable at any time using Scroll. Disabled (de- fault) The display shows Display Variable 1 and does not scroll automatically. The operator can move to the next display variable at any time using Scroll. 2. If you enabled Auto Scroll, set …
-
19, Part II Configuration and commissioning Chapters covered in this part: • Introduction to configuration and commissioning • Configure process measurement • Configure device options and preferences • Integrate the meter with the control system • Complete the configuration Configuration and commissioning Configuration and Use Manual 11
… -
92, Emerson Micro Motion 1700 6.4.1 Configure Discrete Output Source Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > DO SRC ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Assignment Overview Discrete Output Source controls which device condition or process condition is reported via the Discrete Output. Procedure Set Discrete Output Source to the…
-
69, Effects of Update Rate = Special Incompatible features and functions Special mode is not compatible with the following features and functions: • Enhanced events. Use basic events instead. • All calibration procedures. • Zero verification. • Restoring the factory zero or the prior zero. If required, you can switch to Normal mode, perform the desired procedures, and then return to Special mode. Proce…
-
240, Transmitter components and installation wiring 232 Micro Motion Model 1700 Transmitters with Analog Outputs
… -
243, September 2006, version 5.x Modification type Change Expansion • Discrete Output assignable as a flow switch • Discrete Output fault indication configurability • Discrete Input support for multiple action assignments • Added support for querying the display LED status via Mod- bus • Additional HART and Modbus commands • Process comparator expanded to five configurable events • Factory configuration restore function • Factory zero restore function • Alarm history expanded �…
-
48, Procedure Set Density Measurement Unit to the option you want to use. The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter). Options for Density Measurement Unit The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels. Unit description Label Display ProLink III Field Communicator Specific gravity (1) SGU SGU SGU Grams per cubic…
-
231, Transmitter default values and ranges (continued)Table D-1: Type Parameter Default Range Comments Special units Base mass unit g Base mass time sec Mass flow conversion factor 1 Base volume unit L Base volume time sec Volume flow conversion factor 1 Variable map- ping Primary variable Mass flow Secondary variable Volume flow Tertiary variable Mass flow Quaternary variable Volume flow mA Output 1 Primary variable Mass flow LRV –200.00…
-
115, 8 Transmitter operation Topics covered in this chapter: • Record the process variables • View process variables • View transmitter status using the status LED • View and acknowledge status alerts • Read totalizer and inventory values • Start and stop totalizers and inventories • Reset totalizers • Reset inventories 8.1 Record the process variables Micro Motion suggests that you make a reco…
-
216, Connection over multidrop networkFigure B-4: B A C D A. Signal converter B. 250–600 Ω resistance C. Devices on the network D. Master device 5. Start ProLink III. 6. Choose Connect to Physical Device. 7. Set Protocol to HART Bell 202. Tip HART/Bell 202 connections use standard connection parameters. You do not need to configure them here. 8. If you are using a USB signal converter, enable Toggle R…
Table of Contents for Emerson Micro Motion 1700:
-
4.5 Configure density measurement ……………………………………………………………………………………. 45 4.5.1 Configure Density Measurement Unit …………………………………………………………………46 4.5.2 Configure slug flow parameters …………………………………………………………………………47 4.5.3 Configure Density Damping ………………………………………
-
View test result data using ProLink II 1. Choose Tools > Meter Verification > Run Meter Verification and click View Previous Test Results and Print Report. The chart shows test results for all tests stored in the ProLink II database. 2. (Optional) Click Next to view and print a test report. 3. (Optional) Click Export Data to CSV File to save the data to a file on your PC. View test result data using ProLink III 1. Choose Device Tools > Diagnostics > Meter Verification and click Previous T
-
5.4.1 Configure Update Rate Display Not available ProLink II ProLink > Configuration > Device > Update Rate ProLink III Device Tools > Configuration > Process Measurement > Response > Update Rate Field Communicator Configure > Manual Setup > Measurements > Update Rate Overview Update Rate controls the rate at which process data is polled and process variables are calculated. Update Rate = Special produ
-
Connection over local loopFigure C-3: A C D E R1 R3 R2 B A. PC B. Signal converter C. Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements D. DCS or PLC E. Transmitter, with wiring compartment and power supply compartment opened Note This figure shows a serial port connection. USB connections are also supported. 4. To connect over a HART multidrop network: a. Attach the leads from the signal converter to any point on the network. b. Add resistance as necessary. Important HART/Bel
-
Overview Digital Communications Fault Action specifies the values that will be reported via digital communications if the transmitter encounters an internal fault condition. Procedure Set Digital Communications Fault Action as desired. The default setting is None. Options for Digital Communications Fault Action Options for Digital Communications Fault ActionTable 6-15: Label Description ProLink II ProLink III Field Communicator Upscale Upscale Upscale �
-
• Current flowmeter identification data • Current flow and density configuration parameters • Current zero values • Current process values for mass flow rate, volume flow rate, density, temperature, and external pressure • Customer and test descriptions (if entered by the user) If you use ProLink II or ProLink III to run a test, a test result chart and a test report are displayed at the completion of the test. On-screen directions are provided to manipulate the test data or export the data to a CSV file for offline analysis. View
-
Effect of Flow Direction on frequency outputs Flow Direction affects how the transmitter reports flow values via the frequency outputs. The frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set to a flow variable. Effect of the Flow Direction parameter and actual flow direction on frequency outputs Table 4-6: Flow Direction setting Actual flow direction Forward Zero flow Reverse Forward
-
f. Test the gray terminal against all other terminals except the blue one. g. Test the orange terminal against all other terminals except the yellow and violet ones. h. Test the yellow terminal against all other terminals except the orange and violet ones. i. Test the violet terminal against all other terminals except the yellow and orange ones. There should be infinite resistance for each pair. If there is any resistance at all, there is a short between terminals. Postrequisites To return to normal operati
-
Postrequisites For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be controlled by the configuration that is stored for the selected input or output type, and the stored configuration may not be appropriate for your process. After verifying channel an
-
Using the display to view and acknowledge the status alarmsFigure 8-2: SEE ALARM Yes Scroll and Select simultaneously for 4 seconds ACK ALL Yes EXIT Select No Alarm code Scroll ACK Yes Select No Active/ unacknowledged alarms? NoYes Select NO ALARM EXIT Scroll Scroll Select Scroll ScrollSelect Is ACK ALL enabled? Yes No Transmitter operation 124 Micro Motion ® Model 1700 Transmitters with Analog Outputs
-
Configuration menu (continued)Figure B-13: ProLink > Configuration Device • Model • Manufacturer • Hardware Rev • Distributor • Software Rev • ET O • CP Software Rev • CP ETO • Option Board • Firmware Checksum • CP Firmware Checksum • Tag • Date • Descriptor • Message • Sensor type • Transmitter Serial • Floating PT Ordering • Add Comm Resp Delay • Restore Factory Configuration • Digital Comm Fault Set
-
Q quaternary variable (QV) 105 R radio frequency interference (RFI) troubleshooting 194 Rate Factor 91 reference density, See standard density refresh rate display 62 Response Time 70 S safety messages ii scaling frequency outputs 90 mA outputs 83 Scroll Rate 63 secondary variable (SV) 105 security access to display menus 67 sensor coils troubleshooting 201 Sensor Flange Type 79 Sens
Questions, Opinions and Exploitation Impressions:
You can ask a question, express your opinion or share our experience of Emerson Micro Motion 1700 device using right now.