Аппарат для гемодиализа dialog руководство по эксплуатации

B. Braun Dialog+ Service Manual

  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

+

Dialog

SW 9.xx

S e r v i c e M a n u a l

E n g l i s h

Edition 1 -2010

M.KAY Dialog+ SW9xx SM EN 1-2010

loading

Related Manuals for B. Braun Dialog+

Summary of Contents for B. Braun Dialog+

  • Page 1
    Dialog SW 9.xx S e r v i c e M a n u a l E n g l i s h Edition 1 -2010 M.KAY Dialog+ SW9xx SM EN 1-2010…
  • Page 2
    1 / 2 0 1 0 SW 9.xx Contact your Local B. Braun Representative for Service Support 34209 Melsungen B. Braun Avitum AG Germany Tel. No.: +49 5661 713500 (Günter Nissen) E-Mail: guenter.nissen@bbraun.com B. Braun Avitum AG BA-TE-DE08C M.KAY -…
  • Page 3
    S e rvi ce Manu al s w i th a re gi s tr a ti on num be r a re inc lud ed in th e u p da te s erv ic e! B. Braun Avitum AG BA-TE-DE08C M.KAY -…
  • Page 4
    1 / 2 0 1 0 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
  • Page 5
    Flow, Wiring and Tubing Diagrams Spare Parts List Appendix ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information Assembly Instructions Field Service Information Instruction Leaflets Edition/Updates Service Manual B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd…
  • Page 6
    Calibration The symbol appears for necessary calibration measures. Calibration Equipment/Tools symbol gives remarks necessary calibration equipment/tools, i.e. during a calibration, disassembly or assembly. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd…
  • Page 7
    Perform regular technical safety inspections as described in chapter 5 of this Technical Safety Inspection service manual to ensure the safety of the machine. An instructions for use can be ordered at your local B. Braun representative or Instructions for Use dealer.
  • Page 8
    1 — 1 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
  • Page 9
    (fasten with a single ear clamp 19.5 and a tubing clamp 12- 20 mm) • Tubing length: approx. 3 m • Drain height: max. 80 cm 8. Emergency Power Supply/Battery Option Fig.: Dialog+ Rear View B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 10
    1 / 2 0 1 0 1 — 3 SW 9.xx Table of Contents Page Commissioning Check List 1.1.1 Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 11
    …………………………….Operating Hours: ……..h Inventory No.: …………..SW Version: ……….. Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List Note: Text in { } brackets is information for the execution of the check list! Visual Inspection {Machine: clean/complete; no damages/moisture influences or loose assemblies; no moveable parts touching tubings or wires; casters are moveable;…
  • Page 12
    Next Inspection Date: …………………………..Name Service Technician: Name of Company: The commissioning was performed and the machine was hand over to the responsible ………………. organisation (user)…………………………Date/Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 13
    The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 14: Table Of Contents

    Switch Mode Power Supply Microcontroller SMPS-MC 2-19 2.15.1 Legend Switch Mode Power Supply Microcontroller 2-19 2.16 TFT Monitor 2-20 2.16.1 Legend TFT Monitor 2-20 2.17 Front Door 2-21 2.17.1 Legend Front Door 2-21 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 15
    System Integration SMPS-MC 2-43 2.22.3 Component Layout SMPS-MC 2-44 2.22.4 Wiring Diagram SMPS-MC with Battery Option 2-45 2.22.5 Description SMPS-MC 2-46 2.22.6 Fuses 2-48 2.22.7 Signals 2-49 2.22.8 Internal Signals 2-49 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 16: General Information

    All sensor data are sent separately, via two serial bus systems, to the supervisor and controller via the analog board to the digital board. The actuators, motors and valves are driven via the power board valves PBV and power board motors PBM. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 17
    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 3 SW 9.xx For Software ≥ ≥ ≥ ≥ 9.xx For Software 9.xx For Software For Software 9.xx 9.xx B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 18: Overview Sub-Racks

    1 / 2 0 1 0 2 — 5 SW 9.xx Overview Sub-Racks Fig. : Overview Sub-Racks Rear View Dialog+ 2.1.1 Legend Overview Sub-Racks Top Level Sub-Rack DF Sub-Rack UF Sub-Rack Water Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 19: Top Level Sub-Rack

    USB Port: Software Installation LLC/LLS; Service-Switch S1; Jumper JP1/JP2 7 7 7 7 HDF Online Board HOB 8 8 8 8 Power Board Motor PBM 9 9 9 9 Analog Board AB (LLA) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 20: Basic Board Bb

    SMPS-MS (NT) SMPS-MS (NT) SMPS-MS (NT) Power Supply Bedside Link BSL-PWR BSL-PWR BSL-PWR BSL-PWR Power Board Motors PBM P58/P60 P58/P60 Power Supply Hall Sensors P58/P60 P58/P60 Disinfection Valve VD B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 21: Power Board Valves Pbv

    Dialyser Outlet Valve VDA Bypass Valve VBP Option BIC Cartridge Valves VBICP, VBKS, VBKO, VVB Disinfection Valve VD Circulation Valve VZ Venous Tubing Clamp SAKV Arterial Tubing Clamp SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 22: Power Board Motors Pbm

    BIC Piston Pump BICP BICP BICP BICP Concentrate Piston Pump KP KP KP KP UF Piston Pump UFP Level Regulation Pump (Diaphragm Pump) LRP (PPR) LRP (PPR) LRP (PPR) LRP (PPR) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 23: Digital Board Db

    NSVB, BICSS, KSS, MSBK1/2, RDV, SAD, BPADS, BPVDS, Service Switch: BL (Controller/Supervisor Sensor Analog Board) Position 0: Therapy Mode Position 2: TSM Service Program Mode Position 3: Software Installation/Update Mode B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 24: Analog Board Ab

    Fig. : Analog Board AB (LLA) 2.7.1 Legend Analog Board P1/P2 Controller Sensors: PBS, TSHE, TSE, TSBIC, TSD, TSDE, BICLF, ENDLF, PE, LAFS Supervisor Sensors: TSD-S, ENDLF-S Controller/Supervisor Sensors: BL, PBE, PA, PV, PDA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 25: Hdf Online Board Hob

    SW 9.xx HDF Online Board HOB Fig. : HDF Online Board 2.8.1 Legend HDF Online Board P1: Sensors: PSABFS, PSAUS, PSPOSS, FEHDFS, FEDFFS P4: Valves VBE, VDFF, VSAA, VSAE, VSB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 26: Uf Sub-Rack

    Non-Return Valve Outlet Flow Pump RVFPA RVFPA RVFPA RVFPA Inlet Balance Chamber Valve VEBK2 VEBK2 VEBK2 VEBK2 Outlet Balance Chamber Valve VABK1 VABK1 Air Separator Valve VLA VABK1 VABK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 27: Uf Sub-Rack Hdf Online

    Non-Return Valve Outlet Flow Pump RVFPA RVFPA RVFPA RVFPA Substitute Connection Inlet Valve VSAE VSAE VSAE VSAE Outlet Balance Chamber Valve VABK1/2 VABK1/2 VABK1/2 VABK1/2 Filter Vent Valve VBE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 28: Df Sub-Rack

    Dialysate Temperature Sensor TSD Concentrate Pump KP KP KP KP Dialysate Supervisor Temperature Sensor TSD-S TSD-S TSD-S TSD-S Bicarbonate Conductivity Sensor BICLF BICLF UF Pump UFP BICLF BICLF Bicarbonate Non-Return Valve RVB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 29: Water Sub-Rack

    Top BIC Cartridge Valve VBKS VBKS VBKS VBKS Heater Temperature Sensor TSH BIC Pump Valve VBICP VBICP VBICP VBICP Heater Inlet Temperature Sensor TSHE TSHE TSHE TSHE Upline Tank Valve VVB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 30: Rinsing Bridge

    TSDE 2 2 2 2 Pressure Sensor Dialysate Outlet PDA 5 5 5 5 Blood Leak Detector BL BL BL BL 3 3 3 3 Dialyser Inlet Valve VDE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 31: Rear Door

    Rear Door SMPS-MC Fig. : Rear Door 2.14.1 Legend Rear Door 1. Switch Mode Power Supply Microcontroller SMPS-MC SMPS-MC 3. Mains Switch SMPS-MC SMPS-MC 2. Fan 4. Mains Cord B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 32: Switch Mode Power Supply Microcontroller Smps-Mc

    TN/Item-No (Version Number SMPS-MC): TN/Item-No (Version Number SMPS-MC): TN/Item-No (Version Number SMPS-MC): e.g. XXXXXX.5 Ver/SW (SW Version): Ver/SW (SW Version): Ver/SW (SW Version): Ver/SW (SW Version): e.g. Rev. 1.15 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 33: Tft Monitor

    Fig. : TFT Monitor 2.16.1 Legend TFT Monitor Backlight Inverter Board BIB Optical Status Display Board OSDB Front Panel Board FPB Touch Controller Board TCB TFT Monitor TFT Touch Screen B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 34
    Substitution Port 1 1 1 1 PBS/SN PBS/SN Arterial Pressure Sensor PA Connection for concentrate Suction Rods 2 2 2 2 PA PA Venous Pressure Sensor PV PV PV PV B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 35
    Venous Pressure Sensor Valve VPV Inlet Hydrophobic Filter HFE Arterial Pressure Sensor Valve VPA Pump Hydrophobic Filter (Ceramic Filter) HFP Blood Side Test Pressure Sensor Valve VBT Level Regulation Pump LRP B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 36: Front Door

    • Monitoring of the blood side pressures (pressure sensors PA, PV, PBE and PBS) • Test of the blood side pressure sensors in preparation (3/2 way solenoid valve VBT) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 37: Flow Diagrams

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 24 SW 9.xx 2.19 Flow Diagrams 2.19.1 Dialog+ B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 38: Dialog+ With Bic Option And Df Filter Option

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 25 SW 9.xx 2.19.2 Dialog+ with BIC Option and DF Filter Option B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 39: Dialog+ Hdf Online

    Dialog 2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 26 SW 9.xx 2.19.3 Dialog+ HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 40: Legend Flow Diagram

    Membrane Position Sensor Balance Chamber 2 Level Sensor Upline Tank NSVB Arterial Pressure Sensor Pressure Sensor Blood Inlet Blood Pressure Control Sensor Pressure Sensor Dialysate Outlet Degassing Pressure Sensor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 41
    VEBK1 VEBK2 Inlet Valve Balance Chamber 2 Air Separator Valve Upline Tank Valve Upline Tank Inlet Valve VVBE Circulation Valve Water Block Heat Exchanger Dialyser Coupling (to Dialysate) Z.D. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 42: Description Flow Diagram

    The cold inlet water is warmed up via the heat exchanger WT. Thereby the heat consumption to heat up the water is reduced. VVB* VBKS RVBO* Fig. : Water Inlet with Water Block B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 43: Degassing Circuit With Temperature System

    SMPS-MC. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* TSHE VBKO* VVB* VBKS RVBO* Fig. : Degassing Circuit with Temperature System B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 44: Dialysate Processing

    FPE. If the set pressure is reached RVFPE is opened and the fluid can circulate. RVFPE TSD-S TSDE TSBIC BICLF ENDLF ENDLF-S VDEBK1 VEBK1 BICP MSBK1 VDEBK2 VEBK2 MSBK2 Fig. : Dialysate Processing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 45: Central Bicarbonate And Concentrate Supply (Option)

    FPE (VBICP and VBKS are opened). RVFPE TSD-S DBK* TSBIC BICLF ENDLF ENDLF-S BICP VBICP* VBKO* FBIC VBKS VVB* RVBO* BKUS RVBU Fig. : BIC Cartridge Holder (Option) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 46: Balance Chamber System

    2 is drained (see description phase 2). VDEBK1 VABK1 VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 LAFS VEBK2 VDABK2 FBK2 MSBK2 open RVFPA closed Fig. : Phase 1 Balance Chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 47
    VDEBK1 VABK1 VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 LAFS VEBK2 VDABK2 FBK2 MSBK2 open RVFPA closed Fig. : Phase 2 Balance Chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 48: Ultrafiltration And Rinsing Bridge

    VLA is opened and air is removed with positive pressure from LA. VLA is closed once air is not detected anymore by LAFS. The balance chamber is switched to normal operation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 49: Chemical Thermal Disinfection Program

    ENDLF-S VB WT FBK1 Z.D. VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 V.D. BICP LAFS VEBK2 VDABK2 FBK2 MSBK2 RVDA TSHE SBS1 RVFPA SKBS SKKS RVUFP SBS2 Fig. : Phase 2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 50
    C and conductivity is below 0.1 mS/cm. Then the machine gives a message that asks the user to confirm that the machine is free of disinfectant and the disinfection is completed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 51: Block Diagram

    VSAE VBKS* Online VBKO* Board PSAUS PSABFS FEDFFS FEHDFS SAKV SAKA Mains Switch Mains Voltage SMPS-MC * Battery Switch Mode Power Supply * Options Microcontroller Fig. : Block Diagram B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 52: Legend Block Diagram

    TSDE Dialyser Inlet Temperature Sensor Bicarbonate Conductivity BICLF END Conductivity ENDLF Degassing Pressure Sensor Air Separator Level Sensors LAFS Supervisor Dialysate Temperature Sensor Supervisor TSD-S END Conductivity Supervisor ENDLF-S B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 53
    Arterial Pressure Sensor Valve Venous Tubing Clamp SAKV Arterial Tubing Clamp SAKA Heparin Pump Compact Heparin Pump Compact Light Barrier Speed Monitoring Interlocking Sensor KRALO Syringe Plunger Sensor KOLB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 54
    VDFF VSAE Substitution Connection Inlet Valve Substitution Bypass Valve Port Substitution Outlet Sensor PSAUS Port Substitution Drain Sensor PSABFS DF Filter Detection Sensor FEDFFS FEHDFS HD Filter Detection Sensor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 55: Switch Mode Power Supply Microcontroller Smps-Mc

    1 / 2 0 1 0 2 — 42 SW 9.xx 2.22 Switch Mode Power Supply Microcontroller SMPS-MC 2.22.1 Block Diagram SMPS-MC Fig. : Block Diagram Switch Mode Power Supply Microcontroller SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 56: System Integration Smps-Mc

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 43 SW 9.xx 2.22.2 System Integration SMPS-MC Fig. : System Integration B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 57: Component Layout Smps-Mc

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 44 SW 9.xx 2.22.3 Component Layout SMPS-MC Fig. : Layout SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 58: Wiring Diagram Smps-Mc With Battery Option

    2 x 12 V, 7 Ah 10 AT (5×20), breaking capacity 1500 A, 150 Vdc Fuse for Battery Jumper X104 Charge Current for Batteries Default: 50 % (right position) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 59: Description Smps-Mc

    ±12 VD: Digital +24 VGB: Switched blood side +24 VGD: Switched dialysate side +24 VL: Power (uncontrolled) GNDAN: GND analog GNDD: GND digital GNDH: GND aux. voltage GNDL: GND power B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 60
    A DC/DC converter generates the –12 V voltage from the 24 V. The soldered fuse F500 disconnects the 24 V in case of a short circuit in the DC/DC converter. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 61: Fuses

    +24 VGD 3.15 A (TR5) P4/15 P12/14 Note: Soldered fuses (e.g. F300 or F500) must not be changed if they are defective, i.e. the complete SMPS-MC must be exchanged. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 62: Signals

    = 10 s after the switch-on of the switch mode power supply. The watchdog has no safety function and is therefore not tested before the therapy starts. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 63
    3-25 3.3.8.11 Pressure Reduction VABK with UFP 3-26 3.3.8.12 VEBK Test 3-27 3.3.8.13 Pressure Reduction 3-28 3.3.9 Integrity Test HDF Online 3-29 3.3.10 UF Pump 3-31 3.3.11 Conductivity 3-32 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 64
    3-55 3.5.9 Test Run 3-55 3-56 Field Service Utilities FSU 3.6.1 FSU Directories and Subdirectories 3-56 3.6.2 Start Field Service Utility FSU 3-57 3.6.3 Set-Up FSU Main Menu 3-58 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 65
    Measures after Repair 3.8.1 Repair Matrix 3-103 3.8.2 Tests and Measures to be Executed 3-104 3.8.2.1 Install Software 3-104 3.8.2.2 Check System Configuration, Production Report and Set if 3-104 Necessary B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 66
    Test Run UF Comparison Measurement 3-105 3.8.2.6 Test Run 3-105 3.8.2.7 Electrical Safety Check 3-106 3.8.2.8 Perform Disinfection after Repair 3-106 3.8.2.9 Document the Executed Activities 3-106 3.8.3 Execution Protocol 3-107 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 67: Self Test Sw

    If an error is detected during the MPU test the following alarms are displayed: In Therapy – Therapy Selection: • Hardware error RAM/ROM (SUP) In TSM Service Program: • Error in MPU test B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 68: Status Self Test In Service Overview

    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 69: Set-Up Low Level Controller Llc Test Codes

    Format of LLS Test Codes: (Value ≥ 10000) LLS test code to LLC S S S Performed test step Example: 11104 Blood leak detector test: measure level for red and green B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 70: Example Llc/Lls Test Codes For Blood Leak Detector

    Pressure sensors blood side: Equality test for blood side pressure sensors Pressure sensors blood side: Blood side pressure test Disinfection Valve VD Pressure test substitution line S-Online HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 71: Service Mode (Dir_On)

    Set set point for heater to self test code #112 2720x Wait for feedback of test result via BIOS (timeout 15 s) 2730x Set set point of heater to 0 2799x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 72: Eeprom

    Activate buzzer after approx. 1 s for 250 ms 28003 2820x Wait for feedback of test result via BIOS (timeout 15 s) 2830x Set set point of heater to 0 2899x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 73: Alarm System

    Acknowledge with switch-off VGD 16022 – 16027 1640x Wait for D24OFF (timeout 60 s since start of code 200) VGD ON 16028 1650x Wait for D24ON (timeout 10 s) 1699 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 74: Monitoring Of Analog Voltage ±12

    Acknowledge with switch-off of VGD 11005 – 1120x Wait for D24OFF (timeout 30 s) 11011 VGD ON 11012 1130x Wait for D24ON (timeout 30 s) 1199x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 75: Dfs Pressure Test (Table)

    ∆PDA < 25 mmHg, 12077 VBP ———————————- opened VDA, VDE —————————— closed 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 76
    FPA = 500 ml/min Valve Status: VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDE —————————— opened VDA ———————————— closed PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 77
    Pressure reduction FPE = 500 ml/min FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2 ——————- opened VABK1/2, VDABK1/2——————- opened VBP, VDA, VDE ———————— opened Timeout 60 s 1299 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 78: Dfs Pressure Test

    • UFP is started during the pressure build-up phase. • The acknowledgement of the single test steps in the DFS pressure test is performed via a bit in the internal communication (between LLS and LLC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 79: Membrane Movement/Pressure Build-Up Vdabk

    FPA = 20 ml/min Valve Status: VEBK1/2, VDEBK1/2, VABK1/2 ————opened VDABK1/2—————————— closed VDE, VDA, VBP ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 80: Pressure Reduction Vdabk With Ufp

    FPA = 100 ml/min Valve Status: VEBK1/2, VABK1/2 ————————opened VDABK1/2, VDEBK1/2 ——————- closed VDE, VBP —————————— opened VDA ———————————— closed Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 81: Vbp Test

    12036 VDE, VBP —————————— closed 12037 VDA ———————————- opened Final Condition: 12038 Test timeout 30 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 82: Vde/Vda Test

    12076 VBP ———————————- opened 12077 VDA, VDE —————————— closed Final Condition: 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 83: Pressure Build-Up Vdabk

    FPA = 100 ml/min Valve Status: VEBK1/2, VDEBK1/2, VABK1/2 ——— opened VDABK1/2 —————————— closed VBP, VDA, VDE ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 84: Pressure Reduction Membrane Test With Ufp

    FPA = 20 ml/min Valve Status: VEBK1/2, VDABK1/2 ——————- opened VDEBK1/2, VABK1/2 ——————— closed VBP, VDE —————————— opened VDA ———————————— closed Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 85: Membrane Test

    ∆PDA < 25 mmHg, 12131 VBP, VDE, VDA ———————— opened 12132 Final Condition: 12133 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 86: Pressure Build-Up Vabk

    FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2, VDABK1/2 ——— opened VABK1/2 —————————— closed VBP, VDA, VDE ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 87: Membrane Movement To Inlet Side

    VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDA, VDE ———————— opened Final Condition: Test timeout 60 s or no increase of the values of the membrane position sensors B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 88: Pressure Reduction Vabk With Ufp

    Valve Status: VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDE —————————— opened VDA ———————————— closed VZ ———————————— opened Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 89: Vebk Test

    ∆PDA < 25 mmHg, 12181 VBP, VDE, VDA ———————— opened 12182 Final Condition: 12183 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 90: Pressure Reduction

    1280x 12000 VEBK test FPE = 500 ml/min FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2 ——————- opened VABK1/2, VDABK1/2——————- opened VBP, VDA, VDE ———————— opened Timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 91: Integrity Test Hdf Online

    PDA < -100 mmHg, 15033 Test timeout 1 min or wait for switch-off VGD Test delivery rate OSP at 200 ±15 ml/min 15034 – from LLS Acknowledge with switch-off of VGD 15039 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 92
    ∆PDA < 50 mmHg, Test timeout 1 min or wait for switch-off VGD from LLS PDA < -200 mmHg, 15112 – Acknowledge with switch-off of VGD 15118 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 93: Uf Pump

    3 revolutions have been detected Acknowledge with switch-off of VGD 20006 – 20011 2020x Wait for D24OFF (timeout 10 s) VGD ON 20012 2030x Wait for D24ON (timeout 10 s) 2099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 94: Conductivity

    The heater is switched off at TSD-S > 41 °C Acknowledge with switch-off of VGD 22006 – 22011 2220x Wait for D24OFF (timeout 10 s) VGD ON 22012 2230x Wait for D24ON (timeout 10 s) 2299x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 95: Sad Level Test

    ≤ 1689 µs, then acknowledge with switch-off VGB Acknowledge with switch-off VGB 31002 – 31007 3120x Wait for B24OFF (timeout 10 s) VGB ON 31008 3130x Wait for B24ON (timeout 10 s) 3199x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 96: Equality Test For Blood Side Pressure Sensors

    ±40 mmHg. (Error) the test prerequisites from HDFonline-Task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). Completed Close VPD B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 97: Pressure Test Substitution Line S-Online Hdf Online

    Wait for B24ON (timeout 15 s) 4070x (Error) the test prerequisites from HDFonline task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). 4099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 98: Blood Side Pressure Retention Test

    (Timeout 35 s since start of subcode 400) Close VBT Test VBT closed 33039 |PA — PV| ≥ 200 mmHg 33040 Acknowledge with inversion of LLS_SELFTESTBIT 3350x Wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 99: Disinfection Valve Vd

    Wait for B24OFF (timeout 7.5 s) VGB ON 37010 3730x Wait for B24ON (timeout 5 s) 3780x (Error) at start of test one or both couplings connected to rinsing bridge. 3799x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 100: Lx800 Motherboard

    Hard Disk Drive USB8 USB7 USB6 USB5 USB4 Floppy Disk Drive USB2 USB1 USB1 Service SW/FSU PS/2 COM2 COM3 COM4 COM1 USB3 (DIABUS) (ABPM) (Touch Screen) (FPB) Type A B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 101: Ide Interface Lx800 Motherboard

    3.4.5. USB7 USB6 USB5 USB4 USB1 LX800 Assignment USB Ports USB8 USB1 USB2 USB3 USB4-8 USB Service SW/FSU Card Reader DSI (CritLine) Adimea (Kt/V-UV) Fig.: USB Ports B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 102: Detection/Boot Sequence From Bootable Usb Sticks

    3 — 40 SW 9.xx 3.4.4 Detection/Boot Sequence from Bootable USB Sticks Use only USB sticks supplied by B. Braun as tool for the installation of the software. Dialog+ machines detect automatically connected bootable USB sticks. This is indicated by the red Installation text in start screen.

  • Page 103: Bios Settings For Lx800 Motherboard

    3. Press the TAB key when the logo is displayed. Main Processor..4. Press the Pause/Break key to stop the boot routine. The BIOS ..version is displayed, e.g. BIOS Revision 1.04..Pause Break B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 104
    [Normal High] SHFCLK Active Period [Free Running] LP Active Period [Free Running] Onboard Audio [Disabled] Onboard USB1.1 [Enabled] Onboard USB2,0 [Enabled] Onboard DIE [Enabled] Memory Hole At 15M-16M [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 105
    PNP OS Installed [No] Init Display First [PCI Slot] Reset Configuration Data [Disabled] Resources Controlled By [Auto(ESCD)] x IRQ Resources Press Enter x Memory Resources Press Enter PCI/VGA Palette Snoop [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 106: Pc Health Status

    ! ! ! ! » » » » # # # # $ $ $ $ :Move Enter:Select +/-/PU/PD:Value F10:Save Esc:Exit F1:General Help F5: Previous Values F6: Fail-Safe Defaults F7: Optimied Dafaults B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 107
    ! ! ! ! » » » » # # # # $ $ $ $ Esc : Quit : Select Item F10 : Save & Exit Setup Save Data to CMOS Delete Enter Delete Arrow Keys B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 108: Installation Software

    Open rear door. Switch to TSM service program mode (on digital board DB, service switch S1 position 2). Switch on Dialog+. Document default values in default table. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 109
    Position 2: TSM Service Program Mode Supervisor LEDs 0 to 7 (D28 – D35): Position 3: Software Installation/Update Mode Status 0 – 7 for Installation of LLS Software Jumper JP2 SWP: Supervisor Write Protect B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 110
    JP1 and JP2 have to be set to this position (Default CWE and SWE: controller/supervisor write enabled). Set the jumpers back to the CWP/SWP setting after installation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 111
    Bootable USB stick with the TLC/LLC/LLS software 14. Open front door. 15. Connect USB stick A to USB port B on the TLC sub-rack C. 16. Switch on machine. Fig.: Front View TLC Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 112
    USB Stick not Detected The system does not detect the bootable USB stick B BRAUN (indicated by the green logo in the start screen). If necessary disconnect the USB stick and reconnect. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 113
    If YES, press Y or y and <CR> if keyboard is used otherwise press the EQ hardware key! If NO, press N or n and <CR> on keyboard or any other hardware key! Press to confirm the selected CFC. [EQ] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 114
    Don’t forget to set the operation mode switch to TSM mode (2) and check in TSM all subsystems’ version number. Switch the power OFF and REMOVE the USB stick before starting the Dialog… The TLC software was installed successfully. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 115: Boot Machine /Check Version Number

    Enter Customer Specific Default Values The customer specific default values of the machine must be entered after the software installation. 27. Enter customer default values from default table and save B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 116: Enter Specific Parameters

    33. Set system configuration (customer specific values) or if available use the FSU USB stick with the system configuration. 34. Save settings. 35. Switch off machine. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 117: Self Test

    (The ABPM function (if present) can be checked without a test run of the machine in preparation mode, without the concentrate pump.) 12. Terminate dialysis. 13. Perform disinfection/decalcification. 14. Remove blood lines, filter and heparin syringe. 15. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 118: Field Service Utilities Fsu

    The field service utilities has the following directory structure (see left figure). BlueCat DISKTYPE USER 04_Write_Languages 05_Write_Calibration_Data 06_Write_System_Configuration 09_Card_01_Read_Card_Image 09_Card_02_Write_Card_Image 09_Card_03_Read_Patient_Diskette 09_Card_04_Write_Patient_Diskette_to_Card 09_Card_05_Read_Patient_Card 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends 13_Read_Screenshots 14_Get_Error B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 119: Start Field Service Utility Fsu

    To prevent the corruption of USB stick, before removing it, please press AQ, then follow the instructions on the screen! 4. Press the EQ hardware key to execute the utility. Press the AQ to exit the FSU program. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 120: Set-Up Fsu Main Menu

    Card Commands in the main menu. 2. Press the EQ hardware key to execute the utility. 3. Press the BP hardware key and move the cursor down to select a utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 121: Hardware Keys For Fsu

    Press the AQ hardware key to exit the FSU menu or to skip back to the previous menu. Press the EQ hardware key to execute a selected utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 122: Set System Version Number

    BP stop stop hardware keys. Note: (n.nn = the version number) 4. Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 123: Set Working Time Counter (Wtc)

    3. Set the new working time counter with the BP and BP hardware keys. stop stop Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 124: Kill Masterboot Record

    A r e y o u s u r e ? 3. Press AQ if you do not want to delete the masterboot record. AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 125
    A r e y o u R E A L L Y s u r e ? the masterboot record. 5. Press BP to cancel the operation. BP-: no BP+: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 126: Write Utilities

    The last write procedure can be undone with the 07 Undo Last Written utility. The Dialog+ can be switched off and on again between the write procedure and using the 07 Undo Last Written utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 127
    After successfully copying the language to the CFC the version numbers of all languages available on the Dialog+ are displayed. The language selected for treatment is also marked. 4. Press EQ to continue. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 128: Write Calibration Data

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 129
    (SYSver) stored in the calibration data file with the serial number and the system version number from the Dialog+. The calibration data file can be written only if they match. Example Ser-No.= 90832 SYSver = V8_20 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 130
    The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the calibration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 131: Write System Configuration

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 132
    Yes, system configuration file is written to Writing System Configuration to Dialog+ from: TLCver=90832/VX_XX Dialog+ hard disk drive (CFC) A r e y o u s u r e ? AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 133
    The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the system configuration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 134: Undo Last Written

    3. Press EQ if you are sure. This will undo the last written utility from the USB stick to the Dialog+ hard disk drive (CFC). AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 135: Usb Stick Self Test

    Searching for md5sum files….the main menu after the self test. MD5SUMS.XXU.Win found..MD5SUMS.XXU.Lin found..MD5SUMS.SPEC.Lin found..Checking file system integrity..O.K. Calculated CRC = XXXX Reference CRC = XXXX B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 136: Card Reader Read/Write Utilities

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 137: Read Card Image

    E n d o f R e a d C a r d I m a g e P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 138: Write Card Image

    E n d o f Write Card Image P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 139: Read Patient Diskette

    P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 140: Write Patient Diskette To Card

    E n d o f Write Patient Diskette to Card . . . P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 141: Read Patient Card

    P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 142: Write Patient Card

    E n d o f Write Patient Card . . . P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 143: Remove All Patient Data

    05. Read Patient Card (USB Stick → Card) 06. Write Patient Card 07. Remove All Patient Data BP-: move cursor up BP+: move cursor down AQ: back to previous menu execute B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 144
    U S E R / 0 9 _ C a r d _ 0 3 _ R e a d _ P a t i e n t _ D i s k e t t e ….E n d o f Remove All Patient Data . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…
  • Page 145: Read/Get Error Utilities

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 146
    ….displayed….End of Read System Configuration… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 147: Read Calibration Data

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 148
    ……. . End of Read Calibration Data… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 149: Read All Trends

    Year 2007 09_Card_03_Read_Patient_Diskette M M : 09 month September D D : 12 Wednesday 09_Card_04_Write_Patient_Diskette_to_Card h h : 14 hours 09_Card_05_Read_Patient_Card m m : 25 minutes 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 150
    ……. . End of Read Trends… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 151: Read Screenshots

    Dialog+. The vga*.zip files copied to USB stick are deleted from the Dialog+ hard disk drive (CFC) (similarly to the log files) but the subdirectory structure on the Dialog+ hard disk drive (CFC) is left untouched. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 152
    ……. . End of Read Screenshots… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 153: Get Error (Errordisk)

    3 — 91 SW 9.xx 3.6.10.5 Get Error (ErrorDisk) The error files can only be analysed by qualified personnel of B. Braun Avitum (e.g. development department) and should be provided only on special request. The utility gets error information and log files from Dialog+ hard disk drive (CFC) and writes them to the USB stick.

  • Page 154
    Dialog+ hard disk drive (CFC) to the USB stick… The Get Error (ErrorDisk) menu automatically returns (a ..few seconds) to the main menu End of Error Disk… is displayed….. End of Error Disk… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 155: Fsu Handling Of Errors

    USER/XX_Read..USER/14_Get_Error These files can be copied to any location on the PC (e.g. to use them for local examination) or can be deleted from the FSU USB stick. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 156: Troubleshooting Usb Stick

    FSU program. — — — S W I T C H O F F t h e D i a l o g + — — — B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 157
    Extended TMP Limit Range Button is Displayed in Treatment Select Extended TMP Limit Range inactive inactive inactive Arterial Bolus Parameters Arterial Bolus Volume Arterial Bolus with SAKA Support B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 158
    2000 2000 Therapy Time h:min 04:00 04:00 03:30 Gross UF Rate vs. Blood Flow inactive inactive inactive Gross UF Rate vs. Blood Flow active active active UF Profile Editor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 159
    Default Setting for Treatment Mode: Warning of Kt/V Target-Projected Deviation Warning active Warning active Warning inactive Warning for Saving Kt/V Table Timings Parameter Setting Window Disappearing Time Screen Saver Appearance Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 160
    Load Interval inactive inactive inactive Elapsed Month inactive inactive inactive Setting Elapsed Month months Elapsed Working Time Counter WTC inactive inactive inactive Setting Elapsed WTC hours 4000 4000 4000 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 161
    HDF Online Substitution Pump OSP (2.15) Unit Default US Default Modified µl/Head Revolution 8300 8300 Level Reg ulatio n ( 2.16) Unit Default US Default Modified LRP (PPR) Slow LRP (PPR) Fast B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 162
    Morning 06:30 Rinsing Tuesday Night Morning 06:30 Rinsing Wednesday Night Morning 06:30 Rinsing Thursday Night Morning 06:30 Rinsing Friday Night Morning 06:30 Rinsing Saturday Night Morning Sunday Night Morning B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 163
    Temperature Conductivity Name Volume Long Short Time min. (max. 20 Characters) h:min h:min h:min mS/cm Citric Thermal 00:15 00:05 00:05 Decalcification Short 00:05 00:05 00:05 Bleach 00:15 00:05 00:25 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 164
    Lo 15 xevonta Lo 18 xevonta Lo 20 xevonta Lo 23 xevonta Hi 10 xevonta Hi 12 xevonta Hi 15 xevonta Hi 18 xevonta Hi 20 xevonta Hi 23 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 165: Measures After Repair

    4.9.4.19 Calibration Dialyser Inlet Throttle Valve DDE 4.9.4.20 Calibration Pressure Reducer Valve DMV 4.9.4.21 Gap SAKA and SAKV 4.9.4.23 Setting Servomotor for Disinfection Valve VD Inspection Protocol for Automatic Blood Pressure Measurement ABPM B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 166: Tests And Measures To Be Executed

    LLC Calibrations Check and set ratio, tubing constants BPA/BPV, OSP tubing type, check/set PA/PV limit values. Working Time Counter • Set working time counter. Motherboard Check and set BIOS. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 167
    5. Set the blood pump so that no blood side alarms are activated. 6. Start dialysis. 7. Check alarm function of SAD. (The ABPM function can be checked without a test run of the machine in preparation mode.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 168
    1. The measures were executed correctly. Fill in the execution protocol after service (e.g. exchange of a component) and after executing all necessary measures according to the repair matrix. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 169
    Inspection Protocol for Automatic Blood Pressure Measurement ABPM} Name Service Technician: The respective measures were executed correctly after servicing according to the repair matrix…………………………..Date / Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 170
    Timings 4-49 4.7.1.15 Language Selection 4-50 4.7.1.16 Automatic Blood Pressure Measurement ABPM 4-51 ® 4-52 4.7.1.17 bioLogic RR GL Parameters 4.7.1.18 Summer Time Setting 4-53 4.7.1.19 Miscellaneous Parameters 4-54 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 171
    Test 1.14 Bypass and Disinfection Valves 4-88 4.8.2.15 Test 1.15 Flow Pump FPA, Balance Chamber Sensors and Air 4-89 Separator 4.8.2.16 Test 1.16 Pressure Water Part 4-91 4.8.2.17 Test 1.17 UF Pump 4-92 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 172
    4.8.3.10.3.1 Turbidity Calibration with Water 4-144 4.8.3.10.3.2 Calibration with Reference Solution (Red Dye) 4-144 4.8.3.10.4 Actual Value Check Blood Leak Detector 4-145 4.8.3.10.4.1 Check 0 ‰ Blood Concentration 4-145 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 173
    4-170 Currentless Closed SAKV-SG 4.8.3.27 Setting Servomotor for Disinfection Valve VD 4-171 4-173 Production Report 4-174 4.9.1 Version Number Dialog 4-175 4.9.2 Low Level Options 4-176 4.10 Service Reports B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 174: Start Tsm Service Program

    Position 3 = Software Installation/Update 4. Switch on machine. The machine starts in TSM mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 175
    The File Operations menu can be activated with this icon. The current software/firmware versions are displayed in the Dialog Version Number menu. This menu can also be selected in the Production Report. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 176: Quit Tsm Service Program

    3. Close rear door. 4. Switch on machine. The machine starts in therapy mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 177: Structure Of Tsm Service Program

    Calibration 2.13.2 Pump Head Selection {} : not applicable {Calibration 2.14 Substitution Pump} Calibration 2.15 HDF Online Substitution Pump OSP Calibration 2.16 Level Regulation Pump LRP Fig. : TSM Service Program B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 178: Service Overview In Therapy

    7500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 179
    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Second Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 180
    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Third Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 181: Legend Service Overview In Therapy

    LS CURRENTS_S: Loudspeaker current supervisor LS PULSE_S: Loudspeaker pulse supervisor MSBK1: Membrane position sensor balance chamber 1 MSBK2: Membrane position sensor balance chamber 2 NSVB: Level sensors upline tank B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 182
    VDE: Dialyser inlet valve VDEBK1_S: Dialyser inlet valve balance chamber 1 supervisor VDEBK2_S: Dialyser inlet valve balance chamber 2 supervisor VDE-S: Dialyser inlet valve supervisor VDFF: DF filter valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 183
    LLC text messages according to the LLC message hex numbers from LLC L L C M e s s a g e W i n d o w message window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 184: Service Overview In Tsm Service Program

    5500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 185: Legend Service Overview In Tsm Service Program

    Dialysate outlet pressure sensor supervisor Degassing pressure sensor Venous pressure sensor PV_S: Venous pressure sensor supervisor RDV: Venous red detector RDV_S: Venous red detector supervisor RVE: Degassing control valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 186
    VSAE_S: Substitution inlet valve supervisor VSB: Substitution bypass valve VSB_S: Substitution bypass valve supervisor VVBE: Upline tank inlet valve Circulation valve VZ_S: Circulation valve supervisor WTC: Working time counter B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 187: File Operations

    LLC/LLS after every calibration. The calibration data can be downloaded from the compact flash card CFC (hard Down Load Calibration Data disk drive) to the exchanged digital board DB. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 188
    • Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 189: Treatment Support

    The new settings are valid after the restart of the machine. There are five trend groups, which can be selected for the treatment program. Trends and Trend Groups Every group has three treatment parameters. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 190: System Configuration

    1. Click the mmol/l or mS/cm key to select a measurement unit. The acetate mode can be disabled. Acetate Mode Disabled 1. Click the YES key to disable the acetate mode. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 191
    [mS/cm per mmol/l]. Conversion Factor: Conductivity mS/cm (Standard Value) Conversion Factor NaCl Concentration mmol/l (Measured Value) 4. Click the Default Value window if necessary to change the value [mS/cm]. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 192
    [mS/cm]. 4. Click the ACID icon if necessary to edit the following parameters: • Acid Concentrate Name • Conversion Factor [mS/cm per mmol/l] • Default Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 193
    1. Click the ml/min key to set the dialysate flow. You can set the dialysate temperature for therapy. Dialysate Temperature 1. Click the C key to set the dialysate temperature. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 194: Blood Side Parameters

    The selected Chamber-free AV Line does not appear. AV Line with PBE Connector Activate if AV lines with PBE pressure connection are used. Do not activate if AV lines without PBE pressure connection are used. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 195: Min-Max Parameters

    You can set the upper relative alarm limit for the Delta PA max. PA limit window. Delta PA PA max. max. PA ref 200 mmHg PA min. Fig.: Limit Window for PA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 196
    The TMP low limit range is set min. –10 mmHg. If the key is not pressed The TMP limit range is extended to –100 mmHg If the key is pressed (absolute TMP value). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 197
    • a warning message is not displayed anymore TM P 20 mmHg 12 — 28 m m Hg • treatment is interrupted • machine is switched to bypass -100 Bottom TM P Lim it Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 198: Arterial Bolus Parameters

    Arterial Bolus with SAKA Support Arterial Bolus Parameters menu. Prerequisites are: • Arterial tubing clamp SAKA present • An arterial line system with infusion access is present between the SAKA and blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 199: Single-Needle Parameters

    Furthermore the control pressure PV can be set. Thereby the blood pump is started/stopped and the venous tubing clamp currentless closed SAKV-SG is opened/closed. The control pressures for PA and PV can be set for single-needle cross-over. Single-Needle Cross-Over B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 200: Heparin Parameters

    Heparin Bolus Volume Therapy Beginning Bolus An automatic bolus can be administered at the beginning of a therapy. The heparin bolus is performed by the venous red detector RDV. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 201
    You can set the syringe name and the inner diameter of the syringe. Heparin Pump Compact 3. Activate the Selected Syringe menu with the Selected Syringe icon. The following menu is opened. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 202: Rinsing Parameters

    • Rinsing Blood Flow • Rinsing Dialysate Flow • Rinsing Time by UFP • Rinsing Rate by UFP • Rinsing Volume by UFP • Blood Flow for Connecting Patient B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 203
    You can set the blood flow for connecting the patient (in therapy). The blood Blood Flow for Connecting Patient pump BPA is started with this rate when the therapy is started. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 204: Uf Parameters

    The alarm can be acknowledged with the key. Setting Range The alarm limit can not be set lower than the warning limit (even if the warning limit is deactivated). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 205
    A linear adaptation is calculated for the remaining process. EXP (Exponential) If this key is activated only the first interval can be set. An exponential adaptation is calculated for the remaining process. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 206
    The UF process of the activated interval is set to HD with the HD key. UF Mode Sequential SEQ The UF process of the activated interval is set to sequential with the SEQ key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 207: Hdf/Hf Online Parameters

    HDF Dialysate Flow for Predilution HDF/HF Infusion Bolus Volume The HDF/HF infusion bolus volume can be set. You can set predilution for a HDF/HF online therapy by activating the Post/Predilution Selection Predilution key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 208: Disinfection Parameters

    • Auto Switch-On without Daily Confirmation • Maximum Out of Action Time (Day/Hour/ Minutes) • Disinfection Configuration Data Chemical (long), Chemical (short) • Disinfection Configuration Data Central Thermal, Central Chemical, Rinsing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 209
    The machine is rinsed after the preselected switch-on time. The key is only enabled if the function is selected. Switch-On Time The switch-on time can be set (without a weekly schedule). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 210
    The function is activated if the Limit OFF key is not pressed. The Limit OFF key is activated in the default setting, the parameter keys are disabled, i.e. parameter values can not be set. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 211
    • Rinsing Time Central Automatic Chemical: • Inlet Volume • Switch-off for Retention without Automatic Switch-on • Retention Time • Rinsing Flow • Rinsing Time Rinsing: • Inlet Flow • Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 212
    FBK1 VEBK1 VDABK1 M SBK1 VDEBK2 VABK2 BICP LAFS VEBK2 VDABK2 FBK2 M SBK2 RVDA TSHE RVFPA RVUFP Fig.: Flow Diagram — Central Automatic Chemical Disinfection of Water Inlet B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 213
    The rinsing flow can be selected. Rinsing Time The rinsing time can be selected. Central Chemical A central chemical mode can be selected for the automatic switch-on in the weekly schedule. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 214: Filter Parameters

    The operating hours can be set for the intended operation time of the DF/HDF Filter Operation Time [h] filters. The operating hour counter is active when DF Preparation is active. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 215: Dialyser Parameters (Kt/V)

    V (litres)=2.447 – 0.09516 x Age (years) + 0.1074 x Height (cm or feet/inch) + 0.3362 x Dry Weight (kg) Female: V (litres)=-2.097 + 0.1069 x Height (cm or feet/inch) + 0.2466 x Dry Weight (kg) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 216
    The Kt/V is approximately calculated via the Dialyser Filter Urea Clearance Data for the selected dialyser (see table, use scroll bar to see table). Details can be found in Chapter 3 repair instructions. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 217: Kt/V-Uv Parameters

    Default Setting for Treatment Mode: value, a message is activated. Warning of Kt/V Target-Projected Deviation A warning message can be enabled for saving the Kt/V table. Warning for Saving Kt/V Table B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 218: Timings

    Therapy and end of therapy: Time cake, remaining time, status line Disinfection: Time cake, disinfection mode, phase You can set the appearance time of the Screen Saver. Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 219: Language Selection

    English. If German is selected the therapy and TSM language is German. If an additional language is selected this language is used in therapy. The TSM language is English. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 220: Automatic Blood Pressure Measurement Abpm

    The limit values in therapy, e.g. for systole/diastole can be stored on a patient Taking Over Limits from Patient’s chip card. These limit values are taken over if the patient chip card is read. Parameters B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 221: Biologic Rr ® Gl Parameters

    The ID number 200 indicates that the Guide Line Mode is enabled and can be used in therapy. ID Number 100 The ID number 100 indicates that the Guide Line Mode is not enabled and can not be used in therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 222: Summer Time Setting

    If necessary you can set the time difference between summer and winter time. Time Difference Note: The Time Difference should be set to 00:00 for countries where daylight saving time is not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 223: Miscellaneous Parameters

    It is prohibited to run a therapy with a patient if the Skip Self Test function is activated. The self test must be passed successfully prior to a therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 224
    A window is automatically opened at the end of therapy. This window must be Automatic Reinfusion Start at confirmed at the end of the therapy for the reinfusion of the blood. Entering End of Treatment B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 225: Dbi Parameters

    Nexadia functionality, e.g. support of the option bioLogic RR Comfort and Adimea; the mmol/l unit can be used. The Nexadia protocol must be supported by the host computer. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 226: Preventive Maintenance

    9 to 18 month period if the 4,000 working time is more than 4,000 hours after the last reset. W indow 4,000 hours vs 9 to 18 months Elapsed Months B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 227
    Press the Preventive Maintenance Accomplished key and then the Preventive Maintenance Accomplished Save icon after the preventive maintenance was performed. Consequently a new preventive service period is started again after the machine is switched off and on. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 228: Trends And Trend Groups

    Trend and Trend Groups key in the Treatment Support menu. The following menu is opened and displayed. 3. Activate the Trend History menu with the Trend History window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 229
    5. Activate the Edit menu with the Edit icon. 6. The treatment parameters in the six trend groups can be modified. 7. The default parameters can be reset with the Set Defaults window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 230: Manual Test And Calibration

    • Supervisor Window Test • File System Check • ABPM Maintenance (Option ABPM) • DBI Maintenance (Option Nexadia-BSL) • Kt/V-UV Sensor Test (Option Adimea) • DSI Test (Option Crit-Line) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 231: Buttons, Lamps And Sounds

    Check red LED for signal lamp (on optical status display boards OSD). Check yellow LED for signal lamp (on optical status display boards OSD). Check green LED for signal lamp (on optical status display boards OSD). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 232: Monitor Test

    1. Activate the Monitor Test menu with the Monitor Test key in the TL Manual Test and Calibration menu. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 233: Touch Screen Calibration

    (on the keyboard membrane of the TFT housing) three times in the TSM Main Menu. 2. Switch machine off and on again. The old default setting is loaded. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 234: Brightness Calibration

    The day brightness (digits) for the backlight set. Test OSD Red and Yellow (for Signal Lamps) The red and yellow LED of the optical status display OSD can be checked, i.e. switched ON and OFF. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 235: Card Reader And Card Check

    • Use an empty service chip card (art. no. 7703856) for the test and not a patient chip card. • Do not remove the chip card from the card reader during the test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 236: Usb-Port Properties

    The maximum speed of the USB port. Status The connected USB device is in the valid USB device list. Not OK The connected USB device is not in the valid USB device list. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 237: Supervisor Window Test

    Parameter Check key. The following menu is displayed. 3. Press the key to close the supervisor window. Press the key to select and acknowledge the single parameters and close the supervisor window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 238: File System Check

    Hard Disk Check (for the compact flash card) to test the hard disk drive. Press the key of the Application SW Check to test the application software. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 239: Abpm Maintenance

    ABPM system is tested. The measurement chamber is connected to the ABPM system via the pressure tubing. A pressure is built up to 300 mmHg and is held for 3 minutes. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 240: Dbi Maintenance

    5. Switch off machine after successful test. 6. Remove the test plug from the RS 232 port. Pin Assignment RS 232 Test Plug (Soldering Side) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 241: Kt/V-Uv Sensor Test

    (residual digits). This value must be ≤4000 digits (maximum service life). Example: Required current: 1332 (digit) Reserve: 2668 (digit) Total digits: 4000 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 242: Dialog Serial Interface Dsi Test

    If the Sent Data and Received Data are identical, the test was successful. 6. Switch off machine after successful test. 7. Remove the test plug from the DSI interface. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 243: Low Level Manual Test

    Previous and Next Arrow keys. If required, a corresponding calibration menu can be selected in certain LL Manual Test menus. The calibration procedures are described in paragraph 4.8.3 Low Level Manual Calibration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 244: Test 1.1 Staff Call, Alarms And Power Off

    • -12 VAN > -11 V -12 VAN Control / Supervisor ON Test activated ±12 VAN Test: • +12 VAN < 11 V • -12 VAN < -13 V B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 245: Test 1.2 Blood Pump Arterial

    The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPA is stopped. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 246: Test 1.3 Blood Pump Venous

    The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPV/substitution pump is stopped immediately. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 247: Test 1.4 Heparin Pump

    (The clasp nut is assembled on the drive spindle and limits the max. pressure load of the heparin pump). Activates/switches off the heparin pump. HP Motor State Selection between slow and fast forward feed. HP Speed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 248: Test 1.5 Tubing Clamps

    Thereby the switch function and leakage test (pressure — clamp SAKV and SAKA gap, time) can be tested. The closed clamp gap is set to 1.4 mm + 0.1 mm. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 249: Test 1.6 Pressure Sensors (Blood Side)

    Machine with Single Pump The value for PBS is 0. Machine with Double Double Pump must be activated in Production Report, Low Level Options. A value for PBS is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 250: Test 1.7 Air Sensor Sad, Red Sensor

    A tubing filled with fluid is inserted in the SAD If the SAD TEST is activated NO AIR must be detected. The light path is interrupted NO BLOOD (Controller/Supervisor) changes to BLOOD (Controller/Supervisor). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 251: Test 1.8 Coupling Detectors

    You can check the actual status of the reed sensors. Detectors Bicarbonate rinsing connection sensor. BICSS Concentrate rinsing connector sensor. Rinsing bridge connector sensor (dialyser coupling). SBS1 SBS2 Rinsing bridge connector sensor (dialyser coupling). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 252: Test 1.9 Water Inlet, Upline Tank And Flow Pump Fpe

    All balance chamber valves are opened and the outlet flow pump FPA works in main flow with the same speed as the inlet flow pump FPE. The degassing pump EP runs and the degassing bypass valve VEB is open. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 253: Test 1.10 Degassing And Heating

    C or the upline tank VB is empty. The degassing pump EP must be able to build up a flow of > 1200 ml/min at a negative pressure of — 500 mmHg. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 254: Test 1.11 Dialysis Temperature

    Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. The measurement range is as follows: TSE, TSD, TSBIC: 0 to 100 TSD-S: 30 to 55 TSDE: 30 to 55 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 255: Test 1.12 Concentrate Pumps And Conductivity

    Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. Thereby the pulsating flow of the pumps have no influence on the conductivity measurement. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 256: Test 1.13 Balance Chamber Valves

    The following valves can be activated: Valves VDEBK1/2 Dialyser inlet valve 1/2 balance chamber VEBK1/2 Inlet valve 1/2 balance chamber VDABK1/2 Dialyser outlet valve 1/2 balance chamber VABK1/2 Outlet valve 1/2 balance chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 257: Test 1.14 Bypass And Disinfection Valves

    The following valves can be activated: Valves Bypass valve Dialyser inlet valve Dialyser outlet valve Disinfection valve Circulation valve Air separator valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 258: Test 1.15 Flow Pump Fpa, Balance Chamber Sensors And Air

    FPE. Simultaneously the fluid in the right side of the balance chamber is emptied to the drain. Drain Fig. : Filling Cycle Balance Chamber Phase 1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 259
    Thereby FPE delivers fluid into the Fig. : Filling Cycle Balance Chamber Phase 2 dialyser circuit. VLA is only closed and switched to normal operation if the top level sensor LAFSO detects fluid. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 260: Test 1.16 Pressure Water Part

    The pressure can be displayed in mmHg or in mbar, depending on selection. Conversion Factor mmHg to mbar Conversion Factor = 1.3332 Example: -100 mmHg approx. –133 mbar –400 mmHg approx. –533 mbar B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 261: Test 1.17 Uf Pump

    UFP, set by the value of the throttle DDE. Not applicable. Flow Detector Stroke Value UFP The Stroke Value UFP should not be used in a normal case, because UFP must be calibrated (see LL Manual calibration). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 262: Test 1.18 Blood Leak

    If air is visible check the complete system for leaks in menu 1.20 Test Water Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 263
    LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 — 1.44 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 264: Test 1.19 Leak Test

    The test is started with a negative test pressure and then followed by a positive test pressure. The leakage test consists of three test steps. If an error occurs an LLC alarm is LLC Alarm activated (LLC Status 00020000). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 265
    VDEBK1 VABK1 FBK1 Z.D. VEBK1 VDABK1 M SBK1 Dialyser VDEBK2 VABK2 V.D. LAFS VEBK2 VDABK2 FBK2 M SBK2 SBS1 RVUFP RVFPA SBS2 Fig.: Excerpt Flow Diagram for HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 266
    Rinse water circuit to prevent air bubbles. b) Close the outlet. c) Activate test. The FPE builds up a pressure. d) Check pressure drop. The pressure drop must be < 50 mmHg in 30 s. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 267: Test 1.20 Water Part (Overview)

    Valves VDA/VDE are open (valve symbols are black). Valve VBP is closed (valve symbol is white). Bypass Valves VDA/VDE are closed (valve symbols are white). Valve VBP is open (valve symbol is black). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 268: Test 1.21 Emptying

    The actual value, i.e. the fluid level of the upline tank VB is displayed by the level sensors NSVB. Start the Emptying ON at least three times and manually clamp the bypass (tubing) to ensure a sufficient emptying (for transportation). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 269: Test 1.22 Hdf

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 100 SW 9.xx 4.8.2.22 Test 1.22 HDF Not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 270: Test 1.23 Bic Cartridge Valves And Coupling Detectors

    Valves VVB, VBICP, VBKO, VBKS The valves VVB, VBICP, VBKO and VBKS can be switched either separately or together. The status of the reed sensor (OPEN/CLOSED) is displayed. Detectors BKUS, BKUS_S B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 271: Test 1.24 Hdf Online

    DF filter and HDF filter are impermeable to air. Valve VBE can only be opened if PDA < -100 mmHg to prevent fluid penetration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 272: Test 1.25 Hdf Online Valves And Coupling Detectors

    The status of the additional coupling sensor FEDFFS can be checked. FEHDFS (filter recognition HDF filter) The status of the additional coupling sensor FEHDFS can be checked. FEHDFS PSAUS PSABFS FEDFFS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 273: Test 1.26 Option Battery

    Akku OK (Battery OK) time Battery capacity < required capacity for 20 minutes battery running time External State Indicates if machine was switched on externally. Machine was switched on externally B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 274: Test 1.27 Dfs Pressure Test Of Therapy

    The test codes for LLC and LLS are displayed for all test phases. In case of an Test Code error the defective components can be determined with the help of the test description. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 275: Test 1.28 Front Panel Board Fpb (Lls)

    FPB and wait for switching (OFF/ON). • Enter Key Press the key on the front panel board FPB and wait for switching (OFF/ON). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 276: Test 1.29 Patient Leakage Current

    C and is then switched off. The blood pump can be activated additionally if the measurement is performed with tubing set (line). Main Flow The machine can be switched between main flow and bypass. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 277: Test 1.30 Smps-Mc Self Tests

    > 30 s (disconnect mains plug from mains receptacle). #115 Buzzer Test by LLS Ver/SW The software version of the SMPS-MC is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 278: Self Tests Smps-Mc

    • LLC sends #115 to SMPS-MC • LLS must activate AKAL-S within 5 s • Buzzer OK. if not OK: not OK: • Warning is generated • Buzzer not OK. • Therapy still possible B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 279: Test 1.31 Level Regulation

    The delivery of the LRP can be checked with the Test Test key. A measurement chamber must be connected to PV. LRP stops when 500 ml are reached and the required time for the pressure build-up is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 280: Low Level Manual Calibration

    Input values can be increased in increments with the displayed hardware key as an alternative to the software key on the touch screen. The service overview can be selected in the TSM with the hardware key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 281
    If required, the previous calibration menu or the next calibration menu can be selected with the Previous and Next Arrow keys. The menu can be closed with the Exit Door key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 282
    C o n t r o l l e r / S u p e r v i s o r ( 1 2 b i t ) Voltage 0 to 5 V Digits 0 to 4095 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 283: Select Low Level Manual Calibration

    • 2.12 PV Alarm Window, PA Low Limit • 2.13 Flow of BPA and BPV Pump • 2.14 Substitution Pump (not applicable) • 2.15 HDF Online Substitution Pump OSP • 2.16 Level Regulation B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 284: Calibration Pressure Sensors Blood Side Pa, Pv, Pbe And Pbs

    LL Manual Calibration key in the Manual Test and Calibration menu. Activate the Pressure Sensors Blood Side menu by pressing on the line Calibration 2.1 Pressure Sensors Blood Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 285
    Build up a pressure of -100 mmHg (+10 mmHg) with a syringe and clamp the tube. -100 mmHg Press key. Wait until digits are stable. Enter the stable value from the pressure measurement instrument as reference value. Confirm with the key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 286
    Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Confirm and store completed calibration with key. 17. Remove pressure measurement instrument, syringe and tubing. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 287: Heparin Pump

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 118 SW 9.xx 4.8.3.4 Heparin Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 288: Calibration Degassing Pressure Sensor Pe

    You can calibrate the degassing pressure sensor PE. Build up the negative pressure in the degassing circuit with a syringe. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* VBKO* TSHE VVB* VBKS RVBO* Fig.: Degassing Circuit and PE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 289
    TSHE Fig.: Water Sub-Rack with Loop-In for Manometer at PE and Tubing Fig.: Water Sub-Rack with Tubing Clamp between FEP and EP (front Clamp at VEB (rear view) view) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 290
    Manual Test and Calibration menu. 2. Activate the Pressure Sensors (Water Side) menu by pressing on the line Calibration 2.3 Pressure Sensors (Water Side). The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 291
    PE should show a of approx. < -35 mmHg/-50 mbar • Close VEB and increase speed to 2,100 rpm • PE should show a value of approx. < -500 mmHg/-670 mbar B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 292: Calibration Degassing Pressure Sensor Pe For Machines At

    15. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Remove calibration devices. 17. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 293: Calibration Dialysate Outlet Pressure Sensor Pda

    VDA, it must be closed or the tubing must be clamped manually. FBK2 TSDE DF Filter PEV* Z.D. Dialyser V.D. LAFS FBK2 BL PDA SBS1 RVFPA SBS2 Fig.: Dialysate Circuit and PDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 294
    Conversion Factor mmHg to mbar Conversion Factor = 1.3332 Example: –400 mmHg approx. –530 mbar 0 mmHg 0 mbar +400 mmHg approx. +530 mbar B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 295
    LL Manual Calibration key in the Manual Test and Calibration menu. 2. Activate the Pressure Sensors Water Side menu by pressing on the line Calibration 2.3 Pressure Sensors Water Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 296
    21. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 22. Remove calibration devices. 23. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 297: Calibration Degassing Temperature Sensor Tse/Heater Inlet Temperature Sensor Tshe

    KP/BICP TSHE Keep tubings as short as possible TSHE RVDA RVDA Fig.: Original Tubing Water Sub-Rack for TSE/TSHE Fig.: Water Sub-Rack with TSHE in Line with TSE for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 298
    C) is displayed at the temperature measurement instrument. Wait until temperature is stable, i.e. 0.1 C for at least 10 s. Press key. 10. Enter the stable value from the temperature measurement instrument as reference value. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 299
    22. Stop degassing pump EP and flow pump FPE, and disconnect adapter set. 23. Reassemble the heater inlet temperature sensor TSHE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSHE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 300: Calibration Temperature Sensors Tsbic/Tsd/Tsd-S And Tsde

    Measurement Instrument ml/min RVFPE ENDLF-S TSBIC blue mS/cm BICLF ENDLF TSD-S TSDE Measurement Sensor BICP blue FBK1 VEBK1/2 Fig.: Dialysate Preparation with Loop-In Point for TSBIC/TSD/TSD-S and TSDE Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 301
    FBK1 Measurement Sensor blue TSDE Fig.: Tubing Rinsing Bridge with Removed TSDE Fig.: UF Sub-Rack with TSDE in Line with TSBIC/TSD/TSD-S (on DF Sub-Rack) for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 302
    Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. Press key for a combined calibration of TSBIC, TSD/TSD-S and TSDE (The circulation valve VZ stays closed). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 303
    4.8.3.1 General Calibration Information). 21. Reassemble the heater inlet temperature sensor TSDE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSDE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 304: Calibration Conductivity Sensors Bicarbonate Biclf And End Conductivity Endlf/Endlf_S

    If acid concentrates are not available the following B. Braun acetate concentrates can be used as an alternative: e.g. SW44 and SW 174 If acid concentrate from B. Braun is not available acid concentrate from a second source can be used. The composition is listed in table 1. Composition of Calibration Acid concentrate is diluted with deionised water with a ratio of 1:4 (or 1:4.5…

  • Page 305
    FBK1 Measurement VABK2 Sensor VDEBK2 FBK1 blue VEBK2 Fig.: Original Tubing UF Sub-Rack with FBK1 Fig.: UF Sub-Rack with Loop-In Point for BICLF, ENDLF and ENDLF-S Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 306
    Connect measurement instrument between FPE and balance chamber. Connect dialyser couplings to flow meter. Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 307
    14. Press key. Wait until digits are stable. 15. Enter the stable value from the measurement instrument as reference value. 16. Confirm with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 308
    Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 27. Stop FPE and BICP and remove measurement instrument. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 309: Calibration Blood Leak Detector

    SBS2 blue Fig.: Rinsing Bridge with Blood Leak Detector BL TSDE LAFS FBK2 SBS1 blue SBS2 Fig.: Rinsing Bridge with Red Coupling Immerged in Reference Solution for BL Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 310
    Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. Chemical Disinfection with Citric Acid 50% A decalcification must be performed before initial calibration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 311: Blood Leak Detector Limit

    Please Note The same blood leak detector is used for both settings: • The calibration procedure is identical, but with different tolerance ranges • The test procedure is identical B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 312: Check Blood Leak Detector

    LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 — 1.44 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 313: Calibration Blood Leak Detector

    The displayed actual blood concentration value must be in the following limit, to ensure a sufficient resolution: Actual Status Controller > 200 ‰ Blood Concentration [ Confirm and store completed calibration procedure with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 314: Actual Value Check Blood Leak Detector

    The blood concentration values for the controller and supervisor must be in the following limits: Actual Status Actual Status Controller Supervisor Blood Concentration 0.90 to 1.1 0.90 to 1.1 [‰] B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 315: Calibration Of Delivery Rate Ufp, Bicp And Kp

    • Balance or graduated cylinder • Tubing clamp Fill graduated cylinder with 200 ml water. Remove respective concentrate suction rod from rinsing chamber and immerge into graduated cylinder. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 316: Calibration Uf Pump With Balance

    12. Confirm and store completed calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 317: Position Setting For Bicp, Kp And Ufp

    It is recommended to calibrate the UFP with the procedure described (with balance) because the accuracy of the stroke value might be insufficient for a proper UF control of the system. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 318: Calibration Safety Air Detector Sad

    The calibration procedure is completed. 5. Press the key to store the calibration. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 319: Sad Test After Calibration

    Remove the tubing from the SAD. Interrupt the light path in the SAD with a piece of paper. The following settings must be displayed: RDV: NO BLOOD (Controller/Supervisor) changes to RDV: BLOOD (Controller/Supervisor) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 320: Calibration Biclf And Endlf Ratio

    Production Report, LLC Options, Holder for BIC Cartridge. You can determine the limit of the ENDLF ratio (measured at the end conductivity sensor ENDLF). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 321
    Connect the respective concentrate (bicarbonate/acid) after the request. Select service overview to display the current RATIO values. Wait till stable ratio values are reached. Use the mean value if there are slight deviations. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 322
    1. Enter 35 in Desired Setting Controller for the BICLF Cart Ratio. 2. Save the entered value with the Calibration key. 3. Check the bicarbonate ratio value in therapy. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 323: Calibration Load Cell

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 154 SW 9.xx 4.8.3.14 Calibration Load Cell Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 324: Calibration Pv Alarm Window, Pa Low Limit

    The supervisor uses a tolerance of 10 mmHg for monitoring the PV limit values to prevent double alarms. The set PA Absolute Low Limit value (-200 mmHg) is used as limit in therapy PA Absolute Low Limit and TSM service program/system configuration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 325: Calibration Flow Rate Of Bpa And Bpv/Osp

    Pump Head Selection. The following menu is displayed. You can select the pump head roller type used in the blood pumps. • Pump head 8 x 12 • Pump head 7 x 10 (option) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 326: Infusion Density Substitution Pump

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 157 SW 9.xx 4.8.3.17 Infusion Density Substitution Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 327: Stroke Value Hdf Online Substitution Pump Osp

    2.15 HDF Online Substitution Pump OSP. The following menu is displayed. The stroke value must be set to 8300 (for pump head 8 x 12). Stroke Value HDF Online Substitution Pump OSP B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 328: Level Regulation

    The settings are (see default table in chapter 3): • 60 % for the slow mode • 100 % for the fast mode The default settings should not be changed, unless instructed otherwise. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 329: Calibration Inlet Flow Pump Nonreturn Valve Rvfpe

    You can calibrate the inlet flow pump nonreturn valve RVFPE. Fig.: Calibration RVFPE/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVFPE Fig.: DF Sub-Rack with RVFPE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 330
    Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPE. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 331
    VDABK1 MSBK1 VDEBK2 VABK2 VEBK2 VDABK2 FBK2 MSBK2 RVDA Fig.: Calibration RVDA/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVDA Fig.: Water Sub-Rack with RVDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 332
    • 300 mmHg ±37 mmHg • 400 mbar ±50 mbar HDF Online • 375 mmHg ±37 mmHg • 500 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 333: Calibration Outlet Flow Pump Nonreturn Valve Rvfpa

    Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPA. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. 10. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 334: Calibration Dialyser Inlet Throttle Valve Dde

    (30 to 40 °C). Thus uncontrolled pressure conditions are prevented. Select Test 1.20 Water Part (Overview) in TSM. Increase slowly the speed of FPA, until a flow of 500 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 335
    Set a pressure of 300 mmHg/400 mbar with the adjustment screw of DDE. • 300 mmHg ±37 mmHg • 400 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 336: Calibration Pressure Reducer Valve Dmv

    You can calibrate the pressure reducer valve DMV. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* TSHE VBKO* VVB* VBKS RVBO* Fig.: Calibration DMV VVBE Locking Screw for Manometer Fig.: Water Sub-Rack with DMV B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 337
    0.9 ±0.1 bar with the adjustment knob. Close water tap. Press down the adjustment knob to lock pressure reducer valve DMV. 10. Switch off Dialog. 11. Disassemble manometer. 12. Reassemble locking screw. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 338
    Note The 1.4 mm template fits through the gap. Check wit the 1.5 mm template. Note The 1.5 mm template must not fit through the gap. Reassemble cap C. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 339: Arterial Tubing Clamp Saka/Venous Tubing Clamp

    SAKV-SG. • O-ring 7 x 1.5 art. no. 34570624 Check the o-rings C on the V-stopper/stopper D Replace if o-rings show any damages (tears/ruptures/ brittleness). Close front door. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 340: Setting Servomotor For Disinfection Valve Vd

    Do not pull off the bow G from the control disc H, because its position is set ex works. The bow G on the control disc H must be seated in the groove F of the eccentric axle. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 341
    D Light barrier E Bevel on eccentric axle F groove G Bow (only partly visible, is seated in the eccentric groove) H Control disc Rinsing bridge Draw-Up of Disinfectant B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 342: Production Report

    • The activation of the HCT Crit-Line option enables a window in therapy. This window allows the data exchange and display of measurement values to a connected HemaMetrics™ Crit-Line device. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 343
    • Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 344: Low Level Options

    If the option is assembled or the option is retrofitted, it must be selected and WAN = Wide Area Network saved to activate the option. BSL = Bedside Link DBI = Dialog Bedside Link Interface B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 345: Service Reports

    The Operation Mode Report can be selected in the Service Reports menu. Operation Mode Report The switching from therapy mode to TSM mode and back to the therapy mode is documented with date and time. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 346
    5. Technical Safety Inspection and Preventive Maintenance 1 / 2 0 1 0 5 — 1 SW 9.xx This document is the property of B. Braun Avitum AG with all rights reserved. Copyright Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, Service firmware update, retrofitting and commissioning of the Dialog+.
  • Page 347
    Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 Inspection Protocol for Automatic Blood Pressure Measurement ABPM together with the Dialog+ 5.2.1 Execution Procedure for Measurement Inspection ABPM B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 348
    ……………………………. Operating Hours: ……..h Inventory No.: ………….. SW Version: ……….Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List (S = Technical Safety Inspection Points; M = Preventive Maintenance Points) NO YES OK Note: Text in { } brackets is information for the execution of the check list! Technical Safety Inspection …
  • Page 349
    Single-Needle Pressure PBS: (permissible. tolerance ±10 [mmHg]) — Comparison measurement (at approx. 0): Measurement instrument: ……Display on machine: …… [mmHg] — Comparison measurement (at approx. +400): Measurement instrument: ……Display on machine: …… [mmHg] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 350
    {The machine must be in therapy mode. Disconnect mains plug: The buzzer from the SMPS-MC gives three signalling tones as an indication that the machine is running in battery mode.} M 5.10 Disinfection: — Perform by responsible organisation {user} {inform responsible organisation} — Performed by service technician B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 351
    As a supplement this kit includes all parts recommended for exchange during a {34560686 Maintenance Kit/Supplement HDF Online} preventive maintenance of a HDF online machine.} B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 352
    The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 353
    Dialysis Machine Dialog REF {Type/Typ}: SN {Serien-No./Nr.} ………………ABPM Module M2009/2010 Automatic Blood Pressure Measurement ABPM B. Braun Avitum AG, 34209 Melsungen, Germany Manufacturer Address: Responsible Organisation (User) …………………………………………………………………………………………Visual Inspection — Tight seat of all connectors : ———————————————— OK — No damages which could affect the function or the safety (incl.
  • Page 354
    Fig.: Inspection Set-up for Compliance of Error Limits of the 6. Decrease pressure according to test points and enter values in Measurement Display table. 7. Enter deviations in table. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 355
    : —- max. 10 s Pressure reduction from >260 mmHg to <15 mmHg Pressure Tube 50 ml 50 ml Syringe Stopcock System Fig.: Inspection Set-up for Rapid Exhaust B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 356
    6.3.4 Tubing Diagram Water Sub-Rack 6-16 6.3.5 Tubing Diagram BIC Cartridge Holder 6-17 6.3.6 Tubing Diagram Housing HDF Online 6-18 6.3.7 Tubing Diagram UF Sub-Rack HDF Online 6-19 B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 357: Flow Diagrams

    Filter balance chamber (outlet) Filter from dialysate Filter bicarbonate Filter concentrate KtV-UV KtV-UV sensor (option Adimea) Water block with degassing chamber, upline tank, heat exchanger, level sensor, double-stage heater B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 358
    TSD-S Dialysate temperature sensor supervisor Degassing temperature sensor Heater temperature sensor TSHE Heater inlet temperature sensor TSDE Dialyser inlet temperature sensor Ultrafiltration pump RVUFP Ultrafiltration pump nonreturn valve B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 359: Flow Diagram Dialog

    Central Bicarbonate Central Concentrate Supply Supply (Unit Rear) Disinfectant BIC Concentrate Acid Concentrate Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 360: Flow Diagram Dialog+ Bic/Df Option

    Supply Supply (Unit Rear) BKUS Disinfectant BIC Concentrate Acid Concentrate RVBU Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial Options B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 361: Flow Diagram Dialog+ Hdf-Online

    Supply Supply (Unit Rear) BKUS Disinfectant RVBU BIC Concentrate Acid Concentrate Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial Options B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 362: Wiring Diagrams Dialog

    Cable connection from basic board to front door: — Double pump — HDF online Cable connection from basic board to front door: — Double pump — HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 363: Wiring Diagram Front Door Single Pump

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.2 Wiring Diagram Front Door Single Pump Motor Heparin pump DZ/DR, Cover SAKV B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 364: Wiring Diagram Front Door Double Pump

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.3 Wiring Diagram Front Door Double Pump Motor Heparin pump PBE, PBS DZ/DR, Cover SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 365: Wiring Diagram Front Door Hdf Online

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 6-10 SW 9.xx 6.2.4 Wiring Diagram Front Door HDF Online Heparin pump Motor PBE, PBS SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 366
    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s SW9.xx 6.2.5 Wiring Diagram Housing B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 367: Colour Coding Solenoid Valves

    TSHE NSVB Water Block Option BIC VVBE VBKS VBKO VBICP bk black bl blue br brown gn green gr grey rd red vi violet wh white ye yellow B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 368: Tubing Diagrams

    Silicone Tubing Silicone Tubing NSVB 2.9×3 SNP1.5 SNP4 Cable SNP1.5 Silicone Tubing 2.9×3 VVBE UF Sub-Rack RVDA blue Silicone Tubing 6×3 Rinsing Chambers Concentrate Suction Rods Water Inlet B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 369: Tubing Diagram Uf Sub-Rack

    Disposable Filter SNP6 Silicone SNP6 (2x) FBK2 Tubing Silicone Tubing 6×3 SNP6 (3x) Y-Tubing Connector 8 mm VABK2 VDEBK2 FBK1 VEBK2 SNP6 SNP6 Silicone SNP6 Disposable Tubing 6×3 Filter B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 370: Tubing Diagram Df Sub-Rack

    Silicone Silicone from Tubing Tubing WA Block 2.9×3 to BIC Suction Rod T Reduction on UF Sub-Rack to Concentrate to LA (UF) Suction Rod to TSHE/VEB (WA Block) B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 371: Tubing Diagram Water Sub-Rack

    Silicone Tubing NSVB Water Water Water Inlet Inlet Inlet from Piston Pump Rinsing WA-Block from VLA Silicone to DF Tubing Processing Silicone Tubing SNP4 TSHE RVDA Silicone Tubing B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 372: Tubing Diagram Bic Cartridge Holder

    SNP6 SNP4 SNP6 Reduction Piece BKUS Silicone Tubing Silicone Tubing 5×3 2.9×3 SNP6 Silicone Tubing 5×3 VBKO VBKS SNP1.5 VBICP Water Block Silicone Tubing 5×3 RVDA Water Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 373: Tubing Diagram Housing Hdf Online

    Clamp SNP6 13.3 Silicone Silicone Tubing SNP1.5 (2x) Tubing SNP6 6×1.75 Silicone VBKS Tubing Silicone 2.9×3 Tubing Water Water Sub-Rack Inlet BKUS blue Rinsing Chambers Concentrate Suction Rods B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 374: Tubing Diagram Uf Sub-Rack Hdf Online

    Silicone Tubing Silicone Silicone VDABK2 SNP6 Tubing VSAA Tubing Silicone VABK2 Tubing FBK1 VDEBK2 Silicone Tubing 6×1.75 SNP6 VEBK2 (3x) SNP6 Silicone Y-Tubing Tubing 6×1.75 Connector 8 mm B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 375
    (Material Return Form) to B. Braun Avitum AG via the representative of the respective country for the ability to supply refurbished spare parts. The condition and the version of the defective spare parts must allow a qualified refurbishment.
  • Page 376
    …………………………….Disposal and Taking Back of Spare Parts Dispose spare parts (e.g. boards or batteries) according to local disposal guidelines or send back to B. Braun Avitum AG free of charge. Cleaning and Disinfection of Spare Parts Used in Dialog Machines All spare parts sent by the customer to the B.
  • Page 377
    Switch Mode Power Supply Microcontroller SMPS-MC 7-29 7.7.3.1 Compatibility Table SMPS-MC 7-30 Front Door 7-31 7.8.1 Manometer Connection 7-33 7.8.2 Heparin Pump 7-33 7.8.3 Blood Pump 7-34 7.8.4 Safety Air Detector SAD 7-35 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 378
    7-61 7.18.4 Board HDF Online 7-62 7.18.5 DF/HDF Filter Holder HDF Online 7-63 7.19 Automatic Blood Pressure Measurement ABPM Module M2009 7-64 7.19.1 Compatibility Table ABPM Module M2009 7-65 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 379
    SW 9.xx Page 7.20 Emergency Power Supply (Option) 7-66 7.20.1 Battery Sub-Rack 7-67  7-68 7.21 Nexadia -/WAN-BSL (Option) 7.22 Card Reader (Option) 7-69 7.23 Adimea Option (Option) 7-70 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 380: Sub-Racks

    7.6 for single parts UF sub-rack see paragraph 7.5 for single parts DF sub-rack see paragraph 7.4 for single parts Water sub-rack see paragraph 7.3 for single parts B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 381: Solenoid Valves

    C) without loosening the self-locking nut A. The procedure is only permissible if the coil can be turned without much iv 1 x Screw 40×10 effort. v 1 x Washer 4.2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 382
    TSHE 8 34570730 1 Housing complete for temperature sensor 9 3451727B 1 Throttle 0.4 bar 10 3451664B für HDF-Online 1 Throttle 1.3 bar 11 34517782 5 Disposable filter RVDA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 383: Water Sub-Rack

    12 34516069 1 Throttle 0 — 4 bar 1 Tubing connector ∅ 7 x 1/8″ 13 34517243 stainless steel 14 34560289 connector/connector (DPP) 1 Solenoid valve 2/2 way VVBE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 384: Gear Pump For Ep

    1 Motor for gear pump (SW 1.7.4) Micropump Version 2 for EP only 1b 34560319 1 Gear pump (Micropump version 2) (see A in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 385: Df Sub-Rack

    7.4.1 — DF block see separate figure for single parts, paragraph 7.4.1 see separate figure for — Piston pumps single parts, paragraph 7.4.2 BICLF BICP ENDLF TSDS RVFPE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 386: Df Block

    1 Druckregler 1,3 bar 8 34570683 10 O-Ring 4,47 x 1,78 2 Tubing connector ∅ 3.5 x 1/8″ 9 3477103A plastic 5 Tubing connector ∅ 8 x 1/8″ 10 3451916A plastic B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 387
    4e 34771085 5 Seal BICP 5a 34570373 10 Plug (long) 5b 3451614B without connector 1 Temperature sensor ENDLF 5c 3477104A 10 O-ring 8.3 x 2.4 BICLF TSBIC 4b,10 3 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 388: Gear Pump For Fpe/Fpa

    1 Motor for gear pump (SW 1.7.4) 1b 34515330 1 Gear pump (Micropump version 1) Micropump Version 2 for FPA/FPE only (see B in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 389: Piston Pumps

    1 Motor for piston pump 5 34560572 1 Bell joint for piston pump 6 34560602 1 Pump body for piston pump — 7703848 required for motor assembly 1 Assembly jig (motor piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 390: Uf Sub-Rack

    Pos. Art. No. Qty. Description Comments UF Sub-Rack — Air separator complete see separate figure — Balance chamber 2 see separate figure RVFPA VDABK2 VABK1 VDEBK2 VDEBK1 BK1/2 VABK2 VDABK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 391: Balance Chamber

    2 Membrane for balance chamber 20 Tubing clamp SNP-1.5 ∅ 7-8 mm VEBK2 plastic — 34570195 20 Tubing clamp SNP-4 ∅ 10-11 mm — 34570209 plastic VDEBK2 VABK1 VDEBK1 VABK2 VDABK1 VEBK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 392: Air Separator Assembly Group

    1 Throttle 0.4 bar 5/6 34570683 10 O-ring 4.47 x 1.78 6 3451664B 1 Throttle 1.3 bar 5 Tubing connector ∅ 8 x 1/8″ 7 3451916A plastic RVFPA RVFPA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 393: Top Level Sub-Rack

    10 34561277 1 Temperature switch assemble temperature switch on TL sub-rack Basic Board with heat sink paste Analog Board Online Board Motherboard Power Board Valves Online Board Basic Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 394: Lx800 Motherboard

    1 Compact flash card 1 GB The SW for the machine has to be installed again after replacing the CFC. 2 34560610 1 LX800 Motherboard 3 34519327 1 Battery BR2032, 3V Motherboard B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 395: Settings For Tft Display/Cfc

    1 / 2 0 1 0 7 — 21 SW 9.xx Pos. Art. No. Qty. Description Comments 7.6.2 Settings for TFT Display/CFC LX800 Motherboard JLVDSPWR TFT Power 3.3 V: BoeHydis/Samsung JCFSET1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 396: Rear Door

    1 Fan 92 x 92 mm with connector 2 34570284 5 Dust filter 3 3451059A 1 Mains switch 3 34770003 5 Protection cap for mains switch — Mains cord see separate figure (paragraph 7.7.1) SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 397
    4 34518029 1 Strain relief 5 3451441A 1 Tubing holder/fan cover compatible to crank 5a/5b 5a 34518355 1 Crank for blood pump 5b 34560157 1 Crank for blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 398: Mains Cord

    1 Mains cord (Brazil) mains cord with mains plug 2 34561145 mains cord with mains plug 1 Mains cord (Cekon) 3 34561226 1 Mains cord (China) mains cord with mains plug B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 399
    1 Mains cord (Israel) 5 34561170 1 Mains cord (Philipines) mains cord with mains plug 6 34561153 mains cord with mains plug 1 Mains cord (Schuko) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 400
    1 Mains cord (South Africa/India) 8 34561234 mains cord with mains plug 1 Mains cord (Thailand) 9 34561188 1 Mains cord (UK) mains cord with mains plug B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 401: Connection Mains Cord To Smps-Mc

    Central Earth F5B F5A Cable Gland with Lock Nut Rear Door Cable Ties N PE PE LH1LH2 Ferrite Ferrite Brown Green/Yellow Blue Mains Cord with Mains Plug Mains Cord Ferrite B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 402: Compatibility Table For Boards

    The digital board is always delivered with the latest LLS/LLC software versions! 34560734 Analog Board AB (38108641) 34560866 HDF Online Board HOB (38108647) 3451665A MSBK Board (3810671X) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 403
    10 Fuse T3.15 A (TR5) LH1 LH2 N PE PE K3/K4 34514775 2 Relay (12 V) for 24 V C114/C115 34519912 2 Capacitor 1F H104 34560580 1 Buzzer 34560416 10 Plug for buzzer B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 404: Switch Mode Power Supply Microcontroller Smps-Mc

    LH1 LH2 N PE PE 7.7.3.1 Compatibility Table SMPS-MC SMPS-MC Type Plate Version Number SMPS-MC SW Version Dialog+ SW 9.xx (TN/Item-No) (Ver/SW) XXXXXX.5 Rev. 1.15 34560220 (version 5) ex works B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 405
    1 Metal disc ∅ 10 mm for cover 9 34560033 9 34519823 Hinge pin with interior thread hinge pin 9 34516670 1 Seal for blood pump SAKV-SG SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 406: Front Door

    17 34560831 see separate figure (paragraph 7.8.5) 1 Valve block level regulation (double pump) 18 34560823 1 Diaphragm pump with assembly elbow 3 mm 4 mm 4 mm SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 407: Manometer Connection

    5 34560785 1 Housing with seal 5a 34514732 1 Syringe holder 5b 34570640 5 Dummy plug 5c 34570667 1 Clamp 5d 34570659 1 Unlocking lever set 5a 5b 5c 5d B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 408: Blood Pump

    1 Reed sensor with contacts and o-ring 3.68 x 1.78 — 34516670 1 Seal for blood pump ^^^^^^^^^^^^^^^^^^^^^^^^ 100 g tube (to lubricate the shaft of the blood pump) — 7703708 1 Grease SKD 4002 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 409: Safety Air Detector Sad

    SAD with board; do not tamper, exchange only complete red detector If cover for SAD is present: a ball head allen key 2.5/3 is required for bottom screw behind rotary magnet. SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 410: Level Regulation Module

    27 34565205 5m PUR tubing 3.6 x 1.1 (5 m) for manometer connection (filter) 26 27 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 411
    1 Swivel caster 8 3451490A 1 Swivel caster with brake 9 34560548 1 Card-Reader-Board see 7.22 for details 10 3451487C 1 ABPM module M2009 with filter see 7.19 for details B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 412: Basic Housing

    3m Silicone tubing 8 x 3.2 / red (3 m), external high temperature tubing with two single ear clamps 17.5 — 39224538 1 Single ear clamp 17.5 for external high temperature tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 413: Concentrate Suction Rods

    1 Male connector (red) with o-rings 8 34570004 5 O-ring 9.3 x 2.4 9 34565280 3m Silicone tubing 3.2 x 1.6 (3 m) 10 3451728A 1 PE tube 4 x 1 with silicone tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 414: Rinsing Chambers For Concentrate Suction Rods

    3.68 x 1.78 5 3451476A 1 Male connector (red) with o-rings 6 3451475A 1 Male connector (blue) with o-rings 7 34517782 5 Disposable filter 3 4 5 6 7 grey B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 415
    1 PE tube 4 x 1 7 34565280 3m Silicone tubing 3.2 x 1.6 (3 m) 8 34570004 5 O-ring 9.3 x 2.4 9 3451465A 2 Suction nipple with o-rings 9.3 x 2.4 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 416: Central Concentrate Supply (Option)

    Comments Central Concentrate Supply (Option) 3 34565248 10m Silicone tubing 5 x 1.5 / blue (10 m) 5 34565230 10m Silicone tubing 5 x 1.5 / red (10 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 417: Rinsing Bridge 4

    13 34511792 1 Tubing coupling 14 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 15,14,13 12,11,8 15 34771255 20 O-ring 14 x 1.5 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 418
    10 O-ring 5.0 x 1.6 for eccecntric shaft 6 34560858 1 Eccecntric roller 7 34560840 1 Eccecntric shaft 8 3451659D 1 Servomotor with board for VD, servomotor and board must be exchanged together B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 419
    17 34570730 1 Housing complete for temperature sensor 18 34773762 10 O-ring 9.25 x 1.78 19 34511792 1 Tubing coupling 20 34771255 20 O-ring 14 x 1.5 20,19 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 420: Tft Monitor

    2 34570691 2 OSD cover complete with seal and screws 3 34561285 1 Membrane 4 34561366 1 Membrane for Adimea option Dialog start stop stop Dialog start stop stop B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 421
    1 TFT display 15“ (Boe Hydis or Samsung) 9 3451460A 1 Touch controller board TCB version 3 — 34560912 1 Ribbon cable 12 pin FPB — OSD — 34560971 1 Ribbon cable FPB — BIB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 422
    * Balance (e.g. Pocket Balance) For example manufacturer: for details contact manufacturer Kern & Sohn GmbH directly Internet: www:kern-sohn.com * not in the product range of B. Braun Avitum AG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 423: Test And Calibration Equipment And Tools

    1 Male to male connector couplings for the measurement of the patient leakage current — 770284A 1 DCI cable, external required for motor assembly — 7703848 1 Assembly jig (motor for piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 424: Service Kits

    Comments 7.12 Service Kits — 3451897A 1 Repair kit balance chamber for the complete replacement of both membranes, incl. all o-rings and assembly parts (self-locking nuts, screws, springs etc.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 425: Software

    1 CD Remote Bedside Link List ReBeLL, for network version 1.5.13 — 7703562 for WAN-BSL only 1 CD WAN-BSL, firmware version 1.0.4.8 — 7703864 1 Field service utility FSU, version 2.00 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 426
    Software 7.58 (Linux) Software 7.59 (Linux) *1.9 Software-Update 7.59 (Linux) #1.01/ USB Stick with TLC/LLS SW only 1.04 if SW 7.52 is present Software ≥ 8.20 (Linux) Software 9.02 (Linux) 1.04 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 427: Miscellaneous

    1 PVC tubing 10 x 3 / black, (3 m), with a single ear clamp 19.5 and a tubing clamp 12-20 mm — 39234126 1 Tubing clamp 12-20 mm — 34565345 1m Silicone tubing 12 x 2.5 (1 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 428
    — 7703708 1 Grease SKD 4002 100 g tube (to lubricate the shaft of the blood pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 429: Staff Call (Option)

    1 / 2 0 1 0 7 — 55 SW 9.xx Pos. Art. No. Qty. Description Comments 7.16 Staff Call (Option) 1 34517685 1 Staff call board STAT (Default) DYN AUS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 430: Bicarbonate Cartridge Holder (Option)

    10 34514449 1 Cover for bottom housing 11 34517138 1 Reed sensor with contacts and o-ring bottom BIC cartridge holder 3.68 x 1.78 — 34517782 5 Disposable filter filter FBIC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 431: Solenoid Valves And Throttle

    5 O-ring 11.1 x 1.6 4 34771050 10 Pressure spring 500 mmHg 5 34771352 3 Centre disc 5 Tubing connector ∅ 8 x 1/8″ 6 3451916A plastic VBKS VBKO VBICP B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 432: Hdf Online (Option)

    7 — 58 SW 9.xx Pos. Art. No. Qty. Description Comments 7.18 HDF Online (Option) — Substitution port see separate figure — UF sub-rack see separate figure PBS/SN SAKV-SG SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 433: Substitution Port

    1 Substitution connector 5 3477106A 5 O-ring 6.4 x 1.3 6 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 A or A+B (A=Quadring; B=Spacer) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 434: Uf Sub-Rack Hdf Online

    AKM only, connector/connector RVFPA VSAA VSAE 4 3451902C manuf. AKM only, connector/flange 1 Solenoid valve 2/2 way 4 34770852 5 O-ring 11.1 x 1.6 VDABK2 VABK1 VDEBK2 VDEBK1 BK1/2 VDABK1 VABK2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 435: Valve Block Hdf Online

    7 — 61 SW 9.xx Pos. Art. No. Qty. Description Comments 7.18.3 Valve Block HDF Online 1 34560661 1 Hydrophobic filter with tubing adapter 2 3451900A 1 Solenoid valve 3/2 way B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 436: Board Hdf Online

    Pos. Art. No. Qty. Description Comments 7.18.4 Board HDF Online 1 34560866 1 HDF online board Motherboard Online Board Power Board Valves Digital Board Power Board Motors Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 437: Df/Hdf Filter Holder Hdf Online

    Blue 3 3451445A 1 Dialyser coupling (red) 3 3451446A 1 Dialyser coupling (blue) 3 34990445 5 O-ring 12.37 x 2.62 3 34773550 20 Membrane for dialyser coupling Blue Blue B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 438: Automatic Blood Pressure Measurement Abpm Module M2009

    1 34560955 1 Filter (ABPM module M2009) 2 3451487C 1 ABPM module M2009 firmware version of ABPM module on sticker: Version Y (FW):X.X with filter Version YX.X Version YX.X B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 439: Compatibility Table Abpm Module M2009

    Description Comments 7.19.1 Compatibility Table ABPM Module M2009 Position of DIL Switch for DIL Switch: Default LX800 (ex Works) LX800 Motherboard DIL Switch DIL Switch Version YX.X LX800 (default) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 440
    Sulfric acid in the eyes may cause loss of eyesight and acid on the skin will cause burns. Pay attention to the included documentation of the battery supplier! see separate figure — Battery sub-rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 441
    F301 H302 F302 H304 F304 H601 F601 brown blue brown H602 F602 H402 F402 F403 H403 blue green/yellow green/yellow brown blue GNDAKKU XP3.2 brown brown XP3.1 +24VAKKU P9 P7 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 442
    1 / 2 0 1 0 7 — 68 SW 9.xx Pos. Art. No. Qty. Description Comments  7.21 Nexadia -BSL/WAN-BSL (Option) 1 34514554 1 Bedside link board (BSL board) without firmware B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 443
    7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 — 69 SW 9.xx Art. No. Qty. Description 7.22 Card Reader (Option) 1 34560548 1 Card reader board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 444
    7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 — 70 SW 9.xx Art. No. Qty. Description 7.23 Adimea Option 1 3456063A 1 Kt/V-UV board with housing and tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 445
    8 — 1 SW 9.xx ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information TI Assembly Instructions AI Field Service Information FSI Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 446
    In general the mats are made of sturdy two layer material and have reinforced Mat Material edges and corners. The work surface is static dissipative (> 100 MΩ), the other side is conductive (> 100 kΩ). B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 447
    Unprotected electronic parts (e.g. wrapped in normal plastic film or parts padded with foam material) are assumed to be damaged by electrostatic discharge and will be scrapped. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 448
    N, connected to a one or three phase public power supply with a constant current load of the power supply ≥ 100 A per phase, then it is also valid for the Dialog machine with the switch mode power supply microcontroller SMPS-MC. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 449
    The power frequency magnetic field should be measured in the intended installation location to magnetic field acc. to IEC 61000- assure that it is sufficiently low. NOTE U is the a.c. mains voltage prior to application of the test level. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 450
    Dialog machine. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 451
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 7 SW 9.xx Technical Information TI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 452
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 8 SW 9.xx Assembly Instructions AI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 453
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 9 SW 9.xx Field Service Information FSI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 454
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 10 SW 9.xx Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 455
    2011-03-01 4-1-1_1/2010 Revision Chapter 4 TSM Service Program 2011-03-01 4-2-1_1/2010 4-3-1_1/2010 5-1_1/2010 Revision Chapter 5 Technical Safety Inpsection and Preventive Maintenance 2011-03-01 9-1_1/2010 Revision Chapter 9 Edition/Updates 2011-03-01 B. Braun Avitum AG BA-TE-DE08W M.KAY — Dialog+ SW9xx_SM_Chapter 9-1_1-2010.doc/pdf <110301> yymmdd…

Modification: 12 August 2023

Техническая документация (Technical Documentation/Manual) на Гемодиализ 2008K;

Техническая документация, Technical Documentation/Manual на Гемодиализ 2008K

Инструкция по наладке (Adjustment Instruction) на Гемодиализ 2008K Calibration procedures;

Инструкция по наладке, Adjustment Instruction на Гемодиализ 2008K Calibration procedures

Техническая документация (Technical Documentation/Manual) на Гемодиализ 2008K2;

Техническая документация, Technical Documentation/Manual на Гемодиализ 2008K2

Схема электрическая (Electric scheme (circuit)) на Гемодиализ 4008 B/E/S/H;

Схема электрическая, Electric scheme (circuit) на Гемодиализ 4008 B/E/S/H

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 4008 BVM;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 4008 BVM

Техническая документация (Technical Documentation/Manual) на Гемодиализ 4008 E/B/H/S;

Техническая документация, Technical Documentation/Manual на Гемодиализ 4008 E/B/H/S

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 4008B v3.2;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 4008B v3.2

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 4008B v3.2 1996;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 4008B v3.2 1996

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 4008H/S v4.0 OCM Дополнительные инструкции;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 4008H/S v4.0 OCM Дополнительные инструкции

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 4008S;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 4008S

Сервисная инструкция (Service manual) на Гемодиализ 5008;

Сервисная инструкция, Service manual на Гемодиализ 5008

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 5008;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 5008

Техническая документация (Technical Documentation/Manual) на Гемодиализ 5008 (Диагностика гидравлики);

Техническая документация, Technical Documentation/Manual на Гемодиализ 5008 (Диагностика гидравлики)

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ 5008S;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ 5008S

Сервисная инструкция (Service manual) на Гемодиализ AK 10 Blood monitor BMM 10-1;

Сервисная инструкция, Service manual на Гемодиализ AK 10 Blood monitor BMM 10-1

Сервисная инструкция (Service manual) на Гемодиализ AK 200 (1998);

Техническая документация (Technical Documentation/Manual) на Гемодиализ AK 200 S — коды ошибок;

Техническая документация, Technical Documentation/Manual на Гемодиализ AK 200 S - коды ошибок

Инструкция по монтажу и обслуживанию (Installation and Maintenance Guide) на Гемодиализ AK 200 Ultra;

Инструкция по монтажу и обслуживанию, Installation and Maintenance Guide на Гемодиализ AK 200 Ultra

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ AK 200 Ultra S;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ AK 200 Ultra S

Каталог (элементов, запчастей и пр.) (Catalogue, Spare Parts list) на Гемодиализ AK 200S & AK 200 ULTRA S;

Каталог (элементов, запчастей и пр.), Catalogue, Spare Parts list на Гемодиализ AK 200S & AK 200 ULTRA S

Сервисная инструкция (Service manual) на Гемодиализ AK 200S & AK 200 ULTRA S;

Сервисная инструкция, Service manual на Гемодиализ AK 200S & AK 200 ULTRA S

Инструкция пользователя (User manual) на Гемодиализ AK-96;

Инструкция пользователя, User manual на Гемодиализ AK-96

Инструкция пользователя (User manual) на Гемодиализ AK-98;

Инструкция пользователя, User manual на Гемодиализ AK-98

Руководство пользователя (Users guide) на Гемодиализ Artis;

Руководство пользователя, Users guide на Гемодиализ Artis

Сервисная инструкция (Service manual) на Гемодиализ Artis;

Сервисная инструкция, Service manual на Гемодиализ Artis

Инструкция оператора (Operator manual) на Гемодиализ Barkey Plasmatherm (Ver. 1);

Инструкция оператора, Operator manual на Гемодиализ Barkey Plasmatherm (Ver. 1)

Сервисная инструкция (Service manual) на Гемодиализ Barkey Plasmatherm (Ver. 1.1 GB);

Сервисная инструкция, Service manual на Гемодиализ Barkey Plasmatherm (Ver. 1.1 GB)

Инструкция пользователя (User manual) на Гемодиализ Barkey Plasmatherm (Ver. 1.1 GB);

Инструкция пользователя, User manual на Гемодиализ Barkey Plasmatherm (Ver. 1.1 GB)

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ DBB-05;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ DBB-05

Каталог (элементов, запчастей и пр.) (Catalogue, Spare Parts list) на Гемодиализ DBB-05;

Каталог (элементов, запчастей и пр.), Catalogue, Spare Parts list на Гемодиализ DBB-05

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ DBB-05;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ DBB-05

Техническая документация (Technical Documentation/Manual) на Гемодиализ DBB-05;

Техническая документация, Technical Documentation/Manual на Гемодиализ DBB-05

Техническая документация (Technical Documentation/Manual) на Гемодиализ DBB-05 TTraining;

Техническая документация, Technical Documentation/Manual на Гемодиализ DBB-05 TTraining

Техническая документация (Technical Documentation/Manual) на Гемодиализ DBB-06 TTraining;

Техническая документация, Technical Documentation/Manual на Гемодиализ DBB-06 TTraining

Техническая документация (Technical Documentation/Manual) на Гемодиализ DBB-07;

Техническая документация, Technical Documentation/Manual на Гемодиализ DBB-07

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ DBB-07;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ DBB-07

Сервисная инструкция (Service manual) на Гемодиализ DBB-07;

Сервисная инструкция, Service manual на Гемодиализ DBB-07

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Dialog & Dialog ONLINE SW 5.00 (2001);

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Dialog & Dialog ONLINE SW 5.00 (2001)

Сервисная инструкция (Service manual) на Гемодиализ Dialog+ (2003, 6 частей, 1 и 3 на En);

Сервисная инструкция, Service manual на Гемодиализ Dialog+ (2003, 6 частей, 1 и 3 на En)

Сервисная инструкция (Service manual) на Гемодиализ Dialog+ (2003, 7 частей, 1 и 7 на En);

Сервисная инструкция, Service manual на Гемодиализ Dialog+ (2003, 7 частей, 1 и 7 на En)

Сервисная инструкция (Service manual) на Гемодиализ Dialog+ (2010);

Сервисная инструкция, Service manual на Гемодиализ Dialog+ (2010)

Сервисная инструкция (Service manual) на Гемодиализ Dialog+ (часть 3);

Сервисная инструкция, Service manual на Гемодиализ Dialog+ (часть 3)

Инструкция пользователя (User manual) на Гемодиализ Dialog+ SW 7.2x;

Инструкция пользователя, User manual на Гемодиализ Dialog+ SW 7.2x

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Dialog+ SW 7.5x;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Dialog+ SW 7.5x

Инструкция пользователя (User manual) на Гемодиализ Dialog+ SW 9.1x;

Инструкция пользователя, User manual на Гемодиализ Dialog+ SW 9.1x

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Dialog+ SW 9.1x;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Dialog+ SW 9.1x

Каталог (элементов, запчастей и пр.) (Catalogue, Spare Parts list) на Гемодиализ Dialog/Dialog Advanced (Spare Parts List);

Каталог (элементов, запчастей и пр.), Catalogue, Spare Parts list на Гемодиализ Dialog/Dialog Advanced (Spare Parts List)

Инструкция по техническому обслуживанию (Maintenance Instruction) на Гемодиализ Diamax (Nipro);

Инструкция по техническому обслуживанию, Maintenance Instruction на Гемодиализ Diamax (Nipro)

Руководство пользователя (Users guide) на Гемодиализ Formula (F plus, F 2000, F 2000 plus) SW 5.7 (Bellco);

Руководство пользователя, Users guide на Гемодиализ Formula (F plus, F 2000, F 2000 plus) SW 5.7 (Bellco)

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Formula (F plus, F 2000, F 2000 plus) SW 5.8-6 (Bellco);

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Formula (F plus, F 2000, F 2000 plus) SW 5.8-6 (Bellco)

Техническая документация (Technical Documentation/Manual) на Гемодиализ Formula (F plus, F 2000, F 2000 plus, F Therap) (Bellco);

Техническая документация, Technical Documentation/Manual на Гемодиализ Formula (F plus, F 2000, F 2000 plus, F Therap) (Bellco)

Инструкция по техническому обслуживанию (Maintenance Instruction) на Гемодиализ Innova;

Инструкция по техническому обслуживанию, Maintenance Instruction на Гемодиализ Innova

Техническая документация (Technical Documentation/Manual) на Гемодиализ multiFiltrate;

Техническая документация, Technical Documentation/Manual на Гемодиализ multiFiltrate

Сервисная инструкция (Service manual) на Гемодиализ multiFiltrate;

Сервисная инструкция, Service manual на Гемодиализ multiFiltrate

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ multiFiltrate V1.9;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ multiFiltrate V1.9

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ multiFiltrate V5.2;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ multiFiltrate V5.2

Каталог (элементов, запчастей и пр.) (Catalogue, Spare Parts list) на Гемодиализ Phoenix;

Каталог (элементов, запчастей и пр.), Catalogue, Spare Parts list на Гемодиализ Phoenix

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Prometeus серии 4008Н «Прометей»;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Prometeus серии 4008Н "Прометей"

Сервисная инструкция (Service manual) на Гемодиализ Surdial (искусственная почка) (Nipro);

Сервисная инструкция, Service manual на Гемодиализ Surdial (искусственная почка) (Nipro)

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ System 1000 (искусственная почка);

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ System 1000 (искусственная почка)

Руководство пользователя (Users guide) на Гемодиализ Trima Accel [Terumo BCT];

Руководство пользователя, Users guide на Гемодиализ Trima Accel [Terumo BCT]

Сервисная инструкция (Service manual) на Гемодиализ Trima Accel [Terumo BCT];

Сервисная инструкция, Service manual на Гемодиализ Trima Accel [Terumo BCT]

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Диамикс-МФ;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Диамикс-МФ

Инструкция по монтажу (Installation instructions) на Гемодиализ Система Prisma;

Инструкция по монтажу, Installation instructions на Гемодиализ Система Prisma

Инструкция по техническому обслуживанию (Maintenance Instruction) на Гемодиализ Система Prisma (1996);

Инструкция по техническому обслуживанию, Maintenance Instruction на Гемодиализ Система Prisma (1996)

Сервисная инструкция (Service manual) на Гемодиализ Система Prisma ПО v03.05 2002;

Сервисная инструкция, Service manual на Гемодиализ Система Prisma ПО v03.05 2002

Руководство пользователя (Users guide) на Гемодиализ Система Prismaflex c ПО v3.2x;

Руководство пользователя, Users guide на Гемодиализ Система Prismaflex c ПО v3.2x

Руководство оператора (Operators Guide) на Гемодиализ Система Prismaflex Versions 1.07;

Руководство оператора, Operators Guide на Гемодиализ Система Prismaflex Versions 1.07

Сервисная инструкция (Service manual) на Гемодиализ Система Prismaflex ПО v1.05 2004;

Сервисная инструкция, Service manual на Гемодиализ Система Prismaflex ПО v1.05 2004

Инструкция по эксплуатации (Operation (Instruction) manual) на Гемодиализ Системы автопереливания C.A.T.S plus;

Инструкция по эксплуатации, Operation (Instruction) manual на Гемодиализ Системы автопереливания C.A.T.S plus


Архив технической документации медицинского оборудования © 2006 … 2023

  • Page 1
    Dialog SW 9.xx S e r v i c e M a n u a l E n g l i s h Edition 1 -2010 M.KAY Dialog+ SW9xx SM EN 1-2010…
  • Page 2
    1 / 2 0 1 0 SW 9.xx Contact your Local B. Braun Representative for Service Support 34209 Melsungen B. Braun Avitum AG Germany Tel. No.: +49 5661 713500 (Günter Nissen) E-Mail: guenter.nissen@bbraun.com B. Braun Avitum AG BA-TE-DE08C M.KAY -…
  • Page 3
    S e rvi ce Manu al s w i th a re gi s tr a ti on num be r a re inc lud ed in th e u p da te s erv ic e! B. Braun Avitum AG BA-TE-DE08C M.KAY -…
  • Page 4
    1 / 2 0 1 0 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
  • Page 5
    Flow, Wiring and Tubing Diagrams Spare Parts List Appendix ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information Assembly Instructions Field Service Information Instruction Leaflets Edition/Updates Service Manual B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd…
  • Page 6
    Calibration The symbol appears for necessary calibration measures. Calibration Equipment/Tools symbol gives remarks necessary calibration equipment/tools, i.e. during a calibration, disassembly or assembly. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd…
  • Page 7
    Perform regular technical safety inspections as described in chapter 5 of this Technical Safety Inspection service manual to ensure the safety of the machine. An instructions for use can be ordered at your local B. Braun representative or Instructions for Use dealer.
  • Page 8
    1 — 1 SW 9.xx Copyright This document is the property of B. Braun Avitum AG with all rights reserved. Commissioning and Service Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, firmware update, retrofitting and commissioning of the Dialog+.
  • Page 9
    (fasten with a single ear clamp 19.5 and a tubing clamp 12- 20 mm) • Tubing length: approx. 3 m • Drain height: max. 80 cm 8. Emergency Power Supply/Battery Option Fig.: Dialog+ Rear View B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 10
    1 / 2 0 1 0 1 — 3 SW 9.xx Table of Contents Page Commissioning Check List 1.1.1 Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 11
    …………………………….Operating Hours: ……..h Inventory No.: …………..SW Version: ……….. Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List Note: Text in { } brackets is information for the execution of the check list! Visual Inspection {Machine: clean/complete; no damages/moisture influences or loose assemblies; no moveable parts touching tubings or wires; casters are moveable;…
  • Page 12
    Next Inspection Date: …………………………..Name Service Technician: Name of Company: The commissioning was performed and the machine was hand over to the responsible ………………. organisation (user)…………………………Date/Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 13
    The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 14: Table Of Contents

    Switch Mode Power Supply Microcontroller SMPS-MC 2-19 2.15.1 Legend Switch Mode Power Supply Microcontroller 2-19 2.16 TFT Monitor 2-20 2.16.1 Legend TFT Monitor 2-20 2.17 Front Door 2-21 2.17.1 Legend Front Door 2-21 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 15
    System Integration SMPS-MC 2-43 2.22.3 Component Layout SMPS-MC 2-44 2.22.4 Wiring Diagram SMPS-MC with Battery Option 2-45 2.22.5 Description SMPS-MC 2-46 2.22.6 Fuses 2-48 2.22.7 Signals 2-49 2.22.8 Internal Signals 2-49 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 16: General Information

    All sensor data are sent separately, via two serial bus systems, to the supervisor and controller via the analog board to the digital board. The actuators, motors and valves are driven via the power board valves PBV and power board motors PBM. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 17
    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 3 SW 9.xx For Software ≥ ≥ ≥ ≥ 9.xx For Software 9.xx For Software For Software 9.xx 9.xx B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 18: Overview Sub-Racks

    1 / 2 0 1 0 2 — 5 SW 9.xx Overview Sub-Racks Fig. : Overview Sub-Racks Rear View Dialog+ 2.1.1 Legend Overview Sub-Racks Top Level Sub-Rack DF Sub-Rack UF Sub-Rack Water Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 19: Top Level Sub-Rack

    USB Port: Software Installation LLC/LLS; Service-Switch S1; Jumper JP1/JP2 7 7 7 7 HDF Online Board HOB 8 8 8 8 Power Board Motor PBM 9 9 9 9 Analog Board AB (LLA) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 20: Basic Board Bb

    SMPS-MS (NT) SMPS-MS (NT) SMPS-MS (NT) Power Supply Bedside Link BSL-PWR BSL-PWR BSL-PWR BSL-PWR Power Board Motors PBM P58/P60 P58/P60 Power Supply Hall Sensors P58/P60 P58/P60 Disinfection Valve VD B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 21: Power Board Valves Pbv

    Dialyser Outlet Valve VDA Bypass Valve VBP Option BIC Cartridge Valves VBICP, VBKS, VBKO, VVB Disinfection Valve VD Circulation Valve VZ Venous Tubing Clamp SAKV Arterial Tubing Clamp SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 22: Power Board Motors Pbm

    BIC Piston Pump BICP BICP BICP BICP Concentrate Piston Pump KP KP KP KP UF Piston Pump UFP Level Regulation Pump (Diaphragm Pump) LRP (PPR) LRP (PPR) LRP (PPR) LRP (PPR) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 23: Digital Board Db

    NSVB, BICSS, KSS, MSBK1/2, RDV, SAD, BPADS, BPVDS, Service Switch: BL (Controller/Supervisor Sensor Analog Board) Position 0: Therapy Mode Position 2: TSM Service Program Mode Position 3: Software Installation/Update Mode B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 24: Analog Board Ab

    Fig. : Analog Board AB (LLA) 2.7.1 Legend Analog Board P1/P2 Controller Sensors: PBS, TSHE, TSE, TSBIC, TSD, TSDE, BICLF, ENDLF, PE, LAFS Supervisor Sensors: TSD-S, ENDLF-S Controller/Supervisor Sensors: BL, PBE, PA, PV, PDA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 25: Hdf Online Board Hob

    SW 9.xx HDF Online Board HOB Fig. : HDF Online Board 2.8.1 Legend HDF Online Board P1: Sensors: PSABFS, PSAUS, PSPOSS, FEHDFS, FEDFFS P4: Valves VBE, VDFF, VSAA, VSAE, VSB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 26: Uf Sub-Rack

    Non-Return Valve Outlet Flow Pump RVFPA RVFPA RVFPA RVFPA Inlet Balance Chamber Valve VEBK2 VEBK2 VEBK2 VEBK2 Outlet Balance Chamber Valve VABK1 VABK1 Air Separator Valve VLA VABK1 VABK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 27: Uf Sub-Rack Hdf Online

    Non-Return Valve Outlet Flow Pump RVFPA RVFPA RVFPA RVFPA Substitute Connection Inlet Valve VSAE VSAE VSAE VSAE Outlet Balance Chamber Valve VABK1/2 VABK1/2 VABK1/2 VABK1/2 Filter Vent Valve VBE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 28: Df Sub-Rack

    Dialysate Temperature Sensor TSD Concentrate Pump KP KP KP KP Dialysate Supervisor Temperature Sensor TSD-S TSD-S TSD-S TSD-S Bicarbonate Conductivity Sensor BICLF BICLF UF Pump UFP BICLF BICLF Bicarbonate Non-Return Valve RVB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 29: Water Sub-Rack

    Top BIC Cartridge Valve VBKS VBKS VBKS VBKS Heater Temperature Sensor TSH BIC Pump Valve VBICP VBICP VBICP VBICP Heater Inlet Temperature Sensor TSHE TSHE TSHE TSHE Upline Tank Valve VVB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 30: Rinsing Bridge

    TSDE 2 2 2 2 Pressure Sensor Dialysate Outlet PDA 5 5 5 5 Blood Leak Detector BL BL BL BL 3 3 3 3 Dialyser Inlet Valve VDE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 31: Rear Door

    Rear Door SMPS-MC Fig. : Rear Door 2.14.1 Legend Rear Door 1. Switch Mode Power Supply Microcontroller SMPS-MC SMPS-MC 3. Mains Switch SMPS-MC SMPS-MC 2. Fan 4. Mains Cord B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 32: Switch Mode Power Supply Microcontroller Smps-Mc

    TN/Item-No (Version Number SMPS-MC): TN/Item-No (Version Number SMPS-MC): TN/Item-No (Version Number SMPS-MC): e.g. XXXXXX.5 Ver/SW (SW Version): Ver/SW (SW Version): Ver/SW (SW Version): Ver/SW (SW Version): e.g. Rev. 1.15 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 33: Tft Monitor

    Fig. : TFT Monitor 2.16.1 Legend TFT Monitor Backlight Inverter Board BIB Optical Status Display Board OSDB Front Panel Board FPB Touch Controller Board TCB TFT Monitor TFT Touch Screen B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 34
    Substitution Port 1 1 1 1 PBS/SN PBS/SN Arterial Pressure Sensor PA Connection for concentrate Suction Rods 2 2 2 2 PA PA Venous Pressure Sensor PV PV PV PV B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 35
    Venous Pressure Sensor Valve VPV Inlet Hydrophobic Filter HFE Arterial Pressure Sensor Valve VPA Pump Hydrophobic Filter (Ceramic Filter) HFP Blood Side Test Pressure Sensor Valve VBT Level Regulation Pump LRP B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 36: Front Door

    • Monitoring of the blood side pressures (pressure sensors PA, PV, PBE and PBS) • Test of the blood side pressure sensors in preparation (3/2 way solenoid valve VBT) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 37: Flow Diagrams

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 24 SW 9.xx 2.19 Flow Diagrams 2.19.1 Dialog+ B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 38: Dialog+ With Bic Option And Df Filter Option

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 25 SW 9.xx 2.19.2 Dialog+ with BIC Option and DF Filter Option B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 39: Dialog+ Hdf Online

    Dialog 2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 26 SW 9.xx 2.19.3 Dialog+ HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 40: Legend Flow Diagram

    Membrane Position Sensor Balance Chamber 2 Level Sensor Upline Tank NSVB Arterial Pressure Sensor Pressure Sensor Blood Inlet Blood Pressure Control Sensor Pressure Sensor Dialysate Outlet Degassing Pressure Sensor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 41
    VEBK1 VEBK2 Inlet Valve Balance Chamber 2 Air Separator Valve Upline Tank Valve Upline Tank Inlet Valve VVBE Circulation Valve Water Block Heat Exchanger Dialyser Coupling (to Dialysate) Z.D. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 42: Description Flow Diagram

    The cold inlet water is warmed up via the heat exchanger WT. Thereby the heat consumption to heat up the water is reduced. VVB* VBKS RVBO* Fig. : Water Inlet with Water Block B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 43: Degassing Circuit With Temperature System

    SMPS-MC. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* TSHE VBKO* VVB* VBKS RVBO* Fig. : Degassing Circuit with Temperature System B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 44: Dialysate Processing

    FPE. If the set pressure is reached RVFPE is opened and the fluid can circulate. RVFPE TSD-S TSDE TSBIC BICLF ENDLF ENDLF-S VDEBK1 VEBK1 BICP MSBK1 VDEBK2 VEBK2 MSBK2 Fig. : Dialysate Processing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 45: Central Bicarbonate And Concentrate Supply (Option)

    FPE (VBICP and VBKS are opened). RVFPE TSD-S DBK* TSBIC BICLF ENDLF ENDLF-S BICP VBICP* VBKO* FBIC VBKS VVB* RVBO* BKUS RVBU Fig. : BIC Cartridge Holder (Option) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 46: Balance Chamber System

    2 is drained (see description phase 2). VDEBK1 VABK1 VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 LAFS VEBK2 VDABK2 FBK2 MSBK2 open RVFPA closed Fig. : Phase 1 Balance Chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 47
    VDEBK1 VABK1 VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 LAFS VEBK2 VDABK2 FBK2 MSBK2 open RVFPA closed Fig. : Phase 2 Balance Chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 48: Ultrafiltration And Rinsing Bridge

    VLA is opened and air is removed with positive pressure from LA. VLA is closed once air is not detected anymore by LAFS. The balance chamber is switched to normal operation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 49: Chemical Thermal Disinfection Program

    ENDLF-S VB WT FBK1 Z.D. VEBK1 VDABK1 MSBK1 VDEBK2 VABK2 V.D. BICP LAFS VEBK2 VDABK2 FBK2 MSBK2 RVDA TSHE SBS1 RVFPA SKBS SKKS RVUFP SBS2 Fig. : Phase 2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 50
    C and conductivity is below 0.1 mS/cm. Then the machine gives a message that asks the user to confirm that the machine is free of disinfectant and the disinfection is completed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 51: Block Diagram

    VSAE VBKS* Online VBKO* Board PSAUS PSABFS FEDFFS FEHDFS SAKV SAKA Mains Switch Mains Voltage SMPS-MC * Battery Switch Mode Power Supply * Options Microcontroller Fig. : Block Diagram B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 52: Legend Block Diagram

    TSDE Dialyser Inlet Temperature Sensor Bicarbonate Conductivity BICLF END Conductivity ENDLF Degassing Pressure Sensor Air Separator Level Sensors LAFS Supervisor Dialysate Temperature Sensor Supervisor TSD-S END Conductivity Supervisor ENDLF-S B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 53
    Arterial Pressure Sensor Valve Venous Tubing Clamp SAKV Arterial Tubing Clamp SAKA Heparin Pump Compact Heparin Pump Compact Light Barrier Speed Monitoring Interlocking Sensor KRALO Syringe Plunger Sensor KOLB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 54
    VDFF VSAE Substitution Connection Inlet Valve Substitution Bypass Valve Port Substitution Outlet Sensor PSAUS Port Substitution Drain Sensor PSABFS DF Filter Detection Sensor FEDFFS FEHDFS HD Filter Detection Sensor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 55: Switch Mode Power Supply Microcontroller Smps-Mc

    1 / 2 0 1 0 2 — 42 SW 9.xx 2.22 Switch Mode Power Supply Microcontroller SMPS-MC 2.22.1 Block Diagram SMPS-MC Fig. : Block Diagram Switch Mode Power Supply Microcontroller SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 56: System Integration Smps-Mc

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 43 SW 9.xx 2.22.2 System Integration SMPS-MC Fig. : System Integration B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 57: Component Layout Smps-Mc

    2 . T e c h n i c a l System Description 1 / 2 0 1 0 2 — 44 SW 9.xx 2.22.3 Component Layout SMPS-MC Fig. : Layout SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 58: Wiring Diagram Smps-Mc With Battery Option

    2 x 12 V, 7 Ah 10 AT (5×20), breaking capacity 1500 A, 150 Vdc Fuse for Battery Jumper X104 Charge Current for Batteries Default: 50 % (right position) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 59: Description Smps-Mc

    ±12 VD: Digital +24 VGB: Switched blood side +24 VGD: Switched dialysate side +24 VL: Power (uncontrolled) GNDAN: GND analog GNDD: GND digital GNDH: GND aux. voltage GNDL: GND power B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 60
    A DC/DC converter generates the –12 V voltage from the 24 V. The soldered fuse F500 disconnects the 24 V in case of a short circuit in the DC/DC converter. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…
  • Page 61: Fuses

    +24 VGD 3.15 A (TR5) P4/15 P12/14 Note: Soldered fuses (e.g. F300 or F500) must not be changed if they are defective, i.e. the complete SMPS-MC must be exchanged. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 62: Signals

    = 10 s after the switch-on of the switch mode power supply. The watchdog has no safety function and is therefore not tested before the therapy starts. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd…

  • Page 63
    3-25 3.3.8.11 Pressure Reduction VABK with UFP 3-26 3.3.8.12 VEBK Test 3-27 3.3.8.13 Pressure Reduction 3-28 3.3.9 Integrity Test HDF Online 3-29 3.3.10 UF Pump 3-31 3.3.11 Conductivity 3-32 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 64
    3-55 3.5.9 Test Run 3-55 3-56 Field Service Utilities FSU 3.6.1 FSU Directories and Subdirectories 3-56 3.6.2 Start Field Service Utility FSU 3-57 3.6.3 Set-Up FSU Main Menu 3-58 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 65
    Measures after Repair 3.8.1 Repair Matrix 3-103 3.8.2 Tests and Measures to be Executed 3-104 3.8.2.1 Install Software 3-104 3.8.2.2 Check System Configuration, Production Report and Set if 3-104 Necessary B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 66
    Test Run UF Comparison Measurement 3-105 3.8.2.6 Test Run 3-105 3.8.2.7 Electrical Safety Check 3-106 3.8.2.8 Perform Disinfection after Repair 3-106 3.8.2.9 Document the Executed Activities 3-106 3.8.3 Execution Protocol 3-107 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 67: Self Test Sw

    If an error is detected during the MPU test the following alarms are displayed: In Therapy – Therapy Selection: • Hardware error RAM/ROM (SUP) In TSM Service Program: • Error in MPU test B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 68: Status Self Test In Service Overview

    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 69: Set-Up Low Level Controller Llc Test Codes

    Format of LLS Test Codes: (Value ≥ 10000) LLS test code to LLC S S S Performed test step Example: 11104 Blood leak detector test: measure level for red and green B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 70: Example Llc/Lls Test Codes For Blood Leak Detector

    Pressure sensors blood side: Equality test for blood side pressure sensors Pressure sensors blood side: Blood side pressure test Disinfection Valve VD Pressure test substitution line S-Online HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 71: Service Mode (Dir_On)

    Set set point for heater to self test code #112 2720x Wait for feedback of test result via BIOS (timeout 15 s) 2730x Set set point of heater to 0 2799x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 72: Eeprom

    Activate buzzer after approx. 1 s for 250 ms 28003 2820x Wait for feedback of test result via BIOS (timeout 15 s) 2830x Set set point of heater to 0 2899x Completed VGD ON VGB ON B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 73: Alarm System

    Acknowledge with switch-off VGD 16022 – 16027 1640x Wait for D24OFF (timeout 60 s since start of code 200) VGD ON 16028 1650x Wait for D24ON (timeout 10 s) 1699 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 74: Monitoring Of Analog Voltage ±12

    Acknowledge with switch-off of VGD 11005 – 1120x Wait for D24OFF (timeout 30 s) 11011 VGD ON 11012 1130x Wait for D24ON (timeout 30 s) 1199x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 75: Dfs Pressure Test (Table)

    ∆PDA < 25 mmHg, 12077 VBP ———————————- opened VDA, VDE —————————— closed 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 76
    FPA = 500 ml/min Valve Status: VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDE —————————— opened VDA ———————————— closed PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 77
    Pressure reduction FPE = 500 ml/min FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2 ——————- opened VABK1/2, VDABK1/2——————- opened VBP, VDA, VDE ———————— opened Timeout 60 s 1299 Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 78: Dfs Pressure Test

    • UFP is started during the pressure build-up phase. • The acknowledgement of the single test steps in the DFS pressure test is performed via a bit in the internal communication (between LLS and LLC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 79: Membrane Movement/Pressure Build-Up Vdabk

    FPA = 20 ml/min Valve Status: VEBK1/2, VDEBK1/2, VABK1/2 ————opened VDABK1/2—————————— closed VDE, VDA, VBP ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 80: Pressure Reduction Vdabk With Ufp

    FPA = 100 ml/min Valve Status: VEBK1/2, VABK1/2 ————————opened VDABK1/2, VDEBK1/2 ——————- closed VDE, VBP —————————— opened VDA ———————————— closed Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 81: Vbp Test

    12036 VDE, VBP —————————— closed 12037 VDA ———————————- opened Final Condition: 12038 Test timeout 30 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 82: Vde/Vda Test

    12076 VBP ———————————- opened 12077 VDA, VDE —————————— closed Final Condition: 12078 Test timeout 60 s or wait for inversion of Acknowledge with inversion of LLS_SELFTESTBIT LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 83: Pressure Build-Up Vdabk

    FPA = 100 ml/min Valve Status: VEBK1/2, VDEBK1/2, VABK1/2 ——— opened VDABK1/2 —————————— closed VBP, VDA, VDE ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 84: Pressure Reduction Membrane Test With Ufp

    FPA = 20 ml/min Valve Status: VEBK1/2, VDABK1/2 ——————- opened VDEBK1/2, VABK1/2 ——————— closed VBP, VDE —————————— opened VDA ———————————— closed Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 85: Membrane Test

    ∆PDA < 25 mmHg, 12131 VBP, VDE, VDA ———————— opened 12132 Final Condition: 12133 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 86: Pressure Build-Up Vabk

    FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2, VDABK1/2 ——— opened VABK1/2 —————————— closed VBP, VDA, VDE ———————— opened Final Condition: PDA ≥ +425 mmHg for 1 s or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 87: Membrane Movement To Inlet Side

    VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDA, VDE ———————— opened Final Condition: Test timeout 60 s or no increase of the values of the membrane position sensors B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 88: Pressure Reduction Vabk With Ufp

    Valve Status: VEBK1/2, VABK1/2 ——————— closed VDEBK1/2, VDABK1/2 —————— opened VBP, VDE —————————— opened VDA ———————————— closed VZ ———————————— opened Final Condition: PDA ≤ +380 mmHg or timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 89: Vebk Test

    ∆PDA < 25 mmHg, 12181 VBP, VDE, VDA ———————— opened 12182 Final Condition: 12183 Acknowledge with inversion of LLS_SELFTESTBIT Test timeout 60 s or wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 90: Pressure Reduction

    1280x 12000 VEBK test FPE = 500 ml/min FPA = 500 ml/min Valve Status: VEBK1/2, VDEBK1/2 ——————- opened VABK1/2, VDABK1/2——————- opened VBP, VDA, VDE ———————— opened Timeout 60 s B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 91: Integrity Test Hdf Online

    PDA < -100 mmHg, 15033 Test timeout 1 min or wait for switch-off VGD Test delivery rate OSP at 200 ±15 ml/min 15034 – from LLS Acknowledge with switch-off of VGD 15039 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 92
    ∆PDA < 50 mmHg, Test timeout 1 min or wait for switch-off VGD from LLS PDA < -200 mmHg, 15112 – Acknowledge with switch-off of VGD 15118 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 93: Uf Pump

    3 revolutions have been detected Acknowledge with switch-off of VGD 20006 – 20011 2020x Wait for D24OFF (timeout 10 s) VGD ON 20012 2030x Wait for D24ON (timeout 10 s) 2099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 94: Conductivity

    The heater is switched off at TSD-S > 41 °C Acknowledge with switch-off of VGD 22006 – 22011 2220x Wait for D24OFF (timeout 10 s) VGD ON 22012 2230x Wait for D24ON (timeout 10 s) 2299x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 95: Sad Level Test

    ≤ 1689 µs, then acknowledge with switch-off VGB Acknowledge with switch-off VGB 31002 – 31007 3120x Wait for B24OFF (timeout 10 s) VGB ON 31008 3130x Wait for B24ON (timeout 10 s) 3199x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 96: Equality Test For Blood Side Pressure Sensors

    ±40 mmHg. (Error) the test prerequisites from HDFonline-Task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). Completed Close VPD B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 97: Pressure Test Substitution Line S-Online Hdf Online

    Wait for B24ON (timeout 15 s) 4070x (Error) the test prerequisites from HDFonline task are not given during one of the test steps (e.g. autopriming, HDF filter rinsing required after LF error). 4099x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 98: Blood Side Pressure Retention Test

    (Timeout 35 s since start of subcode 400) Close VBT Test VBT closed 33039 |PA — PV| ≥ 200 mmHg 33040 Acknowledge with inversion of LLS_SELFTESTBIT 3350x Wait for inversion of LLS_SELFTESTBIT from LLS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 99: Disinfection Valve Vd

    Wait for B24OFF (timeout 7.5 s) VGB ON 37010 3730x Wait for B24ON (timeout 5 s) 3780x (Error) at start of test one or both couplings connected to rinsing bridge. 3799x Completed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 100: Lx800 Motherboard

    Hard Disk Drive USB8 USB7 USB6 USB5 USB4 Floppy Disk Drive USB2 USB1 USB1 Service SW/FSU PS/2 COM2 COM3 COM4 COM1 USB3 (DIABUS) (ABPM) (Touch Screen) (FPB) Type A B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 101: Ide Interface Lx800 Motherboard

    3.4.5. USB7 USB6 USB5 USB4 USB1 LX800 Assignment USB Ports USB8 USB1 USB2 USB3 USB4-8 USB Service SW/FSU Card Reader DSI (CritLine) Adimea (Kt/V-UV) Fig.: USB Ports B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 102: Detection/Boot Sequence From Bootable Usb Sticks

    3 — 40 SW 9.xx 3.4.4 Detection/Boot Sequence from Bootable USB Sticks Use only USB sticks supplied by B. Braun as tool for the installation of the software. Dialog+ machines detect automatically connected bootable USB sticks. This is indicated by the red Installation text in start screen.

  • Page 103: Bios Settings For Lx800 Motherboard

    3. Press the TAB key when the logo is displayed. Main Processor..4. Press the Pause/Break key to stop the boot routine. The BIOS ..version is displayed, e.g. BIOS Revision 1.04..Pause Break B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 104
    [Normal High] SHFCLK Active Period [Free Running] LP Active Period [Free Running] Onboard Audio [Disabled] Onboard USB1.1 [Enabled] Onboard USB2,0 [Enabled] Onboard DIE [Enabled] Memory Hole At 15M-16M [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 105
    PNP OS Installed [No] Init Display First [PCI Slot] Reset Configuration Data [Disabled] Resources Controlled By [Auto(ESCD)] x IRQ Resources Press Enter x Memory Resources Press Enter PCI/VGA Palette Snoop [Disabled] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 106: Pc Health Status

    ! ! ! ! » » » » # # # # $ $ $ $ :Move Enter:Select +/-/PU/PD:Value F10:Save Esc:Exit F1:General Help F5: Previous Values F6: Fail-Safe Defaults F7: Optimied Dafaults B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 107
    ! ! ! ! » » » » # # # # $ $ $ $ Esc : Quit : Select Item F10 : Save & Exit Setup Save Data to CMOS Delete Enter Delete Arrow Keys B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 108: Installation Software

    Open rear door. Switch to TSM service program mode (on digital board DB, service switch S1 position 2). Switch on Dialog+. Document default values in default table. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 109
    Position 2: TSM Service Program Mode Supervisor LEDs 0 to 7 (D28 – D35): Position 3: Software Installation/Update Mode Status 0 – 7 for Installation of LLS Software Jumper JP2 SWP: Supervisor Write Protect B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 110
    JP1 and JP2 have to be set to this position (Default CWE and SWE: controller/supervisor write enabled). Set the jumpers back to the CWP/SWP setting after installation. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 111
    Bootable USB stick with the TLC/LLC/LLS software 14. Open front door. 15. Connect USB stick A to USB port B on the TLC sub-rack C. 16. Switch on machine. Fig.: Front View TLC Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 112
    USB Stick not Detected The system does not detect the bootable USB stick B BRAUN (indicated by the green logo in the start screen). If necessary disconnect the USB stick and reconnect. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 113
    If YES, press Y or y and <CR> if keyboard is used otherwise press the EQ hardware key! If NO, press N or n and <CR> on keyboard or any other hardware key! Press to confirm the selected CFC. [EQ] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 114
    Don’t forget to set the operation mode switch to TSM mode (2) and check in TSM all subsystems’ version number. Switch the power OFF and REMOVE the USB stick before starting the Dialog… The TLC software was installed successfully. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 115: Boot Machine /Check Version Number

    Enter Customer Specific Default Values The customer specific default values of the machine must be entered after the software installation. 27. Enter customer default values from default table and save B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 116: Enter Specific Parameters

    33. Set system configuration (customer specific values) or if available use the FSU USB stick with the system configuration. 34. Save settings. 35. Switch off machine. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 117: Self Test

    (The ABPM function (if present) can be checked without a test run of the machine in preparation mode, without the concentrate pump.) 12. Terminate dialysis. 13. Perform disinfection/decalcification. 14. Remove blood lines, filter and heparin syringe. 15. Switch off Dialog+. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 118: Field Service Utilities Fsu

    The field service utilities has the following directory structure (see left figure). BlueCat DISKTYPE USER 04_Write_Languages 05_Write_Calibration_Data 06_Write_System_Configuration 09_Card_01_Read_Card_Image 09_Card_02_Write_Card_Image 09_Card_03_Read_Patient_Diskette 09_Card_04_Write_Patient_Diskette_to_Card 09_Card_05_Read_Patient_Card 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends 13_Read_Screenshots 14_Get_Error B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 119: Start Field Service Utility Fsu

    To prevent the corruption of USB stick, before removing it, please press AQ, then follow the instructions on the screen! 4. Press the EQ hardware key to execute the utility. Press the AQ to exit the FSU program. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 120: Set-Up Fsu Main Menu

    Card Commands in the main menu. 2. Press the EQ hardware key to execute the utility. 3. Press the BP hardware key and move the cursor down to select a utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 121: Hardware Keys For Fsu

    Press the AQ hardware key to exit the FSU menu or to skip back to the previous menu. Press the EQ hardware key to execute a selected utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 122: Set System Version Number

    BP stop stop hardware keys. Note: (n.nn = the version number) 4. Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 123: Set Working Time Counter (Wtc)

    3. Set the new working time counter with the BP and BP hardware keys. stop stop Press EQ to accept the entered value. The main menu is displayed after a few seconds. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 124: Kill Masterboot Record

    A r e y o u s u r e ? 3. Press AQ if you do not want to delete the masterboot record. AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 125
    A r e y o u R E A L L Y s u r e ? the masterboot record. 5. Press BP to cancel the operation. BP-: no BP+: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 126: Write Utilities

    The last write procedure can be undone with the 07 Undo Last Written utility. The Dialog+ can be switched off and on again between the write procedure and using the 07 Undo Last Written utility. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 127
    After successfully copying the language to the CFC the version numbers of all languages available on the Dialog+ are displayed. The language selected for treatment is also marked. 4. Press EQ to continue. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 128: Write Calibration Data

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 129
    (SYSver) stored in the calibration data file with the serial number and the system version number from the Dialog+. The calibration data file can be written only if they match. Example Ser-No.= 90832 SYSver = V8_20 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 130
    The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the calibration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 131: Write System Configuration

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 132
    Yes, system configuration file is written to Writing System Configuration to Dialog+ from: TLCver=90832/VX_XX Dialog+ hard disk drive (CFC) A r e y o u s u r e ? AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 133
    The pairs that do not match are displayed with the following message: (not selectable) 5. Press EQ if you are sure. This will write the system configuration data from the USB stick to the Dialog+ hard disk drive (CFC). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 134: Undo Last Written

    3. Press EQ if you are sure. This will undo the last written utility from the USB stick to the Dialog+ hard disk drive (CFC). AQ: no EQ: yes B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 135: Usb Stick Self Test

    Searching for md5sum files….the main menu after the self test. MD5SUMS.XXU.Win found..MD5SUMS.XXU.Lin found..MD5SUMS.SPEC.Lin found..Checking file system integrity..O.K. Calculated CRC = XXXX Reference CRC = XXXX B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 136: Card Reader Read/Write Utilities

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 137: Read Card Image

    E n d o f R e a d C a r d I m a g e P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 138: Write Card Image

    E n d o f Write Card Image P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 139: Read Patient Diskette

    P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 140: Write Patient Diskette To Card

    E n d o f Write Patient Diskette to Card . . . P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 141: Read Patient Card

    P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 142: Write Patient Card

    E n d o f Write Patient Card . . . P r e s s t h e ‘ A Q ’ k e y t o c o n t i n u e . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…

  • Page 143: Remove All Patient Data

    05. Read Patient Card (USB Stick → Card) 06. Write Patient Card 07. Remove All Patient Data BP-: move cursor up BP+: move cursor down AQ: back to previous menu execute B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 144
    U S E R / 0 9 _ C a r d _ 0 3 _ R e a d _ P a t i e n t _ D i s k e t t e ….E n d o f Remove All Patient Data . . . B. Braun Avitum AG BA-TE-DE08C M.KAY…
  • Page 145: Read/Get Error Utilities

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 146
    ….displayed….End of Read System Configuration… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 147: Read Calibration Data

    BP-: move cursor up BP+: move cursor down AQ: exit – see Important Note EQ: execute Important Note: To prevent the corruption of USB stick, before removing it, please press AQ then follow the instructions on the screen! B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 148
    ……. . End of Read Calibration Data… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 149: Read All Trends

    Year 2007 09_Card_03_Read_Patient_Diskette M M : 09 month September D D : 12 Wednesday 09_Card_04_Write_Patient_Diskette_to_Card h h : 14 hours 09_Card_05_Read_Patient_Card m m : 25 minutes 09_Card_06_Write_Patient_Card 10_Read_System_Configuration 11_Read_Calibration_Data 12_Read_All_Trends B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 150
    ……. . End of Read Trends… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 151: Read Screenshots

    Dialog+. The vga*.zip files copied to USB stick are deleted from the Dialog+ hard disk drive (CFC) (similarly to the log files) but the subdirectory structure on the Dialog+ hard disk drive (CFC) is left untouched. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 152
    ……. . End of Read Screenshots… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 153: Get Error (Errordisk)

    3 — 91 SW 9.xx 3.6.10.5 Get Error (ErrorDisk) The error files can only be analysed by qualified personnel of B. Braun Avitum (e.g. development department) and should be provided only on special request. The utility gets error information and log files from Dialog+ hard disk drive (CFC) and writes them to the USB stick.

  • Page 154
    Dialog+ hard disk drive (CFC) to the USB stick… The Get Error (ErrorDisk) menu automatically returns (a ..few seconds) to the main menu End of Error Disk… is displayed….. End of Error Disk… B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 155: Fsu Handling Of Errors

    USER/XX_Read..USER/14_Get_Error These files can be copied to any location on the PC (e.g. to use them for local examination) or can be deleted from the FSU USB stick. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 156: Troubleshooting Usb Stick

    FSU program. — — — S W I T C H O F F t h e D i a l o g + — — — B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 157
    Extended TMP Limit Range Button is Displayed in Treatment Select Extended TMP Limit Range inactive inactive inactive Arterial Bolus Parameters Arterial Bolus Volume Arterial Bolus with SAKA Support B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 158
    2000 2000 Therapy Time h:min 04:00 04:00 03:30 Gross UF Rate vs. Blood Flow inactive inactive inactive Gross UF Rate vs. Blood Flow active active active UF Profile Editor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 159
    Default Setting for Treatment Mode: Warning of Kt/V Target-Projected Deviation Warning active Warning active Warning inactive Warning for Saving Kt/V Table Timings Parameter Setting Window Disappearing Time Screen Saver Appearance Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 160
    Load Interval inactive inactive inactive Elapsed Month inactive inactive inactive Setting Elapsed Month months Elapsed Working Time Counter WTC inactive inactive inactive Setting Elapsed WTC hours 4000 4000 4000 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 161
    HDF Online Substitution Pump OSP (2.15) Unit Default US Default Modified µl/Head Revolution 8300 8300 Level Reg ulatio n ( 2.16) Unit Default US Default Modified LRP (PPR) Slow LRP (PPR) Fast B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 162
    Morning 06:30 Rinsing Tuesday Night Morning 06:30 Rinsing Wednesday Night Morning 06:30 Rinsing Thursday Night Morning 06:30 Rinsing Friday Night Morning 06:30 Rinsing Saturday Night Morning Sunday Night Morning B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 163
    Temperature Conductivity Name Volume Long Short Time min. (max. 20 Characters) h:min h:min h:min mS/cm Citric Thermal 00:15 00:05 00:05 Decalcification Short 00:05 00:05 00:05 Bleach 00:15 00:05 00:25 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 164
    Lo 15 xevonta Lo 18 xevonta Lo 20 xevonta Lo 23 xevonta Hi 10 xevonta Hi 12 xevonta Hi 15 xevonta Hi 18 xevonta Hi 20 xevonta Hi 23 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 165: Measures After Repair

    4.9.4.19 Calibration Dialyser Inlet Throttle Valve DDE 4.9.4.20 Calibration Pressure Reducer Valve DMV 4.9.4.21 Gap SAKA and SAKV 4.9.4.23 Setting Servomotor for Disinfection Valve VD Inspection Protocol for Automatic Blood Pressure Measurement ABPM B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 166: Tests And Measures To Be Executed

    LLC Calibrations Check and set ratio, tubing constants BPA/BPV, OSP tubing type, check/set PA/PV limit values. Working Time Counter • Set working time counter. Motherboard Check and set BIOS. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…

  • Page 167
    5. Set the blood pump so that no blood side alarms are activated. 6. Start dialysis. 7. Check alarm function of SAD. (The ABPM function can be checked without a test run of the machine in preparation mode.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 168
    1. The measures were executed correctly. Fill in the execution protocol after service (e.g. exchange of a component) and after executing all necessary measures according to the repair matrix. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 169
    Inspection Protocol for Automatic Blood Pressure Measurement ABPM} Name Service Technician: The respective measures were executed correctly after servicing according to the repair matrix…………………………..Date / Signature B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd…
  • Page 170
    Timings 4-49 4.7.1.15 Language Selection 4-50 4.7.1.16 Automatic Blood Pressure Measurement ABPM 4-51 ® 4-52 4.7.1.17 bioLogic RR GL Parameters 4.7.1.18 Summer Time Setting 4-53 4.7.1.19 Miscellaneous Parameters 4-54 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 171
    Test 1.14 Bypass and Disinfection Valves 4-88 4.8.2.15 Test 1.15 Flow Pump FPA, Balance Chamber Sensors and Air 4-89 Separator 4.8.2.16 Test 1.16 Pressure Water Part 4-91 4.8.2.17 Test 1.17 UF Pump 4-92 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 172
    4.8.3.10.3.1 Turbidity Calibration with Water 4-144 4.8.3.10.3.2 Calibration with Reference Solution (Red Dye) 4-144 4.8.3.10.4 Actual Value Check Blood Leak Detector 4-145 4.8.3.10.4.1 Check 0 ‰ Blood Concentration 4-145 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 173
    4-170 Currentless Closed SAKV-SG 4.8.3.27 Setting Servomotor for Disinfection Valve VD 4-171 4-173 Production Report 4-174 4.9.1 Version Number Dialog 4-175 4.9.2 Low Level Options 4-176 4.10 Service Reports B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 174: Start Tsm Service Program

    Position 3 = Software Installation/Update 4. Switch on machine. The machine starts in TSM mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 175
    The File Operations menu can be activated with this icon. The current software/firmware versions are displayed in the Dialog Version Number menu. This menu can also be selected in the Production Report. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 176: Quit Tsm Service Program

    3. Close rear door. 4. Switch on machine. The machine starts in therapy mode. Online Board Power Board M otors Power Board Valves Digital Board P1 1 Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 177: Structure Of Tsm Service Program

    Calibration 2.13.2 Pump Head Selection {} : not applicable {Calibration 2.14 Substitution Pump} Calibration 2.15 HDF Online Substitution Pump OSP Calibration 2.16 Level Regulation Pump LRP Fig. : TSM Service Program B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 178: Service Overview In Therapy

    7500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 179
    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Second Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 180
    L L C / S & T L C A l a r m W i n d o w L L C M e s s a g e W i n d o w Fig.: Third Page Service Overview in Therapy B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 181: Legend Service Overview In Therapy

    LS CURRENTS_S: Loudspeaker current supervisor LS PULSE_S: Loudspeaker pulse supervisor MSBK1: Membrane position sensor balance chamber 1 MSBK2: Membrane position sensor balance chamber 2 NSVB: Level sensors upline tank B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 182
    VDE: Dialyser inlet valve VDEBK1_S: Dialyser inlet valve balance chamber 1 supervisor VDEBK2_S: Dialyser inlet valve balance chamber 2 supervisor VDE-S: Dialyser inlet valve supervisor VDFF: DF filter valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 183
    LLC text messages according to the LLC message hex numbers from LLC L L C M e s s a g e W i n d o w message window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 184: Service Overview In Tsm Service Program

    5500 Malfunctions in the balance chamber can occur if: • the values are out of limits • the minimal difference from the detected min. value was not reached B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 185: Legend Service Overview In Tsm Service Program

    Dialysate outlet pressure sensor supervisor Degassing pressure sensor Venous pressure sensor PV_S: Venous pressure sensor supervisor RDV: Venous red detector RDV_S: Venous red detector supervisor RVE: Degassing control valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 186
    VSAE_S: Substitution inlet valve supervisor VSB: Substitution bypass valve VSB_S: Substitution bypass valve supervisor VVBE: Upline tank inlet valve Circulation valve VZ_S: Circulation valve supervisor WTC: Working time counter B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 187: File Operations

    LLC/LLS after every calibration. The calibration data can be downloaded from the compact flash card CFC (hard Down Load Calibration Data disk drive) to the exchanged digital board DB. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 188
    • Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 189: Treatment Support

    The new settings are valid after the restart of the machine. There are five trend groups, which can be selected for the treatment program. Trends and Trend Groups Every group has three treatment parameters. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 190: System Configuration

    1. Click the mmol/l or mS/cm key to select a measurement unit. The acetate mode can be disabled. Acetate Mode Disabled 1. Click the YES key to disable the acetate mode. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 191
    [mS/cm per mmol/l]. Conversion Factor: Conductivity mS/cm (Standard Value) Conversion Factor NaCl Concentration mmol/l (Measured Value) 4. Click the Default Value window if necessary to change the value [mS/cm]. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 192
    [mS/cm]. 4. Click the ACID icon if necessary to edit the following parameters: • Acid Concentrate Name • Conversion Factor [mS/cm per mmol/l] • Default Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 193
    1. Click the ml/min key to set the dialysate flow. You can set the dialysate temperature for therapy. Dialysate Temperature 1. Click the C key to set the dialysate temperature. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 194: Blood Side Parameters

    The selected Chamber-free AV Line does not appear. AV Line with PBE Connector Activate if AV lines with PBE pressure connection are used. Do not activate if AV lines without PBE pressure connection are used. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 195: Min-Max Parameters

    You can set the upper relative alarm limit for the Delta PA max. PA limit window. Delta PA PA max. max. PA ref 200 mmHg PA min. Fig.: Limit Window for PA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 196
    The TMP low limit range is set min. –10 mmHg. If the key is not pressed The TMP limit range is extended to –100 mmHg If the key is pressed (absolute TMP value). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 197
    • a warning message is not displayed anymore TM P 20 mmHg 12 — 28 m m Hg • treatment is interrupted • machine is switched to bypass -100 Bottom TM P Lim it Value B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 198: Arterial Bolus Parameters

    Arterial Bolus with SAKA Support Arterial Bolus Parameters menu. Prerequisites are: • Arterial tubing clamp SAKA present • An arterial line system with infusion access is present between the SAKA and blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 199: Single-Needle Parameters

    Furthermore the control pressure PV can be set. Thereby the blood pump is started/stopped and the venous tubing clamp currentless closed SAKV-SG is opened/closed. The control pressures for PA and PV can be set for single-needle cross-over. Single-Needle Cross-Over B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 200: Heparin Parameters

    Heparin Bolus Volume Therapy Beginning Bolus An automatic bolus can be administered at the beginning of a therapy. The heparin bolus is performed by the venous red detector RDV. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 201
    You can set the syringe name and the inner diameter of the syringe. Heparin Pump Compact 3. Activate the Selected Syringe menu with the Selected Syringe icon. The following menu is opened. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 202: Rinsing Parameters

    • Rinsing Blood Flow • Rinsing Dialysate Flow • Rinsing Time by UFP • Rinsing Rate by UFP • Rinsing Volume by UFP • Blood Flow for Connecting Patient B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 203
    You can set the blood flow for connecting the patient (in therapy). The blood Blood Flow for Connecting Patient pump BPA is started with this rate when the therapy is started. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 204: Uf Parameters

    The alarm can be acknowledged with the key. Setting Range The alarm limit can not be set lower than the warning limit (even if the warning limit is deactivated). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 205
    A linear adaptation is calculated for the remaining process. EXP (Exponential) If this key is activated only the first interval can be set. An exponential adaptation is calculated for the remaining process. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 206
    The UF process of the activated interval is set to HD with the HD key. UF Mode Sequential SEQ The UF process of the activated interval is set to sequential with the SEQ key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 207: Hdf/Hf Online Parameters

    HDF Dialysate Flow for Predilution HDF/HF Infusion Bolus Volume The HDF/HF infusion bolus volume can be set. You can set predilution for a HDF/HF online therapy by activating the Post/Predilution Selection Predilution key. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 208: Disinfection Parameters

    • Auto Switch-On without Daily Confirmation • Maximum Out of Action Time (Day/Hour/ Minutes) • Disinfection Configuration Data Chemical (long), Chemical (short) • Disinfection Configuration Data Central Thermal, Central Chemical, Rinsing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 209
    The machine is rinsed after the preselected switch-on time. The key is only enabled if the function is selected. Switch-On Time The switch-on time can be set (without a weekly schedule). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 210
    The function is activated if the Limit OFF key is not pressed. The Limit OFF key is activated in the default setting, the parameter keys are disabled, i.e. parameter values can not be set. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 211
    • Rinsing Time Central Automatic Chemical: • Inlet Volume • Switch-off for Retention without Automatic Switch-on • Retention Time • Rinsing Flow • Rinsing Time Rinsing: • Inlet Flow • Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 212
    FBK1 VEBK1 VDABK1 M SBK1 VDEBK2 VABK2 BICP LAFS VEBK2 VDABK2 FBK2 M SBK2 RVDA TSHE RVFPA RVUFP Fig.: Flow Diagram — Central Automatic Chemical Disinfection of Water Inlet B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 213
    The rinsing flow can be selected. Rinsing Time The rinsing time can be selected. Central Chemical A central chemical mode can be selected for the automatic switch-on in the weekly schedule. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 214: Filter Parameters

    The operating hours can be set for the intended operation time of the DF/HDF Filter Operation Time [h] filters. The operating hour counter is active when DF Preparation is active. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 215: Dialyser Parameters (Kt/V)

    V (litres)=2.447 – 0.09516 x Age (years) + 0.1074 x Height (cm or feet/inch) + 0.3362 x Dry Weight (kg) Female: V (litres)=-2.097 + 0.1069 x Height (cm or feet/inch) + 0.2466 x Dry Weight (kg) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 216
    The Kt/V is approximately calculated via the Dialyser Filter Urea Clearance Data for the selected dialyser (see table, use scroll bar to see table). Details can be found in Chapter 3 repair instructions. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 217: Kt/V-Uv Parameters

    Default Setting for Treatment Mode: value, a message is activated. Warning of Kt/V Target-Projected Deviation A warning message can be enabled for saving the Kt/V table. Warning for Saving Kt/V Table B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 218: Timings

    Therapy and end of therapy: Time cake, remaining time, status line Disinfection: Time cake, disinfection mode, phase You can set the appearance time of the Screen Saver. Appearing Time B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 219: Language Selection

    English. If German is selected the therapy and TSM language is German. If an additional language is selected this language is used in therapy. The TSM language is English. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 220: Automatic Blood Pressure Measurement Abpm

    The limit values in therapy, e.g. for systole/diastole can be stored on a patient Taking Over Limits from Patient’s chip card. These limit values are taken over if the patient chip card is read. Parameters B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 221: Biologic Rr ® Gl Parameters

    The ID number 200 indicates that the Guide Line Mode is enabled and can be used in therapy. ID Number 100 The ID number 100 indicates that the Guide Line Mode is not enabled and can not be used in therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 222: Summer Time Setting

    If necessary you can set the time difference between summer and winter time. Time Difference Note: The Time Difference should be set to 00:00 for countries where daylight saving time is not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 223: Miscellaneous Parameters

    It is prohibited to run a therapy with a patient if the Skip Self Test function is activated. The self test must be passed successfully prior to a therapy. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 224
    A window is automatically opened at the end of therapy. This window must be Automatic Reinfusion Start at confirmed at the end of the therapy for the reinfusion of the blood. Entering End of Treatment B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 225: Dbi Parameters

    Nexadia functionality, e.g. support of the option bioLogic RR Comfort and Adimea; the mmol/l unit can be used. The Nexadia protocol must be supported by the host computer. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 226: Preventive Maintenance

    9 to 18 month period if the 4,000 working time is more than 4,000 hours after the last reset. W indow 4,000 hours vs 9 to 18 months Elapsed Months B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 227
    Press the Preventive Maintenance Accomplished key and then the Preventive Maintenance Accomplished Save icon after the preventive maintenance was performed. Consequently a new preventive service period is started again after the machine is switched off and on. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 228: Trends And Trend Groups

    Trend and Trend Groups key in the Treatment Support menu. The following menu is opened and displayed. 3. Activate the Trend History menu with the Trend History window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…

  • Page 229
    5. Activate the Edit menu with the Edit icon. 6. The treatment parameters in the six trend groups can be modified. 7. The default parameters can be reset with the Set Defaults window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 230: Manual Test And Calibration

    • Supervisor Window Test • File System Check • ABPM Maintenance (Option ABPM) • DBI Maintenance (Option Nexadia-BSL) • Kt/V-UV Sensor Test (Option Adimea) • DSI Test (Option Crit-Line) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 231: Buttons, Lamps And Sounds

    Check red LED for signal lamp (on optical status display boards OSD). Check yellow LED for signal lamp (on optical status display boards OSD). Check green LED for signal lamp (on optical status display boards OSD). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 232: Monitor Test

    1. Activate the Monitor Test menu with the Monitor Test key in the TL Manual Test and Calibration menu. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 233: Touch Screen Calibration

    (on the keyboard membrane of the TFT housing) three times in the TSM Main Menu. 2. Switch machine off and on again. The old default setting is loaded. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 234: Brightness Calibration

    The day brightness (digits) for the backlight set. Test OSD Red and Yellow (for Signal Lamps) The red and yellow LED of the optical status display OSD can be checked, i.e. switched ON and OFF. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 235: Card Reader And Card Check

    • Use an empty service chip card (art. no. 7703856) for the test and not a patient chip card. • Do not remove the chip card from the card reader during the test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 236: Usb-Port Properties

    The maximum speed of the USB port. Status The connected USB device is in the valid USB device list. Not OK The connected USB device is not in the valid USB device list. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 237: Supervisor Window Test

    Parameter Check key. The following menu is displayed. 3. Press the key to close the supervisor window. Press the key to select and acknowledge the single parameters and close the supervisor window. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 238: File System Check

    Hard Disk Check (for the compact flash card) to test the hard disk drive. Press the key of the Application SW Check to test the application software. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 239: Abpm Maintenance

    ABPM system is tested. The measurement chamber is connected to the ABPM system via the pressure tubing. A pressure is built up to 300 mmHg and is held for 3 minutes. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 240: Dbi Maintenance

    5. Switch off machine after successful test. 6. Remove the test plug from the RS 232 port. Pin Assignment RS 232 Test Plug (Soldering Side) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 241: Kt/V-Uv Sensor Test

    (residual digits). This value must be ≤4000 digits (maximum service life). Example: Required current: 1332 (digit) Reserve: 2668 (digit) Total digits: 4000 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 242: Dialog Serial Interface Dsi Test

    If the Sent Data and Received Data are identical, the test was successful. 6. Switch off machine after successful test. 7. Remove the test plug from the DSI interface. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 243: Low Level Manual Test

    Previous and Next Arrow keys. If required, a corresponding calibration menu can be selected in certain LL Manual Test menus. The calibration procedures are described in paragraph 4.8.3 Low Level Manual Calibration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 244: Test 1.1 Staff Call, Alarms And Power Off

    • -12 VAN > -11 V -12 VAN Control / Supervisor ON Test activated ±12 VAN Test: • +12 VAN < 11 V • -12 VAN < -13 V B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 245: Test 1.2 Blood Pump Arterial

    The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPA is stopped. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 246: Test 1.3 Blood Pump Venous

    The motor torque is displayed by the controller. Torque Cover State If the pump cover is opened the blood pump BPV/substitution pump is stopped immediately. The controller detects and displays the opened pump cover. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 247: Test 1.4 Heparin Pump

    (The clasp nut is assembled on the drive spindle and limits the max. pressure load of the heparin pump). Activates/switches off the heparin pump. HP Motor State Selection between slow and fast forward feed. HP Speed B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 248: Test 1.5 Tubing Clamps

    Thereby the switch function and leakage test (pressure — clamp SAKV and SAKA gap, time) can be tested. The closed clamp gap is set to 1.4 mm + 0.1 mm. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 249: Test 1.6 Pressure Sensors (Blood Side)

    Machine with Single Pump The value for PBS is 0. Machine with Double Double Pump must be activated in Production Report, Low Level Options. A value for PBS is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 250: Test 1.7 Air Sensor Sad, Red Sensor

    A tubing filled with fluid is inserted in the SAD If the SAD TEST is activated NO AIR must be detected. The light path is interrupted NO BLOOD (Controller/Supervisor) changes to BLOOD (Controller/Supervisor). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 251: Test 1.8 Coupling Detectors

    You can check the actual status of the reed sensors. Detectors Bicarbonate rinsing connection sensor. BICSS Concentrate rinsing connector sensor. Rinsing bridge connector sensor (dialyser coupling). SBS1 SBS2 Rinsing bridge connector sensor (dialyser coupling). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 252: Test 1.9 Water Inlet, Upline Tank And Flow Pump Fpe

    All balance chamber valves are opened and the outlet flow pump FPA works in main flow with the same speed as the inlet flow pump FPE. The degassing pump EP runs and the degassing bypass valve VEB is open. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 253: Test 1.10 Degassing And Heating

    C or the upline tank VB is empty. The degassing pump EP must be able to build up a flow of > 1200 ml/min at a negative pressure of — 500 mmHg. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 254: Test 1.11 Dialysis Temperature

    Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. The measurement range is as follows: TSE, TSD, TSBIC: 0 to 100 TSD-S: 30 to 55 TSDE: 30 to 55 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 255: Test 1.12 Concentrate Pumps And Conductivity

    Increase the speed [rpm] until the flow pump FPE has built up a flow of 750 ml/min. Thereby the pulsating flow of the pumps have no influence on the conductivity measurement. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 256: Test 1.13 Balance Chamber Valves

    The following valves can be activated: Valves VDEBK1/2 Dialyser inlet valve 1/2 balance chamber VEBK1/2 Inlet valve 1/2 balance chamber VDABK1/2 Dialyser outlet valve 1/2 balance chamber VABK1/2 Outlet valve 1/2 balance chamber B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 257: Test 1.14 Bypass And Disinfection Valves

    The following valves can be activated: Valves Bypass valve Dialyser inlet valve Dialyser outlet valve Disinfection valve Circulation valve Air separator valve B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 258: Test 1.15 Flow Pump Fpa, Balance Chamber Sensors And Air

    FPE. Simultaneously the fluid in the right side of the balance chamber is emptied to the drain. Drain Fig. : Filling Cycle Balance Chamber Phase 1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 259
    Thereby FPE delivers fluid into the Fig. : Filling Cycle Balance Chamber Phase 2 dialyser circuit. VLA is only closed and switched to normal operation if the top level sensor LAFSO detects fluid. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 260: Test 1.16 Pressure Water Part

    The pressure can be displayed in mmHg or in mbar, depending on selection. Conversion Factor mmHg to mbar Conversion Factor = 1.3332 Example: -100 mmHg approx. –133 mbar –400 mmHg approx. –533 mbar B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 261: Test 1.17 Uf Pump

    UFP, set by the value of the throttle DDE. Not applicable. Flow Detector Stroke Value UFP The Stroke Value UFP should not be used in a normal case, because UFP must be calibrated (see LL Manual calibration). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 262: Test 1.18 Blood Leak

    If air is visible check the complete system for leaks in menu 1.20 Test Water Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 263
    LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 — 1.44 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 264: Test 1.19 Leak Test

    The test is started with a negative test pressure and then followed by a positive test pressure. The leakage test consists of three test steps. If an error occurs an LLC alarm is LLC Alarm activated (LLC Status 00020000). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 265
    VDEBK1 VABK1 FBK1 Z.D. VEBK1 VDABK1 M SBK1 Dialyser VDEBK2 VABK2 V.D. LAFS VEBK2 VDABK2 FBK2 M SBK2 SBS1 RVUFP RVFPA SBS2 Fig.: Excerpt Flow Diagram for HDF Online B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 266
    Rinse water circuit to prevent air bubbles. b) Close the outlet. c) Activate test. The FPE builds up a pressure. d) Check pressure drop. The pressure drop must be < 50 mmHg in 30 s. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 267: Test 1.20 Water Part (Overview)

    Valves VDA/VDE are open (valve symbols are black). Valve VBP is closed (valve symbol is white). Bypass Valves VDA/VDE are closed (valve symbols are white). Valve VBP is open (valve symbol is black). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 268: Test 1.21 Emptying

    The actual value, i.e. the fluid level of the upline tank VB is displayed by the level sensors NSVB. Start the Emptying ON at least three times and manually clamp the bypass (tubing) to ensure a sufficient emptying (for transportation). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 269: Test 1.22 Hdf

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 100 SW 9.xx 4.8.2.22 Test 1.22 HDF Not applicable. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 270: Test 1.23 Bic Cartridge Valves And Coupling Detectors

    Valves VVB, VBICP, VBKO, VBKS The valves VVB, VBICP, VBKO and VBKS can be switched either separately or together. The status of the reed sensor (OPEN/CLOSED) is displayed. Detectors BKUS, BKUS_S B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 271: Test 1.24 Hdf Online

    DF filter and HDF filter are impermeable to air. Valve VBE can only be opened if PDA < -100 mmHg to prevent fluid penetration. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 272: Test 1.25 Hdf Online Valves And Coupling Detectors

    The status of the additional coupling sensor FEDFFS can be checked. FEHDFS (filter recognition HDF filter) The status of the additional coupling sensor FEHDFS can be checked. FEHDFS PSAUS PSABFS FEDFFS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 273: Test 1.26 Option Battery

    Akku OK (Battery OK) time Battery capacity < required capacity for 20 minutes battery running time External State Indicates if machine was switched on externally. Machine was switched on externally B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 274: Test 1.27 Dfs Pressure Test Of Therapy

    The test codes for LLC and LLS are displayed for all test phases. In case of an Test Code error the defective components can be determined with the help of the test description. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 275: Test 1.28 Front Panel Board Fpb (Lls)

    FPB and wait for switching (OFF/ON). • Enter Key Press the key on the front panel board FPB and wait for switching (OFF/ON). B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 276: Test 1.29 Patient Leakage Current

    C and is then switched off. The blood pump can be activated additionally if the measurement is performed with tubing set (line). Main Flow The machine can be switched between main flow and bypass. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 277: Test 1.30 Smps-Mc Self Tests

    > 30 s (disconnect mains plug from mains receptacle). #115 Buzzer Test by LLS Ver/SW The software version of the SMPS-MC is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 278: Self Tests Smps-Mc

    • LLC sends #115 to SMPS-MC • LLS must activate AKAL-S within 5 s • Buzzer OK. if not OK: not OK: • Warning is generated • Buzzer not OK. • Therapy still possible B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 279: Test 1.31 Level Regulation

    The delivery of the LRP can be checked with the Test Test key. A measurement chamber must be connected to PV. LRP stops when 500 ml are reached and the required time for the pressure build-up is displayed. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 280: Low Level Manual Calibration

    Input values can be increased in increments with the displayed hardware key as an alternative to the software key on the touch screen. The service overview can be selected in the TSM with the hardware key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 281
    If required, the previous calibration menu or the next calibration menu can be selected with the Previous and Next Arrow keys. The menu can be closed with the Exit Door key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 282
    C o n t r o l l e r / S u p e r v i s o r ( 1 2 b i t ) Voltage 0 to 5 V Digits 0 to 4095 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 283: Select Low Level Manual Calibration

    • 2.12 PV Alarm Window, PA Low Limit • 2.13 Flow of BPA and BPV Pump • 2.14 Substitution Pump (not applicable) • 2.15 HDF Online Substitution Pump OSP • 2.16 Level Regulation B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 284: Calibration Pressure Sensors Blood Side Pa, Pv, Pbe And Pbs

    LL Manual Calibration key in the Manual Test and Calibration menu. Activate the Pressure Sensors Blood Side menu by pressing on the line Calibration 2.1 Pressure Sensors Blood Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 285
    Build up a pressure of -100 mmHg (+10 mmHg) with a syringe and clamp the tube. -100 mmHg Press key. Wait until digits are stable. Enter the stable value from the pressure measurement instrument as reference value. Confirm with the key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 286
    Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Confirm and store completed calibration with key. 17. Remove pressure measurement instrument, syringe and tubing. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 287: Heparin Pump

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 118 SW 9.xx 4.8.3.4 Heparin Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 288: Calibration Degassing Pressure Sensor Pe

    You can calibrate the degassing pressure sensor PE. Build up the negative pressure in the degassing circuit with a syringe. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* VBKO* TSHE VVB* VBKS RVBO* Fig.: Degassing Circuit and PE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 289
    TSHE Fig.: Water Sub-Rack with Loop-In for Manometer at PE and Tubing Fig.: Water Sub-Rack with Tubing Clamp between FEP and EP (front Clamp at VEB (rear view) view) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 290
    Manual Test and Calibration menu. 2. Activate the Pressure Sensors (Water Side) menu by pressing on the line Calibration 2.3 Pressure Sensors (Water Side). The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 291
    PE should show a of approx. < -35 mmHg/-50 mbar • Close VEB and increase speed to 2,100 rpm • PE should show a value of approx. < -500 mmHg/-670 mbar B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 292: Calibration Degassing Pressure Sensor Pe For Machines At

    15. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 16. Remove calibration devices. 17. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 293: Calibration Dialysate Outlet Pressure Sensor Pda

    VDA, it must be closed or the tubing must be clamped manually. FBK2 TSDE DF Filter PEV* Z.D. Dialyser V.D. LAFS FBK2 BL PDA SBS1 RVFPA SBS2 Fig.: Dialysate Circuit and PDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 294
    Conversion Factor mmHg to mbar Conversion Factor = 1.3332 Example: –400 mmHg approx. –530 mbar 0 mmHg 0 mbar +400 mmHg approx. +530 mbar B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 295
    LL Manual Calibration key in the Manual Test and Calibration menu. 2. Activate the Pressure Sensors Water Side menu by pressing on the line Calibration 2.3 Pressure Sensors Water Side. The following menu is displayed. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 296
    21. Confirm and store complete calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 22. Remove calibration devices. 23. Reconnect tubings. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 297: Calibration Degassing Temperature Sensor Tse/Heater Inlet Temperature Sensor Tshe

    KP/BICP TSHE Keep tubings as short as possible TSHE RVDA RVDA Fig.: Original Tubing Water Sub-Rack for TSE/TSHE Fig.: Water Sub-Rack with TSHE in Line with TSE for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 298
    C) is displayed at the temperature measurement instrument. Wait until temperature is stable, i.e. 0.1 C for at least 10 s. Press key. 10. Enter the stable value from the temperature measurement instrument as reference value. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 299
    22. Stop degassing pump EP and flow pump FPE, and disconnect adapter set. 23. Reassemble the heater inlet temperature sensor TSHE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSHE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 300: Calibration Temperature Sensors Tsbic/Tsd/Tsd-S And Tsde

    Measurement Instrument ml/min RVFPE ENDLF-S TSBIC blue mS/cm BICLF ENDLF TSD-S TSDE Measurement Sensor BICP blue FBK1 VEBK1/2 Fig.: Dialysate Preparation with Loop-In Point for TSBIC/TSD/TSD-S and TSDE Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 301
    FBK1 Measurement Sensor blue TSDE Fig.: Tubing Rinsing Bridge with Removed TSDE Fig.: UF Sub-Rack with TSDE in Line with TSBIC/TSD/TSD-S (on DF Sub-Rack) for Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 302
    Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. Press key for a combined calibration of TSBIC, TSD/TSD-S and TSDE (The circulation valve VZ stays closed). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 303
    4.8.3.1 General Calibration Information). 21. Reassemble the heater inlet temperature sensor TSDE to the original position (see previous page). Note: Pay attention to correct flow direction of temperature sensor TSDE. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 304: Calibration Conductivity Sensors Bicarbonate Biclf And End Conductivity Endlf/Endlf_S

    If acid concentrates are not available the following B. Braun acetate concentrates can be used as an alternative: e.g. SW44 and SW 174 If acid concentrate from B. Braun is not available acid concentrate from a second source can be used. The composition is listed in table 1. Composition of Calibration Acid concentrate is diluted with deionised water with a ratio of 1:4 (or 1:4.5…

  • Page 305
    FBK1 Measurement VABK2 Sensor VDEBK2 FBK1 blue VEBK2 Fig.: Original Tubing UF Sub-Rack with FBK1 Fig.: UF Sub-Rack with Loop-In Point for BICLF, ENDLF and ENDLF-S Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 306
    Connect measurement instrument between FPE and balance chamber. Connect dialyser couplings to flow meter. Increase slowly the speed of FPE, until a flow of 750 ml/min ±50 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 307
    14. Press key. Wait until digits are stable. 15. Enter the stable value from the measurement instrument as reference value. 16. Confirm with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 308
    Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). 27. Stop FPE and BICP and remove measurement instrument. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 309: Calibration Blood Leak Detector

    SBS2 blue Fig.: Rinsing Bridge with Blood Leak Detector BL TSDE LAFS FBK2 SBS1 blue SBS2 Fig.: Rinsing Bridge with Red Coupling Immerged in Reference Solution for BL Calibration B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 310
    Part. After the initialisation phase rinse the blood leak detector approx. 1 min and then perform a leakage test of the system in menu 1.19 Leak Test. Chemical Disinfection with Citric Acid 50% A decalcification must be performed before initial calibration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 311: Blood Leak Detector Limit

    Please Note The same blood leak detector is used for both settings: • The calibration procedure is identical, but with different tolerance ranges • The test procedure is identical B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 312: Check Blood Leak Detector

    LAMP TEST is switched back to OFF. The calculated value must be in the following limits (to the first decimal digit/round off): red LED / green LED 1.15 — 1.44 B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 313: Calibration Blood Leak Detector

    The displayed actual blood concentration value must be in the following limit, to ensure a sufficient resolution: Actual Status Controller > 200 ‰ Blood Concentration [ Confirm and store completed calibration procedure with key. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 314: Actual Value Check Blood Leak Detector

    The blood concentration values for the controller and supervisor must be in the following limits: Actual Status Actual Status Controller Supervisor Blood Concentration 0.90 to 1.1 0.90 to 1.1 [‰] B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 315: Calibration Of Delivery Rate Ufp, Bicp And Kp

    • Balance or graduated cylinder • Tubing clamp Fill graduated cylinder with 200 ml water. Remove respective concentrate suction rod from rinsing chamber and immerge into graduated cylinder. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 316: Calibration Uf Pump With Balance

    12. Confirm and store completed calibration procedure with key. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information see 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 317: Position Setting For Bicp, Kp And Ufp

    It is recommended to calibrate the UFP with the procedure described (with balance) because the accuracy of the stroke value might be insufficient for a proper UF control of the system. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 318: Calibration Safety Air Detector Sad

    The calibration procedure is completed. 5. Press the key to store the calibration. Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed information 4.8.3.1 General Calibration Information). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 319: Sad Test After Calibration

    Remove the tubing from the SAD. Interrupt the light path in the SAD with a piece of paper. The following settings must be displayed: RDV: NO BLOOD (Controller/Supervisor) changes to RDV: BLOOD (Controller/Supervisor) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 320: Calibration Biclf And Endlf Ratio

    Production Report, LLC Options, Holder for BIC Cartridge. You can determine the limit of the ENDLF ratio (measured at the end conductivity sensor ENDLF). B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 321
    Connect the respective concentrate (bicarbonate/acid) after the request. Select service overview to display the current RATIO values. Wait till stable ratio values are reached. Use the mean value if there are slight deviations. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 322
    1. Enter 35 in Desired Setting Controller for the BICLF Cart Ratio. 2. Save the entered value with the Calibration key. 3. Check the bicarbonate ratio value in therapy. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 323: Calibration Load Cell

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 154 SW 9.xx 4.8.3.14 Calibration Load Cell Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 324: Calibration Pv Alarm Window, Pa Low Limit

    The supervisor uses a tolerance of 10 mmHg for monitoring the PV limit values to prevent double alarms. The set PA Absolute Low Limit value (-200 mmHg) is used as limit in therapy PA Absolute Low Limit and TSM service program/system configuration. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 325: Calibration Flow Rate Of Bpa And Bpv/Osp

    Pump Head Selection. The following menu is displayed. You can select the pump head roller type used in the blood pumps. • Pump head 8 x 12 • Pump head 7 x 10 (option) B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 326: Infusion Density Substitution Pump

    4 . T S M S e r v i c e P r o g r a m 1 / 2 0 1 0 4 — 157 SW 9.xx 4.8.3.17 Infusion Density Substitution Pump Not applicable. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 327: Stroke Value Hdf Online Substitution Pump Osp

    2.15 HDF Online Substitution Pump OSP. The following menu is displayed. The stroke value must be set to 8300 (for pump head 8 x 12). Stroke Value HDF Online Substitution Pump OSP B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 328: Level Regulation

    The settings are (see default table in chapter 3): • 60 % for the slow mode • 100 % for the fast mode The default settings should not be changed, unless instructed otherwise. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 329: Calibration Inlet Flow Pump Nonreturn Valve Rvfpe

    You can calibrate the inlet flow pump nonreturn valve RVFPE. Fig.: Calibration RVFPE/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVFPE Fig.: DF Sub-Rack with RVFPE B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 330
    Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPE. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 331
    VDABK1 MSBK1 VDEBK2 VABK2 VEBK2 VDABK2 FBK2 MSBK2 RVDA Fig.: Calibration RVDA/Loop-In Point Loop-In Point for Calibration FBK1 Fig.: UF Sub-Rack with FBK1 RVDA Fig.: Water Sub-Rack with RVDA B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 332
    • 300 mmHg ±37 mmHg • 400 mbar ±50 mbar HDF Online • 375 mmHg ±37 mmHg • 500 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 333: Calibration Outlet Flow Pump Nonreturn Valve Rvfpa

    Set a pressure of 975 mmHg (1300 mbar) with the adjustment screw of RVFPA. • 975 mmHg ±37 mmHg • 1300 mbar ±50 mbar Tighten lock nut again. 10. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 334: Calibration Dialyser Inlet Throttle Valve Dde

    (30 to 40 °C). Thus uncontrolled pressure conditions are prevented. Select Test 1.20 Water Part (Overview) in TSM. Increase slowly the speed of FPA, until a flow of 500 ml/min is reached. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 335
    Set a pressure of 300 mmHg/400 mbar with the adjustment screw of DDE. • 300 mmHg ±37 mmHg • 400 mbar ±50 mbar Tighten lock nut again. Remove manometer and flow meter. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 336: Calibration Pressure Reducer Valve Dmv

    You can calibrate the pressure reducer valve DMV. RVDA SKBS VVBE DBK* TSBIC VB WT VBICP* TSHE VBKO* VVB* VBKS RVBO* Fig.: Calibration DMV VVBE Locking Screw for Manometer Fig.: Water Sub-Rack with DMV B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 337
    0.9 ±0.1 bar with the adjustment knob. Close water tap. Press down the adjustment knob to lock pressure reducer valve DMV. 10. Switch off Dialog. 11. Disassemble manometer. 12. Reassemble locking screw. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 338
    Note The 1.4 mm template fits through the gap. Check wit the 1.5 mm template. Note The 1.5 mm template must not fit through the gap. Reassemble cap C. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 339: Arterial Tubing Clamp Saka/Venous Tubing Clamp

    SAKV-SG. • O-ring 7 x 1.5 art. no. 34570624 Check the o-rings C on the V-stopper/stopper D Replace if o-rings show any damages (tears/ruptures/ brittleness). Close front door. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 340: Setting Servomotor For Disinfection Valve Vd

    Do not pull off the bow G from the control disc H, because its position is set ex works. The bow G on the control disc H must be seated in the groove F of the eccentric axle. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 341
    D Light barrier E Bevel on eccentric axle F groove G Bow (only partly visible, is seated in the eccentric groove) H Control disc Rinsing bridge Draw-Up of Disinfectant B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 342: Production Report

    • The activation of the HCT Crit-Line option enables a window in therapy. This window allows the data exchange and display of measurement values to a connected HemaMetrics™ Crit-Line device. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 343
    • Basic Board BB • Analog Board AB • Power Board Motors BPM • Power Board Valves BPV • HDF Online Board HOB • Digital Board DB B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…
  • Page 344: Low Level Options

    If the option is assembled or the option is retrofitted, it must be selected and WAN = Wide Area Network saved to activate the option. BSL = Bedside Link DBI = Dialog Bedside Link Interface B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 345: Service Reports

    The Operation Mode Report can be selected in the Service Reports menu. Operation Mode Report The switching from therapy mode to TSM mode and back to the therapy mode is documented with date and time. B. Braun Avitum AG BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd…

  • Page 346
    5. Technical Safety Inspection and Preventive Maintenance 1 / 2 0 1 0 5 — 1 SW 9.xx This document is the property of B. Braun Avitum AG with all rights reserved. Copyright Only trained personnel must service the Dialog+, i.e. repair, maintenance, software installation, Service firmware update, retrofitting and commissioning of the Dialog+.
  • Page 347
    Measurement Circuits for the Measurement of the Electrical Safety According to EN 62353/60601-1 Inspection Protocol for Automatic Blood Pressure Measurement ABPM together with the Dialog+ 5.2.1 Execution Procedure for Measurement Inspection ABPM B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 348
    ……………………………. Operating Hours: ……..h Inventory No.: ………….. SW Version: ……….Manufacturer: B. Braun Avitum AG 34209 Melsungen, Germany Check List (S = Technical Safety Inspection Points; M = Preventive Maintenance Points) NO YES OK Note: Text in { } brackets is information for the execution of the check list! Technical Safety Inspection …
  • Page 349
    Single-Needle Pressure PBS: (permissible. tolerance ±10 [mmHg]) — Comparison measurement (at approx. 0): Measurement instrument: ……Display on machine: …… [mmHg] — Comparison measurement (at approx. +400): Measurement instrument: ……Display on machine: …… [mmHg] B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 350
    {The machine must be in therapy mode. Disconnect mains plug: The buzzer from the SMPS-MC gives three signalling tones as an indication that the machine is running in battery mode.} M 5.10 Disinfection: — Perform by responsible organisation {user} {inform responsible organisation} — Performed by service technician B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 351
    As a supplement this kit includes all parts recommended for exchange during a {34560686 Maintenance Kit/Supplement HDF Online} preventive maintenance of a HDF online machine.} B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 352
    The test current must be measured in both directions. Fig. 1: Protective Earth Resistance Equipment Leakage Current: • Differential Measurement L(N) N(L) Fig. 2: Equipment Leakage Current Patient Leakage Current Fig. 3: Patient Leakage Current B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 353
    Dialysis Machine Dialog REF {Type/Typ}: SN {Serien-No./Nr.} ………………ABPM Module M2009/2010 Automatic Blood Pressure Measurement ABPM B. Braun Avitum AG, 34209 Melsungen, Germany Manufacturer Address: Responsible Organisation (User) …………………………………………………………………………………………Visual Inspection — Tight seat of all connectors : ———————————————— OK — No damages which could affect the function or the safety (incl.
  • Page 354
    Fig.: Inspection Set-up for Compliance of Error Limits of the 6. Decrease pressure according to test points and enter values in Measurement Display table. 7. Enter deviations in table. B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 355
    : —- max. 10 s Pressure reduction from >260 mmHg to <15 mmHg Pressure Tube 50 ml 50 ml Syringe Stopcock System Fig.: Inspection Set-up for Rapid Exhaust B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 5-1_1-2010.doc/pdf <110301> yymmdd…
  • Page 356
    6.3.4 Tubing Diagram Water Sub-Rack 6-16 6.3.5 Tubing Diagram BIC Cartridge Holder 6-17 6.3.6 Tubing Diagram Housing HDF Online 6-18 6.3.7 Tubing Diagram UF Sub-Rack HDF Online 6-19 B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 357: Flow Diagrams

    Filter balance chamber (outlet) Filter from dialysate Filter bicarbonate Filter concentrate KtV-UV KtV-UV sensor (option Adimea) Water block with degassing chamber, upline tank, heat exchanger, level sensor, double-stage heater B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 358
    TSD-S Dialysate temperature sensor supervisor Degassing temperature sensor Heater temperature sensor TSHE Heater inlet temperature sensor TSDE Dialyser inlet temperature sensor Ultrafiltration pump RVUFP Ultrafiltration pump nonreturn valve B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 359: Flow Diagram Dialog

    Central Bicarbonate Central Concentrate Supply Supply (Unit Rear) Disinfectant BIC Concentrate Acid Concentrate Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 360: Flow Diagram Dialog+ Bic/Df Option

    Supply Supply (Unit Rear) BKUS Disinfectant BIC Concentrate Acid Concentrate RVBU Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial Options B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 361: Flow Diagram Dialog+ Hdf-Online

    Supply Supply (Unit Rear) BKUS Disinfectant RVBU BIC Concentrate Acid Concentrate Water Dialysate Flow 2 Blood Venous Bicarbonate Dialysate Disinfectant Concentrate Housing Dialysate Flow 1 Blood Arterial Options B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 362: Wiring Diagrams Dialog

    Cable connection from basic board to front door: — Double pump — HDF online Cable connection from basic board to front door: — Double pump — HDF online B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 363: Wiring Diagram Front Door Single Pump

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.2 Wiring Diagram Front Door Single Pump Motor Heparin pump DZ/DR, Cover SAKV B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 364: Wiring Diagram Front Door Double Pump

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 SW 9.xx 6.2.3 Wiring Diagram Front Door Double Pump Motor Heparin pump PBE, PBS DZ/DR, Cover SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 365: Wiring Diagram Front Door Hdf Online

    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s 1 / 2 0 1 0 6-10 SW 9.xx 6.2.4 Wiring Diagram Front Door HDF Online Heparin pump Motor PBE, PBS SAKV SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 366
    6 . F l o w , W i r i n g a n d T u b i n g D i a g r a m s SW9.xx 6.2.5 Wiring Diagram Housing B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…
  • Page 367: Colour Coding Solenoid Valves

    TSHE NSVB Water Block Option BIC VVBE VBKS VBKO VBICP bk black bl blue br brown gn green gr grey rd red vi violet wh white ye yellow B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 368: Tubing Diagrams

    Silicone Tubing Silicone Tubing NSVB 2.9×3 SNP1.5 SNP4 Cable SNP1.5 Silicone Tubing 2.9×3 VVBE UF Sub-Rack RVDA blue Silicone Tubing 6×3 Rinsing Chambers Concentrate Suction Rods Water Inlet B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 369: Tubing Diagram Uf Sub-Rack

    Disposable Filter SNP6 Silicone SNP6 (2x) FBK2 Tubing Silicone Tubing 6×3 SNP6 (3x) Y-Tubing Connector 8 mm VABK2 VDEBK2 FBK1 VEBK2 SNP6 SNP6 Silicone SNP6 Disposable Tubing 6×3 Filter B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 370: Tubing Diagram Df Sub-Rack

    Silicone Silicone from Tubing Tubing WA Block 2.9×3 to BIC Suction Rod T Reduction on UF Sub-Rack to Concentrate to LA (UF) Suction Rod to TSHE/VEB (WA Block) B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 371: Tubing Diagram Water Sub-Rack

    Silicone Tubing NSVB Water Water Water Inlet Inlet Inlet from Piston Pump Rinsing WA-Block from VLA Silicone to DF Tubing Processing Silicone Tubing SNP4 TSHE RVDA Silicone Tubing B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 372: Tubing Diagram Bic Cartridge Holder

    SNP6 SNP4 SNP6 Reduction Piece BKUS Silicone Tubing Silicone Tubing 5×3 2.9×3 SNP6 Silicone Tubing 5×3 VBKO VBKS SNP1.5 VBICP Water Block Silicone Tubing 5×3 RVDA Water Sub-Rack B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 373: Tubing Diagram Housing Hdf Online

    Clamp SNP6 13.3 Silicone Silicone Tubing SNP1.5 (2x) Tubing SNP6 6×1.75 Silicone VBKS Tubing Silicone 2.9×3 Tubing Water Water Sub-Rack Inlet BKUS blue Rinsing Chambers Concentrate Suction Rods B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 374: Tubing Diagram Uf Sub-Rack Hdf Online

    Silicone Tubing Silicone Silicone VDABK2 SNP6 Tubing VSAA Tubing Silicone VABK2 Tubing FBK1 VDEBK2 Silicone Tubing 6×1.75 SNP6 VEBK2 (3x) SNP6 Silicone Y-Tubing Tubing 6×1.75 Connector 8 mm B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd…

  • Page 375
    (Material Return Form) to B. Braun Avitum AG via the representative of the respective country for the ability to supply refurbished spare parts. The condition and the version of the defective spare parts must allow a qualified refurbishment.
  • Page 376
    …………………………….Disposal and Taking Back of Spare Parts Dispose spare parts (e.g. boards or batteries) according to local disposal guidelines or send back to B. Braun Avitum AG free of charge. Cleaning and Disinfection of Spare Parts Used in Dialog Machines All spare parts sent by the customer to the B.
  • Page 377
    Switch Mode Power Supply Microcontroller SMPS-MC 7-29 7.7.3.1 Compatibility Table SMPS-MC 7-30 Front Door 7-31 7.8.1 Manometer Connection 7-33 7.8.2 Heparin Pump 7-33 7.8.3 Blood Pump 7-34 7.8.4 Safety Air Detector SAD 7-35 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 378
    7-61 7.18.4 Board HDF Online 7-62 7.18.5 DF/HDF Filter Holder HDF Online 7-63 7.19 Automatic Blood Pressure Measurement ABPM Module M2009 7-64 7.19.1 Compatibility Table ABPM Module M2009 7-65 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 379
    SW 9.xx Page 7.20 Emergency Power Supply (Option) 7-66 7.20.1 Battery Sub-Rack 7-67  7-68 7.21 Nexadia -/WAN-BSL (Option) 7.22 Card Reader (Option) 7-69 7.23 Adimea Option (Option) 7-70 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 380: Sub-Racks

    7.6 for single parts UF sub-rack see paragraph 7.5 for single parts DF sub-rack see paragraph 7.4 for single parts Water sub-rack see paragraph 7.3 for single parts B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 381: Solenoid Valves

    C) without loosening the self-locking nut A. The procedure is only permissible if the coil can be turned without much iv 1 x Screw 40×10 effort. v 1 x Washer 4.2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 382
    TSHE 8 34570730 1 Housing complete for temperature sensor 9 3451727B 1 Throttle 0.4 bar 10 3451664B für HDF-Online 1 Throttle 1.3 bar 11 34517782 5 Disposable filter RVDA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 383: Water Sub-Rack

    12 34516069 1 Throttle 0 — 4 bar 1 Tubing connector ∅ 7 x 1/8″ 13 34517243 stainless steel 14 34560289 connector/connector (DPP) 1 Solenoid valve 2/2 way VVBE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 384: Gear Pump For Ep

    1 Motor for gear pump (SW 1.7.4) Micropump Version 2 for EP only 1b 34560319 1 Gear pump (Micropump version 2) (see A in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 385: Df Sub-Rack

    7.4.1 — DF block see separate figure for single parts, paragraph 7.4.1 see separate figure for — Piston pumps single parts, paragraph 7.4.2 BICLF BICP ENDLF TSDS RVFPE B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 386: Df Block

    1 Druckregler 1,3 bar 8 34570683 10 O-Ring 4,47 x 1,78 2 Tubing connector ∅ 3.5 x 1/8″ 9 3477103A plastic 5 Tubing connector ∅ 8 x 1/8″ 10 3451916A plastic B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 387
    4e 34771085 5 Seal BICP 5a 34570373 10 Plug (long) 5b 3451614B without connector 1 Temperature sensor ENDLF 5c 3477104A 10 O-ring 8.3 x 2.4 BICLF TSBIC 4b,10 3 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 388: Gear Pump For Fpe/Fpa

    1 Motor for gear pump (SW 1.7.4) 1b 34515330 1 Gear pump (Micropump version 1) Micropump Version 2 for FPA/FPE only (see B in figure) with magnet, without motor B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 389: Piston Pumps

    1 Motor for piston pump 5 34560572 1 Bell joint for piston pump 6 34560602 1 Pump body for piston pump — 7703848 required for motor assembly 1 Assembly jig (motor piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 390: Uf Sub-Rack

    Pos. Art. No. Qty. Description Comments UF Sub-Rack — Air separator complete see separate figure — Balance chamber 2 see separate figure RVFPA VDABK2 VABK1 VDEBK2 VDEBK1 BK1/2 VABK2 VDABK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 391: Balance Chamber

    2 Membrane for balance chamber 20 Tubing clamp SNP-1.5 ∅ 7-8 mm VEBK2 plastic — 34570195 20 Tubing clamp SNP-4 ∅ 10-11 mm — 34570209 plastic VDEBK2 VABK1 VDEBK1 VABK2 VDABK1 VEBK1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 392: Air Separator Assembly Group

    1 Throttle 0.4 bar 5/6 34570683 10 O-ring 4.47 x 1.78 6 3451664B 1 Throttle 1.3 bar 5 Tubing connector ∅ 8 x 1/8″ 7 3451916A plastic RVFPA RVFPA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 393: Top Level Sub-Rack

    10 34561277 1 Temperature switch assemble temperature switch on TL sub-rack Basic Board with heat sink paste Analog Board Online Board Motherboard Power Board Valves Online Board Basic Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 394: Lx800 Motherboard

    1 Compact flash card 1 GB The SW for the machine has to be installed again after replacing the CFC. 2 34560610 1 LX800 Motherboard 3 34519327 1 Battery BR2032, 3V Motherboard B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 395: Settings For Tft Display/Cfc

    1 / 2 0 1 0 7 — 21 SW 9.xx Pos. Art. No. Qty. Description Comments 7.6.2 Settings for TFT Display/CFC LX800 Motherboard JLVDSPWR TFT Power 3.3 V: BoeHydis/Samsung JCFSET1 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 396: Rear Door

    1 Fan 92 x 92 mm with connector 2 34570284 5 Dust filter 3 3451059A 1 Mains switch 3 34770003 5 Protection cap for mains switch — Mains cord see separate figure (paragraph 7.7.1) SMPS-MC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 397
    4 34518029 1 Strain relief 5 3451441A 1 Tubing holder/fan cover compatible to crank 5a/5b 5a 34518355 1 Crank for blood pump 5b 34560157 1 Crank for blood pump B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 398: Mains Cord

    1 Mains cord (Brazil) mains cord with mains plug 2 34561145 mains cord with mains plug 1 Mains cord (Cekon) 3 34561226 1 Mains cord (China) mains cord with mains plug B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 399
    1 Mains cord (Israel) 5 34561170 1 Mains cord (Philipines) mains cord with mains plug 6 34561153 mains cord with mains plug 1 Mains cord (Schuko) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 400
    1 Mains cord (South Africa/India) 8 34561234 mains cord with mains plug 1 Mains cord (Thailand) 9 34561188 1 Mains cord (UK) mains cord with mains plug B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 401: Connection Mains Cord To Smps-Mc

    Central Earth F5B F5A Cable Gland with Lock Nut Rear Door Cable Ties N PE PE LH1LH2 Ferrite Ferrite Brown Green/Yellow Blue Mains Cord with Mains Plug Mains Cord Ferrite B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 402: Compatibility Table For Boards

    The digital board is always delivered with the latest LLS/LLC software versions! 34560734 Analog Board AB (38108641) 34560866 HDF Online Board HOB (38108647) 3451665A MSBK Board (3810671X) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 403
    10 Fuse T3.15 A (TR5) LH1 LH2 N PE PE K3/K4 34514775 2 Relay (12 V) for 24 V C114/C115 34519912 2 Capacitor 1F H104 34560580 1 Buzzer 34560416 10 Plug for buzzer B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 404: Switch Mode Power Supply Microcontroller Smps-Mc

    LH1 LH2 N PE PE 7.7.3.1 Compatibility Table SMPS-MC SMPS-MC Type Plate Version Number SMPS-MC SW Version Dialog+ SW 9.xx (TN/Item-No) (Ver/SW) XXXXXX.5 Rev. 1.15 34560220 (version 5) ex works B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 405
    1 Metal disc ∅ 10 mm for cover 9 34560033 9 34519823 Hinge pin with interior thread hinge pin 9 34516670 1 Seal for blood pump SAKV-SG SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 406: Front Door

    17 34560831 see separate figure (paragraph 7.8.5) 1 Valve block level regulation (double pump) 18 34560823 1 Diaphragm pump with assembly elbow 3 mm 4 mm 4 mm SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 407: Manometer Connection

    5 34560785 1 Housing with seal 5a 34514732 1 Syringe holder 5b 34570640 5 Dummy plug 5c 34570667 1 Clamp 5d 34570659 1 Unlocking lever set 5a 5b 5c 5d B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 408: Blood Pump

    1 Reed sensor with contacts and o-ring 3.68 x 1.78 — 34516670 1 Seal for blood pump ^^^^^^^^^^^^^^^^^^^^^^^^ 100 g tube (to lubricate the shaft of the blood pump) — 7703708 1 Grease SKD 4002 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 409: Safety Air Detector Sad

    SAD with board; do not tamper, exchange only complete red detector If cover for SAD is present: a ball head allen key 2.5/3 is required for bottom screw behind rotary magnet. SAKA SAKV-SG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 410: Level Regulation Module

    27 34565205 5m PUR tubing 3.6 x 1.1 (5 m) for manometer connection (filter) 26 27 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 411
    1 Swivel caster 8 3451490A 1 Swivel caster with brake 9 34560548 1 Card-Reader-Board see 7.22 for details 10 3451487C 1 ABPM module M2009 with filter see 7.19 for details B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 412: Basic Housing

    3m Silicone tubing 8 x 3.2 / red (3 m), external high temperature tubing with two single ear clamps 17.5 — 39224538 1 Single ear clamp 17.5 for external high temperature tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 413: Concentrate Suction Rods

    1 Male connector (red) with o-rings 8 34570004 5 O-ring 9.3 x 2.4 9 34565280 3m Silicone tubing 3.2 x 1.6 (3 m) 10 3451728A 1 PE tube 4 x 1 with silicone tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 414: Rinsing Chambers For Concentrate Suction Rods

    3.68 x 1.78 5 3451476A 1 Male connector (red) with o-rings 6 3451475A 1 Male connector (blue) with o-rings 7 34517782 5 Disposable filter 3 4 5 6 7 grey B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 415
    1 PE tube 4 x 1 7 34565280 3m Silicone tubing 3.2 x 1.6 (3 m) 8 34570004 5 O-ring 9.3 x 2.4 9 3451465A 2 Suction nipple with o-rings 9.3 x 2.4 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 416: Central Concentrate Supply (Option)

    Comments Central Concentrate Supply (Option) 3 34565248 10m Silicone tubing 5 x 1.5 / blue (10 m) 5 34565230 10m Silicone tubing 5 x 1.5 / red (10 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 417: Rinsing Bridge 4

    13 34511792 1 Tubing coupling 14 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 15,14,13 12,11,8 15 34771255 20 O-ring 14 x 1.5 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 418
    10 O-ring 5.0 x 1.6 for eccecntric shaft 6 34560858 1 Eccecntric roller 7 34560840 1 Eccecntric shaft 8 3451659D 1 Servomotor with board for VD, servomotor and board must be exchanged together B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 419
    17 34570730 1 Housing complete for temperature sensor 18 34773762 10 O-ring 9.25 x 1.78 19 34511792 1 Tubing coupling 20 34771255 20 O-ring 14 x 1.5 20,19 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 420: Tft Monitor

    2 34570691 2 OSD cover complete with seal and screws 3 34561285 1 Membrane 4 34561366 1 Membrane for Adimea option Dialog start stop stop Dialog start stop stop B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 421
    1 TFT display 15“ (Boe Hydis or Samsung) 9 3451460A 1 Touch controller board TCB version 3 — 34560912 1 Ribbon cable 12 pin FPB — OSD — 34560971 1 Ribbon cable FPB — BIB B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 422
    * Balance (e.g. Pocket Balance) For example manufacturer: for details contact manufacturer Kern & Sohn GmbH directly Internet: www:kern-sohn.com * not in the product range of B. Braun Avitum AG B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 423: Test And Calibration Equipment And Tools

    1 Male to male connector couplings for the measurement of the patient leakage current — 770284A 1 DCI cable, external required for motor assembly — 7703848 1 Assembly jig (motor for piston pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 424: Service Kits

    Comments 7.12 Service Kits — 3451897A 1 Repair kit balance chamber for the complete replacement of both membranes, incl. all o-rings and assembly parts (self-locking nuts, screws, springs etc.) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 425: Software

    1 CD Remote Bedside Link List ReBeLL, for network version 1.5.13 — 7703562 for WAN-BSL only 1 CD WAN-BSL, firmware version 1.0.4.8 — 7703864 1 Field service utility FSU, version 2.00 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 426
    Software 7.58 (Linux) Software 7.59 (Linux) *1.9 Software-Update 7.59 (Linux) #1.01/ USB Stick with TLC/LLS SW only 1.04 if SW 7.52 is present Software ≥ 8.20 (Linux) Software 9.02 (Linux) 1.04 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 427: Miscellaneous

    1 PVC tubing 10 x 3 / black, (3 m), with a single ear clamp 19.5 and a tubing clamp 12-20 mm — 39234126 1 Tubing clamp 12-20 mm — 34565345 1m Silicone tubing 12 x 2.5 (1 m) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 428
    — 7703708 1 Grease SKD 4002 100 g tube (to lubricate the shaft of the blood pump) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 429: Staff Call (Option)

    1 / 2 0 1 0 7 — 55 SW 9.xx Pos. Art. No. Qty. Description Comments 7.16 Staff Call (Option) 1 34517685 1 Staff call board STAT (Default) DYN AUS B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 430: Bicarbonate Cartridge Holder (Option)

    10 34514449 1 Cover for bottom housing 11 34517138 1 Reed sensor with contacts and o-ring bottom BIC cartridge holder 3.68 x 1.78 — 34517782 5 Disposable filter filter FBIC B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 431: Solenoid Valves And Throttle

    5 O-ring 11.1 x 1.6 4 34771050 10 Pressure spring 500 mmHg 5 34771352 3 Centre disc 5 Tubing connector ∅ 8 x 1/8″ 6 3451916A plastic VBKS VBKO VBICP B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 432: Hdf Online (Option)

    7 — 58 SW 9.xx Pos. Art. No. Qty. Description Comments 7.18 HDF Online (Option) — Substitution port see separate figure — UF sub-rack see separate figure PBS/SN SAKV-SG SAKA B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 433: Substitution Port

    1 Substitution connector 5 3477106A 5 O-ring 6.4 x 1.3 6 34517138 1 Reed sensor with contacts and o-ring 3.68 x 1.78 A or A+B (A=Quadring; B=Spacer) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 434: Uf Sub-Rack Hdf Online

    AKM only, connector/connector RVFPA VSAA VSAE 4 3451902C manuf. AKM only, connector/flange 1 Solenoid valve 2/2 way 4 34770852 5 O-ring 11.1 x 1.6 VDABK2 VABK1 VDEBK2 VDEBK1 BK1/2 VDABK1 VABK2 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 435: Valve Block Hdf Online

    7 — 61 SW 9.xx Pos. Art. No. Qty. Description Comments 7.18.3 Valve Block HDF Online 1 34560661 1 Hydrophobic filter with tubing adapter 2 3451900A 1 Solenoid valve 3/2 way B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 436: Board Hdf Online

    Pos. Art. No. Qty. Description Comments 7.18.4 Board HDF Online 1 34560866 1 HDF online board Motherboard Online Board Power Board Valves Digital Board Power Board Motors Basic Board Analog Board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 437: Df/Hdf Filter Holder Hdf Online

    Blue 3 3451445A 1 Dialyser coupling (red) 3 3451446A 1 Dialyser coupling (blue) 3 34990445 5 O-ring 12.37 x 2.62 3 34773550 20 Membrane for dialyser coupling Blue Blue B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 438: Automatic Blood Pressure Measurement Abpm Module M2009

    1 34560955 1 Filter (ABPM module M2009) 2 3451487C 1 ABPM module M2009 firmware version of ABPM module on sticker: Version Y (FW):X.X with filter Version YX.X Version YX.X B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 439: Compatibility Table Abpm Module M2009

    Description Comments 7.19.1 Compatibility Table ABPM Module M2009 Position of DIL Switch for DIL Switch: Default LX800 (ex Works) LX800 Motherboard DIL Switch DIL Switch Version YX.X LX800 (default) B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…

  • Page 440
    Sulfric acid in the eyes may cause loss of eyesight and acid on the skin will cause burns. Pay attention to the included documentation of the battery supplier! see separate figure — Battery sub-rack B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 441
    F301 H302 F302 H304 F304 H601 F601 brown blue brown H602 F602 H402 F402 F403 H403 blue green/yellow green/yellow brown blue GNDAKKU XP3.2 brown brown XP3.1 +24VAKKU P9 P7 B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 442
    1 / 2 0 1 0 7 — 68 SW 9.xx Pos. Art. No. Qty. Description Comments  7.21 Nexadia -BSL/WAN-BSL (Option) 1 34514554 1 Bedside link board (BSL board) without firmware B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 443
    7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 — 69 SW 9.xx Art. No. Qty. Description 7.22 Card Reader (Option) 1 34560548 1 Card reader board B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 444
    7 . S p a r e P a r t s L i s t 1 / 2 0 1 0 7 — 70 SW 9.xx Art. No. Qty. Description 7.23 Adimea Option 1 3456063A 1 Kt/V-UV board with housing and tubing B. Braun Avitum AG BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd…
  • Page 445
    8 — 1 SW 9.xx ESD/EMC Information 8.1.1 Electrostatic Discharge ESD 8.1.2 Electromagnetic Compatibility EMC Technical Information TI Assembly Instructions AI Field Service Information FSI Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 446
    In general the mats are made of sturdy two layer material and have reinforced Mat Material edges and corners. The work surface is static dissipative (> 100 MΩ), the other side is conductive (> 100 kΩ). B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 447
    Unprotected electronic parts (e.g. wrapped in normal plastic film or parts padded with foam material) are assumed to be damaged by electrostatic discharge and will be scrapped. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 448
    N, connected to a one or three phase public power supply with a constant current load of the power supply ≥ 100 A per phase, then it is also valid for the Dialog machine with the switch mode power supply microcontroller SMPS-MC. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 449
    The power frequency magnetic field should be measured in the intended installation location to magnetic field acc. to IEC 61000- assure that it is sufficiently low. NOTE U is the a.c. mains voltage prior to application of the test level. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 450
    Dialog machine. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 451
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 7 SW 9.xx Technical Information TI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 452
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 8 SW 9.xx Assembly Instructions AI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 453
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 9 SW 9.xx Field Service Information FSI B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 454
    Dialog 8 . A p p e n d i x 1 / 2 0 1 0 8 — 10 SW 9.xx Instruction Leaflets IL B. Braun Avitum AG BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd…
  • Page 455
    2011-03-01 4-1-1_1/2010 Revision Chapter 4 TSM Service Program 2011-03-01 4-2-1_1/2010 4-3-1_1/2010 5-1_1/2010 Revision Chapter 5 Technical Safety Inpsection and Preventive Maintenance 2011-03-01 9-1_1/2010 Revision Chapter 9 Edition/Updates 2011-03-01 B. Braun Avitum AG BA-TE-DE08W M.KAY — Dialog+ SW9xx_SM_Chapter 9-1_1-2010.doc/pdf <110301> yymmdd…

(Ocr-Read Summary of Contents of some pages of the B. Braun Dialog+ Document (Main Content), UPD: 05 May 2023)

  • 232, 4. TSM Service Program 1/2010 4 — 63 BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 4.8.1.2 Monitor Test A table appears with all colours. Every square contains an alphanumerical character with different character fonts. The brightness and the contrast of the monitor can be checked with this test image. 1. Activate the Monitor Test menu with the Monitor Test key in the TL Manual Test a…

  • 326, 4. TSM Service Program 1/2010 4 — 157 BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 4.8.3.17 Infusion Density Substitution Pump Not applicable.

  • 375, 7. Spare Parts List 1/2010 7 — 1 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx Contact your Local B. Braun Representative for Service Support Spare Parts List The spare parts list (edition 1-2010, edition date March 2010) is valid with immediate effect. Refurbished Spare Parts Conditions for Delivery of Refurbished Spare Parts (depending on availability): Defective spare parts must be sent back wi…

  • 118, 3. Repair Instructions 1/2010 3 — 56 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 3.6 Field Service Utilities FSU General The FSU utility can be used in any mode. Exception: if the Dialog+ was switched off during therapy or end of therapy and was switched on again within 15 minutes. Most of the utilities have the following features: • messages are displayed during execution • an End of <utility name> me…

  • 453, 8. Appendix 1/2010 8 — 9 BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 8.4 Field Service Information FSI

  • 254, 4. TSM Service Program 1/2010 4 — 85 BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-2-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 4.8.2.11 Test 1.11 Dialysis Temperature 1. Activate the Dialysis Temperature menu with the Dialysis Temperature key in the LL Manual Test menu. The following menu is displayed. 2. Check the functions with the appropriate keys. You can ski…

  • 160, 3. Repair Instructions 1/2010 3 — 98 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx Unit HD HDF Online US Default Modified Language Selection English — active active active German — inactive inactive inactive Summer Time Setting (enter values accordingly) Begin of Summer Time D/M/Y dd.mm.yyyyy dd.mm.yyyyy dd.mm.yyyyy h:min xx:xx xx:xx xx:xx End of …

  • 452, 8. Appendix 1/2010 8 — 8 BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 8_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 8.3 Assembly Instructions AI

  • 363, 6. Flow, Wiring and Tubing Diagrams 1/2010 6-8 BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_Chapter 6_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 6.2.2 Wiring Diagram Front Door Single Pump SAKV 27 Cover 16 39 603 601 38 26 604 39: DZ/DR, 10 19 Heparin pump SAD BPA LR Motor LR PA PV PBE S 2 3

  • 296, B. Braun Dialog+ 4. TSM Service Program 1/2010 4 — 127 BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx Pay attention to the negative or positive sign during the enter of the reference value. Prevent offset during calibration Pay attention that there is no fluid in the tubing to the manometer. 3. Disconnect tubing (to filter FVD) from rinsing bridge. 4. Connect manometer and a syringe to the l…

  • 215, B. Braun Dialog+ 4. TSM Service Program 1/2010 4 — 46 BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 4.7.1.12 Dialyser Parameters (Kt/V) 1. Activate the System Configuration menu with the System Configuration key in menu Treatment Support. Use the scroll bar until the following menu appears. 2. Activate the Dialyser Parameters (Kt/V) menu with the Dialyser icon. The followi…

  • 122, 3. Repair Instructions 1/2010 3 — 60 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 3.6.4 Set System Version Number The System Version Number for the Dialog+ can be set with this utility. The utility is typically used after updating the LLC software, because the TLC software can not be accessed during the installation of the LLC. There is no file exchange between the USB stic…

  • 83, 3. Repair Instructions 1/2010 3 — 21 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 3.3.8.6 Pressure Build-up VDABK open offen closed geschlossen PDA VDA UFP FPE FPA VEBK VDEBK VABK VDABK VBP VDE RVDA Fig. : Flow Diagram Pressure Build-up VDABK If PDA < 425 mmHg: The speed for FPE and FPA is slowly set in increments of 200 min -1 in 2…

  • 115, 3. Repair Instructions 1/2010 3 — 53 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 18. Switch off the machine after the successful installation of the TLC software. 19. Remove USB stick. 20. Turn service switch S1 (on digital board C ) to position 2 — TSM. 3.5.4 Boot Machine /Check Version Number 21. Switch on machine. 22. Check wheth…

  • 141, B. Braun Dialog+ 3. Repair Instructions 1/2010 3 — 79 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 3.6.9.6 Read Patient Card BlueCat Removable Disk (X: — FSU USB Stick) DISKTYPE USER 05_Write_Calibration_Data 04_Write_Languages 06_Write_System_Configuration 09_Card_01_Read_Card_Image 09_Card_02_Write_Card_Image 09_Card_03_Read_Patient_Diskette 09_Card_04_Wr…

  • 415, 7. Spare Parts List 1/2010 7 — 41 Pos. Art. No. Qty. Description Comments BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 7.9.3 Central Concentrate Supply (Option) 1 3451467A 1 Connection nipple 2 3451459A 1 Suction coupling 2 34570438 5 O-ring 8.0 x 1.25 3 34565248 10m Silicone tubing 5 x 1.5 / blue (10 m) 4 39239829 m Tubing, white per meter 5 34565230 10m Silicone tubing 5 x 1.5 / red (10 m) 6 3451728A …

  • 53, B. Braun Dialog+ 2. Technical System Description 1/2010 2 — 40 BA-TE-DE08C M.KAY Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG Dialog + SW 9.xx Controller/Supervisor BLD Blood Leak Detector PBE Pressure Sensor PA Arterial Pressure Sensor PV Venous Pressure Sensor PDA Pressure Sensor Dialysate Outlet Power Board Motors PBM BICP Bicarbonate Pump KP Concentrate Pump UFP Ultrafiltration Pump LRP Level Regulation Pump Power Board Valves PBV …

  • 5, Service Manual 1/2010 BA-TE-DE08C M.KAY — Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd B. Braun Avitum AG IV Dialog + SW 9.xx Table of Contents Page 1. Installation and Commissioning 1-1 2. Technical System Description 2-1 3. Repair Instructions 3-1 4. TSM Service Program 4-1 5. Technical Safety Inspection with Preventive Maintenance 5-1 6. Flow, Wiring and Tubing Diagrams 6-1 7. Spare Parts List 7-1 8. Appendix 8-1 8.1 ESD/EMC Information 8-2 8.1.1 Electrostatic Discharge ESD …

  • 344, 4. TSM Service Program 1/2010 4 — 175 BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG Dialog + SW 9.xx 4.9.2 Low Level Options Press the Low Level Options key in the Production Report menu to open the Low Level Options menu. Options assembled in the machine must be activated. Further options present in the Dialog must also be activated in this menu or i…

+

Dialog

SW 9.xx

Commissioning and Service

Prevent Electrical Shock Hazard

ESD Information

!

in TFT Monitor

Protective Conductor in TFT Housing

TSM Service Program

Prevent Chemical Burns and Scalding

Contaminated Machines

Cover in Rear Door

Disposal and Taking Back

System Configuration

Function Check

BA-TE-DE08C M.KAY —

Dialog+ SW9xx_sm_TOC_1-2010.doc/pdf <100329> yymmdd

S e r v i c e M a n u a l

Copyright

This document is the property of B. Braun Avitum AG with all rights reserved.

Only trained personnel must service the Dialog+, i.e. repair, maintenance, software

installation, firmware update, retrofitting and commissioning of the Dialog+.

Servicing must only be performed with proper tools, calibration equipment and be in

accordance with the most recent revision of this service manual/technical information,

which must be clearly and thoroughly understood.

Switch off the Dialog+ and disconnect unit from mains if you have to open the

machine for servicing. Do not touch any exposed wiring or conductive surfaces while

the Dialog+ is opened. The voltages present when electrical power is connected to the

Dialog+ can cause serious injury or death.

Pay attention to ESD information, because electronic components are sensitive to

electrostatic discharges.

High Voltage

If a battery option is present in the machine:

High voltage can be present at the backlight inverter board BIB in the TFT monitor,

even if the machine has been disconnected from mains. Pull out the battery

compartment in the base platform and switch off the battery voltage before opening

the machine.

If the TFT housing had to be opened during a service job, the tight seat of the

protective conductors in the TFT housing must be checked.

Only activate the TSM service program for service activities. It is prohibited to connect

a patient to the Dialog+ and to run a therapy if the TSM service program is activated

in the Dialog+. If the TSM service program is activated the complete alarm system is

disabled. The TSM service program is started in the service mode: digital board, service

switch S1, position 2.

Software

The software is installed in the software mode: digital board, service switch S1,

position 3.

Therapy Mode

After completion of all procedures switch back to the therapy mode: digital board,

service switch S1, position 0.

Calibration

Only perform a calibration after the Dialog+ has reached working temperature and the

machine was disinfected and decalcified. Save the calibration data (CFC) before you

exit the TSM service program: TSM Main Menu , File Operations , Save Calibration Data .

Tubing

Tubing must be replaced only by the same tubing type/length and identical installation

manner.

Make sure that the tubings in the machine are not kinked or twisted after servicing

(e.g. if sub-racks are pulled out and inserted again). The tubing must not touch

moving/rotating components (e.g. motors of gear pumps).

Wiring

Wiring must be replaced only by the same cable type/length and identical installation

manner. The cables must not touch moving/rotating components (e.g. motors of gear

pumps).

During servicing on running machines: prevent chemical burns and scalding of the skin

due to the penetration of disinfectant or hot liquid.

Protective gear should be worn in case of servicing of assumed contaminated

machines.

Servicing of mechanical assembly groups (components in contact with fluid): the cover

for the switch mode power supply microcontroller SMPS-MC in the rear door must be

assembled during servicing because it serves as a spray protection.

O-Rings

Always check o-rings from disassembled groups/components and replace if necessary.

Figures

The displayed figures can differ slightly from the machines on site, due to different

hardware statuses.

Fuses

If fuses are replaced they must exactly match the type and rating specified by the

manufacturer in the spare parts list/technical information. Where applicable: fuses

must be approved by UL/CSA.

Spare Parts

Only use original spare parts manufactured and sold by B. Braun Avitum AG.

Dispose spare parts (e.g. boards or batteries) according to local disposal guidelines or

of Spare Parts

send back to B. Braun Avitum AG free of charge (see chapter 7).

The system configuration saved on a diskette must be downloaded to an other Dialog+

machine only if:

the hardware matches and

the identical software version number is present.

Check the respective function of the assembly group/component after servicing. A

complete function check must be performed after every service, according to the

operating manual.

1 / 2 0 1 0

B. Braun Avitum AG

III