Аппарат ивл puritan bennett 560 инструкция

Covidien Puritan Bennett 560 User Manual

  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

User’s Manual

Puritan Bennett

TM

560 Ventilator

loading

Related Manuals for Covidien Puritan Bennett 560

Summary of Contents for Covidien Puritan Bennett 560

  • Page 1
    User’s Manual Puritan Bennett 560 Ventilator…
  • Page 2
    To obtain information about a warranty, if any, contact Covidien Technical Services at 1.800.635.5267 or your local representative. Purchase of this instrument confers no express or implied license under any Covidien patent to use the instrument with any ventilator that is not manufactured or licensed by Covidien.
  • Page 3: Table Of Contents

    Contents Purpose of This Manual ……….Qualification of Personnel .

  • Page 4
    Contents Oxygen ……….. . 4–13 4.8.1 Administering Oxygen .
  • Page 5
    Contents B Modes of Ventilation ……….B–1 Modes of Ventilation .
  • Page 6
    This page intentionally blank. User’s Manual…
  • Page 7
    ECTION Figures ……. Figure 1-1. Locations of Labels – Top-Front View 1–16 .
  • Page 8
    Figures ……… . Figure 5-10.
  • Page 9
    ECTION Tables ……….Table 1-1.
  • Page 10
    This page is intentionally blank. viii User’s Manual…
  • Page 11: Purpose Of This Manual

    Information regarding your product warranty is available from your sales representative or Covidien. Extended Service The Puritan Bennett™ 560 Ventilator offers extended service contracts/warranties for purchase when the ventilator is purchased. Please contact your local Covidien Sales or Service Representative for additional information. For online technical support, visit the SolvIT…

  • Page 12: Technical Support

    Preface Technical Support Technical Service Contacts: Covidien Argentina Covidien Australia Covidien Austria GmbH Aguero 351 52A Huntingwood Drive Campus21 Capital Federal — 1171 ABC, Huntingwood, NSW 2148 Europaring F09402 Argentina Australia Brunn am Gebrige Tel: (5411) 4863-5300 Telephone (+61) 1800 350702 A-2345 Österreich…

  • Page 13
    Tel +34 93 475 86 69 Fax + 46 858 56 05 29 Fax +27 115 429 547 Fax +34 93 373 87 10 Covidien Switzerland Covidien UK & Ireland Covidien Singapore Roosstr. 53 Unit 2, Talisman Business Park Singapore Regional Service…
  • Page 14
    This page is intentionally blank. User’s Manual…
  • Page 15: Safety Information

    1 Safety Information Definitions This manual uses three indicators to highlight critical information: Warning, Caution, and Note. They are defined as follows: Warning Indicates a condition that can endanger the patient or the ventilator operator. Warning Caution Indicates a condition that can damage the equipment. Note: Indicates points of particular emphasis, that make operation of the ventilator more efficient or convenient.

  • Page 16
    If the ventilator fails the alarm tests or if you cannot complete the tests, refer to chapter “Troubleshooting” or call your equipment supplier or Covidien. • When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator.
  • Page 17
    41 °C (106 °F). This may lead to undesirable side effects for the patient. To avoid injury to the patient move the patient and the ventilator to a cooler location. For more information, contact Covidien. •…
  • Page 18
    Safety Information • The operator should connect the ventilator to an AC power source whenever available, for safer operation. • The maximum recommended shelf life of the internal battery is two (2) years. Do not use a battery that has been stored for two years prior to its first use. •…
  • Page 19
    Safety Information • The exhalation block is intended for single use by a single patient . It may periodically be cleaned, but it cannot be disinfected or sterilised. To maintain good measurement quality when used continuously, clean the exhalation block periodically (refer to section 7.3, “Cleaning the Exhalation Block”).
  • Page 20
    For more information, contact Covidien or your equipment supplier. • To connect the ventilator to a Nurse Call device, contact Covidien or your equipment supplier to check the ventilator’s compatibility with the Nurse Call device and order a suitable connection cable.
  • Page 21
    • To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only personnel authorised and qualified by Covidien should attempt to service or make authorised modifications to the Puritan Bennett™ 560 Ventilator.
  • Page 22
    Do not attempt to open, repair or otherwise service the ventilator yourself. Doing so might endanger the patient, damage the ventilator, and/or void your warranty. Only personnel authorised and qualified by Covidien should repair, open or service the ventilator. •…
  • Page 23
    Safety Information • To connect the ventilator to an external power source, first ensure the ventilator’s I/O switch is off (O). Then, connect the desired power cable to the ventilator. Finally, connect the power cable to the external power source. •…
  • Page 24
    The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored…
  • Page 25: Symbols And Markings

    Safety Information Symbols and Markings Table 1-1. Ventilator Symbols Symbols Descriptions It is essential to read, understand, and follow these instructions before using the Puritan Bennett™ 560 Ventilator (ISO 7000-0434A). This symbol appears on the ventilator’s back panel, see Table 1-2, item 5.

  • Page 26
    Safety Information Table 1-1. Ventilator Symbols (Continued) Symbols Descriptions This symbol appears on the ventilator’s front panel ALARM CONTROL key; see Figure 2-3 on page 2-6, item 3. (See my notes, to the left and below, in red). This key is used to: cancel the audible portion of alarms for 60 seconds at a time; cancel an alarm. For more information, refer to section E, “Alarms Tests.”…
  • Page 27
    Safety Information Table 1-1. Ventilator Symbols (Continued) Symbols Descriptions Flow shape (“flow distribution shape”, inspiratory phase) parameter. These symbols appear on the ventilation mode menu screens; selectable for V A/C mode only. For more information, refer to chapter 5, “Operating Procedures”.
  • Page 28
    This symbol indicates a communications port for interfacing with a USB connector. Figure 2-2, item 11. PC connector. This symbol indicates a port that can be used by authorised Covidien product service personnel or Covidien service personnel for software maintenance. See Figure 2-2, item 10.
  • Page 29: Labels / Identification And Instruction Information

    Safety Information Labels / Identification and Instruction Information Various labels or specific markings are affixed to the ventilator that describe precautions to be taken for the correct use of the ventilator and contribute to the traceability of the product. Refer to Table 1-2 and the figures on the following pages for illustrations of these labels and markings and their locations on the…

  • Page 30: Figure 1-1. Locations Of Labels — Top-Front View

    Safety Information Table 1-2. Ventilator Labels and Markings (Continued) 13. FIO Label (Figure 1-1, Figure 1-4) Note: The item number callouts in the following figures refer to those listed in Table 1-2. Figure 1-1. Locations of Labels – Top-Front View 1-16 User’s Manual…

  • Page 31: Figure 1-2. Locations Of Labels — Front-Left View

    Safety Information Figure 1-2. Locations of Labels – Front-Left View Figure 1-3. Location of Labels and Markings – Rear View User’s Manual 1-17…

  • Page 32: Figure 1-4. Location Of Labels — Bottom View

    Safety Information Figure 1-4. Location of Labels – Bottom View 1-18 User’s Manual…

  • Page 33: Ventilator Overview

    2 Ventilator Overview Indications for Use The Device name is indicated for the continuous or intermittent mechanical ventilatory support of patients weighing at least 11 lb (5 kg) who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a doctor. It is essential to read, under- stand, and follow these instructions before using the Puritan Bennett™…

  • Page 34: Contraindications

    Ventilator Overview Target Operators The ventilator may be operated by: • Respiratory therapists • Doctors • Nurses • Homecare providers • Patient and patient’s families Warning This ventilator must be used only under the responsibility and on the prescription of a doctor. Contraindications This ventilator is not for use with anesthetic gases, and is not intended for use as an emergency transport ven- tilator.

  • Page 35: Device Classification

    Ventilator Overview Warning Users must always possess an additional breathing circuit and exhalation valve while using the Device name. Device Classification The ventilator’s IEC / EN 60601-1classification is as follows: • Protection/Insulation class (electric shock): Class II • Protection index of enclosure: IP31 •…

  • Page 36: Front Panel

    Ventilator Overview Front Panel LCD Display – Displays information about the ventilator Exhalation Valve Port – Nipple for providing piloting including patient hours and software version, ventilation pressure to the exhalation valve. Controls the open-closed modes and settings, and monitored and calculated patient position of the exhalation valve.

  • Page 37: Back Panel

    Ventilator Overview Back Panel Ergonomic carrying handle. PC Cable Connector: USB mini-B connector used for Puritan Bennett™ Ventilator Test Software. On/Off (I/O) switch with protective cover: Inlet Port: Device powered on in position I; device Connects the ventilator to a low pressure oxygen switched off in position 0.

  • Page 38: Control Panel

    Ventilator Overview Control Panel Alarm indicators (two LEDs): DOWN/FREEZE key: • Red indicator: Moves the cursor down and decreases parameter values. • • During ventilation, freezes displayed waveform in the Continuous: Very High Priority (VHP) alarm Waveform menu. activated • High priority (HP) alarm activated.

  • Page 39: Ventilation Menu

    Ventilator Overview Ventilation Menu Ventilation menu with ventilation on standby. Ventilation menu during ventilation. General information line: Ventilation settings: Preferences menu access Displays the current ventilation mode, line: Displays the specific ventilation along with the following: parameter values for the Highlight this line and press •…

  • Page 40: Alarm Menu

    Ventilator Overview Alarm Menu Alarm menu with ventilation on standby. Alarm menu when not in standby. Title line: Alarm settings: Access line to Alarm Logs menu. Displays ventilation mode and the Displays the specific alarm parameter following symbols: values for the currently selected Highlight this line and ENTER ventilation mode, which are:…

  • Page 41: Waveforms Menu

    Ventilator Overview 2.10 Waveforms Menu The display of waveforms (Figure 2-6) is optional and can be selected using the Menu key The Wave- form menu is only accessible when ventilation is active. Title line: Graphic zone: Numeric zone: • Displays ventilation mode and the Displays the patient’s pressure and Displays monitored data.

  • Page 42: Usb Memory Device Menu

    Keep in mind that troubleshooting information is available in this manual to assist you in the event of a prob- lem. Refer to chapter 3, “Alarms and Troubleshooting”. If you cannot determine the cause of a problem, contact your equipment supplier or Covidien. Refer to chap- 8.3, “Service Assistance” 2-10 User’s Manual…

  • Page 43: Alarms And Troubleshooting

    3 Alarms and Troubleshooting Warning Setting Alarm limits to extreme values can cause the ventilator alarms to malfunction. When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator. Note: Many of the functions of the ventilator are not accessible when the Locking key is enabled.

  • Page 44: Alarm Level Of Priority

    Alarms and Troubleshooting Alarm Level of Priority The alarm hierarchy for signalling the level of alarm criticality is listed below. • Very High Priority (VHP): Immediate critical situation; ventilation is impossible: Continuous Sound Signaling / With or Without Continuous Red LED Illumination / With or Without Message / With or Without Display Lighting (it is possible for an alarm condition to occur that may not have both a message and lighting).

  • Page 45: Alarm Display

    Alarms and Troubleshooting Alarm Display During operation, when an alarm is activated: • One of the red or yellow alarm indicators to the left of the ALARM CONTROL illuminates and flashes. • An alarm tone sounds. • A message is displayed and flashes in reverse video at the bottom of the Ventilation Menu or Alarm Menu.

  • Page 46: Alarm Logs Menu

    Alarms and Troubleshooting Alarm Logs Menu All alarms are recorded in the internal memory of the ventilator at the time they are activated. The Alarm Logs menu is used to display the last eight (8) alarms activated, along with their date and time of activation.

  • Page 47: Silencing The Audible Portion Of Alarms

    Alarms and Troubleshooting To dismiss the Alarm Logs screen manually: Press the ENTER key when the cursor is on the “Back” line. The Alarm Logs screen is dismissed automatically: • After 15 seconds if no keyboard action is detected • When a High Priority alarm is triggered Note: Only qualified service personnel may access all alarms and events recorded by the ventilator.

  • Page 48: Figure 3-5. Manually Pausing Alarms

    Alarms and Troubleshooting Warning Alarm volume should be adjusted with respect to the ventilator’s operating environment and so that the patient’s caretakers can hear the alarms. The audible alarm vents located at the front of the device should never be obstructed. The alarm can be paused with the Alarm Pause function by pressing the ALARM CONTROL key twice once the alarm has been declared.

  • Page 49: Re-Activating Alarms

    Alarms and Troubleshooting Re-activating Alarms Alarms that have been paused and whose activation conditions continue to exist can be reactivated. To reactivate alarms, proceed as follows: 1. Press the MENU key to access the Alarm Setting menu, if this is not the menu currently displayed.

  • Page 50: Overview Of Alarms

    Alarms and Troubleshooting Overview of Alarms Note: The message: “*IF PERSISTS RESTART/SRVC” will occur only if the alarm condition continues for longer than 30 seconds. Note: Many of the functions of the ventilator are not accessible when the Locking key is enabled.

  • Page 51
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Buzzer Battery Failure. The Battery BUZZER FAULT4 Buzzer Voltage is too low. RESTART/SRVC Internal technical problem that prevents the battery sounding the POWER SUPPLY LOSS alarm.
  • Page 52
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Alarm activation occurs: CHECK Systematically after software • SETTINGS versions have changed. Loss of memorised parameters • Consequence: Locking Key disabled •…
  • Page 53
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Internal battery capacity < 10 min. EMPTY BATTERY or 3%. (battery voltage < 22.5V) Consequence: ventilation comes to a halt. Abnormally high expired flow during EXH VALVE the inspiratory phase of three…
  • Page 54
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Rate measured greater than HIGH RATE Max Rtot set during three consecutive breaths. Alarm activation occurs: After three consecutive breaths. •…
  • Page 55
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Occurs in VALVE configuration when OCCLUSION the tidal volume is measured below CHECK CIRCUIT* 20ml during three consecutive breaths for PSV, CPAP, PA/C and P SIMV modes.
  • Page 56
    Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Faulty internal pressure sensor PRES SENS FLT1 signal. RESTART/SRVC Alarm activation occurs: After 15 seconds. • Faulty proximal pressure sensor PROX SENS FLT2 signal.
  • Page 57: Troubleshooting

    To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only personnel authorised and qualified by Covidien should attempt to service or make authorised modifications to the Puritan Bennett™ 560 Ventilator.

  • Page 58
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Restart ventilator to see if alarm clears. If Battery problem that prevents it from BATTERY FAULT1 not, replace the ventilator and call your operating.
  • Page 59
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Obstruction or abnormal damage of the Clean or replace the exhalation valve and/or exhalation valve. its control tube. Remove moisture from exhalation block and valve.
  • Page 60
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Operating speed of the cooling fan not Restart ventilator to see if alarm clears. If COOLING FAN properly adjusted for the internal ambient not, replace the ventilator and call your RESTART/SRVC temperature of the device.
  • Page 61
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Note: Always consult the clinician before changing PEEP, FIO , pressure, volume or Rate settings. Check and properly connect the patient circuit connections.
  • Page 62
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Note: Ensure that you are operating the ventilator within the proper temperature range (refer to Appendix A, “Specifications”). Put the device in a warmer environment (if the ambient temperature is too low) or in a cooler environment (if the ambient…
  • Page 63
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Note: Always consult the clinician before Adjustment of Max PIP too low (only for changing PEEP, FIO , pressure, volume or V A/C and V SIMV modes).
  • Page 64
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Note: Adjustment of the Max VTI level too low Always consult the clinician before (for PSV, CPAP, P A/C, P SIMV and V SIMV changing PEEP, FIO , pressure, volume or modes).
  • Page 65
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Clean, unblock, and/or properly connect the Patient circuit obstructed. patient circuit. Check and properly connect the patient circuit connections. Leak in the patient circuit.
  • Page 66
    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) OCCLUSION CHECK CIRCUIT Clean, unblock, and/or properly connect the Patient circuit obstructed. patient circuit. *IF PERSISTS RESTART/SRVC A non-vented configuration is being used or Replace the non-vented circuit with a the built-in leak in the mask or in the circuit vented one.
  • Page 67: Additional Troubleshooting

    Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Symptom Possible Reason(s) For The Alarm Event Corrective Action(s) Restart ventilator to see if alarm clears. If Defective inspiratory flow sensor or internal VTI NOT REACHED not, replace the defective device(s) and leak of the machine.

  • Page 68
    Table 3-3. Additional Troubleshooting and Corrective Actions (Continued) Conditions Possible Causes Corrective Actions Replace the ventilator and call your Turbine noise. Light noise customer service representative. Filter and/or turbine silencer Replace the ventilator and call your deteriorated. customer service representative. Whistling noise or Replace the ventilator and call your Valve membranes damaged.
  • Page 69: Installation And Assembly

    41 °C (106 °F). This may lead to undesirable side effects for the patient. To avoid injury to the patient move the patient and the ventilator to a cooler location. For more information, contact Covidien.

  • Page 70: Connecting To External Ac Power

    The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be…

  • Page 71: Figure 4-1. The Power Cable Holder

    Installation and Assembly that is inserted into the notch (Figure 4-1, item 2) of the battery cover: AC Power Cable Holder Figure 4-1. The Power Cable Holder To secure the AC power cable: 1. Insert the power cable holder (Figure 4-2, item 1) into the notch of the battery cover.

  • Page 72: Connecting To An External Dc Power Source

    Installation and Assembly • The AC POWER indicator on the top left corner of the ventilator illuminates. • The indicator flashes while the battery charges and then turns off when the battery is fully charged. Refer to Figure 4-4 on page 4-4. If the AC power cable becomes disconnected or the AC power source fails, an “AC POWER DISCONNEC- TION“…

  • Page 73: Figure 4-5. Connecting The Ventilator To An External Dc Power Source

    Installation and Assembly Note: An alternative means of ventilation should always be available, particularly when the patient is in transit or away from wall power. While using the ventilator on external battery power it is vital that a qualified caregiver (capable of providing necessary corrective actions in the event of alarm conditions) is present.

  • Page 74: Patient Circuit

    Installation and Assembly To connect the DC power cable to the ventilator: 1. Line up the red marker dot on the ventilator’s DC power connector with the marker on the DC power cable (Figure 4-6, item 1). Align the red markers (dots and/or lines) before connecting…

  • Page 75: Choosing The Patient Circuit Type

    To ensure proper performance of the ventilator, use a patient circuit recommended by Covidien in this manual; refer to chapter 4, “Installation and Assembly”…

  • Page 76: Figure 4-7. Single Limb Patient Circuit With Exhalation Valve

    Installation and Assembly See also NOTE: Although shown here, the Figure 4-9 humidifier (item 3), water trap page 4-9. (item 4), and tubes upstream of the single limb patient circuit are not included with the ventilator. Contact your supplier for more information.

  • Page 77: Figure 4-8. Double Limb Patient Circuit

    Installation and Assembly See also Figure 4-9 on page 4-9. Figure 4-8. Double Limb Patient Circuit Note: Although shown here, the humidifier (item 2), water traps (item 3), and their connecting tubes are not included with the patient circuit or ventilator. Contact your supplier for more information. Inhalation Port Exhalation Valve Tube…

  • Page 78: Figure 4-10. Single Limb Patient Circuit Without Exhalation Valve

    Installation and Assembly 5. Place a water trap (item 3) between the outlet port of the humidifier and the patient end. 6. Place a vented (NIV) interface to the end of the patient circuit. (item 5) Figure 4-10. Single limb Patient Circuit Without Exhalation Valve For both types of circuits, shown previously, you should connect the end of the proximal pressure tube as close as possible to the patient (at the mask or cannula entry, if possible) so that the ventilator can account for all load losses due to the circuit and its potential accessories.

  • Page 79: Filters

    Installation and Assembly Filters Warning Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator. If necessary, replace the filter before the recommended replacement period is over (see chapter “Routine Maintenance”). This is particularly important when the ventilator is installed on a wheelchair, because environmental conditions may cause the filter to become dirty more rapidly.

  • Page 80: Humidifier

    Installation and Assembly • Connected to the FROM PATIENT port: This filter protects the internal exhalation flow sensor from the gases exhaled by the patient. Refer to Figure 4-8, item 10. Figure 4-12. Bacteria Filter Refer to the manufacturer’s instructions for more information about the use and maintenance of the bacteria filter(s).

  • Page 81: Exhalation Block

    Installation and Assembly water trap. If you notice any moisture in the patient circuit, you need to replace the wet circuit components with dry ones. Refer to the humidification device’s instruction for information on operating, cleaning, and sterilising the humidifier. Exhalation Block Warning The exhalation block is intended for single use by a single patient…

  • Page 82: Connecting The Oxygen Supply

    Installation and Assembly The Puritan Bennett™ 560 Ventilator can be used with an optional oxygen analyser with minimum and maximum concentration alarms. Always measure the delivered oxygen with a calibrated oxygen analyser (FIO kit) that features a minimum and maximum concentration alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.

  • Page 83: Figure 4-16. Connecting The Oxygen Supply System

    Installation and Assembly To connect the oxygen supply system to the ventilator: Refer to Figure 4-16 as required: 1. Inspect the oxygen supply’s connector (Figure 4-16, item 1) to ensure that connector’s black O-ring (item 2) is not missing. 2. Push the oxygen supply’s oxygen connector (item 1) into the ventilator’s oxygen connector (Figure 4-16, item 3).

  • Page 84
    Installation and Assembly Warning The coupler must not remain connected to the oxygen connector unless it also connected to a leak-proof, external oxygen gas source. When an oxygen supply is not being used with the ventilator, disconnect the oxygen source completely from the ventilator. In the event of an oxygen leak, shut down the supply of oxygen at its source.
  • Page 85: Fitting The Ventilator Into The Dual Bag

    Installation and Assembly Fitting the Ventilator into the Dual Bag The Dual Bag is a carrying bag with a dual function. It allows the Puritan Bennett™ 560 Ventilator to either be mounted onto a wheelchair or carried as a backpack. (see Figure 4-19) Warning…

  • Page 86: Figure 4-19. Using The Dual Bag Accessory

    Installation and Assembly Figure 4-19. Using the Dual Bag Accessory To install the Dual Bag onto a wheelchair do the following: 1. Unclip the two backpack straps from the side clips. 2. Clip the suspension belt onto the central ring. 3.

  • Page 87: Mounting The Ventilator On The Utility Cart

    Installation and Assembly 4.11 Mounting the Ventilator on the Utility Cart Match the mounting holes (item 1) on the bottom of the Puritan Bennett™ 560 Ventilator to the mounting studs (item 2) on the top of the utility cart platform. User’s Manual 4-19…

  • Page 88: Connecting The Nurse Call Cable

    Before using the Nurse Call system, ensure that its connections are secure and it operates properly. For more information, contact Covidien. To connect the ventilator to a Nurse Call device, contact Covidien to check the ventilator’s compatibility with the Nurse Call device and order a suitable connection cable.

  • Page 89: Operating Procedures

    If the ventilator fails the alarm tests or if you cannot complete the tests, refer to section 3.8, “Troubleshooting” or call your equipment supplier or Covidien. Due to its limited internal battery’s reserve capacity, the ventilator should only be operated on the internal battery when no other power source is available. Ensure that the internal battery…

  • Page 90: Figure 5-1. Turning On The Ventilator

    Operating Procedures To turn the ventilator on: • Set the I/O switch (a covered, rocker-type switch located at the rear of the ventilator) to the I position, as shown in Figure 5-1 below. Figure 5-1. Turning on the Ventilator The following events occur: •…

  • Page 91: Usb Menu Parameters

    The USB Menu is not accessible from the Setup Menu or Maintenance menu. To access patient data via a PC, a dedicated software package, Puritan Bennett™ Respiratory Insight Software, is available for Clinicians. Contact Covidien or your Puritan Bennett™ product representative for further infor- mation.

  • Page 92: Usb Menu

    Operating Procedures 5.2.2 USB Menu To access the USB menu when a USB memory device is connected: Press the MENU key several times, until the USB Menu appears: Figure 5-4. Selecting the USB Menu In case of high priority alarm activation the ventilator will automatically display the alarm page. To return to the USB Menu, press the MENU key.

  • Page 93: Transfer Trends

    Operating Procedures • The cursor changes to the plus/minus symbol. • The parameter selected to be modified flashes. 3. Press UP or DOWN to change the selected parameter’s value. 4. Press ENTER to confirm the new parameter setting. • The new parameter setting is displayed continuously. •…

  • Page 94
    Operating Procedures • The parameter selected to be modified flashes. 3. Press UP or DOWN to change the selected parameter’s value. 4. Press ENTER to confirm the new parameter setting. • The new parameter setting is displayed continuously. • The cursor is placed at the STOP position. 5.
  • Page 95: Starting Ventilation

    Operating Procedures Starting Ventilation Before starting ventilation, refer to Appendix C, “Operational Verification Checklist”, and set the parameter values in the Preferences menu.) Warning Verify the functionality of the alarm conditions before connecting the patient to the ventilator. Before starting ventilation, ensure that the device is properly assembled and that the air inlet, cooling vents, and alarm sound diffusion holes are not obstructed.

  • Page 96: Stopping Ventilation

    Operating Procedures Front Panel Keyboard Figure 5-8. Starting Ventilation Stopping Ventilation Warning Do not allow a patient to remain connected to the ventilator when ventilation is stopped, because a substantial quantity of expiratory gas, primarily carbon dioxide, may be inhaled by the patient.

  • Page 97: Turning Off The Ventilator

    Operating Procedures • A new message appears that directs the user to press the key again to confirm ventilation stop. (shown in the graphic below). Figure 5-10. Stopping Ventilation (2) • A double “beep“ sounds. 3. Release the VENTILATION ON/OFF key.

  • Page 98
    This page is intentionally blank. 5-10 User’s Manual…
  • Page 99: Internal Battery

    6 Internal Battery Warning Even though the Puritan Bennett™ 560 Ventilator meets current safety standards, the internal Lithium-ion battery of the device exceeds the 100Wh threshold and is therefore considered to be Dangerous Goods (DG) Class 9 – Miscellaneous, when transported in commerce. As such, the Puritan Bennett™…

  • Page 100: Battery Operation

    Internal Battery Table 6-1. Internal Battery Reserve Capacity Displayed Values Average Operating Time on Internal Battery Power Vt 200 ml (± 5 ml) PIP  10 mbar (± 2 mbar) 11 hours (–10%) Rtot  20 bpm Vt300ml (± 5 ml) PIP …

  • Page 101
    Internal Battery Figure 6-1. Internal Battery Indicator • A loss of external supply alarm is activated. If ventilation is stopped, the internal battery reserve capacity is displayed as a percentage of battery charge. Refer to Figure 6-2. Figure 6-2. Battery Reserve Capacity as a Percentage If the ventilator is running, the internal battery reserve is momentarily displayed as a percentage.
  • Page 102: Testing The Battery

    Internal Battery Warning Due to its limited internal battery’s reserve capacity, the ventilator should only be operated on the internal battery when no other power source is available. Ensure that the internal battery never becomes fully discharged. When the “LOW BATTERY“ alarm is triggered, immediately connect the ventilator to an AC power supply to maintain ventilation and recharge the internal battery.

  • Page 103: Storage

    Internal Battery • The “INTERNAL BATTERY“ indicator flashes (Figure 6-4, item 2). Figure 6-4. Power Indicators When Charging the Battery When the battery charge is complete, the “INTERNAL BATTERY“ indicator turns off. Warning Even if the “INTERNAL BATTERY“ indicator is off, charge of the battery may sometimes be incomplete regardless of charge time when the ambient temperature is above 40 °C (104 °F).

  • Page 104
    Note: When the device is in storage it should be recharged monthly to maximise battery life. If the battery is stored for more than one month at a temperature greater than 21 °C (70 °F), or for more than one or two weeks at a temperature greater than 45 °C (113 °F), the reserve capacity of the battery may be affected.
  • Page 105: Cleaning

    7 Cleaning Warning A patient treated by mechanical ventilation is highly vulnerable to the risks of infection. Dirty or contaminated equipment is a potential source of infection. Clean the ventilator and its accessories regularly and systematically before and after each use and following any maintenance procedure to reduce the risks of infection.

  • Page 106: Cleaning The Accessories

    Cleaning Table 7-1. Approved Cleaning Solutions for Exterior Ventilator Surfaces Description Mild dishwashing detergent 70% isopropyl alcohol (rubbing alcohol) 10% chlorine bleach (90% tap water) Glutaraldehyde Hospital disinfectant cleaners Hydrogen peroxide 15% ammonia (85% tap water) Ammonia-based household cleaners Household cleaners Cleaning the Accessories Follow the accessory manufacturer’s instructions for cleaning the ventilator’s accessories and components, including the patient circuit.

  • Page 107: Routine Maintenance

    Do not attempt to open, repair or otherwise service the ventilator yourself. Doing so might endanger the patient, damage the ventilator, and/or void your warranty. Only personnel authorised and qualified by Covidien should repair, open or service the ventilator. Replacing the Air Inlet Filter Warning Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator.

  • Page 108: Recommended Schedule Of Maintenance

    Routine Maintenance b. The filter is properly installed in its housing. Proper installation of the filter prevents particles from entering the device. Figure 8-1. Replacing the Air Inlet filter Recommended Schedule of Maintenance Consumables and Replacement Intervals When used under normal circumstances — a relatively dust-free atmosphere, and without damage to the device and its components (shocks, cracks, significant dirt) — the intervals for replacing the ventilator’s consum- able elements are as follows: Table 8-1.

  • Page 109
    Routine Maintenance Note: For all additional accessories not necessarily considered as consumables consult the manufacturer’s recommendations. To prevent any risk of cross contamination we recommend the use of DAR™ filters (Ref: 351/5856 or equivalent) to protect the patient outlet port and the exhalation block port. Warning Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator.
  • Page 110: Service Assistance

    In the event of a problem with the ventilator, refer to chapter 3, “Alarms and Troubleshooting”. If you cannot determine the cause of the problem, contact your equipment supplier or Covidien. For more information and local Covidien Technical Service Contact details, See “Technical Support”…

  • Page 111: A Specifications

    A Specifications Physical Table A-1. Physical Description (Excluding Accessories) Ventilator Weight 9.9 lb. (4.5 kg) Ventilator Dimensions 9.25 in wide x 12.40 in deep x 6.0 in high (235 mm wide x 315 mm deep x 154 mm high) Connectors Inspiratory limb connector: ISO 22 mm (OD) conical Exhalation limb connector (on exhalation block): ISO 22 mm (ID) conical…

  • Page 112: Table A-3. Internal Lithium Ion Battery

    Specifications Table A-3. Internal Lithium Ion Battery Voltage 25.2 VDC Full-load capacity 4.8 Ah Ampere-hour rating On standby: 1.5 Ah During ventilation: 0.5 Ah Watt hour rating 124Wh to 126Wh Charging current Standby mode 1.5 A/hr. (duration: < 6 hr.) •…

  • Page 113: Indicators And Alarms

    Specifications Indicators and Alarms Table A-5. Power Indicators Ventilation ON/OFF AC power DC power Internal Battery Blue in standby mode Green Green Flashing if the battery • • charge is in progress. Not lit if ventilation is in • progress. Continuously lit if the •…

  • Page 114: Range, Resolution, And Accuracy

    Specifications Table A-9. Monitored Parameter Tolerances (Continued) Ventilator Parameters Tolerances Exhalation Tidal Volume (VTE) ± (10 ml + 10%VTE)*VTE Total Breath Rate (Rtot) ± 1 bpm I:E Ratio (I:E) ± 50 ms or 10%, whichever is greater I/T Ratio (I/T) ±…

  • Page 115
    Specifications Table A-10. Ventilator Range, Resolution, and Accuracy (Continued) Ventilator Settings Range, Resolution, and Accuracy Pressure support Range: OFF or 5 mbar to 55 mbar in valve configuration (P Support) Range: 6 mbar to 30 mbar in leak in valve configuration Resolution: 1 mbar Accuracy: ±…
  • Page 116
    Specifications Table A-10. Ventilator Range, Resolution, and Accuracy (Continued) Ventilator Settings Range, Resolution, and Accuracy PEEP Range: OFF (0.5 mbar) to 20 mbar Resolution: 1 mbar Accuracy: ± (1 mbar + 10%) mbar Default value: OFF Depends on: PIP in P A/C and PSV modes when Relative Pressure is set to YES Depends on: P Support and P Control in P SIMV mode when Relative Pressure is set to YES…
  • Page 117: Environmental

    Specifications Table A-10. Ventilator Range, Resolution, and Accuracy (Continued) Ventilator Settings Range, Resolution, and Accuracy Minimum Peak Inspiratory Pressure Range: PIP- 20% (not adjustable in pressure breath) (Min PIP) Range: 2-52 in volume breath Resolution: N/A Maximum Peak Inspiratory Pressure Range: PIP+ 20 % (not adjustable in pressure breath) (Max PIP) Range: 12-60 in volume breath…

  • Page 118: Usb

    Specifications Table A-13. USB Memory Device Specifications Characteristics Supported formats USB compatibility USB flash memory USB 2.0 or USB 1.1 Memory file format USB 32 bit format (sector size: 512 — 2,048 bytes) Number of files Maximum 999 USB size 128 MB to 4 GB Table A-14.

  • Page 119: Manufacturer’s Declaration

    Specifications A.10 Manufacturer’s Declaration The following tables, Table A-20 through Table A-23, contain the manufacturer’s declarations for the ventila- tor’s electromagnetic emissions, electromagnetic immunity, and recommended separation distances between the ventilator and portable and mobile RF communications equipment, as well as a list of compliant cables. Warning Portable and mobile RF communications equipment can affect the performance of the Puritan Bennett™…

  • Page 120: Table A-21. Electromagnetic Immunity

    Specifications Table A-21. Electromagnetic Immunity The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should ensure that it is used in such an environment. Electromagnetic Immunity Test IEC / EN 60601 Test Level Compliance Level Environment–Guidance Electrostatic discharge…

  • Page 121: Table A-22. Electromagnetic Immunity — Conducted And Radiated Rf

    Specifications Table A-22. Electromagnetic Immunity – Conducted and Radiated RF The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment. IEC / EN 60601-1-2 Test Electromagnetic Environment–…

  • Page 122: Table A-23. Recommended Separation Distances

    Specifications Table A-22. Electromagnetic Immunity – Conducted and Radiated RF (Continued) Note: • At 80 MHz and 800 MHz, the higher frequency range applies. • These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

  • Page 123: Standards Compliance And Iec Classification

    Specifications Table A-24. Compliant Cables and Accessories Cable or Accessory Maximum length UK AC power cable assembly 1.8 m (5.9 ft) Japan AC power cable assembly 1.8 m (5.9 ft) China AC power cable assembly 1.8 m (5.9 ft) South Africa AC power cable assembly 1.8 m (5.9 ft) India AC power cable assembly 1.8 m (5.9 ft)

  • Page 124
    Specifications Particular Standards Lung Ventilators for Medical Use- Particular Requirements for Basic Safety and Essential • Performance Part 2: Home Care Ventilators for Ventilator-Dependent Patient EN ISO10651- 2:2009. Lung ventilators for medical use – Particular requirements for basic safety and essential •…
  • Page 125: B Modes Of Ventilation

    B Modes of Ventilation Modes of Ventilation This chapter is a general description of the various modes of ventilation and breath types available with the Device name. Note: The default ventilation mode setting is P A/C; for more information, see below. B.1.1 Assist/Control (A/C) Modes When set to an Assist/Control mode, machine-initiated breaths are delivered at a clinician-set volume or pres- sure, inspiratory time, and rate.

  • Page 126: Psv Mode

    Modes of Ventilation B.1.4 PSV Mode PSV mode maintains a constant level of pressure in the patient’s airway during exhalation. In addition, the venti- lator applies a clinician-set pressure (Pressure Support) to each of the patient’s breaths. This has the same benefits as CPAP, with the additional benefit of assisting the patient in moving gas into his or her lungs.

  • Page 127: C Operational Verification Checklist

    • Following maintenance or changes in ventilator settings If the ventilator fails any of the safety checks below, or if you cannot complete these checks, refer to section 3.8, “Troubleshooting” or call the equipment supplier or Covidien (refer to section 8.3, “Service Assistance”).

  • Page 128
    This page is intentionally blank. User’s Manual…
  • Page 129: D Unpacking And Preparation

    Never use a ventilator or any components or accessories that appear to be damaged. If any signs of damage are evident, contact your equipment supplier or Covidien. 4. Clean the ventilator with a mild soap solution, if necessary (refer to chapter “Cleaning”).

  • Page 130
    Unpacking and Preparation Figure D-1. Puritan Bennett™ 560 Ventilator Figure D-2. Dual Bag User’s Manual…
  • Page 131: E Alarms Tests

    Troubleshooting section (refer to chapter 3, “Alarms and Troubleshooting”) of this manual or call your equipment supplier or Covidien (refer to section 8.3, “Service Assistance” The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high enough to allow the PATIENT DISCONNECTION alarm to trigger properly.

  • Page 132: Power Failure Test

    Alarms Tests Power Failure Test Note: If the ventilator is operating on either the external power supply or the internal battery, you must plug it in to an AC power source before beginning this test. 1. Disconnect the ventilator from its AC power supply. Ensure that the following events occur: •…

  • Page 133: Testing The Battery

    Alarms Tests 7. Press and hold the VENTILATION ON/OFF key for three (3) seconds, then release it. Press the VENTILATION ON/OFF key again to confirm stop. • Ventilation stops. Testing the Battery The ventilator is capable of testing the power of the battery (refer to chapter 6, “Internal Battery”).

  • Page 134
    This page is intentionally blank. User’s Manual…
  • Page 135: F Parts And Accessories

    Table F-1 provides a list of accessories that are available for the Device name. To order parts or accessories, contact your equipment supplier or Covidien representative. Note: The ventilator is delivered with the following items: a printed User’s Manual, a CD with Clinician’s Manual (printed copy available upon request);…

  • Page 136: Table F-2. List Of Circuits

    Table F-2 provides a list of consumable parts available for the ventilator. Warning To ensure proper performance of the ventilator, use a patient circuit recommended by Covidien in this manual; refer to chapter 4, “Installation and Assembly” Appendix F, “Parts and Accessories”.

  • Page 137: G Glossary

    G Glossary AC Power Alternating current. Alarm Pause The audible and visual alarms cease and the symbol appears. The symbol will remain until the cause of the alarm is addressed. For example, when the ventilator is running on internal battery, the AC Disconnection alarm may be paused, and the alarm paused symbol will appear until the device is plugged into AC.

  • Page 138
    Glossary Battery Level Display of the remaining battery capacity; located adjacent to the battery symbol. Bias flow Turbine flow during exhalation phase through the patient circuit to avoid rebreathing. An abbreviation for “breaths per minute,” which is the unit of measure for breath rate (see below). Breath Rate (Back Up R) The total number of breaths, both machine and spontaneous, delivered by a ventilator in one minute.
  • Page 139
    Glossary Fraction of Inspired Oxygen (FIO Amount of oxygen delivered to the patient. Sensor The sensor which measures the amount of oxygen being delivered to the patient. Flow Volume of gas delivered by the ventilator compared to time, expressed in litres per minute (lpm). Flow Pattern (Ramp Setting) This is the flow distribution shape during the inspiration phase.
  • Page 140
    Glossary Light Emitting Diode; used as indicator lights on the ventilator’s front panel. Litres Per Minute (a unit of volume flow rate). Machine Hours Counter for the total ventilation time since manufacture or the last CPU board change. Mains AC power supply. Max Leak The maximum alarm setting of a high leakage threshold.
  • Page 141
    Glossary Patient Counter Counter of ventilation time for the patient. Patient effort Inspiratory effort initiated by the patient. Patient circuit Tubing between the ventilator and the patient. Pause Waveforms freezing function. PAW (Peak Airway Pressure) The Peak Airway Pressure is the average peak pressure during the inspiratory phase, measured by each cycle and over the previous 24 hour period.
  • Page 142
    Glossary Rtot Parameter measured by the ventilator equal to the total number of breaths per minute (bpm). Sensitivity This adjustable parameter determines the amount of inspiratory effort required by the patient before the ventilator delivers an assisted breath, or demands flow in the case of a spontaneous breath.
  • Page 143
    Index DEVICE FAULT12….3-10, 3-18 DEVICE FAULT13….3-10, 3-18 AC power DEVICE FAULT3.
  • Page 144
    Index utilisation ……3-1 CONNECT VALVE OR CHANGE PRESS alarm message ventilation ……3-1 3-10, 3-17 Alarms tests Connecting to…
  • Page 145
    Index Front panel ……2-4 LOW VTl alarm message … . 3-12, 3-23 Heat safety device, battery .
  • Page 146
    Index Physical specifications ….A-1 pneumatic……A-8 Placing the ventilator (installing) .
  • Page 147
    Index mounting on a wheelchair … . . 4-17 mounting onto a utility cart … . 4-19 parts and accessories ….F-1 patient outlet port connections (figure).
  • Page 148: Index

    Index This page intentionally blank Index-6 User’s Manual…

  • Page 150
    Part No. 10066883 Rev. B 03/2012 COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and/or internationally registered trademarks of Covidien AG. ™* Trademark of its respective owner. Other brands are trademarks of a Covidien company. ©2012 Covidien. Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.

ИВЛ Puritan Bennett 560

Аппарат искусственной вентиляции легких Puritan Bennett 560 предназначен для постоянной или периодической механической вентиляционной поддержки пациентов с весом тела не ниже 5 кг, которым требуется проведение механической вентиляции. Аппарат ИВЛ Puritan Bennett 560 представляет собой медицинский прибор ограниченного применения, который должен использоваться квалифицированным и обученным персоналом под руководством врача.

Puritan Bennett 560

Данный вентилятор специально предназначен для взрослых и детей, которым необходимо проведение следующих видов инвазивной или неинвазивной вентиляции легких, как предписано наблюдающим врачом:

  • Вентиляция с положительным давлением.
  • Вспомогательная вентиляция с поддержкой/управлением, режимы SIMV или CPAP.
  • Дыхание с управлением по давлению, объему или с поддержкой по давлению.

Вентилятор Puritan Bennett 560 рассчитан на использование в условиях медицинского учреждения, дома и в качестве переносного устройства. Не предполагается использовать его в качестве аппарата ИВЛ для машин скорой помощи.

Использование Puritan Bennett 560 COVIDIEN

Аппаратом искусственной вентиляции легких могут управлять:

  • Врачи – специалисты в области заболеваний органов дыхания;
  • Врачи;
  • Медсестры;
  • Персонал, ухаживающий за больными на дому (сиделки);
  • Сам пациент и его родственники.

Переносной аппарат ИВЛ Puritan Bennett 560 использует микро-турбину для осуществления респираторной поддержки пациентов. Практикующее врачи могут пользоваться различными способами для подключения пациентов к этому аппарату: носовыми масками или лицевыми масками, эндотрахеальными или трахеотомическими
трубками. Пользователь может выбирать следующие режимы вентиляции:

  • С поддержкой/управлением по объему (V A/C).
  • С поддержкой/управлением по давлению (Р A/C).
  • Синхронизированная перемежающаяся принудительная вентиляция (СППВ) с поддержкой по объему (V SIMV).
  • Синхронизированная перемежающаяся принудительная вентиляция (СППВ) с поддержкой по давлению (Р SIMV).
  • Самостоятельное дыхание с созданием постоянного положительного давления (СДППД) в дыхательных путях (СРАР).
  • Вентиляция с поддержкой давлением и вентиляцией при отсутствии дыхания ВПД/ОД) (PSV/ST).

Система безопасности

В вентилятор встроена система сигнализации, которая постоянно отслеживает состояние пациента и машины на предмет сигналов о конкретных ошибках или неполадках, которые могут вызвать опасную ситуацию. Если такие ошибки или неполадки замечены, система сигнализации подает характерный сигнал тревоги, как зрительный, так и звуковой. Параметры срабатывания сигнализации по неполадкам прибора заданы на производстве, в то время как параметры срабатывания, связанные с пациентом, задаются пороговыми значениями величин, которые выбирают операторы (врачи в клинике или сиделки/медперсонал).

Настройки аппарата ИВЛ

Программная клавиша, так называемая клавиша блокировки, предотвращает доступ к настройкам параметров вентиляции, чтобы разграничить «врачебное» использование и «личное» использование самим пациентом.

Обогащение кислородом

Кислород может подаваться из внешнего источника низкого давления, но его расход должен быть не выше 15 л/мин, давление – 50 кПа, 7 PSI. Аппараты искусственной вентиляции легких (ИВЛ) Puritan Bennett 560 автоматически компенсирует избыточный расход, создаваемый при подаче кислорода извне.

Дыхательный контур

Вентилятор можно использовать с контуром пациента, снабженным одним или двумя патрубками (отводами). Если нужно контролировать объем выдыхаемого воздуха (как в случае с пациентами, которые самостоятельно дышать не могут), следует использовать контур с двумя патрубками для контроля дыхательного объема на выдохе.

Производитель Puritan Bennett 560

Производитель аппарата искусственной вентиляции легких Puritan Bennett 560 – Medtonic (COVIDIEN), страна производитель США.

Технические характеристики Puritan Bennett 560

Ниже указаны технические характеристики аппарата Puritan Bennett 560. Физическое описание (включая дополнительные приспособления).

Технические характеристики Puritan Bennett 560

Полный перечень технических характеристик можно найти в инструкции. Скачать инструкцию на аппарат ИВЛ Puritan Bennett 560 COVIDIEN можно в конце статьи.

Видео обзор аппарата ИВЛ Puritan Bennett 560

Ниже представлено видео с обзором на аппарат ИВЛ Puritan Bennett 560 (Covidien, Medtronic).

Передняя панель Puritan Bennett 560

Передняя панель
  1. ЖК- дисплей – отображает информацию о вентиляторе, в том числе время, проведенное пациентом на вентиляторе, версию ПО, режимы и настройки вентиляции, контролируемые и расчетные данные пациента и формы сигналов. Также при помощи дисплея пользователь может просматривать рабочие параметры и параметры системы сигнализации вентилятора и менять их, пользуясь панелью управления.
  2. Панель управления – предоставляет возможность управления настройками и работой вентилятора, а также светодиодную индикацию источника питания вентилятора, состояния «вкл/выкл» и приоритета сигналов тревоги. Функции управления включают в себя выключение и включение вентилятора, настройку режимов работы вентилятора, заглушение и отмену сигналов тревоги и установку параметров срабатывания системы сигнализации и рабочих параметров аппарата.
  3. Подключение датчика FiO2 – для датчика FiO2, который измеряет концентрацию кислорода в контуре пациента.
  4. Порт подключения пациента – предоставляет возможность поступления газа к пациенту через дыхательный контур.
  5. Разъем контроля давления пациента – Патрубок для контроля проксимального давления у пациента.
  6. Порт клапана выдоха – патрубок для подачи управляющего давления на клапан выдоха. Управляет открытием и закрытием клапана выдоха.
  7. Боковые и передние отверстия – вентиляционные отверстия, благодаря которым осуществляется охлаждение внутренних узлов вентилятора. Кроме того, они служат для распространения звукового сигнала тревоги.
  8. Порт «от пациента» – через него осуществляется измерение выдыхаемого объема, т.к. часть выдыхаемого газа через этот порт направляется на датчик расхода. Путем измерения этого расхода вычисляется величина VTE (Дыхательный объем на выдохе).
  9. Выход выдыхаемого газа – сюда подключается клапан выдоха.

Задняя панель

Задняя панель
  1. Эргономичная ручка для переноски.
  2. Выключатель питания «Вкл/Выкл» (I/O) с защитной крышкой: В положении I прибор включен, в положении О — выключен.
  3. Разъем шнура электропитания переменного тока («Сетевой»).
  4. Система крепления шнура электропитания переменного тока («Сетевого»): Закрепляет шнур электропитания переменного тока во избежание случайного отсоединения.
  5. Крышка аккумуляторного отсека.
  6. Разъем для подключения шнура электропитания постоянного тока с направляющим ключом.
  7. Разъем для подключения ПК Мини- USB разъем для подключения тестирующего ПО вентилятора Puritan Bennett 560.
  8. Порт подвода O2: Соединяет вентилятор с источником кислорода низкого давления через переходник, надетый на патрубок ввода O2.
  9. Разъем вывода сигнала на устройство вызова медсестры: Используется для подключения вентилятора к системе вызова медсестры.
  10. Разъем устройства памяти USB: Разъем USB, который используется с программным обеспечением Puritan Bennett по поддержанию работы вентилятора. Имеются два порта USB типа А.
  11. Воздушный фильтр на входе в аппарат: Фильтрует воздух, поступающий в аппарат.

Панель управления

Панель управления
  1. Световые сигналы (два СДИ). Красные: (Непрерывный) Сработал сигнал тревоги ОВУ (очень высокого уровня); Сработал сигнал тревоги ВУ (высокого уровня). Желтый: Сработал сигнал тревоги СУ (среднего уровня).
  2. Клавиша УПРАВЛЕНИЕ СИГНАЛИЗАЦИЕЙ. Однократное нажатие заглушает звуковой сигнал тревоги на 60 секунд. Двойное нажатие останавливает звуковой и световой сигнал тревоги. Если ситуация, вызвавшая сигнал тревоги, исправлена, то сигнализация отменяется (за исключением сигнала по высокому давлению).
  3. Дисплей. Отображает режимы, настройки вентиляции, данные пациента и формы сигналов расхода или давления газа, конфигурацию вентилятора, а также позволяет управлять сигналами тревоги.
  4. Клавиша ВВЕРХ/ВОЗОБНОВИТЬ. Перемещает курсор вверх и увеличивает значение параметра. В ходе вентиляции возобновляет отслеживание форм сигналов расхода или давления в меню форм кривых.
  5. Клавиша ВВОД. Дает доступ к величинам настроек и проверке настроек в случае их изменения. Доступ в подменю.
  6. Клавиша ВНИЗ/ОСТАНОВИТЬ. Перемещает курсор вниз и уменьшает значение параметра. В ходе вентиляции останавливает отображение форм сигналов в меню форм кривых.
  7. Клавиша МЕНЮ. Изменяет отображаемое меню. Нажатием этой клавиши из меню вентиляции можно вызвать экран меню сигнализации. Когда к вентилятору подключается USB-карта памяти, нажатием этой клавиши можно вызвать экран меню устройства USB.
  8. Клавиша ВКЛ/ВЫКЛ ВЕНТИЛЯЦИЮ. Кратковременное нажатие на эту клавишу запускает вентиляцию. Для остановки вентиляции нажатие нужно удерживать в течение трех (3) секунд.
  9. Индикатор состояния вентиляции. Горит голубой индикатор: прибор включен и находится в состоянии готовности (вентиляция отключена). Голубой индикатор отключен: вентиляция включена.
  10. Индикаторы источников электропитания. Горит индикатор ПЕРЕМЕННЫЙ ТОК: подключен источник питания переменного тока. Горит индикатор ПОСТОЯННЫЙ ТОК: подключен источник питания постоянного тока. Индикатор ВСТРОЕННЫЙ АККУМУЛЯТОР горит постоянно: используется встроенный аккумулятор (никакой внешний источник не подключен). Индикатор ВСТРОЕННЫЙ АККУМУЛЯТОР мигает: аккумулятор заряжается.

Сигналы тревоги или неисправности

Сигналы тревоги или неисправности, создаваемые аппаратом Вентилятор Puritan Bennett 560, классифицируются по двум категориям:

  • тревоги вентиляции (или те, что возникают в ходе работы);
  • технические неисправности.

Некоторые из сигналов тревоги вентилятора могут регулироваться в зависимости от режима работы вентилятора. Также существуют нерегулируемые сигналы тревоги, предназначенные для создания сети безопасности и более безопасной вентиляции пациента.

Сигналы тревоги указывают на события, которые могут оказать влияние на ход вентиляции и требуют быстрого вмешательства.

Технические неисправности не влияют непосредственно на работу аппарата. Соответственно, пользователь не ставится в известность о технических неисправностях. Меню технического обслуживания доступно только для обученных техников с правом допуска.

Уровень приоритетности сигналов тревоги

Ниже перечислена иерархия сигналов тревоги в зависимости от того, насколько они критичны:

ОВУ, очень высокий уровень: Наступила критическая ситуация; вентиляция невозможна.

Постоянный звуковой сигнал / с постоянно горящим красным СДИ или без него / с выдачей сообщения или без него / с освещением дисплея или без него (возможны такие аварийные ситуации, при которых не будет как сообщения, так и освещения дисплея).

ВУ – высокий уровень: Критическая ситуация вскоре наступит; потенциальная угроза проведению вентиляции.

Высокочастотный прерывистый звуковой сигнал / мигающий красный СД-индикатор / сообщение / освещение дисплея.

СУ – средний уровень: До начала критической ситуации может пройти значительное время; в ближайшее время вентиляция не прервется.

Прерывистый звуковой сигнал, подаваемый со средней частотой / мигающий желтый СД-индикатор / сообщение / освещение дисплея.

Если не будут приняты какие-либо меры и если звуковой сигнал не будет прерван (заглушением звукового сигнала) или сброшен (путем сброса сигнализации) в течение 60 секунд, то сигнал ВУ будет подаваться на максимальной громкости.

Фильтры Puritan Bennett 560

Аппарат искусственной вентиляции легких Puritan Bennett 560 использует два типа фильтров:

  • воздушный фильтр на входе в аппарат;
  • бактериальный фильтр.

Воздушный фильтр на входе в аппарат

Этот фильтр, состоящий из поролона и специального наполнителя для удаления мелких частиц, расположенный в задней части вентилятора, фильтрует воздух по мере его поступления в прибор.

Бактериальный фильтр

Настоятельно рекомендуется устанавливать бактериальный фильтр как с однопатрубочными, так и с двухпатрубочными контурами. В двухпатрубочном контуре используются два бактериальных фильтра: один в порте К ПАЦИЕНТУ, другой — в порте ОТ ПАЦИЕНТА.

Увлажнитель Puritan Bennett 560

Увлажнитель добавляет влагу (водяной пар) и согревает газ в контуре пациента. Он вставляется в контур пациента между главным выходным патрубком и собственно пациентом.

Увлажнитель Puritan Bennett 560

При использовании увлажнителя весь конденсат, образующийся в контуре пациента, собирается в водяной ловушке. При обнаружении любой влаги в контуре пациента влажные его части нужно заменить сухими.

Подключение датчика FIO2 (концентрации кислорода)

При назначении кислорода рекомендуется использовать датчик концентрации кислорода FiO2, который можно присоединить спереди к аппарату при помощи комплекта для определения FiO2 (концентрации кислорода).

Датчик кислорода требует калибровки, снимать и чистить его разрешается только квалифицированным сотрудникам. Если используется новый датчик, то его нужно выдержать при температуре окружающей среды в течение примерно 20 минут прежде чем устанавливать, калибровать и начинать вентиляцию.

Чтобы установить датчик FiO2 (концентрации кислорода):

датчик FIO2
  1. Выньте устройство из герметичной упаковки.
  2. Установите переходник FiO2 (концентрации кислорода) в разъем FiO2 на вентиляторе (поз. 1).
  3. Соедините датчик FiO2 (концентрации кислорода) (поз.2) с переходником диаметром 15 мм (поз.3).
  4. Установите переходник на выходной порт К ПАЦИЕНТУ, как показано. Закрепите контур пациента за переходником.

Калибровка датчика кислорода FiO2

В видео представлена последовательность действий при калибровке датчика кислорода (FiO2) аппарата ИВЛ Puritan Bennett 560 и для чего это нужно.

Аккумулятор Puritan Bennett 560

Вентилятор Puritan Bennett 560 удовлетворяет всем условиям электробезопасности, внутренний аккумулятор литий-ионная батарея. Максимальный рекомендуемый срок хранения встроенного аккумулятора два года. Не следует брать в работу аккумулятор, который пролежал на складе два года и ни разу не использовался.
Для продления срока службы аккумулятора важно, чтобы его периодически перезаряжали. Не следует хранить аккумуляторы длительное время без перезарядки, это сокращает срок их службы.

Емкость аккумулятора Puritan Bennett 560

Резервная емкость встроенного аккумулятора зависит от уровня заданных параметров вентиляции, от условий окружающей среды (в основном, от температуры) и от физиологических характеристик пациента.

Проверка уровня заряда аккумулятора подразумевает, что на момент проверки вентилятор находится в работе и работает от аккумулятора. Чтобы проверить уровень заряда аккумулятора, следует временно отключить вентилятор от источника переменного тока (либо в режиме готовности, либо в ходе вентиляции) и прочесть значение уровня заряда в процентах, которое показывается рядом с символом аккумулятора, отображаемым в верхней части экрана дисплея.

Работа аккумулятора Puritan Bennett 560

Перед тем, как пользоваться встроенным аккумулятором вентилятора, следует убедиться, что он полностью заряжен и удерживает заряд. Запасные вентиляторы или вентиляторы, находящиеся на хранении, должны быть подключены к источнику переменного тока, чтобы защитить целостность аккумулятора.

Когда прибор впервые запускается после полной перезарядки встроенного аккумулятора, возможно срабатывание сигнального устройства и сигнализации по состоянию аккумулятора. В этом случае нужно подключить прибор к сети переменного тока и перезапустить питание прибора.

В случае прекращения подачи переменного тока или отсоединения от внешнего источника переменного или постоянного тока вентилятор автоматически переключается на встроенный аккумулятор и происходит следующее:

  • Символ аккумулятора отображается в верхней части строки общих сведений.
  • Остаток емкости аккумулятора отображается справа от соответствующего символа.
  • Индикатор «ВСТРОЕННЫЙ АККУМУЛЯТОР» в левой верхней части передней панели индикатора горит постоянно.
  • Срабатывает сигнализация по прекращению подачи электропитания от внешнего источника.

Если вентиляция останавливается, остаток емкости аккумулятора отображается в виде процента заряда аккумулятора.

Если вентилятор работает, то остаток емкости встроенного аккумулятора кратковременно показывается в виде процента. Затем, после того, как вентилятор рассчитает срок работы от аккумулятора (что может занять до двух минут, в зависимости от энергопотребления вентилятора) остаток заряда аккумулятора отображается в часах и минутах (с округлением до пятнадцати минут).

С момента срабатывания сигнала «БАТАРЕЯ РАЗРЯЖЕНА», если не подключить прибор к внешнему источнику тока, могут сработать другие сигналы тревоги в связи с недостаточным уровнем напряжения. В последней стадии разрядки сигнал «БАТАРЕЯ РАЗРЯЖЕНА» становится постоянным, и вентиляция в любой момент может быть прекращена.

Проверка аккумулятора Puritan Bennett 560

Вентилятор постоянно автоматически проверяет состояние встроенного аккумулятора, даже когда он не используется в качестве основного источника энергии. Сигнал «СБОЙ БАТАРЕИ 1» срабатывает в случае обнаружения любой проблемы с аккумулятором или зарядным устройством.

Раз в месяц нужно отсоединять вентилятор от внешнего источника энергии, чтобы проверить целостность соединений между встроенным аккумулятором и другими узлами вентилятора.

Перезарядка аккумулятора Puritan Bennett 560

Когда уровень заряда в аккумуляторе недостаточен, если судить по показателям остаточной емкости, необходимо выполнить перезарядку аккумулятора. В общем рекомендуется ставить вентилятор на зарядку, когда уровень заряда в аккумуляторе падает ниже 80%, а также систематически перезаряжать вентилятор после хранения и перед повторным использованием.

Во избежание перехода с одного источника питания на другой и для продления срока службы аккумулятора, он при присоединении к источнику переменного тока не будет заряжаться до тех пор, пока заряд не упадет ниже 85 — 90%.

Чтобы перезарядить встроенный аккумулятор, нужно подсоединить вентилятор к источнику переменного тока. Когда аккумулятор полностью заряжается, индикатор ВСТРОЕННЫЙ АККУМУЛЯТОР гаснет.

Нет необходимости запускать вентилятор в работу при зарядке аккумулятора, пополнение заряда при работающем вентиляторе приводит к увеличению времени, необходимого для полной зарядки встроенного аккумулятора.

При перезарядке истощенного аккумулятора может понадобиться оставить вентилятор на зарядке на протяжении 6 (шести) часов в режиме готовности или около 13 часов при работающем вентиляторе.

Хранение аккумулятора

Если вентилятор необходимо сдать на хранение на длительный срок, не обязательно вынимать из него аккумулятор. Однако хранить вентилятор следует в сухом прохладном месте, хорошо вентилируемом, при следующих условиях:

  • Температура: около 21 °C (70 °F);
  • Влажность: менее 80% относительной влажности.

Если аккумулятор хранится более месяца при температуре выше 21 °C (70 °F) или более двух недель — при температуре свыше 45 °C (113 °F), то это может сказаться на остаточной емкости аккумулятора. Перед повторным использованием такой аккумулятор следует перезарядить.

Если вентилятор в течение более чем 30 дней хранился подключенным к источнику переменного тока, перед началом вентиляции включите его выключателем I/O (вкл/выкл) в задней части аппарата и дайте ему зарядиться в течение 15 минут.

Очистка Puritan Bennett 560

Протирайте все внешние панели и поверхности перед и после работы с каждым из пациентов, а также по мере надобности, чтобы вентилятор всегда был чистым. Загрязненные или испачканные участки вентилятора нужно очищать перед выполнением технического обслуживания, а также перед отправкой на хранение.

Чтобы почистить поверхность вентилятора, нужно:

  1. Намочить чистую мягкую ткань в некрепком мыльно-водном растворе или в другом разрешенном моющем растворе.
  2. Тщательно отжать ткань, чтобы удалить избыток жидкости.
  3. Слегка протереть наружный корпус вентилятора, следя за тем, чтобы жидкость не попала в какие-либо из отверстий на его поверхности. См. предупреждение выше.
  4. Вытереть насухо поверхность вентилятора чистой мягкой безворсовой тканью.

Растворы, которыми разрешено пользоваться для чистки внешних
поверхностей вентилятора:

  • Слабый раствор средства для мытья посуды;
  • 70% изопропиловый спирт (для протирки);
  • 10% хлорный отбеливатель (90% — вода из крана);
  • Глутаральдегид;
  • Больничные дезинфектанты в форме раствора;
  • Перекись водорода;
  • 15% р-р аммиака (85% — вода из-под крана);
  • Домашние чистящие средства, содержащие аммиак;
  • Домашние чистящие средства.

Техническое обслуживание Puritan Bennett 560

Необходимо регулярно проверять чистоту воздушного фильтра на входе в аппарат,который расположен в задней части вентилятора. При необходимости фильтр заменяют до истечения рекомендованного периода замены. Если вовремя не заменить воздушный фильтр на входе в аппарат или допустить работу вентилятора «без» — Yulia фильтра, это может привести к серьезной поломке аппарата. Воздушные фильтры являются одноразовыми изделиями. Запрещается мыть, чистить или повторно использовать их.

При использовании вентилятора в помещении проверку состояния воздушного фильтра следует проводить раз в месяц. При использовании аппарата на открытом воздухе или в запыленном помещении проверку воздушного фильтра следует выполнять еженедельно и заменять фильтр по необходимости.

Рекомендуемый график технического обслуживания

При использовании в нормальных условиях — относительно незапыленной атмосфере и без повреждений аппарата и его узлов (толчки, трещины, сильная загрязненность) — сроки замены расходных материалов вентилятора приведены ниже.

Расходные материалы и сроки их замены

Воздушный фильтр на входе в аппарат (Поролон + мелкие частицы).

Раз в месяц или чаще, в зависимости от степени загрязнения.

Инспираторный бактериальный фильтр.

Смотреть рекомендации производителя.

Контур пациента.

Смотреть рекомендации производителя. Одноразовый однопользовательский (для одного пациента).

Датчик FiO2.

От 14 до 18 месяцев или чаще в случае постоянных сбоев в калибровке.

Блок выдоха.

Раз в 4 месяца и для каждого нового пациента.

Замена встроенного аккумулятора

Встроенный аккумулятор нужно заменять, когда его емкость падает ниже 3450 мАh. Следует иметь в виду, что по соображениям экологического характера вентилятор и его узлы — включая встроенный аккумулятор — нельзя утилизировать как бытовые отходы. Вентилятор и его узлы следует отправлять на раздельный сбор вторсырья для возможной утилизации и повторного использования с соблюдением всех применимых законодательных норм.

Режимы вентиляции Puritan Bennett 560

Ниже представлено общее описание различных режимов вентиляции и типов дыхания, которые может обеспечивать Вентилятор Puritan Bennett 560. По умолчанию в приборе установлен режим P A/C.

Режимы с поддержкой/управлением (A/C)

При задании режимов с поддержкой/управлением такие параметры аппаратно инициируемых вдохов, как объем, давление, время вдоха и частота дыхания задаются лечащим врачом. Если у пациента между аппаратными вдохами случается самопроизвольный вдох, то вентилятор подает ему воздух в объеме или под давлением, заданными соответствующими установками, а также в течение заданного времени.

Вне зависимости от того, спонтанные это дыхания или аппаратные, каждое дыхание совершается в том же самом заданном объеме (или при том же заданном давлении) и за одно и то же заданное время вдоха.

Режимы с поддержкой/управлением бывают:

  • V A/C (с поддержкой/управлением по объему), если заданным параметром дыхания является объем;
  • P A/C (с поддержкой/управлением по давлению), если заданным параметром дыхания является давление.

Режимы SIMV

При задании режима SIMV (Синхронизированная перемежающаяся принудительная вентиляция, СППВ) такие параметры аппаратно инициируемых вдохов, как объем, давление, время вдоха и частота дыхания задаются лечащим врачом. Эти принудительные вдохи синхронизированы с дыхательными усилиями пациента. Если между аппаратными вдохами пациент делает вдох самостоятельно, вентилятор осуществляет самостоятельное дыхание с поддержкой давлением.

Самостоятельное дыхание с постоянным положительным давлением в дыхательных путях (CPAP) недоступно в режимах СППВ.

Режимы SIMV бывают:

  • V SIMV (СППВ с поддержкой по объему), если заданным параметром принудительного дыхания является объем;
  • P SIMV (СППВ с поддержкой по давлению), если заданным параметром принудительного дыхания является давление.

Режим CPAP

В режиме CPAP вентилятор поддерживает постоянный уровень давления в дыхательных путях пациента.

Режим PSV

В режиме PSV постоянный уровень давления поддерживается в дыхательных путях пациента во время выдоха. Кроме того, каждое из дыханий пациента вентилятор поддерживает заданным врачом давлением поддержки — Pressure Support. Это дает те же преимущества, что и метод CPAP, дополнительно помогая пациенту в поступлении газа в легкие.

Запчасти и расходные материалы

Ниже указан перечень приспособлений для Вентилятор Puritan Bennett 560. Чтобы заказать запчасти или приспособления, свяжитесь с поставщиком оборудования или с представителем компании Covidien.

Вентилятор поставляется в следующей комплектации: отпечатанное Руководство для пользователя, компакт-диск с Руководством для практикующего врача (печатный экземпляр можно заказать отдельно); Один контур пациента с клапаном; один комплект из 6 (шести) комбинированных воздушных фильтров для входного патрубка (поролоновый фильтр/фильтр мелких частиц); одна сумка для переноски; один кислородный переходник; и один шнур электропитания переменного тока.

Перечень расходных материалов

  • Сумка для переноски (серая).
  • Впускной патрубок для кислорода.
  • Тележка для вентилятора.
  • Двойная сумка (синяя или розовая).
  • Шнур электропитания переменного тока (сетевой).
  • Шнур электропитания постоянного тока (для подключения к внешнему источнику постоянного тока, например, к автомобильному разъему постоянного тока напряжением 12 В).
  • Кабель вызова медсестры (5 метров).
  • Блок выдоха, одно пользовательский (голубой).
  • Комбинированный воздушный фильтр, мелкопористый (6 штук в комплекте).
  • Встроенный аккумулятор.
  • Внешний аккумулятор.
  • Комплект для измерения концентрации кислорода.
  • Датчик FIO2 (концентрации кислорода).

Скачать инструкцию на Puritan Bennett 560

Скачать инструкцию по применению и другую документацию на аппарат искусственной вентиляции легких Medtonic (COVIDIEN) Puritan Bennett 560 можно здесь.

Руководство пользователя ( user manual ) на русском языке Puritan Bennett 560 скачать.

Декларация соответсвия Puritan Bennett 560 скачать.

Регистрационное удостоверение Puritan Bennett 560 скачать.

Респираторное обеспечение — каталог скачать.

Трахеостомические трубки и принадлежности — каталог скачать.

Кислородная и аэрозольная терапия — каталог скачать.

Также смотрите Puritan Bennett 840 аппарат ИВЛ.

User Manual

Puritan BennettTM

560 Ventilator

COVIDIEN, COVIDIEN with logo, and the Covidien logo and Positive Results for Life are U.S. and internationally registered trademarks of Covidien AG. ™* brands are trademarks of their respective owners. All other brands are trademarks of a Covidien company.

The information contained in this manual is the sole property of Covidien and may not be duplicated without permission. This manual may be revised or replaced by Covidien at any time and without notice. You should ensure that you have the most current applicable version of this manual; if in doubt, contact Covidien’s Technical Support department or visit the product manuals web page at:

www.medtronic.com/covidien/support/product-manuals

While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.

The ventilator should be operated and serviced only by trained professionals. Covidien’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided.

Nothing in this manual shall limit or restrict in any way Covidien’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Covidien has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.

To obtain information about a warranty, if any, contact Covidien Technical Services at 1 800 635 5267 or your local representative.

Purchase of this instrument confers no express or implied license under any Covidien patent to use the instrument with any ventilator that is not manufactured or licensed by Covidien.

Table of Contents

Preface

Purpose of This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . xi

Qualification of Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . xi

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . xi

Extended Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . xi

Service Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . xii

1

Safety Information

1.1

Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-1

1.2

Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-1

1.2.1

General Warnings Regarding Use of Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 1-1

1.2.2

Warnings Regarding Installation and Environment of Use . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-4

1.2.3

Warnings Regarding Electrical Power Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-7

1.2.4

Warnings Regarding Hoses and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-8

1.2.5

Warnings Regarding Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-11

1.2.6

Warnings Regarding PC Connection and USB Memory Devices . . . . . . . . . . . . . . . . . . . . . .

1-14

1.2.7

Warnings Regarding Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-14

1.2.8

Warnings Regarding Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-17

1.2.9

Warnings Regarding Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-19

1.3

Symbols and Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-19

1.4

Labels (Identification and Instruction Information) . . . . . . . . . . . . . . . . . . . . . . . . . .

1-24

2Ventilator Overview

2.1

Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-1

2.1.1

Target Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 2-1

2.1.2

Target Environments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-1

2.1.3

Target Operators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-2

2.2

Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-2

2.3

Operational Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-3

2.3.1

Safety Net . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-3

2.3.2

Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-3

2.3.3

Oxygen Enrichment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-3

2.3.4

Breathing Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-3

2.4

Device Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-4

2.5

Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-5

2.6

Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-6

2.7

Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-7

2.8

Ventilation Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-8

2.9

Alarm Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-9

2.10

Waveforms Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-10

Table of Contents

2.11

USB Memory Device Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-11

2.12

If Ventilator Failure Occurs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-11

3Alarms and Troubleshooting

3.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 3-1

3.2

Alarm Level of Priority . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

3.3

Alarm Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-3

3.4

Alarm Logs Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-4

3.5

Pausing the Audible Portion of Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

3.6

Pausing and Resetting Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

3.7

Reactivating Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8

3.8

Overview of Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-9

3.9

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-15

3.9.1

Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-15

3.9.2

Additional Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-25

4Installation and Assembly

4.1

Ventilator Startup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-1

4.2

Connecting to External AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-3

4.3

Connecting to an External DC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-6

4.4

Patient Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-8

4.4.1

Choosing the Patient Circuit Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-9

4.4.2

Installing the Patient Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-9

4.5

Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-16

4.5.1

Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-16

4.5.2

Bacteria Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-17

4.6

Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-18

4.7

Exhalation Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-19

4.8

Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-20

4.8.1

Administering Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-20

4.8.2

Connecting the Oxygen Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-21

4.8.3

Connecting the FiO2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-23

4.9

Using the Dual Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-25

4.9.1

Fitting the Ventilator into the Dual Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-25

4.9.2

Wearing the Dual Bag as a Backpack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-26

4.9.3

Securing the Ventilator on a Wheelchair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-26

4.9.4

Securing the Ventilator in a Personal Vehicle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-28

4.10

Mounting the Ventilator on a Utility Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-29

Table of Contents

5Operating Procedures

5.1

Turning on the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5-1

5.2

USB Menu Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5-4

5.2.1

USB Memory Device Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 5-5

5.2.2

USB Memory Device Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 5-5

5.3

Starting Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-9

5.4

Stopping Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-10

5.5

Turning Off the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-11

6Internal Battery

6.1

Battery Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-2

6.2

Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-3

6.3

Testing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-5

6.4

Recharging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-5

6.5

Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-6

7Cleaning

7.1

Cleaning the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-1

7.2

Cleaning the Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-2

7.3

Cleaning the Exhalation Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-3

7.4

Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-4

8Routine Maintenance

8.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-1

8.2

Expected Service Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-1

8.3

Calibrating the Exhalation Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-2

8.4

Calibrating the FiO2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-4

8.5

Replacing the Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-6

8.6

Recommended Schedule of Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 8-8

8.6.1

Preventive Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 8-8

8.6.2

Maintenance of the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8-10

8.6.3

Periodic Test of the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8-10

8.6.4

Replacement of the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8-10

8.7

Service Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8-11

ASpecifications

A.1

Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-1

A.2

Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-1

Table of Contents

A.3

Indicators and Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-2

A.4

Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-3

A.5

Monitored Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-3

A.6

Range, Resolution, and Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-4

A.7

Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-8

A.8

USB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-8

A.9

Pneumatic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. A-9

A.10

Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-10

A.11

Standards Compliance and IEC Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-13

A.11.1

General Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-13

A.11.2

Collateral Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-13

A.11.3

Particular Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-14

A.11.4

Air Transportation Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-14

BModes of Ventilation

B.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

B-1

B.2

Assist/Control (A/C) Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

B-1

B.3

SIMV Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

B-1

B.4

CPAP Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

B-2

B.5

PSV Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

B-2

COperational Verification Checklist

DUnpacking and Preparation

EAlarms Tests

E.1

Low Pressure Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. E-2

E.2

Max Leak Test (Only NIV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-3

E.3

Circuit Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-4

E.3.1

Accessing the Circuit Check Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. E-4

E.3.2

Performing the Circuit Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. E-5

E.3.3

Troubleshooting a Failed Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. E-7

E.3.4

Returning to Ventilation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. E-7

E.4

Power Failure Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-7

E.5

Occlusion Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-8

E.6

Battery Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-9

E.7

Involuntary Stop Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

E-10

F

Parts and Accessories

GGlossary

List of Figures

Figure 1-1. Locations of Labels—Top-Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25 Figure 1-2. Locations of Labels—Front-Left View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25 Figure 1-3. Location of Labels and Markings—Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26 Figure 1-4. Location of Labels—Bottom View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26 Figure 2-1. Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 Figure 2-2. Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Figure 2-3. Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 Figure 2-4. Ventilation Menu Display (on standby at left; during ventilation at right) . . . . . . . . . . . . . . . . . . 2-8 Figure 2-5. Alarm Menu (on standby at left; during ventilation at right) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9 Figure 2-6. Waveforms Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 Figure 2-7. USB Memory Device Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 Figure 3-1. Front Panel (Alarm Control Key). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 Figure 3-2. Alarm Messages (in Ventilation menu at left, in Alarm menu at right) . . . . . . . . . . . . . . . . . . . . . . 3-3 Figure 3-3. Accessing the Alarm Logs Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 Figure 3-4. Alarm Logs Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 Figure 3-5. Alarm Logs Screen (no alarm activated). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 Figure 3-6. Pausing the Audible Portion of Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6 Figure 3-7. Ventilator Screen (alarm paused indicator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7 Figure 3-8. Reactivating Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8 Figure 3-9. Alarm Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8 Figure 4-1. The Power Cable Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4 Figure 4-2. Inserting the Power Cable Holder into the Notch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4 Figure 4-3. Power Cable Connected to the Ventilator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Figure 4-4. Power Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Figure 4-5. Connecting the DC Power Cable to the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7 Figure 4-6. Connecting the Ventilator to an External DC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7 Figure 4-7. Single-Limb Patient Circuit With Exhalation Valve (including accessories). . . . . . . . . . . . . . . . . 4-10 Figure 4-8. Closeup of Exhalation Valve Tube and Proximal Pressure Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11 Figure 4-9. Double-Limb Patient Circuit (including accessories). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12 Figure 4-10. Closeup of Exhalation Valve Tube and Proximal Pressure Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 Figure 4-11. Close-up of Exhalation Bacteria Filter Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 Figure 4-12. Single-Limb Patient Circuit Without Exhalation Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 Figure 4-13. Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16 Figure 4-14. Bacteria Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17 Figure 4-15. Humidifier. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18 Figure 4-16. Removing the Exhalation Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19 Figure 4-17. Rear Panel Oxygen Inlet Port and Coupler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21 Figure 4-18. Connecting the Oxygen Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22 Figure 4-19. Disconnecting the Oxygen Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23 Figure 4-20. Connecting the FiO2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24 Figure 4-21. Using the Dual Bag as a Backpack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26 Figure 4-22. Using the Dual Bag on a Wheelchair (with double-limb circuit on left; with

single-limb circuit on right) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27 Figure 4-23. Using the Dual Bag in a Personal Vehicle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28

List of Figures

Figure 4-24. Mounting the Ventilator on the Utility Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29 Figure 4-25. Securing the Ventilator on the Utility Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29 Figure 4-26. Puritan Bennett™ 560 Ventilator Mounted on Utility Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30 Figure 5-1. Turning on the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Figure 5-2. VENTILATION ON/OFF Button and Standby Indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Figure 5-3. Welcome Menu Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Figure 5-4. Ventilation Menu Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 Figure 5-5. Selecting the USB Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 Figure 5-6. Selecting Transfer Continuously . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6 Figure 5-7. Selecting Transfer Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7 Figure 5-8. Prompt to Start Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9 Figure 5-9. Starting Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10 Figure 5-10. Stopping Ventilation (1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10 Figure 5-11. Stopping Ventilation (2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11 Figure 6-1. Internal Battery Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3 Figure 6-2. Battery Reserve Capacity as a Percentage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4 Figure 6-3. Battery Reserve Capacity in Hours and Minutes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4 Figure 6-4. Power Indicators when Charging the Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5 Figure 7-1. Removing the Exhalation Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 Figure 7-2. Puritan Bennett ™ 560 Ventilator Pneumatic Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Figure 8-1. Blocking the Patient Circuit (single-limb circuit at left; double-limb circuit at right) . . . . . . . . 8-2 Figure 8-2. Calibrating the Exhalation Flow Sensor (1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3 Figure 8-3. Calibrating the Exhalation Flow Sensor (2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3 Figure 8-4. Calibrating the Exhalation Flow Sensor (3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3 Figure 8-5. Calibrating the FiO2 Sensor (1). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Figure 8-6. Calibrating the FiO2 Sensor (2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Figure 8-7. Calibrating the FiO2 Sensor (3). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6 Figure 8-8. Replacing the Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7 Figure E-1. Ventilator Screen (Patient Disconnection alarm shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2 Figure E-2. Ventilator Screen (High Leakage alarm shown). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3 Figure E-3. Circuit Check Screen (before starting). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4 Figure E-4. Blocking the Patient Circuit (single-limb circuit at left; double-limb circuit at right) . . . . . . . . E-5 Figure E-5. Circuit Check (running) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5 Figure E-6. Circuit Check (complete, passed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6 Figure E-7. Circuit Check (complete, failed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6 Figure E-8. Ventilator Screen (AC Power Disconnection alarm shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-8 Figure E-9. Blocking the Patient Circuit (single-limb circuit at left; double-limb circuit at right) . . . . . . . . E-8 Figure E-10. Ventilator Screen (Occlusion alarm shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9 Figure E-11. Ventilator Screen (AC Power Disconnection alarm shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-10

List of Tables

Table 1-1. Ventilator Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19 Table 1-2. Ventilator Labels and Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24 Table 3-1. Overview of Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 Table 3-2. Alarms and Corrective Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16 Table 3-3. Additional Troubleshooting and Corrective Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25 Table 5-1. USB Memory Device Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 Table 5-2. Ventilator to USB Device Data Transfer Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8 Table 6-1. Internal Battery Reserve Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 Table 7-1. Approved Cleaning Solutions for Exterior Ventilator Surfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2 Table 8-1. Preventive Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8 Table A-1. Physical Description (excluding accessories) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 Table A-2. Electrical Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 Table A-3. Internal Lithium Ion Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 Table A-4. Power Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 Table A-5. Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 Table A-6. Audio Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 Table A-7. Performance Parameter Specifications and Tolerances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3 Table A-8. Monitored Parameter Tolerances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3 Table A-9. Ventilator Range, Resolution, and Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4 Table A-10. Environmental Conditions for Storage or Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8 Table A-11. Environmental Conditions for Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8 Table A-12. USB Memory Device Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8 Table A-13. Data Transfer Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8 Table A-14. Airway Resistances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 Table A-15. Patient Circuit Resistances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 Table A-16. Air Inlet Resistance (Filter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 Table A-17. Oxygen Inlet Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 Table A-18. Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 Table A-19. Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11 Table A-20. Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11 Table A-21. Electromagnetic Immunity—Conducted and Radiated RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12 Table A-22. Compliant Cables and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12 Table C-1. Operational Verification Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 Table D-1. Items Included with Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1 Table F-1. List of Consumables and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1 Table F-2. List of Circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2

Page Left Intentionally Blank

x

Preface

Purpose of This Manual

This manual contains important information regarding the safe operation of your Puritan Bennett™ 560 ventilator. Your ventilator is an electrical device that can provide years of useful service with the proper care, as described in this manual.

Ensure that you read and understand the instructions contained in this manual before operating the ventilator.

<WARNING:

Before operating the ventilator, read, understand, and strictly follow the information contained in Chapter 1, Safety Information.

Qualification of Personnel

Installation and maintenance of the device must be made by authorized and trained personnel. In particular, training for the handling of products sensitive to electrostatic discharges must include the use of electrostatic discharge (ESD) protection devices and knowledge of the meaning of the symbol at left, as well as using original spare parts and respecting quality assurance and traceability rules approved by Covidien.

Warranty

Information regarding your product warranty is available from your sales representative or Covidien.

Extended Service

The Puritan Bennett™ 560 ventilator offers extended service contracts/warranties for purchase when the ventilator is purchased. Please contact your local Covidien sales or service representative for additional information.

xi

Preface

For online technical support, visit the SolvITSM Center Knowledge Base by clicking the link at www.medtronic.com/covidien/support/solvit- center-knowledge-base/. Here, you will find answers to frequently asked questions about the product and other Covidien products 24 hours a day, 7 days a week. If you require further assistance, contact your local Covidien representative.

Service Centers

Covidien Argentina

Covidien Asia

Covidien Australia

Covidien Austria GmbH

Pacheco Trade Center

Singapore Regional Service

52A Huntingwood Drive

C-Mill Gebouw K

Marcos Sastre 1990, El Talar,

Centre

Huntingwood, NSW 2148

Jan Campertstraat 21-A

Buenos Aires, Argentina,

15 Pioneer Hub, #06-04

Australia

6416 SG Heerlen, Netherlands

B1610CRH

Singapore 627753

[T] +61 1800 350702

[T] 01 20609 1143

[T] +54 01157898107

[T] +65 6578 5288

[F] +61 2967 18118

[F] 01 20609 2457

[F] +54 114863 4142

[F] +65 6515 5260

[E] techservices.csAustria@

medtronic.com

Covidien Belgium BVBA/SPRL

Covidien Brazil

Covidien Canada

Covidien Chile

C-Mill Gebouw K

Av. Das Nações Undias 12995

19600 Clark Graham

Camino lo Boza (Ex 8395) Pude-

Jan Campertstraat 21-A

Andar 23 — Brooklin

Baie d’Urfe, QC, H9X 3R8

huel

6416 SG Heerlen, Netherlands

São Paulo, SP

Canada

Santiago

[T] 0220 08260

Brasil 04578-000

[T] 514 695 1220 (ext 4004)

Chile

[F] 0270 06690

[T] +5511 2187 6200

[F] 514 695 4965

[T] +562 739 3000

[E]techservices.csBel-

[F] +5511 2187 6380

[F] +562 783 3149

gium@medtronic.com

Covidien China

Covidien Colombia

Covidien Costa Rica

Covidien ECE

2F, Tyco Plaza

Avenida Calle 116 # 7-15 /

Global Park, Parkway 50

Organizačni Složka

99 Tian Zhou Rd

Oficina 1101

La Aurora de Heredia

Prosecká 852/66

Shang Hai 200233

Torre Cusezar, Bogotá, Colombia

Costa Rica

190 00 Praha 9

P.R. China

[T] +57 1 7427300

[T] +506 2239 5386

Czech Republic

[T] +86 4008 1886 86

[F] +571 619 5425

[F] +506 2239 5319

[T] +420 241 095 735

[F] +86 2154 4511 18

[F] +420 239 016 856

Covidien Danmark A/S

Covidien Deutschland GmbH

Covidien ECE

Covidien Finland OY

C-Mill Gebouw K

C-Mill Gebouw K

Galvahiho 7 / A

C-Mill Gebouw K

Jan Campertstraat 21-A

Jan Campertstraat 21-A

832104 Bratislava

Jan Campertstraat 21-A

6416 SG Heerlen, Netherlands

6416 SG Heerlen, Netherlands

Slovakia

6416 SG Heerlen, Netherlands

[T] +45 43 68 21 71

[T] +49 6951709670

[T] +421 2 4821 4573

[T] +358 972519288

[F] +45 43 31 48 99

[F] +49 69299571608

[F] +421 2 4821 4501

[F] +358 972522072

[E]techservices.csDenmark@

[E] techservices.csGermany@

[E] techservices.csFinland@

medtronic.com

medtronic.com

medtronic.com

Service Centers

Covidien France SAS

Covidien Hong Kong

Covidien ECE s.r.o.

Covidien India

C-Mill Gebouw K

Unit 12 — 16, 18/F

Magyarországi Fióktelepe

10th Floor Building No 9B

Jan Campertstraat 21-A

BEA Tower

Mariássy u.7.

DLF Cyber City Phase III Gurgaon

6416 SG Heerlen, Netherlands

Millennium City 5

1095 Budapest

Haryana — 122002

[T] +33 151 323 510

4187 Kwun Tong Road

Hungary

India

[F] +33 157 327 010

Kwum Tong, Kowloon,

[T] +36 1 880 7975

[T] +91 1 244 709800

[E] techservices.csFrance@

Hong Kong

[F] +36 1 778 9459

[F] +91 1 244 206850

medtronic.com

[T] +852 3157 7299

[F] +852 2838 0749

Covidien Ireland

Covidien Israel

Covidien Italia S.p.A

Covidien Japan Inc.

C-Mill Gebouw K

3 HaCarmel Street, Kochav

C-Mill Gebouw K

Technical Support Center

Jan Campertstraat 21-A

Yokneam Building

Jan Campertstraat 21-A

83-1, Takashimadaira 1-Chome

6416 SG Heerlen, Netherlands

Yokneam Elit

6416 SG Heerlen, Netherlands

Itabashi-ku, Tokyo

[T] +353 0 1 4073173

Israel 20692

[T] +39 02 91 483320

175-0082 Japan

[F] +353 0 1 9075668

[T] +972 4 6309423

[F] +39 02 91 294863

[T] +81 0 3 6859 0120

[E] techservices.csIreland@

[F] +97 2774704093

[E} techservices.csItaly@

[F] +81 0 3 6859 0142

medtronic.com

medtronic.com

[E] service.repair.israel@

medtronic.com

Covidien Korea

Covidien Mexico

Covidien Nederland BV

Covidien New Zealand

5F, Hibrand Living Gwan, #215,

Autopista México-Querétaro

C-Mill Gebouw K

Cnr Manu Tapu Dr & Joseph

Yangjae-Dong, Seocho-Gu

KM 34.5 Nave 3 Cortina 113

Jan Campertstraat 21-A

Hammond Pl.

Cuautitlán Izcalli 54740

Seoul, Korea

6416 SG Heerlen, Netherlands

Auckland Airport

[T] +822 570 5459

México, Estado de México

[T] +31 202061470

New Zealand

[F] +822 570 5499

[T] 5255 5804 1524 (ext. 1410)

[F] +31 707709229

[T] +64 508 489 264

[F] 5255 5536 1326

[E] techservices.csItaly@

medtronic.com

Covidien Norge AS

Covidien Panama

Covidien Polska

Covidien Portugal Lda.

C-Mill Gebouw K

Parque Industrial Costa del Esta

C-Mill Gebouw K

C-Mill Gebouw K

Jan Campertstraat 21-A

Calle Primera, Edificio # 109

Jan Campertstraat 21-A

Jan Campertstraat 21-A

6416 SG Heerlen, Netherlands

Panama City, Panama

6416 SG Heerlen, Netherlands

6416 SG Heerlen, Netherlands

[T] +47 24159887

[T] +507 264 7337

[T] +48 223060034

[T] +351 21 761 62 44

[F] +47 23024955

[F] +507 236 7408

[F] +48 223060853

[F] +351 800 781385

[E] techservices.csNorway@

[E] techservices.csPoland@

[E] techservices.csPortugal@

medtronic.com

medtronic.com

medtronic.com

Covidien Puerto Rico

Covidien Russia

Covidien Saglik A.S.

Covidien South Africa

Palmas Industrial Park

Tupikovy proezd, Building 1,

LTD.ŞTI

Waterfall Distribution Campus

Road 869 Km 2.0 Bdlg. #1

Marushinsky, Krekshino village

Akçaburgaz Mah. 1567 Sok. No. 2

Cnr Bridal Veil Road & K101 Pre-

Cataño, PR 00962

Moscow, Russia

DHL Depolama Tesisleri

toria Main Road

[T] 787 993 7250 (ext. 7221/22)

[T] +7 495 995 1898

Esenyurt Istanbul, Turkey

Midrand

[F] 787 993 7234

[F] +7 495 933 6468

[T] +90 212 6223 500

South Africa 1685

[E] service.repair.russia@

[F] +90 212 6720 722

[T] +27 11 542 9584

medtronic.com

[E] service.repair.turkey@

[F] +27 86 604 8360

medtronic.com

[E]service.repair.southafri-

ca@medtronic.com

Preface

Covidien Spain S.L.

Covidien Sverige AB

Covidien Switzerland

Covidien Thailand

C-Mill Gebouw K

C-Mill Gebouw K

C-Mill Gebouw K

99 Soi Rubia

Jan Campertstraat 21-A

Jan Campertstraat 21-A

Jan Campertstraat 21-A

Sukhumvit 42 Road

6416 SG Heerlen, Netherlands

6416 SG Heerlen, Netherlands

6416 SG Heerlen, Netherlands

13-14 Fl., Berli Jucker Building

[T] +34 91 275 48 54

[T] +46 8517 61573

[T] +41 44 511 82 71

Prakanong, Klongtoey

[F] +34 91 276 89 33

[F] +46 8502 52110

[F] +41 44 511 16 34

Bangkok 10110, Thailand

[E] techservices.csSpain@

[E] techservices.csSweden@

[E] techservices.csSwitzer-

[T] +66 2 207 3100

medtronic.com

medtronic.com

land@medtronic.com

[F] +66 2 657 6325

Covidien UK

C-Mill Gebouw K

Jan Campertstraat 21-A

6416 SG Heerlen, Netherlands

[T] +44 0 2030271757

[F] +44 0 2036848869

[E] techservices.csUK@medtronic.com

1Safety Information

1.1Definitions

This manual uses three indicators to highlight critical information: warning, caution, and note. They are defined as follows:

WARNING

Indicates a condition that can endanger the patient or the ventilator operator.

Caution

Indicates a condition that can damage the equipment.

Note

Indicates points of particular emphasis, that make operation of the ventilator more efficient or convenient.

It is essential to read, understand and follow these instructions before using the Puritan Bennett™ 560 ventilator.

In order to use the ventilator correctly and efficiently and to help prevent incidents, please pay particular attention to section 1.2, Warnings, as well as all warnings and cautions contained throughout this manual.

:Note:

Many ventilator functions are not accessible when the Locking key is enabled. For additional assistance contact your clinician or equipment representative.

1.2Warnings

1.2.1General Warnings Regarding Use of Equipment

<WARNING:

The ventilator must be used only under the responsibility and on the prescription of a doctor.

1-1

Safety Information

<WARNING:

The ventilator must be used according to its intended use. Refer to section 2.1, Indications for Use.

<WARNING:

Be aware this manual describes how to respond to the ventilator, but does not tell you how to respond to the patient.

<WARNING:

While the ventilator is in use, an alternative means of ventilation should always be available in the event of a ventilator problem. This is particularly true for ventilator-dependent patients. Supplementary observation, appropriate for the patient’s condition, is also recommended.

<WARNING:

To ensure that ventilation continues uninterrupted, ensure alternative power sources are available (AC power source, extra batteries, or an auxiliary DC car adapter). Be prepared for the possibility of power failure by having an alternative means of ventilation ready for use—particularly for ventilatordependent patients.

<WARNING:

Do not allow a patient to remain connected to the ventilator when ventilation is stopped, because a substantial quantity of exhalation gas, primarily carbon dioxide, may be inhaled by the patient. In some circumstances, inhaling carbon dioxide may lead to under-ventilation, suffocation, and serious injury or death.

<WARNING:

Always have immediate access to an alternative means of ventilation, which is ready for use, to avoid patient death or serious injury.

<WARNING:

The ventilator must not be used with flammable anesthetic substances.

<WARNING:

Do not start ventilation until you ensure that the device is suitably assembled, that the air inlet filter is properly installed and is not obstructed, and that there is proper clearance all around the unit. Also ensure that the patient circuit is suitably connected to both the ventilator and the patient and that the patient circuit, including all hoses, is not damaged or obstructed.

<WARNING:

A ventilator-dependent patient should always be monitored by trained and competent medical personnel. Ensure that the patient’s caregiver is able and prepared to take suitable action in the event the ventilator identifies an alarmed condition or experiences a problem.

<WARNING:

Do not use a patient circuit with a leak accessory for ventilator-dependent patients.

Warnings

<WARNING:

Refer to this manual for equipment compatible with this ventilator. It may be unsafe to interconnect this equipment with other equipment not described in this manual.

<WARNING:

Before dispensing the ventilator to caregivers or the patient for home use, ensure the Locking Key is activated so that critical ventilator settings are not modified.

<WARNING:

Do not perform ventilator alarm tests while the patient is connected to the ventilator. Provide the patient with an alternate means of ventilation before conducting these tests.

<WARNING:

Verify the functionality of the alarms before connecting the patient to the ventilator. Refer to Appendix

E, Alarms Tests.

<WARNING:

If the ventilator fails the alarm tests or if you cannot complete the tests, refer to Chapter 3, Alarms and Troubleshooting or call your equipment supplier or Covidien.

<WARNING:

When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator.

<WARNING:

A continuous alarm condition will be activated if the ventilator power switch is turned off while ventilation is in progress. When the power switch is turned back on again, ventilation will resume without having to press the VENTILATION ON/OFF button.

<WARNING:

To reduce the risk of infection, wash your hands thoroughly before and after handling the ventilator or its accessories.

<WARNING:

A patient treated by mechanical ventilation is highly vulnerable to the risks of infection. Dirty or contaminated equipment is a potential source of infection. Clean the ventilator and its accessories regularly and systematically before and after each use and following any maintenance procedure to reduce the risks of infection. The use of a bacterial filter at the ventilator’s outlet (TO PATIENT) port—or both ports if a double-limb circuit is used—is recommended. Refer to Chapter 7, Cleaning.

<WARNING:

Handle the ventilator with care during and after use, particularly when ambient temperatures are high. Some ventilator surfaces may become hot, even if safety specifications are not exceeded.

Safety Information

<WARNING:

Do not connect the ventilator to any device other than a PC with a dedicated compatible Puritan Bennett™ software package.

<WARNING:

The ventilator system is not intended to be a comprehensive monitoring device and does not activate alarms for all types of conditions. For a detailed understanding of ventilator operations, be sure to thoroughly read this manual before attempting to use the ventilator system.

1.2.2Warnings Regarding Installation and Environment of Use

<WARNING:

Even though the Puritan Bennett™ 560 ventilator meets current safety standards, the internal Lithiumion battery of the device exceeds the 100Wh threshold and is therefore considered to be Dangerous Goods (DG) Class 9 – Miscellaneous, when transported in commerce. As such, the Puritan Bennett™ 560 ventilator and/or the associated Lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International Air Transport Association), International Maritime Dangerous Goods code for sea and the European Agreement concerning the International Carriage of Dangerous Goods by Road (ADR) for Europe. Private individuals who transport the device are excluded from these regulations although for air transport some requirements apply. For air transport; the Puritan Bennett™ 560 ventilator is permitted as checked-in or carry-on baggage. Two spare batteries per person may be taken on board as carry-on luggage only, with the prior approval of the airline. This classification and regulatory requirements may vary depending upon the country and mode of transport. Therefore it is recommended that users verify with the carrier / airline as to which measures to take before the voyage.

<WARNING:

To minimize the risk of damage, you must use the ventilator’s dual bag to transport the ventilator. Ventilator accessories are listed in Table F-1.

<WARNING:

When using the ventilator in a carrying case, only use a carrying case that is listed in the instructions for use to prevent adverse ventilator performance, which can consequently result in patient death.

<WARNING:

Regularly clean the ventilator’s dual bag according to manufacturer’s recommendations.

<WARNING:

The ventilator should never be immersed in any liquid, and any liquid on the surface of the device should be wiped away immediately.

Warnings

<WARNING:

To avoid damage to the ventilator, in particular the batteries or electrical components, fluids must not be allowed to enter the device, particularly through the air inlet filter or the cooling apertures located in the side, rear, and bottom panels of the ventilator.

<WARNING:

To ensure correct and lasting operation of the device, ensure that the ventilator is installed and operated in the environmental conditions recommended in Appendix A, Specifications.

<WARNING:

Do not leave power cables lying on the ground where they may pose a hazard.

<WARNING:

Do not operate the ventilator in a magnetic resonance imaging (MRI) environment. Doing so could cause a ventilator malfunction.

<WARNING:

Do not operate the ventilator in the presence of active high frequency (HF) surgical equipment. Doing so could cause a ventilator malfunction.

<WARNING:

Do not operate the ventilator in direct sunlight, near heat sources, outdoors, or near installations where liquid may pose a risk without first providing adequate protection for the device.

<WARNING:

Avoid using the ventilator, if possible, in dusty environments. Dusty environments may require more vigilant monitoring, cleaning, and/or replacement of air intake and other filters.

<WARNING:

Ensure that the ventilator’s immediate surroundings allow for the proper operational connection of the device without folding, pinching, or damaging any of the required cables or tubes, and that the connection of the patient circuit to the patient provides for a secure, comfortable fit.

<WARNING:

Ensure that the ventilator is not positioned or located such that the AC and DC connections at the back of the ventilator are difficult to access.

<WARNING:

Do not cover the ventilator or place in a position that affects proper operation, e.g., blocking a front or lateral opening.

Safety Information

<WARNING:

Place the ventilator in a safe place when ventilating and according to the recommendations in this manual.

<WARNING:

Do not place the ventilator in a position where a child, pet or pest can reach it, or in any position that might cause it to fall on the patient or someone else.

<WARNING:

To ensure correct and lasting operation of the ventilator, ensure that its air circulation holes (main inlet or cooling) are never obstructed. Place the device in an area where air can freely circulate around the ventilator and avoid installing it near floating fabrics, such as curtains.

<WARNING:

If the ventilator has been transported or stored at a temperature that differs more than ±20°C (±36°F) from the temperature in which it will be operating, the ventilator should be allowed to stabilize in its operating environment for at least 2 hours prior to use. When the ambient temperature is 20°C, 2 hours are required to warm the ventilator from the minimum storage temperature or to cool the ventilator from the maximum storage temperature prior to use.

<WARNING:

If the ambient temperature where the device is operated is greater than 35°C (95°F), the temperature of the patient circuit or the flow supplied at the device outlet may exceed 41°C (106°F), and the patient circuit may reach up to 60°C (140°F). This may lead to undesirable side effects for the patient. To avoid injury to the patient move the patient and the ventilator to a cooler location. For more information, contact Covidien.

<WARNING:

The default setting for altitude compensation is YES. Altitude compensation should always be set to YES for accurate volume delivery calculations at all elevations.

<WARNING:

To reduce the risk of a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (such as flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses.

<WARNING:

Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator. If necessary, replace the filter before the recommended replacement period is over (see Chapter 8, Routine Maintenance). This is particularly important when the ventilator is installed on a wheelchair, because environmental conditions may cause the filter to become dirty more rapidly.

<WARNING:

Handle the ventilator with care during and after use, particularly when ambient temperatures are high. Some ventilator surfaces may become hot, even if safety specifications are not exceeded.

Warnings

<WARNING:

Exercise care to avoid any potential significant risks of reciprocal interference posed by the ventilator and accessories during specific investigations or treatments.

1.2.3Warnings Regarding Electrical Power Supplies

<WARNING:

The operator should connect the ventilator to an AC power source whenever available, for safer operation.

<WARNING:

The maximum recommended shelf life of the internal battery is 2 years. Do not use a battery that has been stored for 2 years prior to its first use.

<WARNING:

Periodic recharging is important to help maximize useful life of the battery. Do not store the internal battery for extended periods, without recharging, as this may reduce the maximum life.

<WARNING:

For the AC (“mains”) power cable to be properly secured, the attachment located on the power cable must be fitted into the power cable holder incorporated in the battery access cover and located under the AC (mains) power socket. Refer to section 4.2, Connecting to External AC Power.

<WARNING:

The power supply to which the ventilator is connected (both AC and DC) must comply with all applicable standards and provide electrical power corresponding to the voltage characteristics inscribed on the rear of the ventilator to ensure correct operation. Refer also to the electrical specifications found in Appendix A, Specifications.

<WARNING:

Ensure that the ventilator’s internal battery is fully charged before connecting the ventilator to an external DC power source. Powering the ventilator using an external 12–30 VDC power source (via the DC power cable) does not enable its internal battery to recharge.

<WARNING:

Due to the internal battery’s limited reserve capacity, the ventilator should only be operated on the internal battery when no other power source is available. Ensure that the internal battery never becomes fully discharged.

<WARNING:

When using a car auxiliary adapter (cigarette lighter) ensure the car has been started prior to plugging in the ventilator’s DC adapter. Refer to section 4.3, Connecting to an External DC Power Source.

Safety Information

<WARNING:

Even if the internal battery charging indicator is off, charging of the battery may sometimes be incomplete if the ambient temperature is above 40°C (104°F) because of the battery’s internal heat safety device.

<WARNING:

When the Low Battery alarm is triggered, immediately connect the ventilator to an AC power supply to maintain ventilation and recharge the internal battery.

<WARNING:

Batteries should be disposed of according to environmental legislation in your country and locality.

<WARNING:

Never expose any batteries to direct flame.

<WARNING:

Ensure that the AC power cable is in perfect condition and not compressed. The device should not be turned on if the AC power cable is damaged.

1.2.4Warnings Regarding Hoses and Accessories

<WARNING:

The ventilator must not use, nor be connected to, any anti-static or electrically conductive hoses, tubing, or conduits.

<WARNING:

Minimum and maximum VTE alarm parameters must be properly set to warn in the event of patient disconnection.

<WARNING:

Before opening the packaging for the patient circuit, ensure that no damage is evident to the packaging or its contents. Do not use if evidence of damage exists.

<WARNING:

The patient circuit should not be changed during ventilation.

<WARNING:

On a DAILY basis, inspect the patient circuit to ensure that it shows no signs of damage, is properly connected, and is operating correctly without leakage.

<WARNING:

Single use accessories should not be reused.

Warnings

<WARNING:

The exhalation block is intended for single use by a single patient . It may periodically be cleaned, but it cannot be disinfected or sterilized. To maintain good measurement quality when used continuously, clean the exhalation block periodically (refer to section 7.3, Cleaning the Exhalation Block). The exhalation block should be changed every 4 months and cannot be reused with any other patient.

<WARNING:

During invasive ventilation (when an artificial airway bypasses the patient’s upper respiratory system), the patient’s upper respiratory system cannot humidify the incoming gas. For this reason, a humidifier, to minimize drying of the patient’s airway and subsequent irritation and discomfort, must be used.

<WARNING:

If exhaled tidal volume measurements are required to ensure correct patient ventilation a double-limb patient circuit configuration must be used in order to detect leaks. In this case, both the minimum and maximum VTE alarm parameters must be properly set to warn in the event of patient disconnection.

<WARNING:

Failing to replace a dirty air inlet filter or operating the ventilator without a filter may cause serious damage to the ventilator.

<WARNING:

Before cleaning the ventilator, first disconnect the ventilator and the patient circuit.

<WARNING:

If the ventilator is used indoors, the condition of the air inlet filter should be checked monthly. If the ventilator is used outdoors or in a dusty environment, the filter should be checked weekly and replaced as necessary.

<WARNING:

The air inlet filter is not reusable; do not attempt to wash, clean, or reuse it.

<WARNING:

The patient circuit should always be positioned to avoid hindering the patient’s movements, to prevent accidental disconnection or leakage, and to minimize the risk of patient strangulation.

<WARNING:

For pediatric use, ensure that the patient circuit type fits, and, in all respects, is suitable for use with a child. Use a pediatric circuit for patients that weigh under 53 lb. (23 kg). To ensure proper performance of the ventilator, use a recommended patient circuit; see Table F-2.

<WARNING:

Resistance of the exhalation valve and accessories (water traps, filters, HMEs, etc.) must be as low as possible.

Safety Information

<WARNING:

Adding attachments to the ventilator breathing system can cause the pressure during exhalation at the patient connection port to increase.

<WARNING:

The exhalation valve must allow rapid discharge of the circuit pressure. Ensure that the exhalation valve is always clean and its evacuation aperture (exhaust port) is never obstructed.

<WARNING:

Users must always possess an additional breathing circuit and exhalation valve while using the Puritan Bennett™ 560 ventilator.

<WARNING:

Always ensure that the humidification device is positioned lower than both the ventilator and the patient. Use water traps, if necessary, to limit water in the patient circuit and periodically empty these water traps. Take precautions when discarding the fluid in the water trap. Discard per local ordinance for proper disposal.

<WARNING:

Use of a nebulizer or humidifier can lead to an increase in the resistance of inspiratory and exhalation filters. Monitor the filters frequently for increased resistance or blockage.

<WARNING:

If a heated humidifier is used, you should always monitor the temperature of the gas delivered to the patient. Gas delivered from the ventilator that becomes too hot may burn the patient’s airway.

<WARNING:

Adding accessories to the ventilator breathing circuit, such as a humidifier and water trap(s), may result in a decrease in tidal volume delivered to the patient due to the added compressible volume of the accessory. Always assure that the patient is receiving the appropriate inspired volume when altering the breathing circuit configuration.

<WARNING:

The level of inspiratory resistance of the circuit and accessories (bacteria filter, humidifier, HMEs, etc.) must be as low as possible. Settings—particularly the Patient Disconnection alarm, maximum inspired volume (Max VTI), and minimum inspired volume (Min VTI) settings—must be periodically adjusted according to changes in the patient circuit resistance—especially when filters are replaced.

<WARNING:

To ensure proper performance of the ventilator, use a patient circuit recommended by Covidien in this manual; refer to Chapter 4, Installation and Assembly and Appendix F, Parts and Accessories. The total specified length of the patient circuit tubing as measured from the ventilator outlet to the ventilator inlet is 1.1 meters (3.6 feet) to 2.0 meters (6.6 feet). The tubing must conform to all applicable standards and must be fitted with Ø 22 mm terminals that also conform to all applicable standards. Ensure that

Warnings

both the length and the internal volume of the patient circuit are appropriate for the tidal volume: a corrugated tube of Ø 22 mm for adult patients, and a corrugated tube of Ø 15 mm for pediatric patients with a tidal volume lower than 200 ml.

<WARNING:

To ensure proper performance of the ventilator, use only accessories (including oxygen accessories) approved and recommended by Covidien. See Appendix F, Parts and Accessories or contact your customer services.

<WARNING:

To reduce the likelihood of disconnection and to prevent adverse ventilator performance, use only accessories compatible with the ventilator. Compatibility is determined by reviewing the instructions for use of either the ventilator or the accessories.

<WARNING:

When using non-invasive ventilation (NIV) without an exhalation valve, use a vented nose or face mask or a non vented combined with a leak accessory. When using non-invasive ventilation (NIV) with an exhalation valve, use a non-vented mask.

1.2.5Warnings Regarding Settings

<WARNING:

Before starting ventilation, always verify that all settings are properly set in accordance with the required prescription.

<WARNING:

Before starting ventilation, ensure that the device is properly assembled and that the air inlet, cooling vents, and alarm sound diffusion holes are not obstructed. Ensure also that the patient circuit is of the proper configuration (double or single limb), properly connected to the ventilator, and that the circuit hoses are neither damaged nor compressed and contain no obstructions or foreign bodies.

<WARNING:

The CPAP mode does not provide a set respiratory rate. Do not use this mode for ventilator-dependent patients.

<WARNING:

Do not allow a patient to remain connected to the ventilator when ventilation is stopped, because a substantial quantity of exhalation gas, primarily carbon dioxide, may be inhaled by the patient.

<WARNING:

Alarm volume should be adjusted with respect to the ventilator’s operating environment and so that the patient’s caretakers can hear the alarms. The audible alarm vents located at the front of the device should never be obstructed. The alarm can be paused with the Alarm Pause function by pressing the ALARM CONTROL key twice once the alarm has been declared.

Safety Information

<WARNING:

Ensure that the I Sens setting is not set to OFF when ventilating patients capable of triggering spontaneous breaths.

<WARNING:

The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient, based on clinical judgment, the condition and needs of the patient, and the benefits, limitations, and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether those are best for the patient’s current needs.

<WARNING:

In adult or pediatric use ensure that the adjusted tidal volume is compatible with the needs of the patient.

<WARNING:

When changing the mode during ventilation, significant transitions of pressure, flow or cycling rate might occur, depending on the difference between the modes. Before setting the new mode, first ensure that the settings between the different modes are compatible. This reduces the risk of discomfort and harm to the patient.

<WARNING:

Do not conduct the ventilator alarm test while the patient is connected to the ventilator. Switch the patient to an alternate means of ventilation before testing.

<WARNING:

The Min PIP alarm setting must be adjusted for the patient, but must also be set high enough to allow the Patient Disconnection alarm to trigger properly. Perform the low pressure test to ensure the alarm is properly set.

<WARNING:

The Max Leak alarm setting must be adjusted for the patient, but must also be set low enough to allow the High Leakage alarm to trigger properly. Perform the max leak test to ensure the alarm is functioning properly. This alarm only applies to leak configuration (NIV).

<WARNING:

If Apnea Time is set to a value higher than 60/Control R then the Apnea alarm will not activate.

<WARNING:

If an Apnea alarm is required, set the Apnea setting to YES in the Preferences Menu.

Warnings

<WARNING:

The Apnea alarm should be set to YES for ventilator dependent patients.

<WARNING:

Setting any alarm limits to OFF or extreme high or low values can cause the associated alarm not to activate during ventilation, which reduces its efficacy for monitoring the patient and alerting the clinician to situations that may require intervention.

<WARNING:

Ensure the Insp Time setting is compatible with the physiological requirements of the patient.

<WARNING:

Adjustable alarms should not be systematically canceled; instead, they should be adjusted according to the needs and condition of the patient.

<WARNING:

Do not pause, disable, or decrease the volume of the ventilator’s audible alarm if patient safety could be compromised.

<WARNING:

A continuous alarm condition will be activated if the ventilator power switch is turned off while ventilation is in progress. When the power switch is turned back on again, the ventilation will resume without having to press the VENTILATION ON/OFF button.

<WARNING:

In the SIMV mode the use of a double-limb circuit is recommended. The Min VTE setting should remain active in the event that pressure losses are present on the patient circuit downstream from the proximal pressure link. In such cases the Patient Disconnection alarm would not be systematically activated in case of a disconnection of the circuit.

<WARNING:

The inspiration trigger threshold should be carefully modified in order to avoid the risk of false triggering or “autotriggering” of the ventilator. For example, Level 0P, the most sensitive mode, is recommended for pediatric use. However, for an adult, this setting may result in autotriggering.

<WARNING:

The sound level of the alarms should be adjusted according to the installation environment and the size of the area monitored by the patient’s caregiver. Ensure that the alarm sound apertures at the front of the device are never obstructed.

Safety Information

1.2.6Warnings Regarding PC Connection and USB Memory Devices

<WARNING:

Do not connect the ventilator to any device other than a PC with a dedicated compatible Puritan Bennett™ software package.

<WARNING:

Always verify the file ID before using a USB memory device to transfer data between the ventilator and a PC.

<WARNING:

USB connections are not intended for connection to any devices other than the specified USB flash storage (see section 5.2.1, USB Memory Device Specifications).

1.2.7Warnings Regarding Maintenance

<WARNING:

Never use a ventilator or any components or accessories that appear to be damaged. If any signs of damage are evident, contact your equipment supplier or Covidien.

<WARNING:

To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only personnel authorized and qualified by Covidien should attempt to service or make authorized modifications to the Puritan Bennett™ 560 ventilator.

<WARNING:

If you cannot determine the cause of a problem with your ventilator, contact your equipment supplier. Do not use the ventilator until the problem has been corrected.

<WARNING:

To ensure proper performance of the ventilator, the preventative maintenance schedule should be followed. For further information contact Covidien.

<WARNING:

On a daily basis, ensure the proper connection and operation of the patient circuit.

<WARNING:

If a problem with the ventilator is suspected, FIRST CHECK THAT THE PATIENT IS NOT IN DANGER. If necessary, remove the patient from the ventilator and provide an alternative means of ventilation.

Warnings

<WARNING:

After assembling, cleaning, or reassembling the patient circuit, and on a daily basis, inspect the hoses and other components to ensure that there are no cracks or leaks and that all connections are secure.

<WARNING:

Use all cleaning solutions and products with caution. Read and follow the instructions associated with the cleaning solutions you use to clean your ventilator. Use only those solutions listed in Table 7-1.

<WARNING:

Never use a liquid cleaner inside the patient circuit, or on any component of a gas pathway. Clean the patient circuit only as specified by the manufacturer’s instructions.

<WARNING:

Do not attempt to open, repair or otherwise service the ventilator yourself. Doing so might endanger the patient, damage the ventilator, or void your warranty. Only personnel authorized and qualified by Covidien should repair, open or service the ventilator.

<WARNING:

If the ventilator is damaged, or its external housing is not correctly closed, or it behaves in a way that is not described in this manual (excessive noise, heat emission, unusual odor, alarms not triggered during the start-up procedure), the oxygen and power supplies should be disconnected and use of the device stopped immediately.

<WARNING:

The exhalation block is intended for single use by a single patient . It may periodically be cleaned, but it cannot be disinfected or sterilized. To maintain good measurement quality when used continuously, clean the exhalation block periodically (refer to section 7.3, Cleaning the Exhalation Block). The exhalation block should be changed every 4 months and cannot be reused with any other patient.

<WARNING:

Ensure that the exhalation block is completely dried after cleaning and prior to use.

<WARNING:

When an exhalation block is set up, each time it is removed, or after installing a new exhalation block on the machine, it is essential that the exhalation flow sensor be recalibrated before the exhalation block is used. Refer to section 8.3, Calibrating the Exhalation Flow Sensor.

<WARNING:

The patient circuit is intended for single use by a single patient and should be changed according to the manufacturer’s recommendations and according to the patient circuit lifetime. Refer to the instructions for use supplied by the manufacturer of the patient circuit (included with the ventilator) and Chapter 4, Installation and Assembly.

Safety Information

<WARNING:

A patient treated by mechanical ventilation is highly vulnerable to the risks of infection. Dirty or contaminated equipment is a potential source of infection. Clean the ventilator and its accessories regularly and systematically before and after each use and following any maintenance procedure to reduce the risks of infection. The use of a bacterial filter at the ventilator’s outlet (TO PATIENT) port—or both ports if a double-limb circuit is used—is recommended. Refer to Chapter 7, Cleaning.

<WARNING:

Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator. If necessary, replace the filter before the recommended replacement period is over (see Chapter 8, Routine Maintenance). This is particularly important when the ventilator is installed on a wheelchair, because environmental conditions may cause the filter to become dirty more rapidly.

<WARNING:

For environmental protection, the ventilator and its components, whatever their respective conditions of operation, cannot be disposed of with household waste and must be submitted for suitable selective collection and possible recycling. Observe all applicable regulations when disposing of the ventilator and any of its components.

<WARNING:

Before using the ventilator’s internal battery, ensure that the battery is fully charged and that the charge holds. Back up ventilators or those in storage should be connected to an AC power source to protect the integrity of the battery.

<WARNING:

The maximum recommended shelf life of the internal battery is 2 years. Do not use a battery that has been stored for 2 years prior to its first use. Periodic recharging is important to help maximize useful life of the battery. Do not store the internal battery for extended periods, without recharging, as this may reduce the maximum life.

<WARNING:

To connect the ventilator to an external power source, first ensure the ventilator’s I/O (power) switch is off (O). Then, connect the desired power cable to the ventilator. Finally, connect the power cable to the external power source.

<WARNING:

To disconnect the ventilator from an external power source, first power down the ventilator. Then, disconnect the power cable from the external power source and, finally, the ventilator.

<WARNING:

Connect the external DC power source by first connecting the power cable to the ventilator and then to the external DC source. Follow the reverse procedure to disconnect the device from the external DC power source.

Warnings

<WARNING:

Connect the external electrical power source by first connecting the power cable to the ventilator and then to the external power source. Follow the reverse procedure to disconnect the device from electrical power sources.

1.2.8Warnings Regarding Oxygen

<WARNING:

The ventilator must not be used with flammable anesthetic substances.

<WARNING:

Oxygen therapy for patients with respiratory failure is a common and effective medical prescription. However, be aware that inappropriate oxygen use may potentially lead to serious complications, including, but not limited to, patient injury.

<WARNING:

Strictly follow the instructions provided in section 4.8.2, Connecting the Oxygen Supply, which include the use of a flow regulator and special oxygen connector.

<WARNING:

To avoid injury to the patient and/or possible damage to the ventilator: before connecting the ventilator to the oxygen supply, ensure a flow meter (flow regulator) is connected to the ventilator to regulate the oxygen supply to the required specification.

<WARNING:

The Puritan Bennett™ 560 ventilator can be used with an optional oxygen analyzer with minimum and maximum concentration alarms. Always measure the delivered oxygen with a calibrated oxygen analyzer (FiO2 kit) that features a minimum and maximum concentration alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.

<WARNING:

The Puritan Bennett™ 560 ventilator is designed to deliver a percentage of oxygen equal or lower than 50%. Do not exceed this value as this may cause the ventilator to malfunction and put the patient at risk.

<WARNING:

Ensure that the oxygen supply pressure to the machine never exceeds 7 psi (50 kPa) or a flow of 15 lpm. For volume and sensitivity tolerances, refer to Table A-7.

<WARNING:

In the event of an oxygen leak, shut down the supply of oxygen at its source. In addition, remove and/ or keep any incandescent source away from the device, which may be enriched with oxygen. Circulate fresh air into the room to bring the oxygen level down to normal.

Safety Information

<WARNING:

The hose connecting the ventilator to the oxygen source must be designed exclusively for use with medical-grade oxygen. Under no circumstances should the oxygen hose be modified by the user. In addition, the hose must be installed without the use of lubricants.

<WARNING:

Ensure that the only gas supplied to the ventilator through the dedicated oxygen supply connector is medical-grade oxygen.

<WARNING:

The coupler must not remain connected to the oxygen inlet unless it is also connected to a leakproof, external oxygen gas source. When an oxygen supply is not being used with the ventilator, disconnect the oxygen source completely from the ventilator.

<WARNING:

To prevent any interference with the internal sensors of the ventilator, do not install a humidifier upstream of the ventilator.

<WARNING:

To ensure stability, when the Puritan Bennett™ 560 ventilator is mounted on a cart, the weight of the oxygen bottle should not exceed 14 kg (30 lbs).

<WARNING:

The oxygen supply hose ages even when it is not in use and should be replaced periodically. Follow the expiration date, if any.

<WARNING:

The oxygen supply must be regulated using a flow meter connected to the source gas outlet.

<WARNING:

The oxygen supply must be shut off when ventilation is interrupted. Before disconnecting the oxygen hose, allow the ventilator to continue for a few cycles without oxygen to flush the patient circuit of excess oxygen.

<WARNING:

Before connecting the oxygen supply, ensure that the stud on the oxygen inlet is protruding outwards.

<WARNING:

Inspect the oxygen coupler before use to ensure it has its black O-ring attached and in good condition. Do not use an oxygen coupler with a missing, damaged, or worn O-ring.

Symbols and Markings

1.2.9Warnings Regarding Electromagnetic Interference

<WARNING:

The Puritan Bennett™ 560 ventilator requires special precautions for electromagnetic compatibility and should be installed and started according to the recommendations found in Appendix A, Specifications. In particular, the use of nearby mobile and portable communications equipment using radio frequencies, such as mobile telephones or other systems exceeding the levels set in the IEC 60601-1-2 standard, may affect its operation. Refer to section A.10, Manufacturer’s Declaration.

<WARNING:

Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the ventilator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

<WARNING:

The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored to verify normal operation.

1.3 Symbols and Markings

Table 1-1. Ventilator Symbols

Symbols Descriptions

It is essential to read, understand, and follow these instructions before using the Puritan Bennett™ 560 ventilator (ISO 7000-0434A).

This symbol appears on the ventilator’s back panel, see item 5 in Table 1-2.

It is mandatory to read, understand, and follow these instructions before using the Puritan Bennett™ 560 Ventilator (ISO 7010-M002). This symbol appears on the ventilator’s air inlet label, see item 5 in Table 1-2.

Type BF applied part (IEC 60417-5333).

A regulatory standard classification for protection against electrical shock for the part of the device that contacts the patient.

This symbol appears on the ventilator’s back panel; see item 5 in Table 1-2.

Direct current, DC (IEC 60417-5031).

This symbol appears on the ventilator’s front panel and back panel; see Figure 1-1, and item 9 in Figure 1-3.

Alternating current, AC (IEC 60417-5032).

This symbol appears on the ventilator’s front panel and back panel; see item 8 in Figure 1-3, and item 10 in

Figure 2-3 (page 2-7).

Internal battery.

This symbol appears on the ventilator’s front panel; see item 10 in Figure 2-3 (page 2-7).

Safety Information

Table 1-1. Ventilator Symbols (Continued)

Symbols Descriptions

Insulation class II equipment (IEC 60417-5172).

A regulatory standard classification for protection against electric shock. Class II equipment relies on double insulation rather than protective earthing.

This symbol appears on the ventilator’s back panel; see item 5 in Table 1-2.

Index of Protection rating for the ventilator’s enclosure, defined in IEC 60529 (BSEN60529).

The first digit, 3, indicates protection against the intrusion of small foreign bodies (including fingers, tools, wires, etc. with a diameter greater than 2.5 mm) into the ventilator. The second digit, 2, indicates protection against water dripping or falling vertically when the enclosure is tilted at an angle up to 15° from its normal position, as well as an environment featuring water vapor condensation and/or light rain.

This rating appears on the ventilator’s back panel; see item 5 in Table 1-2.

CSA—Canadian Standards Association.

This symbol appears on the ventilator’s back panel; see item 5 in Table 1-2.

CE—Conformity European

Signifies compliance with the medical device directive 93/42/EEC as amended by 2007/47/EC.

This symbol appears on the ventilator’s back panel; see item 5 in Table 1-2.

This combined symbol appears on the ventilator’s UP/UNFREEZE key; see item 4 in Figure 2-3 (page 2-7).

This key is used to: move the LCD display’s cursor upwards, line-by-line; increase the value of displayed and selected parameter settings; restart (“unfreeze”) waveforms tracing.

This combined symbol appears on the ventilator’s DOWN/FREEZE key; see item 6 in Figure 2-3 (page 2-7).

This key is used to: move the LCD display’s cursor downwards, line-by-line; decrease the value of displayed and selected parameter settings; stop (“freeze”) waveforms tracing.

This symbol appears on the ventilator’s ENTER key; see item 5 in Figure 2-3 (page 2-7).

This key is used to confirm command actions.

This combined symbol appears on the ventilator’s ALARM CONTROL key; see item 2 in Figure 2-3 (page 2-7).

This key is used to: cancel the audible portion of alarms for 60 seconds at a time; cancel an alarm.

For more information, see Appendix E, Alarms Tests.

This symbol appears on the ventilator’s MENU key; see item 7 in Figure 2-3 (page 2-7).

This key is used to access the ventilator’s menus via the ventilator’s front panel LCD display.

This symbol (IEC 60417–5009) appears on the ventilator’s VENTILATION ON/OFF button; see item 8 in Figure 2-3 (page 2-7).

This button is used to start and stop ventilation.

TO PATIENT port.

This symbol appears on the front right of the ventilator, adjacent to the TO PATIENT port; see item 1 in Figure 1-1.

FROM PATIENT port (double-limb option).

This symbol appears on the front left of the ventilator, adjacent to the FROM PATIENT port; see item 4 in Figure 1-1.

Patient proximal pressure port.

This symbol appears on the front right of the ventilator, adjacent to the proximal pressure and TO PATIENT ports; see Figure 1-1, and item 3 in Figure 1-4.

Symbols and Markings

Table 1-1. Ventilator Symbols (Continued)

Symbols Descriptions

Exhalation valve pilot port.

This symbol appears on the front right of the ventilator, adjacent to the exhalation valve and TO PATIENT ports, indicating the connection of the tubing between the patient circuit exhalation valve; see Figure 1-1, and item 3 in Figure 1-4.

Oxygen inlet.

This marking appears on the back panel of the ventilator, adjacent to the oxygen inlet port; see item 2 in Figure 1-3.

Nurse call connection.

This symbol appears on the back panel of the ventilator, adjacent to the nurse call receptacle; see item 12 in

Figure 1-3.

Note: Nurse call is not intended for use in the home environment.

Switch in “Off” position (IEC 60417-5008).

This symbol appears on the I/O (power on/off) switch on the back panel of the ventilator to indicate the switch’s “Off” position. See item 2 in Figure 2-2 (page 2-6).

Switch in “On” position (IEC 60417-5007).

This symbol appears on the I/O (power on/off) switch on the back panel of the ventilator to indicate the switch’s “On” position. See item 2 in Figure 2-2 (page 2-6).

Software lock enabled.

This symbol appears at upper left in the ventilator’s LCD display when the keyboard Locking key is enabled.

Internal battery.

This symbol appears at top center in the ventilator’s LCD display to indicate that the ventilator is being powered by its internal battery. See item 1 in Figure 2-4 (page 2-8), and Chapter 6, Internal Battery, for more information.

Pressure rise times (inspiratory phase) parameter.

These symbols appear on the ventilation mode menu screens. In pressure ventilation modes, you can select one of four rise times with setting 1 representing the fastest rise time and setting 4 representing the slowest.

Flow shape (“flow distribution shape”, inspiratory phase) parameter.

These symbols appear on the ventilation mode menu screens; selectable for V A/C mode only. In volume ventilation mode you can select between Square (SQ), Descending (D) or Sinusoidal (S) flow patterns.

Selected line (filled square).

When making menu choices, this graphic indicates the line on which the cursor is currently positioned.

Non-selected line (empty square).

When making menu choices, this graphic indicates a line on which the cursor is currently not positioned.

Locked parameter line.

When making menu choices, this graphic indicates a line that cannot be selected (the Locking key is enabled).

Active parameter line.

When making menu choices, this graphic indicates that the current parameter is selected and can be changed. See Chapter 5, Operating Procedures.

Puritan Bennett 560 User manual

Safety Information

Table 1-1. Ventilator Symbols (Continued)

Symbols Descriptions

Inspiratory effort detected.

This symbol appears in the front panel display’s Status window when the patient triggers a breath.

Parameter adjustment bar.

This graphic shows the current setting for parameters such as display contrast and alarm volume in the Preferences menu.

WEEE (Waste Electrical and Electronic Equipment).

This symbol means that this product must not be disposed of with household waste. Observe local ordinances for proper disposal. See item 5 in Table 1-2.

Year of manufacture.

Manufacturer.

Authorized representative.

Audio paused (ALARM CONTROL key pressed once).

This symbol means the sounding of audible alarms is currently disabled. This period lasts 60 seconds. For more information, see section 3.5, Pausing the Audible Portion of Alarms.

Alarm paused (ALARM CONTROL key pressed twice).

This symbol means one or more alarms have been paused, or reset/canceled. The alarm is paused until the alarm condition is corrected and the condition reoccurs. For more information, see section 3.6, Pausing and Resetting Alarms.

Alarm off (Apnea off).

This symbol means that the Apnea alarm has been set to OFF in the Preferences menu. For more information, see section 3.6, Pausing and Resetting Alarms.

Exhalation valve detected.

This symbol means that an exhalation valve has been detected during ventilation.

No exhalation valve detected.

This symbol means that no exhalation valve has been detected during ventilation.

Single patient use only (ISO 7000-1051).

This symbol means that the labeled device is for use by a single patient only.

Freeze waveforms.

This symbol means the tracing of patient pressure and flow waveforms is currently paused or “frozen.”

Follow instructions for use (ISO 7000-1641).

This symbol directs the user to observe and adhere to the instructions contained in the product’s user manuals.

USB port.

This symbol indicates a communications port for interfacing with a USB connector. See item 11 in Figure 1-3.

Symbols and Markings

Table 1-1. Ventilator Symbols (Continued)

Symbols Descriptions

PC connection.

This symbol indicates a port that can be used by authorized Covidien product service personnel or Covidien service personnel for software maintenance. See item 10 in Figure 1-3.

Atmospheric pressure limitations. See section A.7 for specifications.

Humidity limitations. See section A.7 for specifications.

Temperature limitations. See section A.7 for specifications.

Fragile.

Keep dry.

Keep away from direct sunlight.

This side up.

Stacking limitation.

The number shown (represented by “n”) indicates the maximum number of additional identical packages that may be stacked on top of a package containing this device, when this device is correctly packaged.

For the Puritan Bennett™ 560 ventilator, n = 2.

Lithium battery.

This symbol Indicates that the contents of the package contain lithium batteries.

Safety Information

1.4 Labels (Identification and Instruction Information)

Various labels or specific markings are affixed to the ventilator that describe precautions to be taken for the correct use of the ventilator and contribute to the traceability of the product. See Table 1-2 and the figures on the following pages for illustrations of these labels and markings and their locations on the ventilator. Use the item numbers in Table 1-2 to locate the labels in Figures 1-1 through

1-4.

Table 1-2. Ventilator Labels and Markings

1. TO PATIENT port label

2. Oxygen inlet marking and label

3. Exhalation valve and patient

4. FROM PATIENT port, exhala-

(Figures 1-1 and 1-4)

(Figure 1-3)

pressure connection label

tion limb connection of patient

circuit—single use exhalation

(Figures 1-1 and 1-4)

block label

(Figures 1-1, 1-2, and 1-4)

1—Location of AC power cable

receptacle

5. Air inlet label

6. Exhaled gas outlet label

7. Identification label

8. AC power (mains) cable

(Figure 1-3)

(Figure 1-2)

(Figure 1-4)

receptacle marking

(Figure 1-3)

1—Location of DC power cable

receptacle

9. External DC cable receptacle

10. PC connection marking

11. USB port marking

12. Nurse call cable receptacle

marking

(Figure 1-3)

(Figure 1-3)

marking

(Figure 1-3)

(Figure 1-3)

13. FiO2 label

(Figures 1-1 and 1-4)

Labels (Identification and Instruction Information)

:Note:

The item number callouts in the following figures correspond to those listed in Table 1-2.

Figure 1-1. Locations of Labels—Top-Front View

Figure 1-2. Locations of Labels—Front-Left View

Safety Information

Figure 1-3. Location of Labels and Markings—Rear View

Figure 1-4. Location of Labels—Bottom View

2Ventilator Overview

2.1Indications for Use

The Puritan Bennett™ 560 Ventilator is indicated for the continuous or intermittent mechanical ventilatory support of patients weighing at least 11 lb (5 kg) who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a doctor. It is essential to read, understand, and follow these instructions before using the Puritan Bennett™ 560 Ventilator.

2.1.1 Target Patients

Specifically, the ventilator is applicable for adult and pediatric patients who require the following general types of invasive or non-invasive ventilatory support, as prescribed by an attending doctor:

Positive Pressure ventilation

Assist/Control, SIMV, or CPAP modes of ventilation

Breath types including Volume Control, Pressure Control, and Pressure Support

2.1.2Target Environments

The ventilator is suitable for use in institutional, home, and portable settings. It is not intended for use in Emergency Medical Service (EMS) such as an emergency transport.

The Puritan Bennett™ 560 Ventilator is suitable for use on commercial aircraft, per FAA requirements. See section A.11, Standards Compliance and IEC Classification. Patients traveling with the Puritan Bennett™ 560 Ventilator may be required by their airline to demonstrate evidence of compliance with the RTCA/DO-160F standard, as well as other requirements. Contact your airline prior to travel to determine airline specific requirements and documentation.

<WARNING:

Even though the Puritan Bennett™ 560 Ventilator meets current safety standards, the internal Lithium-ion battery of the device exceeds the 100Wh threshold and is therefore considered to be Dangerous Goods (DG) Class 9 – Miscellaneous, when transported in commerce. As such, the Puritan Bennett™ 560 Ventilator and/or the associated Lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International Air Transport

2-1

Ventilator Overview

Association), International Maritime Dangerous Goods code for sea and the European Agreement concerning the International Carriage of Dangerous Goods by Road (ADR) for Europe. Private individuals who transport the device are excluded from these regulations although for air transport some requirements apply. For air transport; the Puritan Bennett™ 560 Ventilator is permitted as checked-in or carry-on baggage. Two spare batteries per person may be taken on board as carry-on luggage only, with the prior approval of the airline. This classification and regulatory requirements may vary depending upon the country and mode of transport. Therefore it is recommended that users verify with the carrier / airline as to which measures to take before the voyage.

2.1.3Target Operators

<WARNING:

This ventilator must be used only under the responsibility and on the prescription of a doctor.

The ventilator may be operated by the following caregivers:

Respiratory therapists

Doctors

Nurses

Homecare providers

Patient and patient’s families

For more details on knowledge and skill requirements for operating the Puritan Bennett™ 560 ventilator, please consult your clinician.

It is the responsibility of the clinician or clinical educator to ensure that both the patient and the caregiver fully understand the topics necessary for operation of the ventilator.

2.2 Contraindications

This ventilator is not for use with anesthetic gases, and is not intended for use as an emergency transport ventilator.

Operational Use

2.3 Operational Use

The Puritan Bennett™ 560 ventilator uses a micro-turbine to provide ventilatory support to patients. Clinicians may use a variety of interfaces to connect patients to the ventilator for continuous or intermittent ventilatory support. Some examples include mouthpieces; nasal masks or full face masks; endotracheal tubes or tracheotomy tubes. User-selectable ventilation modes are:

Assisted Controlled Volume (V A/C)

Assisted Controlled Pressure (P A/C)

Volume Synchronized Intermittent Mandatory Ventilation (V SIMV)

Pressure Synchronized Intermittent Mandatory Ventilation (P SIMV)

Continuous Positive Airway Pressure (CPAP)

Pressure Support Ventilation with apnea ventilation (PSV/ST)

2.3.1Safety Net

Incorporated in the ventilator design is an alarm system that continuously monitors both patient and machine for signs of specific errors or faults that could lead to an unsafe condition. Should any of these errors or faults be detected, the alarm system announces the specific alarm condition both audibly and visually. The machine-related alarm conditions are factory set, whereas the patientrelated alarm conditions are defined by alarm-threshold values selected by an operator (a clinician or a caregiver). For more information, see Chapter 3, Alarms and Troubleshooting.

2.3.2 Settings

A software key, known as the Locking key, restricts access to ventilation parameter settings and ventilation mode changes in order to distinguish between clinician usage and patient usage.

2.3.3 Oxygen Enrichment

Oxygen may be supplied from an external, low pressure source, but the oxygen flow must be limited to 15 lpm (50 kPa, 500 mbar). The ventilator automatically compensates for the extra flow created by the external oxygen supply (see Chapter 4, Installation and Assembly.)

2.3.4 Breathing Circuit

The ventilator can be used with a singleor double-limb patient circuit. If exhaled volume monitoring is required (such as ventilator dependent patients), use the double-limb circuit for exhaled tidal volume monitoring. For more information, see section 4.4, Patient Circuit.

Ventilator Overview

<WARNING:

Users must always possess an additional breathing circuit and exhalation valve while using the Puritan Bennett™ 560 ventilator.

2.4Device Classification

The ventilator’s IEC/EN 60601-1classification is as follows:

Protection/insulation class (electric shock): Class II

Protection index of enclosure: IP32

Degree of protection against risk of electric shock: BF

Power: External (AC–mains, or DC–cigarette lighter) or internal (DC–battery)

Operation mode: Continuous operation

For additional information, see Appendix A, Specifications.

Front Panel

2.5 Front Panel

Figure 2-1. Front Panel

1

LCD display—Shows information about the ventilator, including

6 Exhalation valve port—Nipple for providing piloting

patient hours and software version, ventilation modes and settings,

pressure to the exhalation valve. Controls the open-closed

and monitored and calculated patient data and waveforms. The

position of the exhalation valve.

display also allows the user to view and, using the control panel,

adjust the ventilator’s operating and alarm configuration settings.

2

Control panel—Features the controls for setting up and operat-

7 Lateral and front openings—Vents that allow for air cir-

ing the ventilator, and LEDs to indicate the ventilator’s power

culation to cool the ventilator’s internal components. In

source, ventilation on/off status, and alarm priority level. Control

addition, these openings function as sound ports for

functions include turning on and off the ventilation, configuring

audible alarms.

ventilation modes, pausing audible alarms, canceling alarms, and

WARNING: Do not cover or obstruct these openings.

setting device and alarm parameters.

3

FiO2 sensor connection—Connection for FiO2 sensor, which mon-

8 From patient port—Exhaled volume measurements are

itors the amount of oxygen in the patient circuit.

taken from this port, through which a portion of the

exhaled gas is diverted to the exhalation flow sensor.

VTE is calculated from this flow measurement.1

4

Patient connection port—Provides an outlet for the gas to be

9 Exhaled gas outlet—Exhalation valve connects here.

delivered to the patient via the patient circuit.

5

Patient pressure monitoring port—Nipple for monitoring proxi-

mal patient pressure.

1.If exhaled tidal volume monitoring is required, use the double-limb circuit.

Ventilator Overview

2.6 Back Panel

Figure 2-2. Back Panel

1

Ergonomic carrying handle.

7

PC cable connector:

USB mini-B connector used for Puritan Bennett™ ventilator

test software.

WARNING: Do not connect the ventilator to any device other

than a PC with a dedicated compatible Puritan Bennett™ soft-

ware package.

2

I/O (power) switch with protective cover:

8

O2 inlet port:

Device powered on in position I; device switched off in posi-

Connects the ventilator to a low pressure oxygen source via an

tion 0.

adapter connected to the O2 inlet (see section 4.8, Oxygen).

3

AC power (mains) cable connector.

9

Nurse call output connector:

Used to connect the ventilator to the nurse call system.

Note: Nurse call is not intended for use in the home environ-

ment.

4

AC power (mains) cable holding system:

10

USB memory device connection:

Secures AC power cable to avoid accidental disconnection.

USB connection to be used with the Puritan Bennett™ Respira-

tory Insight software package. There are two USB type A ports.

WARNING: USB connections are not intended for connection

to any devices other than the specified USB flash storage (per

section 5.2.1, USB Memory Device Specifications).

5

Access cover for the internal battery.

11

Air inlet filter:

Filters air as it enters the ventilator.

6

DC power cable connector with key.

Control Panel

2.7 Control Panel

Figure 2-3. Control Panel

1 Alarm indicators (two LEDs):

6

DOWN/FREEZE key:

Red indicator:

• Moves the cursor down and decreases parameter values.

• Continuous: Very high priority (VHP) alarm activated.

• During ventilation, freezes the waveform shown in the Wave-

• Flashing: High priority (HP) alarm activated.

form menu.

Yellow indicator:

• Flashing: Medium priority (MP) alarm activated.

• Continuous: Low priority (LP) alarm activated.

2

ALARM CONTROL key:

7

MENU key:

• Press once to pause an audible alarm for 60 seconds.

Changes the menu shown. From the Ventilation menu screen,

• Press twice to halt visual and audible alarms. If alarm is reme-

press this key to show the Alarm menu screen.

died, the alarm is canceled (other than the high pressure alarm).

When a USB memory device is inserted into the ventilator,

press this key to show the USB memory device screen.

3

Display screen:

8

VENTILATION ON/OFF button:

Shows modes, ventilation settings, patient data and waveforms,

• ON: Press briefly and release to start ventilation.

configuration of the ventilator, and alarm management.

• OFF: Press and hold for 3 seconds, then press again to stop

ventilation.

4

UP/UNFREEZE key:

9

Ventilation status indicator:

• Moves the cursor up and increases parameter values.

• Blue indicator illuminated: Device is turned on and ventilation

• During ventilation, reactivates waveform tracing in the Wave-

is off (on standby).

form menu.

• Blue indicator off: Ventilation is on.

5

ENTER key:

10

Electrical power source indicators:

• Access to a setting value and validation of the modification of

• AC power indicator lit: AC power source connected.

this setting.

• DC power indicator lit: DC power source connected.

• Access to a sub-menu.

• Internal battery indicator lit continuously: Internal battery in

use (no external power source connected.)

• Internal battery indicator flashing: Battery charging.

Ventilator Overview

2.8 Ventilation Menu

Figure 2-4. Ventilation Menu Display (on standby at left; during ventilation at right)

1

General information line:

2

Ventilation settings:

3 Preferences menu access line:

Shows the current ventilation mode,

Shows the specific ventilation parameter

Highlight this line and press the

along with the following:

values for the currently selected ventilation

ENTER

key to show the Preferences

: Battery symbol if the ventilator is

mode.

menu.

powered by the internal battery.

: Audio paused symbol if an alarm is

currently inhibited.

: Alarm paused symbol if an alarm

has been canceled manually and the

cause of the alarm remains.

: Apnea alarm deactivation symbol.

: Exhalation valve symbol.

: No exhalation valve symbol.

ABS: Absolute symbol.

REL: Relative symbol.

4

Bargraph:

5

Status/monitored data window:

6 Alarm conditions window:

Shows pressure generation during ven-

• Ventilation stopped (standby): Shows the

• For active alarms, scrolls through active

tilation.

message, “PRESS

TO START VENTILA-

alarm messages in flashing reverse video.

TION.”

• For inactive alarms, shows the last alarm

• Ventilation on: Parameters are monitored

along with its trigger date and end-of-

event time.

and shown.

• The Inspiratory effort detected symbol

See Chapter 3, Alarms and Troubleshooting

for details.

appears adjacent to the monitored I:E ratio

when the patient actively triggers a breath.

Alarm Menu

2.9 Alarm Menu

Figure 2-5. Alarm Menu (on standby at left; during ventilation at right)

1

Title line:

2

Alarm settings:

3 Access line to Alarm Logs menu.

Shows ventilation mode and the follow-

Shows the specific alarm parameter

Highlight this line and press the ENTER

ing symbols:

values for the currently selected ventila-

key to show the Alarm Logs menu.

: Battery symbol if the ventilator is

tion mode, which are:

See section 3.4, Alarm Logs Menu.

• Minimum and maximum alarm thresh-

powered by the internal battery.

old settings

: Audio paused symbol if an alarm is

• Current monitored patient readings, or

currently inhibited.

hyphen (–) when ventilation is in standby

: Alarm paused symbol if an alarm has

been canceled manually and the cause of

the alarm remains.

: Apnea alarm deactivation symbol.

: Exhalation valve symbol.

: No exhalation valve symbol.

4

Status/monitored data window:

5

Alarm message window:

• Ventilation stopped (standby): Shows

• For active alarms, scrolls through active alarm messages in flashing reverse video.

the message, “PRESS

TO START VENTI-

• For inactive alarms, shows the last alarm along with its trigger date and end-of-event

LATION.”

time.

• Ventilation on: Parameters are moni-

See Chapter 3, Alarms and Troubleshooting for details.

tored and shown.

• The Inspiratory effort detected

symbol

appears adjacent to the mon-

itored I:E ratio when the patient actively

triggers a breath.

Ventilator Overview

2.10 Waveforms Menu

The display of waveforms (see Figure 2-6) is optional and can be selected using the Menu key.

The Waveform menu is only accessible when ventilation is active.

Figure 2-6. Waveforms Menu

1 Title line:

2 Graphic zone:

3 Numeric zone:

Shows ventilation mode and the follow-

Shows the patient’s pressure and flow

Shows monitored data.

ing symbols:

waveforms as a function of time.

: Battery symbol if the ventilator is

powered by the internal battery.

: Audio paused symbol if an alarm is

currently inhibited.

: Alarm paused symbol if an alarm has

been canceled manually and the cause of

the alarm remains.

: Apnea alarm deactivation symbol.

: Freeze waveforms symbol if the

tracing of patient waveforms has been

halted during ventilation.

: Exhalation valve symbol.

: No exhalation valve symbol.

Loading…

  • About
  • Blog
  • Projects
  • Help
  • Donate

    Donate icon
    An illustration of a heart shape

  • Contact
  • Jobs
  • Volunteer
  • People

Bookreader Item Preview

texts

Puritan Bennett 560 Service Manual

Puritan Bennett 560 Service Manual

Addeddate
2020-05-19 23:37:00
Classification
Clinical;Ventilator;Puritan Bennett Ventilator;Puritan Bennett 560
Identifier
manual_Puritan_Bennett_560_Service_Manual
Identifier-ark
ark:/13960/t5q90375k
Ocr
ABBYY FineReader 11.0 (Extended OCR)
Ppi
300
Scanner
Internet Archive Python library 1.9.0

comment

Reviews

There are no reviews yet. Be the first one to
write a review.

SIMILAR ITEMS (based on metadata)

  • Главная страница
  • Каталог
  • Медицинское оборудование
  • Анестезиология, ИВЛ и Реаниматология
  • Аппараты ИВЛ
  • Стационарные аппараты ИВЛ

Портативный аппарат ИВЛ Puritan Bennett 560 для инвазивной и неинвазивной вентиляционной поддержки дыхания.

Вес и возраст пациентов: от 5 кг, взрослые, дети.

Преимущества: универсальность, высокая чувствительность потокового триггера к снижению усилия вдоха, доступность для пациентов, безопасность, период автономной работы 11 часов, лёгкость и портативность (вес 4,5 кг), низкий уровень шума.

Функционально: выбор режимов и типа вентиляции лёгких, блокировка клавиатуры, совместимость с автономным оборудованием подачи кислорода низкого давления, выбор специфических конфигураций для детей и взрослых.

Использование ИВЛ Puritan Bennett 560: в лечебных учреждениях, на дому, при транспортировке.

Варьирование уровня тревог позволяет с высокой точностью контролировать параметры вентиляции и минимизировать риски.

Программное обеспечение: просмотр отчётов и диаграмм за сутки на экране, анализ и архивирование данных о пациенте за год, перенос данных с помощью USB на ПК врача, подробные диаграммы за 2 суток.

Оборудование отвечает российским стандартам, имеет РУ, а также Сертификат RTCA/DO-160F (позволяет использовать его на борту коммерческих рейсов).

В избранное

В сравнение

Доставка по всей России

Сроки уточняйте у оператора

  • Описание

Портативный аппарат искусственной вентиляции легких Puritan Bennett 560 предназначен для механической вентиляционной поддержки дыхания взрослых и детей при оказании экстренной помощи на месте, во время перевозки, в условиях ЛПУ, а также для продолжительного жизнеобеспечения в медучреждениях и на дому.

К его основным достоинствам относятся:

  • Универсальность (может применяться при большинстве патологических состояний, сопровождаемых дыхательной недостаточностью, и вызванных нервно-мышечными поражениями);
  • Простота и скорость запуска, доступность для использования пациентами без помощи медперсонала;
  • Чувствительность, безопасность, большой объём предоставляемых данных, которые можно перенести на ПК врача с помощью USB.

ИВЛ Puritan Bennett 560 – разработка известного американского производителя медоборудования Medtronic и отвечает всем современным стандартам и нормам, в том числе он имеет Сертификат RTCA/DO-160F, благодаря чему может применяться на борту коммерческих рейсов.

Он применим и для взрослых, и для детей (весом от 5 кг), при этом допускается домашнее использование, так как оборудование максимально просто и безопасно в эксплуатации.

Преимущества аппарата ИВЛ Puritan Bennett 560

  • До 11 часов автономной работы на внутренней литий-ионной батарее – можно использовать вдали от источника тока, при транспортировке (в оптимальных условиях, без отягощающих обстоятельств).
  • Контроль заряда батареи – специальный индикатор показывает оставшееся время в часах и минутах на основе ситуационного режима.
  • Универсальность – применение при большинстве патологических состояний с хронической недостаточностью дыхания и дыхательных патологиях, вызванных нервно-мышечными заболеваниями (ХОБЛ, боковой амиотрофический склероз, спинальная мышечная атрофия и проч.).
  • Варьирование уровня тревог повышает точность контроля параметров вентиляции и сводит к минимуму риски клинических нарушений.
  • Соответствие стандартам РФ: Puritan Bennett 560 имеет Регистрационное удостоверение и отвечает всем требованиям к оборудованию этого типа в России.
  • Портативность, небольшой вес (4,5 кг) и наличие в комплекте сумки для переноски позволяют безопасно переносить и перевозить оборудование.
  • Высокий уровень самостоятельности больных обеспечивает безопасность в отсутствие рядом медработников.
  • Уровень шума – менее 30 дБА на расстоянии 1 метра.

Технические характеристики Puritan Bennett 560

Общие

Вес аппарата – 4,5 кг.

Вес пациентов – от 5 кг.

Диапазон дыхательного объёма – 50-2000 мл.

Уровень шума – менее 30 дБА на расстоянии 1 метра.

Настройки

Типы вентиляции – инвазивный и неинвазивный.

Режимы – P A\C, V A\C, P SIMV, V SIMV, PSV, CPAP.

Выбор специфических конфигураций для детей и взрослых.

Безопасность

Блокировка клавиатуры.

Регулируемость потокового триггера на снижение усилия вдоха.

Автоматическое определение клапана при выборе режима в соответствии с конфигурацией контура.

Совместимость

С автономным оборудованием подачи кислорода низкого давления извне.

Анализ и управление информацией

Создание и хранение диаграмм состояния пациента за 2 суток.

Просмотр отчётов и диаграмм за 1 сутки на экране.

Хранение информации о пациенте до 1 года.

Перенос информации и диаграмм на ПК с помощью USB.

Инструкция производителя по эксплуатации Puritan Bennett 560

Разработчик аппарата рекомендует не пренебрегать дополнительной техникой безопасности при эксплуатации:

  • Вести тщательное наблюдение в соответствии с тяжестью состояния больного;
  • Иметь в ближайшем доступе другие средства вентиляции лёгких для работы с пациентами, не способными дышать самостоятельно;
  • Иметь в наличии альтернативные источники питания (переменного тока, запасные аккумуляторы, вспомогательные автомобильные переходники постоянного тока).

Приобрести Puritan Bennett 560 на оптимальных условиях и по выгодной цене вы можете в Тиара-Медикал: специалисты проконсультируют вас, помогут оформить заказ и проследят за его оперативной доставкой. Кроме того, компания способна обеспечить техническое гарантийное и постгарантийное обслуживание аппаратов ИВЛ и их ремонт квалифицированными мастерами.

  • Снят с производства: Нет
  • Товар готов к отгрузке: Нет

Мы гарантируем

Гарантия производителя и продавца

Профессиональные консультации и поддержка на каждом этапе

Собственный сервисный центр

Демократичная цена и лизинг

Быстрая и бережная доставка

Частые вопросы


Стоимость

Вопрос: Почему на многие товары не указана цена?
Ответ:  Итоговая стоимость оборудования зависит от множества факторов:

1) Конфигурация. Многие модели медицинского оборудования являются модульными системами. По желанию клиента некоторые модули могут быть добавлены или исключены из поставки. Яркий пример – ультразвуковые сканеры, каждый из которых может комплектоваться различными наборами датчиков (на выбор из нескольких десятков) и дополнительными модулями (например, для расчетов и 4d-исследований). Таким образом, один и тот же УЗ-сканер может иметь несколько десятков конфигураций, значительно различающихся по цене.

2) Стоимость доставки. Мы предлагаем несколько вариантов доставки, из которых наши клиенты могут выбрать наиболее приемлемый по скорости и цене. Подробнее…

3) Установка и наладка. Многие виды оборудования требуют обязательной установки и наладки с помощью сертифицированного специалиста, выдающего акт ввода в эксплуатацию, что так же сказывается на стоимости.

4) Курс валюты, сроки поставки и прочие менее значимые факторы.

Совет: Если вы видите в каталоге какой-либо компании точную цену на медицинское оборудование – обязательно уточняйте, что входит в эту сумму!

Скидки! У нас действует гибкая система скидок, постоянно проводятся специальные акции и действуют другие привлекательные предложения. Следите за новостями!


Доставка

Территория доставки?

ТИАРА-МЕДИКАЛ осуществляет доставку медицинского оборудования в пределах Таможенного Союза (ЕврАзЭС) транспортными компаниями. За 10 лет работы мы установили тесные партнерские отношения с различными транспортными компаниями и предлагаем нашим покупателям наиболее выгодные варианты доставки.

В каких случаях бесплатная доставка?

Доставка по Санкт-Петербургу – БЕСПЛАТНО.
Доставка до транспортных компаний – БЕСПЛАТНО.


Лизинг

Компания ТИАРА-МЕДИКАЛ имеет многолетний опыт продажи медицинского оборудования в лизинг. Мы сотрудничаем с лизинговыми компаниями, выбранными покупателем, или можем порекомендовать наших проверенных партнеров.

Какое оборудование можно купить в лизинг?

В лизинг предоставляется оборудование для УЗИ, томографии, рентгенологии, эндоскопии, офтальмологии, косметологии. А также любое медицинское оборудование стоимостью от 1 000 000 рублей. Обратитесь за расчетом выгодного приобретения в лизинг к нашим специалистам по телефону: 8 (800) 500-26-76

Как быстро принимаем решение?

Срок рассмотрения от 1 дня.

С какими лизинговыми компаниями мы сотрудничаем?

В основном с «Элемент лизинг» и «Балтийский лизинг», также готовы работать с другими компаниями, которые выгодны и удобны для Вас.


Сервис

Мы создали лучшую систему сервисной поддержки медицинского оборудования, на протяжении всего срока службы. В нашей команде работают высококвалифицированные инженеры, систематически совершенствующие свои навыки на заводах производителей мед. оборудования. Мы оказываем исчерпывающий спектр услуг по поддержке и ремонту оборудования.

При поставке мы предлагаем

Установку, настройку, ввод в эксплуатацию (по всей территории РФ).

Обслуживание после поставки

Наш собственный лицензированный сервисный центр производит:
— Гарантийное и пост-гарантийное комплексное обслуживание медицинской техники.
— Гарантийный и пост-гарантийный ремонт.
— Выездной инструктаж пользователей.
— Поддержку документацией и учебными материалами.
— Консультации на любом этапе использования.

Отдел запчастей медицинского оборудования

Подбор и продажа оригинальных запчастей для медицинской техники.


Гарантии

ТИАРА-МЕДИКАЛ осуществляет продажу медицинского оборудования, инструментов и материалов в соответствии с законодательством РФ. Наше оборудование имеет всю необходимую разрешительную документацию, гарантию производителя и продавца.

Гарантийный срок на медицинское оборудование

Срок базовой гарантии на мед. оборудование составляет 12 месяцев со дня покупки и может быть увеличен в зависимости от индивидуальных гарантийных условий производителя!

Как заказать гарантийное обслуживание

Гарантийное сервисное обслуживание осуществляется по запросу в сервисный центр ТИАРА-МЕДИКАЛ. Звоните по тел.: 8 (800) 500-26-76 или оставьте заявку на странице сервисного центра

Кто проводит обслуживание медицинского оборудования

Мы имеем собственный лицензированный сервисный центр для обслуживания и устранения неисправностей и команду сертифицированных специалистов выездного обслуживания техники. Работы проводятся согласно стандартам производителя. Доставляем оборудование в сервисный центр — бесплатно!

Наши специалисты готовы на них ответить, рассказать об оборудовании и помочь с выбором.

Понравилась статья? Поделить с друзьями:
  • Тифи таблетки из тайланда инструкция по применению
  • Левофлоксацин 500 мг инструкция по применению цена отзывы при пневмонии
  • Мини камера md81s инструкция на русском
  • Acer aspire v3 772g руководство
  • Картинка демократический стиль руководства