Operating manual
Cond 3110
Conductivity meter
ba75792e02
11/2008
Related Manuals for wtw Cond 3110
Summary of Contents for wtw Cond 3110
-
Page 1
Operating manual Cond 3110 Conductivity meter ba75792e02 11/2008… -
Page 2
© Copyright Weilheim 2008, WTW GmbH Reproduction in whole — or even in part — is prohibited without the express written permission of WTW GmbH, Weilheim. Printed in Germany. ba75792e02 11/2008… -
Page 3: Table Of Contents
Contents Cond 3110 — Contents Overview ……..5 Keypad .
-
Page 4
Contents Cond 3110 What to do if……..33 Technical data . -
Page 5: Overview
Cond 3110 Overview Overview The Cond 3110 compact precision conductivity meter enables you to perform conductivity measurements quickly and reliably. The Cond 3110 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications. The proven procedures for determining or adjusting the cell constant support your work with the conductivity meter.
-
Page 6: Keypad
Overview Cond 3110 Keypad Cond 3110 In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g.
-
Page 7: Display
Cond 3110 Overview Display Measured parameter Sensor symbol S/cm (calibration µ evaluation) Measured value 1/cm Time display C ° °F Temperature Tref20 Tref25 display, function ARng LoBat display indicators, status display indicators Status display Stability control (AutoRead) is active indicators ARng Automatic range switching;…
-
Page 8: Socket Field
Overview Cond 3110 Socket field Connectors: Conductivity measuring cell Service interface Caution Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary measuring cells fulfill these conditions.
-
Page 9: Safety
Cond 3110 Safety Safety This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual before working with the meter. The operating manual must always be available within the vicinity of the meter.
-
Page 10: Authorized Use
Safety Cond 3110 Authorized use Authorized use of the meter consists exclusively of the measurement of conductivity, temperature and salinity in a laboratory or field environment. The technical specifications as given in chapter 7 T must ECHNICAL DATA be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized.
-
Page 11
Safety datasheets of the chemical manufacturers. Caution In addition to the safety instructions mentioned here, also follow the safety instructions of the sensors used. The operating manuals of the sensors are available on the supplied CD and on the Internet under www.WTW.com. ba75792e02 11/2008… -
Page 12
Safety Cond 3110 ba75792e02 11/2008… -
Page 13: Commissioning
Cond 3110 Commissioning Commissioning Scope of delivery Conductivity meter Cond 3110 4 batteries 1.5 V Mignon type AA Short instructions CD-ROM with detailed operating manual Initial commissioning Perform the following activities: Insert the supplied batteries Switch on the meter. 3.2.1 Inserting the batteries Unscrew the two screws (1) on the underside of the meter.
-
Page 14: Switching On The Meter
Commissioning Cond 3110 Mignon AA). In order to charge the batteries, an external charging device is required. Caution Make sure that the poles of the batteries are the right way round. ± ± signs on the batteries must correspond to the signs in the battery compartment.
-
Page 15: Operation
Cond 3110 Operation Operation General operating principles This section contains basic information on the operation of the Cond 3110. 4.1.1 Operating modes The meter has the following operating modes: Measurement The display indicates the measurement data in the measured value…
-
Page 16: Measuring
Operation Cond 3110 Measuring Preparatory activities Perform the following preparatory activities when you want to measure: Connect a measuring cell to the meter. Calibrate or check the meter with the measuring cell. Select the measured parameter with <M>. Stability control…
-
Page 17: Measuring The Conductivity
Cond 3110 Operation 4.2.1 Measuring the conductivity Perform the preparatory activities according to section 4.2. Immerse the conductivity measuring cell in the test sample. If necessary, scroll with <M> until the measured parameter with the unit mS/cm or µS/cm is displayed.
-
Page 18
Operation Cond 3110 C ° ARng ba75792e02 11/2008… -
Page 19: Determining/Setting Up The Cell Constant [C]
Cond 3110 Operation Determining/setting up the cell constant [C] Why determine/set up Due to aging, the cell constant slightly changes. As a result, an inexact the cell constant? measured value is displayed. Calibration determines the current value of the cell constant and stores this value in the meter.
-
Page 20
Operation Cond 3110 1/cm Tref25 Immerse the measuring cell in the control standard solution, 0.01 mol/l KCI. Start the calibration with <ENTER>. The determination of the cell constant with stability control starts. The display indicator flashes until there is a stable signal. -
Page 21: Using The Last Calibrated Cell Constant
Cond 3110 Operation Calibration evaluation After the calibration, the meter automatically evaluates the current status. The evaluation appears on the display. Display Cell constant [cm in the range 0.450 … 0.500 cm 0.800 … 0.880 cm You are working with a correctly calibrated measuring cell.
-
Page 22: Setting The Cell Constant Manually
Operation Cond 3110 If necessary, press <CAL> repeatedly until CAL and the last calibrated cell constant is displayed. 1/cm Tref25 Confirm the selection with <ENTER>. The displayed cell constant is used. The meter switches to the measured value display. 4.3.3 Setting the cell constant manually…
-
Page 23
Cond 3110 Operation S/cm µ 1/cm Tref25 ARng Set the cell constant to be used with <▲><▼>, e.g. 0.846 cm S/cm µ 1/cm Tref25 ARng Confirm the selection with <ENTER>. The new cell constant is used from now on. The meter switches to the measured value display. -
Page 24
Operation Cond 3110 Selecting the cell Press the <CAL> key repeatedly until USE CELL is displayed. constant 0.475 cm Confirm the selection with <ENTER> or <CAL >. If necessary, press <CAL> repeatedly until the cell constant 0.475 cm is displayed. -
Page 25: Temperature Compensation Tc
Cond 3110 Operation Temperature compensation TC The calculation of the temperature compensation is based on the preset reference temperature, Tref 20 or Tref 25 (see section 4.5 S ETTINGS As the temperature compensation, the nonlinear temperature compensation «nLF» according to DIN 38404 or EN 27 888 respectively is permanently set.
-
Page 26: Settings
Operation Cond 3110 Settings You can adapt the meter to your individual requirements. The settings are done in the following menus: System settings (<ENTER >) – Switch-off interval (tOff) Measurement settings (<M >) – Reference temperature (Tref25 or Tref20) – Temperature unit (°C / °F) –…
-
Page 27: Measurement Settings
Cond 3110 Operation 4.5.2 Measurement settings These settings apply to the determination of the cell constant and measurement (the default condition is printed in bold). Reference temperature t25, t20 Temperature unit (UnI) °C, °F Cleaning interval (Int.C) 0 … 180 … 999 d Reference temperature Open the menu for measurement settings with <M…
-
Page 28
Operation Cond 3110 Using <▲><▼>, toggle between °C and °F. Confirm with <ENTER>. Int.C, the setting of the cleaning interval is displayed. Cleaning interval (Int.C) Time Set the interval with <▲><▼>. Confirm with <ENTER>. The measurement settings are completed. The meter switches to the measuring mode. -
Page 29: Reset
Cond 3110 Operation Reset 4.6.1 Resetting the cell constant This function serves to erase the last determined cell constant. Subsequently, the meter uses the last manually set cell constant in the range 0.800 … 0.880 cm or the fixed cell constant, 0.475 cm Based on the last erased cell constant the meter decides to which of the two manually set cell constants the cell constant is reset.
-
Page 30: Resetting All Meter Settings
Operation Cond 3110 4.6.2 Resetting all meter settings This function resets all meter settings to the default condition. The relevant values are given in the following sections: System settings section 4.5.1 Measurement settings section 4.5.2 The following settings are also reset to the default condition:…
-
Page 31: Maintenance, Cleaning, Disposal
Cond 3110 Maintenance, cleaning, disposal Maintenance, cleaning, disposal Maintenance The only maintenance activity required is replacing the batteries. Note See the relevant operating manuals of the measuring cells for instructions on maintenance. 5.1.1 Replacing the batteries Unscrew the two screws (1) on the underside of the meter, Open the battery compartment (2) on the underside of the meter.
-
Page 32: Cleaning
Maintenance, cleaning, disposal Cond 3110 Caution Make sure that the poles of the batteries are the right way round. ± ± signs on the batteries must correspond to the signs in the battery compartment. Close the battery compartment (2) and tighten the screws (1).
-
Page 33: What To Do If
Cond 3110 What to do if… What to do if… Error message Cause Remedy OFL, UFL – Measured value outside the – Use suitable measuring cell measuring range Error message, Cause Remedy – Measuring cell contaminated – Clean cell and replace it if necessary –…
-
Page 34
What to do if… Cond 3110 ba75792e02 11/2008… -
Page 35: Technical Data
Cond 3110 Technical data Technical data General data Dimensions approx. 180 x 80 x 55 mm Weight approx. 0.4 kg Mechanical structure Type of protection IP 67 Electrical safety Protective class Test certificates Ambient Storage — 25 °C … + 65 °C…
-
Page 36
Technical data Cond 3110 Measuring ranges, Variable Measuring range Resolution resolution 0.0 … 70.0 according to the IOT table − 5.0 … + 105.0 T [°C] + 23.0 … + 221.0 T [°F] Cell constants Cell constant C Values Can be calibrated in 0.450 … -
Page 37: Lists
Cond 3110 Lists Lists This chapter provides additional information and orientation aids. Specialist terms The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here. Index The index helps you to find the topics that you are looking for.
-
Page 38
Lists Cond 3110 Reset Restoring the original condition of all settings of a measuring system. Resistance Short name for the specific electrolytic resistance. It corresponds to the reciprocal value of the electrical conductivity. Resolution Smallest difference between two measured values that can be displayed by a measuring instrument. -
Page 39
Cond 3110 Lists Index Authorized use ….. . 10 Reset ……29 AutoRead . -
Page 40
Lists Cond 3110 ba75792e02 11/2008… -
Page 41
Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: +49 (0) 881 183-0 +49 (0) 881 183-100 Fax: +49 (0) 881 183-420 E-Mail: Info@WTW.com Internet: http://www.WTW.com…
-
Page 1
Operating manual Cond 3110 Conductivity meter ba75792e02 11/2008… -
Page 2
© Copyright Weilheim 2008, WTW GmbH Reproduction in whole — or even in part — is prohibited without the express written permission of WTW GmbH, Weilheim. Printed in Germany. ba75792e02 11/2008… -
Page 3: Table Of Contents
Contents Cond 3110 — Contents Overview ……..5 Keypad .
-
Page 4
Contents Cond 3110 What to do if……..33 Technical data . -
Page 5: Overview
Cond 3110 Overview Overview The Cond 3110 compact precision conductivity meter enables you to perform conductivity measurements quickly and reliably. The Cond 3110 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications. The proven procedures for determining or adjusting the cell constant support your work with the conductivity meter.
-
Page 6: Keypad
Overview Cond 3110 Keypad Cond 3110 In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g.
-
Page 7: Display
Cond 3110 Overview Display Measured parameter Sensor symbol S/cm (calibration µ evaluation) Measured value 1/cm Time display C ° °F Temperature Tref20 Tref25 display, function ARng LoBat display indicators, status display indicators Status display Stability control (AutoRead) is active indicators ARng Automatic range switching;…
-
Page 8: Socket Field
Overview Cond 3110 Socket field Connectors: Conductivity measuring cell Service interface Caution Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary measuring cells fulfill these conditions.
-
Page 9: Safety
Cond 3110 Safety Safety This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual before working with the meter. The operating manual must always be available within the vicinity of the meter.
-
Page 10: Authorized Use
Safety Cond 3110 Authorized use Authorized use of the meter consists exclusively of the measurement of conductivity, temperature and salinity in a laboratory or field environment. The technical specifications as given in chapter 7 T must ECHNICAL DATA be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized.
-
Page 11
Safety datasheets of the chemical manufacturers. Caution In addition to the safety instructions mentioned here, also follow the safety instructions of the sensors used. The operating manuals of the sensors are available on the supplied CD and on the Internet under www.WTW.com. ba75792e02 11/2008… -
Page 12
Safety Cond 3110 ba75792e02 11/2008… -
Page 13: Commissioning
Cond 3110 Commissioning Commissioning Scope of delivery Conductivity meter Cond 3110 4 batteries 1.5 V Mignon type AA Short instructions CD-ROM with detailed operating manual Initial commissioning Perform the following activities: Insert the supplied batteries Switch on the meter. 3.2.1 Inserting the batteries Unscrew the two screws (1) on the underside of the meter.
-
Page 14: Switching On The Meter
Commissioning Cond 3110 Mignon AA). In order to charge the batteries, an external charging device is required. Caution Make sure that the poles of the batteries are the right way round. ± ± signs on the batteries must correspond to the signs in the battery compartment.
-
Page 15: Operation
Cond 3110 Operation Operation General operating principles This section contains basic information on the operation of the Cond 3110. 4.1.1 Operating modes The meter has the following operating modes: Measurement The display indicates the measurement data in the measured value…
-
Page 16: Measuring
Operation Cond 3110 Measuring Preparatory activities Perform the following preparatory activities when you want to measure: Connect a measuring cell to the meter. Calibrate or check the meter with the measuring cell. Select the measured parameter with <M>. Stability control…
-
Page 17: Measuring The Conductivity
Cond 3110 Operation 4.2.1 Measuring the conductivity Perform the preparatory activities according to section 4.2. Immerse the conductivity measuring cell in the test sample. If necessary, scroll with <M> until the measured parameter with the unit mS/cm or µS/cm is displayed.
-
Page 18
Operation Cond 3110 C ° ARng ba75792e02 11/2008… -
Page 19: Determining/Setting Up The Cell Constant [C]
Cond 3110 Operation Determining/setting up the cell constant [C] Why determine/set up Due to aging, the cell constant slightly changes. As a result, an inexact the cell constant? measured value is displayed. Calibration determines the current value of the cell constant and stores this value in the meter.
-
Page 20
Operation Cond 3110 1/cm Tref25 Immerse the measuring cell in the control standard solution, 0.01 mol/l KCI. Start the calibration with <ENTER>. The determination of the cell constant with stability control starts. The display indicator flashes until there is a stable signal. -
Page 21: Using The Last Calibrated Cell Constant
Cond 3110 Operation Calibration evaluation After the calibration, the meter automatically evaluates the current status. The evaluation appears on the display. Display Cell constant [cm in the range 0.450 … 0.500 cm 0.800 … 0.880 cm You are working with a correctly calibrated measuring cell.
-
Page 22: Setting The Cell Constant Manually
Operation Cond 3110 If necessary, press <CAL> repeatedly until CAL and the last calibrated cell constant is displayed. 1/cm Tref25 Confirm the selection with <ENTER>. The displayed cell constant is used. The meter switches to the measured value display. 4.3.3 Setting the cell constant manually…
-
Page 23
Cond 3110 Operation S/cm µ 1/cm Tref25 ARng Set the cell constant to be used with <▲><▼>, e.g. 0.846 cm S/cm µ 1/cm Tref25 ARng Confirm the selection with <ENTER>. The new cell constant is used from now on. The meter switches to the measured value display. -
Page 24
Operation Cond 3110 Selecting the cell Press the <CAL> key repeatedly until USE CELL is displayed. constant 0.475 cm Confirm the selection with <ENTER> or <CAL >. If necessary, press <CAL> repeatedly until the cell constant 0.475 cm is displayed. -
Page 25: Temperature Compensation Tc
Cond 3110 Operation Temperature compensation TC The calculation of the temperature compensation is based on the preset reference temperature, Tref 20 or Tref 25 (see section 4.5 S ETTINGS As the temperature compensation, the nonlinear temperature compensation «nLF» according to DIN 38404 or EN 27 888 respectively is permanently set.
-
Page 26: Settings
Operation Cond 3110 Settings You can adapt the meter to your individual requirements. The settings are done in the following menus: System settings (<ENTER >) – Switch-off interval (tOff) Measurement settings (<M >) – Reference temperature (Tref25 or Tref20) – Temperature unit (°C / °F) –…
-
Page 27: Measurement Settings
Cond 3110 Operation 4.5.2 Measurement settings These settings apply to the determination of the cell constant and measurement (the default condition is printed in bold). Reference temperature t25, t20 Temperature unit (UnI) °C, °F Cleaning interval (Int.C) 0 … 180 … 999 d Reference temperature Open the menu for measurement settings with <M…
-
Page 28
Operation Cond 3110 Using <▲><▼>, toggle between °C and °F. Confirm with <ENTER>. Int.C, the setting of the cleaning interval is displayed. Cleaning interval (Int.C) Time Set the interval with <▲><▼>. Confirm with <ENTER>. The measurement settings are completed. The meter switches to the measuring mode. -
Page 29: Reset
Cond 3110 Operation Reset 4.6.1 Resetting the cell constant This function serves to erase the last determined cell constant. Subsequently, the meter uses the last manually set cell constant in the range 0.800 … 0.880 cm or the fixed cell constant, 0.475 cm Based on the last erased cell constant the meter decides to which of the two manually set cell constants the cell constant is reset.
-
Page 30: Resetting All Meter Settings
Operation Cond 3110 4.6.2 Resetting all meter settings This function resets all meter settings to the default condition. The relevant values are given in the following sections: System settings section 4.5.1 Measurement settings section 4.5.2 The following settings are also reset to the default condition:…
-
Page 31: Maintenance, Cleaning, Disposal
Cond 3110 Maintenance, cleaning, disposal Maintenance, cleaning, disposal Maintenance The only maintenance activity required is replacing the batteries. Note See the relevant operating manuals of the measuring cells for instructions on maintenance. 5.1.1 Replacing the batteries Unscrew the two screws (1) on the underside of the meter, Open the battery compartment (2) on the underside of the meter.
-
Page 32: Cleaning
Maintenance, cleaning, disposal Cond 3110 Caution Make sure that the poles of the batteries are the right way round. ± ± signs on the batteries must correspond to the signs in the battery compartment. Close the battery compartment (2) and tighten the screws (1).
-
Page 33: What To Do If
Cond 3110 What to do if… What to do if… Error message Cause Remedy OFL, UFL – Measured value outside the – Use suitable measuring cell measuring range Error message, Cause Remedy – Measuring cell contaminated – Clean cell and replace it if necessary –…
-
Page 34
What to do if… Cond 3110 ba75792e02 11/2008… -
Page 35: Technical Data
Cond 3110 Technical data Technical data General data Dimensions approx. 180 x 80 x 55 mm Weight approx. 0.4 kg Mechanical structure Type of protection IP 67 Electrical safety Protective class Test certificates Ambient Storage — 25 °C … + 65 °C…
-
Page 36
Technical data Cond 3110 Measuring ranges, Variable Measuring range Resolution resolution 0.0 … 70.0 according to the IOT table − 5.0 … + 105.0 T [°C] + 23.0 … + 221.0 T [°F] Cell constants Cell constant C Values Can be calibrated in 0.450 … -
Page 37: Lists
Cond 3110 Lists Lists This chapter provides additional information and orientation aids. Specialist terms The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here. Index The index helps you to find the topics that you are looking for.
-
Page 38
Lists Cond 3110 Reset Restoring the original condition of all settings of a measuring system. Resistance Short name for the specific electrolytic resistance. It corresponds to the reciprocal value of the electrical conductivity. Resolution Smallest difference between two measured values that can be displayed by a measuring instrument. -
Page 39
Cond 3110 Lists Index Authorized use ….. . 10 Reset ……29 AutoRead . -
Page 40
Lists Cond 3110 ba75792e02 11/2008… -
Page 41
Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: +49 (0) 881 183-0 +49 (0) 881 183-100 Fax: +49 (0) 881 183-420 E-Mail: Info@WTW.com Internet: http://www.WTW.com…
Русский
Обзор
Кнопочная
панель
Дисплей
50
RU Русский
Кондуктометр Cond 3110
Этот компактный прецизионный кондуктометр Cond 3110позволяет быстро
и надёжно измерять проводимость растворов.
Прибор карманного формата Cond 3110 подходит для использования в
любых растворах и обеспечивает максимум удобства, надёжности и
точности в работе.
Названия кнопок выделены угловыми скобками <..>.
Как правило, символ на кнопке (например <ENTER>) в руководстве по
эксплуатации означает короткое нажатие кнопки (менее 2 с).
Продолжительное нажатие кнопки (ок. 2 с) обозначается знаком
_
подчёркивания вслед за указанием символа на кнопке (например
<ENTER
>).
_
<On/Off>:
Включить/выключить измерительный прибор
<On/Off
>:
Сбросить калибровочные данные
_
<M>:
Выбрать измеряемый параметр
<M
>:
Открыть меню настройки калибровки и измерений
_
<CAL>:
Вызвать способ калибровки
<CAL
>:
Показать калибровочные данные
< >:
Увеличить значение, перейти к следующему
< >:
Уменьшить значение, перейти к следующему
_
<ENTER>:
Подтвердить ввод значений
<ENTER
>:
Открыть меню системных настроек
8
—
8
Time
8
8 8
Tref20
Tref25
LoBat
nLF
Sal
8
8
m
S/cm
µ
1/cm
8
C °
/K
%
°F
ARng
Cal
AR
Измеряемый параметр
Символ датчика
(обработка
калибровочных данных)
Показание результата
измерения
TP
Показание температуры,
функции, состояния
ba75783y03
Cond 3110
05/2017
Простой в управлении, прочный кондуктометр WTW Cond 3110 для измерений электропроводности и солености. Упрощенная клавиатура для простой и безопасной работы. Прочный корпус и 100% защита от влаги. Большой жидкокристаллический экран для легкого чтения. Экран кондуктометра одновременно отображает измеряемую величину и температуру. Автоматическая термокомпенсация для всех видов природных вод. Кондуктометр работает с 4-полюсной ячейкой TetraCon® 325 или с 2-полюсной ячейкой KLE 325. Легкая работа и надежный результат гарантирован как для природной воды, так и на химическом производстве.
Кондуктометр WTW Cond 3110 SET 1 поставляется в кейсе с ячейкой TetraCon 325 и принадлежностями (кондуктометрический стандарт 1413 мкСм / см при 25 ° C, штатив, стаканчик, короткое руководство по эксплуатации, CD-ROM и батарейки).
Технические характеристики кондуктометра WTW Cond 3110:
Диапазон измерений (разрешение) проводимости: 0,0 … 1000 мСм/см ± 0,5 % значения
Константы ячеек:
-заданные: 0,475 см–1, 0,880 см–1
-калибруемые: 0,450 … 0,500 см–1, 0,800 … 0,880 см–1
Соленость: 0.0 — 70.0
Температура: – 5,0 … 105,0 °C ± 0,1 °C
Контроль дрейфа: автоматически
Tемпература сравнения: 20°C/25°C переключаемая
Компенсация Т: nLF
Память калибровки: вызов текущих значений
Дисплей: сегментный ЖК
Память: —
Сохранение:-
Интерфейс:-
Питание: 4 x 1,5 В AA или 4 x 1,2 V NiMH
Автономная работа: до 1000 часов.
-
Добавить к сравнению
-
Кондуктометрическая ячейка WTW TetraCon 325 с графитовыми электродами
- 301960
WTW
шт 20% 941.40 евро -
Кондуктометр WTW Cond 3110 SET 1 (портативный, с датчиком TetraCon 325)
Производитель: WTW
Артикул: 00000033030
Цена опт.: по запросу
Цена розн.: по запросу
В том числе НДС 20%
Описание
Кондуктометр WTW Cond 3110 SET 1 (портативный, с датчиком TetraCon 325)
Портативный кондуктометр ProfiLineCond 3110 в кейсе и с новым 4-электродным датчиком TetraCon 325 общего назначения для измерений
проводимости.
· Прочный корпус и 100%-ная защита от влаги
· Четкие нажатия кнопок
· Встроенный таймер для напоминания о
калибровке
Легкая работа и надежный результат гарантирован как для природной воды, так
и на химическом производстве.
Область применения:
Контроль технологических процессов
Экспедиционное оборудование
Модель |
Cond 3110 |
Проводимость |
0,0 … 1000 |
Удельное |
– |
Константы |
заданные: |
Соленость |
0,0 … 70,0 |
TDS |
– |
Температура |
– 5,0 … |
Контроль |
автоматически |
Tref |
20 °C/25 |
Компенсация |
nLF |
Память |
вызов |
Дисплей |
сегментный |
Память |
– |
Сохранение |
– |
Интерфейс |
– |
Питание |
4 x 1,5 В |
Автономная |
до 1000 |
Документы и сертификаты
Сопутствующие товары
Онлайн-заявка
Если вы заинтересованы в приобретении товара «Кондуктометр WTW Cond 3110 SET 1 (портативный, с датчиком TetraCon 325)», или хотите задать вопрос, то сделайте это прямо здесь:
— физическое лицо,
— юридическое лицо
Уважаемые покупатели, информируем Вас, что производитель может изменить цвет, внешний вид и характеристики товара без дополнительного уведомления продавца (НВ-Лаб). Поэтому размещенные на нашем сайте характеристики и фотографии являются справочными. Характеристики и внешний вид купленного товара иногда могут отличаться от опубликованных. Мы стараемся поддерживать описания в актуальном состоянии и обновляем информацию по мере получения её от производителей.
Вся размещённая на сайте информация опубликована исключительно для ознакомительных целей и ни при каких условиях не является публичной офертой.