Contents
-
Table of Contents
-
Bookmarks
Quick Links
5502E
Multi-Product Calibrator
Getting Started
PN 4238530
November 2012
© 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
Related Manuals for Fluke 5502E
Summary of Contents for Fluke 5502E
-
Page 1
5502E Multi-Product Calibrator Getting Started PN 4238530 November 2012 © 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. -
Page 2
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. -
Page 3
OPERATOR SAFETY SUMMARY WARNING HIGH VOLTAGE is used in the operation of this equipment LETHAL VOLTAGE may be present on the terminals, observe all safety precautions! To prevent electrical shock hazard, the operator should not electrically contact the output HI or sense HI terminals or circuits connected to these terminals. During operation, lethal voltages of up to 1020 V ac or dc may be present on these terminals. -
Page 5: Table Of Contents
Connect to Line Power ………………8 Select Line Frequency ………………8 Placement ………………….10 Airflow Considerations ………………11 Instruction Manuals ………………. 11 5502E Getting Started Manual …………..11 5502A Operators Manual …………….11 General Specifications ………………12 Detailed Specifications ………………13 DC Voltage ………………..13 DC Current ………………..
-
Page 6
5502E Getting Started AC Voltage, Triangle Wave Characteristics (typical) ……..24 AC Current (Non-Sine Wave) …………… 24 AC Current (Non-Sine Wave) (cont.) …………25 AC Current, Square Wave Characteristics (typical) ……..25 AC Current, Triangle Wave Characteristics (typical) ……..25… -
Page 7
Title Page Symbols ……………………4 Standard Equipment ………………..7 Mains-Power Cord Types Available from Fluke Calibration …….. 10 Verification Tests for DC Voltage (Normal) …………26 Verification Tests for DC Current (AUX) …………. 27 Verification Tests for Resistance …………….28 Verification Tests for AC Voltage (Normal) ………… -
Page 8
5502E Getting Started… -
Page 9
List of Figures Figure Title Page 5502E Multi-Product Calibrator …………….2 RS-232 Remote Connection ………………. 6 Access the Fuse and Select Line Voltage …………… 9 Mains-Power Cord Types Available from Fluke Calibration …….. 10 Allowable Duration of Current >11 A …………..14… -
Page 10
5502E Getting Started… -
Page 11: Getting Started
The 300 MHz and 600 MHz options are not compatible with the 5502E. The 5502E does not have SCOPE and TRIG OUT terminals. The SCOPE and TRIG OUT buttons are on the front panel of the 5502E, but they are non-functional and beep when pushed.
-
Page 12
488.1-1987 and 488.2-1987. • A EIA Standard RS-232 serial data interface to print, show, or move internally stored calibration constants, and for remote control of the 5502E. • A pass-through RS-232 serial data interface to send data to the Unit Under Test… -
Page 13: Safety Information
Multi-Product Calibrator Safety Information Safety Information In this manual, A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: •…
-
Page 14: Symbols
WEEE Directive Annex I, this product is classed as category 9 «Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Risk of Danger. Important …
-
Page 15: Contact Fluke Calibration
• Brazil: +55-11-3759-7600 • Anywhere in the world: +1-425-446-6110 To see product information and download the latest manual supplements, visit Fluke Calibration’s website at www.flukecal.com. To register your product, visit http://flukecal.com/register-product. Overload Protection The Calibrator supplies reverse-power protection, fast output disconnection, and/or fuse protection on the output terminals for all functions.
-
Page 16: Remote Operation (Rs-232)
5502E Getting Started Remote Operation (RS-232) There are two rear-panel serial data RS-232 ports: SERIAL 1 FROM HOST, and SERIAL 2 TO UUT (see Figure 2). Each port is dedicated to serial-data communications to operate and control the Product during calibration procedures. For complete information on remote operation, see Chapter 5 of the Operators Manual.
-
Page 17: Unpack And Inspection
Calibration Service Center (see “Contact Fluke Calibration”). A performance test is in the “Maintenance” section of Chapter 7 in the Operators Manual. If you ship the calibrator to Fluke Calibration, use the initial container. If it is not available, you can get a new container from Fluke Calibration with the Calibrator model and serial number.
-
Page 18: Connect To Line Power
If it is necessary to use a different type, refer to Table 3 and Figure 4 for a list and illustration of the line-power plug types available from Fluke Calibration. After you make sure that the line voltage selection is set correctly and that the correct fuse for that line voltage is installed, connect the calibrator to a correctly-grounded three-prong outlet.
-
Page 19
Multi-Product Calibrator Select Line Frequency Line Voltage Indicator Changing Line Fuse Changing Line Voltage gjh004.eps Figure 3. Access the Fuse and Select Line Voltage… -
Page 20: Placement
5502E Getting Started Table 3. Mains-Power Cord Types Available from Fluke Calibration Fluke Calibration Option Type Voltage/Current Number North America 120 V/15 A LC-1 North America 240 V/15 A LC-2 Universal Euro 220 V/15 A LC-3 United Kingdom 240 V/13 A…
-
Page 21: Airflow Considerations
• 5502E Getting Started (PN 4238530) One of each manual above comes with the instrument. Refer to the Fluke Calibration Catalog or speak to a Fluke Calibration sales representative (see “Contact Fluke Calibration”) if more printed copies are necessary. The manuals are also available on the Fluke Calibration website.
-
Page 22: General Specifications
The following tables list the 5502E specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502E has been turned off. (For example, if the 5502E has been turned off for 5 minutes, the warm-up period is 10 minutes.) All specifications apply for the temperature and time period indicated.
-
Page 23: Detailed Specifications
Weight (without options) ……..22 kg (49 lb) Absolute Uncertainty Definition ……The 5502E specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total specification of the 5502E for the temperature range indicated.
-
Page 24: Dc Current
Formula 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amperes. For example, 17 A, at 23 °C could be provided for 60-23-17 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
-
Page 25: Resistance
Multi-Product Calibrator Detailed Specifications Resistance Absolute Uncertainty, tcal ±5 °C ±(% of output + floor) Resolution Allowable Current Floor (Ω) Time and temp since ohms Range % of output (Ω) zero cal 90 Day 1 Year 12 hrs ±1 °C 7 days ±5 °C 0 to 10.999 Ω…
-
Page 26: Ac Voltage (Sine Wave)
5502E Getting Started AC Voltage (Sine Wave) Absolute Uncertainty, tcal Max Distortion and ±5 °C ±(% of output + μV) Noise 10 Hz to Range Frequency Resolution Max Burden 5 MHz Bandwidth ±(% of output + 90 Day 1 Year…
-
Page 27: Ac Current (Sine Wave)
60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
-
Page 28: Ac Current (Sine Wave) (Cont.)
60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
-
Page 29: Capacitance
Multi-Product Calibrator Detailed Specifications Capacitance Absolute Uncertainty, tcal ±5 °C Allowed Frequency or Charge-Discharge Rate [1] [2] [3] ±(% of output + floor) Range Resolution Min and Max to Typical Max for Typical Max for 90 Day 1 Year Meet <0.5 % Error <1 % Error Specification…
-
Page 30: Temperature Calibration (Thermocouple)
5502E Getting Started Temperature Calibration (Thermocouple) Absolute Uncertainty Absolute Uncertainty Source/Measure tcal ±5 °C ± Source/Measure tcal TC Type Range °C Range °C Type °C ±5 °C ± °C 90 Day 1 Year 90 Day 1 Year 600 to 800 0.42…
-
Page 31: Temperature Calibration (Rtd)
The subsequent paragraphs provide additional specifications for the 5502E Calibrator ac voltage and ac current functions. These specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502E has been turned off. All extended range specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more than 5 °C.
-
Page 32: Ac Voltage (Sine Wave) Extended Bandwidth
5502E Getting Started AC Voltage (Sine Wave) Extended Bandwidth 1-Year Absolute Uncertainty Range Frequency Max Voltage Resolution tcal ±5 °C Normal Channel (Single Output Mode) 1.0 to 33 mV Two digits, e.g., 25 mV ±(5.0 % of output 34 to 330 mV 0.01 to 9.99 Hz…
-
Page 33: Ac Voltage (Non-Sine Wave) (Cont.)
Multi-Product Calibrator Additional Specifications AC Voltage (Non-Sine Wave) (cont.) 1-Year Absolute Uncertainty, tcal Square Wave Frequency Max Voltage Resolution ±5 °C, ±(% of output + % of range) Range (p-p) Normal Channel (Single Output Mode) 0.01 to 10 Hz 5.0 + 0.5 Two digits on each range 10 to 45 Hz 0.25 + 0.5…
-
Page 34: Ac Voltage, Square Wave Characteristics
5502E Getting Started AC Voltage, Square Wave Characteristics Risetime @ Overshoot Settling Time @ 1 kHz @ 1 kHz Duty Cycle Range Duty Cycle Uncertainty 1 kHz Typical Typical Typical ±(0.02 % of period + 100 ns), 50 % duty cycle <10 μs to 1 % of…
-
Page 35: Ac Current (Non-Sine Wave) (Cont.)
Multi-Product Calibrator Additional Specifications AC Current (Non-Sine Wave) (cont.) 1-Year Absolute Uncertainty tcal ±5 °C Max Current Square Wave Range p-p Frequency ±(% of output + % of range) Resolution 10 to 45 Hz 0.25 + 0.5 0.047 to 0.65999 mA 45 Hz to 1 kHz 0.25 + 0.25 Six digits…
-
Page 36: Performance Tests
The limits were calculated by adding or subtracting the 90-day specification from the output value. There is no built-in factor for measurement uncertainty. These performance tests replace those found in the 5502A Operators Manual and only apply to the 5502E. Table 4. Verification Tests for DC Voltage (Normal)
-
Page 37: Verification Tests For Dc Current (Aux)
Multi-Product Calibrator Performance Tests Table 5. Verification Tests for DC Current (AUX) Range Output Lower Limit Upper Limit 329.999 μA 0.000 μA -0.020 μA 0.020 μA 329.999 μA 190.000 μA 189.957 μA 190.043 μA 329.999 μA -190.000 μA -190.043 μA -189.957 μA 329.999 μA 329.000 μA…
-
Page 38: Verification Tests For Resistance
5502E Getting Started Table 6. Verification Tests for Resistance Range Output Lower Limit Upper Limit 10.999 Ω 0.000 Ω -0.0010 Ω 0.0010 Ω 10.999 Ω 2.000 Ω 1.9989 Ω 2.0011 Ω 10.999 Ω 10.900 Ω 10.8980 Ω 10.9019 Ω 32.999 Ω…
-
Page 39
Multi-Product Calibrator Performance Tests Table 6. Verification Tests for Resistance (cont.) Range Output Lower Limit Upper Limit 109.999 kΩ 33.000 kΩ 32.9971 kΩ 33.0028 kΩ 109.999 kΩ 109.000 kΩ 108.9910 kΩ 109.0089 kΩ 329.999 kΩ 119.000 kΩ 118.9872 kΩ 119.0127 kΩ 329.999 kΩ… -
Page 40: Verification Tests For Ac Voltage (Normal)
5502E Getting Started Table 7. Verification Tests for AC Voltage (Normal) Range Output Frequency Lower Limit Upper Limit 32.999 mV 3.000 mV 45 Hz 2.977 mV 3.022 mV 32.999 mV 3.000 mV 10 kHz 2.977 mV 3.022 mV 32.999 mV 30.000 mV…
-
Page 41
Multi-Product Calibrator Performance Tests Table 7. Verification Tests for AC Voltage (Normal) (cont.) Range Output Frequency Lower Limit Upper Limit 3.29999 V 3.00000 V 100 kHz 2.99437 V 3.00563V 3.29999 V 3.00000 V 450 kHz 2.98659 V 3.01340 V 3.29999 V 3.29000 V 1 MHz 2.250 V… -
Page 42: Verification Tests For Ac Current
5502E Getting Started Table 8. Verification Tests for AC Current Range Output Frequency Lower Limit Upper Limit 329.99 μA 33.00 μA 32.87 μA 33.13 μA 1 kHz 329.99 μA 33.00 μA 32.60 μA 33.40 μA 10 kHz 329.99 μA 33.00 μA 32.20 μA…
-
Page 43
Multi-Product Calibrator Performance Tests Table 8. Verification Tests for AC Current (cont.) Range Output Frequency Lower Limit Upper Limit 32.999 mA 19.0000 mA 30 kHz 18.935 mA 19.065 mA 32.999 mA 32.9000 mA 10 Hz 32.849 mA 32.951 mA 32.999 mA 32.9000 mA 1 kHz 32.886 mA… -
Page 44
5502E Getting Started Table 8. Verification Tests for AC Current (cont.) Range Output Frequency Lower Limit Upper Limit 2.99999 A 2.99000 A 10 kHz 2.92520 A 3.05480 A 20.5000 A 3.3000 A 500 Hz 3.2954 A 3.3046 A 20.5000 A 3.3000 A… -
Page 45: Verification Tests For Capacitance
Multi-Product Calibrator Performance Tests Table 9. Verification Tests for Capacitance Test Frequency Range Output Lower Limit Upper Limit or Current 0.3999 nF 0.2200 nF 5 kHz 0.2192 nF 0.2308 nF 0.3999 nF 0.3500 nF 1 kHz 0.3387 nF 0.3613 nF 1.0999 nF 0.4800 nF 1 kHz…
-
Page 46: Verification Tests For Thermocouple Simulation
5502E Getting Started Table 9. Verification Tests for Capacitance (cont.) Test Frequency Range Output Lower Limit Upper Limit or Current 540 μA dc 3.299 mF 2.000 mF 1.9902 mF 2.0098 mF 800 μA dc 3.299 mF 3.000 mF 2.9868 mF 3.0132 mF…
-
Page 47: Verification Tests For Frequency
Multi-Product Calibrator Performance Tests Table 12. Verification Tests for Frequency Range, Normal Output, Normal, V Frequency Lower Limit [1] Upper Limit [1] Output, V 119.00 Hz 118.99602 Hz 119.00398 Hz 120.0 Hz 119.99600 Hz 120.00400 Hz 3.29999 3.00000 1000.0 Hz 999.974000 Hz 1000.026000 Hz 100.00 kHz…
-
Page 48
5502E Getting Started…
Fluke 5502E Test Equipment PDF User Guides and Manuals for Free Download: Found (1) Manuals for Fluke 5502E Device Model (Getting Started)
More Test Equipment Device Models:
-
Phoenix Contact
PROFIPOL
PROFIPOL®PHOENIX CONTACT GmbH & Co. KG Flachsmarktstraße 8 · D-32825 Blomberg Telefon +49 (0) 5235-3-00 · Fax +49 (0) 5235-3-41200 www.phoenixcontact.com · [email protected] 3456D Bedienungsanleitung Operating manualF Mode d‘emploiEManuel de instrucciones � …
PROFIPOL Test Equipment, 19
-
Fluke
i1000s
Manual Supplement© 2001 Fluke Corporation. All rights reserved. Printed in U.S.A.Manual Title: i1000s Calibration Supplement Issue: 1Part Number: 1574933 Issue Date: 7/01Print Date: May 2001 Page Count: 2Revision/Date:This supplement contains information necessary to ensure the accuracy ofthe above manual. Enter the c …
i1000s Measuring Instruments, 3
-
PASCO
ME-8236
Instruction Manual012-13762DMaterials Testing MachineME-8236Part of theComprehensive Materials Testing SystemME-8244 7100 N Load CellLeadscrewsCalibration Rod and NutLoad Bar (crosshead)Tensile Samples* (60)Sensor CableLoad Bar Round NutVelcro® Loop MaterialSafety Shield (2)Velcro® Hook MaterialWARNING: Provide prope …
ME-8236 Test Equipment, 18
-
Aeroflex
PSD60-2R
OPERATION MANUAL PSD60-2R FUEL QUANTITY TEST SET MANUAL NUMBER:E6-0597-01 (75322) REVISION:A1 DATE:08/24/2011 WARNING: INFORMATION SUBJECT TO EXPORT CONTROL LAWS This document contains controlled technology or technical data under the jurisdiction of the Export Administration Regulations (EAR), 15 CF …
PSD60-2R Test Equipment, 22
Recommended Documentation:
Types of Manuals:
The main types of Fluke 5502E instructions:
- User guide — rules of useing and characteristics
- Service manual — repair, diagnostics, maintenance
- Operation manual — description of the main functions of equipment
Test Equipment Instructions by Fluke:
-
Fluke 1550C
September 2010 Rev. 1, 6/18 © 2010-2018 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. 1550C/1555 Insulation Tester Calibration Manual …
1550C Test Equipment, 26
-
Fluke P3000 Series
PN 3952260 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies. P3000 Series Pneumatic Deadweight Tester Users Manual w ww . . co [email protected] …
P3000 Series Test Equipment, 40
-
Myron L ULTRAPEN PT4
4 5ULTRAPEN™ PT4Free Chlorine Equivalent (FCE TM)& Temperature PenS ACCURATE S RELIABLE S SIMPLE Professional Water Analysis• Accuracy of up to ±0.3 ppm FCE • Proprietary Sensor Design• Calibration with pH and ORP Standard Solutions• Automatic Temperature Compensation • …
ULTRAPEN PT4 Measuring Instruments, 6
-
ERICHSEN 243/I
testing equipment for quality managementFlowcupsModels 243A Simple Methodto Measure ViscosityASTM D 1200SNV 37 110SIS 18 41 15DIN 53 211ISO 2431DIN EN ISO 2431Technical Description and Operating Instructions …
243/I Test Equipment, 4
-
ITech IT5101
Internal Resistance Tester IT5101 Programming Guide Model: IT5101/IT5101E/IT5101H Version: 1.0 …
IT5101 Test Equipment, 48
-
Gerus GERUTEST NTB5
GERUS Seite 1 von 9 Bedienungsanleitung zum Düsenprüfstand GERUTEST NTB5 Operation manual for Nozzle test bed GERUTEST NTB5 GERUS Apparatebau GmbH & Co KG Engelschalkstrass …
GERUTEST NTB5 Test Equipment, 9
-
BRUEL & KJAER 4128
Product DataBrüel & KjærBKHead and Torso Simulator — Type 4128Head and Torso Simulator Type4128 has been developed for objectivein-situ research and evaluation of:❍ telephones and headsets❍ audioconference devices❍ microphones❍ headphones❍ hearing aids and hearing protectorsThe design of Type 4128 is …
4128 Test Equipment, 6
Калибраторы многофункциональные Fluke 5502E и Fluke 5502A со встраиваемыми модулями поверки осциллографов 300 МГц, 600 МГц зарегистрированы в Госреестре средств измерений РФ под № 55804-13 и имеют Свидетельство об утверждении типа СИ №53341 (срок действия до 12 октября 2023 г.).
ВНИМАНИЕ: Калибраторы многофункциональные Fluke 5502E и Fluke 5502A в базовом исполнении (без встроенных модулей поверки осциллографов 300 МГц, 600 МГц) сняты с производства и к заказу недоступны. К заказу доступно ограниченное количество калибраторов Fluke 5502A/3 и Fluke 5502A/3.
Особенности:
- Калибровка / поверка широкого спектра оборудования для электрических измерений
- Надежная защита цепей предотвращает ошибки оператора
- Прочный переносной футляр со встроенными ручками и колесами и съемные стенки для доступа к передней и задней панелям прибора.
- Выходное значение тока до 120 А при одновременном использовании с опциональным усилителем тока Fluke 52120A, управляемым напряжением
- Эргономичные прочные ручки облегчают транспортировку прибора
- Доступная цена
Калибратор Fluke 5502А позволяет решить распространенные задачи калибровки / поверки, включая:
- портативные и настольные аналоговые и цифровые мультиметры с количеством разрядов до 4,5;
- токовые клещи;
- щитовые измерители;
- электронные термометры;
- диаграммные самописцы;
- осциллографы (при наличии соответствующей опции);
- координатные самописцы;
- регистраторы данных.
Универсальный калибратор Fluke 5502А подходит для использования с наиболее распространенными цифровыми мультиметрами с количеством разрядов 3,5 и 4,5 и другими устройствами. Конструкция прибора позволяет удобно транспортировать его и проводить калибровку на рабочем месте. Модель 5502A можно полностью автоматизировать с помощью ПО MET/CAL® Plus для управления процессом калибровки. Это идеальный калибратор для профессионалов в области метрологии, которым требуется решение для калибровки электрических параметров с низкой или средней точностью.
Калибратор Fluke 5502A является источником постоянного напряжения и тока, переменного напряжения и тока с несколькими формами сигнала и гармониками, имеет два параллельных выхода по напряжению или напряжению и току для имитации источников постоянного и переменного тока с управлением фазой, сопротивлением, емкостью, термопарами и термометрами сопротивления. Многофункциональный калибратор Fluke 5502А может также использоваться с термопарами и моделями термопар. Две опции предоставляют дополнительную возможность калибровки осциллографов с частотой до 300 МГц или до 600 МГц. Также с помощью усилителя тока Fluke 52120А, управляемого напряжением, и токовых катушек Fluke 5500А/COIL можно калибровать токоизмерительные клещи, рассчитанные на 6000 А.
Более подробная информация по Fluke 5502A представлена в брошюре, которую можно скачать по ссылке: Брошюра Fluke 5502A.pdf (2.6 Mb).
Компания FLUKE предлагает экономичную версию многофункционального калибратора Fluke 5502Е, позволяющую сэкономить до 50% от стоимости базовой модели Fluke 5502A. Отличительной особенностью данной модификации является невозможность дооснащения калибратора такими дополнительными опциями, как:
- Калибровка / поверка Осциллографов
- Калибровка / поверка Измерителей мощности постоянного и переменного тока
- Калибровка / поверка Анализаторов качества электрической энергии и анализаторов гармоник
С подробными техническими характеристиками Вы можете познакомиться, скачав файл размером 2 Mb по ссылке: TTX Fluke 5502E.
Fluke 5502A и Fluke 5502Е Технические характеристики Многофункциональных калибраторов
Fluke 5502A и Fluke 5502E Модели многофункциональных калибраторов
Номер в Госреестре СИ 55804-13
Fluke 5502Е экономичная версия многофункционального калибратора Fluke 5502A, позволяющую сэкономить до 50% от стоимости базовой модели.
Отличительной особенностью данной модификации являются отсутствие следующих функций:
· Функции питания постоянного и переменного тока (одновременный выходной сигнал по напряжению и току);
· Функция двойного напряжения (одновременный выходной сигнал по напряжению/напряжению);
· Функция «Фаза»;
· Функция «Гармоники».
Разъемы опорного сигнала 10 МГц IN (вход) и OUT (выход) на задней панели не функциональные. Клеммы предназначены только для обслуживания.
У модели Fluke 5502E нет клемм SCOPE и TRIG OUT на передней панели.
Опции калибровки / поверки осциллографов полосой 300 МГц и 600 МГц несовместимы с моделью Fluke 5502E.
Госреестр СИ №55804-13
Fluke 5502А полная версия многофункционального калибратора Fluke.
Отличительной особенностью данной модификации являются наличие следующих функций:
· Функции питания постоянного и переменного тока (одновременный выходной сигнал по напряжению и току);
· Функция двойного напряжения (одновременный выходной сигнал по напряжению/напряжению);
· Функция «Фаза»;
· Функция «Гармоники».
Разъемы опорного сигнала 10 МГц IN (вход) и OUT (выход) на задней панели функциональные.
Номер в Госреестре СИ №55804-13
Fluke 550А/3 полная версия многофункционального калибратора Fluke 5502A со встроенной опцией поверки осциллографов до 300 МГц.
Отличительной особенностью данной модификации являются наличие встроенной опции SC300, предназначенной для поверки и калибровки аналоговых и цифровых осциллографов полосой до 300 МГц.
Разъемы опорного сигнала 10 МГц IN (вход) и OUT (выход) на задней панели.
Наличие клемм SCOPE и TRIG OUT на передней панели.
Номер в Госреестре СИ №55804-13
Fluke 550А/6 полная версия многофункционального калибратора Fluke 5502A со встроенной опцией поверки осциллографов до 600 МГц.
Отличительной особенностью данной модификации являются наличие встроенной опции SC600, предназначенной для поверки и калибровки аналоговых и цифровых осциллографов полосой до 600 МГц.
Разъемы опорного сигнала 10 МГц IN (вход) и OUT (выход) на задней панели.
Наличие клемм SCOPE и TRIG OUT на передней панели.
Номер в Госреестре №61033-15
Усилитель тока, управляемый напряжением, Fluke 52120A обеспечивает на выходе постоянный ток до 100 А и переменный ток до 120 А частотой до 10 КГц с точностью до 140 миллионных долей.
Принимает входные сигналы постоянного или переменного тока от любого калибратора, генератора сигналов или источника питания: напряжение до 2 В или тока до 200 мА.
Применение опциональных токовых катушек позволяет создавать выходной ток до 3 000 А или до 6 000 А.
Прочный кейс 5522A/CARRYCASE предназначен для транспортировки калибраторов Fluke 5502Е, 5502А и Fluke 5522А с любыми встроенными опциями.
Основные особенности:
- Прочность
- Герметичность
- Удобство транспортировки благодаря колесам и выдвижной ручке
- Наличие съемных передней и задней панелей для обеспечения работы с калибратором в полевых условиях без извлечения из кейса.
Прочный переносной кейс 55XX/CASE с колесами
Прочный кейс для транспортировки калибраторов Fluke 5080A, Fluke 5502E, 5502A и Fluke 5522A с любыми встроенными опциями.
Основные особенности:
- Прочность
- Герметичность
- Удобство транспортировки благодаря колесам и выдвижной ручке.
5500A/LEADS Термопара и набор тестовых проводов в сумке / Leads Set for 5500A
Госреестр СИ РФ №61596-15
Катушка для калибровки бесконтактных измерителей тока Fluke 5500А/COIL имеет 50 витков и предназначена для поверки и калибровки измерительных токовых клещей и других бесконтактных измерителей тока.
Опция Fluke 9100-200 представляет собой 2 токовых катушки на 10 и на 50 витков в одном корпусе и предназначена для калибровки измерительных токоизмерительных клещей и других бесконтактных измерителей тока.
Не внесена в Госреестр СИ РФ.
Y8021 IEEE-488 Shielded Interface Cable, 1 m / Экранированый интерфейсный кабель IEEE-488 длина 1м
Y8022 IEEE-488 Shielded Interface Cable, 2 m / Экранированый интерфейсный кабель IEEE-488 длина 2м
ПО MET/CAL/TEAMXP компании Fluke Calibration состоит из двух приложений: MET/CAL® и MET/TEAM®.
Основные задачи ПО MET/CAL®:
- Автоматизация калибровки/поверки всех типов контрольно-измерительных приборов и оборудования, в том числе приборов постоянного тока и низких частот, радиочастотных и СВЧ приборов;
- Создание, редактирование, тестирование и документирование процедур калибровки;
- Настройка и составление отчетов для параметров погрешности измерений, занесение данных поверки в протокол.
ПО MET/TEAM® предназначено для организации процесса калибровки и управления парком контрольно-измерительного оборудования. Его основные особенности:
- Отслеживание информации о парке оборудования, включая историю калибровок и технического состояния, список пользователей и заказчиков.
- Анализ и составления отчета о парке оборудования, изготовление фирменных печатных сертификатов и протоколов.
- Импорт данных о парке приборов и результатов калибровки в ПО MET/CAL
- Обеспечения соответствия требованиям стандартов качества, например, ГОСТ ISO 9000-2011, ГОСТ ИСО/МЭК 17025-2009 и других.
MET/CAL-IEEE USB 2 контроллера National Instruments с интерфейсным кабелем USB-IEEE-488 / 2 National Instruments USB-To-GPIB Controllers.
Fluke 5502A и Fluke 5502E Перечень моделей
Fluke 5502A и Fluke 5502E Перечень опций
Fluke 5502A и Fluke 5502E Документация
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under
normal use and service. The warranty period is one year and begins on the date of shipment.
Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the
original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses,
disposable batteries, or to any product which, in Fluke’s opinion, has been misused, altered,
neglected, contaminated, or damaged by accident or abnormal conditions of operation or
handling. Fluke warrants that software will operate substantially in accordance with its functional
specifications for 90 days and that it has been properly recorded on non-defective media. Fluke
does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user
customers only but have no authority to extend a greater or different warranty on behalf of Fluke.
Warranty support is available only if product is purchased through a Fluke authorized sales outlet
or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for
importation costs of repair/replacement parts when product purchased in one country is submitted
for repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of
charge repair, or replacement of a defective product which is returned to a Fluke authorized
service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that service center, with a description of the
difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in
transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid
(FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination,
alteration, accident, or abnormal condition of operation or handling, including overvoltage failures
caused by use outside the product’s specified rating, or normal wear and tear of mechanical
components, Fluke will provide an estimate of repair costs and obtain authorization before
commencing the work. Following repair, the product will be returned to the Buyer transportation
prepaid and the Buyer will be billed for the repair and return transportation charges (FOB
Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM
ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or
exclusion or limitation of incidental or consequential damages, the limitations and exclusions of
this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or
unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not
affect the validity or enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands