Fluke 725 руководство по эксплуатации на русском

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

®

725Ex

Multifunction Process Calibrator

Users Manual

January 2005 Rev.1, 8/05

© 2005 Fluke Corporation, All rights reserved. Printed in USA

All product names are trademarks of their respective companies.

loading

Related Manuals for Fluke 725Ex

Summary of Contents for Fluke 725Ex

  • Page 1
    ® 725Ex Multifunction Process Calibrator Users Manual January 2005 Rev.1, 8/05 © 2005 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies.
  • Page 2
    LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to…
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction ……………………1 Contacting Fluke ………………….1 Standard Equipment………………….. 3 Safety Information ………………….3 Ex Hazardous Areas …………………. 3 Faults and Damage ………………..8 Safety Regulations ………………..9 Certification Information………………… 10 Getting Acquainted with the Calibrator …………….10 Input and Output Terminals ………………

  • Page 4
    725Ex Users Manual Current Measurement with Loop Power…………..19 Measuring Electrical Parameters (Lower Display) ………… 21 Measuring Temperature……………….. 22 Using Thermocouples ………………22 Using Resistance-Temperature Detectors (RTDs) ……….25 Measuring Pressure ………………..28 Zeroing with Absolute Pressure Modules…………..29 Using Source Mode ………………….. 31 Sourcing 4 to 20 mA………………..
  • Page 5
    (continued) Contents Service Center Calibration or Repair …………….. 50 Replacement Parts ………………..51 Accessories ……………………53 Specifications ……………………. 55 DC Voltage Measurement ………………55 DC Voltage Source………………..55 Millivolt Measurement and Source* …………….55 DC mA Measurement and Source…………….56 Ohms Measurement ………………..
  • Page 6
    725Ex Users Manual…
  • Page 7: List Of Tables

    Symbols……………………..8 Input/Output Terminals and Connectors…………….11 Key Functions……………………13 Thermocouple Types Accepted………………… 23 RTD Types Accepted ………………….26 mA Step Values……………………40 Approved Batteries …………………… 50 Replacement Parts …………………… 52 Fluke Pressure Module Compatibility ………………53 Pressure Modules ……………………. 54…

  • Page 8
    725Ex Users Manual…
  • Page 9: List Of Figures

    List of Figures Figure Title Page Standard Equipment………………….7 Input/Output Terminals and Connectors…………….10 Keys ……………………….12 Elements of a Typical Display ………………..15 Voltage-to-Voltage Test………………….17 Adjusting the Contrast ………………….18 Measuring Voltage and Current Output …………….. 19 Connections for Supplying Loop Power…………….. 20 Measuring Electrical Parameters……………….

  • Page 10
    725Ex Users Manual Connections for Sourcing Pressure ………………38 Calibrating a Thermocouple Transmitter …………….42 Calibrating a Pressure-to-Current (P/I) Transmitter ………….. 44 Calibrating a Current-to-Pressure (I/P) Transmitter ………….. 46 Calibrating a Chart Recorder ………………..48 Replacing the Batteries ………………….49…
  • Page 11: Introduction

    To order accessories, receive operating assistance, or get electrical and physical parameters. For a summary of the location of the nearest Fluke distributor or Service source and measurement functions, see Table 1. Center, call:…

  • Page 12
    Pt200 Ω (385) Pt500 Ω (385) Pt1000 Ω (385) Pressure Fluke 700PEx series modules ranging from Fluke 700PEx series modules ranging from 10 in. H2O to 3,000 psi 10 in. H2O to 3,000 psi using an external pressure source (hand pump)
  • Page 13: Standard Equipment

    Replacement Parts in Table 9. 500 or CSA C22.1 Section 18. • TL75 test leads (one set) The Model 725Ex Calibrator has been designed for use in • AC72 alligator clips (one set) Ex Hazardous Areas. These are areas where potentially •…

  • Page 14
    30 V and Isc not exceed 100 mA. the 725Ex calibrator Voc or of the barrier Voc. The maximum current will be the sum of the 725Ex calibrator The Model 725Ex calibrator will itself be a source of Isc and the barrier Isc.
  • Page 15
    To avoid electric shock, injury, damage to the Calibrator, or ignition of an explosive atmosphere, follow all equipment safety procedures. • 725E Use the Calibrator only as described in this User Manual and the Fluke x CCD (Concept Control Drawing) or the protection provided by the Calibrator may be impaired. •…
  • Page 16
    725Ex Users Manual • When servicing the Calibrator, use only specified replacement parts. Do not open the Calibrator case. Opening the case invalidates the Calibrator’s Ex Approval. • Do not allow water inside the case. • Before each use, verify the Calibrator’s operation by measuring a known voltage.
  • Page 17
    Ex Hazardous Areas AC72 Alligator Clips TL75 Test Lead Set Stackable Test Leads aly01f.eps Figure 1. Standard Equipment…
  • Page 18: Faults And Damage

    725Ex Users Manual Table 2. Symbols Symbol Meaning DC — Direct Current Power ON/OFF Earth ground Conforms to ATEX requirements. Battery Risk of Danger. Important information. Refer to manual. Double insulated Conforms to relevant Canadian and US Standards. Conforms to relevant European Union directives.

  • Page 19: Safety Regulations

    When the calibrator is used in an Ex hazardous area, unless the area is known to be safe, do not connect to any circuits that exceed the entity parameters defined on Fluke 725Ex CCD Control Drawing. See “Ex Hazardous Areas”.

  • Page 20: Certification Information

    725Ex Users Manual Certification Information ( II 1 G EEx ia IIB 171 C 0344 KEMA 04ATEX1303X Class I Div. 1 Groups B,C, and D MULTIFUNCTION 725Ex PROCESS CALIBRATOR Class I Zone 0 Aex/Ex ia IIB 171 °C LR110460 2004.1573226 Ta = -10 C……

  • Page 21
    Getting Acquainted with the Calibrator Table 3. Input/Output Terminals and Connectors Name Description Pressure module Connects the Calibrator to a pressure module. connector B, C MEASURE V, mA Input terminals for measuring voltage, current, and supplying loop power. terminals TC input/output Terminal for measuring or simulating thermocouples.
  • Page 22: Keys

    II 2 G EEx ia IIC T4 ZELM 02 ATEX 0120 REFER TO CONTROL DRAWING: ANSCHLUSSSCHEMA BEACHTEN: SUIVRE LE SCHÉMA DE CONNEXION: FLUKE 725Ex CCD 30V MAX ALL TERMINALS 30V MAX ALL TERMINALS SOURCE / MEASURE SOURCE / MEASURE MEASURE…

  • Page 23
    Getting Acquainted with the Calibrator Table 4. Key Functions Description Turns the power on and off. Selects voltage, mA, or loop power measurement functions in the upper display. Clears the switch test. See “Switch Test”. Selects the pressure measurement function in the upper display. Repeated pushes cycle through the different pressure units.
  • Page 24
    725Ex Users Manual Table 4. Key Functions (cont.) Description Disables Shut Down Mode Enables Shut Down Mode XW Y Increases or decreases the source level. Cycles through the 2-, 3-, and 4-wire selections. Moves through the memory locations of calibrator setups.
  • Page 25: Display

    Getting Acquainted with the Calibrator Display Figure 4 shows the elements of the display. Memory Locations Loop for Calibrator Setups Annunciator Low Battery Symbol Units Display Mode Indicator Auto Ramp sh07f.eps Figure 4. Elements of a Typical Display…

  • Page 26: Getting Started

    725Ex Users Manual Press to select dc voltage sourcing. Getting Started Press to select a digit to change. Press This section introduces some basic operations of the to select 1 V for the output value. Press and hold Calibrator. to enter 1 V as the 0 % value.

  • Page 27
    Shut Down Mode aly39f.eps Figure 5. Voltage-to-Voltage Test…
  • Page 28: Contrast Adjustment

    725Ex Users Manual Contrast Adjustment To adjust the contrast, proceed as follows: Press C and O until Contst Adjust is displayed as shown in Figure 6. MULTIFUNCTION 725Ex PROCESS CALIBRATOR Press and hold Xto darken contrast. Press and hold W to lighten contrast.

  • Page 29: Using Measure Mode

    Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, MULTIFUNCTION 725Ex PROCESS CALIBRATOR or to measure the output of a pressure instrument, use the upper display and proceed as follows: Press to select volts or current.

  • Page 30
    725Ex Users Manual MULTIFUNCTION 725Ex PROCESS CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ 2004.1573266 LR110460 TEST DC PWR Zone 0 AEx ia IIB 171 C – + – I.S. Class I Div 1, Groups B,C,D STORE…
  • Page 31: Measuring Electrical Parameters (Lower Display)

    Using Measure Mode Measuring Electrical Parameters (Lower Display) To measure the electrical parameters using the lower display, proceed as follows: Connect the Calibrator as shown in Figure 9. MULTIFUNCTION 725Ex PROCESS CALIBRATOR If necessary, press for MEASURE mode (lower display). Press for dc voltage or current, or frequency or resistance.

  • Page 32: Measuring Temperature

    725Ex Users Manual WCaution Measuring Temperature One thermocouple pin is wider than the other. To Using Thermocouples avoid possible damage to Calibrator or to The Calibrator supports twelve standard thermocouples, equipment under test do not try to force a including types E, N, J, K, T, B, R, S, L, XK, BP, and U.

  • Page 33
    Using Measure Mode Table 5. Thermocouple Types Accepted Positive Lead (H) Type Positive Lead Color Negative Lead Specified Range Material Material (°C) ANSI* IEC** Chromel Purple Violet Constantan -200 to 950 Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300 Iron White Black Constantan -200 to 1200…
  • Page 34
    725Ex Users Manual 30V 24mA MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + Process Temperature LOOP mA — Warning 30 V maximum to TC Miniplug aly12f.eps Figure 10. Measuring Temperature with a Thermocouple…
  • Page 35: Using Resistance-Temperature Detectors (Rtds)

    Using Measure Mode Using Resistance-Temperature Detectors (RTDs) To measure temperature using an RTD input, proceed as follows: The Calibrator accepts RTD types shown in Table 6. RTDs If necessary, press for MEASURE mode. are characterized by their resistance at 0 °C (32 °F), which is called the “ice point”…

  • Page 36
    725Ex Users Manual Table 6. RTD Types Accepted α RTD Type Ice Point (R Material Range (°C) 100 Ω 0.003926 Ω/°C Pt100 (3926) Platinum -200 to 630 100 Ω 0.00385 Ω/°C Pt100 (385) Platinum -200 to 800 120 Ω 0.00672 Ω/°C…
  • Page 37
    Using Measure Mode 30V 24mA MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP Resistance to mA — be measured 30V 24mA MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + Resistance to LOOP mA — be measured 30V 24mA MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA +…
  • Page 38: Measuring Pressure

    Users Manual Measuring Pressure Many ranges and types of pressure modules are available Gage Differential from Fluke. See “Accessories” near the back of this High manual. Before using a pressure module, read its PRESSURE MODULE PRESSURE MODULE instruction sheet. The modules vary in use, media, and…

  • Page 39: Zeroing With Absolute Pressure Modules

    Using Measure Mode • To avoid damaging the pressure module from Zeroing with Absolute Pressure Modules corrosion, use it only with specified materials. To zero, adjust the Calibrator to read a known pressure. Refer to the printing on the pressure module This can be barometric pressure, if it is accurately known.

  • Page 40
    725Ex Users Manual Isolation Valve Gage Module 725Ex MULTIFUNCTION PROCESS CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ Differential 2004.1573266 LR110460 Zone 0 AEx ia IIB 171 C I.S. Class I Div 1, Groups B,C,D STORE 100%…
  • Page 41: Using Source Mode

    Using Source Mode Enter the desired current by pressing Using Source Mode keys. In SOURCE mode, the Calibrator generates calibrated Sourcing Other Electrical Parameters signals for testing and calibrating process instruments; supplies voltages, currents, frequencies, and resistances; Volts, ohms, and frequency are also sourced and shown simulates the electrical output of RTD and thermocouple in the lower display.

  • Page 42
    RECALL Loop Black Power Supply Readout or Controller – Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly17f.eps Figure 14. Connections for Simulating a 4 to 20- mA Transmitter in a non-Ex hazardous Area…
  • Page 43
    30V 24mA MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA — Device Under Test – Black Common Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly16f.eps Figure 15. Electrical Sourcing Connections…
  • Page 44: Simulating Thermocouples

    725Ex Users Manual Simulating Thermocouples Simulating RTDs Connect the Calibrator TC input/output to the instrument Connect the Calibrator to the instrument under test as under test with thermocouple wire and the appropriate shown in Figure 17. Proceed as follows to simulate an…

  • Page 45
    Zone 0 AEx ia IIB 171 C I.S. Class I Div 1, Groups B,C,D STORE 100% SETUP RECALL TC Miniplug Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly20f.eps Figure 16. Connections for Simulating a Thermocouple…
  • Page 46
    725Ex Users Manual Sensor Terminals Black Black Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly40f.eps Figure 17. Connections for Simulating 3-Wire RTD…
  • Page 47: Source Pressure Mode

    Nm) of torque between the pressure module pressure in the SOURCE field. Figure 18 shows how to fittings, or between the fittings and the body connect a pump to a Fluke pressure module which makes of the module. Always apply appropriate it a calibrated source.

  • Page 48
    725Ex Users Manual Connect a pressure module to the Calibrator as shown in Figure 18. The threads on the pressure modules accept standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary. Press (lower display). The Calibrator…
  • Page 49: Setting 0 % And 100 % Output Parameters

    Setting 0 % and 100 % Output Parameters This setting may now be used for the following: Setting 0 % and 100 % Output • Parameters Manually stepping an output with 25 % increments. • Jump between the 0 and 100 % span points by For current output, the Calibrator assumes that 0 % momentarily pushing corresponds to 4 mA and 100 % corresponds to 20 mA.

  • Page 50: Auto Ramping The Output

    725Ex Users Manual Auto Ramping the Output Storing and Recalling Setups Auto ramping gives the ability to continuously apply a Store up to eight settings in a nonvolatile memory and varying stimulus from the Calibrator to a transmitter, while recall the settings for later use. A low battery condition or a your hands remain free to test the response of the battery change does not jeopardize the stored settings.

  • Page 51: Calibrating A Transmitter

    Calibrating a Transmitter Set the zero and span parameters by pressing Calibrating a Transmitter keys. Enter these parameters by pressing and Use the measurement (upper display) and source (lower holding . For more information on display) modes to calibrate a transmitter. This section setting parameters, see “Setting 0 % and 100 %”…

  • Page 52
    Zone 0 AEx ia IIB 171 C – + – STORE I.S. Class I Div 1, Groups B,C,D 100% SETUP RECALL – Black Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly44f.eps Figure 19. Calibrating a Thermocouple Transmitter…
  • Page 53: Calibrating A Pressure Transmitter

    Calibrating a Pressure Transmitter Calibrating a Pressure Transmitter The following example shows how to calibrate a pressure transmitter. Connect the Calibrator to the instrument under test as shown in Figure 20. Proceed as follows: Press for current (upper display). If required, press again to activate loop power.

  • Page 54
    Zone 0 AEx ia IIB 171 C STORE I.S. Class I Div 1, Groups B,C,D 100% SETUP RECALL Black Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly34f.eps Figure 20. Calibrating a Pressure-to-Current (P/I) Transmitter…
  • Page 55: Calibrating An I/P Device

    Calibrating an I/P Device Calibrating an I/P Device The following test allows calibration of a device that controls pressure. Proceed as follows: Connect the test leads to the instrument under test as shown in Figure 21. The connections simulate a current-to-pressure transmitter and measures the corresponding output pressure.

  • Page 56
    Zone 0 AEx ia IIB 171 C I.S. Class I Div 1, Groups B,C,D STORE 100% SETUP RECALL Black Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly28f.eps Figure 21. Calibrating a Current-to-Pressure (I/P) Transmitter…
  • Page 57: Switch Test

    Switch Test Note Switch Test Pressure the device slowly to ensure accurate To perform a switch test, follow these steps: readings. Run the test several times to confirm Note repeatability. This example used a normally closed switch. OPEN displays once the switch is open. Bleed the The procedure is the same for an open switch pump slowly until the pressure switch closes.

  • Page 58: Testing An Output Device

    Zone 0 AEx ia IIB 171 C I.S. Class I Div 1, Groups B,C,D STORE 100% SETUP RECALL 0 to 1 V dc Input Black Refer to Fluke 725Ex CCD Control Drawing when using in an Ex hazardous area. aly25f.eps Figure 22. Calibrating a Chart Recorder…

  • Page 59: Replacing The Batteries

    Replacing the Batteries Replacing the Batteries WWarning Hex Key • To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator (M) appears. • Remove the Calibrator from the Ex- hazardous area before opening the battery door.

  • Page 60: Approved Batteries

    725Ex Users Manual Approved Batteries Table 8. Approved Batteries Battery Manufacturer Type (All Batteries Alkaline- AA 1.5 V) Duracell MN1500 Eveready (Energizer) Panasonic Powerline LR6A Rayovac Varta 4906 Ucar Gold Clean the Calibrator and pressure modules with a soft Maintenance cloth dampened with water or water and mild soap.

  • Page 61: Replacement Parts

    “Contacting Fluke” at the beginning of the manual. The Fluke 725Ex Calibrator covered by the warranty will be promptly repaired or replaced (at Fluke’s option) and Replacement Parts returned to you at no charge. See the back of the title Table 9 lists the part number of each replaceable part.

  • Page 62
    Test lead, black 688066 AC72 alligator clip, red 1670641 AC72 alligator clip, black 1670652 Input Decal 690948 Fluke 725Ex CD ROM, contains Fluke 725Ex 2406548 User Manual Fluke 725Ex Control Drawing 6800032 Fluke 725Ex Safety Information 2151996 Fluke 725Ex Calibration Manual…
  • Page 63: Accessories

    The output of Fluke 700PEx pressure modules can cause prices, contact a Fluke representative. Pressure Modules the 725Ex 5 digit display to overflow, or else produce and Fluke model numbers (see Table 10) are listed values that are too low to be read if inappropriate units below.

  • Page 64
    725Ex Users Manual Table 11. Pressure Modules Fluke Model Number Range Type and Media Fluke-700P01Ex 0 to 10” H differential, Low: dry High: dry Fluke-700P24Ex 0 to 15 psi differential, Low: dry High: wet Fluke-700P05Ex 0 to 30 psi gage, wet…
  • Page 65: Specifications

    Multifunction Process Calibrator Specifications Specifications DC Voltage Source All specifications apply from +18 °C to +28 °C unless Accuracy, Range Resolution stated otherwise. All specifications assume a 5 minute (% of Reading + Counts) warmup period. 100 mV 0.01 mV 0.02 % + 2 DC Voltage Measurement 10 V…

  • Page 66: Dc Ma Measurement And Source

    725Ex Users Manual DC mA Measurement and Source Ohms Source Accuracy, Excitation Current from Accuracy Range Resolution Ohms Range ± Ω (% of Reading + Counts) Measurement Device 15 to 400 Ω 24 mA 0.001 mA 0.15 to 0.5 mA 0.15…

  • Page 67: Frequency Source

    Multifunction Process Calibrator Specifications Frequency Source Measure and Source Type Range Accuracies Accuracy Range Resolution (% of output 600 to 800 °C 2.2 °C 800 to 1000 °C 1.8 °C frequency) 1000 to 1800 °C 1.4 °C ± 0.05 % 2.0 to 1000.0 CPM 0.1 CPM -200 to 0 °C…

  • Page 68: Rtd Excitation (Simulation)

    725Ex Users Manual RTD Excitation (simulation) Allowable Excitation by RTD type Ni 120 0.15 to 3.0 mA Pt 100-385 0.15 to 3.0 mA Pt 100-3926 0.15 to 3.0 mA Pt 100-3916 0.15 to 3.0 mA Pt 200-385 0.05 to 0.80 mA Pt 500-385 0.05 to 0.80 mA…

  • Page 69: Pressure Measurement

    Multifunction Process Calibrator Specifications Pressure Measurement Range Resolution Accuracy Units Determined by 5 digits Determined by psi, inH2O@4 C, inH2O@20 C, inH2O@60 F, kPa, pressure module pressure module cmH2O@4 C, cmH2O@20 C, bar, mbar, kg/cm , mmHg, inHg General Specifications Operating temperature -10 C to 55 C Storage temperature…

  • Page 70
    725Ex Users Manual Entity Parameters For Entity Parameters, Refer to Fluke 725Ex CCD, Control Drawing for use in Ex hazardous areas.

background image

®

725Ex

Multifunction Process Calibrator

Руководство

пользователя

January 2005 Rev. 2, 5/09 (Russian)

© 2005, 2009 Fluke Corporation, All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

  • Page 1
    ® Multifunction Process Calibrator Users Manual October, 1998 Rev.3, 5/04 © 1998-2004 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies.
  • Page 2
    LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke’s opinion, has been misused, altered,…
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction……………………1 Contacting Fluke………………….1 Standard Equipment …………………. 3 Safety Information………………….3 Getting Acquainted with the Calibrator …………….8 Input and Output Terminals ………………8 Keys …………………….. 10 Display ……………………13 Getting Started ………………….14 Shut Down Mode ………………….

  • Page 4
    Users Manual Using Thermocouples………………. 20 Using Resistance-Temperature Detectors (RTDs) ……….23 Measuring Pressure ………………..26 Zeroing with Absolute Pressure Modules …………..27 Using Source Mode………………….29 Sourcing 4 to 20 mA ………………..29 Simulating a 4- to 20-mA Transmitter……………. 29 Sourcing Other Electrical Parameters ……………
  • Page 5
    Contents (continued) External Fluke Pressure Module Compatibility …………53 Specifications……………………. 56 DC Voltage Measurement………………56 DC Voltage Source ………………..56 Millivolt Measurement and Source*…………….56 DC mA Measurement and Source …………….57 Ohms Measurement………………..57 Ohms Source ………………….57 Frequency Measurement ………………57 Frequency Source…………………
  • Page 6
    Users Manual…
  • Page 7
    RTD Types Accepted………………….24 mA Step Values ……………………38 Remote Control Upper Display ………………… 46 Remote Control Lower Display ………………… 46 «S» Commands Select Sensor Type ………………48 Replacement Parts ………………….. 51 Fluke Pressure Module Compatibility ………………. 53 Pressure Modules ……………………. 54…
  • Page 8
    Users Manual…
  • Page 9
    List of Figures Figure Title Page Standard Equipment ………………….6 Input/Output Terminals and Connectors …………….8 Keys……………………….10 Elements of a Typical Display………………..13 Voltage-to-Voltage Test ………………….15 Adjusting the Contrast ………………….16 Measuring Voltage and Current Output …………….17 Connections for Supplying Loop Power…………….
  • Page 10
    Users Manual Connections for Sourcing Pressure………………36 Calibrating a Thermocouple Transmitter …………….40 Calibrating a Pressure-to-Current (P/I) Transmitter…………… 42 Calibrating a Current-to-Pressure (I/P) Transmitter…………… 44 Calibrating a Chart Recorder………………..45 Replacing the Battteries ………………….50 Replacement Parts…………………… 52 viii…
  • Page 11: Introduction

    In addition to the functions in Table 1, the calibrator has To order accessories, receive operating assistance, or the following features and functions: get the location of the nearest Fluke distributor or Service A split-screen display. The upper display allows you Center, call: to measure volts, current, and pressure only.

  • Page 12
    Users Manual Table 1. Summary of Source and Measure Functions Function Measure Source dc V 0 V to 30 V 0 V to 10 V dc mA 0 to 24 mA 0 to 24 mA Frequency 1 CPM to 10 kHz 1 CPM to 10 kHz Resistance to 3200…
  • Page 13: Standard Equipment

    Standard Equipment Standard Equipment Safety Information The items listed below and shown in Figure 1 are included The calibrator is designed in accordance with IEC1010-1, with your calibrator. If the calibrator is damaged or ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1- something is missing, contact the place of purchase 92.

  • Page 14
    Users Manual W Warning To avoid possible electric shock or personal injury: Do not apply more than the rated voltage, as marked on the calibrator, between the terminals, or between any terminal and earth ground (30 V 24 mA max all terminals). Before each use, verify the calibrator’s operation by measuring a known voltage.
  • Page 15
    Safety Information W Warning When using a pressure module, make sure the process pressure line is shut off and depressurized before you connect it or disconnect it from the pressure module. Use only 4 AA batteries, properly installed in the calibrator case, to power the calibrator. Disconnect test leads before changing to another measure or source function.
  • Page 16
    Users Manual AC72 Alligator Clips TL75 Test Lead Set Stackable Test Leads sh01f.eps Figure 1. Standard Equipment…
  • Page 17
    Safety Information Table 2. International Symbols AC — Alternating current Double insulated DC — Direct current Battery Earth ground Refer to the manual for information about this feature. Pressure ON/OFF Conforms to Canadian Standards Conforms to European Union directives Association directives…
  • Page 18: Getting Acquainted With The Calibrator

    Users Manual Getting Acquainted with the Calibrator Input and Output Terminals Figure 2 shows the calibrator input and output terminals. MULTIFUNCTION CALIBRATOR Table 3 explains their use. V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ STORE 100% SETUP RECALL sh05f.eps Figure 2.

  • Page 19
    Getting Acquainted with the Calibrator Table 3. Input/Output Terminals and Connectors Name Description Pressure module Connects the calibrator to a pressure module or the calibrator to a PC for a remote connector control connection. MEASURE V, mA Input terminals for measuring voltage, current, and supplying loop power. terminals TC input/output Terminal for measuring or simulating thermocouples.
  • Page 20: Keys

    Users Manual Keys Figure 3 shows the calibrator keys and Table 4 explains their use. MULTIFUNCTION CALIBRATOR MULTIFUNCTION CALIBRATOR V mA ZERO V mA ZERO LOOP LOOP V mA V mA MEAS MEAS SOURCE SOURCE ˚ ˚ ˚ ˚ STORE STORE 100% 100%…

  • Page 21
    Getting Acquainted with the Calibrator Table 4. Key Functions Name Description Turns the power on or off. Selects voltage, mA or Loop Power measurement function in the upper display. Selects the pressure measurement function in the upper display. Repeated pushes cycle through the different pressure units.
  • Page 22
    Users Manual Table 4. Key Functions (cont.) Name Description Cycles through : Slow repeating 0 % — 100 % — 0 % ramp Fast repeating 0 % — 100 % — 0 % ramp Repeating 0 % — 100 % — 0 % ramp in 25 % steps Disables Shut Down Mode Enables Shut Down Mode Increases or decreases the source level.
  • Page 23: Display

    Getting Acquainted with the Calibrator Display Figure 4 shows the elements of a typical display. Memory Locations Loop for Calibrator Setups Annunciator Low Battery Symbol Units Display Mode Indicator Auto Ramp sh07f.eps Figure 4. Elements of a Typical Display…

  • Page 24: Getting Started

    Users Manual Press to increase the output to 5 V. Press and Getting Started hold to enter 5 V as the 100 % value. This section acquaints you with some basic operations of Press to step between 0 and 100 % the calibrator.

  • Page 25
    Shut Down Mode sh39f.eps Figure 5. Voltage-to-Voltage Test…
  • Page 26: Contrast Adjustment

    Users Manual Contrast Adjustment Note Available with V2.1 Firmware or greater. To MULTIFUNCTION CALIBRATOR identify firmware version, press and hold J when powering up. The firmware version will be shown in the upper units display for about 1 second after initialization. To adjust the contrast, proceed as follows: V mA ZERO…

  • Page 27: Using Measure Mode

    Using Measure Mode Using Measure Mode Measuring Electrical Parameters (Upper Display) MULTIFUNCTION CALIBRATOR To measure the current or voltage output of a transmitter, or to measure the output of a pressure instrument, use the upper display and proceed as follows: Press to select volts or current.

  • Page 28
    Users Manual MULTIFUNCTION CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ TEST DC PWR – + – STORE 100% SETUP RECALL – Black sh18f.eps Figure 8. Connections for Supplying Loop Power…
  • Page 29: Measuring Electrical Parameters (Lower Display)

    Using Measure Mode Measuring Electrical Parameters (Lower Display) To measure the electrical parameters using the lower display, proceed as follows: Connect the calibrator as shown in Figure 9. MULTIFUNCTION CALIBRATOR If necessary, press for MEASURE mode (lower display) Press for dc voltage or current, or frequency or resistance.

  • Page 30: Measuring Temperature

    Users Manual Measuring Temperature Note If the calibrator and the thermocouple plug are at Using Thermocouples different temperatures, wait one minute or more The calibrator supports ten standard thermocouples, for the connector temperature to stabilize after including type E, N, J, K, T, B, R, S, L, or U. Table 5 you plug the miniplug into the TC input/output.

  • Page 31
    Using Measure Mode Table 5. Thermocouple Types Accepted Positive Lead (H) Type Positive Lead Color Negative Lead Specified Range Material Material ( C) ANSI* IEC** Chromel Purple Violet Constantan -200 to 950 Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300 Iron White Black Constantan…
  • Page 32
    Users Manual 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + Process Temperature LOOP mA — Warning 30 V maximum to TC Miniplug sh12f.eps Figure 10. Measuring Temperature with a Thermocouple…
  • Page 33: Using Resistance-Temperature Detectors (Rtds)

    Using Measure Mode Using Resistance-Temperature Detectors (RTDs) To measure temperature using an RTD input, proceed as follows: The calibrator accepts RTD types shown in Table 6. RTDs If necessary, press for MEASURE mode. are characterized by their resistance at 0 °C (32 °F), which is called the “ice point”…

  • Page 34
    Users Manual Table 6. RTD Types Accepted RTD Type Ice Point (R Material Range ( C) Pt100 (3926) Platinum 0.003926 -200 to 630 Pt100 (385) Platinum 0.00385 -200 to 800 Ni120 (672) Nickel 0.00672 -80 to 260 Pt200 (385) Platinum 0.00385 -200 to 630 Pt500 (385)
  • Page 35: Measuring Temperature With Rtd/Measuring 2-, 3-/4-Wire

    Using Measure Mode 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + Resistance LOOP to be mA — measured 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA — 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP…

  • Page 36: Measuring Pressure

    Users Manual Measuring Pressure Gage Many ranges and types of pressure modules are available Differential from Fluke. See “Accessories” near the back of this High manual. Before you use a pressure module, read its 700 P 06 700 P 04…

  • Page 37: Zeroing With Absolute Pressure Modules

    Using Measure Mode To avoid damaging the pressure module from O@20 C, inH O@4 C, inH O@20 C, mbar, overpressure, never apply pressure above the bar, kg/cm , or kPa. rated maximum printed on the pressure Zeroing with Absolute Pressure Modules module.

  • Page 38
    Users Manual Isolation Valve Gage Module MULTIFUNCTION CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ Differential STORE 100% Module SETUP RECALL Tank sh37f.eps Figure 13. Connections for Measuring Pressure…
  • Page 39: Using Source Mode

    Using Source Mode If necessary, press for SOURCE mode. Using Source Mode Press until both mA and SIM display. In SOURCE mode, the calibrator generates calibrated signals for testing and calibrating process instruments; Enter the desired current by pressing supplies voltages, currents, frequencies, and resistances; keys.

  • Page 40: Simulating 4- To 20-Ma Transmitter Connections

    Users Manual MULTIFUNCTION CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ +24 V STORE 100% SETUP RECALL Loop Black Power Supply Readout or Controller – sh17f.eps Figure 14. Connections for Simulating a 4- to 20-mA Transmitter…

  • Page 41
    Using Source Mode 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE + mA mA + LOOP mA — – Black Common 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA — – Black Common sh16f.eps Figure 15. Electrical Sourcing Connections…
  • Page 42: Simulating Thermocouples

    Users Manual Simulating Thermocouples Simulating RTDs Connect the calibrator TC input/output to the instrument Connect the calibrator to the instrument under test as under test with thermocouple wire and the appropriate shown in Figure 17. Proceed as follows to simulate an thermocouple mini-connector (polarized thermocouple RTD: plug with flat, in-line blades spaced 7.9 mm [0.312 in]…

  • Page 43
    Using Source Mode MULTIFUNCTION CALIBRATOR TEST DC PWR Color depends – on type of TC V mA ZERO LOOP – V mA MEAS SOURCE ˚ ˚ STORE 100% SETUP RECALL TC Miniplug sh20f.eps Figure 16. Connections for Simulating a Thermocouple…
  • Page 44: Connections For Simulating 3-Wire Rtd

    Users Manual SENSOR TERMINALS BLACK BLACK sh40f.eps Figure 17. Connections for Simulating 3-Wire RTD…

  • Page 45: Sourcing Pressure

    10 ft.-lb. (13.5 Nm) of torque between the pressure in the SOURCE field. Figure 20 shows how to pressure module fittings, or between the connect a pump to a Fluke pressure module which makes fittings and the body of the module. Always it a calibrated source.

  • Page 46
    Users Manual Connect a pressure module to the calibrator as shown in Figure 18. The threads on the pressure modules accept standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary. Press (lower display). The calibrator automatically senses which pressure module is attached and sets its range accordingly.
  • Page 47: Setting 0 % And 100 % Output Parameters

    Setting 0 % and 100 % Output Parameters You can now use this setting for the following: Setting 0 % and 100 % Output Parameters Manually stepping an output with 25 % increments. For current output, the calibrator assumes that 0 % Jump between the 0 and 100 % span points by corresponds to 4 mA and 100 % corresponds to 20 mA.

  • Page 48: Auto Ramping The Output

    Users Manual Auto Ramping the Output Storing and Recalling Setups Auto ramping gives you the ability to continuously apply a You can store up to eight of your settings in a nonvolatile varying stimulus from the calibrator to a transmitter, while memory and recall the settings for later use.

  • Page 49: Calibrating A Transmitter

    Calibrating a Transmitter Set your zero and span parameters by pressing Calibrating a Transmitter keys. Enter these parameters by pressing and Use the measurement (upper display) and source (lower holding . For more information on display) modes to calibrate a transmitter. This section setting parameters, see “Setting 0 % and 100 %”…

  • Page 50
    Users Manual MULTIFUNCTION CALIBRATOR V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ TEST DC PWR – + – STORE 100% SETUP RECALL – Black sh44f.eps Figure 19. Calibrating a Thermocouple Transmitter…
  • Page 51: Calibrating A Pressure Transmitter

    Calibrating a Pressure Transmitter Calibrating a Pressure Transmitter The following example shows how to calibrate a pressure transmitter. Connect the calibrator to the instrument under test as shown in Figure 20. Proceed as follows: Press for current (upper display). If required, press again to activate loop power.

  • Page 52
    Users Manual Hand Pump S I G N A L Measure mA – Source Pressure 24 V Loop Power Enabled T E S T MULTIFUNCTION CALIBRATOR Pressure Module V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ STORE 100% SETUP RECALL Black…
  • Page 53: Calibrating An I/P Device

    Calibrating an I/P Device Calibrating an I/P Device The following test allows you to calibrate a device that controls pressure. Proceed as follows: Connect the test leads to the instrument under test as shown in Figure 21. The connections simulate a current-to-pressure transmitter and measures the corresponding output pressure.

  • Page 54
    Users Manual S I G N A L Measure – Pressure Source mA T E S T MULTIFUNCTION CALIBRATOR Pressure Module V mA ZERO LOOP V mA MEAS SOURCE ˚ ˚ STORE 100% SETUP RECALL Black sh28f.eps Figure 21. Calibrating a Current-to-Pressure (I/P) Transmitter…
  • Page 55: Testing An Output Device

    Testing an Output Device Testing an Output Device Use the source functions to test and calibrate actuators, recording, and indicating devices. Proceed as follows: MULTIFUNCTION CALIBRATOR Connect the test leads to the instrument under test as shown in Figure 22. Press for current or dc voltage, or frequency or resistance (lower display).

  • Page 56: Remote Control Commands

    The calibrator will initialize with its remote port enabled. The terminal emulator connected to The Fluke 700SC Serial Interface Cable Assembly (PN the calibrator should be set up for: 9600 baud, no parity, 667425) plugs into the pressure module connector and 8 data bits, and 1 stop bit.

  • Page 57
    Remote Control Commands Table 8B. Remote Control Lower Display (cont) Serial Input Description KHz source Hz measurement Hz source CPM measurement CPM source Ohms measurement (default 2W) Select Ohms source 2-wire measurement (Ohms and RTDs) 3-wire measurement (Ohms and RTDs) 4-wire measurement (Ohms and RTDs) Thermocouple measurement (default Type J) use “S”…
  • Page 58
    Users Manual Table 8B. Remote Control Lower Display (cont) Serial Input Description Enter a source value using ascii characters 0,1,2,…9,-,.terminated by <CR> (carriage return) <CR> Single Broadcast of most recent lower display value and units Table 8C. «S» Commands Select Sensor Type Selection Entry Serial Thermocouple Type…
  • Page 59: Replacing The Batteries

    Replacing the Batteries Replacing the Batteries Turn the calibrator off, remove the test leads from the terminals, and hold the calibrator face down. Warning Using a flat-blade screwdriver, turn the battery door To avoid false readings, which could lead to screws 1/4-turn counterclockwise and remove the possible electric shock or personal injury, battery door.

  • Page 60: Maintenance

    Users Manual Maintenance Cleaning the Calibrator Warning To avoid personal injury or damage to the calibrator, use only the specified replacement parts and do not allow water into the case. Caution To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers. Battery and Clean the calibrator and pressure modules with a soft Compartment…

  • Page 61: Replacement Parts

    Service Center. Fluke 690963 assumes no responsibility for damage in transit. Keypad 690955 The Fluke 725 calibrator covered by the warranty will be Case bottom 664235 promptly repaired or replaced (at Fluke’s option) and AA alkaline batteries 376756 returned to you at no charge.

  • Page 62
    Users Manual zi45f.eps Figure 24. Replacement Parts…
  • Page 63: Accessories

    External Fluke Pressure Module Compatibility Accessories For more information about these accessories and their The output of Fluke 700P pressure modules can cause prices, contact your Fluke representative. Pressure the 725’s 5 digit display to overflow, or else produce Modules and Fluke model numbers (see Table 10) are values that are too low to be read if inappropriate units listed below.

  • Page 64
    Users Manual Table 11. Pressure Modules Fluke Model Number Range Type and Media Fluke-700P00 0 to 1” H differential, dry Fluke-700P01 0 to 10” H differential, dry Fluke-700P02 0 to 1 psi differential, dry Fluke-700P22 0 to 1 psi differential, wet…
  • Page 65
    Accessories Table 11. Pressure Modules (cont.) Fluke Model Number Range Type and Media Fluke-700P29 0 to 3,000 psi gage, wet Fluke-700P30 0 to 5,000 psi gage, wet Fluke-700P31 0 to 10,000 psi gage, wet Fluke-700PA3 0 to 5 psi absolute, wet…
  • Page 66: Specifications

    Users Manual DC Voltage Source Specifications Accuracy, Specifications are based on a one year calibration cycle Range Resolution (% of Reading + Counts) and apply from +18 C to +28 C unless stated otherwise. All specifications assume a 5 minute warmup period. 100 mV 0.01 mV 0.02 % + 2…

  • Page 67: Dc Ma Measurement And Source

    Specifications DC mA Measurement and Source Ohms Source Accuracy, Excitation Current from Accuracy Range Resolution (% of Reading + Counts) Ohms Range Measurement Device 24 mA 0.001 mA 15 to 400 0.15 to 0.5 mA 0.15 0.02 % + 2 15 to 400 0.5 to 2 mA Temperature coefficient -10 C to 18 C, +28 C to…

  • Page 68: Frequency Source

    Users Manual Frequency Source 600 to 800 C 2.2 C 800 to 1000 C 1.8 C Accuracy 1000 to 1800 C 1.4 C (% of output -200 to 0 C 0.85 C 0 to 900 C 0.7 C Range Resolution frequency) -200 to 0 C 1.1 C…

  • Page 69: Loop Power Supply

    Specifications Loop Power Supply RTD Excitation (simulation) Allowable Excitation by RTD type Voltage: 24 V Ni 120 0.15 to 3.0 mA Pt 100-385 0.15 to 3.0 mA Maximum current: 22 mA Pt 100-392 0.15 to 3.0 mA Short circuit protected Pt 100-JIS 0.15 to 3.0 mA Pt 200-385…

  • Page 70: Pressure Measurement

    Users Manual Pressure Measurement Range Resolution Accuracy Units Determined by 5 digits Determined by psi, inH O@4 C, inH O@20 C, kPa, cmH O@4 C, pressure module pressure module O@20 C, bar, mbar, kg/cm , mmHg, inHg General Specifications Operating temperature -10 C to 55 C Storage temperature — 20 C to 71 C…

  • Page 71
    Index —0— —B— —E— 0% output parameter, setting, 36 Battery, replacing, 48 Electrical parameters measurement, 18 sourcing, 28 —1— —C— 100% output parameter, setting, 36 Calibration, 49 —G— Cleaning calibrator, 49 Connections Getting started, 14 —4— for pressure measurement, 25 for pressure sourcing, 35 4 to 20 mA transmitter —I—…
  • Page 72
    Users Manual —K— Pressure modules available, 52 Source mode, 28 Sourcing Pressure modules, zeroing, 26 4 to 20 mA, 28 Key functions (table), 11 Pressure transmitter, calibrating, 40 electrical parameters, 28 Keys, 10 pressure, 34 —R— thermocouples, 31 —L— Specifications, 55 Recalling setups, 37 Standard equipment, 3 Loop power…
Точность измерений
Напряжение постоянного тока
30,000 В  0,02%+ 2 единицы (верхняя часть дисплея)
30,000 В  0,02%+ 2 единицы (нижняя часть дисплея)
100,00 мВ  0,02% + 2 единицы
-10,00 мВ — 75,00 мВ  0,025 % + 1 единица (через TC разъем)
Сила постоянного тока
24,000 мА  0,02% + 2 единицы
Сопротивление
0,0 — 400,0 Ω  0,1 Ω (4-хжильный), 0.15 Ω (2-х- и 3-хжильный)
401 — 1500 Ω  0,5 Ω (4-хжильный), 1 Ω (2-х- и 3-хжильный)
1500 — 3200 Ω  1 Ω (4-хжильный), 1,5 Ω (2-х- и 3-хжильный)
Частота
2,0 — 1000,0 CPM  0,05 % + 1 единица
1,0 — 1100,0 Гц  0,05 % + 1 единица
1,00 кГц — 10,00 кГц  0,05 % + 1 единица
Чувствительность  Минимум двойной амплитуды 1 В
Давление
Точность  От 0,025% диапазона при использовании любого из 29 безопасных модулей давления.
   (для получения подробных спецификаций см. модули давления в дополнительном оборудовании и аксессуарах)
   Имеются модули для дифференциального, манометрического, абсолютного, двойного, высокого давления и для вакуума.
Точность возбуждения
Напряжение постоянного тока
100,00 мВ  0,02%+2 единицы
10,000 В  0,02%+2 единицы
-10,00 мВ — 75,00 мВ  0,025 % + 1 единица (через TC разъем)
Сила постоянного тока
24,000 мА (Возбуждение)  0,02% + 2 единицы
24,000 мА (Моделирование)  0,02% + 2 единицы
Сопротивление
15,0 — 400,0 Ω  0.15 Ω (ток. возб. 0,15 — 0.5 мА), 0,1 Ω (ток возб. 0,5 — 2 мА)
401 — 1500 Ω  0,5 Ω (ток возбуждения 0,05 — 0,8 мА)
1500 — 3200 Ω  1 Ω (ток возбуждения 0,05 — 0,4 мА)
Частота
2,0 — 1000,0 CPM  0,05%
1,0 — 1100,0 Гц  0,05%
1,00 кГц — 10,00 кГц  0,25%
Форма сигнала:  Прямоуг. импульс двойной амплитуды 5 В, отклонение -0.1 В
Импульсные RTD трансмиттеры и термопары
Точность измерений:
NI-120  0,2 °C
PT-100 (385)  0,33 °C
PT-100 (393)  0,3 °C
PT-100 (JIS)  0,3 °C
PT-200 (385)  0,2 °C
PT-500 (385)  0,3 °C
PT-1000 (385)  0,2 °C
Разрешение:  0,1 °C
J:  0,7 °C
K:  0,8 °C
T:  0,8 °C
E:  0,7 °C
R:  1,8 °C
S:  1,5 °C
B:  1,4 °C
L:  0,7 °C
U:  0,75 °C
N:  0,9 °C
Разрешение  J, K, T, E, L, N, U: 0,1 °C, 0,1 °F B, R, S: 1 °C, 1 °F
XK  0.6°C
BP  1.2°C
Точность возбуждения:
NI-120  0,2 °C
PT-100 (385)  0,33 °C
PT-100 (393)  0,3 °C
PT-100 (JIS)  0,3 °C
PT-200 (385)  0,2 °C
PT-500 (385)  0,3 °C
PT-1000 (385)  0,2 °C
Разрешение:  0,1 °C
Примечание:  Точность указана для измерений, произведенных с помощью 4 проводов.
J:  0,7 °C
K:  0,8 °C
T:  0,8 °C
E:  0,7 °C
R:  1,4 °C
S:  1,5 °C
B:  1,4 °C
L:  0,7 °C
U:  0,75 °C
N:  0,9 °C
Разрешение  J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C
XK  0.6°C
BP  1.2°C
Характеристики
Линейно-нарастающие функции
Функции возбуждения:  Напряжение, ток, сопротивление, частота, температура, давление.
Функции линейного изменения  Функция медленного, быстрого (пилообразного) линейного изменения, пошаговая функция (шаг 25%)
Функция мощности сети
Напряжение:  24 В
Точность:  10%
Максимальный ток:  22 мА, защита от короткого замыкания
Функции задания шага
Функции возбуждения:  Напряжение, ток, сопротивление, частота, температура, давление.
Шаги  25% диапазона, 100% диапазона
Требования к окружающей среде
Рабочая температура
Температура хранения
Высота над уровнем моря при эксплуатации
Электрический стандарт безопасности
Полученные одобрения
   EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Механические и общие характеристики
Размер
Вес
Батареи
Замена аккумуляторных батарей
   Отдельный аккумуляторный отсек для быстрой смены батареи без нарушения отметки о калибровке
Присоединения к боровому порту
   Соединительное устройство блока давления

Fluke 725 Калибратор

Производитель:

Модель:

725

Дата:

1998

Категория:

Группа:

Описание:

Multifunction Process Calibrator

Информация

The Fluke 725 Multifunction Process
Calibrator is a powerful yet easy-to-use field
calibrator. Use the measure and source functions
to test and calibrate almost any process
parameter.
• Compact, streamlined shape makes it easy
to carry
• Rugged, reliable design stands up to field use
• Easy to read measure/source screen lets you
view input and output simultaneously
• Measure volts, mA, RTDs, thermocouples,
frequency, and ohms to test sensors and
transmitters
• Source/simulate volts, mA, thermocouples, RTDs,
frequency, and ohms to calibrate transmitters
• Measure/source pressure using any of 28
Fluke 700Pxx Pressure Modules
• Source mA with simultaneous pressure
measurement to conduct valve and I/P tests
• Support flow meter testing with frequency
and CPM functions
• Perform fast linearity tests with 25% and
100% steps
• Execute remote tests with auto step and auto
ramp
• Power transmitters during test using loop
supply with simultaneous mA measurement
• Store frequently-used test setups for later use
• Backlight lets you work in poor light
• Remote Interface allows benchtop automated
operations
• Large battery capacity of four AA cells
• Battery door for easy changes

показать больше

Руководство пользователя

Hасто́льная кни́га тип:

Руководство пользователя

Страницы:

75

Размер:

1.27 Mbytes (1326640 Bytes)

Язык:

english

Пересмотр:

Hасто́льная кни́га ID:

Дата:

Качество:

Электронный документ, ни сканирование, очень хорошо читается.

Дата загрузки:

2014 11 29

MD5:

05e7a6529bab0acbfff41ee1ba5f975c

Загрузки:

723

Информация

Introduction …
1
Contacting
Fluke …
1
Standard
Equipment …
3
Safety
Information …
3
Getting Acquainted with the Calibrator

8
Input and Output Terminals

8
Keys …
10
Display

13
Getting Started

14
Shut Down Mode

14
Contrast Adjustment

16
Using Measure
Mode …
17
Measuring Electrical Parameters (Upper Display)
… 17
Current Measurement with Loop
Power …
17
Measuring Electrical Parameters (Lower Display)
… 19
Measuring
Temperature …
20
Using
Thermocouples …
20
Using Resistance-Temperature Detectors (RTDs)
… 23
Measuring Pressure

26
Zeroing with Absolute Pressure Modules

27
Using Source
Mode …
29
Sourcing 4 to 20
mA …
29
Simulating a 4- to 20-mA
Transmitter …
29
Sourcing Other Electrical Parameters

29
Simulating
Thermocouples …
32
Simulating
RTDs …
32
Sourcing Pressure

35
Setting 0 % and 100 % Output Parameters

37
Stepping and Ramping the
Output …
37
Manually Stepping the mA
Output …
37
Auto Ramping the Output

38
Storing and Recalling
Setups …
38
Calibrating a
Transmitter …
39
Calibrating a Pressure
Transmitter …
41
Calibrating an I/P
Device …
43
Testing an Output
Device …
45
Remote Control
Commands …
46
Replacing the Batteries

49
Replacing the Fuses

49
Maintenance …
50
Cleaning the
Calibrator …
50
Service Center Calibration or
Repair …
50
Replacement Parts

51
Accessories

53
External Fluke Pressure Module Compatibility
… 53
Specifications …
56
DC Voltage
Measurement …
56
DC Voltage Source

56
Millivolt Measurement and
Source* …
56
DC mA Measurement and Source

57
Ohms
Measurement …
57
Ohms Source

57
Frequency Measurement

57
Frequency
Source …
58
Temperature,
Thermocouples …
58
Loop Power
Supply …
59
RTD Excitation (simulation)

59
Temperature, RTD Ranges, and Accuracies (ITS-90)
… 59
Pressure
Measurement …
60
General Specifications

60

показать больше

Руководство по техническому обслуживанию

Hасто́льная кни́га тип:

Руководство по техническому обслуживанию

Страницы:

79

Размер:

1.89 Mbytes (1978303 Bytes)

Язык:

english

Пересмотр:

2

Hасто́льная кни́га ID:

PN 667581

Дата:

Качество:

Электронный документ, ни сканирование, очень хорошо читается.

Дата загрузки:

2014 11 29

MD5:

5abf0bd5f323599cad06510518e760ae

Загрузки:

2166

Информация

Calibration Manual

Introduction …
1
Contacting
Fluke …
1
Read First – Safety
Information …
2
International Symbols

3
Specifications …
4
DC Voltage Measurement (724 and
725) … 4
DC Voltage Measurement and Source
(726) … 4
DC Voltage Source (724 and 725)

4
DC mA Measurement and Source (724 and 725)
(Measurement only for
724) …
4
DC mA Measurement and Source (726)
… 5
Ohms Measurement (724 and 725)

5
Ohms Measurement (726)

5
Ohms Source (724 and 725)

5
Ohms Source
(726) …
6
Frequency Measurement
(725) …
6
Frequency Measurement
(726) …
6
Frequency Source (725)

6
Frequency Source (726)

7
Millivolt Measurement and Source (724 and
725) … 7
Temperature, Thermocouples (724 and
725) … 7
Temperature, Thermocouples (726)
… 8
Temperature, RTD Ranges, and Accuracies (724 and 725)
… 9
RTD Accuracy (Read and Source) (ITS-90)
(726) … 10
Loop Power supply

11
Pressure Measurement (725 and 726)
… 11
Pulse Read and Pulse Source (726)

11
Pressure Units Availability (725)

11
General
Specifications …
12
Cleaning the
Calibrator …
12
Replacing the Batteries

12
Fuses

13
Remote Control Interface (725 and 726)
… 14
Serial Command List
(726) …
17
Required Equipment

22
Performance
Tests …
24
Preparing for the Performance Tests
… 24
Upper Display Voltage Measurement Tests
… 25
Lower Display mV/TC Measurement Tests
… 26
Lower Display Voltage Measurement
Tests … 27
Upper Display mA Measurement
Tests …
28
Lower Display mA Measurement Tests (725 and
726) … 29
Lower Display Frequency Measurement Test (725 and 726)
… 30
Lower Display Frequency Source Test (725 and 726)
… 31
Lower Display 4-Wire Resistance Measurement
Tests … 32
Lower Display 3-Wire RTD Measurement
… 33
Lower Display Thermocouple Measurement
Tests … 34
Lower Display Thermocouple Source Test
… 35
Lower Display mA Source Tests (725 and 726)
… 36
Lower Display mV Source
Tests …
37
Lower Display Voltage Source Tests
… 37
Lower Display Ohms Source Tests

38
Pressure Module Input
(725/726) …
39
Calibration
Adjustment …
39
Setup …
39
Initiating Communication (724 and 725)
… 40
Calibration Adjustment Procedures (724 and 725)
… 41
Cal Volts Input

41
Cal Volts Output

42
Cal mA Input (Fluke 725 Only)

43
Cal mA Output (725 Only)

44
Cal mV
Input …
45
Cal mV Output

46
Cal Thermocouples

47
Cal Ohms Hi
Source …
48
Cal Ohms Low Source

49
Cal RTD Low
Range …
50
Cal RTD Hi Range

51
Cal ISO Volts

52
Cal ISO
mA …
53
Calibration Adjustment Procedures (726)
… 54
Initiating Communication

54
Calibrating Isolated mA
Input …
55
Calibrating Isolated Voltage Input
… 56
Calibrating mA Input

57
Calibrating mA
Output …
58
Calibrating Voltage
Input …
59
Calibrating Voltage Output

60
Calibrating Low Ohms
Source …
61
Calibrating High Ohms Source

62
Calibrating Ohms
Measure …
63
Calibrating Thermocouple mV Output
… 64
Calibrating Thermocouple mV Input
… 65
Calibrating Thermocouple
CJC: …
66
Replaceable
Parts …

показать больше

Технический паспорт

Hасто́льная кни́га тип:

Технический паспорт

Страницы:

2

Размер:

81.77 Kbytes (83737 Bytes)

Язык:

english

Пересмотр:

B

Hасто́льная кни́га ID:

1268002

Дата:

Качество:

Электронный документ, ни сканирование, очень хорошо читается.

Дата загрузки:

2014 11 30

MD5:

df0e5840867765255b2d5065394e41a6

Загрузки:

796

Добавить комментарий

Понравилась статья? Поделить с друзьями:
  • Липоферон инструкция по применению цена в россии
  • Должностная инструкция заведующего поликлиникой районной больницы
  • Руководство по эксплуатации ниссан тиида с11
  • Аквасол с пантенолом спрей инструкция по применению
  • Увлажнитель воздуха огонек инструкция по применению