-
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Related Manuals for Flexim FLUXUS G704
Summary of Contents for Flexim FLUXUS G704
-
Page 1
User Manual UMG70XV3-4EN Ultrasonic Flowmeter for Gas FLUXUS G704 FLUXUS G704 A2 FLUXUS G709 Firmware V5.xx… -
Page 2
MS-DOS, Excel, Windows are registered trademarks of Microsoft Corporation. FLUXUS is a registered trademark of FLEXIM GmbH. FLEXIM GmbH Wolfener Strasse 36 12681 Berlin Germany Tel.: +49 (30) 936 67 660 Fax: +49 (30) 936 67 680 E-mail: flexim@flexim.de www.flexim.com… -
Page 3
Die Sprache, in der die Anzeigen auf dem FLUXUS erscheinen, kann einge- stellt werden (siehe Abschnitt 9.4). FLUXUS can be operated in the language of your choice (see section 9.4). Il est possible de sélectionner la langue utilisée par FLUXUS à l’écran (voir section 9.4). -
Page 5: Table Of Contents
Safety Precautions………………..9 Warranty ………………….10 Safety Instructions according to Directive 94/9/EC……..11 Safety Instructions for the Use of Flowmeter FLUXUS G704 A2 in Explosive Atmosphere ………………11 Use of the Junction Box JB01S#E3# in Explosive Atmosphere ……13 Use of Ultrasonic Transducers in Explosive Atmosphere (Zone 1/Zone 21) ..15 Handling ………………….24…
-
Page 6
Connection of the Serial Interface …………… 56 6.10 Connection of the Sensor Module (SENSPROM) ……….56 Installation of FLUXUS G709…………….58 Location………………….. 58 Connection of the Transducers — Connection System TS ……..58 Connection of the Transducers — Connection System AS ……..66 Connection of the Power Supply …………….. -
Page 7
Displaying the Measured Values …………..100 11.1 Selection of the Physical Quantity and of the Unit of Measurement ….100 11.2 Toggling between the Channels……………. 101 11.3 Adjustment of the Display …………….. 102 11.4 Status Line ………………….103 11.5 Transducer Distance ………………104 Advanced Measuring Functions ………….. -
Page 8
Inputs ………………….. 148 17.1 Linking the Temperature Inputs to the Measuring Channels……148 17.2 Selection of the Temperature Probe (FLUXUS G704) ……..149 17.3 Linking Other Inputs to the Measuring Channels ……….149 17.4 Activation of the Inputs ………………150 17.5… -
Page 9: Introduction
The content of this manual is subject to changes without prior notice. All rights reserved. No part of this manual may be reproduced in any form without FLEXIM’s written permis- sion. Safety Precautions…
-
Page 10: Warranty
• unsuitable or insufficient maintenance • repair of FLUXUS by unauthorized personnel FLEXIM assumes no responsibility for injury to the customer or third persons proximately caused by the material owing to defects in the product which were not predictable or for any indirect damages.
-
Page 11: Safety Instructions According To Directive 94/9/Ec
The conformance with all relevant standards will be guaranteed only if the flowmeter is connected only with the transducers specified by FLEXIM for the corresponding application. Observe the specified operating and explosion protection tempera- tures (see Technical Data on the following pages).
-
Page 12
2 Safety Instructions according to Directive 94/9/EC Technical Data For further technical data see annex A, section Technical Data. FLUXUS G704 A2 manufacturer FLEXIM GmbH Wolfener Str. 36 12681 Berlin marking II3G Ex nA II T4 Ta -20…+60 °C II3D Ex tD A22 IP66 T100°C… -
Page 13: Use Of The Junction Box Jb01S#E3# In Explosive Atmosphere
If other cable glands are used than those supplied by FLEXIM, they have to be approved for the application area and the cable diameters have to be observed.
-
Page 14
2 Safety Instructions according to Directive 94/9/EC The cover of the junction box has to be closed firmly. Technical Data For further information see annex A, section Technical Data. Junction Box JB01S#E3# FLEXIM GmbH manufacturer Wolfener Str. 36 12681 Berlin II2G 0044;… -
Page 15: Use Of Ultrasonic Transducers In Explosive Atmosphere (Zone 1/Zone 21)
2 Safety Instructions according to Directive 94/9/EC Use of Ultrasonic Transducers in Explosive Atmosphere (Zone 1/Zone 21) Observe the specified operating and explosion protection tempera- tures of the transducers. For the explosion protection temperatures see table on the following page. The transducers **M1N31, **P1N31, **Q1N31,…
-
Page 16
2 Safety Instructions according to Directive 94/9/EC The transducers **K1N41, **G1N41, **K1N43, **G1N43, **H1N43, **M2N41, **P2N41, **M1N43, **P1N43, **M2E45, **P2E45, **K1N81, **G1N81, **K1N83, **G1N83, **H1N83, **M2N81, **P2N81, **M1N83, **P1N83, **K1LI1, **G1LI1, **M2LI1, **P2LI1, **M2E85, **P2E85, **Q2E85, **Q2N81, **Q1N83 must always be used in the delivered transducer shoe to protect them against mechanical stress. -
Page 17
2 Safety Instructions according to Directive 94/9/EC surface explosion protection ambient temperature temperature of transducers temperature (pipe/vessel) **M1N31, **P1N31, **Q1N31 -20 …+120 °C -20…+120 °C **M2N41, **P2N41, -55…+180 °C -55…+180 °C **K1N41, **G1N41, **M2N81, **P2N81, **K1N81, **G1N81, **Q2N81 **K1N31, **G1N31, **K1N71, -40…+180 °C -40…+180 °C… -
Page 18
2 Safety Instructions according to Directive 94/9/EC Technical Data For further information see annex A, section Technical Data. Type **M1N31, **P1N31, **Q1N31 manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin marking 0044; II2G EEx m II T6…T4 T -20…+120 °C… -
Page 19
2 Safety Instructions according to Directive 94/9/EC Type **K1N41, **G1N41, **M2N41, **P2N41 manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin marking 0044; II2G Ex q II T6…T3 T -55…+180 °C II2D Ex tD A21 IP65 TX certification IBExU04ATEX1011 X type of protection… -
Page 20
2 Safety Instructions according to Directive 94/9/EC Type **K1N43, **G1N43, **H1N43, **M1N43, **P1N43 manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin marking 0044; II2G Ex q II T6…T3 T -55…+140 °C II2D Ex tD A21 IP65 TX certification IBExU04ATEX1011 X… -
Page 21
2 Safety Instructions according to Directive 94/9/EC Type **K1N71, **G1N71 manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin II2G marking 0044 II2D Ex e q II T6…T3 Ex tD A21 IP65 TX certification IBExU07ATEX1168 X IECEx IBE08.0007X type of protection… -
Page 22
2 Safety Instructions according to Directive 94/9/EC Type **K1N73, **G1N73, **H1N73 manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin II2G marking 0044; II2D Ex e q II T6…T3 Ex tD A21 IP65 TX certification IBExU07ATEX1168 X IECEx IBE08.0007X type of protection… -
Page 23
2 Safety Instructions according to Directive 94/9/EC Type **K1LI1, **G1LI1, **M2LI1, **P2LI1 FLEXIM GmbH manufacturer Wolfener Str. 36 12681 Berlin marking 0044; II 2G 2D Ex q II T6…T3 Ex tD A21 IP68 TX IBExU07ATEX1168 X certification IECEx IBE08.0007X type of protection… -
Page 24: Handling
• Do not open the housing without authorization. The degree of protection of the flowme- ter FLUXUS G704 A2, G704 is given only if the cover of the housing is tightly screwed. • Keep the transducers clean. Manipulate the transducer cables cautiously. Avoid ex- cessive cable bend.
-
Page 25: Flowmeter
4 Flowmeter Flowmeter Measuring Principle The flow of the medium is measured by ultrasonic signals using the transit time differ- ence method. Ultrasonic signals are emitted by a transducer installed on one side of a pipe, reflected on the opposite side and received by a second transducer. The signals are emitted alter- natively in and against the flow direction.
-
Page 26: Description Of The Flowmeter
The front plate has to be removed to access the command panel. keyboard 2x 16-digit LCD serial interface display (backlit) RS232 transducers power supply cover equipotential bonding terminal (only FLUXUS G704 A2) inputs outputs Fig. 4.3: Command panel of FLUXUS G704 A2 and FLUXUS G704 UMG70XV3-4EN 12.01.2009…
-
Page 27
4 Flowmeter 4.2.2 Design of FLUXUS G709 The flowmeter is designed as 19 » rack (42 DU, 3 HU). 2x 16-digit LCD display serial interface RS232 (backlit) sensor module sensor module measuring measuring keyboard channel B channel A inputs T 4 B S B 1 P 7 a T 4 B… -
Page 28
4 Flowmeter 4 Flowmeter Keyboard The keyboard has three function keys ENTER, BRK and C and ten numerical keys. Several keys have double functions. They can be used for data entry and for navigation through scroll lists. The arrow shaped keys will be used as cursor keys for scrolling through scroll lists and for input of digits and letters in input mode. -
Page 29
4 Flowmeter Table 4.4: Input of text positioning of the cursor changes the currently selected character to an «A» DISP changes the currently selected character to a «Z» DISP changes between small and capital letters selection of the precedent/next ASCII character deletes the character and inserts a blank scrolls automatically upwards/downwards through the restricted ASCII charac- N E X T… -
Page 30: Selection Of The Measuring Point
5 Selection of the Measuring Point Selection of the Measuring Point The correct selection of the measuring point is crucial for achieving reliable measure- ment results and a high measurement accuracy. A measurement must take place on a pipe in which •…
-
Page 31: Undisturbed Flow Profile
5 Selection of the Measuring Point The following conditions have to be observed at the measuring point: • no material deposits in the pipe • no accumulation of liquid (condensate), e.g. before orifice plates or at pipe sections lo- cated lower If liquid or solids accumulate on the bottom of the pipe, the transducers have to be fixed on the sides of the pipe (see Fig.
-
Page 32
5 Selection of the Measuring Point Table 5.1: Recommended distance from disturbance sources D = nominal pipe diameter at the measuring point, L = recommended distance disturbance source: 2x 90 ° elbows in one plane supply: l 50D return: l 10 D disturbance source: 2x 90 °… -
Page 33
5 Selection of the Measuring Point Table 5.1: Recommended distance from disturbance sources D = nominal pipe diameter at the measuring point, L = recommended distance disturbance source: valve supply: l 80D return: l 20D disturbance source: reducer supply: l … -
Page 34: Influence Of Noise
5 Selection of the Measuring Point Influence of Noise The ultrasonic waves do not propagate only in the medium but also in the pipe wall (see Fig. 5.2). They will be reflected at flanges. ultrasonic waves in the medium (measuring signal) ultrasonic waves at the pipe wall (pipe wall signal) Fig.
-
Page 35
5 Selection of the Measuring Point Table 5.2: Measuring points to be avoided in the distance of l ± 2 D from the reflection point Pipe wall signal and measuring signal reach the transducer simultaneously. disadvantageous —— — D — — — distance to reflection point D — outer pipe diameter… -
Page 36: Selection Of The Measuring Point Taking Into Account Flow Profile And Influence Of Noise
5 Selection of the Measuring Point Selection of the Measuring Point Taking into Account Flow Profile and Influence of Noise • Select a measuring point area where the flow profile is fully developed (see section 5.2). • Select the measuring point within this area so that the influence of noise can be ne- glected (see 5.3).
-
Page 37
5 Selection of the Measuring Point Both demands can not be observed always at the same time. In these cases, the mea- suring point has to be selected so that the influence of noise is min. and the measuring point is as far as possible from disturbance sources of the flow profile. example: medium: natural gas pipe material: stainless steel… -
Page 38: Installation Of Fluxus G704
The flowmeter FLUXUS G704 must be operated outside of explosive Attention! atmospheres. The flowmeter FLUXUS G704 A2 must be operated outside of ATEX zone 0 and 1. The degree of protection of the flowmeter is only guaranteed if the Attention! cable glands are firmly tightened and the housings have been tightly screwed.
-
Page 39: Pipe Installation
6 Installation of FLUXUS G704 • Remove the cover of the housing. • At the chosen location, drill 4 holes in the wall (see Fig. 6.1). • Insert dowels in the holes. • Fix the housing with screws to the wall.
-
Page 40: Connection Of The Transducers — Connection System Ts
6 Installation of FLUXUS G704 Connection of the Transducers — Connection System TS If transducers are replaced or added, the sensor module has to be Note! replaced or added, too (see section 6.10). It is recommended to run the cables from the measuring point to the flowmeter before the connection of the transducers to avoid load on the connection points.
-
Page 41
6 Installation of FLUXUS G704 equipotential bonding transducers transducers terminal measuring channel A measuring channel B (FLUXUS G704 A2) basic part cable gland with double hole sealing compres- sion part cap nut transducer cable transducers ATEX zone 1 Fig. 6.3: Transducers ATEX zone 1 — direct connection Table 6.1: Terminal assignment… -
Page 42: Transducers (Atex Zone 1) — Connection Via Junction Box
6 Installation of FLUXUS G704 6.4.2 Transducers (ATEX Zone 1) — Connection via Junction Box Connection of the Extension Cable with the Flowmeter • Remove the left blind plug for the connection of the transducers (see Fig. 6.4 ). • Open the cable gland of the extension cable. The compression part remains in the cap nut.
-
Page 43: Connection Of The Extension Cable With The Junction Box
6 Installation of FLUXUS G704 Connection of the Extension Cable with the Junction Box The outer shield of the extension cable must not have electrical con- Attention! tact to the junction box. The extension cable has to remain insulated completely up to the shield terminal of the junction box.
-
Page 44
6 Installation of FLUXUS G704 basic part cable gland compres- sion part cap nut extension cable transducer cable transducers ATEX zone 1 Fig. 6.5: Transducers (ATEX zone 1) — connection via junction box, connection of extension cable and transducer cable to the junction box UMG70XV3-4EN 12.01.2009… -
Page 45: Connection Of The Transducer Cable With The Junction Box
6 Installation of FLUXUS G704 TV TVS TG TG TRS TR VS RS shield terminal equipotential bonding terminal Fig. 6.6: Connection of extension cable and transducer cable (ATEX zone 1) Connection of the Transducer Cable with the Junction Box • Remove the blind plug from the junction box (see Fig. 6.5).
-
Page 46: Transducers (Atex Zone 2, Fm Div. 2 And Without Explosion Protection)
6 Installation of FLUXUS G704 6.4.3 Transducers (ATEX Zone 2, FM Div. 2, without Explosion Protection) — Direct Connection Observe the safety instructions in chapter 2. • Remove the left blind plug for the connection of the transducers (see Fig. 6.7).
-
Page 47: Transducers (Atex Zone 2, Fm Div. 2 And Without Explosion Protection)
6 Installation of FLUXUS G704 Table 6.5: Terminal assignment terminal connection X_AV SMB connector (brown cable, marked white) X_AR SMB connector (brown cable, marked black) 6.4.4 Transducers (ATEX Zone 2, FM Div. 2, without Explosion Protection) — Connection via Junction Box Connection of the Extension Cable with the Flowmeter •…
-
Page 48: Connection Of The Extension Cable With The Junction Box
6 Installation of FLUXUS G704 Table 6.6: Terminal assignment terminal connection white or marked cable (core) white or marked cable (inner shield) brown cable (inner shield) brown cable (core) Connection of the Extension Cable with the Junction Box The outer shield of the extension cable must not have electrical con- Attention! tact to the junction box.
-
Page 49
6 Installation of FLUXUS G704 TV TVS TG TG TRS TR basic part cable gland compres- sion part cap nut extension cable transducer cable transducers: shield terminal • ATEX zone 2 • FM Div. 2 • without explosion protection Fig. 6.9: Transducers (ATEX zone 2, FM Div. 2, without explosion protection) -… -
Page 50: Connection Of The Transducers — Connection System As
6 Installation of FLUXUS G704 Connection of the Transducers — Connection System AS • Remove the left blind plug for the connection of the transducers (see Fig. 6.10). • Insert the extension cable with the AMP-Quick and SMB connectors in the housing.
-
Page 51: Connection Of The Power Supply
If the flowmeter FLUXUS G704 A2 is used in explosive atmosphere, the switch should be installed outside the explosive atmosphere. If this is not possible, the switch should be installed in the area with the least risk.
-
Page 52: Connection Of The Outputs
6 Installation of FLUXUS G704 Connection of the Outputs Configure the outputs (see chapter 18). The terminals I1 active loop to be used for the connection of the output are dis- Terminal:P1+,P1- played at the end of the configuration dialog (here: P1+ and P1- for the active current loop).
-
Page 53
6 Installation of FLUXUS G704 Table 6.11: Circuits of the outputs output FLUXUS terminal circuits semi-active = 0…24 V current loop (red) > 0.021A [] LOAD operated as passive cur- rent loop (black) passive = 5…25 V current loop (red) >0.021A… -
Page 54: Connection Of The Inputs
6 Installation of FLUXUS G704 Connection of the Inputs Max. 4 temperature probes Pt100/Pt1000 (4-wire) can be connected to the flowmeter (option). Configure the temperature inputs (see chapter 17). Table 6.12: Terminal assignment of the flowmeter terminal connection with connection without…
-
Page 55
6 Installation of FLUXUS G704 If the flowmeter has a passive current input, an external power supply is necessary. The power supply U must be able to • supply sufficient power for the energy requirements of the passive current source •… -
Page 56: Connection Of The Serial Interface
6 Installation of FLUXUS G704 Table 6.13: Technical data of the current input active passive measuring range 0…+20 mA -20…+20 mAâ 0.1% v. MW 10 µA 0.1% v. MW 10 µA accuracy 50 /0.6 W 50 /0.6 W input impedance…
-
Page 57
6 Installation of FLUXUS G704 • Stop the measurement. • Insert the sensor module in the lower row of terminal strip KL1 (see Fig. 6.16). The transducers of measuring channel A are connected to SA1…SA4, the transducers of measuring channel B to SB1…SB4. -
Page 58: Installation Of Fluxus G709
7 Installation of FLUXUS G709 Installation of FLUXUS G709 Location Select the measuring point according to the recommendations in chapter 5. The ambient temperature must be within the operating temperature range of the transducers. Select the location of the flowmeter within cable reach of the measuring point. The ambi- ent temperature must be within the operating temperature range of the flowmeter (see annex A, section Technical Data).
-
Page 59: Transducers (Atex Zone 1) — Connection Via Junction Box
7 Installation of FLUXUS G709 7.2.1 Transducers (ATEX Zone 1) — Connection via Junction Box Connection of the Extension Cable with the Flowmeter • Prepare the extension cable. Cut the outer shield and brush it back. • Push the extension cable through the shield terminal to terminal strip KL6 for measur- ing channel A and to terminal strip KL8 for measuring channel B (see Fig.
-
Page 60: Connection Of The Extension Cable With The Junction Box
7 Installation of FLUXUS G709 Connection of the Extension Cable with the Junction Box The outer shield of the extension cable must not have electrical con- Attention! tact to the junction box. The extension cable has to remain insulated completely up to the shield terminal of the junction box. •…
-
Page 61
7 Installation of FLUXUS G709 basic part cable gland compres- sion part cap nut extension cable transducer cable transducers ATEX zone 1 Fig. 7.2: Transducers (ATEX zone 1) — connection via junction box, connection of extension cable and transducer cable to the junction box UMG70XV3-4EN 12.01.2009… -
Page 62: Connection Of The Transducer Cable With The Junction Box
7 Installation of FLUXUS G709 TV TVS TG TG TRS TR VS RS shield terminal equipotential bonding terminal Fig. 7.3: Connection of extension cable and transducer cable (ATEX zone 1) Connection of the Transducer Cable with the Junction Box • Remove the blind plug from the junction box (see Fig. 7.2). •…
-
Page 63: Connection Via Junction Box
7 Installation of FLUXUS G709 7.2.2 Transducers (ATEX Zone 2, FM Div. 2, without Explosion Protection) — Connection via Junction Box Connection of the Extension Cable with the Flowmeter • Prepare the extension cable. Cut the outer shield and brush it back. •…
-
Page 64: Connection Of The Extension Cable With The Junction Box
7 Installation of FLUXUS G709 Connection of the Extension Cable with the Junction Box The outer shield of the extension cable must not have electrical con- Attention! tact to the junction box. The extension cable has to remain insulated completely up to the shield terminal. •…
-
Page 65
7 Installation of FLUXUS G709 TV TVS TG TG TRS TR basic part cable gland compres- sion part cap nut extension cable transducer cable transducers: shield terminal • ATEX zone 2 • FM Div. 2 • without explosion protection Fig. 7.5: Transducers (ATEX zone 2, FM Div. 2, without explosion protection) — connection via junction box, connection of extension cable and transducer cable to the junction box UMG70XV3-4EN 12.01.2009… -
Page 66: Connection Of The Transducers — Connection System As
7 Installation of FLUXUS G709 Connection of the Transducers — Connection System AS • Open the cable gland of the extension cable (see Fig. 7.6). • Push the basic part of the cable gland towards the AMP-Quick and SMB connectors, the cap nut and the compression part in the other direction.
-
Page 67: Connection Of The Power Supply
7 Installation of FLUXUS G709 Table 7.7: Terminal assignment terminal connection measuring channel A measuring channel B X6AV X8BV SMB connector (white or marked cable) X6AR X8BR SMB connector (brown cable) AMP-Quick connector Connection of the Power Supply A switch according to IEC 61010-1:2001 has to be provided in the Attention! building installation which must be near the instrument, easily acces- sible for the user and marked as disconnection device of the instru-…
-
Page 68: Connection Of The Outputs
7 Installation of FLUXUS G709 Connection of the Outputs Configure the outputs (see chapter 18). The terminals I1 active loop to be used for the connection of the output are dis- Terminal:P1+,P1- played at the end of the configuration dialog (here: P1+ and P1- for the active current loop).
-
Page 69
7 Installation of FLUXUS G709 Table 7.9: Circuits of the outputs output FLUXUS terminal circuits semi-active = 0…24 V current loop, (red) > 0.021A [] LOAD operated as passive cur- rent loop (black) passive = 5…25 V current loop (red) >0.021A []+4V LOAD… -
Page 70: Connection Of The Inputs
7 Installation of FLUXUS G709 Connection of the Inputs Max. 4 temperature probes Pt100 (4-wire) can be connected to the flowmeter (option). Configure the temperature inputs (see chapter 17). Table 7.10: Terminal assignment of the flowmeter terminal connection with connection without extension cable extension cable gray…
-
Page 71
7 Installation of FLUXUS G709 If the flowmeter has a passive current input, an external power supply is necessary. The power supply U must be able to • supply sufficient power for the energy requirements of the passive current source •… -
Page 72: Connection Of The Serial Interface
7 Installation of FLUXUS G709 Table 7.11: Technical data of the active current input current input active passive measuring range 0…20 mA -20…+20 mA 0.1 % of reading 10 µA 0.1 % of reading 10 µA accuracy 50 /0.6 W 50 /0.6 W input impedance short circuit current…
-
Page 73: Connection Of The Sensor Module (Sensprom)
7 Installation of FLUXUS G709 Connection of the Sensor Module (SENSPROM) The sensor module contains important transducer data for the operation of the flowmeter with the transducers. It is connected with the corresponding terminals of the flowmeter. If transducers are replaced or added, the sensor module has to be replaced or added, too.
-
Page 74: Installation Of The Temperature Probe (Option)
— Apply a film of thermal conductivity paste (no FLEXIM supply) onto the contact sur- face of the temperature probe. — Take the spring end of the ball chain and insert the last ball in one of the slots on the top of the temperature probe.
-
Page 75: Connection Of The Temperature Probe
8 Installation of the Temperature Probe (Option) Connection of the Temperature Probe • Connect the temperature probe with the temperature input of the flowmeter. Tabelle 8.1: Terminal strip KL2 of the flowmeter terminal connection of connection of temperature probe extension cable T1a…T4a T1A…T4A red/blue…
-
Page 76: Start-Up
9 Start-up Start-up Switching on As soon as the flowmeter is connected with the power sup- FLEXIM FLUXUS ply it will be displayed which transducer has been detected G70X-XXXXXXXX at which channel. Afterwards the serial number of the flowmeter will be dis- played for a short time.
-
Page 77
9 Start-up 9.2.2 The Program Branches and their Menu Items • The pipe and medium parameters will be input in the program branch PARAMETER. • The steps for the measurement will be processed in the program branch MEASURING. • Physical quantity and unit of measurement as well as all parameters necessary for the measured value output will be defined in the program branch OUTPUT OPTIONS. -
Page 78: Hotcodes
9 Start-up 9.2.4 Navigation If a vertical arrow is displayed, the menu item contains a scroll list. The current list item is displayed in the lower line. Scroll with key to select a list item in the low- Parameter for Channel er line.
-
Page 79: Language Selection
ROM as soon as the measurement begins. The operation of the flowmeter will be inter- rupted if the power supply fails. All input data remain saved.. After return of the power supply, the serial number will be FLEXIM FLUXUS displayed for a few seconds. G70X-XXXXXXXX The stopped measurement will be continued.
-
Page 80: Basic Measurement
10 Basic Measurement Basic Measurement The pipe and medium parameters will be entered for the selected measuring point (see chapter 5). The ranges are limited by the properties of the transducers and of the flowme- ter. The transducers have to be connected to the flowmeter during input Note! of the parameters.
-
Page 81
10 Basic Measurement 10.1.2 Pipe Wall Thickness Enter the pipe wall thickness. The range depends on Wall Thickness the connected transducers. Press ENTER. The inner pipe diameter (= outer pipe diameter — 2x pipe wall thick- Note! ness) will be calculated internally. If the value is not within the inner pipe diameter range of the connected transducers, an error message will be displayed. -
Page 82: Input Of The Medium Parameters
10 Basic Measurement If OTHER MATERIAL is selected, the sound velocity has to be entered. Enter the sound velocity of the lining material. Press c-Material ENTER. 3200.0 For the sound velocity of some materials see annex C, Table C.1. Enter the thickness of the liner. Press ENTER. Liner Thickness The inner pipe diameter (= outer pipe diameter — 2x pipe wall thick- Note!
-
Page 83
10 Basic Measurement • min. and max. sound velocity • kinematic viscosity • density • gas compressibility factor 10.2.1 Sound Velocity The sound velocity of the medium is used for the calculation of the transducer distance at the beginning of the measurement. However, the sound velocity does not influence the measuring result directly. -
Page 84: Other Parameters
Select SPECIAL VERSION to enter the transducer pa- rameters. The parameters have to be provided by the transducer manufacturer. Press ENTER. If standard transducer parameters are used, FLEXIM can not guaran- Note! tee for the precision of the measured values. A measurement might be even impossible.
-
Page 85: Selection Of The Channels
10 Basic Measurement 10.3.2 Extension Cable If the transducer cable has to be extended, enter the Additional cable additional cable length (e.g. between the junction box 65.0 and the flowmeter). Press ENTER. 10.4 Selection of the Channels The channels on which will be measured can be activated individually. This is only possi- ble if the flowmeter has more than one measuring channel.
-
Page 86: Define Number Of Sound Paths
10 Basic Measurement 10.5 Define Number of Sound Paths The number of transits of the ultrasonic waves through the medium depends on the placement of the transducers on the pipe. If the number of transits is odd (diagonal mode), the transducers will be mounted on op- posite sides of the pipe.
-
Page 87: Installation Of The Damping Mats
10 Basic Measurement 10.6 Installation of the Damping Mats Before the transducers are mounted, damping mats will be installed. The damping mats reduce propagation of noise in the pipe wall. • Select the measuring point according to the recommendations in chapter 5. •…
-
Page 88
10 Basic Measurement • Fix the damping mat to the pipe: remove the protection sheet bit by bit from the damp- ing mat and attach the damping mat to the pipe at the same time. Air pockets must be avoided. Note! improvement of adhesion and removal of air pockets: •… -
Page 89
10 Basic Measurement The damping mat adheres strongly to the pipe. Later alignment is not Note! possible. The damping mat has to be positioned correctly immediate- The damping mat can not be reused after removal. Note! • Depending on the transducer type, another layer of the damping mat is necessary (see annex A, Technical Data). -
Page 90
10 Basic Measurement • Cut the damping mat to size, if necessary (see annex A, section Technical Data). • Apply the a bead of ultrasound gel along the center line onto the inner side of the damping mat. • Spread the ultrasound gel evenly on the surface of the damping mat by means of a spattle (see Fig. -
Page 91: Mounting And Positioning The Transducers
10 Basic Measurement 10.7 Mounting and Positioning the Transducers Observe the safety instructions in chapter 2. The transducers will be fixed to the pipe by means of the supplied transducer mounting fixture. • Install the damping mats (see section ).10.6 •…
-
Page 92
10 Basic Measurement 10.7.2 Mounting the Transducers with Mounting Shoes and Tension 10 Basic Measurement Straps Mounting the Transducers with Clasps • Cut the tension straps (pipe circumference + 120 mm). • Make sure that part (2) of the clasp is on top of part (1) (see Fig. 10.8). The hooks of part (2) must be on the outer side of the clasp. -
Page 93
10 Basic Measurement • Mount the second transducer in the same way. Adjust the displayed transducer dis- tance by means of the measuring tape. • Tighten the screws of the clasps. • Push the transducers in the mounting shoes. Press the transducer firmly on the pipe. There should be no air pockets between transducer surface and pipe wall. -
Page 94
Place the clasp on the side of the pipe. Turn the tension strap around the pipe (see Fig. 10.12) Fig. 10.12: Mounted transducer in mounting shoe, fixed with FLEXIM clasp UMG70XV3-4EN 12.01.2009… -
Page 95
10 Basic Measurement The clasp and the metal spring must be in contact completely with Note! the pipe to ensure a good fixation. • Place the mounting shoe on the pipe. Hold clasp and mounting shoe with one hand while pushing the tension strap through parts (2) and (1) of the clasp (see Fig. 10.11). •… -
Page 96
(only FLUXUS G709). Shift a transducer slightly in the range of the recommend- ed transducer distance • FLUXUS G704: until the bar graph has max. length (max. 6 squares) • FLUXUS G709: until the LED of the measuring channel lights green. -
Page 97
10 Basic Measurement 10.7.4 Consistency Check If a wide range for the sound speed has been entered in the program branch PARAMETER or the exact parameters of the medium are not known, a consistency check is recom- mended. The transducer distance can be displayed during measurement by scrolling with key DISP The optimum transducer distance (here: 50.0 mm) will be L=(50.0) 54.0… -
Page 98: Start Of Measurement
10 Basic Measurement Enter the new adjusted transducer distance. Press EN- Transd.Distance? TER. 50.0 Scroll with key again until the transducer distance is DISP L=(51.1) 50.0 mm displayed and check the difference between optimum and 54.5 m3/h entered transducer distance. Repeat the steps if neces- sary.
-
Page 99: Detection Of Flow Direction
10 Basic Measurement The measuring time depends on the measuring conditions. If, e.g. the measuring signal can not be detected immediately, the measurement might be > 1 s. The outputs and the serial interface continuously get the measuring results of the as- signed channel.
-
Page 100: Displaying The Measured Values
11 Displaying the Measured Values Displaying the Measured Values The physical quantity will be set in the program branch OUTPUT OPTIONS (see section 11.1). During measurement, the designation of the physical quantity will be displayed in the up- per line, its value in the lower line. The display can be adapted (see section 11.3). 11.1 Selection of the Physical Quantity and of the Unit of Measurement The following quantities can be measured:…
-
Page 101: Toggling Between The Channels
11 Displaying the Measured Values 11.1.1 Measurement of the Standard Volume Flow If the medium is a gas, the standard volume flow can be selected as physical quantity ad- ditionally to the operating volume flow. The standard volume flow will be calculated as follows: = V * p/p /T * 1/K…
-
Page 102: Adjustment Of The Display
11 Displaying the Measured Values The AutoMux mode is the default display mode. It will be activated automatically after a cold start. All Channels The measured values of all channels (measuring and calculation channels) will be dis- played. The next active channel is selected after min. 1.5 s. Only Calculation Channels The measured values of the calculation channels will be displayed only.
-
Page 103: Status Line
11 Displaying the Measured Values The following information can be displayed in the lower line: • flow velocity • sound velocity • mass flow • volume flow • heat flow DISP The display in the upper line can be changed during measurement with key .
-
Page 104: Transducer Distance
11 Displaying the Measured Values flow profile information about the flow profile based on the Reynold’s number fully turbulent flow profile fully laminar flow profile the flow is in the transition range between laminar and turbulent flow unknown, can not be calculated flow velocity comparison of the measured flow velocity with the flow limits of the system …
-
Page 105: Advanced Measuring Functions
12 Advanced Measuring Functions Advanced Measuring Functions 12.1 Damping Factor Each displayed measured value is the floating average of all measured values of the last x seconds, where x is the damping factor. A damping factor of 1 s means that the mea- sured values are not averaged as the measuring rate is approx 1/s.
-
Page 106
12 Advanced Measuring Functions Table 12.1: Keys for display of the totalizers activation press key once during measurement deactivation press key three times during measurement display of the totalizer for the press key once during measurement positive flow direction display of the totalizer for the press key once during measurement negative flow direction… -
Page 107
12 Advanced Measuring Functions When the Measurement is Interrupted The behavior of the totalizers after an interruption of the measurement or after a RESET of the flowmeter will be set in SPECIAL FUNCTION\SYSTEM SETTINGS\MEASUR- ING\QUANTITY RECALL. If ON is selected, the totalizer values will be stored and Quantity recall used for the next measurement. -
Page 108: Upper Limit Of The Flow Velocity
12 Advanced Measuring Functions The overflow of a totalizer influences all output channels, e.g. data Note! memory, online output. The output of the sum of both totalizers (the throughput Q) via an output will not be valid after the first overflow (wrapping) of one of the corresponding totalizers.
-
Page 109: Cut-Off Flow
12 Advanced Measuring Functions 12.4 Cut-off Flow The cut-off flow function automatically sets all measured flow velocities to zero that are below a preset value. All values derived from this measured value will also be set to zero. The cut-off flow can depend on the flow direction or not. The cut-off value will be set in SPECIAL FUNCTION\SYSTEM SETTINGS\MEASURING\CUT-OFF FLOW.
-
Page 110: Uncorrected Flow Velocity
12 Advanced Measuring Functions 12.5 Uncorrected Flow Velocity For special applications, the uncorrected flow velocity might be of interest. The profile correction of the flow velocity will be activated in SPECIAL FUNCTION\SYS- TEM SETTINGS\MEASURING\FLOW VELOCITY. Select NORMAL to display and output the flow velocity Flow Velocity with profile correction.
-
Page 111: Measurement Of Transient Processes (Fastfood Mode)
12 Advanced Measuring Functions 12.6 Measurement of Transient Processes (FastFood mode) The FastFood mode allows to measure flows with high dynamics. The continuous adap- tation to changing measuring conditions which takes place in the normal measuring mode is only partially realized in FastFood mode. The sound velocity of the medium is not measured.
-
Page 112: Calculation Channels
12 Advanced Measuring Functions The values of the current measuring data set will be deleted when Note! the FastFood mode is deactivated and activated again without inter- rupting the measurement. The values of the current measuring data set are kept if the measure- Note! ment has been interrupted before the FastFood mode is activated again.
-
Page 113
12 Advanced Measuring Functions Three scroll lists are displayed in the upper line: >CH1< funct ch2 • selection of the first measuring channel (CH1) • selection of the calculation function (FUNCT) • selection of the second measuring channel (CH2) Select a scroll list with key The list items will be displayed in the lower line. -
Page 114
12 Advanced Measuring Functions example: The difference of the volume flows of the channels A and B has to be deter- mined. The physical quantity of measuring channel A can be the volume flow or the mass flow, but not the flow velocity. The physical quantities of the two measuring channels do not need to be identical (channel A = mass flow, channel B = volume flow). -
Page 115: Change Of Limit For The Inner Pipe Diameter
12 Advanced Measuring Functions If a calculation channel has been activated, HumanMux Y:Flow Velocity mode (see section 11.2.2) is selected at the beginning of 53.41 the measurement and the values of the calculation chan- nel are displayed. If AutoMux mode is selected, the measured values of the measuring channels, but not of the calculation channels will be displayed alternately.
-
Page 116
12 Advanced Measuring Functions Select in the program branch SPECIAL FUNCTION the Special Funct. menu item PROGRAM CODE. Set program code Enter a program code with max. 6 digits. Press ENTER. Program Code —— An error message is displayed if a reserved number has INVALID CODE! been entered (e.g. -
Page 117: Storing And Output Of Measured Values
13 Storing and Output of Measured Values Storing and Output of Measured Values Storing The following data can be stored: • date • time • measuring point number • pipe parameters • medium parameters • transducer parameters • sound path (reflection or diagonal) •…
-
Page 118: Data Memory
13 Storing and Output of Measured Values 13.1 Data Memory 13.1.1 Activating/Deactivating of the Data Memory Select in the program branch OUTPUT OPTIONS the Output Options channel for which the data memory has to be activated. for Channel Press ENTER. This display will not be indicated, if the flowmeter has only one measuring channel.
-
Page 119: Output Of The Measured Values
13.2 Output of the Measured Values The measured values can be output via the serial interface. For the connection of the in- terface to the flowmeter see section 6.9 (FLUXUS G704), 7.7 (FLUXUS G709). 13.2.1 RS232 Interface • output of the measured values in ASCII format •…
-
Page 120
13 Storing and Output of Measured Values 13.2.3 Online Output The measured values will be transmitted via the serial interface to a PC directly during measurement. If the data memory is activated, the measured values will be stored addi- tionally. •… -
Page 121
13 Storing and Output of Measured Values Fig. 13.1: Selection of the serial interface Fig. 13.2: Receive measured values UMG70XV3-4EN 12.01.2009… -
Page 122
13 Storing and Output of Measured Values Fig. 13.3: Display of the received measuring data sets Offline Output to a Terminal Program • Select the program branch SPECIAL FUNCTION on the flowmeter. Press ENTER. Select the list item PRINT MEAS.VAL. Press ENTER. Special Function… -
Page 123
13 Storing and Output of Measured Values 13.2.5 Data Format The header will be transmitted at the beginning of the measurement. The first four lines contain general information about the flowmeter and the measurement. The following lines contain the configuration parameters output for each channel in a data block. example: \DEVICE :G70X-XXXXXXXX… -
Page 124
13 Storing and Output of Measured Values One data line is transmitted per storage interval and per activated measuring channel. The line «???» will be transmitted if there are no measured values availabe for the stor- age interval. example: With a storage rate of 1 s, 10 lines «???» will be transmitted, if the mea- surement is restarted after an interruption for transducer positioning of 10 s. -
Page 125
13 Storing and Output of Measured Values Offline Output During offline output, columns will only be generated if at least one measured value is stored in the data set. The columns Q_POS and Q_NEG will not be generated if the total- izers are deactivated. -
Page 126: Delete The Measured Values
13 Storing and Output of Measured Values Select the serial interface for the offline output. Send Offline via This display will be indicated only if the flowmeter has RS232 >RS485< an RS485 interface. 13.3 Delete the Measured Values Select the program branch SPECIAL FUNCTION. Press ENTER. Select the list item DELETE MEAS.VAL.
-
Page 127
13 Storing and Output of Measured Values 13.4.2 Sample Mode Select the sample mode. Storage mode If SAMPLE is selected, the displayed measured value >SAMPLE< average will be used for storing and online output. If AVERAGE is selected, the mean value of all measured values during a storage interval will be used for storing and online output. -
Page 128: Available Data Memory
13 Storing and Output of Measured Values 13.4.4 Storing of the Signal Amplitude Select SPECIAL FUNCTION\SYSTEM SETTINGS\STORING\STORE AMPLITUDE. If ON is selected and the data memory is activated, the Store Amplitude amplitude of the measured signal will be stored to- >ON<…
-
Page 129: Libraries
14 Libraries Libraries The internal material database of the flowmeter contains parameters for pipe and lining materials as well as for media. It can be extended by user defined materials or media. User defined materials and media will always be displayed in the scroll lists of the pro- gram branch PARAMETER.
-
Page 130
14 Libraries This error message will be displayed if the entered number MAXIMAL: of data sets for a category of material data exceeds the ca- Materials: pacity of the coefficient memory. Enter the number of the user defined materials. Press EN- Format USER-AREA TER. -
Page 131: Input Of Material/Medium Properties Without The Extended Library
14 Libraries 14.2 Input of Material/Medium Properties without the Extended Library To enter the material/medium properties as constants, the extended library has to be de- activated. Select in SPECIAL FUNCTION\SYSTEM SETTINGS\LI- Libraries BRARIES the list item EXTENDED LIBRARY. Press EN- Extended Library Select OFF to deactivate the extended library.
-
Page 132: Extended Library
14 Libraries Enter the roughness of the material. Press ENTER. Roughness For the typical roughness of some materials see annex C, Table C.2. Medium Properties Enter the min. and max. value of the sound velocity for the c-Medium medium in m/s. Press ENTER. 1400.0 c-Medium 1550.0…
-
Page 133
If, however, the process conditions fluctuate strongly and the medium properties depend strongly on the temperature (e.g. viscosity of hydraulic oil), polynomials or customized in- terpolation functions should be used. Contact FLEXIM to find the best solution for the measuring task. -
Page 134
Y=M*X+N: linear function of the temperature Y=POLYNOM: y = k x + k Y=F(X,Z): customized interpolation function (only for ex- perienced users or after consultation of FLEXIM) … GO BACK: return to the precedent menu item Select a user defined material/medium. USER Material… -
Page 135
14 Libraries Medium Properties Enter for the medium: • longitudinal sound velocity (in m/s) • kinematic viscosity (in mm • density (in kg/m • gas compressibility factor. One to five values depending on the selected function have to be entered. Press ENTER after each input. -
Page 136: Deleting A User Defined Material/Medium
14 Libraries Enter the 10 heat flow coefficients: a0…a4, r0…r4. Press Heat-flow Coeffs ENTER after each input. 0.0 a0 Select YES to store the heat flow coefficients. Press EN- Heat-flow Coeffs TER. Save? no >YES< 14.3.5 Input of Steam and Concentration Coefficients Use the program FluxKoef (option).
-
Page 137
14 Libraries Select MATERIAL LIST to edit the material scroll list or Libraries MEDIUM LIST to edit the medium scroll list. Material list Select …GO BACK to return to SYSTEM SETTINGS. Press ENTER. Select FACTORY if all materials/media of the internal data- Material list base are to be displayed in the scroll list. -
Page 138
14 Libraries The materials/media are displayed in the order in which they have Note! been added. 14.5.3 Adding all Materials/Media to the Scroll List Select ADD ALL. Press ENTER to add all materials/media Material list of the internal database to the current scroll list. >Add all 14.5.4 Removing a Material/Medium from the Scroll List Select REMOVE MATERIAL or REMOVE MEDIUM to remove… -
Page 139: Settings
15 Settings Settings 15.1 Time and Date The flowmeter has a battery buffered clock. Measured values will be automatically stored with date and time. 15.1.1 Time Select in SPECIAL FUNCTION\SYSTEM SETTINGS the SYSTEM settings list item SET CLOCK. Press ENTER. Set Clock The actual time is displayed.
-
Page 140
15 Settings 15.2.1 Pipe Circumference Select ON if the pipe circumference has to be entered in- Pipe Circumfer. stead of the pipe diameter in the program branch PARAM- >ON< ETER. Press ENTER. If ON is selected for PIPE CIRCUMFER., the outer pipe di- Outer Diameter ameter will be requested in the program branch PARAME- 100.0… -
Page 141
15 Settings 15.2.2 Medium Pressure The dependency of the properties of a medium on the pressure can be taken into ac- count. This display is indicated only if in SPECIAL FUNCTION\SYSTEM SETTINGS\MEASUR- ING the GAS MEASURING has been deactivated. If GAS MEASURING is activated, the medium pressure will always be requested in the program branch PARAMETER. -
Page 142: Measurement Settings
15 Settings 15.2.5 Steam in Inlet Select ON if the medium in the supply line may be vapor- Steam in inlet ous for the heat flow measurement of a liquid medium. In >ON< the program branch PARAMETER, the supply pressure has to be entered then.
-
Page 143
15 Settings This menu item will be displayed only, if a WaveInjector is Wave Injector in the scope of supply (see user manual of the WaveInjec- >ON< tor). Select YES if the actually measured sound velocity is to be Compare c-fluid compared to the theoretical or expected value. -
Page 144: Settings Of The Standard Conditions For The Gas Measurement
15 Settings 15.4 Settings of the Standard Conditions for the Gas Measurement Select SPECIAL FUNCTION\SYSTEM SETTINGS\GAS- SYSTEM settings MEASURING. Gas-Measuring This display is indicated only if in SPECIAL FUNC- TION\SYSTEM SETTINGS\MEASURING the gas measur- ing has been activated. Enter the pressure for the local standard conditions. Normal pressure 1.01325 Enter the temperature for the local standard conditions.
-
Page 145: Instrument Information
15 Settings 15.6 Instrument Information Select SPECIAL FUNCTION\INSTRUM. INFORM. to ob- Special Function tain information about the flowmeter. Press ENTER. Instrum. Inform. Type and serial number of the flowmeter will be displayed G70X-XXXXXXXX in the upper line. Free: 18327 The available data memory will be displayed in the lower line (here: 18 327 measured values can be stored).
-
Page 146: Superuser-Mode
16 SuperUser-Mode SuperUser-Mode The SuperUser mode allows experimental work. Features of the SuperUser mode are: • Defaults will not be observed. • There are no plausibility checks when parameters are being entered. • There is no check whether the entered parameters are within the limit determined by physical laws and technical data.
-
Page 147: Malfunctions In Superuser Mode
16 SuperUser-Mode 16.3 Malfunctions in SuperUser Mode As the SuperUser mode operates without any plausibility check, absurd entries may re- sult in an automatic switching-off of the flowmeter or in a crash of the internal software. An absurd input is e.g. 0 (zero) for the number of sound paths or 0.1 mm for the outer pipe diameter.
-
Page 148: Inputs
LINK TEMPERATURE linking other inputs to measuring channels LINK OTHERS selection of a temperature probe (FLUXUS G704) PT100/PT1000 return to the precedent menu item …GO BACK 17.1 Linking the Temperature Inputs to the Measuring Channels 17.1.1 Linking the Temperature Inputs…
-
Page 149: Selection Of The Temperature Probe (Fluxus G704)
The configuration of a measuring channel will be stored when the Note! next channel is selected. The configuration dialog of a channel has to be finished to store the changes. 17.2 Selection of the Temperature Probe (FLUXUS G704) Select SPECIAL FUNCTION\SYSTEM SETTINGS\PROC. SYSTEM settings INPUTS. Press ENTER.
-
Page 150: Activation Of The Inputs
17 Inputs The configuration of a measuring channel will be stored when the Note! next channel is selected. The configuration dialog of a channel has to be finished to store the changes. 17.4 Activation of the Inputs The activation of the inputs in program branch OUTPUT OPTIONS will Note! be displayed only if the flowmeter has inputs of the corresponding type and they are linked to a measuring channel.
-
Page 151: Temperature Correction
17 Inputs 17.5 Temperature Correction A correction value (offset) for the temperature can be defined for each temperature input. If a correction value is defined, it will be automatically added to the measured tempera- ture. This function is useful if e.g.: •…
-
Page 152: Outputs
18 Outputs Outputs If the flowmeter is equipped with outputs, they have to be installed and activated before they can be used: • assigning a measuring channel (source channel) to the output (if the flowmeter has more than one measuring channel) •…
-
Page 153
18 Outputs The source items and their scroll lists are described in Table 18.1. Table 18.1: Configuration of the outputs source item list item output physical quantity selected in the program branch MEASURED VALUE ACTUAL MEASURE OUTPUT OPTIONS flow independently of the physical quantity select- FLOW ed in the program branch OUTPUT OPTIONS heat flow independently of the physical quantity… -
Page 154
18 Outputs 18.1.1 Output Range When configuring an analog output, the output range will I1:Output range be defined now. Select a list item or OTHER RANGE.. to 4/20 enter the output range manually. If OTHER RANGE.. is selected, enter the values OUTPUT … -
Page 155
18 Outputs Table 18.3: Examples for the error output list item for the error output output signal I [ m A ] Error-value Minimum (4.0mA) I [ m A ] Error-value Hold last value I [ m A ] Error-value Maximum (20.0mA) … -
Page 156
18 Outputs Select a list item for the error output. Press ENTER. Error-value Minimum (4.0mA) If OTHER VALUE … is selected, enter an error value. It Error-value has to be within the limits of the output. Press ENTER. The settings will be stored now at the end of the dialog. Note! The terminals to be used for the connection of the output I1 active loop… -
Page 157: Error Value Delay
18 Outputs 18.2 Error Value Delay The error value delay is the time interval after which the error value will be transmitted to the output in case no valid measured values are available. The error value delay can be entered in the program branch OUTPUT OPTIONS if this menu item has been previously activated in the program branch SPECIAL FUNCTION.
-
Page 158: Activation Of A Pulse Output
18 Outputs Enter in FULL-SCALE VAL. the highest measured value Full-Scale Val. expected. 300.00 m3/h FULL-SCALE VAL. is the measured value corresponding to the upper limit of the output range as defined in section 18.1.1. example: The output range 4…20 mA was selected for a current loop, the zero-scale value was set to 0 m /h and the full-scale value to 300 m If the volume flow is 300 m…
-
Page 159: Activation Of An Alarm Output
18 Outputs The max. flow that the pulse output can work with will be displayed now. This value is cal- culated from the entered pulse value and pulse width. If the flow exceeds this value, the pulse output will not function properly. In this case, pulse value and pulse width have to be adapted to the flow conditions.
-
Page 160
18 Outputs Select YES to activate the alarm output. Press ENTER. Alarm Output >YES< Three scroll lists will be displayed: R1=FUNC<typ mode • FUNC for the switching condition Function: • TYP for the holding behavior • MODE for the switching function A scroll list will be selected in the upper line with key . -
Page 161
18 Outputs example 2: lower limit = -10 m A measured value of e.g. -11 m /h falls below the limit. The alarm switch- A measured value of e.g. -9.9 m /h does not fall below the limit. The alarm does not switch. -
Page 162: Behavior Of The Alarm Outputs
18 Outputs 18.6 Behavior of the Alarm Outputs 18.6.1 Apparent Switching Delay Measured values and totalizer values will be displayed rounded to two decimal places. The limits, however, will be compared to the non-rounded measured values. This might cause an apparent switching delay when the measured value changes marginally (less than two decimal places).
-
Page 163
18 Outputs Alarm outputs with switching condition ERROR will be activated only after several unsuc- cessful measuring attempts. Therefore, typical short-term disturbances of the measure- ment (e.g. switching on of a pump) will not activate the alarm. Alarm outputs with switching condition +- -+ and of type NON-HOLD will be activated with each change of the flow direction for approx. -
Page 164: Deactivating The Outputs
18 Outputs Table 18.8: Pictograms for the alarm state indication switching holding behavior switching actual condition (FUNC) (TYPE) function (MODE) state closed NON-HOLD NO CONT. open HOLD NC CONT. + — — + QUANTITY ERROR 18.7 Deactivating the Outputs If the programmed outputs are no longer required, they can be deactivated.
-
Page 165: Troubleshooting
• the transducers were used for a prolonged period at a high temperature (several months >130 °C for normal transducers or > 200 °C for high temperature transducers). The flowmeter has to be sent to FLEXIM for recalibration under reference conditions. The display does not work at all or always fails Make sure that the correct voltage is available at the terminals.
-
Page 166: Problems With The Measurement
Date and time are wrong, the measured values will be deleted when the flowmeter is switched off The data backup battery has to be replaced. Send the flowmeter to FLEXIM. An output does not work Make sure that the outputs are configured correctly. Check the function of the output as described in section 18.1.3.
-
Page 167: Correct Selection Of The Measuring Point
Otherwise, the pressure of the transducer will be insufficient when the tempera- ture is low. It is recommended to use FLEXIM clasps with springs if the tempera- Note! ture fluctuates strongly. Contact FLEXIM.
-
Page 168: Maximum Acoustic Contact
Measurements on porous pipe materials (e.g. concrete or cast iron) are only possi- ble under certain conditions Contact FLEXIM. The pipe liner may cause problems during measurement if it is not attached tightly to the inner pipe wall or consists of acoustically absorbing material Try measuring on a liner free section of the pipe.
-
Page 169: Problems With The Totalizers
19 Troubleshooting The flow velocity to be measured is outside the measuring range of the flowmeter The measuring point is not appropriate Select another measuring point to check whether the results are better. The cross-sec- tion of the pipe is never perfectly circular, thus influences the flow profile. Change the transducer position according to the pipe deformation.
-
Page 170: A Technical Data
A Technical Data Flowmeter FLUXUS G704 G704 A2 G709 design standard field device field device for ATEX 19 » module zone 2 measurement measuring principle transit time difference correlation principle flow velocity 0.01…35 m/s, pipe diameter dependent repeatability 0.15 % of reading ±0.01 m/s accuracy — volume flow ±…
-
Page 171
FLUXUS G704 G704 A2 G709 measuring functions physical quantities operational volume flow, standard volume flow, mass flow, flow velocity totalizers volume, mass calculation functions average, difference, sum data logger loggable values all physical quantities and totalized values capacity > 100 000 measured values communication interface — process integration:… -
Page 172
FLUXUS G704 G704 A2 G709 binary output Reed relay 48 V/0.25 A open collector (OC) 24 V/4 mA optorelay 32 V/100 mA binary output as alarm output — functions limit, change of flow direction or error limit, change of flow direction or error binary output as pulse output… -
Page 173
Dimensions (in mm) FLUXUS G704 M20 (6x) screw holes for wall mount- Ø 4.5 UMG70XV3-4EN 12.01.2009… -
Page 174
FLUXUS G709 198.1 208.25 dimensions (in mm) UMG70XV3-4EN 12.01.2009… -
Page 175: Transducers
A Technical Data Transducers Shear Wave Transducers (for ATEX zone 1) technical type GDK1N31 GDM1N31 GDG1N31 order code GSK-NA1TS GSM-NA1TS GSG-NA1TS transducer frequency MHz 0.5 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 metal pipe: 30 plastic pipe: 1 plastic pipe: 1 plastic pipe: 1 outer pipe diameter…
-
Page 176
A Technical Data technical type GDK1N31 GDM1N31 GDG1N31 explosion protection transducer GSK-NA1TS GSM-NA1TS GSG-NA1TS zone explosion protection temperature min. °C max. °C +180 +120 +180 marking 0044; II 2G 0044; II 2G 0044; II 2G Ex q II T6…T3 EEx m II T6…T4 Ex q II T6…T3 Ta -40…+180 °C Ta -20…+120 °C… -
Page 177
A Technical Data Shear Wave Transducers (for ATEX zone 1) technical type GDP1N31 GDQ1N31 order code GSP-NA1TS GSQ-NA1TS transducer frequency MHz 2 medium pressure min. extended min. metal pipe: 30 plastic pipe: 1 outer pipe diameter min. extended min. recommended max. -
Page 178
A Technical Data technical type GDP1N31 GDQ1N31 explosion protection transducer GSP-NA1TS GSQ-NA1TS zone explosion protection temperature min. °C max. °C +120 +120 marking 0044; II 2G 0044; II 2G EEx m II T6…T4 EEx m II T6…T4 Ta -20…+120 °C Ta -20…+120 °C certification IBExU98ATEX… -
Page 179
A Technical Data Shear Wave Transducers (for ATEX zone 1, IP 68) technical type GDK1LI1 GDM2LI1 GDP2LI1 GDG1LI1 order code GSK-NA1TS/ GSM-NA1TS/ GSP-NA1TS/ GSG-NA1TS/ IP68 IP68 IP68 IP68 transducer frequency MHz 0.5 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 metal pipe: 30 metal pipe: 30… -
Page 180
A Technical Data technical type GDK1LI1 GDM2LI1 GDP2LI1 GDG1LI1 explosion protection transducer GSK-NA1TS/ GSM-NA1TS/ GSP-NA1TS/ GSG-NA1TS/ IP68 IP68 IP68 IP68 zone explosion protection temperature min. °C max. °C +180 +180 +180 +180 marking 0044; II 2G 0044; II 2G 0044; II 2G 0044;… -
Page 181
A Technical Data Shear Wave Transducers (for ATEX zone 1/2 (gas/dust), high temperature) technical type GDM2E85 GDP2E85 order code GSM-EA1TS GSP-EA1TS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 plastic pipe: 1 plastic pipe: 1 outer pipe diameter min. -
Page 182
A Technical Data technical type GDM2E85 GDP2E85 explosion protection transducer GSM-EA1TS GSP-EA1TS zone explosion protection temperature min. °C max. °C +225 +225 marking 0044; II 2G 0044; II 2G Ex eq II T6…T2 Ex eq II T6…T2 Ta -45…+225 °C Ta -45…+225 °C II 3D Ex tD II 3D Ex tD… -
Page 183
A Technical Data Shear Wave Transducers (for ATEX zone 2, FM or without Explosion Protection) technical type GDK1N52 GDM1N52 GDP1N52 GDG1N52 order code GSK-NA2TS GSM-NA2TS GSP-NA2TS GSG-NA2TS GSK-NF2TS GSM-NF2TS GSP-NF2TS GSG-NF2TS GSK-NNNTS GSM-NNNTS GSP-NNNTS GSG-NNNTS transducer frequency MHz 0.5 medium pressure min. -
Page 184
A Technical Data technical type GDK1N52 GDM1N52 GDP1N52 GDG1N52 explosion protection transducer GSK-NA2TS GSM-NA2TS GSP-NA2TS GSG-NA2TS zone explosion protection temperature min. °C max. °C +190 +130 +130 +190 marking II 3G II 3G II 3G II 3G Ex nA II T6…T3 Ex nA II T6…T4 Ex nA II T6…T4 Ex nA II T6…T3… -
Page 185
A Technical Data Shear Wave Transducers (for ATEX zone 2 and without Explosion Protection, IP technical type GDK1LI8 GDM2LI8 GDP2LI8 GDG1LI8 order code GSK-NA2TS/ GSM-NA2TS/ GSP-NA2TS/ GSG-NA2TS/ IP68 IP68 IP68 IP68 GSK-NNNTS/ GSM-NNNTS/ GSP-NNNTS/ GSG-NNNTS/ IP68 IP68 IP68 IP68 transducer frequency MHz 0.5 medium pressure min. -
Page 186
A Technical Data technical type GDK1LI8 GDM2LI8 GDP2LI8 GDG1LI8 explosion protection transducer GSK-NA2TS/ GSM-NA2TS/ GSP-NA2TS/ GSG-NA2TS/ IP68 IP68 IP68 IP68 zone explosion protection temperature min. °C max. °C marking II 3G II 3G II 3G II 3G Ex nA II T6…T5 Ex nA II T6…T5 Ex nA II T6…T5 Ex nA II T6…T5… -
Page 187
A Technical Data Shear Wave Transducers (without Explosion Protection and with Connection System AS) technical type GDK1NZ7 GDM1NZ7 GDP1NZ7 GDG1NZ7 order code GSK-NNNAS GSM-NNNAS GSP-NNNAS GSG-NNNAS transducer frequency MHz 0.5 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 metal pipe: 30 metal pipe: 30 plastic pipe: 1… -
Page 188
A Technical Data Shear Wave Transducers (high temperature, without Explosion Protection and with Connection System AS) technical type GDM2EZ7 GDP2EZ7 order code GSM-ENNAS GSP-ENNAS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 plastic pipe: 1 plastic pipe: 1 outer pipe diameter min. -
Page 189
A Technical Data Shear Wave Transducers (high temperature, for ATEX zone 2 or without Explo- sion Protection) technical type GDM2E52 GDP2E52 GDQ2E52 order code GSM-EA2TS GSP-EA2TS GSQ-EA2TS GSM-ENNTS GSP-ENNTS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: 30 metal pipe: 30 plastic pipe: 1 plastic pipe: 1… -
Page 190
A Technical Data technical type GDM2E52 GDP2E52 GDQ2E52 explosion protection transducer GSM-EA2TS GSP-EA2TS GSQ-EA2TS zone explosion protection temperature min. °C max. °C +235 +235 +235 marking II 3G II 3G II 3G Ex nA II T6…T2 Ex nA II T6…T2 Ex nA II T6…T2 Ta -45…+235 °C Ta -45…+235 °C… -
Page 191
A Technical Data Lamb Wave Transducers (for ATEX Zone 1) technical type GRH1N33 GRK1N33 GRG1N33 order code GLH-NA1TS GLK-NA1TS GLG-NA1TS transducer frequency MHz 0.3 medium pressure min. extended metal pipe: 10 metal pipe: metal pipe: 10 10 (> DN 120) 5 (<… -
Page 192
A Technical Data technical type GRH1N33 GRK1N33 GRG1N33 explosion protection transducer GLH-NA1TS GLK-NA1TS GLG-NA1TS zone explosion protection temperature min. °C max. °C +140 +140 +140 marking 0044; II 2G 0044; II 2G 0044; II 2G Ex q II T6…T3 Ex q II T6…T3 Ex q II T6…T3 Ta -40…+140 °C Ta -40…+140 °C… -
Page 193
A Technical Data Lamb Wave Transducers (for ATEX Zone 1) technical type GRM1N83 GRP1N83 GRQ1N83 order code GLM-NA1TS GLP-NA1TS GLQ-NA1TS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: metal pipe: metal pipe: 10 (> DN 60) 10 (> DN 35) 10 (>… -
Page 194
A Technical Data technical type GRM1N83 GRP1N83 GRQ1N83 explosion protection transducer GLM-NA1TS GLP-NA1TS GLQ-NA1TS zone explosion protection temperature min. °C max. °C +140 +140 +140 marking 0044; II 2G 0044; II 2G 0044; II 2G Ex eq II T6…T3 Ex eq II T6…T3 Ex eq II T6…T3 Ta -55…+140 °C Ta -55…+140 °C… -
Page 195
A Technical Data Lamb Wave Transducers (for ATEX Zone 2, FM or without Explosion Protection) technical type GRH1N52 GRK1N52 GRG1N52 order code GLH-NA2TS GLK-NA2TS GLG-NA2TS GLH-NF2TS GLK-NF2TS GLG-NF2TS GLH-NNNTS GLK-NNNTS GLG-NNNTS transducer frequency MHz 0.3 medium pressure min. extended metal pipe: 10 metal pipe: metal pipe: 10 10 (>… -
Page 196
A Technical Data technical type GRH1N52 GRK1N52 GRG1N52 explosion protection transducer GLH-NA2TS GLK-NA2TS GLG-NA2TS zone explosion protection temperature min. °C max. °C +150 +150 +150 marking II 3G Ex II 3G Ex II 3G Ex nA II T6…T3 nA II T6…T3 nA II T6…T3 Ta -55…+150 °C Ta -55…+150 °C… -
Page 197
A Technical Data Lamb Wave Transducers (for ATEX Zone 2 or without Explosion Protection) technical type GRM1N52 GRP1N52 GRQ1N52 order code GLM-NA2TS GLP-NA2TS GLQ-NA2TS GLM-NNNTS GLP-NNNTS GLQ-NNNTS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: metal pipe: metal pipe: 10 (>… -
Page 198
A Technical Data technical type GRM1N52 GRP1N52 GRQ1N52 explosion protection transducer GLM-NA2TS GLP-NA2TS GLQ-NA2TS zone explosion protection temperature min. °C max. °C +150 +150 +150 marking II 3G II 3G II 3G Ex nA II T6…T3 Ex nA II T6…T3 Ex nA II T6…T3 Ta -55…+150 °C Ta -55…+150 °C… -
Page 199
A Technical Data Lamb Wave Transducers (without Explosion Protection and with Connection System AS) technical type GRH1NC3 GRK1NC3 GRG1NC3 order code GLH-NNNAS GLK-NNNAS GLG-NNNAS transducer frequency MHz 0.3 medium pressure min. extended metal pipe: 10 metal pipe: metal pipe: 10 10 (>… -
Page 200
A Technical Data Lamb Wave Transducers (without Explosion Protection and with Connection Sys- tem AS) technical type GRM1NC3 GRQ1NC3 GRP1NC3 order code GLM-NNNAS GLQ-NNNAS GLP-NNNAS transducer frequency MHz 1 medium pressure min. extended min. metal pipe: metal pipe: metal pipe: 10 (>… -
Page 201: Junction Box
A Technical Data Junction Box technical type JB01S4E3M JB02 JB03 dimensions see dimensional draw- see dimensional draw- see dimensional draw- fixation wall mounting wall mounting wall mounting option: option: option: 2 » pipe mounting 2 » pipe mounting 2 » pipe mounting material housing stainless steel 316L…
-
Page 202
A Technical Data 2 » Pipe Mounting Kit (option) Terminal Assignment JB01 Transducers terminal strip KL1 terminal connection signal shield T V S T R S shield K L 2 K L 1 signal Extension Cable (Flowmeter) terminal strip KL2 terminal connection signal… -
Page 203
A Technical Data JB02, JB03 Transducers terminal connection SMB connector SMB connector T V S T R S K L 2 Extension Cable (Flowmeter) terminal strip KL2 terminal connection signal shield shield signal UMG70XV3-4EN 12.01.2009… -
Page 204
Technical Data Technical Data Damping Mats (Option) Damping mats will be used for the gas measurement to reduce noise influences on the measurement. Transducer damping mats will be installed below the transducers. Pipe damping mats will be installed at reflection points, e.g. flange, welding. pipe damping mat transducer damping mat D — outer pipe diameter… -
Page 205
A Technical Data A Technical Data Units of Measurement standard/ flow mass flow totalizers sound heat heat flow operating velocity velocity quantity volume mass volume flow kg/h cm/s kg/min /min inch/s ml/min lb/d l/min lb/h lb/min hl/h lb/s hl/min hl/s Ml/d bbl/d bbl/h… -
Page 206
A Technical Data Flow Nomogram (metrical) volume flow volume flow V o l u m e n f l u s s V o l u m e n f l u s s 1 0 0 1 0 0 1 0 0 1 0 0 1 0 0… -
Page 207
A Technical Data Flow Nomogram (imperial) volume flow volume flow V o l u m e n f l u s s V o l u m e n f l u s s 1 0 0 1 0 0 1 0 0 1 0 0 1 0 0… -
Page 208: B Menu Structure
B Menu Structure Menu Structure cold start resistant Program Branch PARAMETER main menu: selection of the program branch >PAR< mea opt sf PARAMETER Parameter selection of a measuring channel (A, B) or of Parameter a calculation channel (Y, Z) for Channel This display will not be indicated, if the flow- meter has only one measuring channel.
-
Page 209
B Menu Structure cold start resistant selection whether the pipe is lined Lining >YES< selection of the lining material Lining This display is indicated only if LINING = YES Bitumen has been selected. input of the sound velocity of the lining mate- c-Material rial 3200.0… -
Page 210
B Menu Structure cold start resistant input of the kinematic viscosity of the medium kinem. Viscosity range: 0.01…30 000 mm 1.00 mm2/s This display is indicated only if OTHER ME- DIUM or NATURAL GAS has been selected. input of the operating density of the medium Density range: 0.1…20 000… -
Page 211
B Menu Structure cold start resistant input of the length of an extension cable Additional cable 65.0 When a calculation channel is selected (Y, Z) Calculation channels are only available if the flowmeter has more than one measuring channel. display of the current calculation function Calculation: Y= A — B selection of the calculation function… -
Page 212
B Menu Structure cold start resistant input of the number of sound paths A: Sound Path display of the transducer distance to be ad- Transd. Distance justed between the inner edges of the trans- A:54 mm Reflec ducers Program Branch OUTPUT OPTIONS main menu: selection of the program branch par mea >OPT<… -
Page 213
B Menu Structure cold start resistant input of the duration over which a floating av- Damping erage of the measured values has to be de- termined range: 1…100 s activation of the data memory Store Meas.Data >YES< activation of the measured value output via Serial Output the serial interface to a PC or a printer >YES<… -
Page 214
B Menu Structure cold start resistant input of the lowest/highest measured value to Zero-Scale Val. be expected for the current output 0.00 These values will be assigned to the lower/ upper limit of the output range. These displays are indicated only if CURRENT Full-Scale Val. -
Page 215
B Menu Structure cold start resistant Alarm Output activation of an alarm output Alarm Output This display is indicated only if an alarm out- >YES< put has been installed in SPECIAL FUNC- TION\SYSTEM SETTINGS\PROC. OUT- PUTS. selection of the switching condition (FUNC), of R1=FUNC<typ mode the holding behavior (TYP) and of the switch- Function:… -
Page 216
B Menu Structure cold start resistant Program Branch SPECIAL FUNCTION main menu: selection of the program branch par mea opt >SF< SPECIAL FUNCTION Special Function SYSTEM SETTINGS selection of SPECIAL FUNCTION\SYSTEM Special Funct. SETTINGS SYSTEM settings SYSTEM SETTINGS\SET CLOCK selection of the displays for the input of date SYSTEM settings … -
Page 217
B Menu Structure cold start resistant input of the number of user defined materials Format USER-AREA Materials: input of the number of user defined media Format USER-AREA Media: input of the number of user defined data sets Format USER-AREA for the heat flow coefficients Heat-Coeffs: input of the number of user defined data sets Format USER-AREA… -
Page 218
B Menu Structure cold start resistant SYSTEM SETTINGS\DIALOGS/MENUS selection of the displays for activation/deacti- SYSTEM settings vation or setting of menu items in the other Dialogs/Menus program branches activation of the menu item for the input of the Pipe Circumfer. pipe circumference in the program branch >ON<… -
Page 219
B Menu Structure cold start resistant activation of the menu item for the input of a Tx Corr.Offset correction value (offset) for each temperature >ON< input in the program branch MEASURING selection of the error value delay Error-val. delay • DAMPING: the damping factor will be used. damping >EDIT<… -
Page 220
B Menu Structure cold start resistant activation of the display for the difference be- Compare c-fluid tween measured and expected sound veloci- >YES< ty of a selected reference medium during measurement selection whether the flow velocity is dis- Flow Velocity played and output with or without profile cor- normal >UNCORR.<… -
Page 221
B Menu Structure cold start resistant input of an upper limit of the flow velocity Velocity limit range: 0.1…25.5 m/s All measured values exceeding the limit will be marked as outliers. Input of 0 (zero) switches off the detection for outliers. -
Page 222
B Menu Structure cold start resistant SYSTEM SETTINGS\PROC. OUTPUTS selection of the displays for the setting of the SYSTEM settings outputs of the flowmeter Proc. outputs selection of the output to be installed Install Output Current I1 SYSTEM SETTINGS\STORING selection of the displays for storing of mea- SYSTEM settings … -
Page 223
B Menu Structure cold start resistant activation of storing of the concentration of Store Concentr. the medium >ON< This display is indicated only if SYSTEM SETTINGS\MEASURING\ENABLE CONCEN- TR. (option) has been activated. The value will be stored only if the data mem- ory is activated. -
Page 224
B Menu Structure cold start resistant input of the instrument address Device adress: confirmation or change of the transmission Serial protocol parameters default >SETUP< change of the baud rate, parity or number of >BAUD< parity stop bits 1200 EVEN SYSTEM SETTINGS\MISCELLANEOUS selection of the display for the setting of the SYSTEM settings … -
Page 225
B Menu Structure cold start resistant start of the measured values transmission Send HEADER This display will be indicated only if measured ….values are stored in the data memory and the flowmeter is connected with the PC by a seri- al cable. -
Page 226
B Menu Structure cold start resistant input of the sound velocity of the material c-Material range: 600…6553.5 m/s 1590.0 input of the roughness of the material Roughness INSTALL MATERIAL with SPECIAL FUNCTION\SYSTEM SET- TINGS\LIBRARIES\EXTENDED LIBRARY = ON selection of the function for the temperature … -
Page 227
B Menu Structure cold start resistant input of the roughness of the material Roughness confirmation that the changes are to be Save changes stored >YES< This display is indicated only if a new material has been entered or the properties of an ex- isting material have been changed. -
Page 228
B Menu Structure cold start resistant input of the kinematic viscosity of the medium kinem. Viscosity range: 0.01…30 000.00 mm 1.01 mm2/s input of the operating density of the medium Density 60.00 kg/m3 INSTALL MEDIUM with SPECIAL FUNCTION\SYSTEM SET- TINGS\LIBRARIES\EXTENDED LIBRARY = ON selection of the function for the temperature … -
Page 229
B Menu Structure cold start resistant input of the gas compressibility factor GASFACTOR confirmation that the changes are to be Save changes stored >YES< This display is indicated only if a new medium has been entered or the properties of an ex- isting medium have been changed. -
Page 230: C Reference
The following tables provide assistance for the user. The accuracy of the data depends on the composition, the temperature and the manufacturing process of the material. FLEXIM does not assume liability for any inaccuracies. Table C.1: Sound Velocity of Selected Pipe and Lining Materials at 20 °C…
-
Page 231
C Reference Table C.2: Typical Roughness Coefficients of Pipes The values are based on experience and measurements. material absolute roughness [mm] drawn pipes of non-ferrous metal, 0…0.0015 glass, plastics and light metal drawn steel pipes 0.01…0.05 fine-planed, polished surface max. 0.01 planed surface 0.01…0.04 rough-planed surface… -
Page 232
C Reference Table C.3: Typical Properties of Selected Media at 20 °C and 1 bar medium sound velocity kinematic density viscosity [m/s] [g/cm acetone 1190 0.73 ammonia (NH 1386 0.613 gasoline 1295 0.88 beer 1482 0.998 BP Transcal LT 1365 20.1 0.876 BP Transcal N… -
Page 233
C Reference Table C.4: Properties of Methane medium medium density sound kinematic compressibility temperature pressure velocity viscosity factor [°C] [bar] [kg/m [m/s] [cSt] (AGA8-DC92) 31.177 415.43 0.358693909 0.9062727 29.683 425.18 0.38628171 0.9182674 28.354 434.39 0.414403611 0.928556 27.159 443.13 0.44309437 0.9374469 26.076 451.46 0.472426753… -
Page 234
C Reference Table C.5: Chemical Resistance of Autotex Autotex (keyboard) is resistant according to DIN 42115, part 2 against the following chemicals for a contact time of more than 24 h without visible modification: • ethanol • nitric acid <10 % •… -
Page 235: D Certificates
D Certificates Certificates UMG70XV3-4EN 12.01.2009…
-
Page 236
D Certificates UMG70XV3-4EN 12.01.2009… -
Page 240
D Certificates UMG70XV3-4EN 12.01.2009…
Особенности
• Диапазон измерений скорости потока 0,01…25 м/с
• Погрешность измерений ±1 % (опция: ± 0,5 %)
• Метод измерений: время-импульсный корреляционный метод и «зондирующий» (доплеровский), функция автоматического переключения между методами измерений
• Рабочие среды: Все звукопроводящие жидкости: вода, нефть, бензин, гидравлическое масло, конденсат, суспензии, кислоты, продукты нефтехимического производства с содержанием газообразных и твердых включений до 10 % от объема
• Материалы труб: все звукопроводящие материалы: стали, пластик, стекло и др.
Применение
• Измерение расхода питьевой и технологической воды
• Измерение расхода канализационных стоков
• Измерение расхода нефтепродуктов в процессах переработки нефти
• Измерение расхода агрессивных и ядовитых жидкостей: кислот, реагентов, в том числе в пластиковых и стеклянных трубах
• Измерение расхода питательной воды на АЭС при температурах до 400 °С
• Измерение количества тепла в процессах повышения энергетической эффективности крупных зданий и сооружений
Основные технические характеристики
Масса | 2,8 кг |
Электропитание | 100…240 В 50/50 Гц, 18…36 В пост. тока |
Рабочая температура |
Блока электроники: -10 °С…+60 °С Накладных датчиков: -40 °С…+200 °С (в зависимости от типа) |
Измерительные каналы | 1 или 2 канала |
Степень пылевлагозащиты | IP65 |
Аналоговых входов и выходов |
Basic: выходы: 1× токовый, 3× реле опции: выходы: 0/4…20 мА, 0…1 (10) В, 0…1 (10) кГц, оптореле |
Интерфейс связи | Modbus RTU (по интерфейсу RS485), HART |
Маркировка взрывозащиты | 2ExnAIIT6…T4 |
Накладные ультразвуковые датчики | диаметр трубы от 6 мм до 6500 мм, взрывоопасные зоны 1 или 2 |
Регистрационный номер Госреестра СИ РФ |
56831-14, межповерочный интервал 4 года |
Комплект поставки: расходомер жидкости F704, выбранные накладные датчики, акустическая паста/фольга/изоляционные маты, крепления датчиков на трубопровод, руководство по эксплуатации на русском языке, копии всех сертификатов, копии разрешительных документов.
Flexim FLUXUS G704 Measuring Instruments PDF User Guides and Manuals for Free Download: Found (1) Manuals for Flexim FLUXUS G704 Device Model (Operation & User’s Manual)
More Measuring Instruments Device Models:
-
RIDGID
Scout
Patents Pending• Français – 21• Castellano – pág. 41WARNINGRead this Operator’s Manualcarefully before using thistool. Failure to understandand follow the contents of thismanual may result in electri-cal shock, fire and/or seriouspersonal injury.Sc out™Operator’s ManualScoutTMSonde and Line Locator www. …
Scout Measuring Instruments, 66
-
Mod-tronic
Love Controls UPT 1000
Model UPT 1000 – Universal Process-Temperature IndicatorSpecifications — Installation and Operating Instructions949-0542PHYSICAL DATASelectable Inputs: Thermocouple, RTD, DC Voltage, or DC cur-rent.Input Impedance: Thermocouple = 3 megohms, Current = 10ohms, Voltage = 5000 ohms.RTD Excitation Current: 200µA.Display: …
Love Controls UPT 1000 Measuring Instruments, 4
-
AXIOMET
AX-C705
AX-C7051. Introduction1.1. GeneralizationThe Meter is a battery-operated process maintenace field device which combined both digital-parameter function withprocess-signal source together.Functioning as follows:• Measurement function of TC voltage, DC voltage , Ohms, DC, Continuity and Diode; Data reserving function;� …
AX-C705 Measuring Instruments, 11
-
IME
CE4DMID31
LE11109AA_01_W1MDPMDTempo di media Average time Temps de moyenne MittlerezeitVelocità comunicazione Baud rate Vitesse de communication ÜbertragungsdatengeschwindigkeitIndirizzo comunicazione Communication address Adresse de communication KommunikationsadressePotenza attiva media Active power demand Puissance moyenne …
CE4DMID31 Measuring Instruments, 14
Recommended Documentation: