Elcometer 345 инструкция по эксплуатации на русском

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Elcometer 345

Coating Thickness Gauge

Operating Instructions

loading

Related Manuals for Elcometer 345

Summary of Contents for Elcometer 345

  • Page 1
    Elcometer 345 Coating Thickness Gauge Operating Instructions…
  • Page 2
    (electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission of Elcometer Limited. A copy of this Instruction Manual is available for download on our Website via www.elcometer.com. Doc No. TMA-0431 Issue 03…
  • Page 3: Table Of Contents

    CONTENTS Section Page About your gauge …………. . . 5 1.1 Gauge features .

  • Page 4
    Calibration adjustment …………16 5.1 Calibration Adjustment Error .
  • Page 5: About Your Gauge

    Our products cover all aspects of coating inspection, from development through application to post application inspection. The Elcometer 345 Coating Thickness Gauge is a world beating product. With the purchase of this gauge you now have access to the worldwide service and support network of Elcometer. For more information visit our website: www.elcometer.com…

  • Page 6: Gauge Features

    MCS.215 (82/84), IMO MSC.244(83), ISO 1461, ISO 2063, IS0 2808-7C, ISO 2808-12, ISO 19840, NF T30-124, SSPC- PA2 (2004). For IMO, PSPC , the thickness coating, should be measured and recorded. Your Elcometer 345 can be used for this. 1.3 WHAT THE BOX CONTAINS •…

  • Page 7
    Consult your local Environmental Authority for further guidance. To maximise the benefits of your new Elcometer 345 Coating Thickness Gauge please take some time to read these Operating Instructions. Do not hesitate to contact Elcometer or your Elcometer supplier if you have any…
  • Page 8: Getting Started

    Note: Before you use your gauge for the first time, remove the paper strip from the battery compartment. 2.1 FITTING BATTERIES The Elcometer 345 may be used with dry cell batteries or rechargeable batteries. 2 x LR03 (AAA) alkaline batteries are supplied with the gauge. To fit or replace batteries: 1.

  • Page 9: Fitting The Harness

    2.4 CONNECTING THE SEPARATE PROBE The Elcometer 345 separate probe gauge has a 4-pin socket at the base of the gauge. Line up the 2 red marks when inserting the plug.

  • Page 10: Switching The Gauge On/Off

    2.5 SWITCHING THE GAUGE ON/OFF To switch on, press Gauges with an integral probe will switch on automatically by placing the probe against a hard surface. To switch off, press and hold until OFF is displayed. The gauge will switch off automatically after one minute of inactivity. 2.6 DISPLAY The display will show the measurement value.

  • Page 11: Backlight Operation

    2.9 BACKLIGHT OPERATION: Elcometer 345 gauges have a backlit display. The backlight is designed for low light operations and can be turned on and off via the keypad. Turning the backlight off will save battery power. To toggle the backlight on/off, press and release .

  • Page 12: Quick Operation Guide

    3 QUICK OPERATION GUIDE separate gauges Switch on 4 seconds Switch off >>>> >>>> Metric <=> >>>> µm or mm <=> mil English bare metal on foil Calibration >4 seconds adjustment for 4 seconds Smooth >>>> >>>> >>>> >>>> to set. Substrates adjust to foil…

  • Page 13
    on thin foil >4 seconds >>>> >>>> >>>> >>>> adjust to foil Calibration adjustment for Rough and on thick foil Special >4 seconds Substrates >>>> to set >>>> >>>> >>>> adjust to foil >>>> to set. >>>> >>>> >>>> keep pressing until first statistic Set User List for Statistical…
  • Page 14
    To Turn >>>> >>>> >>>> >>>> until Statistical Data ‘Off’ >>>> >>>> Stats off Align IR’s >>>> >>>> Print Data Output to continue measurements bare metal To Reset >>>> to adjust to 10 µm (0.5 mil) [or 100 µm (4.0 mil) for scales F2 Factory Calibration and F3] >>>>…
  • Page 15: Taking A Measurement

    4 TAKING A MEASUREMENT Before starting to take measurements it is good practice to check the performance of the gauge using the calibration adjustment procedure — see “Calibration adjustment” on page 16. 1. Hold the gauge with the probe pressed Integral firmly against…

  • Page 16: Calibration Adjustment

    5 CALIBRATION ADJUSTMENT The Elcometer 345 should be adjusted before use, or at least checked to ensure that the calibration adjustment is correct for the conditions of use. Calibration adjustment should be carried out with the appropriate probe, on the same type of metal, the same curvature and similar surface finish as that of the item to be measured.

  • Page 17: Calibration Adjustment For Smooth Surfaces

    5.2 CALIBRATION ADJUSTMENT FOR SMOOTH SURFACES Use this calibration adjustment procedure for measurements on smooth, flat surfaces. Note: The key-lock feature requires the keys used to access these routines to be pressed for 4 seconds before any response is made. This is to prevent accidental operation. 1.

  • Page 18: Special Calibration Adjustment

    6. Confirm that the gauge reads correctly at zero and at the calibration foil value. If required the calibration adjustment can be repeated to ensure its accuracy. To re-adjust the gauge, repeat all the steps in Section 5.2. Note: As well as calibration foils, certified coated standards are available — see “Spares and accessories” on page 32.

  • Page 19: The Correction Value Method

    4. Press and hold for more than 5 seconds to adjust this average to the foil value. 5. Press to accept the low calibration or wait 7 seconds to accept the calibration automatically. The display will now show CH1 (= Calibrate High). 6.

  • Page 20: To Recall The Factory Calibration

    The correction value is determined according to the roughness of the surface per ISO 8501, the Surface Roughness Comparator Elcometer 125. If the roughness of the surface under the coating is unknown a default correction value of 25 µm (1.0 mil) is subtracted from each reading. For further details see EN19840 ‘Paints and Varnishes — Corrosion protection of steel structures by protective paint systems — Measurement and acceptance criteria of the dry film thickness”.

  • Page 21: Statistical Summary

    6.1 STATISTICS The statistics display on the Elcometer 345 shows a preset combination of the five statistical functions which may be shown on the display in any user defined order by pressing one key. The feature is switched on by selecting at least one statistical value to be displayed — see “Setting The User List For Statistical Values”…

  • Page 22: Clearing Statistical Data

    is pressed the gauge displays the next statistical value on the user list for a minimum of 3 seconds. is pressed the gauge displays the previous statistical value on the user list for a minimum of 3 seconds. • To display the rest of the list of values must be pressed while the current value is on the display.

  • Page 23: Setting The User List For Statistical Values

    6.4 SETTING THE USER LIST FOR STATISTICAL VALUES • To create a user list press >>>> >>>> . The display will show CLS to indicate ‘Clear Statistics?’ • Press to enter the menu for setting the user list of statistical values. The gauge displays the symbol of the first Statistics item on the statistics menu with its numerical position in the user list shown in the top left of the display.

  • Page 24: Example Of Statistical User List

    • To enter the statistical value displayed at the next available position on the list press . If this is the first entry then the current user list is deleted and replaced with a new user list and it appears in the top left of the display.

  • Page 25: Data Output

    Your gauge is fitted with an infrared data output feature. This operates with the Infrared Portable Printer — see “Spares and accessories” on page 32 for sales part number. Ensure the Elcometer 345 is within 500 mm of the printer, and that the Elcometer 345 IR LED is aligned with the IR detector on the printer. PRINTING To print the statistical values, press >>>>…

  • Page 26: Storage

    The gauge does not contain any user-serviceable components. In the unlikely event of a fault, the gauge should be returned to your local Elcometer supplier or directly to Elcometer. The warranty will be invalidated if the instrument has been opened.

  • Page 27: Error Codes

    Gauge over-range reading. Memory hardware fault. Contact Elcometer or your local Elcometer supplier. Gauge is not configured after repair. Contact Elcometer or your local Elcometer supplier. 8 Err The printer is off line, out of paper or printer or cable fault. Correct printer fault or press…

  • Page 28: Technical Specification

    11 TECHNICAL SPECIFICATION Range: 0 µm to 1500 µm (0 mil to 60 mil) Accuracy: ± 3% of the reading or ± 2.5 µm (0.1 mil) whichever is the greater. ± 1% of the reading or ± 2.5 µm (0.1 mil) whichever is the greater when calibrated close to the thickness to be measured.

  • Page 29
    Battery Life: 30 to 40 hours continuous with alkaline dry batteries (15 000 to 20 000 readings at an average of 8 readings per minute). Battery life is reduced when using the back light. Minimum Substrate Thickness: 300 µm (12 mil) unless special adjustment is made — see section 5.3. Gauges will have reduced range when adjusted for thin substrates.
  • Page 30: Probes

    12 PROBES Gauges with serial number TA—- onward are fitted with a new design of probes. The part numbers for replacements for separate probes are shown in Table 1 on page 31. Right Angle Probe Telescopic Probe…

  • Page 31
    Table 1: Probe part numbers Replacement Original Current Scale Range Resolution Accuracy probe no. probe no. probe types 0 — 1500 µm ±3% or ±2.5 µm 1 µm F1 Standard T3457286- T34515844 ±1% or ±2.5 µm Below 20 µm, 0.1 µm F1 Right Angle T3457650- T34515845…
  • Page 32: Spares And Accessories

    13 SPARES AND ACCESSORIES The following spare parts and optional accessories are available from your local Elcometer supplier or direct from Elcometer. 13.1 CALIBRATION FOIL SETS Set of µm mils Part number for set 8 foils 12, 25, 50, 125, 175, 250, 500, 1020 0.5, 1, 2, 5, 7, 10, 20, 40 T9904199F…

  • Page 33
    13.4 COATED THICKNESS STANDARDS INCLUDING CERTIFICATE Ferrous Standard (4 Values): T995111261 13.5 INFRARED PORTABLE PRINTER Infrared Portable Printer: T99913877 Thermal Printer Paper: T99913878 13.6 PROBE ADAPTORS Jumbo Hand Grip: T9997766- V Adaptor for pipes: T9997381-…
  • Page 34
    13.7 HOLSTER Integral Probe Version: T9999516- Separate Probe Version: T3459515- 13.8 RAINCOAT Protective gauge covers (pack of 5): T99913537 13.9 WRIST HARNESS Wrist harness: T99916063…
  • Page 35: Related Equipment

    14 RELATED EQUIPMENT Elcometer produces a wide range of coating thickness gauges and associated paint inspection equipment. Users of the Elcometer 345 may also benefit from the following Elcometer products: • Elcometer 456 Coating Thickness Gauge • Elcometer 406L Statistical Mini-Glossmeter For further information contact Elcometer, your Elcometer supplier or visit the Elcometer website, www.elcometer.com…

  • Page 36
    15 INDEX Backlight Error codes Outputting data Batteries Fitting Life Factory calibration Print Data Output Low voltage Foils Printer Type Ordering Infrared Blast cleaned surfaces Printing Probe Connecting Gauge features Calibration adjustment Probes Error Adapters for Rough and Special Substrates Wear of Hand grip, Jumbo Protection for gauge…
  • Page 37
    Clearing 13, 22 Displaying List of 13, 23 Turning off Storage Substrate thickness Minimum Switching on/off Technical specification Test certificates Thin substrates Units Changing V Adaptor, for pipes…

Page 1: Elcometer 345 Manual

Eng

lish

Elcometer 345

Coating Thickness Gauge

Operating Instructions

Page 2: Elcometer 345 Manual

Eng

lish

R

Equipment described in these instructions is covere d by the following Patents:FNF UK Patent No: GB2306009B and FNF US Patent No: 5886522Simple Statistics Patent Applications: UK No. 9903992.7, GmbH No. 19945836.7

This product meets the Electromagnetic Compatibility Directive.The product is Class B, Group 1 ISM equipment according to CISPR 11

Group 1 ISM product: A product in which there is intentionally generated and/or used conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.Class B product are suitable for use in domestic establishments and in establishments directly connected to a lowvoltage power supply network which supplies buildings used for domestic purposes.

and are registered trademarks of Elcometer Limited.All other trademarks acknowledged.© Copyright Elcometer Limited. 2009.All rights reserved. No part of this Document may be reproduced, transmitted, transcribed, stored (in a retrieval systemor otherwise) or translated into any language, in any form or by any means (electronic, mechanical, magnetic, optical,manual or otherwise) without the prior written permission of Elcometer Limited.A copy of this Instruction Manual is available for download on our Website via www.elcometer.com.

Doc No. TMA-0431 Issue 03Text with Cover No. 20607

Page 3: Elcometer 345 Manual

R

3

CONTENTS

Section Page

1 About your gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Gauge features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.3 What the box contains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82.1 Fitting batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82.2 Low battery voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82.3 Fitting the Harness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.4 Connecting the Separate Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.5 Switching the gauge on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.6 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.7 Changing units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.8 Key-Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.9 Backlight Operation: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112.10 Simple Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112.11 Delete Last Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3 Quick Operation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

4 Taking a measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Page 4: Elcometer 345 Manual

R

4

5 Calibration adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165.1 Calibration Adjustment Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165.2 Calibration Adjustment for Smooth Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175.3 Special Calibration Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185.4 The Correction Value Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195.5 To Recall The Factory Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

6 Statistical summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216.1 Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216.2 Displaying Statistical Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216.3 Clearing Statistical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226.4 Setting The User List For Statistical Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236.5 Example of Statistical User List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

7 Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

8 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

9 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

10 Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

11 Technical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

12 Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

13 Spares and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

14 Related equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Page 5: Elcometer 345 Manual

R

5

Thank you for your purchase of this Elcometer 345 Coating Thickness Gauge. Welcome to Elcometer.

Elcometer are world leaders in the design, manufacture and supply of inspection equipment for coatingsand concrete. Our products cover all aspects of coating inspection, from development through applicationto post application inspection.

The Elcometer 345 Coating Thickness Gauge is a world beating product. With the purchase of this gaugeyou now have access to the worldwide service and support network of Elcometer. For more information visitour website: www.elcometer.com

1 ABOUT YOUR GAUGE

The Elcometer 345 Coating Thickness Gauge is a hand-heldcoating thickness gauge for fast and accurate measurement of thethickness of coatings on metal substrates.

The gauge is available either with a built-in integral probe or with aseparate probe.

The gauge can be switched to display metric (µm/mm) or imperial(mils) units of measurement.

Your gauge was calibrated at the factory, however you can adjustthe calibration using the set of calibration foils supplied with thegauge.

The Elcometer 345 Coating Thickness Gauge is powered byinternal dry cell batteries.

R R

Elcometer345 withIntegral

Probe

Elcometer 345 withSeparate Probe

Page 6: Elcometer 345 Manual

R

6

1.1 GAUGE FEATURES

• User adjustable calibration

• Switchable between metric and imperial units

• Bleep (when valid reading taken)

• Infrared data output (to mini-printer)

• Simple Statistics

1.2 STANDARDSThe Elcometer 345 can be used in accordance with the following National andInternational Standards, ASTM B 499, ASTM D 1186-B, ASTM G 12, BS 3900 (C5), BS 5411 (11), DIN 50981, IMOMCS.215 (82/84), IMO MSC.244(83), ISO 1461, ISO 2063, IS0 2808-7C, ISO 2808-12, ISO 19840, NF T30-124, SSPC-

PA2 (2004).For IMO, PSPCa, the thickness coating, should be measured and recorded. Your Elcometer 345 can beused for this.

1.3 WHAT THE BOX CONTAINS

• Elcometer 345 Coating Thickness Gauge with batteries fitted

• Set of calibration foils

• Neck harness

• Carrying case

• Operating instructions

a. International Marine Organisation, Performance Standard for Protective Coatings.

IR

R

0

Page 7: Elcometer 345 Manual

R

7

The Elcometer 345 Coating Thickness Gauge is packed in a cardboard and foam package. Please ensurethat this packaging is disposed of in an environmentally sensitive manner. Consult your local EnvironmentalAuthority for further guidance.

To maximise the benefits of your new Elcometer 345 Coating Thickness Gauge please take some time to re adthese Operating Instructions. Do not hesitate to co ntact Elcometer or your Elcometer supplier if you h ave anyquestions.

Page 8: Elcometer 345 Manual

R

8

2 GETTING STARTED

Note: Before you use your gauge for the first time, remove the paper strip from the battery compartment.

2.1 FITTING BATTERIES

The Elcometer 345 may be used with dry cell batteries or rechargeable batteries. 2 x LR03 (AAA) alkalinebatteries are supplied with the gauge. To fit or replace batteries:

1. Switch off the gauge.

2. Lift off battery compartment cover at rear of gauge.

3. Place batteries in the battery clips ensuring correct polarity.

4. Replace battery compartment cover.

Note: Remove the batteries from the gauge if it is to remain unusedfor a long period of time. This will prevent damage to the gauge inthe event of malfunction of the batteries.

2.2 LOW BATTERY VOLTAGE

When the battery voltage is low the battery symbol illuminates onthe display. The gauge will still function but new batteries arerecommended.

When the battery symbol flashes and the gauge beeps every tenseconds, the batteries must be replaced.

When the batteries are exhausted, the gauge will beep three times and then switch off.

Page 9: Elcometer 345 Manual

R

9

2.3 FITTING THE HARNESS

Your gauge is supplied with a neck harness which should bethreaded through the bracket on the side of the gauge as shown.

2.4 CONNECTING THE SEPARATE PROBE

The Elcometer 345 separate probe gauge has a 4-pin socket at thebase of the gauge. Line up the 2 red marks when inserting the plug.

To disconnect a probe, pull on the outer knurled sleeve of the plug- do not twist. This will release the lock that prevents the plug frombeing pulled out of the socket accidentally.

• Do not operate the Elcometer 345 without a probe connected.

• Do not remove the probe while the instrument is switched on.

Page 10: Elcometer 345 Manual

R

10

2.5 SWITCHING THE GAUGE ON/OFF

To switch on, press .

Gauges with an integral probe will switch on automatically by placing the probe against a hard surface.

To switch off, press and hold until OFF is displayed.

The gauge will switch off automatically after one minute of inactivity.

2.6 DISPLAY

The display will show the measurement value.

In addition, a number of symbols can appear on the display:

2.7 CHANGING UNITS

To change the units of measurement displayed on the screen(µm/mm or mils):

With the gauge switched on, press . The symbol on the display will start to flash. Press . Theunits will change. To change units again, repeat these steps.

2.8 KEY-LOCK

The gauge is fitted with a ‘Key-Lock’ feature, which is a form of password, to prevent accidental operationof calibration adjustment — see “Calibration adjustment” on page 16.

Page 11: Elcometer 345 Manual

R

11

2.9 BACKLIGHT OPERATION:

Elcometer 345 gauges have a backlit display. The backlight is designed for low light operations and can beturned on and off via the keypad. Turning the backlight off will save battery power.

To toggle the backlight on/off, press and release . The symbol will flash on the display. Press andrelease once more. The will stop flashing and the backlight will be toggled on or off.

When switched on, the backlight comes on when a reading is taken and remains on for 3 seconds.

2.10 SIMPLE STATISTICS

This feature is normally switched off when the gauge is dispatched from the factory.

Your gauge can display statistical summary information in a preset sequence. The user can select any valuein any sequence from the list of five values, plus an OFF option — see “Statistical summary” on page 21.

2.11 DELETE LAST READING

This operation will only delete the last reading taken. Previous readings can not be deleted individually.

To delete the last reading from the statistical calculations when the statistics feature is switched on, pressand hold for more than 2 seconds but less than 4 seconds. The display will show dEL. Release to delete the last reading,

Note: If is held for longer than 4 seconds, dEL will change to OFF. Releasing will switch thegauge off and the last reading will not be deleted — see “Switching the gauge on/off” on page 10.Switch on and try again.

Page 12: Elcometer 345 Manual

R

12

3 QUICK OPERATION GUIDE

Switch on or

Switch off >>>> >>>>

Metric <=> English

>>>> µm or mm <=> mil

Calibration adjustment for Smooth Substrates

>>>> >>>> >>>> >>>> to set.

separategauges

R

0

4 seconds

R

0

bare metalon

4 secondsR

0

on foil

>4 seconds

oradjust to foil

Page 13: Elcometer 345 Manual

R

13

Calibration adjustment for Rough and Special Substrates

>>>> >>>> >>>> >>>>

>>>> to set >>>> >>>> >>>>

>>>> to set.

Set User List for Statistical Values

>>>> >>>> >>>> keep pressing until first statistic

wanted appears, then press to accept. Repeat for all statistics wanted. Press to set.

View Statistics >>>> etc. to view each statistic

To Clear Statistical Data

>>>> >>>> >>>> >>>> to clear, to retain

R

0

on thin foil

>4 seconds

oradjust to foil

R

0

on thick foil

>4 seconds

oradjust to foil

Page 14: Elcometer 345 Manual

R

14

To Turn Statistical Data ‘Off’

>>>> >>>> >>>> >>>> until

>>>> >>>> Stats off

Print Data Output

Align IR’s >>>> >>>>

to continue measurements

To Reset Factory Calibration

>>>> to adjust to 10 µm (0.5 mil) [or 100 µm (4.0 mil) for scales F2

and F3] >>>> >>>> >>>> to clear 1 Err and reset factory

calibration.

R

0

bare metalon

Page 15: Elcometer 345 Manual

R

15

4 TAKING A MEASUREMENT

Before starting to take measurements it is good practice to check the performance of the gauge using thecalibration adjustment procedure — see “Calibration adjustment” on page 16.

1. Hold the gauge with the probe pressedfirmlyb against the surface you aremeasuring.

2. The display will blinkc and then display thecoating thickness.

If required, this reading can be adjusted forthe conditions of the measurement — see“Calibration adjustment” on page 16.

If the following display appears

the value is just greater than the range ofthe gauge, or the gauge is out ofcalibration.

b. The reading may be inaccurate if the probe is not held flat against the surface.c. If the display does not blink then the gauge has not recognised that the probe is on the surface, usually because the

coating is much too thick for the range of the gauge.

R

R

0

Integral probe

Separate probe

F

Page 16: Elcometer 345 Manual

R

16

5 CALIBRATION ADJUSTMENT

The Elcometer 345 should be adjusted before use, or at least checked to ensure that the calibrationadjustment is correct for the conditions of use.

Calibration adjustment should be carried out with the appropriate probe, on the same type of metal, thesame curvature and similar surface finish as that of the item to be measured. It is best to use an uncoateditem that is similar to the coated item to be tested.

When the gauge is shipped from the factory it comes with a preset calibration. See “To Recall The FactoryCalibration” on page 20 for instructions on how to restore this calibration.

5.1 CALIBRATION ADJUSTMENT ERROR

If an error message 1 Err is displayed during adjustment it indicates that the adjustment has not taken placecorrectly. This may be due to incorrect substrate, incorrect values or units, etc. Press to cancel theerror message. This will automatically restore the gauge to the factory calibration setting. Check and repeatthe adjustment procedure. See “To Recall The Factory Calibration” on page 20 for a method to deliberatelyreset the factory calibration setting.

Page 17: Elcometer 345 Manual

R

17

5.2 CALIBRATION ADJUSTMENT FOR SMOOTH SURFACES

Use this calibration adjustment procedure for measurements on smooth, flat surfaces.

Note: The key-lock feature requires the keys used to access these routines to be pressed for 4 secondsbefore any response is made. This is to prevent accidental operation.

1. Place the probe on the bare metal, wait for a reading to be displayed. Lift the probe from the surface.

2. Press and hold for 4 seconds to set the display to zero. The bleeper will sound to indicate theinstruction has been accepted.

3. Place the calibration foil closest to the thickness to be measured on the baremetal surface and take a reading.

Note: If the probe is left on the bare substrate or moved quickly to the calibrationfoil the display will show a sequence of three ‘rising’ ‘-’ symbols and sound aseries of bleeps as a warning. The probe must be lifted into free air before takinga measurement on the calibration foil.

4. Press or for more than 4 seconds to adjust the display to thecalibration foil value, When close to the correct reading, a few quick presses of

or will step to the correct reading. If the statistics feature is switchedon, the display will show one of the statistical values (“Statistical summary” onpage 21). After 4 seconds C appears in the left-hand side of the display and thevalue will start to change. The bleeper will sound to indicate that the calibrationroutine has been started.

5. When the value in the display is correct, press to accept the newcalibration or wait 7 seconds for the gauge to accept the calibration automatically. The display will blinkthen re-display the reading. The bleeper will sound to indicate acceptance of the new calibration.

R

0

Page 18: Elcometer 345 Manual

R

18

6. Confirm that the gauge reads correctly at zero and at the calibration foil value. If required thecalibration adjustment can be repeated to ensure its accuracy.

To re-adjust the gauge, repeat all the steps in Section 5.2.

Note: As well as calibration foils, certified coated standards are available — see “Spares and accessories”on page 32.

5.3 SPECIAL CALIBRATION ADJUSTMENT

Use this calibration adjustment procedure for measurements on rough surfaces, small diameters, specialsteel alloys, thin substrates below 300 µm (0.012″), etc.

Two calibration foils are required; one below the thickness to be measured (the thin foil) and one above (thethick foil).

Note: In this routine if, after the key-lock feature is operated, 7 seconds elapses without pressing a key ortaking a reading, the gauge will automatically continue to the next step, i.e. the gauge will move from lowcalibration CLO to high calibration CH1 or from high calibration to measurement mode.

1. Switch the gauge on and press , the symbol in the display will flash. Press and release tobegin the special calibration sequence. The display will show CLO (= Calibrate Low) and the roughsurface icon will flash at the rate of once per second.

2. Place the chosen thin calibration foil onto the bare metal surface.

3. Repeated measurements can now be taken on this foil and the display will show the accumulatedaverage of these readings. This is useful when there is significant variation from one reading to thenext, such as would occur on a rough surface or profile. Averaging a number of readings will give amore stable value and it is this average that should be adjusted for calibration.

Page 19: Elcometer 345 Manual

R

19

4. Press and hold or for more than 5 seconds to adjust this average to the foil value.

5. Press to accept the low calibration or wait 7 seconds to accept the calibration automatically.

The display will now show CH1 (= Calibrate High).

6. Place the chosen high calibration foil on the bare metal surface and take a number of readings until astable average is displayed.

7. Press and hold or for more than 4 seconds to adjust this average to the foil value.

8. Press to accept the calibration or wait 7 seconds for the gauge to accept the calibrationautomatically. The display will blink and then re-display the reading. The bleeper will sound 3 times toindicate that this routine is complete. will stop flashing and will remain visible on the display.

9. Confirm that the average gauge reading on each calibration foil is correct. An intermediate foilthickness can also be used to confirm the performance over the calibrated range.

The calibration values are stored in the gauge until new values are entered, even when the gauge isswitched off. on the display indicates the special calibration on switch on.

To re-calibrate, repeat the procedure. Re-calibrating using the smooth surface calibration method (see“Calibration Adjustment for Smooth Surfaces” on page 17) to set zero will remove from the display.

5.4 THE CORRECTION VALUE METHOD

The smooth surface adjustment method (see “Calibration Adjustment for Smooth Surfaces” on page 17)can be used for blast cleaned surfaces if a correction value is applied to each reading taken on theroughened surface.

Page 20: Elcometer 345 Manual

R

20

When coating thickness gauges are adjusted on a smooth surface, readings on a blast cleaned or otherwiseroughened surface will be high and the correction value must be subtracted from the reading to obtain thethickness of the coating,

The correction value is determined according to the roughness of the surface per ISO 8501, the SurfaceRoughness Comparator Elcometer 125. If the roughness of the surface under the coating is unknown adefault correction value of 25 µm (1.0 mil) is subtracted from each reading. For further details see EN19840’Paints and Varnishes — Corrosion protection of steel structures by protective paint systems — Measurementand acceptance criteria of the dry film thickness”.

5.5 TO RECALL THE FACTORY CALIBRATION

Set zero in the normal way. Then take a reading on the same bare substrate and adjust the gauge to read10 µm (0.5 mil). This will cause the gauge to display 1 Err. The factory setting will be recalled automaticallywhen is pressed to clear the error message.

Page 21: Elcometer 345 Manual

R

21

6 STATISTICAL SUMMARY

6.1 STATISTICS

The statistics display on the Elcometer 345 shows a preset combination of the five statistical functions whichmay be shown on the display in any user defined order by pressing one key. The feature is switched on byselecting at least one statistical value to be displayed — see “Setting The User List For Statistical Values” onpage 23. The feature is switched off by selecting OFF. The menu from which the list is created is as follows:

• Number of Readings taken

• Mean value of these Readings

• Standard deviation of these Readings

• Highest Reading taken

• Lowest Reading

• OFF (Switches the statistics off.)

6.2 DISPLAYING STATISTICAL VALUES

• To display a statistical value from the normal measurement mode, press and release or .This will display for 3 seconds either (+) the first statistical value in the user list or (-) the last statisticalvalue in the user list.

• If or is released within 4 seconds the gauge will display the value for a further 3 secondsduring which time or can be pressed again. If the key is held for more than 5 seconds thecalibration routine is activated.

Page 22: Elcometer 345 Manual

R

22

If is pressed the gauge displays the next statistical value on the user list for a minimum of 3seconds.

If is pressed the gauge displays the previous statistical value on the user list for a minimum of 3seconds.

• To display the rest of the list of values or must be pressed while the current value is on thedisplay.

• To exit the list press .

STATISTICAL SYMBOLS

= Number of readings

= Mean (Average)

= Standard Deviation

H = Highest reading

L = Lowest reading

6.3 CLEARING STATISTICAL DATA.

• To clear the statistical data when starting a new set of data press >>>> >>>> . Thedisplay will show CLs to indicate “Clear Statistics?”.

• Press to clear statistics memory or press to exit without clearing the statistics memory.

Note: 4 Err is displayed when the memory for statistical values is full.

Page 23: Elcometer 345 Manual

R

23

6.4 SETTING THE USER LIST FOR STATISTICAL VALUES

• To create a user list press >>>> >>>> . The display will show CLS to indicate ‘ClearStatistics?’

• Press to enter the menu for setting the user list of statistical values. The gauge displays thesymbol of the first Statistics item on the statistics menu with its numerical position in the user list shownin the top left of the display. If the numeric display is blank then the statistic is not selected in thecurrent user list. (To exit from the setting menu and retain the current user list press . This returnsthe gauge to normal measurement mode.)

DISPLAY SYMBOLS

= Number of readings

= Average (Mean)

= Standard Deviation

= Highest reading

= Lowest reading

= Turn statistics off

F

F

F

F

LoF

Of fF

Page 24: Elcometer 345 Manual

R

24

• To enter the statistical value displayed at the next available position on the list press . If this is thefirst entry then the current user list is deleted and replaced with a new user list and it appears in thetop left of the display. This action also serves to switch the statistics feature on.

• To view the next statistical value on the user list press . Repeat presses of will scroll roundthe menu. Press to add the statistic value to the list. Press to exit menu or exit automaticallywhen all 5 items have been entered.

Selecting OFF and pressing will switch the statistics feature off.

6.5 EXAMPLE OF STATISTICAL USER LIST

To set statistics for Average, Lowest, Highest and Number of readings, in that order:

Press >>>> >>>>

Display will show CLS.

Keep pressing until is displayed, press to accept

Keep pressing until is displayed, press to accept

Keep pressing until is displayed, press to accept

Keep pressing until is displayed, press to accept

Press . The statistics are now set in this order.

F1

LoF2

F3

F4

Page 25: Elcometer 345 Manual

R

25

7 DATA OUTPUT

Your gauge is fitted with an infrared data output feature. This operates with the Infrared Portable Printer -see “Spares and accessories” on page 32 for sales part number. Ensure the Elcometer 345 is within500 mm of the printer, and that the Elcometer 345 IR LED is aligned with the IR detector on the printer.

PRINTING

To print the statistical values, press >>>> >>>> . The display will show PrF.

To print these values press >>>> again.

Press to return the gauge to measurement mode.

Page 26: Elcometer 345 Manual

R

26

8 STORAGE

This gauge incorporates a Liquid Crystal Display (LCD). If the display is heated above 50°C (120°F)it may be damaged. This can happen if the gauge is left in a car parked in strong sunlight.

Always store the gauge in its carrying pouch when it is not being used.

9 MAINTENANCE

You own one of the finest hand-held coating thickness gauges in the world. If looked after, it will last alifetime.

The gauge does not contain any user-serviceable components. In the unlikely event of a fault, the gaugeshould be returned to your local Elcometer supplier or directly to Elcometer. The warranty will be invalidatedif the instrument has been opened.

Regular calibration checks are a requirement of quality management procedures e.g. ISO 9000 and otherstandards. For gauge checks and certification contact Elcometer or your local Elcometer supplier.

Details of Elcometer offices around the world are given on the outside cover of these Operating Instructions.Alternatively visit the Elcometer website, www.elcometer.com

PROBES

The probe will wear with repeated use. Probe life depends on the number of measurements taken and themanner in which readings are taken. To extend probe life, always set the probe down so that it isperpendicular to the panel surface. Never drag the probe along the surface, as doing so will drasticallyreduce the life of the probe.

Replacement of integral probes must be carried out by Elcometer. Separate probes can be fitted by theuser.

Page 27: Elcometer 345 Manual

R

27

10 ERROR CODES

Press to clear error codes.

Error code Explanation

0 Err Unable to obtain a stable reading. This is either due to a very high electrical noise environment, beyond emc specification limits, or could indicate a gauge fault.

1 Err Calibration factors are out of range. Clear the error message and repeat the adjustment procedure.

2 Err Averaging error.

3 Err No readings taken for average calculation.

4 Err Statistics memory full.

5 Err Statistics overflow, more than 9,999 readings have been taken. Statistics data must be cleared (see Section 6.3) before further readings can be taken for statistical analysis.

6 Err Gauge over-range reading.

7 Err Memory hardware fault. Contact Elcometer or your local Elcometer supplier.

8 Err Gauge is not configured after repair. Contact Elcometer or your local Elcometer supplier.

9 Err The printer is off line, out of paper or printer or cable fault. Correct printer fault or press to cancel print function.

Page 28: Elcometer 345 Manual

R

28

11 TECHNICAL SPECIFICATION

Range: 0 µm to 1500 µm (0 mil to 60 mil)

Accuracy: ± 3% of the reading or ± 2.5 µm (0.1 mil)whichever is the greater.

± 1% of the reading or ± 2.5 µm (0.1 mil)whichever is the greater when calibrated close to the thickness to bemeasured.

Resolution, metric: 1 µmBelow 20 µm, 0.1 µm

Resolution, imperial: 0.1 milBelow 1 mil, 0.01 mil

Display: Liquid Crystal Display (LCD) 3½ Digits, Character height: 10 mm (0.4″)

Operating Temperature: 0°C to 50°C (32°F to 120°F)

Dimensions: 56 mm x 24 mm x 120 mm (2.2″ x 0.95″ x 4.75″)

Weight (incl. Dry Batteries): 115 gms. (4 oz).

Measurement speed: > 40 readings per minute.

Battery Type: 2 x LR03 (AAA), alkalined dry batteries or rechargeablee equivalents

Page 29: Elcometer 345 Manual

R

29

Battery Life: 30 to 40f hours continuous with alkaline dry batteries (15 000 to 20 000readings at an average of 8 readings per minute). Battery life is reducedwhen using the back light.

Minimum Substrate Thickness: 300 µm (12 mil) unless special adjustment is made — see section 5.3.Gauges will have reduced range when adjusted for thin substrates.

d. Alkaline batteries must be disposed of carefully to avoid environmental contamination. Please consult your localenvironmental authority for information on disposal in your region.Do not dispose of any batteries in fire.

e. Rechargeable batteries can be used if they are charged outside the gauge.f. Battery life is reduced to approximately 25% of the dry battery life when using rechargeable batteries. Follow the

instructions provided by the battery manufacturer when charging and disposing of rechargeable batteries.

Page 30: Elcometer 345 Manual

R

30

12 PROBES

Gauges with serial number TA—- onward are fitted with a new design of probes. The part numbers forreplacements for separate probes are shown in Table 1 on page 31.

Right Angle Probe Telescopic Probe

Page 31: Elcometer 345 Manual

R

31

Table 1: Probe part numbers

Scale Range Accuracy a ResolutionReplacement probe types b

Original probe no.

Current probe no.

F1 0 — 1500 µm ±3% or ±2.5 µm±1% or ±2.5 µmc

1 µmBelow 20 µm, 0.1 µm

F1 StandardF1 Right AngleF1 Telescopic

T3457286-T3457650-T3457651-

T34515844T34515845T34515846

0 — 60 mil ±3% or ±0.1 mil±1% or ±0.1 milc

0.1 milBelow 1 mil, 0.01 mil

a. Whichever is the greaterb. Separate probe versions onlyc. Whichever is the greater when calibrated close to the thickness to be measured.

Table 2: Probe capabilities

Probe TypeMinimum

Convex Surface Diameter

Minimum Concave

Surface RadiusHeadroom

Minimum Sample

DiameterCal Foil Value a

Separate

F1 4 mm (0.16″) 25 mm (0.98″) 85 mm (3.35″) 4 mm (0.16″) 250 µm (10 mil)

F1 Right Angle 4 mm (0.16″) 25 mm (0.98″) 25 mm (0.98″) 4 mm (0.16″) 250 µm (10 mil)

F1 Telescopic 4 mm (0.16″) 25 mm (0.98″) 30 mm (1.18″) 4 mm (0.16″) 250 µm (10 mil)

Integral F1 4 mm (0.16″) 25 mm (0.98″) 122 mm (4.8″) 4 mm (0.16″) 250 µm (10 mil)

a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurementconditions.

Page 32: Elcometer 345 Manual

R

32

13 SPARES AND ACCESSORIES

The following spare parts and optional accessories are available from your local Elcometer supplier or directfrom Elcometer.

13.1 CALIBRATION FOIL SETS

Individual calibration foils and customised sets are also available. Consult your local Elcometer supplier.

13.2 CALIBRATION CERTIFICATES FOR CALIBRATION FOILS

Certificates traceable to National Standards including UKAS and NIST are available on request.

13.3 TEST CERTIFICATES

A certificate with results of a standard test on known calibration foil values over the full range of the probe.Order using sales part number TEST-345.

Set of µm mils Part number for set

8 foils 12, 25, 50, 125, 175, 250, 500, 1020 0.5, 1, 2, 5, 7, 10, 20, 40 T9904199F

3 foils 25, 250, 1020 1, 10, 40 T9904199G

Page 33: Elcometer 345 Manual

R

33

13.4 COATED THICKNESS STANDARDS INCLUDING CERTIFICA TE

13.5 INFRARED PORTABLE PRINTER

13.6 PROBE ADAPTORS

Ferrous Standard (4 Values): T995111261

Infrared Portable Printer: T99913877

Thermal Printer Paper: T99913878

Jumbo Hand Grip: T9997766-

V Adaptor for pipes: T9997381-

Page 34: Elcometer 345 Manual

R

34

13.7 HOLSTER

13.8 RAINCOAT

13.9 WRIST HARNESS

Integral Probe Version: T9999516-

Separate Probe Version: T3459515-

Protective gauge covers (pack of 5): T99913537

Wrist harness: T99916063

Page 35: Elcometer 345 Manual

R

35

14 RELATED EQUIPMENT

Elcometer produces a wide range of coating thickness gauges and associated paint inspection equipment.

Users of the Elcometer 345 may also benefit from the following Elcometer products:

• Elcometer 456 Coating Thickness Gauge

• Elcometer 406L Statistical Mini-Glossmeter

For further information contact Elcometer, your Elcometer supplier or visit the Elcometer website,www.elcometer.com

Page 36: Elcometer 345 Manual

R

36

15 INDEX

BBacklight 11Batteries

Fitting 8Life 29Low voltage 8Type 28

Blast cleaned surfaces 19

CCalibration adjustment 16

Error 16for Rough and Special Substrates 13Procedure 12Rough surfaces 18Small diameters 18Smooth surfaces 17Special steel alloys 18Thin substrates 18

Calibration certificates 32Coated thickness standards

Spares 33Contents of box 6Correction value 19

DData output 25Delete Last Reading 11Display symbols 10

EError codes 27

FFactory calibration 20Foils

Ordering 32

GGauge features 6

HHand grip, Jumbo 33Harness

Fitting 9Wrist (spare part) 34

Holster (belt) 34

IImperial units 10Infrared printer 33

KKey-Lock 10

MMaintenance 26Measurement 15Metric units 10

OOutputting data 25

PPrint Data Output 14Printer

Infrared 33Printing 25Probe

Connecting 9Probes 30

Adapters 33Wear of 26

Protection for gauge 34

QQuick Operation Guide 12

RRange

Warning 15Reset 14

SService 26Shims

Ordering 32Spare parts 32Standards 6

Coated thickness 33Statistics 11, 21

Page 37: Elcometer 345 Manual

R

37

Clearing 13, 22Displaying 21List of 13, 23Turning off 14

Storage 26Substrate thickness

Minimum 29Switching on/off 10

TTechnical specification 28Test certificates 32Thin substrates 18

UUnits

Changing 10

VV Adaptor, for pipes 33

Описание
345 снят с производства! Замена Толщиномер Elcometer 456

Толщиномер покрытий Elcometer 345 выпускается в небольшом и износоустойчивом корпусе и снабжен ЖК-дисплеем с легко читаемыми цифрами. Это позволяет использовать данный прибор в многочисленных областях.

Особенности

  • Прибор прост и удобен в использовании и имеет функцию блокировки клавиш во избежание случайного нарушения калибровки
  • Двухточечная калибровка обеспечивает высокую точность
  • Прибор имеет жесткую конструкцию и поставляется со специальным износоустойчивым датчиком
  • Имеется возможность выполнения измерений на плоских, изогнутых и круглых подложках
  • Прибор может переключаться между отображением показаний в мкм, мм и mils/thou
  • Имеются варианты прибора со встроенным (I) и внешним датчиком (S)
  • Прибор отображает простые статистические данные (среднюю величину, стандартное отклонение, количество показаний, наибольшее и наименьшее показание)
  • Точность измерений составляет ±1-3% по полному диапазону; ±1% при калибровке близкой к измеряемому показанию (по двум точкам)
  • Высокая скорость получения показаний: >40 показаний в минуту
  • Вывод данных через ИК-порт для печати на принтере HP
  • Датчики выпускаются для измерения толщины покрытий методом электромагнитной индукции и с помощью вихревых токов

* ±1% при калибровке близкой к толщине измеряемого покрытия; ±3% при калибровке по полному рабочему диапазону толщины.

** Патент № GB 2306009В; патент США 5886522.

Используется в соответствии с

  • Датчики Типа F:
    ASTMB 499, BS 5411-11, BS 3900-C5-6Aa, BSENISO 1461, DIN 50981, ISO 2178, ISO 2178-6Aa, prEnISO 19840
  • Датчики Типа N:
    ASTMD 1400, ASTMB 244, BS 5411-3, BS 3900-C5-6Ba, BS5599, DIN 50984, ISO 2360, ISO 2808-6Ba
  • Датчики Типа FNF:
    Все перечисленные для F и N типа, а также ASTM E 376

Таблица для приборов измерения толщины покрытий

  • Комбинация покрытие-подложка
  • Выбор правильного датчика для измерения покрытия

Таблица показывает типичные комбинации покрытие/подложка.

Покрытие

Подложка

Al

Латунь

Бронза

Cu

Сталь

Mg

Нерж. сталь

Ti

U

Zn

Алюминий

F

Анодированное

NF NF

Латунь

F

Бронза

F

Кадмий

F

Керамика

F

Хром (твердый)

NF* NF* F

Медь

F

Eloxal

NF

Эпоксидное

NF NF NF NF F NF NF NF

Гальваническое

F

Лакированное

NF NF NF NF F NF NF

Напыление металлом

F

Дисульфид

F NF

Восстановленный никель (Electroless Nickel)

NF* NF* NF* F*

Краска

NF NF NF NF F NF NF NF NF NF

Пластик

NF NF NF NF F NF NF NF NF NF

Металлизация

F

Резина

NF F NF

Резистивные (Resist)

NF

Олово

F

Лак

NF NF NF NF F

Цинк

F

Примечания

NF — Используется датчик для немагнитных подложек

F — Используется датчик для магнитных подложек

* — Необходима калибровка по образцам

ELCOMETER 300, ELCOMETER 345, ELCOMETER 355 , tolshhinomer-pokryitij-elcometer-345
, заказать и узнать цену Толщиномер elcometer 345


Технические характеристики

Описание

Диапазон

Разрешение

Точность*

Код для заказа

Elcometer 345 для черных металлов — встроенный датчик

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 FB-I1

Elcometer 345 для черных металлов — внешний датчик

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 FB-S1

Elcometer 345 для цветных металлов — встроенный датчик

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 NB-I1

Elcometer 345 для цветных металлов — внешний датчик

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 FNBF-S1

Elcometer 345 для цветных и черных металлов — встроенный датчик**

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 FNFB-I1

Elcometer 345 для цветных и черных металлов — внешний датчик**

0-1500 мкм

0,1 мкм до 20 мкм
1 мкм с 20 мкм

±1-3 % или ±2,5 мкм

A345 FNFB-S1

Elcometer 345 для черных металлов — встроенный датчик

0-5 мм

1 мкм до 1 мм
10 мкм с 1-5 мм

±1-3 % или ±0,02 мм

A345 FB-I2

Elcometer 345 для черных металлов — внешним датчик

0-5 мм

1 мкм до 1 мм
10 мкм с 1-5 мм

±1-3 % или ±0,02 мм

A345 FB-S2

Elcometer 345 для черных металлов — внешним датчик

0-13 мм

1 мкм до 2 мм
10 мкм с 2-13 мм

±1-3 % или ±0,05 мм

A345 FB-S3

  • Home
  • Brands
  • Elcometer Manuals
  • Measuring Instruments
  • Elcometer 345

1 docs – User Manuals, Help Guides and Specs – for the Elcometer 345 product are present in our data base.

  • Elcometer 345 Operating Instructions Manual (11 pages)

    Elcometer 345 Operating Instructions Manual (11 pages)

    Free Measuring Instruments Manuals: Elcometer 345

    Pages: 11 | Size:

Related Products:

  • Elcometer 319 S

  • Elcometer Protovale 331

  • Elcometer 2290

  • Elcometer 456 Model E

  • Elcometer 331 Series

  • Elcometer 106

  • Elcometer 207

  • Elcometer 205

  • Elcometer CG70ABDL

  • Elcometer MTG2

Product Types:

  • Elcometer Monitor

  • Elcometer Medical Equipment

  • Elcometer Washer

  • Elcometer Security Sensors

  • Elcometer Kitchen Appliances

  • Elcometer Music Mixer

  • Elcometer Test Equipment

  • Elcometer Measuring Instruments

  • Elcometer Power Tool

  • Elcometer Voice Recorder

  • Elcometer Flashlight

  • Elcometer Tools

  • Elcometer Microscope

  • Elcometer Thermometer

  • Elcometer Metal Detector

  • Elcometer Analytical Instruments

  • Elcometer Data Loggers

  • Elcometer Industrial Equipment

  • Elcometer Cleaning Equipment

Tips for Finding Manuals:

This web-page provides a list of 1 accessible operating manuals and information books describing Elcometer 345.
All manuals and instructions for Elcometer 345 are introduced in an easy-to-use PDF format and may be gratuitously downloaded or looked through directly from the site.

The page offers the following types of manuals: Measuring Instruments.

Helpful hints: While selecting a necessary guide for Elcometer 345 one should pay special attention to the type of the document.
We try to supply you with the fullest possible set of papers we or our users are able to find. These may be overviews and specifications of the device, mounting and installing instructions, the unit operating rules and maintenance regulations and much more.

Haven’t found a required manual for your Elcometer 345?

Check in a while. We update our guides collection and add new documents on a daily basis for you to be always able to find the very paper you need on our web-site. In case you own a directory or an instruction for Elcometer 345, which is absent on our site, and you’d like to share it with the public, please send it to us as a scanned copy or a PDF file, and we’ll definitely place it on our page while providing your name as a supplier of the doc. Lots of our users will be grateful for your assistance!

File Specifications:

2282/2282592-345.pdf file (17 Feb 2023)

Accompanying Data:

Elcometer 345 Measuring Instruments, Music Mixer PDF Operating Instructions Manual (Updated: Friday 17th of February 2023 02:53:13 PM)

Rating: 4.6 (rated by 91 users)

Compatible devices: 2080, UG20DL, 1615, CG70ABDL, 2300 Series, 510, SG80BDL, 456 3.

Recommended Documentation:

Operating Instructions Manual (Text Version):

(Ocr-Read Summary of Contents of some pages of the Elcometer 345 Document (Main Content), UPD: 17 February 2023)

  • 2, English R The Elcometer Portable Miniprinter has been tested in accordance with EEC regulations governing Electro-magnetic compliance and it meets the required directives. is a registered trademark of Elcometer Instruments Ltd. All other trademarks acknowledged. © Copyright Elcometer Instruments Ltd. 1992-2008. All rights reserved. No part of this Document may be reproduced, transmitted, transcribed, stored (in a retrieval system or otherwise) or…

  • 3, R 1 CONTENTS Section Page 1 About this printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 2 The power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3 Switching the printer on and off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 4 Connecti…

  • 4, R 2 Thank you for your purchase of this Elcometer Portable Miniprinter. Welcome to Elcometer. Elcometer are world leaders in the design, manufacture and supply of coatings inspection equipment. Our products cover all aspects of coating inspection, from development through application to post application inspection. The Elcometer Portable Miniprinter is a world beating product. With the purchase of this instrument you now have access to the …

  • 5, Elcometer 345 R 3 To maximise the benefits of your new Elcometer Portable Miniprinter please take some time to read these Operating Instructions. Do not hesitate to contact Elcometer or your Elcometer supplier if you have any questions. 2 THE POWER SUPPLY Your instrument is powered by a rechargeable battery pack. Use only the battery charger supplied — use of any other battery charger may cause damage to the printer and will invalidate the warranty. Your Elcometer Portable Miniprinter is …

  • 6, R 4 4 CONNECTING TO YOUR GAUGE Connect your gauge to your printer via the 25 way D-type socket at the rear of the printer. Plug the Elcometer printer/gauge connection cable supplied into the 25 way D-type socket on the printer and the socket on the gauge (3 way for 345 models, 25 way ‘D’ for 355 models, 5 way for 224 and 456 Models) before switching on either the printer or gaug…

  • 7, R 5 5 LOADING PAPER 1. Switch on the printer, open the lid and remove the old paper still left in the mechanism by pressing the paper feed button — do not pull the paper out of the back of the mechanism. 2. Pull off a few centimetres of paper from the new paper roll and check that the end of the paper is square. If the paper is not square, then cut the end to ensure that it is. 3. Place the paper roll in the well within the printer, with the …

  • 8, R 6 6 CHANGING THE RIBBON 1. Switch off the printer and then open the lid. 2. To remove the old ribbon cassette, push down at the right-hand end, unclip from the left-hand end and then lift the cassette out of the printer. 3. Take the new ribbon cassette and check that the ribbon is taut by turning the small wheel in a clockwise direction. Clip the cassette into position, ensuring that the pa…

  • 9, R 7 7 CHANGING THE BAUD RATE If you need to change the baud rate of the printer (for example, when using the printer with a different type of gauge) refer to the table shown on page 4. To change the baud rate setting: 1. Hold down the Set Up button and at the same time press the On button to switch the printer on. The printer will print out the current settings. 2. Each time the Set Up button is pressed, another parameter will be printed. 3. Whe…

  • 10, Elcometer 345 R 8 8 DATA CONNECTIONS 9 MAINTENANCE The printer does not contain any user-serviceable components. In the unlikely event of a fault, the instrument should be returned to your Elcometer supplier or directly to Elcometer. Contact details can be found on the outside cover of these operating instructions and at www.elcometer.com 10 SPARE PARTS The following spare parts are available from Elcometer, or your Elcometer sup…

  • 11, R 9 11 RELATED EQUIPMENT In addition to the Elcometer Portable Miniprinter, Elcometer produces a wide range of other equipment for testing, measuring and recording the characteristics of coatings. Users of the Elcometer Portable Miniprinter may also benefit from the following Elcometer product: • Elcometer Infrared Printer • ElcoMaster Software For further information contact Elcometer, your Elcometer supplier or visit www.elcometer.com

  • Elcometer 345 User Manual

  • Elcometer 345 User Guide

  • Elcometer 345 PDF Manual

  • Elcometer 345 Owner’s Manuals

Recommended: F5D7130CE1, EasyCoder PF2i, TruFlat 20F671T, SPORTTRAINER SD-800E

Links & Tools

Operating Impressions, Questions and Answers:

Понравилась статья? Поделить с друзьями:
  • Bronchoforton salbe инструкция по применению на русском
  • Руководство по эксплуатации бмс воркер
  • Нитроглицерин таблетки показания к применению инструкция таблетки
  • Дулаглутид инструкция по применению цена отзывы аналоги
  • Диециклен инструкция по применению при эндометриозе